78
linois State University Bioelectrical Impedance Analysis (BIA) Chapter 6

Illinois State University Bioelectrical Impedance Analysis (BIA) Chapter 6

Embed Size (px)

Citation preview

  • Slide 1
  • Illinois State University Bioelectrical Impedance Analysis (BIA) Chapter 6
  • Slide 2
  • Illinois State University BIA n In the last decade, the use of bioelectric impedance and conductivity methods for prediction of body composition has grown rapidly.
  • Slide 3
  • Illinois State University BIA n BIA is a rapid, noninvasive, and relatively inexpensive method for evaluating body composition in field and clinical settings.
  • Slide 4
  • Illinois State University BIA n BIA is now regarded as either a substitute or supplement to conventional anthropometry in field studies.
  • Slide 5
  • Illinois State University BIA n With this method, low-level electrical current is passed through the clients body, and the impedance (Z), or opposition to the flow of the current, is measured with a BIA analyzer.
  • Slide 6
  • Illinois State University BIA n The individuals TBW can be estimated from the impedance measurements because the electrolytes in the bodys water are excellent conductors of electrical current.
  • Slide 7
  • Illinois State University BIA n When the volume of TBW is large, the current flows more easily through the body with less resistance.
  • Slide 8
  • Illinois State University BIA n The resistance to current flow is greater in individuals with large amounts of body fat, given that adipose tissue is a poor conductor of electrical current due to its relatively small water content.
  • Slide 9
  • Illinois State University BIA n Because the water content of the FFB is relatively large (~73% water), FFM can be predicted from TBW estimates.
  • Slide 10
  • Illinois State University BIA n Individuals with large FFM and TBW have less resistance to current flowing through their bodies than do those with a smaller FFM.
  • Slide 11
  • Illinois State University Assumptions n The use of BIA to estimate body composition is based on the different conductive and dielectric properties of various biological tissues at various frequencies of current.
  • Slide 12
  • Illinois State University Assumptions n Tissues that contain a lot of water and electrolytes such as cerebrospinal fluid, blood, or muscle are highly conductive whereas fat, bone, and air-filled spaces such as lung are highly resistive or dielectric tissues.
  • Slide 13
  • Illinois State University Assumptions n An applied electric current always follows the path of least resistance, and in the human body this will include extracellular fluid, blood, muscle, and other conductive tissues that comprise the majority of fat-free mass.
  • Slide 14
  • Illinois State University Assumptions n The volume of these tissues can be deduced from measurement of their combined resistance.
  • Slide 15
  • Illinois State University Assumptions Assumptions are made that do not apply perfectly to the human body as a conductor, and it is important to understand these limitations when using BIA to estimate body composition.
  • Slide 16
  • Illinois State University BIA Impedance (Z) is the frequency- dependent opposition of a conductor to the flow of alternating electric current and is composed of two components, resistance (R) and reactance (Xc).
  • Slide 17
  • Illinois State University BIA Resistance is the pure opposition of the conductor to the flow of the current. Reactance is the storage of an electrical charge by a condenser for a brief moment in time.
  • Slide 18
  • Illinois State University BIA Electrical conduction in biological tissues is mainly ionic: that is, electric charges are transferred between ionized salts, bases, and acids dissolved in the body fluids.
  • Slide 19
  • Illinois State University BIA The conventional approach is to measure whole body resistance or impedance between the wrist and the ipsilateral ankle and to use stature (S) as an index of the length of the conductor.
  • Slide 20
  • Illinois State University BIA Thus, S 2 /R or S 2 /Z is the basic variable used in BIA equations for predicting total body water or fat-free mass.
  • Slide 21
  • Illinois State University BIA Several limitations to this assumption are immediately apparent:
  • Slide 22
  • Illinois State University BIA Geometrically, the body is not a cylinder with uniform cross-sectional area, but is better represented as five cylinders (two arms, two legs, and a trunk) connected in series that have large differences in their cross-sectional areas.
  • Slide 23
  • Illinois State University BIA When such a set of conductors is connected in series, the conductor with the smallest cross-sectional area (i.e., the arm) will determine most of the resistance of the series.
  • Slide 24
  • Illinois State University BIA Thus, whereas an arm is about 4% and a leg about 17% of body weight, they account for about 47% and 50%, respectively, of whole-body resistance: conversely, the trunk comprises about 46% of body weight but may have little if any influence on whole body resistance when measured conventionally from the right ankle to the right wrist.
