1 29 th March 2007 C.P. Ward Pat Ward University of Cambridge Status of Far Detector Linearity Calibrations

Status of Far Detector Linearity Calibrations

Embed Size (px)

DESCRIPTION

Status of Far Detector Linearity Calibrations. Pat Ward University of Cambridge. History. Original 20-point Gain Curves had too few points in linear region Feb 06: 60-point GC used to make first set of FD linearity calibrations July 06: start of regular 40-point GC - PowerPoint PPT Presentation

Citation preview

Page 1: Status of Far Detector Linearity Calibrations

1 29th March 2007 C.P. Ward

Pat Ward

University of Cambridge

Status of Far Detector Linearity Calibrations

Page 2: Status of Far Detector Linearity Calibrations

2 29th March 2007 C.P. Ward

History

Original 20-point Gain Curves had too few points in linear region

Feb 06: 60-point GC used to make first set of FD linearity calibrations

July 06: start of regular 40-point GC Second set of calibrations made from July

06 GC (in September 06) This week: have added calibrations from

Aug 06 and Sep 06 GC to offline database

Page 3: Status of Far Detector Linearity Calibrations

3 29th March 2007 C.P. Ward

Constants in Offline Database

TimeStart InsertDate GainCurve # points

1970-01-01 2006-04-03 Feb 06 60

2006-07-07 2006-09-21 Jul 06 40

2006-08-04 2007-03-27 Aug 06 40

2006-09-01 2007-03-28 Sep 06 40

All have TimeEnd = 2038-01-01

Page 4: Status of Far Detector Linearity Calibrations

4 29th March 2007 C.P. Ward

Stability:Jul06 – Sep 06

88% channels good fits both times and adcmax>12000

Ratio of corrections at ADC=12000: mean = 1.003 sigma = 0.051

Bay 11 rack 4 tube 2; Hardware change??

Page 5: Status of Far Detector Linearity Calibrations

5 29th March 2007 C.P. Ward

Have asked Howie to process remaining data – not yet done

Code in cvs:

PulserCalibration/PulserGainFit

PulserCalibration/macros/write_calpulserfits.C Write-up: MINOS-doc-1886 I will have to pass this on to Ryan now –