ENT 412 Bioelectrical Instrumentation Design_4

Embed Size (px)

Citation preview

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    1/49

    ENT 412 BIOELECTRICAL

    INSTRUMENTATION DESIGN

    BIOELECTRODES

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    2/49

    A BRIEF H ISTORY OF BIOELECTRODES Electrochemistry studies on electrode polarization

    Electrode polarization is an interfacial phenomenonoccurring at the electrode-electrolyte interface

    Research started from 1826 Current research: Extensive work on tissue

    impedance measurements was done by Schwancommencing in 1951

    Schwan also engaged in extensive studies onpolarization phenomena involving platinumelectrodes, including platinum black electrodes overboth the linear and non-linear range

    2

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    3/49

    SOURCES OF BIOELECTRIC SIGNALS Endogenous and Exogenous signals

    Endo - arise from natural physiological processes andare measured within or on living creatures

    Exo- applied from without (generally noninvasively)to measure internal structures and parameters

    Bioelectric signals arise from the time-varyingtransmembrane potentials seen in nerve cells(neuron action potentials and generator

    potentials) and in muscle cells

    3

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    4/49

    4

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    5/49

    EMG EMG recording is used to diagnose some causes of

    muscle weakness or paralysis, muscle or motorproblems such as tremor or twitching, motor nervedamage from injury or osteoarthritis, and pathologies

    affecting motor end plates Carried out on of skeletal muscles and superficial

    muscles

    A skeletal muscle fiber action potential propagates at3 to 5 m/sec; its duration is 2 to 15 msec, depending

    on the muscle, and it swings from a resting value ofapproximately -85 mV to a peak of approximately +30mV. At the skin surface, it appears as a triphasicspike of 20- to 2000-mV peak amplitude (Guyton,1991)

    5

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    6/49

    6

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    7/49

    EMG amplifier gains are typically X1000 andtheir bandwidths reflect the transient nature of

    the single motor units (SMU) action potentials -reactively coupled with low and high -3-dBfrequencies of 100 and 3 kHz, respectively

    EMGs can be viewed in the time domain (most

    useful when single fibers or SMUs are beingrecorded), in the frequency domain (the FFT istaken from an entire, surface-recorded EMGburst under standard conditions), or in the timefrequency (TF) domain

    7

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    8/49

    ECG

    8

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    9/49

    The QRS spike in the ECG is seen to be associated

    with the rapid rate of depolarization of ventricularmuscle just preceding its contraction. The P wave iscaused by atrial depolarization and the T wave isassociated with ventricular muscle repolarization

    ECG QRS spike can range from a 400-mV to 2.5-mV

    peak - the gain required for ECG amplification isapproximately 103

    9

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    10/49

    EEGThe largest EEG potentials recorded on the scalp

    are approximately150 mV at peak

    The standard 10 to 20 EEG electrode array uses

    19 electrodes; some electrode arrays used inbrain research use 128 electrodes

    EEG amplifiers must work with low-frequency,low amplitude signals; consequently, they mustbe low noise types with low 1/f noise spectrums.

    EEG amplifiers can be reactively coupled; their -3-dB frequencies should beabout 0.2 and 100 Hz.Amplifier midband gain needs to be on the orderof 104 to 105 10

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    11/49

    11

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    12/49

    The most noteworthy features of biopotentials are

    Small amplitudes (10 mV to 10 mV)

    Low frequency range of signals (dc to several hundredhertz)

    The most noteworthy problems of such acquisitionsare

    Presence of biological interference (from skin,electrodes, motion, etc.),

    Noise from environmental sources (power line, radiofrequency, electromagnetic, etc.).

    12

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    13/49

    PRINCIPLES OF BIOPOTENTIAL

    MEASUREMENTS Electrode design and its attachment suited to the

    application;

    Amplifier circuit design for suitable amplificationof the signal and rejection of noise andinterference;

    Good measurement practices to mitigateartifacts, noise, and interference.