  • Slide 25
  • Illinois State University BIA Differences in the structure as well as the relative proportions of the trunk versus the limbs also affect the conduction of the current.
  • Slide 26
  • Illinois State University BIA The assumption that whole-body resistance is linearly related to the conductive volume and its electrolyte concentration may not be strictly true.
  • Slide 27
  • Illinois State University BIA Some have reported that BIA predicts fat- free mass less well at the extremes of body fatness.
  • Slide 28
  • Illinois State University BIA The prediction of total body water or fat- free mass using the conventional whole body BIA approach is dependent to a large extent on their strong associations with the mass and bioelectric characteristics of the appendicular skeletal muscle.
  • Slide 29
  • Illinois State University BIA Individuals who deviate markedly from the norm for the size of the trunk in proportion to the limbs are more likely to have erroneous estimates.
  • Slide 30
  • Illinois State University BIA The changes in impedance due to changes in fluid or hydration status are more complex than they appear using the single frequency approach.
  • Slide 31
  • Illinois State University Applicability The BIA method of estimating body composition is best suited to epidemiological studies.
  • Slide 32
  • Illinois State University Applicability It can improve population estimates of obesity and can be used to supplement other field methods in assessing levels of protein-energy malnutrition.
  • Slide 33
  • Illinois State University Applicability BIA can be used also in clinical settings to quantify body composition.
  • Slide 34
  • Illinois State University Accuracy The accuracy of the estimated variables is complicated by factors that may produce shifts in body fluids or electrolytes.
  • Slide 35
  • Illinois State University Accuracy The ability of BIA to detect small changes in body composition has practical limitations as well.
  • Slide 36
  • Illinois State University BIA The BIA method is applicable technically to all subjects regardless of age, sex, ethnicity, or health status.
  • Slide 37
  • Illinois State University BIA In the tetrapolar technique, the paired source and receiving electrodes must be separated by at least 5 cm to avoid interaction.
  • Slide 38
  • Illinois State University BIA In adults, another possible physical limitation to accurate impedance measurements could be extreme obesity.
  • Slide 39
  • Illinois State University BIA The main limitation to the general applicability of BIA is the availability of appropriately calibrated, cross-validated predictive equations.
  • Slide 40
  • Illinois State University BIA It is most important to make a careful selection of equations that were developed from a sample that is similar in age, sex, ethnicity, and health status to the subjects under study.
  • Slide 41
  • Illinois State University BIA Factors that affect the distribution of fluids and electrolyte concentrations between intra- and extracellular compartments can be expected to affect resistance.
  • Slide 42
  • Illinois State University BIA Factors that have acute, temporary effects on fluid and electrolyte equilibrium in healthy subjects, such as exercise, need to be controlled or significant errors may result.
  • Slide 43
  • Illinois State University BIA Pregnancy and menstruation may affect fluid balance and the accuracy of BIA predictions of body composition also.
  • Slide 44
  • Illinois State University BIA The prediction of fat-free mass using BIA is subject to additional complications that may limit applicability over and above those noted so far.
  • Slide 45
  • Illinois State University BIA The most important is considered to be variation in the concentration of water within the FFM, or the ratio of the total body water to FFM (TBW/FFM).
  • Slide 46
  • Illinois State University Equipment: All BIA devices consist essentially of: 1. an alternating electrical current source. 2. cables and electrodes for introducing the current into the body and for sensing the voltage drop due to impedance. 3. a system for measuring impedance.
  • Slide 47
  • Illinois State University BIA Two very different approaches have been used most frequently: two-electrode and four-electrode techniques.
  • Slide 48
  • Illinois State University BIA Each approach has specific advantages and disadvantages.
  • Slide 49
  • Illinois State University BIA In the two-electrode bridge technique, the electrodes that sense the voltage drop are the same as those that introduce the current.
  • Slide 50
  • Illinois State University BIA The main advantages of this approach are that highly accurate measures can be obtained with a very low amplitude current and that electromagnetic leakage toward nearby metallic objects is minimal.
  • Slide 51
  • Illinois State University BIA There are two major disadvantages to this method: 1. the impedance measured reflects both the impedance of the body as well as that due to electrode polarization, which may be high at low frequency.
  • Slide 52
  • Illinois State University BIA 2. Needle electrodes must be used to avoid the high impedance of the skin. These must be inserted subcutaneously in a standardized fashion and may result in minor pain and local tissue trauma that reduce both the acceptability and the accuracy of the impedance measurements.
