DP7700 Trouble Shooting Guide V1.0

Embed Size (px)

Citation preview

  • 8/8/2019 DP7700 Trouble Shooting Guide V1.0

    1/15

    DP7700-2107 Troubleshooting Guide

    Last Revised: September 29, 2007

  • 8/8/2019 DP7700 Trouble Shooting Guide V1.0

    2/15

    SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

    Last Modified: September 29, 2007 Page 2 of 15

    REVISION CONTROL

    Document Title: DP7700-2107 Troubleshooting Guide

    Author: Mindray Service Department

    Date By Action Pages

    September 3rd,

    2007

    Rudolf Pan New document All

    Review/Approval History

    Date By Action Pages

  • 8/8/2019 DP7700 Trouble Shooting Guide V1.0

    3/15

    SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

    Last Modified: September 29, 2007 Page 3 of 15

    Table of Contents

    Page1. Categories _______________________________________________________________5

    2. Boot up__________________________________________________________________6

    2.1 Symptoms_____________________________________________________________________ 6

    2.2 Check route ___________________________________________________________________ 6

    2.3 Reason of defect________________________________________________________________ 6

    2.4 Defective parts statistics _________________________________________________________ 6

    3. Black screen _____________________________________________________________7

    3.1 Symptoms_____________________________________________________________________ 7

    3.2 Check route ___________________________________________________________________ 7

    3.3 Reason of defect________________________________________________________________ 7

    3.4 Defective parts statistics _________________________________________________________ 7

    4. Keyboard and trackball related_______________________________________________8

    4.1 Symptoms_____________________________________________________________________ 8

    4.2 Solution (keyboard)__________________________________________________________ 8

    4.3 Reason of defect________________________________________________________________ 8

    4.4 Defective parts statistics _________________________________________________________ 95. Lock up __________________________________________________________________10

    5.1 Symptoms____________________________________________________________________ 10

    5.2 Solution _____________________________________________________________________ 10

    5.3 Reason of defect_______________________________________________________________ 10

    6. Image related _____________________________________________________________11

    6.1 Symptoms____________________________________________________________________ 11

    6.2 Check route __________________________________________________________________ 11

    6.3 Reason of defect_______________________________________________________________ 12

    6.4 Defective parts statistics ________________________________________________________ 13

    7. Intermittent problems _______________________________________________________14

    7.1 Symptoms____________________________________________________________________ 14

    7.2 Solution _____________________________________________________________________ 14

    8. Screen flicker _____________________________________________________________15

    8.1 Symptoms____________________________________________________________________ 15

    8.2 Solution _____________________________________________________________________ 15

  • 8/8/2019 DP7700 Trouble Shooting Guide V1.0

    4/15

    SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

    Last Modified: September 29, 2007 Page 4 of 15

    8.3 Reason of defect_______________________________________________________________ 15

  • 8/8/2019 DP7700 Trouble Shooting Guide V1.0

    5/15

    SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

    Last Modified: September 29, 2007 Page 5 of 15

    1. Categories

    This document offers several methods to isolate defects or problems according to the symptoms observed.

    We can classify the symptoms to several categories. The categories are very helpful for you to find

    problems in different systems and its a general tool, not only used in the system described in thisdocument.

    For different categories, we have different methods to solve problems. Most of the methods are

    commonly used in all Mindray ultrasound systems but there might be slight differences when you use

    them on different systems.

    No. Name Description

    Category 1 Boot up related Fail to boot, or halted in certain

    point of the boot progress.Category 2 Black screen Keyboard is functioning but there

    is nothing displayed on the

    monitor.

    Category 3 Keyboard and trackball related Certain keys dont work.

    Category 4 Lock up System halted during certain

    operation.

    Category 5 Image related The system is running well but

    the image is abnormal.

    Category 6 Intermittent problems The problem cannot be

    reproduced according to the same

    operation or handling. However,

    the problem happens in a random

    rate during using.

