NordicNeuroLab

InroomViewingDevice LCD 3.0

User Manual

Page 2: fMRI Hardware System

2 Revision 1

m a k i n g f u n c t i o n a l M R I e a s y

Page 3: fMRI Hardware System

Revision date: May 2016 3

m a k i n g f u n c t i o n a l M R I e a s y

Introduction

NordicNeuroLab’s LCD monitor (NNL MRI InroomViewingDevice) is a complete, state-of-the-art solution designed specifically for use in the MR environment.

The LCD monitor is compatible with MR systems up to 3 T.

The LCD monitor uses a standard HDMI display signal, and can be connected to any computer or video output device with digital outputs.

This User Manual covers all important information regarding operation of the LCD monitor. We strongly recommend that you carefully read this manual to understand the recommended safe operation practices of the equipment and to be able to take full advantage of the features offered by the hardware.

The LCD Monitor complies with the EU regulations for active medical devices, Medical Device Directive 93/42/EEC, classified as a Medical Device Class I product.

Page 4: fMRI Hardware System

4 Revision 1

m a k i n g f u n c t i o n a l M R I e a s y

Page 5: fMRI Hardware System

Revision date: May 2016 5

m a k i n g f u n c t i o n a l M R I e a s y

Contents

Introduction ............................................................................................................................................. 3 Contents .................................................................................................................................................. 5

Safety Information .......................................................................................................................... 6 1.1. Intended use ............................................................................................................................ 6

1.2. General Safety Information ..................................................................................................... 6

1.3. Electrical Safety ....................................................................................................................... 7

1.4. Safety Precautions and Maintenance...................................................................................... 8

1.5. Product Marking and Labeling................................................................................................. 9

General Information ..................................................................................................................... 11 2.1. Inspection of Delivered Goods .............................................................................................. 11

2.2. Power Connection Information ............................................................................................. 11

System Overview and Installation Guide ...................................................................................... 12 3.1. System overview ................................................................................................................... 12

3.2. Installation ............................................................................................................................. 13

3.2.1. Parts Included ................................................................................................................ 13

3.2.2. LCD Stand Assembly Instructions .................................................................................. 16

3.2.3. System installation instruction ...................................................................................... 18

User Information .......................................................................................................................... 20 4.1. Adjusting Height of Stand ...................................................................................................... 20

4.2. In-built front surface camera................................................................................................. 20

4.3. USB connectors ..................................................................................................................... 20

4.4. On Screen Menu .................................................................................................................... 20

4.5. Flip and mirror adjustment ................................................................................................... 21

4.6. Remote control of screen settings ........................................................................................ 21

Maintenance and Disposal ........................................................................................................... 22 5.1. Cleaning Information ............................................................................................................. 22

5.1.1. LCD Monitor .................................................................................................................. 22

5.1.2. LCD Power Supply .......................................................................................................... 22

5.1.3. LCD stand and wires ...................................................................................................... 22

5.2. Correct Disposal of the LCD ................................................................................................... 22

Technical Specifications ................................................................................................................ 23 Contact .......................................................................................................................................... 24

Page 6: fMRI Hardware System

6 Revision 1

m a k i n g f u n c t i o n a l M R I e a s y

Safety Information

1.1. Intended use

The LCD monitor must only be operated by personnel properly trained in MR Safety. This device must not be used for a life support system. This device must not be used for diagnostics. Intended Use (Does not apply to US)

The LCD monitor is intended to be used by trained professionals as a viewing monitor within the MR environment.

Intended Use (Applies to US ONLY)

The LCD monitor is intended to be used by trained professionals as an fMRI viewing monitor within the MR environment.

1.2. General Safety Information

It is important that the monitor is used within the MR environment where the temperature and humidity are kept at a stable level. The background room illumination must be at a minimum level of 100 lux. The LCD monitor must only be operated by personnel properly trained in identifying interference problems such as artifacts, streaks and distortions in image data. It is required that all personnel handling the LCD monitor are familiar with the safety instructions given in this manual and other documentation provided by NordicNeuroLab to ensure the safe operation of the LCD monitor and associated equipment.

Warning

If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:

- Reorient or relocate the receiving antenna.

- Increase the separation between the equipment and receiver.

- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.

- Consult the dealer or NordicNeuroLab for help.