  • Slide 53
  • Illinois State University BIA The four-electrode technique overcomes the main disadvantages of the two- electrode approach.
  • Slide 54
  • Illinois State University BIA In this method, the current is applied with one pair of electrodes located distally while a second pair located proximally measures the electrical potential across a segment of the conductor.
  • Slide 55
  • Illinois State University BIA Because the electrodes that inject the current are separate from those that detect the potential, impedance due to electrode polarization can be eliminated.
  • Slide 56
  • Illinois State University BIA The use of spot or band electrodes that are attached to the surface of the skin, rather than penetrating it, avoids problems associated with pain and tissue trauma.
  • Slide 57
  • Illinois State University BIA The high cutaneous impedance is overcome by using current with higher amplitudes.
  • Slide 58
  • Illinois State University BIA The two main disadvantages are considered to be the control of stray capacitance produced by the paired wires and electromagnetic leakage to external metallic objects when very high frequencies of current are used.
  • Slide 59
  • Illinois State University Measurement Procedures: It is very important to adhere to standardized measurement procedures when using BIA.
  • Slide 60
  • Illinois State University Measurement Procedures: BIA is measured in a standard fashion while the subject lies supine on a flat, nonconductive bed, cot, or couch.
  • Slide 61
  • Illinois State University Measurement Procedures: The arms are abducted slightly so that they do not touch the sides of the trunk.
  • Slide 62
  • Illinois State University Measurement Procedures: The legs are separated so that the ankles are at least 20 cm apart and, ideally, the thighs do not touch.
  • Slide 63
  • Illinois State University Measurement Procedures: Complete separation of the thighs may be difficult to attain in extremely obese subjects.
  • Slide 64
  • Illinois State University Measurement Procedures: The head should be flat against the bed or elevated minimally with a thin pillow.
  • Slide 65
  • Illinois State University Measurement Procedures: It is important to adhere to this standard body positioning, which has been used in virtually all calibration studies, because deviations produce large differences in the measured impedance.
  • Slide 66
  • Illinois State University Measurement Procedures: The subject may wear clothing, with the exception of shoes and socks, and must not wear metallic jewelry.
  • Slide 67
  • Illinois State University Measurement Procedures: Generally, care should be taken to ensure that the subject and the impedance analyzer are isolated from all metallic objects and other electronic devices by a distance of at least 50 cm.
  • Slide 68
  • Illinois State University Measurement Procedures: For whole-body measurements of adults using the four-electrode technique, the electrodes are attached in the standard fashion to the dorsal surface of the hand and anterior surface of the ipsilateral foot.
  • Slide 69
  • Illinois State University Measurement Procedures: One voltage-sensing electrode is attached to the wrist midway between the styloid processes; the other is attached to the ankle midway between the malleoli.
  • Slide 70
  • Illinois State University Measurement Procedures: The electrodes for introducing the current (source electrodes) are attached to the foot and hand at least 5 cm distal to the sensing electrodes: generally on the third metatarso-phalangeal and third metacarpo-phalangeal joints, respectively.
  • Slide 71
  • Illinois State University Measurement Procedures: The skin should be lightly washed with alcohol before attaching the electrodes.
  • Slide 72
  • Illinois State University Measurement Procedures: The side of the body to which the electrodes are attached does not usually have a significant effect on impedance measurements in healthy subjects.
  • Slide 73
  • Illinois State University Measurement Procedures: Impedance measurements should be taken after a minimum 2 hour fast and at least 8 to 12 hours after any strenuous exercise, alcohol, or other factors that may alter hydration.
  • Slide 74
  • Illinois State University Measurement Procedures: It is recommended that the stage of the menstrual cycle be recorded in longitudinal studies of premenopausal females, although some investigators have not found effects of menstruation impedance measurements.
  • Slide 75
  • Illinois State University Measurement Procedures: Although few studies have demonstrated significant effects, it is recommended that the subject void the bladder completely immediately prior to measurement.
  • Slide 76
  • Illinois State University Measurement Procedures: To minimize changes in impedance due to gravity-induced fluid shifts in healthy subjects, it is recommended that impedance measurements be taken within 5 to 10 minutes after lying down.
  • Slide 77
  • Illinois State University BIA BIA accuracy for prediction of percent BF is 3.5-5%
  • Slide 78
  • Illinois State University BIA These figures should be regarded as ideal or minimal. Actual errors are likely to be greater but will be unknown unless a cross-validation study is performed.