71
The microdosimetric variance-covariance method used for beam quality characterization in radiation protection and radiation therapy Jan Erik Lillhök Stockholm University

The microdosimetric variance-covariance method used for ...196874/FULLTEXT01.pdf1994, Tilikidis et al 1996, Wambersie et al 2002, Gerlach et al 2002, De Nardo et al 2004]. The linear

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

  • The microdosimetric variance-covariance method used for beam quality characterization

    in radiation protection and radiation therapy

    Jan Erik Lillhök

    Stockholm University

  • The variance-covariance method used for beam quality characterization J. Lillhök

    2

    © Jan Erik Lillhök, Stockholm 2007 ISBN (91-7155-391-6) Printed in Sweden by Universitetsservice AB, Stockholm 2007 Distributor: Medical Radiation Physics, Stockholm University, Karolinska Institutet

  • The variance-covariance method used for beam quality characterization J. Lillhök

    3

    Till Sofia, Gösta och den lillae

    ”Epäusko eilispäivän

    tänään on totinen tieto,

    tämän päivän taikausko

    huomenna on huomenrusko.”

    Eino Leino 1924

  • The variance-covariance method used for beam quality characterization J. Lillhök

    4

  • The variance-covariance method used for beam quality characterization J. Lillhök

    5

    Contents

    ABSTRACT....................................................................................................................................7

    1. INTRODUCTION .......................................................................................................................9

    2. THEORY..................................................................................................................................11 2.1. The variance-covariance method ............................................................................................................ 11 2.2. The tissue-equivalent proportional counter.............................................................................................. 14 2.3. Nanodosimetry and radiation quality ....................................................................................................... 15

    3. MATERIALS AND METHODS.................................................................................................19 3.1. Detector construction .............................................................................................................................. 19 3.2. Algorithms ............................................................................................................................................... 22 3.3. Separation of dose components.............................................................................................................. 27 3.4. Monte Carlo simulations.......................................................................................................................... 33

    4. RESULTS ................................................................................................................................37 4.1. Instrument response................................................................................................................................ 37 4.2. Dose equivalent measurements .............................................................................................................. 45 4.3. Pulsed radiation fields ............................................................................................................................. 48 4.4. Separation of dose components.............................................................................................................. 50 4.5. A low-dose rate application ..................................................................................................................... 51 4.6. Uncertainties ........................................................................................................................................... 54 4.7. Nanodosimetry in radiation therapy ......................................................................................................... 57

    5. CONCLUSION AND OUTLOOK .............................................................................................61

    ACKNOWLEDGEMENTS............................................................................................................62

    REFERENCES ............................................................................................................................64

    APPENDIX...................................................................................................................................69

  • The variance-covariance method used for beam quality characterization J. Lillhök

    6

    PAPERS This thesis is based on the following papers. I. Kyllönen J-E, Lindborg L and Samuelson G, 2001. The response of the Sievert instrument in neutron beams up to 180 MeV. Radiation Protection Dosimetry 94 (3) 227 – 232 II. Kyllönen J-E and Lindborg L. Photon and neutron dose discrimination using low pressure proportional counters with graphite and A150 walls. Radiation Protection Dosimetry (2007) doi:10.1093/rpd/ncm162 III. Kyllönen J-E, Lindborg L and Samuelson G, 2001. Cosmic radiation measurements on-board aircraft with the variance method. Radiation Protection Dosimetry 93 (3) 197 – 205 IV. Lillhök J, Beck P, Bottolier-Depois J F, Latocha M, Lindborg L, Roos H, Roth J, Schraube H, Spurny F, Stehno G, Trompier F and Wissmann F. A comparison of ambient dose equivalent meters and dose calculations at constant flight conditions. Accepted 18 Dec 2006 for publication in Radiation Measurements V. Lillhök J E, Grindborg J E, Lindborg L, Gudowska I, Alm Carlsson G, Söderberg J, Kopeć M and Medin J. Nanodosimetry in a clinical neutron therapy beam using the variance-covariance method and Monte Carlo simulations. Submitted to Physics in Medicine and Biology 19 Feb 2007

  • The variance-covariance method used for beam quality characterization J. Lillhök

    7

    ABSTRACT Radiation quality is described by the RBE (relative biological effectiveness) that varies with the ionizing ability of the radiation. Microdosimetric quantities describe distributions of energy imparted to small volumes and can be related to RBE. This has made microdosimetry a powerful tool for radiation quality determinations in both radiation protection and radiation therapy. The variance-covariance method determines the dose-average of the distributions and has traditionally been used with two detectors to correct for beam intensity variations. Methods to separate dose components in mixed radiation fields and to correct for beam variations using only one detector have been developed in this thesis. Quality factor relations have been optimized for different neutron energies, and a new algorithm that takes single energy deposition events from densely ionizing radiation into account has been formulated. The variance-covariance technique and the new methodology have been shown to work well in the cosmic radiation field onboard aircraft, in the mixed photon and neutron fields in the nuclear industry and in pulsed fields around accelerators. The method has also been used for radiation quality characterization in therapy beams. The biological damage is related to track-structure and ionization clusters and requires descriptions of the energy depositions in nanometre sized volumes. It was shown that both measurements and Monte Carlo simulation (condensed history and track-structure) are needed for a reliable nanodosimetric beam characterization. The combined experimental and simulated results indicate that the dose-mean of the energy imparted to an object in the nanometre region is related to the clinical RBE in neutron, proton and photon beams. The results suggest that the variance-covariance technique and the dose-average of the microdosimetric quantities could be well suited for describing radiation quality also in therapy beams.

  • The variance-covariance method used for beam quality characterization J. Lillhök

    8

  • The variance-covariance method used for beam quality characterization J. Lillhök

    9

    1. INTRODUCTION A common device used to characterize radiation fields in terms of the linear energy transfer (LET) is the low-pressure proportional counter of tissue-equivalent material [ICRU 1984, Waker 1995, Gerdung et al 1995]. The measured quantity, lineal energy, can be used as a measure of the LET although the validity of the approximation depends on energy and type of particle, and on the size of the simulated volume [Kellerer and Chmelevsky 1975, Leuthold and Burger 1985]. Experimental microdosimetry has therefore been successful in radiation protection [Menzel et al 1989, Schmitz et al 1995, EURADOS 2004] and for beam characterization in radiation therapy [Menzel et al 1994, Tilikidis et al 1996, Wambersie et al 2002, Gerlach et al 2002, De Nardo et al 2004]. The linear energy transfer (LET) is a macroscopic quantity defined by the average energy lost by a particle while the lineal energy (y) is a stochastic quantity describing the energy distribution in small volumes. The microdosimetric quantities can therefore spot regions of concentrated ionizations, which is important since track-structure and electron clusters are crucial for describing the mechanisms of radiation damage [Goodhead 1994, Nikjoo et al 1997, Brahme et al 1997, Hill 1999]. If the probability of damage on the nanometre level is related to the energy imparted to a relevant volume, then it might be better described by the microdosimetric distribution for the same object size than by the LET. Measurements of the energy deposition from single events can be difficult at high dose rates and in pulsed radiation fields due to pile-up of several events. Experimental nanodosimetry with large gaseous detectors also faces the problem of a non-uniform gas multiplication region that distorts the measured spectrum [Ségur et al 1994]. Smaller detectors reduce this problem and single-event measurements have been reported for simulated objects in the nanometre region [Kliauga 1990, Anachkova et al 1997, Cesari et al 2002]. Some of these experimental limitations are also reduced when using the variance-covariance method in which the dose-average of the lineal energy (Dy ) is determined from the multiple-event energy distribution [Kellerer 1968, Bengtsson 1970, Lindborg and Bengtsson 1971, Lindborg et al 1988]. The measurements are based on charge measurements during specific time intervals and the method can be used in pulsed beams without pile-up effects. The technique can also be used without gas multiplication and is therefore suitable for measurements at low pressures and in high dose-rate applications [Goldhagen et al 1990, Grindborg et al 1995, Lindborg and Grindborg 1997]. The questions addressed in this thesis are how well the variance-covariance method can be used to determine ambient dose-equivalent in radiation protection, and if it can be used for beam quality characterization in radiation therapy. For this purpose an instrument based on the variance method was built [Paper I]. The method was further developed for some of the requirements in radiation protection applications. An algorithm correcting for beam intensity variations using one instead of the usual two detectors was formulated [summary]. An improved algorithm for the quality factor in mixed fields and methods to separate low- and high-LET dose components were developed [Paper II, III, summary]. The instrument and methods were evaluated in

  • The variance-covariance method used for beam quality characterization J. Lillhök

    10

    measurements onboard aircraft in the cosmic radiation field [Papers III, IV], in the nuclear industry [Paper II] and in pulsed fields around accelerators [summary]. Experiments and Monte Carlo simulations using both macroscopic (condensed history) and microscopic (track-structure) codes of the energy deposition in the nanometre region were compared and the relation between the dose-mean lineal energy (Dy ) and radiation quality in radiation therapy was investigated [Paper V, summary].

