2
How to select the correct RF amplifier Example estimation of the required power for a requested field strength of 10 V/m in the desired frequency range Cables, connector, filter, switches –1.5 dB 1) + 9 dB Antenna gain Field uniformity – 6 dB 3) 43 dBm = P –1 dB – Power 0 dBm Signal generators 3 m test distance E = √ˉˉˉˉˉ 30×P –––– d Directonal coupler – 0.3 dB 2) 1) Harmonics in the field must be below 6 dBc, filters are typically not required. 2) This formula does not cover influences from chamber or test setup. An additional margin may be useful. 3) Field uniformity needs up to 6 dB more amplifier power (for a single measurement point). The test distance between antenna and uniform field area has to include the distance to the antenna phase center. According to IEC/EN 61000-4-3 the amplifier must be operated below its 2 dB compression point. Due to amplifier variation the 1 dB compression point should be selected for the estimation. The uniform field area (UFA) IEC/EN 61000-4-3 field calibration 150 cm 150 cm Measurement points 80 cm D = 1 m to 3 m Uniform field area Anechoic chamber Uniform field area (UFA) criteria Size: 1.5 m × 1.5 m –0 dB/+6 dB variation over UFA plane More than 75 % of points have to be within tolerance Larger sizes and partial illumination allowed for larger EUT Smaller sizes allowed for smaller EUT (including cabling) Located 0.8 m above floor Distance from antenna tip: minimum 1 m, preferred 3 m Uniform area criteria versus beamwidths and distance Beamwidth: Θ = 2tan –1 [ ] W ––– 2D W ––– 2D Distance: D = W ––––––– 2tan(Θ/2) Window: W = 2Dtan [ ] Θ –– 2 28° 41° 74° 3 m 1 m 2 m Amplifier harmonics versus intercept point Relationship between the power level of the fundamental and harmonic frequencies. P Fundamental ~ P Generator P 2nd harmonics ~ (P Generator ) 2 P 3rd harmonics ~ (P Generator ) 3 Remark: The saturation power of an amplifier can be much higher than the output power at P 1 dB compression. Fundamental signal output level in dBm 3rd order product 3:1 Slope Intercept diagram +60 +40 +20 0 –20 –40 –60 –80 –100 –120 –140 –80 –60 –40 –20 0 20 40 60 dBm 2nd order product 2:1 slope 1:1 slope 2nd order intercept 3rd order intercept Intercept signal level Field strength % of nom. value Nominal value 18 V/m Fundamental 14 V/m 78 % 2nd harmonics 7.3 V/m 40 % 3rd harmonics 8.7 V/m 49 % Harmonics in the field must be below 6 dBc, filters are typically not required The formulas are only valid for the linear amplifier power range. In the nonlinear range, harmonics increase at a higher rate EUT monitoring capabilities CANoe™ CANalyzer™ Customers monitoring PC Video monitoring system TTL I/O converter Measurement or stimulus device LAN (TCP/IP) CAN, MOST, LIN, FlexRay™ Camera Customer-specific TTL Physical parameters (U, l, f, T...) EUT CON (DCOM) LAN (TCP/IP) USB/LAN LAN/GPIB/RS-232 R&S®EMC32-S Signal processing Signal conversion Signal source EMS radiated test setup ¸SMB100A RF and microwave signal generator ¸OSP open switch and control platform ¸EMC32 EMC measurement software ¸NRP-Zxx power sensors R&S®BBA150 broadband amplifier R&S®HL046E log-perodic antenna Formulas s VSWR r Reflection coefficient a R Return loss s (VSWR) V V max min r V V P refl in % a dB V V R in lg 20 dB to ratio conversion table 1 1.005 1.010 1.015 1.020 1.025 1.030 1.035 1.040 1.045 