34
NIRScout 816 Users’ Guide NIRx Medizintechnik GmbH Baumbachstr. 17, 13189 Berlin, Germany www.nirx.de NIRx Document NSC-816, Rev 1. 4, Mar 2010

NIRScout816 Users_Manual-082010

Embed Size (px)

Citation preview

Page 1: NIRScout816 Users_Manual-082010

NIRScout 816 Users’ GuideNIRx Medizintechnik GmbH

Baumbachstr. 17, 13189 Berlin, Germany

www.nirx.de

NIRx Document NSC-816, Rev 1. 4, Mar 2010

Page 2: NIRScout816 Users_Manual-082010
Page 3: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

1 General Information

Attention, consult accompanying documents.

Type BF equipment.

Warning, high voltage.

Protective earth (ground).

Do not dispose off in household garbage. Contact Manufacturer for proper disposal information.

Mains disconnect OFF.

Mains disconnect ON.

3

Page 4: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

1.1 Electrical Safety

Properly ground the instrument

This instrument is grounded through the power cord. It is a Class 1 device. Use only the power cord specified for your instrument (see section “Specifications”). Use only a power cord that is in good condition.

WARNING:To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth.

Do not operate without covers and panels installed

To avoid personal injury and equipment damage, the user needs to turn off the mains disconnect before removing covers, panels, or grounding connections. Reinstall covers, panels, or grounding connections before turning the mains disconnect back on.

WARNING! Modification of the equipment is not permitted.

2 Handling and Maintenance

2.1 Transport and Storage

DYNOT compact is a mobile instrument and may be moved by the user between different investigation sites (i.e., labs, exam rooms, etc.) within the institution of the original installation. When storing the instrument, it should be covered to protect it from dust. Furthermore, the equipment must not be exposed to fluids or condensation humidity conditions.

During storage or transport, the equipment should be kept at a temperature between -15°C to +70°C and at a relative humidity of 20% to 80% (non-condensing). Ambient pressure should be within 700 hPa and 1060 hPa.

2.2 Safe Performance

The equipment should only be handled and operated by trained personnel. Operators must familiarize themselves with the latest revision of the user’s manual before using the device.

The DYNOT compact instrument is intended for continuous operation.

2.3 Cleaning, sterilization, and disinfection

The outside of the instrument may be cleaned with a damp cloth. A mild soap or a window cleaner may be used to aid cleaning; do not use solvents. Allow the instrument to thoroughly dry before use.

4

Page 5: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

The fiber optic probe ends that come in contact with the skin can be cleaned and disinfected by wiping them after use with isopropyl alcohol 70%.

DO NOT USE ETHANOL OR METHANOL FOR CLEANING OF THE FIBER-OPTIC SURFACES. THE FIBERS MUST NOT BE STERILIZED IN AN AUTOCLAVE.

2.4 Routine Maintenance

The fiber optic probes are delicate optical devices. Any strain, kinking, pinching, and tight bending (bending radius < 2 cm) should be avoided since this can damage the fibers. Before each measurement, the fibers should be visually inspected for signs of physical damage. In the case of suspected fiber damage, please contact the manufacturer.

No regular service, calibration, or inspection of the instrument by the user, vendor, manufacturer, is required for proper operation of the instrument. In order to maintain adequate signal quality, the user is advised to regularly perform the system check peocedures described in detail in Chapter XXX.

2.5 Waste Products and Disposal

There are no waste products from the equipment

Please contact the manufacturer for help with the disposal of the instrument or its components at the end of its useful life.

2.6 Operating Environment

NIRScout Instruments is for indoor use only.

NIRScout instruments are not safe for the use in presence of inflammable anesthetics.NIRScout instruments are not to be used in an operating room.

The equipment should be operated at a temperature of -10°C to +40°C and at a relative humidity of 20% to 80% (non-condensing). Ambient pressure should be within 700 hPa and 1060 hPa. System Description

2.7 Use of Device

The NIRx NIRScout Near Infrared Spectroscopy (NIRS) imaging system is intended for non-invasive transcranial Near-Infrared Spectroscopy (NIRS) of blood perfusion dynamics in the human head. NIRScout is a research instgrument. The instrument and all accessories are to be used for research purposes only.

NIRx products are NOT “medical devices” intended to be used for the purpose of diagnosis of disease or other conditions, or in the cure, mitigation, treatment, or prevention of disease, or used as a life support device. Use of NIRx products are solely for research purposes.

5

Page 6: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

3 Functional Description

NIRScout performs dual-wavelength continuous-wave (CW) near infrared (NIR) diffuse tomographic measurements on large tissue structures or other scattering media at high (= several Hz) sampling rates. The system facilitates NIR illumination of multiple target locations in a time-multiplexed, scanning fashion. Using LED emitters at two distinct wavelengths allows discrimination of the two oxygenation states of tissue hemoglobin. Both wavelengths are emitted simultaneously and are distinguished by modulating and demodulating each at distinct frequencies in the low kHz-range.

The instrument employs parallel readout of multiple optical detector channels, each of which uses adaptive gain switching to maximize the dynamic measurement range (>109). The device features 4 parallel, optically isolated digital input channels (TTL/CMOS logic) for the acquisition of event trigger signals. The NIRScout instrument is operated through a graphical user interface (GUI) on a personal computer (PC), to which it is connected by USB 2.0.