  • The variance-covariance method used for beam quality characterization J. Lillhök

    11

    2. THEORY The fundamental difference between quantities like absorbed dose and specific energy, or LET and lineal energy, is their relation to the stochastic nature of radiation interaction. Absorbed dose and LET are macroscopic quantities defined by average energy depositions, while specific and lineal energy are stochastic quantities described by distributions of energy deposition events (section 2.1). One of the important microdosimetric quantities is the dose-mean lineal energy Dy that can be determined

    with the variance-covariance method (section 2.1) Experimentally Dy is measured using tissue-equivalent low-pressure proportional counters (TEPC) or ion chambers to simulate the energy deposition in small volumes (section 2.2). There are also nanodosimetric arguments for relating single-event spectra and Dy to the probability of biological damage and radiation quality (section 2.3). Comprehensive descriptions of microdosimetry are for the interested reader given by Kellerer [1975], ICRU [1984] and Rossi and Zaider [1995].

    2.1. The variance-covariance method The fundamental dosimetric quantities absorbed dose (D) and linear energy transfer (L) are defined by

    dm

    dD

    ε= (2.1)

    and

    dl

    dEL = (2.2)

    where εd is the average energy imparted to the small mass dm and dE is the mean energy lost by a particle traversing a distance dl [ICRU 1993a]. These quantities are sometimes referred to as macroscopic although they are point functions. The definitions in terms of the average energy imparted do not reflect the stochastic nature of radiation interaction with matter. The quantities can vary in a large volume but they do not fluctuate like the energy deposition from individual ionizing particles. The energy imparted (ε) to a slightly larger volume of mass m is described by the microdosimetric quantity specific energy (z)

    mz

    ε= (2.3)

    The energy imparted in (2.3) varies with the absorbed dose, and can be due to interactions of several particles, or alternatively but not equivalently, several events. An

  • The variance-covariance method used for beam quality characterization J. Lillhök

    12

    event denotes the interaction of statistically correlated particles including also secondary particles such as delta-electrons [ICRU 1984, ICRU 1993a]. Both the number of events and the energy imparted in each event are described by Poisson statistics. The distribution of z-values at dose D is denoted f(z,D) and the average specific energy equals the absorbed dose in the limit

    DzDzzfzm →= →

    ∫ 00

    d),( (2.4)

    The lineal energy (y) is defined in terms of energy imparted in a single event and the mean chord-length of the object (l ).

    ly

    ε= (2.5)

    The single-event frequency and dose distributions f1(z) and d1(z) (index 1 denotes it is a single event) are independent of the absorbed dose and they are characteristic of the radiation field itself. The average values for the lineal energy are defined by

    ∫∞

    =0

    1 d)( yyyfyF (2.6)

    F

    F

    FD y

    yyyfy

    yyyydy

    2

    0

    12

    0

    1 d)(1

    d)( === ∫∫∞∞

    (2.7)

    Frequency and dose averages of the energy imparted (ε) and the specific energy (z) are defined analogously. The relative variance (V1) of a single-event distribution f1(z) can like any distribution be written in terms of its first and second moments.

    112

    2

    2

    22

    2

    2

    2

    2

    1 −=−=−===

    F

    D

    F

    F

    F

    FF

    F

    y

    F

    z

    y

    y

    y

    y

    y

    yy

    yzV

    σσ (2.8)

    In a multiple-event spectrum there are several events that contribute to the energy imparted. For a Poisson distribution with n events on average, the relative variance in the event number is given by

    nn

    n

    nV nn

    122

    2

    ===σ

    (2.9)

    From statistical theory it is known that the variance of the average of independent stochastic variables decreases with the number of events [e.g. Rice 1995]. Since the events discussed here are independent by definition, the variance (V1) of the single-event spectrum has also decreased with a factor of n . The total relative variance in a multiple event spectrum is then according to (2.8) and (2.9) given by

  • The variance-covariance method used for beam quality characterization J. Lillhök

    13

    Dm

    yl

    zm

    yl

    znm

    yl

    yn

    yV

    n

    VV DD

    F

    D

    F

    DnR ⋅

    ⋅=⋅

    ⋅=⋅⋅==+= 1 (2.10)

    where m is the mass, l is the mean chord-length and D is the absorbed dose delivered by n events. The dose-mean lineal energy is then from (2.10) given by

    −=== z

    z

    z

    l

    m

    zl

    mzV

    l

    my zRD

    22σ (2.11)

    All variance sources such as the contribution from the chord-length distribution, energy straggling, electronic noise and varying beam intensity, contribute to the total variance. Kellerer and Rossi [1984] published a formulation of how to correct for a beam intensity variation by subtracting the relative covariance (CR) between two detectors (A and B) exposed in the same field. For detector A the expression is

    ( )

    −=−=

    B

    BA

    A

    AARRD z

    zz

    z

    z

    l

    mzCV

    l

    my

    2

    (2.12)

    The purpose of the second detector is to monitor beam variations but the detectors do not necessarily have to be identical. Response differences might however influence the correction CR if the beam variations are not the same for different radiation components. For example a variation in a slow neutron component is not necessarily monitored correctly by a graphite-walled ionization chamber with low neutron sensitivity. The variance-covariance method was developed further by Kellerer [1996a–c] by including corrections also for slow spatial variations, i.e. changes in the signal ratio between two detectors. Slow means the influence of the spatial variation in successive measurement intervals is negligible. The dose-average lineal energy is expressed in terms of dose in two detectors A and B and the dose for the preceding measurement intervals denoted with indices A-1 and B-1. For detector A the expression is

    ( )

    +⋅

    −=−−

    11

    12 2

    ABBAA

    AABA

    A

    AD

    zzzzz

    zzzz

    z

    z

    l

    my (2.13)

    An observed variation in the signal ratios between two detectors could be caused by detectors moving through an inhomogeneous field, spatial variations in the field itself or variations in the gas-gain of the proportional counter.

  • The variance-covariance method used for beam quality characterization J. Lillhök

    14

    2.2. The tissue-equivalent proportional counter An established method in microdosimetry is to use a low-pressure proportional counter to simulate the energy imparted to a microscopic tissue-volume [ICRU 1984, Rossi and Zaider 1995, Waker 1995, Gerdung et al 1995]. The detector is often spherical and made of tissue-equivalent A-150 plastic and filled with propane or methane based tissue-equivalent gas [ICRU 1989]. The mass stopping powers for charged particles in the detector gas and in a microscopic tissue object are equal if differences in density and atomic composition are neglected.

    µdet

    =

    ρρSS

    (2.14)

    Consider a detector filled to a gas pressure corresponding to a density given by

    det

    µµ

    det

    µµdet r

    r

    l

    lρρρ == (2.15)

    where µl , detl , rµ, and rdet are the mean chord-lengths and the radii of the tissue object

    and the detector, respectively. A particle passing the detector volume will then loose the same amount of energy as when passing the tissue object.

    ( ) ( ) µµµµ

    lS

    lS ερ

    ρρ

    ρε =

    =

    = detdetdet

    det (2.16)

    The cross sectional areas, the volumes and the masses of the detector and the tissue object are related as

    2

    2det

    detµ

    µ r

    rAA = (2.17)

    3

    3det

    det

    µ

    µr

    rvv = (2.18)

    2

    2det

    2

    2det

    3

    3det

    detdetdetdet

    µ

    µ

    µ

    µµ

    µ

    µµ

    µr

    rm

    r

    rv

    r

    rv

    r

    rvm ==== ρρρ (2.19)

    For a constant particle fluence Φ in both volumes, a constant average stopping power S , and delta-particle equilibrium, the energy imparted to the detector is related to the energy imparted to the tissue volume as

    2

    2det

    2

    2det

    detdetµ

    µµ

    µ r

    rES

    r

    rASAE =Φ=Φ= (2.20)

  • The variance-covariance method used for beam quality characterization J. Lillhök

    15

    The absorbed doses in the two objects are then the same as seen from (2.1), (2.19) and (2.20)

    µµ

    µ

    µµ

    µµ

    Dm

    E

    r

    rm

    r

    rE

    m

    ED ====

    2

    2det

    2

    2det

    det

    detdet (2.21)

    For a Poisson distributed random variable, like the number of particles crossing the detector volume, the relative variance decrease with the average number of particles according to (2.9). The relative variance in the detector and the micrometer volume are then given by

    2det

    2

    detdetdet r

    rV

    A

    AV

    N

    NVV µµ

    µµ

    µµ =Φ

    Φ== (2.22)

    Equations (2.11), (2.12), (2.13) and (2.14) then gives an expression for the dose-average lineal energy determined from the relative variance (Vdet) and the absorbed dose (Ddet) in repeated measurements.

    µµ

    µ

    µ

    µ

    µµµD l

    DVm

    l

    Dr

    rV

    r

    rm

    l

    DVmy detdetdet

    det2

    2det

    det2det

    2

    det ⋅⋅=⋅⋅

    =⋅⋅

    = (2.23)

    It is also useful to express the total Dy -value of a radiation field in terms of the

    absorbed dose fractions and Dy -values of its components, for example the photon and neutron component.