Page 7: fMRI Hardware System

Revision date: May 2016 7

m a k i n g f u n c t i o n a l M R I e a s y

1.3. Electrical Safety

Warning

Do not operate the device near water, where it can become moist, or near excessive heat. Operation under such conditions could result in failure of the LCD monitor, possible electric shock, or fire. If any foreign substances enter the device such as liquid, metal chips or dust, immediately turn off power. This could result in failure of the device, electric shock, or fire.

If smoke, noxious odors or unusual noise should come from the device, immediately turn off power. If the device is incapable of displaying a picture after it has been turned on, turn it off and do not attempt to operate. Disconnect power cord from mains outlet.

Should any cable become damaged or frayed, it must only be replaced by original cable provided by NNL.

If moving the LCD monitor is necessary, switch off the power, disconnect all cables, and then move it to its new position with care.

To disconnect the equipment from the power supply, the power cord must first be removed from the mains outlet.

Do not attempt to remove the back panel of the LCD monitor, as this presents an electric shock hazard.

All service should be conducted by qualified NNL service personnel only.

Switch off the LCD monitor when not in use.

Warning

If this equipment is connected to a mains supply that have a rating voltage 150V to 240V and 60Hz, a protective earth cable has to be permanently installed.

This cable should not exceed the length of 5 meter and should have a dimension of least 2.5mm2

Warning

To avoid risk of electric shock, this equipment must only be connected to the mains supply with a protective earth.

Page 8: fMRI Hardware System

8 Revision 1

m a k i n g f u n c t i o n a l M R I e a s y

1.4. Safety Precautions and Maintenance

Caution

Eye injury can be caused by looking directly into the light beam of the Fiber Transmitter.

Warning

Handle the LCD monitor with care; do not subject it to excessive vibrations or movement of any kind.

Do not obstruct or cover the LCD monitor or the LCD Power Supply unit with any type of material or objects. This could prevent proper cooling of the electronics.

Consult a service technician if the display does not operate normally when the operating instructions are followed.

Warning

For a small percentage of the population, viewing certain types of video that contain flashing patterns of light may trigger epileptic seizures.

The following users are cautioned to consult a physician before using the LCD:

- children under 5 years

- anyone with a history of epilepsy, or a family history of epilepsy

- anyone who has ever experienced epileptic seizures or sensory disturbances triggered by flashing light effects

Warning

Self-heating of cables may occur due to improper cable routing! This may cause injury to equipment and/or operating personnel.

- Cables should not be looped or crossed.

Caution

Parts of the LCD system may get hot during operation! Burn hazard.

LCD monitor should not be touched for more than 10 seconds during normal use.

Page 9: fMRI Hardware System

Revision date: May 2016 9

m a k i n g f u n c t i o n a l M R I e a s y

Caution

No or incorrect visual feedback during MR guided or MR monitored procedure- Risk of patient injury. It is in the responsibility of the performing physician to plan appropriate emergency exit measures prior start of a MR guided or MR monitored procedure.

Caution

Do not use the LCD monitor for diagnosis.

1.5. Product Marking and Labeling

Consult Accompanying Documentation

This is a general symbol used to encourage users to get familiar with the accompanying documentation and safety precautions before using the system.

Attention

Consult the User Manual for more information.

Follow instructions for use

MR Conditional

An item that has been demonstrated to pose no known hazards in a specified MR environment with specified conditions of use.

This marking applies to:

- LCD

- LCD Power Supply (LPS)

- Power cable for LCD monitor

- LCD monitor stand

See installation guide for each product for safe operating conditions.

MR Unsafe

An item that is known to pose hazards in all MR environments and must not be brought into the MR suite.

This marking applies to:

Page 10: fMRI Hardware System

10 Revision 1

m a k i n g f u n c t i o n a l M R I e a s y

- Fiber Transmitter

Power ON

The power on symbol, appearing on one end of a toggle switch indicates that the control places the equipment into a fully powered state.

Power OFF

The power off symbol on a toggle switch indicates that using the control will disconnect power to the device.

Power

The power symbol on a toggle switch indicates that using the control will connect/disconnect

power to the device.

Protective earth

Additional protective earth connection. Must be used at voltage range from 150V – 240V @ 60Hz

Fragile; Handle with Care

Keep dry

This way up

Do not stack

Storage temperature

Not below -20°C and not above +60°C

Storage humidity

Not below 20% and not above 90%

Kensington Lock

A Kensington lock is provided for safe locking of the LPS when installed. This is to help avoid the LPS being moved by untrained personnel within the MR environment.

