Introduction Rotary Table Calibration Laser Tracker Investigation ADM Tests Conclusion

Introduction Rotary Table Calibration Laser Tracker Investigation ADM Tests Conclusion

Laser Tracker Test Facility at SLACLaser Tracker Test Facility at SLAC

Page 2: February, 2008 IWAA08 G. Gassner, R. Ruland gassner@slac.stanford.edu 1 Georg Gassner Robert Ruland, SLAC February, 2008 Introduction Rotary Table Calibration

2February, 2008IWAA08

G. Gassner, R. [email protected]@slac.stanford.edu

Introduction

Particle Physics experiments at SLAC require high accuracy positioning, e. g. 100 µm over a distance of 140 m or 25 µm in a 7 x 10 x 3 meter room

Realization -> Laser Tracker Measurements

Rotary calibration table (Kugler GmbH) providing an accuracy of better than 0.2 arcsec

Interferometer bench

Page 3: February, 2008 IWAA08 G. Gassner, R. Ruland gassner@slac.stanford.edu 1 Georg Gassner Robert Ruland, SLAC February, 2008 Introduction Rotary Table Calibration

3February, 2008IWAA08

G. Gassner, R. [email protected]@slac.stanford.edu

Rotary Table

Renishaw Signum RESM angle encoder system 1 arcsec graduation accuracy

0.01 arcsec resolution

Page 4: February, 2008 IWAA08 G. Gassner, R. Ruland gassner@slac.stanford.edu 1 Georg Gassner Robert Ruland, SLAC February, 2008 Introduction Rotary Table Calibration

4February, 2008IWAA08

G. Gassner, R. [email protected]@slac.stanford.edu

Calibration of the rotary table (1)

Rosette technique calibrate precision polygon prismssum of the angles measured must result in 360 degrees Moeller-Wedel electronic autocollimator (0.1arcsec accuracy, 0.05arcsec reproducibility)

Page 5: February, 2008 IWAA08 G. Gassner, R. Ruland gassner@slac.stanford.edu 1 Georg Gassner Robert Ruland, SLAC February, 2008 Introduction Rotary Table Calibration

5February, 2008IWAA08

G. Gassner, R. [email protected]@slac.stanford.edu

Calibration of the rotary table (2)

Derived Technique set the fixture with the mirrors in a way that mirror ‘a’ is in line with autocollimator ‘a’

rotate the table together with the fixture until mirror ‘b’ is in line with autocollimator ‘b’

Page 6: February, 2008 IWAA08 G. Gassner, R. Ruland gassner@slac.stanford.edu 1 Georg Gassner Robert Ruland, SLAC February, 2008 Introduction Rotary Table Calibration

6February, 2008IWAA08

G. Gassner, R. [email protected]@slac.stanford.edu

Calibration of the rotary table (3)

Absolute value of an angle between two positions of the rotary table can be determined with ±0.2 arcsec std.

Page 7: February, 2008 IWAA08 G. Gassner, R. Ruland gassner@slac.stanford.edu 1 Georg Gassner Robert Ruland, SLAC February, 2008 Introduction Rotary Table Calibration

7February, 2008IWAA08

G. Gassner, R. [email protected]@slac.stanford.edu

Laser Tracker Horizontal Angle Measurement System Investigation

Test Setupmirror as a target for laser tracker

parallelism of the two rotation axes

axial displacement of the rotation axes

Rotary table

Faro Laser Tracker

Mirror

Page 8: February, 2008 IWAA08 G. Gassner, R. Ruland gassner@slac.stanford.edu 1 Georg Gassner Robert Ruland, SLAC February, 2008 Introduction Rotary Table Calibration

8February, 2008IWAA08

G. Gassner, R. [email protected]@slac.stanford.edu

Comparison SMR vs. SMM

Test of Mirror (SMM) and Retroreflector (SMR)

Only small deviations can be found

Page 9: February, 2008 IWAA08 G. Gassner, R. Ruland gassner@slac.stanford.edu 1 Georg Gassner Robert Ruland, SLAC February, 2008 Introduction Rotary Table Calibration

9February, 2008IWAA08

G. Gassner, R. [email protected]@slac.stanford.edu

Calibration Results

Two calibration runs with

Faro Laser Tracker Xi

Two calibration runs with

Faro Laser Tracker Xi V2

Page 10: February, 2008 IWAA08 G. Gassner, R. Ruland gassner@slac.stanford.edu 1 Georg Gassner Robert Ruland, SLAC February, 2008 Introduction Rotary Table Calibration

10February, 2008IWAA08

G. Gassner, R. [email protected]@slac.stanford.edu

ADM Calibration on Interferometer Bench

Calibration of the scale factor of the ADM in the field.

Tests of the general performance and tests to detect malfunctions in the laboratory.

Laser Tracker

Carriage Interferometerd

Horizontal Bench

Page 11: February, 2008 IWAA08 G. Gassner, R. Ruland gassner@slac.stanford.edu 1 Georg Gassner Robert Ruland, SLAC February, 2008 Introduction Rotary Table Calibration

11February, 2008IWAA08

G. Gassner, R. [email protected]@slac.stanford.edu

Calibration Results

Calibration run with

Faro Laser Tracker Xi

Calibration runs with

Faro Laser Tracker Xi V2

Page 12: February, 2008 IWAA08 G. Gassner, R. Ruland gassner@slac.stanford.edu 1 Georg Gassner Robert Ruland, SLAC February, 2008 Introduction Rotary Table Calibration

12February, 2008IWAA08

G. Gassner, R. [email protected]@slac.stanford.edu

Conclusion

Rotary table - absolute angle accuracy < 0.2 arcsec

Rotary table to test any horizontal angle measurement instrument

Faro laser trackers angle measurement system deviations stable and correctible

With corrections, angle accuracies of 1 arcsec are possible.

Test of the ADM showed its usability as a replacement for the interferometer mode