    ( ) nDnDnnDnDD ydydydyddyyydy ,,,, 1)( +−=+== ∫ γγγ (2.24)

    2.3. Nanodosimetry and radiation quality One reason to study energy depositions in variable volume sizes is to see if a match between a measured quantity and some beam quality characteristics can be found for some specific size. The theory of dual radiation action [Kellerer and Rossi 1972] proposed that energy depositions in a micrometer sized object could account for the curvature of survival curves. The underlying assumption was that the damage was combined by two sublesions, which is equivalent with a damage probability proportional to the square of the energy imparted to the object (z2). Equation (2.11) gives that the average of z2 can be written

  • The variance-covariance method used for beam quality characterization J. Lillhök

    16

    222 Dm

    yDlz

    m

    yzlz DD +=+= (2.25)

    Since the probability of a damage was proportional to z2, the average biological effect due to this damage is at absorbed dose D given by the familiar linear-quadratic expression

    +=+== 222)( DD

    m

    ylD

    m

    yDlzD DD ββββε (2.26)

    Such a mechanistic derivation of a linear-quadratic dose-effect relation is not realistic [Kellerer and Rossi 1978, Goodhead 1987, Zaider 1998]. A therapeutic situation might well be described by a linear-quadratic model, but it is far too complex to be defined only by the first two moments of the energy imparted to a target volume of a few micrometer. However, it will be argued below that the concept of relating energy imparted to an object to the production of a biological damage can account for at least a part of empirical observations determined by damages on the nanometre level. Consider a number of very small objects inside a large macroscopic medium exposed to an absorbed dose D. The probability of having no interaction in one of the objects increases with decreasing D and with decreasing object size. Most of the nanometre sized objects do actually not receive any dose at all, while the affected volumes receive an extremely high dose on average. The absorbed dose to an affected volume is also most likely due to energy depositions from a single event, i.e. the frequency-mean specific energy Fz in microdosimetric terms. An increasing absorbed dose D is

    therefore only related to an increasing number of affected volumes, while Fz in each

    object remains the same. For example, Fy for 100 keV electrons in a 100 nm diameter volume is 1.6 keV/µm [Nikjoo et al 1994]. An absorbed dose D=2 Gy gives then

    Fz =33 Gy to 6% of all 100 nm objects, while the remaining 94% are unaffected. At D=100 mGy, only 0.3% of the volumes are affected but they still recive 33 Gy. Assume that a biological effect (ε) is the result of damages on the nanometre level and that the production of damage is somehow related to the energy imparted to a corresponding volume. The damage probability (p) can then be expressed as a function of a microdosimetric quantity such as the lineal energy. As noted above the average number of affected nanometer volumes (n ) increases linearly with absorbed dose D, and the biological effect generated by damages on the nanometre scale can be written

    +++=== ∫ ...)()()()( 210 D

    FF

    ykky

    kDdyypyf

    y

    DypnDε (2.27)

    where ki are the constants of a serial expansion of p(y). If it is assumed that the damage frequency has a Poisson distribution, and that one damage is enough to give an effect, then the probability of no effect is P(ε=0) given by

  • The variance-covariance method used for beam quality characterization J. Lillhök

    17

    ( )

    +++−=−== ...)()0(ln 210 D

    F

    ykky

    kDDP εε (2.28)

    As already mentioned it not realistic to understand an empirical survival relation only in terms of the initial damage probability like in (2.28). But a biological effect generated by initial damages on the nanometre level might however account for a part of the linear term in such a relation. Equation (2.28) would then be included in an exponential survival probability S(D) that expanded in a power series becomes

    ( )

    ++++

    +++−= ...´...)(ln 32210 DDDykky

    kDDS D

    F

    γβα (2.29)

    Another important nanodosimetric feature is that the relative dose contribution from delta-electrons increases at small object sizes [ICRU 1984, Leuthold and Burger 1988]. The energy distribution of delta-electrons is essentially determined by the velocity and not the LET of the primary particle. Low- and high-LET radiations become therefore more similar in terms of the energy imparted to small objects. The differences between the single-event distribution (and Dy ) decrease, and the shape of a biological weighting function RBE(L) must be different from a function RBE(Dy ). The maximum RBE occurs typically around and LET of 100–200 keV/µm, which with a mean-energy to produce an ion pair of W=30 eV corresponds to 30–80 ionizations in a 10 nm volume. For comparison the largest cluster sizes, i.e. the maximum number of ionizations, produced by a 4He ion travelling 10 nm is about 40 at 0.17 MeV/u [Grosswendt 2002]. If ionization clusters are assumed to never be larger than this, i.e. there is no saturation in the production of initial damage, and the probability of damage is proportional to the square of the lineal energy (or specific energy), then (2.27) and (2.28) gives

    ( ) DFF

    ykDy

    ykDdzkyyf

    y

    DP −=−=∫−==

    22)()0(ln ε (2.30)

    If for some reason the other contribution (α´) to the linear term in D in (2.29) can be disregarded in comparison with the initial damage probability (2.28), then a linear-quadratic expression considering only the first two terms in (2.29) becomes

    ( ) ( ) ( )22 (()(ln DDykDDDS D ββα +−=+−= (2.31)

    In (2.26) the average energy imparted to the micrometre volume (2z ) defines both the first and second terms and relates them through

    DDD z

    mm

    yl βεββα === (2.32)

    Or in other words, the curvature (α/β) of the survival curve is defined by the dose-mean of one of the microdosimetric quantities in a micrometer sized volume. This is however

  • The variance-covariance method used for beam quality characterization J. Lillhök

    18

    fundamentally different from (2.31) where the linear and the quadratic terms are if not independent, at least not necessarily dependent. The linear term is defined by the initial damages on the nanometre volume only, while the quadratic term can account for other mechanisms (e.g. cell repair) of the complex biological system.

  • The variance-covariance method used for beam quality characterization J. Lillhök

    19

    3. MATERIALS AND METHODS Investigating the applicability of the variance-covariance method in radiation protection requires a reliable instrument for the measurements (Paper I). The method determines the dose mean of the lineal energy (Dy ), which can be used to determine a quality factor depending linearly on the lineal energy. This is inappropriate for high-LET radiation and an improved algorithm with dose rate limits for its application was formulated (Paper III and section 3.2). The low- and high-LET dose components are not directly accessible from a multiple-event spectrum and methods to separate the components were developed (Paper II and section 3.3). A practical instrument cannot be too large or heavy and using one detector instead of two is therefore attractive and a covariance correction for one detector was developed (section 3.2). Monte Carlo methods are useful not only for response simulations, but also for verification and extension of measurement results in difficult applications such as nanodosimetry (section 3.4).

    3.1. Detector construction Two tissue-equivalent proportional counters (TEPC) and a graphite walled proportional counter (GPC) of identical design were constructed for the purpose of this investigation [Kyllönen et al 1997, Paper I, Paper II]. The TEPCs are made of A-150 plastic [ICRU 1989] with a wall-thickness of 5 mm. The sensitive volume of 1.2 dm3 is cylindrical with a height and a diameter of 11.5 cm. With a propane-based tissue-equivalent gas filling of 1.45 kPa it simulates a tissue-volume of 2 µm mean chord-length. The polarity effect, i.e. the difference in measured saturation current using positive or negative voltage, was determined to 0.2%. The A-150 plastic is hygroscopic due to its polyamide (nylon®) content and the detectors must be evacuated at an elevated temperature (baked) before gas-filling [Braby et al 1995, Gerdung et al 1995]. In a badly evacuated system the gas gain will decrease slowly with time due to an increasing pressure and the poor multiplication characteristics of the water vapour diffusing out from the walls. A proper baking is also of importance for the temperature dependence of the system since the water vapour content in the detector wall varies with temperature. An improved heating procedure for baking the tissue-equivalent material has been established and the long term stability is improved by refilling the detectors to their final working pressure after a few days when the diffusion rate back into the plastic has decreased. The saturation current of a sealed detector, i.e. the gas pressure with a constant We (mean energy to produce an ion pair), increased with 2% during 2 years. The gas multiplication at the same high-voltage decreased with -15% in the same time period. The temperature dependence of the detectors was determined to -0.4% per ºC. The baking procedure is not necessary for the graphite detector but it needs to be carefully evacuated at a low pressure.

  • The variance-covariance method used for beam quality characterization J. Lillhök

    20

    Figure 3.1. Cross-section of the low-pressure proportional counter with tissue-equivalent or graphite walls used in radiation protection applications.

    The charge created in the detector volume is measured with capacitor feed-back electrometers developed by Combinova AB in cooperation with Göran Samuelson at the Swedish Radiation Protection Authority (SSI). The voltage over the 1000 pF capacitor is measured 10 times per second with a 24-bit analog-to-digital converter (ADS 1210U). One bit is used for polarity and the other 23 bits for dividing the maximum capacitor -voltage of 5V into steps of 0.6 µV. The electronic noise is dominated by the 0.6 µV (rms) contribution from the analog-to-digital-converter. The capacitor is reset with a current in the opposite direction, every 57 minutes at 1 pA. The highest reset frequency is about one reset every second, which with a gas multiplication of 1000 corresponds to a dose rate of about 10 mGy/h. This is of importance in e.g. cosmic radiation measurements where the atmospheric dose rates can suddenly increase in connection with strong solar eruptions and ground level events [Lantos 1993, Spurný and Daschev 2002]. Using a lower gas gain increases of course the maximum measurable dose rate correspondingly.