AC Power

CE Marking

Page 11: fMRI Hardware System

Revision date: May 2016 11

m a k i n g f u n c t i o n a l M R I e a s y

General Information

2.1. Inspection of Delivered Goods

Each LCD Monitor has been tested prior to delivery and is ready for immediate use. Upon receipt, please report any transportation damage or missing accessories immediately. NNL can only accept liability for such damage if it is claimed prior to initial operation. In case of transportation damage, please contact the NNL customer service department. For faster support please have ready all shipment details, LCD monitor serial number and damage description. It is recommended to keep and store the original crate/box used to ship the LCD monitor for all future transportation needs.

2.2. Power Connection Information

The LCD monitor operates on the line voltages 90 – 250V AC and 50/60 Hz. If this equipment is used on line voltages from 150V to 240V and 60Hz, a protective earth cable has to be permanently installed. If the supply voltage in your location is not 90 – 250 V AC, please consult your NNL Sales representative or NNL customer service. Always use the power supply cord supplied in the original shipping carton. If the enclosed power cord could not be used due to a different standard in your country, use a power cord that conforms to your regional standards.

In order to prevent electrical hazard, the cord set must use a three-core cable of at least 6A/0.75mm2,

and have an earth/ground contact on both the free socket and the plug.

Warning

The fuse holder for the main fuse is integrated in the power inlet in the LCD Power Supply. Replace the fuse with same type, characteristics and voltage rating (Littlefuse 02153.15).

Disconnect power before changing the main fuse.

Note: Incorrect cable connections may result in irregular LCD function, damage to the LCD monitor quality and/or components and may shorten the life of the unit.

Page 12: fMRI Hardware System

12 Revision 1

m a k i n g f u n c t i o n a l M R I e a s y

System Overview and Installation Guide

3.1. System overview

The LCD monitor consists of four components:

- Monitor: A powerful and flexible solution which utilizes LED backlight technology to present visual stimuli in the MR room. The monitor has a camera on the front and USB connectors on the back. The USB connectors are for HID devices only.

- LCD Power Supply (LPS): Shielded enclosure containing the Power Supply Unit for the LCD Monitor which is located in the MR environment.

- Fiber Transmitter: Translating the video signal from a computer (or similar video output device) into an optical signal.

- LCD Stand: An aluminium stand for the monitor.

Figure 1 gives an overview over the set-up of the LCD Monitor. All signals entering the shielded magnet room are transmitted by use of fiber optics only.

Figure 1. High level overview of the signal path and set-up of the LCD Monitor.

Page 13: fMRI Hardware System

Revision date: May 2016 13

m a k i n g f u n c t i o n a l M R I e a s y

3.2. Installation

3.2.1. Parts Included

NNL LCD Monitor , LCD 3.0

Presentation of visual stimuli and video materials inside the MR room.

The LCD monitor contains trace amounts of ferromagnetic material compared to the total weight of the system. This poses no risk or (hazard) of being pulled in by the magnetic field. All assembly, maintenance and servicing activities must be performed outside the Magnet Room as ferromagnetic parts may be exposed.

LCD Power Supply (LPS)

Shielded enclosure for the Power Supply Unit.

The LPS contains ferromagnetic materials and must be kept at a safe distance from the MR scanner (outside the 20 mT (200 Gauss) line).

LCD power and fiber cable

Link between the LCD monitor and LPS.

One end of the cable contains ferromagnetic material and can be identified by its warning label and rectangular shape. This end must be kept at a safe distance from the MR scanner (outside the 20 mT (200 Gauss) line) and must be the end connected to the LPS rather than the LCD display.

Fiber Transmitter

Converts video signal into optical signal for transmission.

Contains ferromagnetic materials, keep out of the Magnet Room.

EDID adapter

Makes the computer “see” the NNL LCD monitor even if it’s turned off

Contains ferromagnetic materials, keep out of the Magnet Room.

Fiber cable

Connection between the Fiber Transmitter and the LPS through an available waveguide.

This cable is fragile!

Minimum bending radius: 10 cm

Page 14: fMRI Hardware System

14 Revision 1

m a k i n g f u n c t i o n a l M R I e a s y

DVI-to-HDMI adapter

Adapts DVI-D signal to an HDMI signal; used if the PC or video output device only has a DVI port available

Contains ferrormagnetic materials, keep out of the Magnet Room.