    13

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    14/49

    ELECTRODES FOR BIOPOTENTIAL

    RECORDINGS SilverSilver Chloride Electrodes

    consists of a highly conductive metal, silver,interfaced to its salt, silver chloride, and connectedvia an electrolytic gel to the human body

    design to produce the lowest and most stable junctionpotentials - J unction potentials are the result of thedissimilar electrolytic interfaces, and are a serioussource of electrode-based motion artifacts

    additionally, an electrolytic gel typically based on

    sodium or potassium chloride is applied to theelectrode

    A gel concentration in the order of 0.1M (molarconcentration) results in a good conductivity and lowjunction potential without causing skin irritation

    14

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    15/49

    Reusable silversilver chloride electrodes are made ofsilver disks coated electrolytically by silver chloride,or, alternatively, particles of silver and silver chlorideare sintered together to form the metallic structure ofthe electrode.

    suited for acute studies or basic researchinvestigations

    15

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    16/49

    Disposable electrodes are made similarly, althoughthe use of silver may be minimized (for example,thesnap-on button itself may be silver coated andchlorided).

    To allow for a secure attachment, a large foam padattaches the electrode body with adhesive coating onone side.

    Suited for ambulatory or long term use. 16

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    17/49

    Gold Electrodes

    have the advantages of high conductivity andinertness desirable in reusable electrodes

    commonly used in EEG recordings

    Small reusable electrodes are designed so that theycan be securely attached to the scalp

    The electrode body is also shaped to make a recessedspace for electrolytic gel, which can be appliedthrough a hole in the electrode body

    17

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    18/49

    The electrodes are attached in hair-free areas by useof a strong adhesive

    Disadvantagesof using gold electrodes over silversilver chloride electrodes - greater expense, higher

    junction potentials, and greater susceptibility tomotion artifacts

    Advantages - maintain low impedance, inert andreusable, and good for short-term recordings as longas a highly conductive gel is applied and they are

    attached securely

    18

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    19/49

    Conductive Polymer Electrodes

    Certain polymeric materials

    have adhesive properties andby attaching monovalent

    metal ions can be made conductive

    19

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    20/49

    The polymer is attached to a metallic backing madeof silver or aluminum foil, which allows electriccontact to external instrumentation

    This electrode does not need additional adhesive orelectrolytic gel

    The conductive polymeric electrode performsadequately as long as its relatively higher resistivity

    (over metallic electrodes) and greater likelihood ofgenerating artifacts are acceptable

    20

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    21/49

    Needle Electrodes

    comprise a small class of invasive electrodes, usedwhen it is absolutely essential to record from theorgan itself

    The most common application is in recording frommuscles or muscle fibers

    21

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    22/49

    A metallic, typically steel, wire is delivered via aneedle inserted at the site of the muscle fiber. Thewire is hooked and hence fastens to the muscle fiber,even as the needle is removed. Small signals such as

    motor unit potentials can be recorded in this manner

    use is limited to only highly specialized andsupervised clinical or research applications

    22

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    23/49

    ELECTRIC CHARACTERISTICS

    The electric characteristics of biopotentialelectrodes are generally nonlinear and a functionof the current density at their surface

    electrodes can be represented by an equivalentcircuit

    23

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    24/49

    Rd andCdar e componen tsthat represent theimpedance associated with the electrodeelectrolyte interface and polarization at thisinterface.

    Rs i s the ser i es resi stance associ at ed w i th

    i n ter faci al effects and the r esi stance of the

    electrodematerials themselves

    The battery Ehc r epr esen ts th e hal f-cel l poten ti al

    24

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    25/49

    25

    An example of biopotential electrodeimpedance as a function of frequency.Characteristic frequencieswill be somewhat different for electrodedifferent geometries and materials.