  • The variance-covariance method used for beam quality characterization J. Lillhök

    21

    The relative variance decrease with an integration time below 300 ms. The charge collection time in the detector is of the order of microseconds and the variance decrease is attributed to electronic time constants. The decrease is also independent of the radiation field (neutron, photon or mixed fields) and can be taken into account by applying a correction of 1.1 to measurements with 100 ms integration time. This is preferable for example in pulsed radiation fields where a high sampling frequency is desireable (section 4.3). The quantity of interrest in radiation protection is the ambient dose-equivalent H*(10), i.e. the dose-equivalent produced at 10 mm depth in the ICRU sphere in an expanded and aligned field [ICRU 1993]. The dose-equivalent (H) is given by the product of the absorbed dose and the quality factor and the measured dose-equivalent (Hm) is determined from a calibration factor (N), the absorbed dose to detector gas (Dgas) and the average quality factor (DQ )

    ),,,,,,,,( 2 Mmle

    WbaqNfQNDH gasqDgasm σ== (3.1)

    In practice Hm is calculated from a number of variables and coefficients. The average electric charge (q ) and its variance ( 2qσ ) created during 100 ms in the detector volume are measured with the electrometers described above and traceable to national standards. The parameters a and b were determined from H*(10)-response investigations in neutron fields of different energies [Paper I]. The gas mass (gasm ) and

    mean chord-length (l ) are determined from the physical detector volume, a gas density of 1.798 kg.m-3 at 20°C and 100 kPa [ICRU 1984] and the gas pressure measured with a calibrated barometer traceable to national standards. The gas amplification (M) is determined as the current measured at the working voltage (typically -1000V) divided with the saturation current, i.e. with no gas gain and complete ion collection. The ambient dose-equivalent calibration factor relates the instrument reading to a reference ambient dose-equivalent value in a 137Cs-field.

    m

    ref

    H

    HN

    )10(*= (3.2)

    This calibration factor is used in all applications but with different W/e-values, 26.8 J/C in low-LET fields and 31 J/C in high-LET fields [Lindborg 1985, IAEA 1995, Grosswendt 2002]. In mixed radiation fields the combined value 28 J/C is used. The detectors used in the nanodosimetric investigations (section 4.7) are described in detail in Paper V.

  • The variance-covariance method used for beam quality characterization J. Lillhök

    22

    3.2. Algorithms An important development of a practical instrument is the reduction of size and weight. The variance-covariance method would for example gain from using only one detector, which requires a different solution for the covariance correction. The method of using successive measurements in (2.13) is therefore reformulated for one detector. If the intensity changes are slow compared to the sampling time, it is possible to use the covariance between the data series formed by successive intervals in one detector. The expression for the dose-average lineal energy for detector A corrected for a covariance between two series (odd and even) becomes in analogy with (2.12)

    − →

    ⋅−= ∞→

    A

    evenodd

    A

    An

    evenodd

    evenoddA

    A

    AD

    z

    zz

    z

    z

    l

    m

    zz

    zzz

    z

    z

    l

    my

    22

    (3.3)

    A radiation protection instrument must also be able to determine the quality factor (Q) reasonably well. The average quality factor can be calculated from a single-event dose distribution d(y) by for example approximating L with y in the quality factor expression QICRP60(L) from ICRP60 [ICRP 1991].

    ( ) ∫∫∫

    ∫∫∞

    ∞∞

    +++=

    =≈≡

    100

    100

    10

    10

    0

    0

    60

    0

    6060,

    )(300

    )(2.232.0)(

    )()()()()(

    dyydy

    dyydydyyd

    dyydyQdLLdLQLQ ICRPICRPICRPD (3.4)

    A first order linear approximation is given by

    ( ) DlinlinD ybadyydyQdyydbyayQ +==+= ∫∫∞∞

    00

    , )()()()( (3.5)

    The linear approximation (3.5) will overestimate the average quality factor (3.4) in the high-LET region. This overestimation can be substantial in neutron radiation fields above a few MeV due to the contributions from secondary alpha particles and heavier recoils [Paper I]. The substitution L=y is also only approximate due to the differences between the two quantities [Kellerer and Chmelevsky 1975, Leuthold and Burger 1985]. This is however often acceptable since instrument characteristics can be taken into account when calibrating the quality factor expression for an optimal H*(10)-response.

  • The variance-covariance method used for beam quality characterization J. Lillhök

    23

    0

    1

    10

    100

    1000

    0 50 100 150 200

    Energy deposition per mean chord-length [keV/µm]

    Num

    ber

    of c

    ount

    s

    Figure 3.2. A multiple-event spectrum in a mixed low- and high-LET radiation field with dose rate 11 µGy/h and Dy =5 keV/µm. The ordinate shows the number of measurements in 0.17 keV/µm bins. The abscissa shows the energy imparted in 100 ms divided by the mean chord-length.

    A different formulation of the quality factor combines single- and multiple-events as described in Paper III. Figure 3.2 shows a typical multiple-event spectrum in a mixed photon and neutron field. The distribution is dominated by multiple events below 50 keV/µm, and single-events above 50 keV/µm. Energy depositon from high-LET particles can easily be distinguished in the spectrum and the probability of having more than one high-LET event in a 100 ms measurement is negligible. The quality factor for the spectrum can therefore be composed by one term for the multiple-events, and another for the single-events. The linear function (3.5) is used up to 150 keV/µm where (3.5) intersects with (3.4). For the events above 150 keV/µm the formulation in (3.4) is used (Figure 3.3).

    ( )

    ( ) ( )∑

    ∫∫∫

    =

  • The variance-covariance method used for beam quality characterization J. Lillhök

    24

    Here dy

  • The variance-covariance method used for beam quality characterization J. Lillhök

    25

    energy imparted (Figure 3.4). A larger detector that simulates the same mean chord-length will also shift the distribution towards higher values since it has a larger cross-sectional area and a larger energy imparted for the same particle fluence (2.20). With increasing radiation quality (Dy ) the variance increases (2.10), which broadens the spectrum.

    0

    20

    40

    60

    80

    100

    120

    10 100 1000

    Energy deposition per mean chord-length [keV/µm]

    Num

    ber

    of c

    ount

    s

    yD=10 keV/µm

    yD=2 keV/µm

    5 cm detector

    yD=2 keV/µm

    230 µGy/h77 µGy/h230 µGy/h

    10 cm detector

    Figure 3.4. Shift of the multiple-event spectrum with different detector sizes, Dy -values and dose rates. The simulated mean chord-length is 2 µm. The ordinate shows the number of measurements in 1 keV/µm bins. The abscissa shows the energy imparted in 100 ms divided by the mean chord-length.

    Assume that a high-LET event (yhigh) must be more than k standard deviations (

    µlE /σ )

    higher the average value to be clearly resolved.

    µµ lEµ

    lEµ

    high kl

    mDk

    l

    Ey

    /det

    /σσ +=+= (3.7)

    Rewrite (2.23) in terms of the average energy per measurement interval.

    E

    l

    l

    E

    lE

    l

    EV

    l

    DVmy

    µlE

    µ

    µ

    lE

    µµD

    µµ

    2/

    2

    2/detdetdetdet

    σσ=

    === (3.8)

  • The variance-covariance method used for beam quality characterization J. Lillhök

    26

    Equations (3.7) and (3.8) gives a second order equation for µlE /

    σ with the solution

    DhighDD

    lEyy

    ykykµ

    +

    ±−=

    2

    / 22σ (3.9)

    With (3.7) and (3.9) we can express the maximum dose-rate at which an event yhigh can be resolved, in terms of the detector gas density (ρdet) and measured total Dy for a detector with radius rdet.

    det2

    det

    22

    det

    /

    max

    22

    ρπσ

    r

    yyyk

    kyk

    y

    m

    kylD

    DhighDD

    highlEhigh µ

    +

    −+

    =−

    = (3.10)

    Figure 3.5 shows Dmax for yhigh=150 keV/µm as a function of Dy of the radiation field for two detector sizes and for k=3. Equation (3.6) can for example with a 10 cm detector be used to correct for high-LET events above 150 keV/µm up to about 60 µGy/h in a radiation field with Dy = 5 keV/µm.

    0

    100

    200

    300

    400

    1 10 100

    Dose-mean lineal energy of radiation field [keV/µm]

    Max

    imum

    dos

    e ra

    te

    [µG

    y/h]

    _

    d=10 cm

    d=5 cm

    Figure 3.5. Maximum dose rate for resolving a high-LET event of 150 keV/µm in a multiple-event spectrum as a function of Dy of the radiation field. The limit is defined as 3 standard deviations (k=3) above the average energy imparted. Results for detectors with 10 cm diameter (––) and 5 cm diameter (----) are shown. A single high-LET event of 150 keV/µm or higher can be resolved below the lines.

  • The variance-covariance method used for beam quality characterization J. Lillhök

    27

    Figure 3.6 shows (3.10) applied to multiple-event spectra measured in a 97 MeV neutron beam at the iThemba-lab (section 4.1). With k=4 the estimated lowest single-events that can be resolved at these dose rates are 174, 576 and 863 keV/µm, indicated by lines in the figure.

    1

    10

    100

    1000

    1 10 100 1000 10000

    Energy deposited per mean chord-length [keV/µm]

    Num

    ber

    of c

    ount

    s 7 µGy/h80 keV/µm

    50 µGy/h90 keV/µm 86 µGy/h

    98 keV/µm

    Figure 3.6. Measured multiple-event spectra in a 97 MeV neutron beam and estimated lowest LET-value (vertical lines) a single-event can have to be resolved in the spectra according to (3.10). The ordinate shows the number of measurements in energy bins between 0.5 and 1 keV/µm width (constant for each spectrum). The abscissa shows the energy imparted during 100 ms divided by the mean chord-length.