Fig. 1 shows a block diagram depicting the NIRScout hardware components and main signal connections. All components except for the power supply plug into the instrument front and interconnect through a backplane. The USB controller establishes data and control signal connection with a host PC and contains

4

……

fmod1

fmod2

gains

timing

USB controller signals

Backplane

Detector channel

LED driver

MUX ADC

DO

DAC1,2

CTR DI

CTRL

Trigger input

USB 2.0

Power supply

LP

4x…

4x

8x

PD

LED

DI

Opto Iso

FOC

Fig. 1: Block diagram of NIRScout functional components and interconnections. FOC: Fiber optic cable; PD: Photo diode; LP: Low-pass filter; LED: Light emitting diode; DI: Digital input; Opto Iso: Opto-isolator; USB: Universal serial bus; MUX: Multiplexer; DAC: Digital-to-analog converter; DO: Digital out; CTR: Counter; CTRL: Controller; fmod1,2: Modulation frequencies 1, 2.

Page 7: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

all data acquisition and control hardware necessary for instrument operation. This includes digital I/O lines, counters, an analog-to-digital converter, and digital-to-analog converters.

The LED driver card provides multiple parallel precision current sources which allow the desired intensity control of the LED emitters. Each optode incorporates two LED wavelengths, which receive distinct current modulation frequencies in the low kHz range. A tissue scan is facilitated by sequentially turning on and off LED optodes under digital control of the LED driver, at a rate of about 60 Hz.

Detection fiber optics are used to guide the transmitted and back reflected light from the target to the detection card, each of which houses 4 detector channels. The optical signal transmitted by the fiber optic cable is converted to a voltage proportional to light intensity by means of a photo detector (photo diode) followed by electrical amplification. The gain of the amplifier is automatically adjusted by the instrument to provide optimal signal-to-noise ratio depending on the intensity of the received light. The instrument is therefore capable of dealing with optical intensities over an extremely large dynamic range (>109), which allows diffuse transmission measurements on tissue over distances from under 1 cm to about 4 cm. This makes the instrument suitable for optical tomographic ('multi-distance') imaging. The amplifier output is low-pass filtered and converted to a digital signal with 16-bit resolution by the controller board.

To allow time stamping of the measured data for the purpose of event synchronization with concurrent measurements or experimental protocols, the instrument provides a trigger input card, which offers four parallel digital input channels. Each of these is optically isolated from the rest of the system. Each input is over-voltage protected and contains a comparator circuit ('Schmitt Trigger') to improve noise immunity. The inputs are TTL/CMOS compatible and are negative-edge triggered.

4 Hardware Components

4.1 Main Frame

The instrument's form factor is based on an industry standard 19-inch rack enclosure accommodating 42 horizontal partitions (= 8.4” / 213.4 mm usable front panel width) of 525” (133.35 mm = 3 height units). All the main functional components are implemented as modular electronic cards, which can be inserted as required by the system's configuration. The back of the enclosure contains the AC mains power entry as well as the main On/Off switch. Also situated in the rear panel is a cooling fan exhaust.

Fig. 2 shows the front and rear view of the instrument.

7

Page 8: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

4.2 USB Controller Card

The USB controller performs all functions of instrument control and signal acquisition (Fig. 3). Communication to the host PC and instrument operation software is established through USB 2.0 connection. Once connected, the system goes through a startup sequence during which the green status indicator (2) lights up and then starts to blink. At this point, the data link to the host PC is established.

The red indicator (3) signals the power on state of the instrument. Only when the status indicator is blinking and the power LED is lit is the instrument ready for operation.

The order in which the system is powered and the USB is connected does not matter.

4

Fig. 3: USB Controller Card USB-1618. (1) USB 2.0 connector to host PC; (2) Data connection status indicator (green); (3) Power indicator (red).

Fig. 4: LED driver card LED-82. (1) Connector for 8-channel LED optode set.

Fig. 5: Detection card QCD-10. The unit contains four parallel optical sensing channels. Each of the four photo sensor located in the fiber port (1) serves one fiber optic cable, which is secured by individual set screws.

Fig. 6: Trigger card TRG-42. BNC terminals (1) afford connections to four negative-edge triggered parallel digital input channels.

1

2 3

1

1 1

41 2 3

Fig. 2: (a) Front view of the NIRScout 816 imaging system. (1) Quad-channel detection cards QDC-10; (2) Trigger input card TRG-42; (3) USB controller card USB-1618; (4) Led driver card LED-82. (b) Rear View showing AC power inlet and power main On/Off switch

Page 9: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

4.3 LED Driver Card

The 8-channel LED optode set is connected to the front panel connector of the the LED driver (Fig. 4). Connection should be established before powering up and operating the instrument. Connecting and disconnecting the optode set during operation should be avoided.

4.4 Detector Card

The device is shown in Fig. 5. The fiber port connector accommodates four individual fiber optic cables, which are secured by individual set screws. The basic version of the instrument can operate one to four QCD modules, or 4 – 16 optical detection channels. A version of the instrument for up to 24 detectors is available.