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    26/49

    26

    The Effect of Electrode Properties on Electrode Impedance

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    27/49

    BIOPOTENTIAL

    AMPLIFIERS

    27

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    28/49

    The Instrumentation Amplifier

    28

    The instrumentation amplifier. This amplifier has a very high input impedance, high CMRR, and adifferential gain set by the resistors in the two amplifier stages. The gain of the first stage (amplifiers A1 and A2)is 1 +2R2/R1, the second stage (amplifier A3) is R4/R3, and the third stage (amplifier A4) is 1 + R7/R6. Thelower cornerfrequency is 1/(2R5C1) and the upper corner frequency is 1/(2R7C2). The variable resistor R isadjusted to maximize the CMRR. Electrodes E1 and E2 are the recording electrodes while E3 is the referenceor the ground electrode.

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    29/49

    The key design component of all biopotentialamplifiers is the instrumentation amplifier

    This design results in the desired differentialgain distributed over two stages of the amplifier

    I t also achieves a very high input resistance as aresult of the noninverting amplifier front end

    I t exhibits a very high CMRR as a result of thedifferential first stage followed by a second-stagedifferential amplifier - The CMRR is enhanced byadjusting one of the matching resistors and byselecting high CMRR op amps

    29

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    30/49

    ECG AMPLIFIERS

    Active filters with a lower corner frequency of0.05 Hz and an upper corner frequency of 100 Hzare also typically added

    leakage from the amplifier is required to be belowthe safety standard limit of 10 mA

    safety of the patient is achieved by providingelectrical isolation from the power line and theearth ground, which prevents passage of leakage

    current from the instrument to the patient undernormal conditions or under reasonable failureconditions

    30

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    31/49

    Electrical isolation is achieved by usingtransformer or optical coupling components

    In use with defib - the amplifier circuit must be

    protected against the high defibrillation voltagesand must be augmented by circuit componentssuch as current-limiting resistors, voltage-limiting diodes, and spark gaps

    31

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    32/49

    EEG AMPLIFIERS

    The distinguishing feature of an EEG amplifier isthat it must amplify very small signals

    all components of the amplifier must have a very

    low thermal noise and in particular low electronic(voltage and current) noise at the front end of theamplifier

    EEG amplifiers used in clinical applications mustbe electrically isolated and protected against high

    defibrillation voltages

    32

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    33/49

    EMG AMPLIFIERS

    EMG amplifiers are often used in theinvestigation of muscle performance,neuromuscular diseases, and in building certainpowered or smart prostheses - enhancedamplifier bandwidth suffices

    postprocessing circuits are almost always needed

    33

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    34/49

    CIRCUIT ENHANCEMENTS

    These enhancements include circuits for reducingelectric interference, filtering noise, reduction ofartifacts, electrical isolation of the amplifier, andelectrical protection of the circuit againstdefibrillation shocks

    34

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    35/49

    Electrical Interference Reduction

    Sources of interference include induced signals frompower lines and electric wiring; RF fromtransmitters, electric motors, and other appliances;magnetically induced currents in lead wires; and soon

    Interference induced on the body common to the

    biopotential sensing electrodes is called the commonmode interference (as distinguished from thebiopotential that is differential to the sensingelectrodes)

    The common mode interference is principally rejected

    by a differential or instrumentation amplifier with ahigh CMRR. Further improvement is possible by useof the driven right leg circuit.

    35

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    36/49

    Electrical Interference Reduction

    The driven right leg circuit employs the clever idea ofnegative feedback of the common mode signal into this lead.The common mode signal is sensed from the first stage ofthe instrumentation amplifier, amplified and inverted, andfed back into the right leg lead

    At this stage the common mode signal is reduced to(idR0)/ (1 + 2R2/ R1)

    The driven right leg circuit along with a high CMRR of theamplifier and filtering permit very high quality biopotentialmeasurements

    36

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    37/49

    37

    The schematic on the left shows electric interference induced by thedisplacement current id fr om the power l i ne. Thi s cur r ent f lows in to thegroun d el ectr odelead generating common-mode voltageVc. The dr i venr i ght l eg ci r cui t on th e r i ght uses negati ve feedback i nt o th e

    right leg electrode to reduce the effective common-mode voltage.

