MRX002 Pressure testing - November 2016

PET Aerosol Vial - MRX002 Pressure test results

Embed Size (px)

Citation preview

MRX002Pressuretesting- November2016

WhypressuretestMRX002?MRX002vialshavebeendesignedaslaboratorytoolsforthestudyofHFApMDIformulations.Theseweretestedforpressureresistance.TheburstpressuresrecordedinthisdocumentestablishthepressurelimitsandofMRX002vials.

ExperimentalTheburstpressuresof35MRX002vialswererecordedwiththeDHIndustriesLtdburstpressuretester(0325/100).MRX002vialswerefastenedsecurelytothetester,filledwithwaterthatwaspressurisedataconstantrate(1.3bar(g)/s)untilburstingofthevial.Thevialpressurewasrecordedevery100ms,andthepressuregaugewasaccuratewithin0.01bar.

ConclusionTheaverageburstpressureis50.1±4.2bar(g),rangingfrom41to56bar(g).

ThisiswellabovethevapourpressureofHFAsatroomtemperature(max5.7bar(g)forHFA134a),andat60oC(16.8bar(g)),asrequiredforleaktestingbytheFDA.

AboutMerxin:Merxinisagenericmedicaldevicedevelopmentcompany.Developmentandsupplyofgenericinhalationdevicesas“offtheshelfandreadytouse”forfinishedproductdevelopmentisourcorebusiness.Bysupplyingstandardised off-the-shelfdevices,Merxinprovidesclientswithaquickerlowercostandlowerriskroutetomarket.

Contact:Merxincanbecontactedat:[email protected] andviaourwebsite:www.merxin.com

Generic MedicalDevicesMadeEasyCopyrights:MerxinLtd2016© November2016

Thecontentofthisdocumentcanbechangedatanytime.PleasecontactMerxinforlatestversion.

BurstpressureBar(g)

Max 56.24Min 41.19

Average 50.1StDev 4.2

Median 51.51stquartile 46.23rdquartile 52.8

Results

35

40

45

50

55

60

65Bar(g)

Max

Min

AverageMedian

3Q

1Q

N=35vials,fromonesinglebatchRoomTemperature:22oC

Units:bars HFA227 HFA134a Propane Isobutane

20oC 3.90 5.72 7.51 2.75

60oC 11.75 16.82 18.54 7.76

ThesemeasurementswerecarriedoutinpartnershipwithDHIndustriesLtd

MRX002vialsburstpressuresrangefrom41.2bar(g)to56.2bar(g),withanaverageof50.1bar(g)andamedianof51.5bar(g).Thevapourpressureofcommonpropellantsare(bar):

MRX002vialsburstpressuresarehigherthanthevapourpressuresofHFA227,HFA134a,PropaneandIsobutane.TheFDArequiresthatHFApMDIproductsbetestedatupto60oCforupto3minutes.ThevapourpressuresofthepropellantsatthistemperaturearelowerthantheMRX002burstpressure.AmorphousPETsoftensat67oC,thereforecareshouldbetakenwhentestingMRX002samplesat60oC.