4.5 Digital I/O Trigger Card

Fig. 6 shows the trigger card front panel. Four parallel digital TTL/CMOS level compatible inputs are available, which are negative-edge triggered. Connections are made with standard 50 Ohm BNC connectors. All inputs are optically isolated from the instrument electronics to enhance safety and signal integrity.

4.6 Optical Probes

The system relies on optical probes ('optodes'), emitters and receivers, to be located on the target tissue, usually the intact human head (Fig. 7). Optical sources are realized by dual-wavelength LED emitters placed directly on the skin. The sensors are formed by fiber optic cables that are placed on the skin, and which guide the received light to the optical sensor situated on the detection card inside the instrument.

4.6.1 Connecting the Fiber Optics to the Imager

Insert the metal ferrule of the fiber optic cable into the appropriate fiber port input. Carefully push in the fiber until mechanical resistance is met. Do not exert strong force. It may help to perform slight turning

9

Fig. 7: Sensors used with the NIRScout system. (a) Detection fiber optical cabels, bundle of four. The black plastic tips are placed on tissue, the metal end ferules are inserted into the detection card. (b) Bundle of eight dual-wavelength LED emitters. Black plastic ends with LED lens are placed on the skin, and the ribbon cable connects to the LED driver card. (3) Example of optode placement on a dummy head.

a b cb

Page 10: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

motions to ease the fiber connector into the port. Once the fiber is seated turn the set screw next to it clockwise to lock in the ferrule. A small amount of torque is sufficient to prevent the fiber from slipping back out. Do not use strong force as this can damage the fiber ferrule. To release fiber, turn set screw counterclockwise (a fraction of a turn will suffice) and carefully extract the ferrule.

Hints: All fiber ferrules should protrude into the port by the same amount. If some fibers are observed to protrude out more than others, these may not be connected properly. Make sure the set screw is not blocking the path for the ferule before connecting the fiber optic cable.

4.6.2 Connecting the LED Emitters to the Imager

The LED POD emitter flat ribbon cable is connected to the instrument through a flat ribbon cable. The connector in the LED driver card is polarized to prevent reversed connection. Carefully insert the cable until it is fully seated.

The cable should only be connected or disconnected while the instrument is powered down. Otherwise damage may occur to the LED optodes.

4.7 Optodes and Head Gear

A variety of fiber optic and LED-based optodes as well as probe placement systems are available for NIRScout. An example of a fabric-cap based solution is shown in Fig. 7c. For detailed instructions and handling advice, please refer to the specific operators' manuals.

4

Page 11: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

4.8 System Setup

Fig. 8 shows an example of a complete instrument setup. A host PC (shown is a Laptop) connects to the USB controller. For purposes of safety and data integrity, this connection may be electro-optically isolated, as shown in Fig. 4. Optodes, i.e. emitters (LED) and detectors (fiber optics), are connected to the target under investigations (shown: dummy head phantom). AC mains connections are made to the PC, the imager, and the USB receiver.

11

1

23

4

5 6

78

9

10

AC Mains

PS PS

Fig. 8.: Photograph of complete System Setup. Setup components: (1) NIRScout device; (2) target (test phantom); (3) controll PC; (4) USB optical isolator transceiver (Tx); (5) USB optical isolator receiver (Rx). Setup connections: [6] USB cabel from PC to USB Tx; [7] fiber optic cable from USB Tx to Rx; [8] USB cable from USB Rx to NIRScout device; [9] Flat ribbon to LED illumination probes; [10] fiber optical bundles for detection; PS: power supply.

Page 12: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

5 User Interface

The NIRScout imaging system includes the NIRStar data acquisition and control software, a LabVIEW™ (National Instruments Corp., Austin, TX) based graphical user interface. Instrument functions are represented by graphical elements like buttons, sliders, and displays through which the user controls the instrument. Fig. 8 shows the NIRStar home screen.

5.1 Start and Stop Software

When calling the software, it is automatically started in running mode, as indicated by a black arrow in the top left screen corner. Always exit software using the EXIT button on the lower right corner. The arrow will turn white.

Clicking the white arrow will re-start the software.

4

2

1

4

5

63

7

8

Fig. 8: The NIRStar user interface home screen. (1) Software start button; (2) System setup screen; (3) Setup screen display selectors; (4) Status display; (5) Experimental notes editor; (6) Source-detector configuration; (7) Main control elements and elapsed time/frame counter; (8) Tab control access to Pre-Scan check and Scan screen.

Page 13: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

5.2 Screen Selector

Using the tabs control, the user can select one of three screen views: Setup, Pre-Scan, and Scan.

5.3 Control Panel

The Control Panel remains visible throughout operation and provides access to the basic imager functions. The following controls are available:

CALIBRATE: Prior to each measurement, a calibration procedure must be performed during which the instrument determines the optimum gain (or signal amplification) setup for each source-detector combination. Gain levels are numbered 0 … 7 with each step representing a 10-fold amplification factor.

During calibration, the Gain Map will be updated to reflect the current gain settings. The gain setup is automatically followed by a scan to determine the achieved signal quality.