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    38/49

    Filtering filtering at the front end of the amplifier and limiting

    the bandwidth of the biopotential amplifier canfurther help to reduce the interference

    Small inductors or ferrite beads in the lead wireshelp to block very high frequency electromagneticinterference

    Small capacitors between each electrode lead andground filter the RF interference

    use of high-pass filtering in the early stages ofamplification is recommended - dc potentials arisingat the electrodeskin interface

    Low-pass filtering at several stages of amplification

    is recommended to attenuate residual RFinterference as well as muscle signal interference

    a 50 or 60 Hz notch filter to remove the power lineinterference

    38

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    39/49

    39

    Amplifier front end filters T1: RF choke; R0

    andC0: RF fi l ter ; R1 and C1: high-pass fi l ter ; R2and C2: low-pass fi l ter .

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    40/49

    40

    Notch filter for power line interference

    (50 or 60 Hz): twin T notch filter inwhich notch frequency is governed byR1,R2, R3, C1, C2, and C3, and notch

    tuning byR4.

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    41/49

    Artifact Reduction

    computerized processing may be necessary to identifyan artifact and delete it from display and processing

    41

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    42/49

    42

    Baseline restoration circuit: the high-

    pass filter capacitor C1 i s di scha r ged byfi el d effect t r ansi stor

    F when activated manually orautomatically by a baseline restorationpulse.

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    43/49

    Electrical Isolation

    Electrical isolation limits the possibility of thepassage of any leakage current from the instrumentin use to the patient

    patient safety must be ensured by electrical isolationto reduce the prospect of leakage of current from any

    other sensor or instrument attached to the patient tothe Earth ground of the instrument being tested

    Electrical isolation can be done electrically byinserting a transformer in the signal path or opticallyby introducing an optical coupler

    43

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    44/49

    44

    Electrical isolation: transformercoupled using the transformer T (top) or optical using thediode D and the photodetector P (bottom). Note that theisolator separates circuit common on the amplifier sidefrom the Earth ground on the output side.

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    45/49

    Defibrillation Protection

    Biopotential-measuring instruments can encountervery high voltages, such as those from electricdefibrillators, that can damage the instrument

    Therefore, the front end of the biopotentialinstrument must be designed to withstand these highvoltages

    Use of resistors in the input leads can limit the

    current in the lead and the instrument.

    45

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    46/49

    Protection against high voltages is achieved by theuse of diodes or Zener diodes. These componentsconduct at 0.7 V (diode conduction voltage) or 10 to15 V (depending on the Zener diode breakdown

    voltage), thus protecting the sensitive amplifiercomponents

    As a final line of protection, the isolation components(optical isolator or transformer) must be protected by

    a spark gap that activates at several thousand volts.The spark gap ensures that the defibrillation pulsedoes not breach the isolation.

    46

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    47/49

    47

    Electrical protection circuit: resistanceR l im i t sth e cur r ent , rever se-biased di odes D l im i t t he i nput

    vol tage, and th e spar k gap Sprotects againstdefibrillation pulse-related breakdown of theisolation transformer T

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    48/49

    MEASUREMENT PRACTICES

    Electrode use

    Skin Preparation

    Reduction of environmental interference

    48

  • 7/28/2019 ENT 412 Bioelectrical Instrumentation Design_4

    49/49

    CONCLUSION

    Biopotential source presents its own distinct challenge interms of electrode interface, amplifier design, pre- orpostprocessing, and practical implementation and usage

    ECG signals can be best acquired using AgAgCl electrodes,

    although good experimental/clinical practice is needed toreduce biological and environmental interference. Furthercircuit protection and isolation are necessary in clinical usage

    EEG signals are distinguishable by their very low amplitude,and hence EEG electrodes must be securely attached via avery small electrodeskin resistance and the amplifier must

    exhibit exceptionally low noise

    For EMG acquisition, electrodes are needed that can beattached for long periods of time to the muscle groups understudy. The EMG signal inevitably needs postprocessing, suchas integration, to derive a measure of muscle activity

    49