DisplayPort-to-HDMI adapter

Adapts a DisplayPort signal to an HDMI signal; used if the PC or video output device only has a DisplayPort port available

Contains ferrormagnetic materials, keep out of the Magnet Room.

Power cable

For powering the LPS, a means to isolate me equipment from supply mains. The power cable delivered with your system may vary depending on your type of mains outlet .

This cable should NOT be accessible in case of an emergency?

Power cables contain ferromagnetic materials, keep at safe distance from the MR scanner (outside the 20mT / 200 Gauss line).

Page 15: fMRI Hardware System

Revision date: May 2016 15

m a k i n g f u n c t i o n a l M R I e a s y

The stand consists of following items:

2 x LCD stand leg

The LCD stand contains trace amounts of ferromagnetic compared to the total weight of the system, and poses no risk (hazard) of being pulled in by the magnetic field. However, all assembly, maintenance and servicing activities must be performed outside the Magnet Room as ferromagnetic parts may be exposed.

1 x lower vertical stabilizer

The LCD stand contains ferromagnetic material. All assembly, maintenance and servicing must be performed outside the Magnet Room.

1 x vertical stabilizer with VESA fixation

The LCD stand contains ferromagnetic material. All assembly, maintenance and servicing must be performed outside the Magnet Room.

1x vertical stabilizer

The LCD stand contains ferromagnetic material. All assembly, maintenance and servicing must be performed outside the Magnet Room.

14 x M8x50 Allen screw

The LCD stand contains ferromagnetic material. All assembly, maintenance and servicing must be performed outside the Magnet Room.

Page 16: fMRI Hardware System

16 Revision 1

m a k i n g f u n c t i o n a l M R I e a s y

3.2.2. LCD Stand Assembly Instructions

Warning

The LCD stand contains ferromagnetic material. All assembly, maintenance and servicing must be performed outside the Magnet Room.

Do not bring the LCD stand inside the Magnet Room until the LCD monitor has been securely mounted on the stand.

The assembly of the LCD stand is divided into three steps:

Step 1:

- Use the lower stabilizer bar and two M8x50 Allen screws to connect the two legs.

- Unscrew the tightening knob on both legs in order to pull the inner telescopic leg out. Pull them up to a height of around 35cm.

Step 2:

- Mount the stabilizer leg with VESA brackets to the upper part of the telescopic legs with four M8x50 Allen screws.

Optional:

- If the stand is to be used in a location where the lower bar needs to be removed, then a second upper stabilizer bar must be used. This bar is mounted below the first bar using four M8x50 Allen screws.

Page 17: fMRI Hardware System

Revision date: May 2016 17

m a k i n g f u n c t i o n a l M R I e a s y

Step 3:

- Place the LCD monitor according to the drawing and insert M8x50 screws from the back of the stand. It is recommended that two persons perform this task to avoid damage to the LCD monitor or the stand.

Warning

When adjusting the height, two persons are required for safe operation

Page 18: fMRI Hardware System

18 Revision 1

m a k i n g f u n c t i o n a l M R I e a s y

3.2.3. System installation instruction

Operator Room

Connect the Fiber Transmitter to the desired PC or video output device HDMI port. To utilize the computer’s DVI or DisplayPort interface, use the DVI/DisplayPort-to-HDMI adapter that comes with the system.

Connect the EDID adapter straight to the fiber transmitter, this will give the right screen settings to the computer and also make the computer remember the NNL monitor when it’s turned off.

Connect the fiber cable to the Fiber Transmitter and feed the fiber cable through the waveguide into the MR suite.

Connect the micro USB to the Fiber Transmitter and to the computers USB port. If a DVD/BluRay-player is used as a source, connect the USB cable to the supplied USB power supply.

The Fiber Transmitter also acts as an USB hub, providing access to the monitor camera and the USB connectors at the rear of the monitor.

Magnet Room

Danger: The LPS contains ferrous material and extreme caution must be used while moving and installing the device in the MR suite. Improper handling of LPS within a magnetic field of greater than 20mT may result in serious injury or death. Only technicians with knowledge of the risk of introducing ferrous objects into the MR are allowed to perform the installation.

