ARCHITECT c4000 stand-alone System specifications:
1.
System specifications
Primary components System control center, sample handler, and
processing module
Operator interface System control center
Touch-screen monitor
Keyboard and pointing device
Processing module keypad
Optional bar code scanner
Throughput c4000 stand alone:
Photometric assays only
ICT (integrated chip technology) assaysonly
Photometric and ICT assay mix
Individual assay time
Warm-up time from cold start
Up to 400 tests per hour.
Up to 600 tests per hour
Up to 800 tests per hour
Up to 10 minutes
Approximately 30 minutes
7/24/2019 c4000 stand alone specifications.pdf
2/6
Service Department
Inter Bio-Lab, Inc.
Document Nr.
TP A4000-01
Sample types Serum, plasma, whole blood, or other body
fluids.
Sample probe carryover performance
Less than 0.1 ppm (serum, plasma, urine, orcerebrospinal fluid)
Quality control Levey-Jennings and Westgard rules, Control
range tracking
Onboard data storage Hard drive
Stored data protection Optional UPS (uninterruptible power supply)
2.
C4000 System processing module specifications
Bar code readers Reagent supply center and RSH
Dispensing volume:
Sample (Photometric):
Sample (ICT):
Reagent
1.5 - 35.0 L per test in 0.1 L increments.
15 L per sample.
20 - 345 L per test in 1 L increments
Temperature:
Onboard reagent refrigerator
Reaction mixture
Sample carousel
2C to 10C.
36.7C to 37.3C.
Minimum of 10C below room temperature.
3.
C4000 System processing module capacities
Bulk solutions:
ICT Reference Solution:
Alkaline Wash:
Acid Wash:
2L
500 mL
500 mL
Reagent supply center:
Outer carousel:
Inner carousel:
40 - 60 positions
25 - 30 positions
Reaction carousel: 99 cuvettes
Reaction cuvettes:
7/24/2019 c4000 stand alone specifications.pdf
3/6
Service Department
Inter Bio-Lab, Inc.
Document Nr.
TP A4000-01
Minimum volume:
Maximum volume for photometric
reaction:
160 L
360 L
4. Sample handler capacities
Sample carriers: 5 positions per carrier
RSH (robotic sample handler) bays: 4 bays on a stand-alone c4000
RSH (routine sections) Up to 20 sections on a stand-alone c4000
RSH (priority sections) 0-7 sections (up to 35 positions) on a stand-
alone c4000
5. Physical specifications:
Module Depth Width Height Weight
SCC 21" (53.3 cm) 15.5" (39.4 cm) 22" (55.9 cm) 55 lbs. (25 kg)
c4000
processing module and
sample handler
36" (90.7 cm) 63" (160 cm) 49" (125.1 cm) 1132 lbs. (513.5
kg)
6. Weight and force specifications
7/24/2019 c4000 stand alone specifications.pdf
4/6
Service Department
Inter Bio-Lab, Inc.
Document Nr.
TP A4000-01
Total weight 1132.0 lbs. (513.5 kg)
Weight at each foot 1. 264.5 lbs. (120.0 kg)
2.
291.5 lbs. (132.2 kg)
3.
328.5 lbs. (149.0 kg)
4.
247.5 lbs. (112.3 kg)
Force at each foot 1. 60.5 lbs./sq. in. (4.3 kg/sq. cm)
2.
66.7 lbs./sq. in. (4.7 kg/sq. cm)
3.
75.2 lbs./sq. in. (5.3 kg/sq. cm)
4.
56.6 lbs./sq. in. (4.0 kg/sq. cm)
7. Electrical specifications and requirements
AC power: Voltage: 200, 208, 220, 230, or 240 10% VAC
(180 - 264 VAC)
Frequency: 50 or 60 Hz (47 - 63 Hz)
Circuit breaker that can be reset: North America: 20 amp
Outlet (one)
North America
20 amps - NEMA L6-20R (250 VAC, 20A, twist
lock)
With 30 amp UPS - NEMA L6-30R (250 VAC,
30A, twist lock)
Power cord length 13' (4.0 m)
Rated power consumption: 3000 volt amp (3 kVA) maximum
8.
Optical specifications
Light source: Tungsten-halogen lamp
Detector: Silicon photodiode array
Light path length: 5 mm
Photometric method: Concave diffraction grating
Reaction cuvette: Rectangular glass cuvette
Wavelengths: 16 wavelengths (340, 380, 404, 416, 450, 476,
500, 524, 548, 572, 604, 628, 660, 700, 748,
and 804 nm)
Photometric range: -0.1 to 3.2 Abs (converted to 10 mm light path
length)
7/24/2019 c4000 stand alone specifications.pdf
5/6
Service Department
Inter Bio-Lab, Inc.
Document Nr.
TP A4000-01
Linearity: 2% at 2.0 Abs
9. Water and liquid waste specifications and requirements.
Deionized water consumption: 15 L per hour during normal operation
25 L per hour (maximum)
Average liquid waste output: 15 L per hour during normal operation
25 L per hour (maximum)
Water quality:
Maximum microbial contamination:
Minimum resistivity:
Pressure:
IMPORTANT: Do not process samples on theSystem when you are performing maintenance on
your DI water system. After maintenance is
complete, check the pressure to be sure it is within
specifications.
1000 colony-forming units/mL
1 Meg Ohm - cm @ 25C (77F)
15 - 25 psi
Drainage port:
You must have a drainage port that meets the
following specifications or you must use the
external waste pump (optional accessory) if the
drain is located in a sink or otherwise elevated.
Location
Height Within 9.5 ft. (2.9 m) from the rear of the
system
4" (10.0 cm) above floor level
Drainage capacity: 200 L per hour
Liquid waste configuration:
Two gravity-fed waste tubes:
High-concentration waste tube:
Drain into the drainage port or the external
waste pump
Drains into the high-concentration waste
bottle, drainage port, or the external waste
pump.
7/24/2019 c4000 stand alone specifications.pdf
6/6
Service Department
Inter Bio-Lab, Inc.
Document Nr.
TP A4000-01
10.
External waste pump
The external waste pump is optional for use with the c4000 System when a floor drain is unavailable and
a sink drain is used. This accessory pumps waste from the processing module(s) to an elevated drainlocated in a sink.
11.Environmental specifications and requirements
Heat dissipation while in the typical Running
mode:
3050 BTU/hour (894 watts)
Noise level: Does not exceed 85 dBA during normal operation
above a reference sound pressure of 20 Pa.
Temperature, operating: 15C to 30C (59F to 86F)
Humidity: 10% - 85% (non-condensing) RH (relative humidity)
at 25C (77F)
Altitude: 8500 ft. (2590.7 m)
Placement: For indoor use. Do not place in direct sunlight.
Avoid drafts from heating and cooling vents.
Storage and Transport: Keep dry. Fragile - Handle with care.
12.Consumables
Chemistry:
a.
Large, 90 mL cartridge: White or natural.
b.
20 mL cartridge: White.
c.
Small, 55 mL cartridge: White or natural.
d.
Sample cups.
e.
Detergent A.
f.
Detergent B.
g.
Alkaline wash.
h.
Acid wash.
i.
Water bath additive.
Electrolytes (ICT)
a.
ICT module (electrode set).
b.
ICT cleaning fluid.
c.
ICT reference solution.