1.050 1.055 1.060 1.065 1.070 1.075 1.080 1.085 1.090 1.095 60 50 40 35 34 33 32 31 30 29 28 27 a R dB 0.002 0 0.005 0.045 0.024 0.022 0.012 0.015 0.010 0.027 0.020 0.017 0.007 0.029 0.041 0.038 0.036 0.031 0.034 0.043 s r 1.01 0.005 46.1 dB Amplitude ratio Power ratio 1.02 0.010 0.01 40.1 –100 dB 10 –5 10 –10 1.03 0.015 0.02 36.6 –50 dB 0.00316 0.00001 1.04 0.020 0.04 34.2 –40 dB 0.010 0.0001 1.05 0.024 0.06 32.3 –30 dB 0.032 0.001 1.06 0.029 0.08 30.7 –20 dB 0.100 0.010 1.07 0.034 0.11 29.4 –10 dB 0.316 0.100 1.08 0.038 0.15 28.3 –6 dB 0.501 0.251 1.09 0.043 0.19 27.3 –3 dB 0.708 0.501 1.10 0.048 0.23 26.4 –2 dB 0.794 0.631 1.11 0.052 0.27 25.6 –1 dB 0.891 0.794 1.12 0.057 0.32 24.9 0 dB 1.000 1.000 1.13 0.061 0.37 24.3 1 dB 1.122 1.259 1.14 0.065 0.43 23.7 2 dB 1.259 1.585 1.15 0.070 0.49 23.1 3 dB 1.413 2 ≈ 1.995 1.16 0.074 0.55 22.6 6 dB 2 ≈ 1.995 3.981 1.17 0.078 0.61 22.1 10 dB 3.162 10.000 1.18 0.083 0.68 21.7 20 dB 10.000 100.000 1.19 0.087 0.75 21.2 30 dB 31.623 1000.000 1.20 0.091 0.83 20.8 40 dB 100.000 10000.000 1.30 0.130 1.70 17.7 50 dB 316.228 100000.000 1.40 0.167 2.78 15.6 100 dB 10 5 10 10 1.50 0.200 4.00 14.0 P P Ratio dB 1 0 10 10 = Ratio P P dB = 10 10 1 0 log V V Ratio dB 1 0 20 10 = Ratio V V dB = 20 10 1 0 log Wavelength – frequency conversion 300 f m MHz λ 1.60 0.231 5.33 12.7 1.70 0.259 6.72 11.7 1.80 0.286 8.16 10.9 1.90 0.310 9.63 10.2 2.00 0.333 11.10 9.5 2.20 0.375 14.1 8.5 2.40 0.412 17.0 7.7 2.60 0.444 19.8 7.0 2.80 0.474 22.4 6.5 3.00 0.500 25.0 6.0 3.50 0.556 30.9 5.1 4.00 0.600 36.0 4.4 5.00 0.667 44.4 3.5 6.00 0.714 51.0 2.9 7.00 0.750 56.2 2.5 8.00 0.778 60.5 2.2 10.0 0.818 66.9 1.7 20.0 0.905 81.9 0.9 50.0 0.961 92.3 0.3 R R 1 r s 1 r 0.05a 10 1 s 0.05a 10 1 + = + = = + = R s1 r s1 1 r 0.05 a 10 + = = R R s1 a 20 lg s1 1 a 20 lg r RF fundamentals in commercial EMS testing www.rohde-schwarz.com/ad/EMS Power sensors and accessories Thermal power sensors Outstanding linearity and maximum accuracy Outstanding performance for reference applications Excellent impedance matching. Sophisticated connector concept Internal calibration test Frequency range from DC to 110 GHz Level range from –35 dBm to +20 dBm Excellent impedance matching Innovative connector design for improved ease of use Outstanding performance for reference applications and calibration labs Average power sensors Specially designed for EMC applications. 90 dB dynamic range thanks to innovative three-path concept. Power measurements independent of the modulation type Frequency range from 9 kHz to 6 GHz Level range from –67 dBm to +33 dBm Support of low frequencies down to 9 kHz Antennas for EMS applications Antenna type Frequency range Gain Permissible input power Half-power beamwidth (E plane) R&S®HL562E ultralog antenna 30 MHz to 6 GHz –8 dBi at 30 MHz 8.5 dB at > 200 MHz 225 W CW at 30 MHz 900 W CW at 80 MHz 150 W CW at 6 GHz 80 ° to 30 ° R&S®HL223 log-periodic antenna 200 MHz to 1.3 GHz ≥ 6 dBi 1500 W CW at 200 MHz 600 W CW at 1.3 GHz approx. 60 ° R&S®HL046E log-periodic antenna 80 MHz to 3 GHz typ. 8.0 dBi 2100 W CW at 80 MHz 900 W CW at 500 MHz 375 W CW at 3 GHz approx. 60 ° R&S®HL050E log-periodic antenna 750 MHz to 6 GHz typ. 8.5 dBi 100 W CW approx. 60 ° R&S®HF907 double-ridged horn antenna 800 MHz to 18 GHz 5 dBi to 14 dBi 300 W CW at 800 MHz 150 W CW at 18 GHz 100 ° to 40 °