Page 19: fMRI Hardware System

Revision date: May 2016 19

m a k i n g f u n c t i o n a l M R I e a s y

First decide where to place the LCD Monitor. It is normally located at the back of the scanner or at the side of the patient bed. Consider the length of the power cable from the LPS to the LCD.

Then decide where to place the LPS. Use a gauss meter to measure the magnetic field at this location or use the technical drawings for the magnet room to ensure that the LPS location is at below 20 mT (200 Gauss). Note that the LPS must be located near an AC power outlet.

Before taking the LPS into the magnet room, verify that the path which the LPS will be moved to the installation location, is outside the 20 mT (200 Gauss) limit.

Once the final location for installation has been selected and the path for moving the LPS

into the magnet room has been determined, only then should the installation take place. For safety reasons, only the installer should be present in the magnet room during the installation.

Secure the LPS in a fixed location where the magnetic field is less than 20 mT (200 Gauss). A Kensington Lock should be attached to the back of the LPS (4) to avoid unwanted movement by untrained personnel. The Kensington Lock should restrict movement of the LPS to always be outside a 20mT (200 Gauss) limit.

Connect the fiber cable to the fiber input on the LPS (7).

Connect the LCD power cable between the LCD Monitor and the LPS. Start by connecting the end that is to be connect to the LPS end (1).

Connect the fibers that are bundled into the power cable. On the LCD monitor, use the connector that is marked with “Fiber input”only.

Connect the power cable to the LPS (5) and to the mains outlet. Switch on the LPS (3) and confirm that the backlight of the LCD lights up and the NNL splash screen is present.

Thanks

Warning

IMPORTANT! The end with the ferrite bead is to be connected to the LPS first. It can be identified by the rectangular shape and a warning label relating to ferromagnetic material. This has to be done before connecting the cable to the LCD monitor.

3

4

6

7

8

5 1

2

Page 20: fMRI Hardware System

20 Revision 1

m a k i n g f u n c t i o n a l M R I e a s y

Caution

Do not move the LCD monitor by dragging or pulling the power cable. This could result in malfunction and damage to the system.

Turn off the system before disconnecting the LCD power cable!

If this is done while the system is on, or the LCD power cable is inserted while the LPS is on, then the system needs to be restarted to operate normally.

User Information

4.1. Adjusting Height of Stand

The LCD monitor height above the floor can be adjusted: While one person is holding the monitor, loosen the knobs on both telescopic legs. Gently lower or raise the section that the LCD monitor is mounted on. Tighten both knobs on the telescopic legs to lock the LCD monitor into position.

Warning

Crush hazard. Keep hands away from moving telescopic legs when adjusting height.

Caution

Do not unscrew the tightening wheel on the telescopic legs all the way out. This may damage the stand and/or cause problems securing the LCD monitor in a suitable position.

4.2. In-built front surface camera

The in-built camera is working as a standard USB “web camera”. This camera is only accessible when the Fiber Transmitter is connected to a PC with the USB cable. A standard web camera software can be used to capture the video from this camera.

4.3. USB connectors

The USB connectors located on the back of the LCD monitor can be used for connecting HID units, like keyboard and mouse. Since this USB is a USB1.1 link, no mass storage device and high speed devices can be used.

4.4. On Screen Menu

The On Screen Menu allows for adjustments of picture settings to optimize for a particular application or environment. Note: Most adjustments are already optimized and there is no need to change them.

Page 21: fMRI Hardware System

Revision date: May 2016 21

m a k i n g f u n c t i o n a l M R I e a s y

The On Screen Menu touch buttons are located on the front top left corner of the LCD Monitor. The control functions of these buttons are described below

Button Function

Activate menu

Exit Menu

Exit submenu

Enter submenu

Enter

Move up in menu

Increase level

Move down in menu

Decrease level

4.5. Flip and mirror adjustment

The LCD monitor has the option to flip and mirror the image on the screen. The image can be rotated 180 degrees and mirrored, in total four modes. The selected setting is stored and will be retained after power cycling. To active the flip mirror, enter the On Screen Menu and enter the “other” submenu. Then enter the “Display rotation” option and choose one of the four desire option. Active with the enter button.

4.6. Remote control of screen settings

It’s possible to remotely control all screen settings through the inbuilt USB link. This can be used for advanced users who wants to have full control of the monitor from the operator-room. Please contact support for guidelines how to set up this control. Contact information found under “Contact”

Page 22: fMRI Hardware System

22 Revision 1

m a k i n g f u n c t i o n a l M R I e a s y

Maintenance and Disposal

5.1. Cleaning Information

Warning

Unplug all power connections before cleaning.