The gain is deemed optimal if a channel's signal falls between 0.1 V and 1.0 V. Because the same gain is applied to both wavelengths' signals, sometimes a compromise must be achieved so as to provide sufficient amplification for one wavelength, while not saturating the other. Therefore, the gain setting is satisfactory if at least one signal lies in the optimum signal range, while the other may be above or below the stated voltage thresholds.

The START button initiates a scan without writing data to the computer disc for test measurements and display optimization. The test scan may be stopped by clicking the START button a second time or with the STOP button.

The RECORD button initiates a scan for which the data, gain settings, trigger events and various system settings get recorded. If first a test scan is initiated with the START button, and at some point afterwards during the scan the RECORD button is clicked, all data (i.e., up to that time point and from then on forward) will be saved. Once recording, the scan can only be stopped using the STOP button.

The STOP button will finish any ongoing scan.

The Frame No. counter displays the number of completed scans, or frames, which is defined as the number of full illumination cycles. Th nth illumination cycle is completed once all illumination points (LEDs) have blinked n times during the measurement.

13

Page 14: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

The source switching frequency fss is a fixed instrument parameter, typically 62.5 Hz. The sampling (or frame, or scan) rate can be calculated as: fs = fss / (N+2), where N is the number of sources.

The timer shows the time elapsed since starting a scan and can be reset to zero. The timer reading is not recorded, and resetting has no effect on the data.

The EXIT button exits the application.

5.4 Setup Screen

The setup screen is used to judge the system setup and signal quality obtained from the calibration procedure prior to running a measurement.

5.4.1 Setup Display

The setup display is a multi-purpose indicator that provides several measures of signal quality and appropriateness of the setup. The DISPLAY knob and WAVELENGTH selector are used to select the desired quantity.

4

Fig. 9: The Setup display in the setup tab maps various signal quality measures for each source-detector combination in a 2D false color map. (a) Gains settings from 0-7;2(b) Signal level in Voltage (log scale), (c) Signal-to-noise ratio as coefficient of variation (C.V.). Source-detector combinations masked out in the Hardware Configuration menu are blocked in the display.

Page 15: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

The GAIN MAP shows the gains obtained from the system calibration procedure in a color-coded map. Gains are mapped in a logarithmic fashion, each step representing a 10-fold increase in amplification.

The gain has an inverse relationship with the intensity of the the transmitted light. For typical source-detector separations (2.5 – 3 cm), typically, high gains of 4 and higher are required. The lower gain stages are to be expected only for short source-detector separation values, as experienced in multi-distance, or tomographic setups (there are, however exceptions, such as measurements on children, or on the forehead).

The gain setting for a chosen optode setup generates a characteristic pattern, which should be quantitatively be reproducible across different subjects.

The SIGNAL LEVELS are mapped on a logarithmic scale in units of voltage. Satisfactory signal levels fall in the range of 0.1 – 1.0 V (light green to yellow).

Channels that are masked out from display in the Hardware Configuration dialog (see Chapter 5.7.1), are suppressed in the setup display.

5.4.2 Status History and Notes Editor

The System Status History window will prompt an account of all user actions with time stamp to help keep track of the experimental conditions. A timestamped history is recorded in a text file automatically (see file formats, below).

The Experimental Notes editor allows the user to enter plain text comments, which are recorded in the data .hdr file (see file formats, below).

15

Page 16: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

5.5 Pre-Scan Screen

The pre-scan screen is intended for setting up the real time display preferences and checking signal levels and trigger inputs.

5.5.1 Level Display

The real-time level display shows bar graphs for each channel indicating the received signal level. The lower and upper boundary (0.1 and 1.0 v, respectively) of the optimal signal strength are indicated by red horizontal lines. The wavelength or hemoglobin (Hb) state displayed is toggled tieht he Wavelength button (7).

5.5.2 Real-Time Curve Display

The measured signals for all channels selected in the hardware setup masking screen (see 5.7.1) are displayed in a scrolling curve plot. The wavelength or hemoglobin (Hb) state displayed is toggled with the Wavelength button (7).

4

Fig. 10: The NIRStar user interface Pre-Scan Screen. (1) Real-time level display for all channels (log-Volts), levels between 0.1 V and 1.0 V are optimal. (2) Real-time curve display of measurement channels for one wavelength or Hb state. (3) Selector for display type. (4) Low-pass filter cutoff selector. (5) Curve stacking controls. (6) Trigger input indicator. (7) Wavelength selector for display. (8) Clear display button.

2

1

4

5

63

7

8

Page 17: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

5.5.3 Display Type

Use this drop down menu to select one of the following option:

Wavelength raw [mV]

Displays raw detector readings for each wavelength in units of mV. No low-pass filtering or normalization is applied.

Wavelength filtered [mV]

Displays the low-pass filtered detector readings for each wavelength in units of mV. A moving-average low-pass filter is applied, whose cutoff frequency can be set by the knob labeled 'LP Filter Freq.' This feature may be used to remove high-frequency noise and heart beat artifacts from the signal.

Hb filtered [uM]

Displays the estimated changes in oxy- or deoxy-hemoglobin concentration. A moving-average low-pass filter is applied, whose cutoff frequency can be set by the knob labeled 'LP Filter Freq.' This feature may be used to remove high-frequenvy noise and heart beat artifacts from the signal.

