Upload
others
View
8
Download
0
Embed Size (px)
Citation preview
Micro Gas Chromatograph
POKE meeting Tallin, Sabrina Schmidt, Laboratory of Industrial Chemistry and Reaction Engineering,
Åbo Akademi
Classical Gas Chromatograph (GC)
•Injection
•Separation
•Detection
Micro-GC Agilent 3000A
Micro-GC : Elements
Each element contains:
Micro-injector
Sample column (+pre column)
Reference column
Detector
Electronic pressure control
Gas flow solenoids
Hardware
Control board
The portable module
• Delivered also as portable module with a battery
and an inbuilt gas-cylinder
For workers/rescuers in coal mines
• Coal mine explosion: indicated by C2H6, C2H4 and C2H2.
• Micro GC can analyze these compounds in less than 60 seconds at the necessary detection limit of 1ppm
General features
• Mobile
• Modular setup, easy exchange of column types
• Online analysis with programmable sampling intervals
• Very fast analysis, typically less than 180 s
General features
• Fixed/variable volume injector
• Back flush: removes unwanted contaminants from your analysis
• Thermal conductivity detector: 10 times more sensitive than traditional TCDs, allowing you to detect common gas components in the low ppm range.
General features: Precautions
• Only for clean gas samples Micro injector is sensitive towards plugging
• No liquids, no condensibles
• No particles
• Agilent supplies sample filters and conditioners
• Gas liquid separator
• Pressure reducer
Our micro GC
• Four separate modules
• Self-made valve system to connect two reactor setups and two calibration gases
Columns
• Plot U: hydrocarbons, CO2, air/CO, water, polar compounds
• OV-1: Hydrocarbons, phenol, amines, non-polar
• Alumina: petrochemicals (acetylene, ethylene, propadiene)
• Molesieve 5A : Hydrogen, CO, oxygen, nitrogen, methane
Two practical examples from our lab:
Online Analysis
Example 1: Online analysis for methyl chloride synthesis
Analysis of methanol, dimethyl ether, methyl chloride
No liquids: Lines are heated to 95 °C to prevent condensation of methanol
GC signals and analysis time
13
• Sample analysis in less than 1 min
• Separation is not optimal
PlotU
OV-1
Stability of methanol flow
14
0
0,1
0,2
0,3
0,4
0,5
0,6
0,7
0,8
0 200 400
Are
a /
µV
*s*1
06
flow
stop
t/min
•Standard deviation from average flow: 1.6 %
•Fixed injection volume is reliable: easy calibration
•Information on residence time distribution
Switch to standard GC due to breakdown of micro GC
Analysis using a standard GC
Separation: +
Sample analysis : - (10 min instead of 1 min)
Sampling: - (running from the reactor to the GC with a gas tight
syringe heated to 100 °C)
Maintenance: - (septum of reactor sampling unit and GC)
Standard GC is sufficient
Micro GC vs. normal GC
Example 2: Aqueous phase reforming of sugar alcohols
Analysis of CO, CO2, H2 , CH4, linear hydrocarbons C1-C4,
He, N2
All analyzed in one sample/run
Inert gas: 1 wt% He in N2 as internal standard for micro GC
Analysis works perfectly – No change in signal since four years
Summary: Advantages /Disadvantages
• Modular Design: Column, Injector and Detector in a box
No possibility to self-repair
• Micro injector is delicate! Micro injector was plugged
– The whole module has to be replaced
• Mobile
• Online analysis
• Easy calibration due to fixed injection volume
• Parallel analysis with four columns
Thank you for your attention!