RF fundamentals in commercial EMS testing · 2020. 1. 16. · 3607.1713.82 01.00 PDP 1 en About Rohde & Schwarz The Rohde & Schwarz electronics grou is a leading sulier of solutions

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: RF fundamentals in commercial EMS testing · 2020. 1. 16. · 3607.1713.82 01.00 PDP 1 en About Rohde & Schwarz The Rohde & Schwarz electronics grou is a leading sulier of solutions

How to select the correct RF amplifierExample estimation of the required power for a requested field strength of 10 V/m in the desired frequency range

Cables, connector,filter, switches

–1.5 dB 1) + 9 dB

Antenna gain Field uniformity

– 6 dB 3)43 dBm =

P–1 dB – Power

0 dBm

Signal generators 3 m test distance

E = √ˉˉˉˉˉ30×P––––

d

Directonal coupler

– 0.3 dB

2)

1) Harmonics in the field must be below 6 dBc, filters are typically not required.2) This formula does not cover influences from chamber or test setup. An additional margin may be useful.3) Field uniformity needs up to 6 dB more amplifier power (for a single measurement point).

❙ The test distance between antenna and uniform field area has to include the distance to the antenna phase center.

❙ According to IEC/EN 61000-4-3 the amplifier must be operated below its 2 dB compression point. Due to amplifier variation the 1 dB compression point should be selected for the estimation.

The uniform field area (UFA)IEC/EN 61000-4-3 field calibration

150 cm

150 cm

Measurementpoints 80 cm

D = 1 m to 3 m

Uniform field area

Anechoicchamber

Uniform field area (UFA) criteria ❙ Size: 1.5 m × 1.5 m ❙ –0 dB/+6 dB variation over UFA plane ❙ More than 75 % of points have to be within tolerance ❙ Larger sizes and partial illumination allowed for larger EUT ❙ Smaller sizes allowed for smaller EUT (including cabling) ❙ Located 0.8 m above floor ❙ Distance from antenna tip: minimum 1 m, preferred 3 m

Uniform area criteria versus beamwidths and distance

Beamwidth: Θ = 2tan–1 [ ]W–––2DW

–––2D

Distance: D =W

–––––––2tan(Θ/2)

Window: W = 2Dtan [ ]Θ––2

28° 41° 74°

3 m 1 m2 m

Amplifier harmonics versus intercept point

Relationship between the power level of the fundamental and harmonic frequencies.

PFundamental ~ PGenerator

P2nd harmonics ~ (PGenerator)2

P3rd harmonics ~ (PGenerator)3

Remark:The saturation power of an amplifier can be much higher than the output power at P1 dB compression.

Fundamental signal output level in dBm

3rd order product3:1 Slope

Intercept diagram

+60

+40

+20

0

–20

–40

–60

–80

–100

–120

–140

–80 –60 –40 –20 0 20 40 60

dB

m

2nd order product2:1 slope

1:1 slope

2nd order intercept

3rd order intercept

Intercept signal level

Field strength % of nom. value

Nominal value 18 V/m

Fundamental 14 V/m 78 %

2nd harmonics 7.3 V/m 40 %

3rd harmonics 8.7 V/m 49 %

❙ Harmonics in the field must be below 6 dBc, filters are typically not required ❙ The formulas are only valid for the linear amplifier power range. In the nonlinear range, harmonics increase at a higher rate

EUT monitoring capabilitiesEUT monitoring capabilities

CANoe™CANalyzer™

Customersmonitoring PC

Video monitoringsystem

TTL I/Oconverter

Measurement orstimulus device

LAN (TCP/IP)

CAN, MOST, LIN, FlexRay™

Camera

Customer-specific

TTL

Physical parameters (U, l, f, T...)