5.5.4 Low-Pass Filter

The knob is used to set the cutoff frequency for th erunning-average low-pass filter, which is applied to the displayed data if display type is set to either 'Wavelengths filtered' or 'Hb Filtered.'

NOTE: The filter setting does not affect the saved data.

5.5.5 Curve Stacking

Curve stacking is a feature, by which the user may separate the displayed curves by equal amounts to improve visualization of features. This is of advantage in particular for displaying the nomalized Hb data, for which the superposition of the unstacked curves makes discriminating signal features difficult. By checking 'Stack Curves' the feature is enabled.

The 'Stacking Factor' determines the amount by which the curves are separated.

NOTE: For the wavelength display, selecting 'Stack Curves' results in plots that are normalized to the baseline.

5.5.6 Wavelength / Clear Display

The button labeled 'WAVELENGTH' selects the display wavelength or Hb state for the level and the curves display in the pre-scan screen.

17

Page 18: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

5.5.7 Trigger Indicators

These correspond, from left to right, to the for trigger input channels of the instrument. When a falling edge is received on a trigger input, the corresponding indicator lights up.

5.6 Scan Screen

The scan screen is intended as main display during an experiment and dedicates much of the screen space to displaying data at both wavelengths/Hb states and trigger events. The data are plotted for all channels that are selected in the masking feature of the Hardware Configuration menu (see 5.7.1). All plot features selected in the pre-scan screen (filtering, wavelength/Hb, curve stacking, etc.) will be applied to the scan screen.

Trigger events are indicated by vertical markers that are plotted between and synchronously with the optical readings.

The x-axis range of the display can be changed ion the Display Time Range control.

4

Fig. 11: The NIRStar user interface Scan Screen. (1), (2) Real-time curve plots for Wavelength 1 and 2 / oxy- and deoxy-hemoglobin. Amplitudes are in Volts or μmol/L. (3) Trigger events display. (4) Display time range (x axis) control.

1

3

2

Page 19: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

5.7 Menu

5.7.1 Configure Hardware

Measurement Setup

Allows the basic setup of the measurement for a given instrument configuration. The number of (1) detectors and (2) sources to be used in the measurement can be changed. With the 'All' buttons, the respective maximum number of channels is set.

The maximum values are limited to the capacity of the imager, e.g., if one 8-channel LED driver and four detector channels are present, the maximum values allowed are 8 Sources and 16 detectors.

The screen also allows the setup of specific illumination patterns (see below).

Channel Masking

Allows selection of specific source-detector-combinations (i.e., measurement channels) for display in the Setup Display and the curve plots. Rows represent source channels, and columns represent detector channels. Selected channels are marked by a filled square, and will appear in plots. Unchecked (white) boxes mark those channels which are excluded from plots.

The masking pattern is automatically saved and retrieved when entering hardware configuration.

The boxes marked with red, right pointing arrows are used to toggle complete rows (i.e. all detectors for a specific source). The boxes marked with green, downward pointing arrows are used to select/un-select complete columns (i.e. all sources for a specific detector).

The masking options allow to mask or clear all channels, and to load configuration files containing pre-

19

Fig. 12: Measurement Setup panel.

Fig. 13: Channel Masking panel. Top: All channels selected. Bottom: Detector 2, all sources selected.

2

Page 20: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

defined masking pattern. The 'Auto' function masks out all channels that have a signal level below the specified value in Volts (default 0 0.1 V).

Hardware Configuration

The hardware Configuration panel defines the basic hardware capability of the instrument. Update of these parameters is usually needed only for the initial installation of the instrument, and in case of hardware changes (e.g., upgrades).

'Available detector cards' should be set to the number of detector channel cards physically present in the instrument. This number multiplied by the number of channels per card (i.e., 4) equals the maximum number of detection channels available.

'Available Sources' may be set to '8' or '16' depending on the type of LED driver installed.

WARNING! It is important that the 'Available Number of Sources' value reflect the configuration of the LED driver, even if fewer sources than available are to be used. Wrong setup will lead to data loss!

'Trigger inputs available' should be checked if your system is configured with trigger input channels. If those are not present, this option MUST be unchecked!

Advanced Setup

Advanced Setup allows changing basic instrument parameters affecting timing, LED modulation, etc.

These should only be changed by developers and professional staff and are password-protected to prevent users from inadvertently changing parameters.

WARNING! Inappropriate settings in the Advanced Setup may severely interfere with data quality or may cause damage to the instrument!

4

Fig. 14: Hardware Configuration Panel.

Page 21: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

Illumination Patterns

Measurement setup allows the specification of custom scanning patterns, including simultaneous illumination at multiple sites for increasing scan rates.

Fig. 14a shows the default illumination pattern for a 16 source measurement. Each LED will be illuminated sequentially, resulting in 16 illumination cycles, at a scan rate of about 3.5 Hz. The number of acquired channels in this example is 16 x16 = 256.

This setting can always be achieved by clicking the 'Standard' button.