5.1.1. LCD Monitor

Dust and other matter that collect on the LCD monitor screen may eventually degrade picture quality. Dampen (do not saturate) a clean, lint-free cloth with environmentally friendly glass cleaner. Clean the monitor screen using circular motions with the cloth to avoid streams. Remove fingerprints, grease, dirt and dust. Carefully dry the screen with a second, lint-free cloth.

5.1.2. LCD Power Supply

Clean the ventilation holes every six months. This cycle period safely prevents the holes from being covered with dust and particles. Open holes guarantee reliable cooling of the internal electronic circuits.

5.1.3. LCD stand and wires

Use a cloth with disinfectant (dampen, do not saturate) to clean the stand, casing and wires. Remove fingerprints, dirt, grease and dust. Do not use any oil or grease on the mechanical parts or wheels.

5.2. Correct Disposal of the LCD

To prevent possible harm to the environment due to uncontrolled waste disposal, please separate this product from other types of waste and recycle it at the end of its working life. Users should contact their local government office for details regarding where and how the LCD monitor can be safely recycled. Alternatively, the product can be returned to NordicNeuroLab for appropriate disposal.

Page 23: fMRI Hardware System

Revision date: May 2016 23

m a k i n g f u n c t i o n a l M R I e a s y

Technical Specifications

LCD Monitor Display type TFT active matrix liquid crystal panels

Screen Diagonal 40’’

Active area 878.112mm(H) x 485.352mm(V)

Pixels 3840 x 2160

Pixel Pitch 0.076225 (H) x 0.2247 (V)

Pixel Arrangement R,G,B vertical stripe

Display Mode Normally black

Display colors 10bit(8 + FRC), 1073.7M

Surface Luminance 350 c/m2 typ.

Contrast ratio 5000:1 typ.

Refresh rate 60Hz@UHD 120Hz@FHD

Backlight LED

Response Time 8,5 ms (G to G)

Color Gamut NTSC 88%

Size; weight 933.1 x 550.4 x 38.4 mm; 16.7kg

VESA fixation 2x 100x100, D: 400mm, M4

Alternative fixation 40mm x 300mm, M6

Camera 640 x 480 MAX, 30fps

USB connectors Dual Type-A, Low Speed (1.5 Mbit/s), 0.5A, 10W @ +5V

Fiber Transmitter Input signal TMDS Signal (DVI 1.0 standard)

Electrical connector HDMI receptacle type

Graphical Resolution Support 4096x2160@30Hz HMDI 1.4b

Optical connector LC Simplex Multi Mode 50/125µm [OM3]

Max Data Transfer rate 2.97Gpbs

Size 55 x 54 x 15 mm

Power consumption 0.45A, 0.85W @ +5V

Power connector Micro USB

LCD Power Supply (LPS) Voltage range 90 - 250 VAC

Frequency 50/60 Hz

Fuse T 3.15AH 250V

Power consumption Approx. 140 VA

Size 235 x 315 x 132 mm

Optical In/Output 2xLC-connector /MPO-connector

Environmental Data Storage ambient temperature -20 to +60 degrees Celsius

Operation ambient temperature 0 to +50 degrees Celsius

Relative Humidity 20% to 90%

Maximum operation altitude 4000 m. above sea level

Panel Surface Temperature 65 degrees Celsius

Static magnetic field Max. 3 Tesla

Lifetime 30000 hrs

Approvals and legal requirements Safety IEC60601-1 3’rd edition

EMC IEC60601-1-2 3’rd edition

NordicNeuroLab will provide circuit diagrams, component part lists, descriptions and calibration instructions to assist authorized service personnel in parts repair.

Page 24: fMRI Hardware System

24 Revision 1

m a k i n g f u n c t i o n a l M R I e a s y

Contact

If you have any questions or problems, please contact us through any of the following communication channels: If you purchased your system through one of our international distributors, please contact the distributor first, especially if your system is under warranty. In all other cases, please contact [email protected] if you have any questions or to get assistance with technical problems. For Sales related questions, please contact your local distributor, or contact us at [email protected].

NordicNeuroLab AS Møllendalsveien 65 C 5009 Bergen Norway www.nordicneurolab.com