    Category 7 Screen flicker Screen glinting or trembling

  • 8/8/2019 DP7700 Trouble Shooting Guide V1.0

    6/15

    SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

    Last Modified: September 29, 2007 Page 6 of 15

    2. Boot up

    2.1 Symptoms

    Stops at Mindray LOGO, and nothing else happens

    2.2 Check route

    2.3 Reason of defect

    Software initialization fails

    2.4 Defective parts statistics

    54.50%

    45.50%power supply

    board

    main board

  • 8/8/2019 DP7700 Trouble Shooting Guide V1.0

    7/15

    SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

    Last Modified: September 29, 2007 Page 7 of 15

    3. Black screen

    3.1 Symptoms

    When power on the machine, screen is black.

    3.2 Check route

    3.3 Reason of defect

    Main board failure, power supply board failure, monitors failure

    3.4 Defective parts statistics

    55.00%

    45.00%power supply

    board

    monitor

  • 8/8/2019 DP7700 Trouble Shooting Guide V1.0

    8/15

    SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

    Last Modified: September 29, 2007 Page 8 of 15

    4. Keyboard and trackball related

    4.1 Symptoms

    Some keys are not functioning, trackball out of control, and nothing else abnormal

    4.2Solution (keyboard)

    Note:

    The flow chart above only tell how to deal with the problem that keyboard cant function,

    When meeting trackball out of control and nothing else abnormal, please check the connection between

    Trackball and keyboard or connection between trackball (keyboard) and DSC board.

    Change the keyboard PCB, or trackball

    4.3 Reason of defectKeyboard failure, trackball failure.

  • 8/8/2019 DP7700 Trouble Shooting Guide V1.0

    9/15

    SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

    Last Modified: September 29, 2007 Page 9 of 15

    4.4 Defective parts statistics

    22.00%

    48.00%

    28%2%

    clean or adjustsilicon rubber

    pad position

    silicon rubber

    pad

    keyboard PCB

    power supply

    board

  • 8/8/2019 DP7700 Trouble Shooting Guide V1.0

    10/15

    SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

    Last Modified: September 29, 2007 Page 10 of 15

    5. Lock up

    5.1 Symptoms

    When you press some keys the system locks up, and nothing else abnormal. Or can not wake up from

    system dormant

    5.2 Solution

    Change the main board

    5.3 Reason of defect

    Main board failure

  • 8/8/2019 DP7700 Trouble Shooting Guide V1.0

    11/15

    SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

    Last Modified: September 29, 2007 Page 11 of 15

    6. Image related

    6.1 Symptoms

    Machine is working normally, but image is abnormal

    6.2 Check route

  • 8/8/2019 DP7700 Trouble Shooting Guide V1.0

    12/15

    SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

    Last Modified: September 29, 2007 Page 12 of 15

    Note: Some customers complain that the image quality is not good; generally, these problems can be

    solved by adjusting the machine (such as monitor brightness, image parameters .)

    Sometimes the defect of monitor itself can also cause the image display not good.

    6.3 Reason of defect

    Some electronic or magnetic interference, probe failure, channel board failure, and main board failure.

  • 8/8/2019 DP7700 Trouble Shooting Guide V1.0

    13/15

    SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

    Last Modified: September 29, 2007 Page 13 of 15

    6.4 Defective parts statistics

    22.00% 11.00%

    28.00%

    22.00%17.00%

    machine is not

    adjusted well

    probe

    monitor

    main board

    power supply

    board

  • 8/8/2019 DP7700 Trouble Shooting Guide V1.0

    14/15

    SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

    Last Modified: September 29, 2007 Page 14 of 15

    7. Intermittent problems

    7.1 Symptoms

    Image with some ripples

    7.2 Solution

    Check the surrounding, including AC power input, is there any machines which will interfere with the

    system. Do remember keep a good environment.

  • 8/8/2019 DP7700 Trouble Shooting Guide V1.0

    15/15

    SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

    Last Modified: September 29, 2007 Page 15 of 15

    8. Screen flicker

    8.1 Symptoms

    Screen glinting or trembling

    8.2 Solution

    Exchange the CRT. Or check the surrounding, including AC power input, is there any machines which

    will interfere with the system. Do remember keep a good environment.

    8.3 Reason of defect

    Some wrong with CRT