Fig. 14b shows a scan sequence in which always two LEDs are switched on at the same time, i.e., 1+9, 2+10, 3+11, and so on. This is permissible if the LEDs that are on during the same scan cycle are sufficiently separated to avoid optical cross talk, i.e. no detector is within measuring distance (typically, closer than 4-5 cm) of both LEDs. This situation can be achieved through careful experimental planning, e.g., in bilateral head measurements, where half the sources illuminate each hemisphere.

The shown pattern can be achieved by clicking 'Bilateral.' The sampling rate for 8 illumination cycles is 6.25Hz. The number of acquired channels in this example is 2 x 8 x 8 = 128.

Custom scan patterns can be achieved by choosing the number of desired cycles and editing the LED sequence in the matrix. The scan frequency is given as:

fscan = 62.5 Hz / (no. of cycles + 2)

and the number of measured channels is given by:

Nch = No. of cycles x No. of detectors

The example in Fig 15b shows a case of 4 scan cycles, for which 4 LEDs are turned on simultaneously (fscan = 10.4 Hz, 64 measurement channels).

21

Fig. 15: Examples of different illumination scenarios. (a) Standard pattern for 16 sources. (b) Bilateral pattern for bilateral 8-source illumination. (c) Custom setup example.

Page 22: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

5.7.2 Diagnostics

The diagnostics window is used to test the dark noise performance of each detector, at each gain setting for both wavelengths. Before running the dark noise test by clicking (1) make sure to have the detector inputs well shielded from ambient light. The bar graph (2) displays the noise or offset level for each detector. The y-axis is in units of mV. select the gain with slider (3) to observe the noise values, which are the root-mean-square (RMS) values of a sequence of detector readings. The noise should stay well below 0.1 mV for Gains 0-5. For Gain 6, the noise level will be around 0.5 mV, and for Gain 7 about 5 mV noise is to be expected.

Button (4) toggles from noise to DC offset display, and button (5) toggles between the wavelengths.

5.7.3 Options

The option window allows selection of a root directory, under which all data files will be saved.

Also, a prefix for data file names can be specified.

5.7.4 About

Information about the software and copyright notices revision can be obtained through menu item 'About'.

4

2

3

4

1

5

Fig. 16: System Diagnostics screen (dark noise measurement).

Fig. 17: Options screen.

Fig. 18: About dialog.

Page 23: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

6 Data File Formats

23

Page 24: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

7 Quick Start Guide

7.1 Turning on the System

1. Make sure all system components are connected to mains power and that the host PC is turned on. The imager may already be connected to the PC through USB.

2. Turn on the NIRScout imager using the power switch on the rear panel next to the power inlet. The Power LED in on the USB controller must light up.

3. Connect the imager to the PC USB port and make sure that a valid USB link between both devices is established as indicated by continuous blinking of the green Status LED on the USB controller front panel.

4. Launch NIRStar by either double-clicking on a link on the Windows™ Desktop or by calling the software from the Windows™ Start menu.

5. Start NIRStar by clicking the arrow in the software's toolbar. Successful system setup is indicated by the 'Ready' prompt in the status window.

7.2 System Configuration

Make sure the system's configuration (especially, numbers of sources, detectors, and trigger channels used) match the requirement of the planned measurement. If necessary, adjust the hardware configuration as explained in Chapter XXX.

7.3 Probe Setup

The optodes to be used for the measurement should be placed on the target, as described in the optodes' user manual.

7.4 Initializing a Measurement

Use the 'CALIBRATE' function to determine the gain setup for the measurement. This process will take several seconds. The progress of the calibration procedure will be apparent by the changing gain display, indicating the system's testing of different gain values. After finishing the setup, the gain map will provide several measures of the achieved signal quality as discussed in detail in Chapter XXX.

7.5 Pre Scan

If desired, the real-time display may be optimized in the Pre Scan screen, as described in Chapter XXX. This does not affect the measured data and is entirely optional.

A test scan for visual inspection without saving of the acquired data may be initiated through the 'START' button. Data can be viewed in either the Pre Scan or Scan screen. A second push of the button will stop the acquisition.

4

Page 25: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

7.6 Perform Scan (Measurement)

The measurement is initiated by pressing the 'RECORD' button. Adjust viewing options in Pre Scan screen and view data in Pre Scan or Scan sreen as desired.

Use 'STOP' button to finish data recording.

7.7 Shut Down System

1. Exit software through 'EXIT' button and close window if desired.

2. Turn off imager using main switch on rear panel.

3. Disconnect USB link if desired

4. Shut down PC if desired.

25

Page 26: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

8 Error Messages and Trouble Shooting

Table 1: Error messages and trouble shooting

4

Error / Warning / Problem Potential Cause Recomm ended ActionPow er LED does not light up No pow er to instrument Check if pow er sw itch is turned on

Check if AC line is properly connected.Check fuses and replace if necessary

No data connection to PC Reconnect USB cableTry dif ferent cable or dif ferent portMake sure softw are and drivers are installed

No data connection to PC

Low or no signal in all channels Check connection of LED PODCheck if LEDs are blinking during scanningCheck connection of the f iber optic bundles

Low signal f rom the source

Low signal f rom the detector

Data overly noisy Ambient light interference

Status LED does is of f or on continuously

Softw are error message: 'Device may not be connected'

Make sure USB connection is established (Status LED blinking)