    3.3. Separation of dose components Separation of single high-LET events with the variance method is typically possible only above a hundred keV/µm (Figure 3.2), which is not sufficient to separate photon and neutron absorbed dose fractions. The components are related via their Dy -values in

    (2.24), which has as the consequence that the neutron dose-mean lineal energy (nDy , )

    and the neutron quality factor ( nDQ , ) are very sensitive on the neutron absorbed dose

    fraction dn, while the neutron dose-equivalent (Hn) is almost independent of dn.

  • The variance-covariance method used for beam quality characterization J. Lillhök

    28

    { } ( )

    { } { }n

    nDnDn

    nDn

    nDD

    nn

    DnD

    nnDn

    n

    nD

    nDnD

    nD

    d

    dyyd

    yd

    yy

    dd

    ydy

    dyd

    d

    y

    yy

    y

    ∆−≈>>≈∆

    −==

    =∆

    −−∂∂==∆

    ∂∂

    =∆

    γγ

    γ

    ,,,

    ,,

    ,

    ,

    ,

    ,,

    ,

    )24.2(

    11)24.2(

    1

    (3.11)

    { } ( ) ( )

    { } ( ) { } nn

    DnDnn

    nDD

    nD

    nn

    nD

    nDn

    n

    nD

    nDnD

    nD

    d

    dyyd

    d

    yy

    yba

    b

    dd

    yb

    ybad

    d

    yba

    QQ

    Q

    ∆−≈>>≈∆

    −+

    ==

    =∆∂∂

    +=∆

    ∂+∂

    ==∆

    γγ

    ,,,,

    ,

    ,

    ,

    ,

    ,,

    ,

    )11.3(

    11)5.3(

    (3.12)

    { } ( )

    { } 0)12.3(1

    11)1.3(

    ,

    ,

    ,,

    ,

    =

    ∆−

    ∆≈≈∆

    ∂∂

    +=

    =∆

    ∂∂

    +=∆∂

    ∂==

    n

    n

    n

    nn

    nnD

    nD

    n

    nn

    nDnnD

    nn

    n

    nDn

    nn

    n

    d

    d

    d

    dd

    dQ

    Q

    d

    dd

    QDdQD

    Hd

    d

    QDd

    HH

    H

    (3.13)

    If only the neutron dose-equivalent (Hn) is of interrest it is not very crucial that dn is determined with high accuracy. It is however important for example in comparisons that also the other quantities are determined with reasonable precision. A known method in mixed-field dosimetry is to use two detectors with different sensitivities for the different radiation components [Attix 1987, Bewley 1989]. Consider two low-pressure proportional counters, one with tissue-equivalent A-150 plastic walls (TEPC) and one with graphite walls (GPC) [Paper II]. The absorbed doses in the two detectors can be written

    γγ

    γ

    DrDrD

    DDD

    nnGPC

    nTEPC

    +=+=

    (3.14)

    where rn and rγ are the neutron and photon absorbed dose ratios between the GPC and the TEPC. The relative neutron absorbed dose fraction (dn) in the TEPC is given by

    ( ) ( )nTEPC

    GPC

    nTEPC

    GPCTEPC

    TEPC

    nn rr

    D

    Dr

    rrD

    DDr

    D

    Dd

    −=

    −−

    ==γ

    γ

    γ

    γ (3.15)

  • The variance-covariance method used for beam quality characterization J. Lillhök

    29

    0

    0.5

    1

    1.5

    0.0001 0.001 0.01 0.1 1 10 100

    Energy [MeV]

    DG

    PC/D

    TE

    PC

    Figure 3.7. Ratio between dose to the GPC (graphite) and the TEPC detector as function of energy for neutrons (●) and photons (□).

    Figure 3.7 shows rn and rγ as a function of photon and neutron energy. The results were simulated with the Monte Carlo code MCNP 4C as described in section 3.4. In photon fields the two detector responses are equal within ±20%. For neutrons rn increases above 5 MeV due to an increasing neutron non-elastic cross-section in graphite. At most workplaces the neutron fluence above 5 MeV is relatively low while the thermal fluence can be considerable and rn is typically 0.05. The two quantities nDy , and nDQ , are very sensitive on the uncertainty in rγ but not on

    the uncertainty in rn, which is seen directly from differentiation of (3.15) and using that dn~0.1.

    ( ) ( ) { }

    ( ) ( )( )

    ( ) ( )( )( )n

    n

    n

    nn

    nn

    nn

    nn

    n

    rr

    rr

    rr

    rrd

    rr

    dr

    r

    d

    dd

    d

    −∆+∆

    ∆+∆

    =

    ==

    ∂∂

    +∆

    ∂∂

    =∆

    γ

    γ

    γ

    γ

    γγ

    2/122

    2/1

    2

    222

    2/1

    2

    2

    2

    2

    8111

    )12.3(1

    (3.16)

  • The variance-covariance method used for beam quality characterization J. Lillhök

    30

    Typical values are rγ=1 and rn=0.05 (Figure 3.7) that with ∆rγ=0.05 results in a 50% uncertainty in dn, while ∆rn=0.05 results in only 5% uncertainty in dn. The same numbers apply for nDy , and nDQ , as seen from (3.11) and (3.12).

    A second method to separate photon and neutron absorbed dose fractions was presented in Paper II. Specific Dy -values were assigned to the two components, which is

    motivated by that the neutron Dy -value is essentially determined by the fast neutrons for many distribution. Both the absorbed dose to detector gas per unit neutron fluence δn(E) and the energy dependent neutron dose-mean lineal energy )(n, EyD decreases strongly with decreasing energy (section 4.1). The total neutron nDy , given by

    ∫∫=

    dEE

    dEEyEy

    nEn

    nDnEn

    nD)(

    )()(

    ,

    ,,

    , δφ

    δφ (3.17)

    decreases even more sharply. Here Φn,E is the neutron fluence differential in energy. Figure 3.9 shows )(Enδ , )(n, EyD and the product )()( n, EyE Dnδ as a function of neutron energy. The functions have been normalized to unity at 2 MeV.

    0.001

    0.01

    0.1

    1

    10

    0.001 0.01 0.1 1 10

    Neutron energy [MeV]

    Rel

    ativ

    e va

    lues

    _

    Figure 3.9. The neutron energy dependence of the dose-average lineal energy

    )(n, EyD (---), the dose to detector gas δn(E)(– – –), and the product δn(E) )(n, EyD (–––). The functions are normalized to unity at 2 MeV.

  • The variance-covariance method used for beam quality characterization J. Lillhök

    31

    The product δn(E) )(n, EyD decreases strongly below 1 MeV and the low energy neutrons contribution to nDy , in (3.17) is small for most fluence distributions Φn,E. The neutron dose fraction can according to (2.24) be determined from the measured Dy -value

    ( )( )γ

    γ

    ,,

    ,

    DnD

    DDn yy

    yyd

    −−

    = (3.18)

    with typical values γ,Dy = 3 – 1.5 keV/µm and nDy , = 60 – 80 keV/µm (Figure 4.3).

    At third (unpublished) method to separate low- and high-LET absorbed dose components is related to the observation that the peak in a multiple-event spectrum is mainly due to a large number of low-LET interactions if the dose rate is not too high and the high-LET component is not totally dominating. According to the Central limit theorem a sum of independent equally distributed variables will approach a normal distribution for a sufficient number of variables [e.g. Rice 1995]. The required number of variables depends on the symmetry of the original distribution. Energy depositions from single events are distributed on a logarithmic rather than on a linear scale and the sum distribution more likely to be well represented by a lognormal distribution. This is exemplified in Figure 3.10 with both a normal and a lognormal distribution fitted to a measured multiple-event distribution in a 137Cs field of 6 µSv/h.

    0

    0.2

    0.4

    0.6

    0.8

    1

    1.2

    0 5 10 15

    Energy per mean chord-length [keV/µm]

    Num

    ber

    of c

    ount

    s [

    a.u.

    ]

    Figure 3.10. Multiple-event spectrum measured in a 6 µSv/h Cs-137 photon field (○) with fitted normal (---) and lognormal (––) distributions. The ordinate shows the number of measurements in 0.14 keV/µm bins. The abscissa shows the energy imparted during 100 ms divided by the mean chord-length.

  • The variance-covariance method used for beam quality characterization J. Lillhök

    32

    A lognormal distribution fitted to the peak in a multiple-event distribution is used to estimate the absorbed dose fraction and the Dy -value of the low-LET component (Figure 3.11). Data on the high-LET (right) side of the peak, where high-LET events give a significant contribution, should not be included in the fit. The choice of what data to include is given by the best correlation between the fit and the measurements. A Dy -value for the peak that is higher than 1.5 – 3 keV/µm (Figure 4.3) indicates that the width of the fitted distribution is too large and includes a non-negligible number of high-LET events.

    20

    40

    60

    80

    100

    120

    140

    160

    0 50 100 150 200 250 300 350 400

    Energy per mean chord-lenght [keV/µm]

    Num

    ber

    of c

    ount

    s [a

    .u.]

    Figure 3.11. Separation of low-LET component using a lognormal distribution fitted to a multiple-event distribution. The absorbed dose rate is 47 µGy/h and the dose-mean lineal energy 9.2 keV/µm. The ordinate shows the number of counts in 1.4 keV/µm bins. The abscissa shows the energy imparted during 100 ms divided by the mean chord-length.