EUT

CON (DCOM)

LAN (TCP/IP)

USB/LAN

LAN/GPIB/RS-232

R&S®EMC32-S

Signal processing Signal conversion Signal source

EMS radiated test setup

¸SMB100ARF and microwave signal generator

¸OSPopen switch and control platform

¸EMC32EMC measurement software

¸NRP-Zxxpower sensors

R&S®BBA150 broadband amplifier

R&S®HL046E log-perodic antenna

Formulass VSWRr Reflection coefficientaR Return loss

s

(VSWR)

V

V

max

min

r

V

V

←←→→

Prefl in % a dB

V

V

R in

lg 20←←→→

dB to ratio conversion table

1

1.005

1.010

1.015

1.020

1.025

1.030

1.035

1.040

1.045

1.050

1.055

1.060

1.065

1.070

1.075

1.080

1.085

1.090

1.095

60

50

40

35

34

33

32

31

30

29

28

27

aR

dB

0.002

0 ∞

0.005

0.045

0.024

0.022

0.012

0.015

0.010

0.027

0.020

0.017

0.007

0.029

0.041

0.038

0.036

0.031

0.034

0.043

s r 1.01 0.005 46.1 dB Amplitude ratio Power ratio

1.02 0.010 0.01 40.1 –100 dB 10–5 10–10

1.03 0.015 0.02 36.6 –50 dB 0.00316 0.00001

1.04 0.020 0.04 34.2 –40 dB 0.010 0.0001

1.05 0.024 0.06 32.3 –30 dB 0.032 0.001

1.06 0.029 0.08 30.7 –20 dB 0.100 0.010

1.07 0.034 0.11 29.4 –10 dB 0.316 0.100

1.08 0.038 0.15 28.3 –6 dB 0.501 0.251

1.09 0.043 0.19 27.3 –3 dB 0.708 0.501

1.10 0.048 0.23 26.4 –2 dB 0.794 0.631

1.11 0.052 0.27 25.6 –1 dB 0.891 0.794

1.12 0.057 0.32 24.9 0 dB 1.000 1.000

1.13 0.061 0.37 24.3 1 dB 1.122 1.259

1.14 0.065 0.43 23.7 2 dB 1.259 1.585

1.15 0.070 0.49 23.1 3 dB 1.413 2 ≈ 1.995

1.16 0.074 0.55 22.6 6 dB 2 ≈ 1.995 3.981

1.17 0.078 0.61 22.1 10 dB 3.162 10.000

1.18 0.083 0.68 21.7 20 dB 10.000 100.000

1.19 0.087 0.75 21.2 30 dB 31.623 1000.000

1.20 0.091 0.83 20.8 40 dB 100.000 10000.000

1.30 0.130 1.70 17.7 50 dB 316.228 100000.000

1.40 0.167 2.78 15.6 100 dB 105 1010

1.50 0.200 4.00 14.0

PP

Ratio dB

1

0

1010=RatioPPdB = ⋅

10 10

1

0

log

VV

Ratio dB

1

0

2010=RatioVVdB = ⋅

20 10

1

0

log

Wavelength – frequency conversion

300fm

MHz

λ

1.60 0.231 5.33 12.7

1.70 0.259 6.72 11.7

1.80 0.286 8.16 10.9

1.90 0.310 9.63 10.2

2.00 0.333 11.10 9.5

2.20 0.375 14.1 8.5

2.40 0.412 17.0 7.7

2.60 0.444 19.8 7.0

2.80 0.474 22.4 6.5

3.00 0.500 25.0 6.0

3.50 0.556 30.9 5.1

4.00 0.600 36.0 4.4

5.00 0.667 44.4 3.5

6.00 0.714 51.0 2.9

7.00 0.750 56.2 2.5

8.00 0.778 60.5 2.2

10.0 0.818 66.9 1.7

20.0 0.905 81.9 0.9

50.0 0.961 92.3 0.3

R

R

1 r s

1 r

0.05a10 1

s0.05a

10 1

+=

+=

−=

+

=R

s 1r

s 11

r 0.05 a

10

+ = −

=

R

R

s 1a 20 lg

s 1

1a 20 lg

r

RF fundamentals in commercial EMS testing

www.rohde-schwarz.com/ad/EMS

Power sensors and accessoriesThermal power sensors ❙ Outstanding linearity and maximum accuracy ❙ Outstanding performance for reference applications