Large number or all channels go to highest gain setting

Check proper placement of probes on subject's headCheck for damage on LED illuminator or f iber optic cable

All detectors show maximum gain for the same source

Check optical coupling of the af fected source to the skin, if needed part hair, apply coupling gel, apply gentle pressure through outer cap or bandage

No or low reading for one source, at one w avelength

Damage to LED driver or LED optode

Restart instrument, and w ithout performing performing calibration start scan and point af fected LED directly into a detection f iber tip or detector channel port. If OK there should be a stable reading for both w avelength plots in the scan screen (display set to raw data )

All sources show maximum gain for the same detector channel

Check optical coupling of the af fected detector to the skin, if needed part hair, apply coupling gel, apply gentle pressure through outer cap or bandageCheck connection of af fected f iber optic cable to the instrumentMake sure head is w ell shielded and/or not directly illuminated by ambient light, in particular f luorescent tubes.

Page 27: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

9 Accessories, Replacement Parts, and Disposables

All Accessories and replacement parts are to be obtained only through NIRx directly, or from authorized sales partners. Disposables may only be used if obtained from NIRx or authorized dealers, or with the expressed approval by NIRx.

9.1 Fiber optic probes and Led illuminators

A wide range of fiber optic probes and LED illuminators is available for the NIRScout Series to suit the divers range of intended user applications. To provide up-to-date information about available products, please inquire with NIRx or one of its authorized vendors.

9.2 Probe holders / Head Gear

A wide range of head caps and probe holders is available for the NIRScout Series to suit the divers range of intended user applications. To provide up-to-date information about available products, please inquire with NIRx or one of its authorized vendors.

9.3 Cables

For the data connection, a generic cable of suitable length conforming to USB 2.0 gudelines may be used.

For power mains connection, a generic power cord may be used that adheres to the following: Three-wire cabel with neutral, live phase, and protective earth wires of sufficient cross-section (1mm2 minimum for each wire), IEC C13 type connector for instrument connection, and applicable approvals or certification (e.g., EU: VDE, TÜV, GS,... US: CSA, UL, … etc.).

Connection to trigger input may be established with generic coaxial cables (BNC RU58, or similar).

9.4 Disposables

Third-party disposables such as spatulas (for parting hair), contact gel, etc. may be used as long as the applicable directives for medical safety (biocompatibility, etc.) are complied with. To assure uncompromised data quality, we strongly advise use of disposables recommended by NIRx.

27

Page 28: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

9.5 List of Approved Accessories

Table 2: List of approved accessories

4

# Part No. Nam e Description1 FOB-XX Fiber optic bundle

2 NLP-8-XX LED POD

3 L2F-8-XX LED-to-f iber coupler

P2B-8-XX Parallel-to-BNC adapter

P2IDC-10-XX Parallel-to10-pin

4 NMA-XX Mechanical arm Articulated arm for support of f iberoptics and LED cables5 NBP-XX Base plate

Fiber optic cable for detection or illumination,various lengths, diameters, and tip designs LED illuminator (8 channels, 2 w avelengths)various lengths and tip designsCouples LED light into illumination f iber bundles (8 channels, 2 w avelengths)Adapter box to connect parallel (25-pin) to BNC trigger input connectorsAdapter cable to connect parallel-cable (25-pin) to 10-pin IDC trigger input

Mounting plate to stabilize NIRScout and to provice mounting point for mechanical arm

Page 29: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

10 NIRScout Technical Specifications

Optical Detection

No. of Channels.......................................................4-24 (Parallel Digital Demodulation)Sensitivity................................................................< 1pW NEPDynamic Range........................................................>90dBopt

Sensor Type.............................................................Si Photodiode

Optical Illumination

No. of Illumination Sources.....................................8/16 (Time-Multiplexed)Wavelengths............................................................760 nm, 850 nm (Frequency-Encoded)Power......................................................................5 mW / WavelengthMultiplexing Rate....................................................62.5 Hz / (No. of Sources +2)Emitter Type............................................................LED

Communication

Host.........................................................................USB 2.0Experiment..............................................................TTL/CMOS (1-4 Trigger In, falling edge)

Electrical

Supply Voltage.........................................................90-250 VAC (50-60Hz)Power consumption.................................................175 W max.Phases.....................................................................1

Physical

Size (W × L × H)........................................................257 mm (42HP) × 326 mm × 165 mm (3U)Weight.....................................................................5 kg

Environmental

Temperature range..................................................10°C to 40°C (operating)-15°C to +70°C (storage and transport)

Humidity....................................................................20 to 80% relative humidity non-condensing Ambient pressure.......................................................700 hPa to 1060 hPa

Classification

Degree of electric shock protection.........................Class IType of electric shock protection.............................Type BFDegree of liquid ingress protection.........................IPX 0 (not protected)

29

Page 30: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

Mode of operation..................................................Continuous

Optical Safety

The device uses low-power NIR LED illumination and falls under Laser Class I (eye safe).

Methods of sterilization or disinfection

Applied parts (fiber optic probe tips) can be disinfected with diluted isopropyl ('rubbing') alcohol 70%. Fiber optics are NOT suitable for autoclaving. Fiber optics /LED probes and fabric caps are suitable for cold-gas sterilization. Please see also section “Cleaning, sterilization, and disinfection” on p.5.