  • The variance-covariance method used for beam quality characterization J. Lillhök

    33

    3.4. Monte Carlo simulations The Monte Carlo code MCNP version 4C [Briesmeister 2000] was used to simulate the absorbed dose responses of the tissue-equivalent (TEPC) and the graphite walled (GPC) detectors in photon and neutron fields. Simulated energies were 20 keV to 1 GeV for photons and thermal to 20 MeV for neutrons. In the photon simulations the absorbed dose to the detector gas was determined by scoring the energy imparted to the sensitive volume by secondary electrons. The neutron simulations were based on the kerma approximation, i.e. the absorbed dose was calculated from the neutron fluence and kerma coefficients under the assumption of charged particle equilibrium (CPE) [e.g. Attix 1987]. The solid materials in the detector geometry (Figure 3.12) have the same atomic composition as the physical detector (Figure 3.1), and the gas volumes are propane-based tissue-equivalent gas. The exception is the sensitive volume of the detector that in the neutron simulations was given the same atomic composition as the wall but with the same density as the other gas volumes. In this way the kerma of the wall material was determined with reasonable scattering properties retained. The simulated absorbed doses might be 2 – 3 % too low due to stopping power differences between A-150 plastic and the tissue-equivalent gas [ICRU 2005, ICRU 1993b].

    Figure 3.12. Cross-section of the detector geometry used in the MCNP4C simulations.

  • The variance-covariance method used for beam quality characterization J. Lillhök

    34

    Neutron interaction in the low pressure TE-gas cavity were hence neglected, which is reasonable at high energies [Newhauser and Brede 1997, Rubach and Bichsel 1982, Pszona and Makarewicz 1982]. At lower energies the neutron interaction in the gas increases and the detector is no longer a good Bragg-Gray cavity. However, the differences in stopping power and kerma coefficients between the gas and A150 are of the order of a few per cent for the neutron secondaries 1H and 4He [Chadwick et al 1999, ICRU 2005] and not corrected for here. The neutron interactions in the gas are not very important for the separation of the low-LET component using the TEPC+GPC detector combination either. The neutron absorbed dose fraction dn is not very sensitive to the uncertainty in the ratio of the GPC and TEPC neutron absorbed doses rn (3.16). Monte Carlo methods were also used in Paper V for verification and extrapolation of measured data to very small objects. The task was approached in two steps. The total

    Dy -value of a radiation field can like in (2.24) be written in terms of the Dy -values and the absorbed dose fractions of its charged secondaries

    ...)( ,,,, HeDHeHDHDiDiD ydydydyddyyydy ++=== ∫ ∑ γγ (3.19) The dose-mean lineal energy for the whole distribution of each particle type (iDy , ) was

    calculated from the mass-stopping powers and energy dependent dose-mean lineal energies ( )(, Ey iD ).

    ∫∞

    Φ

    Φ=

    0

    ,

    0

    ,,

    iD,)(

    )()(

    dEES

    dEEyES

    yi

    iE

    iDi

    iE

    ρ

    ρ (3.20)

    The code SHIELD-HIT [Gudowska et al 2004, Geithner et al 2006] was used to generate the dose fractions (di) and the fluence of the charged secondaries (ΦE,i). Corrections for material differences between the geometries in the simulation and the experiments were in Paper V introduced using Bragg-Gray cavity theory, i.e. with particle fluences and mass-stopping power ratios. The dose-mean lineal energy values

    )(, Ey iD in (3.20) were taken from Kyllönen et al [2003] and Nikjoo et al [2001 and

    2002]. These data were obtained from ion-track simulations in water vapor normalized to unit density using the Monte Carlo code PITS [Wilson and Nikjoo 1999] coupled with the electron transport code KURBUC [Uehara et al 1993]. Figure 3.13a shows the dose distributions in a volume with 100 nm diameter for 1H, 4He, 7Li and 12C ions of energy 5 MeV/u. Figure 3.13b shows the dose distributions in a volume with 100 nm for 1H in the energy range 0.3–30MeV. The frequency (Fy ) and dose average (Dy ) lineal energies are presented inside the figure. A selected data set from Kyllönen et al [2003],

    Dy - and Fy -values for cylinders with equal diameter and length, are summarized in Appendix.

  • The variance-covariance method used for beam quality characterization J. Lillhök

    35

    0

    0.5

    1

    1.5

    2

    2.5

    1 10 100 1000

    Lineal energy [ev/nm]

    yd(y

    )12C yF=38 eV/nm

    yD=250 eV/nm

    7Li yF=24 eV/nm

    yD=68 eV/nm4He yF=18 eV/nm

    yD=35 eV/nm

    1H yF=7 eV/nm

    yD=12 eV/nm

    Figure 3.13a. Simulated nanodosimetric distributions for ions with energy 5 MeV/u in a cylindrical volume with diameter and length 100 nm. 1H(×), 4He(●), 7Li(□) and 12C(◊).

    0

    0.2

    0.4

    0.6

    0.8

    1

    1.2

    1.4

    1.6

    1.8

    0.1 1 10 100 1000

    Lineal energy [ev/nm]

    yd(y

    )

    0.3 MeV 1H

    yF=52 eV/nm

    yD=77 eV/nm

    1 MeV 1H

    yF=23 eV/nm

    yD=35 eV/nm

    10MeV 1H

    yF=4 eV/nm

    yD=8 eV/nm 30MeV 1H

    yF=2eV/nm

    yD=7eV/nm

    Figure 3.13b. Simulated nanodosimetric distributions for protons in a cylindrical volume with diameter and length 100 nm. Protons 0.3MeV (×), 1MeV(●), 10MeV(□) and 30MeV(◊).

  • The variance-covariance method used for beam quality characterization J. Lillhök

    36

  • The variance-covariance method used for beam quality characterization J. Lillhök

    37

    4. RESULTS Section 4.1 presents the absorbed dose, dose-mean lineal energy, quality factor and dose-equivalent responses in neutron and photon fields for the TEPC detector described in section 3.1. Results from monoenergetic and quasi-monoenergetic neutron beams have been published in Paper I, Kyllönen and Mayer [2002] and Kyllönen et al [2002]. New results from defolding the peak responses, unpublished experimental data from iThemba-lab in Capetown, and unpublished Monte Carlo simulations are presented. In section 4.2 the results of using the variance-covariance method for radiation protection measurements in the cosmic radiation field and in the nuclear industry is evaluated. The performance of the different algorithms for covariance correction in pulsed fields is evaluated in section 4.3 and the different methods to separate low- and high-LET dose in section 4.4. In section 4.5 the instrument performance at extremely low levels of radiation is investigated, and in particular the influence on the result from a small radioactive content in the A-150 plastic wall discussed. Section 4.6 is devoted to an uncertainty analysis and the final section 4.7 is concerned with the variance-covariance method used for nanodosimetric investigations and beam quality characterization.

    4.1. Instrument response The TEPC was in 2003 exposed in a 7Li(p,n)7Be neutron beam with 97 MeV peak energy at iThemba-lab outside Capetown, South Africa (unpublished). The neutron fluence was monitored with time-of-flight techniques and the characteristic low-energy (tail) contribution could be subtracted by using an additional measurement at 16º angle to the primary beam. Often the tail is not measured and the peak value δn(Epeak) must be defolded from the detector response and the fluence distribution.

    ∑−

    =

    ∆Φ+∆Φ=1

    1,, )()()()(

    N

    iiiniEnpeakpeaknpeakEn EEEEEED δδ (4.1)

    Here δn(E) is the energy dependent conversion coefficient from neutron fluence to absorbed dose to detector gas, D is the measured total absorbed dose, Φn,E is the neutron fluence distribution differential in energy, ∆Ei the energy bins and N the total number of bins in the spectrum. The dose-mean lineal energy of the peak )(, peaknD Ey is

    determined analogously from

    =

    =

    ∆Φ

    ∆Φ+∆Φ=

    N

    iiiniEn

    N

    iiiiniEnpeakpeakpeaknpeakEn

    EEE

    EEyEEEEyEEy

    1,

    1

    1nD,,nD,,

    D

    )()(

    )()()()()()(

    δ

    δδ (4.2)

    where )(, Ey nD is the energy dependent dose-mean lineal energy function for the

    detector.

  • The variance-covariance method used for beam quality characterization J. Lillhök

    38

    Table 4.1 shows the results from iThemba and defolded results from measurements at UCL (Université catholique de Louvain) and TSL (The Svedberg Laboratory, Uppsala) [Paper I, Kyllönen et al 2002]. The neutron distributions at TSL were extrapolated to thermal energy using a constant distribution below the lowest calculated energy bin [Prokofiev 2001]. For the highest beam energy of 172 MeV there were no calculations available and the distribution was assumed to be the same as at 94 MeV, but shifted to higher energies.

    Table 4.1. Defolded results for the peak in quasi-monoenergetic 7Li(p,n)7Be neutron beams (unpublished).