❙ Excellent impedance matching. ❙ Sophisticated connector concept ❙ Internal calibration test

❙ Frequency range from DC to 110 GHz ❙ Level range from –35 dBm to +20 dBm ❙ Excellent impedance matching ❙ Innovative connector design for improved ease of use

❙ Outstanding performance for reference applications and calibration labs

Average power sensors ❙ Specially designed for EMC applications. ❙ 90 dB dynamic range thanks to innovative three-path concept.

❙ Power measurements independent of the modulation type

❙ Frequency range from 9 kHz to 6 GHz ❙ Level range from –67 dBm to +33 dBm ❙ Support of low frequencies down to 9 kHz

Antennas for EMS applicationsAntenna type

Frequency range

Gain Permissible input power

Half-power beamwidth (E plane)

R&S®HL562E ultralog antenna

30 MHz to 6 GHz –8 dBi at 30 MHz8.5 dB at > 200 MHz

225 W CW at 30 MHz900 W CW at 80 MHz150 W CW at 6 GHz

80 ° to 30 °

R&S®HL223 log- periodic antenna

200 MHz to 1.3 GHz

≥ 6 dBi 1500 W CW at 200 MHz600 W CW at 1.3 GHz

approx. 60 °

R&S®HL046E log- periodic antenna

80 MHz to 3 GHz typ. 8.0 dBi 2100 W CW at 80 MHz900 W CW at 500 MHz375 W CW at 3 GHz

approx. 60 °

R&S®HL050E log- periodic antenna

750 MHz to 6 GHz typ. 8.5 dBi 100 W CW approx. 60 °

R&S®HF907 double-ridged horn antenna

800 MHz to 18 GHz 5 dBi to 14 dBi 300 W CW at 800 MHz150 W CW at 18 GHz

100 ° to 40 °

RF_fundamentals_po_en.indd 1 03.03.2015 11:51:09

Page 2: RF fundamentals in commercial EMS testing · 2020. 1. 16. · 3607.1713.82 01.00 PDP 1 en About Rohde & Schwarz The Rohde & Schwarz electronics grou is a leading sulier of solutions

R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG

Trade names are trademarks of the owners

PD 3607.1713.82 | Version 01.00 | March 2015 (as)

Data without tolerance limits is not binding | Subject to change

© 2015 Rohde & Schwarz GmbH & Co. KG | 81671 Munich, Germany

Service that adds value❙ Worldwide ❙ Local and personalized❙ Customized and flexible❙ Uncompromising quality ❙ Long-term dependability

3607.1713.82 01.00 PD

P 1 en

About Rohde & SchwarzThe Rohde & Schwarz electronics group is a leading supplier of solutions in the fields of test and measurement, broadcast and media, secure communications, cyber­security, and radiomonitor ing and radiolocation. Founded more than 80 years ago, this independent global company has an extensive sales network and is present in more than 70 countries. The company is headquartered in Munich, Germany.

Sustainable product design ❙Environmental compatibility and eco­footprint ❙Energy efficiency and low emissions ❙Longevity and optimized total cost of ownership

Certified Environmental Management

ISO 14001Certified Quality Management

ISO 9001

Regional contact ❙Europe, Africa, Middle East | +49 89 4129 12345 customersupport@rohde­schwarz.com

❙North America | 1 888 TEST RSA (1 888 837 87 72) [email protected]­schwarz.com

❙Latin America | +1 410 910 79 88 customersupport.la@rohde­schwarz.com

❙Asia Pacific | +65 65 13 04 88 customersupport.asia@rohde­schwarz.com

❙China | +86 800 810 82 28 | +86 400 650 58 96 customersupport.china@rohde­schwarz.com

Rohde & Schwarz GmbH & Co. KGwww.rohde­schwarz.com

Te

st &

Mea

sure

men

t

Post

er |

01.0

0

3607171382

RF fundamentals in commercial EMS testing

RF_fundamentals_po_en.indd 2 03.03.2015 11:51:14