Degree of safety in the presence of flammable anesthetics

The equipment is not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.

Declaration of Electromagnetic Emissions/Immunity

Medical electrical equipment needs special precautions regarding electromagnetic compatibility (EMC). The NIRScout equipment needs to be installed and put into service according to the EMC information provided below.

Mobile and portable RF communications devices can affect the equipment.

Accessories and cables compliant with the EMC test results

– Unshielded power cord, 250VAC, 6A, 5 m length max.

– NIRx LED illumination probe Type NLP-8

– USB 2.0 Optical Line Isolator Type 2244

WARNING! Accessories and cables other than those specified, with the exception of parts sold by NIRx as replacement parts for internal components, may result in increased emissions or decreased immunity of the NIRScout instrument.

WARNING! The NIRScout should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the instrument should be observed to verify normal operation in the configuration in which it will be used.

4

Page 31: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

31

Guidance and manufacturer’s declaration – electromagnetic emissions

The NIRx DYNOT-232 is intended for use in the electromagnetic environment specified below. The customer or the user of the NIRx DYNOT-232 should assure that is used in such an environment.

Emissions test Compliance Electromagnetic environment - guidance

RF emissions

CISPR 11

Group 1 The NIRx DYNOT-232 uses RF energy only for its internal function. Therefore, RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

RF emissions

CISPR 11Class A

The NIRx DYNOT-232 is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes

Harmonic emissions

IEC 61000-3-2Class A

Voltage fluctuations/ flicker emissions

IEC 61000-3-3

Complies

Page 32: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

4

Guidance and manufacturer’s declaration – electromagnetic immunity

The NIRx DYNOT-232 is intended for use in the electromagnetic environment specified below. The customer or the user of the NIRx DYNOT-232 should assure that is used in such an environment.

Immunity test IEC 60601 test level Compliance level Electromagnetic environment -

guidance

Electrostatic discharge (ESD)

IEC 61000-4-2

±6 kV contact

±8 kV airComplies

Floor should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.

Electrical fast transient/burst

IEC 61000-4-4

±2 kV for power supply lines

±1 kV for input/output lines

CompliesMains power quality should be that of a typical commercial or hospital environment.

Surge

IEC 61000-4-5

±1 kV differential mode

±2 kV common mode

CompliesMains power quality should be that of a typical commercial or hospital environment.

Voltage dips, short interruptions, and voltage variations on power supply input lines

IEC 61000-4-11

<5% UT (>95% dip in UT) for 0.5 cycle

40% UT (60% dip in UT) for 5 cycles

70% UT (30% dip in UT) for 25 cycles

<5% UT (>95% dip in UT) for 5 sec

Complies

Mains power quality should be that of a typical commercial or hospital environment. If the user of the NIRx DYNOT-232 requires continued operation during power mains interruption, it is recommended that the equipment be powered from an uninterruptible power supply or battery.

Power frequency (50/60 Hz) magnetic field

IEC 61000-4-8

3 A/m Complies

Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

NOTE: UT is the a.c. mains voltage prior to application of the test level.

Page 33: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

33

Guidance and manufacturer’s declaration – electromagnetic immunity

The NIRx DYNOT-232 is intended for use in the electromagnetic environment specified below. The customer or the user of the NIRx DYNOT-232 should assure that is used in such an environment.

Immunity test IEC 60601 test level

Compliance level Electromagnetic environment - guidance

Portable and mobile RF commutations equipment should be used no closer to any part of the NIRx DYNOT-232, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

Recommended separation distance

Conducted RFIEC 61000-4-5

3 Vrms150 kHz to 80 MHz

3 Vrms 1.17d P=

Radiated RFIEC 61000-4-3

3 V/m80 MHz to 2.5 GHz 3 V/m

1.17d P= 80 MHz to 800 MHz

2.33d P= 800 MHz to 2.5 GHz

where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).

Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,a

should be less than the compliance level in each frequency range.b

Interference may occur in the vicinity of equipment marked with the following symbol:

NOTE 1: At 8 MHz and 800 MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

a Field strengths from fixed transmitters, such as base stations for radio (cellular, cordless) telephones and land mobile radios, amateur radio, AM and FM broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the equipment is used exceeds the applicable RF compliance level above, the equipment should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting the equipment.

b Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V/m

Page 34: NIRScout816 Users_Manual-082010

NIRScout 816 Users' Guide

4

Recommended separation distances between portable and mobile RF communications equipment and the NIRx DYNOT-232 instrument

The NIRx DYNOT-232 is intended for use in an electromagnetic environment in which radiated RF disturbances are controller. The customer or the user of the NIRx DYNOT-232 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the NIRx DYNOT-232 as recommended below, according to the maximum output power of the communications equipment.

Rated maximum output power of transmitter

W

Separation distance according to frequency of transmitterm

150 kHz to 80 MHz

1.17d P=

80 MHz to 800 MHz

1.17d P=

800 MHz to 2.5 GHz

2.33d P=

0.01 0.12 0.12 0.230.1 0.37 0.37 0.741 1.17 1.17 2.33

10 3.70 3.70 7.37100 11.70 11.70 23.30

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated by using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.