    Laboratory

    Epeak (MeV)

    δn(Epeak) (pGy.cm2)

    )(, peaknD Ey

    (keV/µm) UCL 60 18.6 154

    TSL 68 65.7 110

    TSL 94 48.9 111

    iThemba 97 92 106

    TSL 172 9.1 40

    Figure 4.1 shows the experimental δn(E) from Paper I, δn(Epeak) from Table 4.1 and MCNP simulations of the absorbed dose to TEPC gas (section 3.4), conversion coefficients to ambient absorbed dose D*(10) [Ferrari and Pelliccioni 1998, Leuthold et al 1992] and A-150 kerma [Chadwick et al 1999, Savitskaya and Sannikov 1995]. D*(10) is defined as the absorbed dose at 10 mm depth in the ICRU-sphere in an expanded and aligned field [ICRU 1993]. The D*(10) coefficients from Ferrari and Pelliccioni [1998] were derived as the ratio of the reported H*(10)-values and the effective quality factor ffeQ at 10 mm depth in the ICRU-sphere. The data of Ferrari and

    Pelliccioni do not assume charge-particle equilibrium and the build-up layer consists only of the 10 mm ICRU-tissue in the definition of the quantities. The difference between the TEPC (MCNP simulations) and the ambient absorbed dose D*(10) at low energies is due to an increasing contribution from thermalized neutron interactions 14N(n,p)14C and 1H(n,γ)2H in the larger mass of the ICRU-sphere compared to the TEPC. Above 20 MeV the secondary proton ranges become comparable with the TEPC wall thickness and the condition of charged-particle equilibrium (CPE) fails. The TEPC absorbed dose is not expected to be a good measure of A150 kerma, but rather the D*(10) without CPE reported by Ferrari and Pelliccioni. The best estimate of the absorbed dose per unit neutron fluence δn(E) is hence using the experimental data between 71 keV and 19 MeV, MCNP simulations below 71 keV (normalized to the measurements) and the conversion coefficients to D*(10) above 19 MeV. The two experimental data points at 60 MeV (UCL) and 172 MeV (TSL) are significantly lower than expected. For the 172 MeV beam the energy distribution is uncertain and the results could be explained by the presence of low-energy neutrons with lower absorbed

  • The variance-covariance method used for beam quality characterization J. Lillhök

    39

    dose conversion coefficients. For the 60 MeV beam the fluence distribution was however carefully measured and the low absorbed dose conversion coefficient is possibly related to an error in the absolute neutron fluence. These data points are not included in δn(E).

    1E-03

    1E-01

    1E+01

    1E+03

    1E-08 1E-06 1E-04 1E-02 1E+00 1E+02 1E+04

    Neutron energy [MeV]

    Con

    vers

    ion

    coef

    ficie

    nts

    [fG

    y.m

    2]_

    Figure 4.1. Conversion coefficients from neutron fluence to absorbed dose and kerma. Measured TEPC (♦), MCNP simulated TEPC (–––), ambient absorbed dose D*(10) (+) and A-150 kerma (- - -). See text for details.

    The TEPC photon absorbed dose response was measured in gamma and x-ray beams between 15 keV and 7 MeV at SSI and at PTB (unpublished). Figure 4.2 shows the measured and MCNP simulated conversion coefficients from photon fluence to absorbed dose to detector gas. Included are coefficients for ambient dose-equivalent [ICRP 1996, Pelliccioni 2000] and tissue-kerma [ICRU 1992]. At low photon energies the absorbed dose and ambient dose-equivalent are lower than kerma due to attenuation. The dose to detector gas is lower than the ambient dose-equivalent due to higher attenuation in the detector. For example, using exponential attenuation and mass-attenuation coefficients [ICRU 1992] show that 0.2 cm Al + 0.5 cm A-150 plastic attenuates 36% more than 1 cm ICRU tissue, which is in good agreement with the 30% difference between the detector dose and ambient dose-equivalent in Figure 4.2. At high photon energies both the measured and the simulated TEPC absorbed dose is lower than kerma due to lack of build-up and charged-particle equilibrium. The ambient dose-equivalent conversion coefficients from ICRP74 are based on a CPE assumption and follow the kerma coefficient closely. Peliccioni [2000] does not presume external

  • The variance-covariance method used for beam quality characterization J. Lillhök

    40

    CPE in his reported absorbed dose at 1 cm depth in the ICRU sphere and these values are consequently lower than the kerma coefficients. The TEPC absorbed dose is higher than H*(10) from Pelliccioni due a better build-up and the TEPC response should be reasonably well estimated by the MCNP simulated results for the whole energy range. The H*(10) response of the TEPC is within ±20% for photon energies usually encountered.

    0.001

    0.01

    0.1

    1

    10

    100

    0.01 0.1 1 10 100

    Photon energy [MeV]

    Con

    vers

    ion

    coef

    ficie

    nts

    [fG

    y.m

    2 an

    d fS

    v.m

    2]_

    Figure 4.2. Photon fluence-to-dose, kerma and dose-equivalent conversion coefficients. Experimental dose to TEPC detector gas (♦), MCNP simulated dose to TEPC detector gas (—), ICRP74 ambient dose-equivalent (□), ambient dose-equivalent from Pelliccioni [2000] (---) and tissue kerma from ICRU46 (−−−).

    The dose-mean lineal energy of this detector was not simulated and values at energies not covered by the experiments must be estimated by other means. A thermal neutron

    thDy , -value of 46 keV/µm is calculated from the results of the non-moderated and

    moderated 252Cf -field reported in Paper I using (2.24). At high neutron energies the simulations by Ferrari and Pelliccioni [1998] show relatively small (10%) variations in the effective quality factor ffeQ at 10 mm depth in the ICRU-sphere. It is assumed that

    the relative variation in Dy above 100 MeV is the same as in ffeQ even though )(ff LQe

    is not a linear function of L (or y). At intermediate neutron energies the dominating dose contribution comes from elastically scattered hydrogen recoils. Due to their short ranges most of these secondary protons deposit all their energy in the detector volume and their dose-mean lineal energy ( elpDy _, ) is approximately proportional to the neutron energy. The MCNP

    simulations (Figure 4.1) show no gamma contribution generated by neutrons above 50

  • The variance-covariance method used for beam quality characterization J. Lillhök

    41

    keV and Dy is determined by the protons only. The proton elpDy _, is therefore scaled in

    proportion to neutron energy from the experimental Dy at the lowest neutron energy 71 keV. The MCNP simulations show also that for this particular detector the gamma contribution from 1H(n,γ)2H is about 60% of the total thermal neutron absorbed dose. It is assumed that the gamma dose at intermediate neutron energies is due to thermalized neutrons in the detector and then the MCNP simulated gamma dose gives directly the thermalized neutron dose fraction (dth) at different intermediate neutron energies. An estimation of the intermediate neutron Dy -value is then given by the contributions from elastic protons and thermalized neutrons according to

    thDthelpDththDthelpDelpD ydydydydy ,_,,_,_ )1( +−=+= (4.3) Figure 4.3 shows the estimated dose-mean lineal energy function )(, Ey nD for the TEPC

    together with the measured data from the monoenergetic beams (Paper I, Kyllönen and Mayer 2002) and the defolded data from Table 4.1. The data from Paper I has been corrected for the W/e-difference between a pure neutron field (31 J/C) and a mixed radiation field (28 J/C). Included are also results reported by other groups [Rodgers and Gross 1981, Wissmann 2002, Nunomiya et al 2002, Nguyen et al 1980, Menzel and Schuhmacher 1981]. The data represent measured and simulated values for mean chord lengths between 1.3 and 2.7 µm, and all results are for spherical objects except for the variance-covariance TEPC. Included in Figure 4.3 are also experimental Dy -values from gamma and x-ray beams. The nuclei used were 238Pu, 241Am, 137Cs-137 and 60Co-60 with average gamma energies 15 keV, 60 keV, 662 keV and 1250 keV. The x-ray HVL-energies were between 32 keV and 242 keV.

  • The variance-covariance method used for beam quality characterization J. Lillhök

    42

    1

    10

    100

    1000

    0.0001 0.001 0.01 0.1 1 10 100 1000

    Energy [MeV]

    Dos

    e av

    erag

    e lin

    eal e

    nerg

    y [

    keV

    /µm

    ]_

    Photons

    Neutrons

    Figure 4.3. Experimental Dy for neutrons (■) and photons (□) for the variance-covariance instrument. Also shown is data from the literature [Rodgers and Gross 1981, Wissmann 2002, Nunomiya et al 2002, Nguyen et al 1980 and Menzel and Schuhmacher 1981]. The solid line represents the derived energy dependence of the neutron Dy -values for the variance-covariance instrument.

    The defolded neutron Dy -values at 60 MeV and 172 MeV value (154 keV/µm and 40 keV/µm) deviate from the expected trend as they did also for the absorbed dose (Figure 4.1). As already mentioned the 172 MeV results could be explained by the presence of low-energy neutrons with lower absorbed dose conversion coefficients but also lower

    Dy -values. For the 60 MeV beam there could be some problems with the absolute

    neutron fluence, but since the Dy -value depends only on the fluence energy distribution the value 154 keV/µm may be fairly reliable. Partial A-150 kerma coefficients from Meulders et al [2000] show that the relative kerma contribution from secondary alpha-particles peaks around 30 MeV, which could result also in a maximum in the neutron

    Dy -value due to the high-LET of the alpha-particles. This would imply that there is a

    Dy -maximum between 19 MeV and 60 MeV, but the uncertainties are large and a constant value is assumed at energies in between. The measured neutron Dy -values are 20–30% higher than reported by the other groups,

    and the photon Dy -values are 20% higher than reported in ICRU36 [ICRU 1984]. One

    source of uncertainty is that the mean chord-length l is determined directly from the physical detector size and the gas pressure (section 3.1). Another is the variance contribution from the chord-length distribution, which however should be small [Kellerer 1975]. This difference require further investigations to be fully understood.

  • The variance-covariance method used for beam quality characterization J. Lillhök

    43

    The neutron energy dependence of the linear quality factor Qlin(y) in (3.5) is given directly using )(, Ey nD . The combined quality factor Qcomb(y) in (3.6) requires however

    that the high-LET events are considered. A correction is obtained from single-event measurements [Wissmann 2002] up to 19 MeV. At higher energies Dy was assumed not to vary much (Figure 4.3) and the correction obtained at 19 MeV is applied. Figure 4.4 shows the result together with the effective quality factor ffeQ at 10 mm depth in the

    ICRU-sphere. The effective quality factor was derived from H*(10) and D*(10) conversion coefficients up to 20 MeV from ICRP [1996] and Leuthold et al [1992], and

    ffeQ from Ferrari and Pelliccioni [1998] above 20 MeV.

    1

    10

    100

    0.00000001 0.000001 0.0001 0.01 1 100 10000

    Neutron energy [MeV]

    Qua

    lity

    fact

    or

    [Sv/

    Gy]

    Figure 4.4. Quality factors as a function of neutron energy. The effective quality factor ffeQ at 10 mm depth in the ICRU-sphere (▬); Q(y) =0.52+0.28y (– –); Q(y) =0.73+0.17y (––); Q(y) = 0.88+0.09y (---) and the Qcomb(y) relation that corrects for single-events above 150 keV/µm (○).

    At high neutron energies Qlin(y) give very high values on the average quality factor due to the presence of high-LET secondaries (Figure 3.3). In the intermediate energy region the recoil protons are essentially stoppers in the detector cavity and their LET and quality factor are underestimated [Schuhmacher et al 1990, 1995]. In the thermal region the lower effective quality factor ffeQ is due to a larger photon contribution (low-LET)

    from thermalized neutrons in the large mass of the ICRU-sphere.

  • The variance-covariance method used for beam quality characterization J. Lillhök

    44

    Both the absorbed dose and quality factor responses are low in the intermediate energy region (Figures 4.1 and 4.4), and the effect is even more pronounced for the ambient dose-equivalent (Figure 4.5). The conversion coefficients from fluence to H*(10) are from ICRP74 below 20 MeV and from Ferrari and Pelliccioni [1998] for higher energies.

    0.01

    0.1

    1

    10

    100

    1000

    1E-08 0.000001 0.0001 0.01 1 100 10000

    Neutron energy [MeV]

    H*

    conv

    ersi

    on c

    oeff

    icie

    nts

    [fS

    v.m

    2]_

    Figure 4.5. Fluence-to-dose-equivalent conversion coefficients as a function of neutron energy. The H*(10)/Φ-values for the ICRU-sphere are from ICRP74 below 20 MeV and from Ferrari and Pelliccioni [1997] at higher energies (▬). The other lines for Q(y) = 0.52+0.28y (– –); Q(y) = 0.73+0.17y (––); Q(y) = 0.88+0.09y (---) and Qcomb(y) in (3.6) (○).

    The combined quality factor relation Qcomb(y) in (3.6) has improved the Q- and H*(10)- responses at high neutron energies compared to the linear relations Qlin(y) in (3.5). The large underestimation at intermediate energies however remains. It is a combination of scattering differences between the detector and the ICRU-sphere leading to a lower absorbed dose (Figure 4.1), and of an underestimation of the proton LET leading to a lower Q-value (Figure 4.4). The latter could be improved by reducing the gas pressure but the first is difficult since it is related to the physical size of the detector.

  • The variance-covariance method used for beam quality characterization J. Lillhök

    45

    4.2. Dose equivalent measurements One of the most challenging applications for radiation protection dosimetry is aircrew exposure to cosmic radiation. The energetic cosmic radiation produces a wide range of secondary particles in the atmosphere with energies extending into the GeV-region. Contributions to the ambient dose-equivalent in the atmosphere at high latitudes are 50–60% from neutrons, 20–30% from electrons, 10% from protons and 10% from photons and muons [Roesler et al 2002, Ferrari et al 2002, EURADOS 2004]. The LET distribution in tissue-like media consists of both a low- and high-LET component, where the latter is mainly due to neutron secondaries such as alpha-particles, carbon, nitrogen and oxygen recoils. Instrument calibration and response characterization are consequently quite intricate, which can result in significant differences in the determined H*(10) [Paper IV]. The variance-covariance instrument was used in two research contracts for cosmic radiation dosimetry supported by the European Commission [European Commission 1995, 2000]. The results show that the method gives results in good agreement (±10%) with other active and passive instruments and calculations using Qlin(y) from high-energy neutron calibration, or Qcomb(y) [Paper III and IV, Bottollier-Depois et al 2004, EURADOS 2004]. A systematic difference is that combQ in (3.6) is 10% lower than

    )(60, yQ ICRPD in (3.4), which is expected from the underestimation in the proton region

    20–150 keV/µm (Figure 3.3). A corresponding underestimation would also be expected in the dose-equivalent, which was also observed in Paper III. During the devoted in-flight comparison in Paper IV such an effect related to the quality factor was hidden by other systematic differences related to calibration methods used by different groups. Calculations of the route doses are also extremely valuable as a complement in comparisons and for aircrew dose assessment. A number of codes have been developed for this application where one of the most well-established is EPCARD (European Program Package for the Calculation of Aviation Route Doses) [Schraube et al. 2000, Schraube et al. 2002]. Figure 4.6 shows the dose-equivalent measured with the variance-covariance instrument on flights between 1998 and 2003 as a function of the EPCARD calculated value. The average ratio between the measured and the calculated value is 0.99 ± 0.13 (1s), but despite this very good agreement there are some significant differences worth noting. The ratio varies with time from 0.80 ± 0.05 in 1998 (low solar activity) to 1.05 ± 0.07 in 2002/2003 (high solar activity). The same difference of 20% between low and high solar activity is also seen in published data compilations [EURADOS 2004]. Variations on a daily basis is also expected since EPCARD uses a monthly average of the signal from the ground-based neutron monitor CLIMAX in Colorado, and is therefore not sensitive to short-term variations in the cosmic radiation intensity. In Paper IV it was also noted that the measured altitude variation was stronger than calculated with EPCARD. The TEPC detectors measured an increase of 80% – 90% compared to 40 – 50 % for EPCARD in the high-LET (neutron) component of H* between flight levels FL400 (~12.2 km) and FL320 (~9.8 km).

  • The variance-covariance method used for beam quality characterization J. Lillhök

    46

    1

    10

    100

    1 10 100

    H*EPCARD [µSv]

    H* m

    Sv]

    Figure 4.6. Measured dose-equivalent as a function of the calculated value using EPCARD. Flights in 1998 from Paper III (□); flights in 2002 from Bottollier-Depois et al 2004 (×); flights between 2002 and 2003 (unpublished) (○) and in 2003 from Paper IV (∆).

    Another high-energy comparison was done at the CERF facility, available at CERN (European Organization for Nuclear Research) until recently. The concrete moderated high-energy neutron field resembled the atmospheric neutron spectrum and was used for response studies, comparisons and calibrations of dose meters and instruments for aircrew dosimetry [Höfert and Stevenson 1994, Mitaroff and Silari 2002]. The CERF results obtained with the variance-covariance TEPC were in good agreement with other instruments and simulations [Paper I, Mayer 2003, Mayer et al 2004]. The dose-equivalent using a linear relation Q(y)=0.88+0.09y agreed with a single-event TEPC system (HANDI) within ± 5 %, which is well within the uncertainties. The measured

    Dy -value of 37 ± 3 keV/µm is also in good agreement with the value 33 keV/µm derived from a single-event spectra measured by Höfert et al [1994]. The good response for the neutron dose-equivalent at high and thermal neutron energies is expected to get poorer with an increasing intermediate neutron contribution. Within the EVIDOS project (Evaluation of individual dosimetry in mixed neutron and photon radiation fields) a large number of measurements were made in simulated and real workplace fields [Paper II, d’Errico et al 2005, Schuhmacher et al 2005]. The neutron reference values were obtained by Bonner sphere spectrometry and the photon reference values evaluated from ion chamber measurements and from the variance-covariance instrument using the TEPC+GPC method (section 3.3). Figure 4.7 shows the ratio of the

  • The variance-covariance method used for beam quality characterization J. Lillhök

    47

    TEPC neutron dose-equivalent (Hn) and the Bonne-sphere neutron dose-equivalent (H*n(10)) as a function of the intermediate neutron dose fraction for all the EVIDOS fields and confirms the dependence on the intermediate neutron component.

    0

    1

    2

    0.01 0.1 1

    Intermediate neutron contribution to H *(10)

    Hn/

    H* n

    (10)

    Figure 4.7. Ratio of measured and reference neutron ambient dose-equivalent as a function of the intermediate neutron contribution. Intermediate neutrons are here defined by the energy interval 0.5 eV to 0.1 MeV. The quality factor was calculated according to (3.5) with a=0.52 and b=0.28 µm/keV.

    It seems that H*n(10) is difficult to determine with better accuracy than 30–50% in unkown neutron radiation fields in the nuclear industry. But note that H*n(10) is usually only about half of the total ambient dose-equivalent, and the remaining photon component is determined correctly within 10–20% (Figure 4.2). The uncertainty (1s) in the total H*(10) in