32
S. No. List of Equipments for CBPST - Kochi Qty Testing Equipment 1. Digital Vernier Calipers Each One No 2. Digital Ball Ended micrometer 3. Digital Thickness Gauge 4. Pi Tapes 5. Analytical Weighing Balance 6. Moisture Analyzer 7. Vacuum Oven with Pump 8. Hydrostatic Pressure testing System for Pipes (a) Hydrostatic Pressure Tester (b) Hot Water bath (c) Low temperature Cabinet (d) End fittings for Pipes 9. Constant Temperature Oil Bath for Reversion Test 10. Falling Weight Impact Tester for Pipe 11. Environmental Stress Cracking resistance Tester 12. Free Falling Dart Impact Tester with Built- In Vacuum System, for Plastics Films Testing 13. Izod and Charpy Impact tester 14. Ultrasonicator 15. Compact Microprocessor-Controlled Bench-Top Laboratory Fermenter-Bioreactor 16. Autoclave 17. Incubator 18. Optical Microscope with Programmable Hot Stage 19. FTIR with Microscope 20. Ultra Violet Visible Double Beam spectrophotometer 21. Universal Testing Machine 22. Gel Permeation Chromatography 23. CHN Analyzer 24. Horizontal Laminar air flow Specification 25. Water Vapour Transmission Rate 26. Gas Permeability Tester PROCESSING MACHINERY 1. Extrusion Blown Film Unit with Carona Treater

CBPST, Kochi Testing Technical Specifications.pdf

Embed Size (px)

Citation preview

Page 1: CBPST, Kochi Testing Technical Specifications.pdf

S.

No.

List of Equipments for CBPST - Kochi Qty

Testing Equipment

1. Digital Vernier Calipers

Each One

No

2. Digital Ball Ended micrometer

3. Digital Thickness Gauge

4. Pi Tapes

5. Analytical Weighing Balance

6. Moisture Analyzer

7. Vacuum Oven with Pump

8. Hydrostatic Pressure testing System for Pipes

(a) Hydrostatic Pressure Tester

(b) Hot Water bath

(c) Low temperature Cabinet

(d) End fittings for Pipes

9. Constant Temperature Oil Bath for Reversion Test

10. Falling Weight Impact Tester for Pipe

11. Environmental Stress Cracking resistance Tester

12. Free Falling Dart Impact Tester with Built- In Vacuum

System, for Plastics Films Testing

13. Izod and Charpy Impact tester

14. Ultrasonicator

15. Compact Microprocessor-Controlled Bench-Top Laboratory

Fermenter-Bioreactor

16. Autoclave

17. Incubator

18. Optical Microscope with Programmable Hot Stage

19. FTIR with Microscope

20. Ultra Violet Visible Double Beam spectrophotometer

21. Universal Testing Machine

22. Gel Permeation Chromatography

23. CHN Analyzer

24. Horizontal Laminar air flow Specification

25. Water Vapour Transmission Rate

26. Gas Permeability Tester

PROCESSING MACHINERY

1. Extrusion Blown Film Unit with Carona Treater

Page 2: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 1

Digital Vernier Calipers

Technical Specifications

Range : 0 - 150 mm & 0 – 300 mm Accuracy : ± 0.01 mm Repeatability : ± 0.01 mm Display : LCD Construction material : Carbide-tipped jaws for outside and inside measurement Power Supply : Lithium battery Calibration certificate traceable to NIST should be provided

Page 3: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 2

Digital Ball Ended micrometer

Technical Specifications

Range : 0 - 25 mm Resolution : ± 0.001 mm Measuring surface (Ball) : 3.2 mm diameter Measuring force : 7 to 9 N Measuring system : Electromagnetic induction type sensor Power supply : Lithium battery (CR2032) x 1 Calibration certificate traceable to NIST should be provided

Page 4: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 3

Digital Thickness Gauge

Technical Specifications

Measuring Range : 0 – 10 mm / 0.4 inch Resolution : 0.001 mm / 0.0001 inch Depth of Jaw : 50 mm Display : LCD Measuring force : < 1.4 N Lifting device : Operating by lifting device – self closing Power Supply : Lithium battery Calibration certificate traceable to NIST should be provided

Page 5: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 4

Pie Tapes

Technical Specifications

Size : 20-300, 300-700, 700-1100 mm L.C : 0.1 mm Construction material : High grade of SS material Calibration certificate traceable to NIST should be furnished

Page 6: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 5

Analytical Weighing Balance

Technical Specifications

Capacity : 220 g Accuracy : 0.1 mg Display : Backlit display Size : 20 (W) x 35 (H) x 30 (D) cm Repeatability : 0.1 mg Power Supply : 230 V, 50 hz Pan Size : 8 – 10 cm Weighing units : mg, g, & kg Stabilization time : 2 sec Stability Filters : Three filters to match weighing environment Output : RS 232 Accessories : Calibration Weights & Density measurement kit

Page 7: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 6

Moisture Analyzer

Technical Specifications

Sample weight : 0.01 g to 200 g

Reproducibility : ± 0.05 mg for (1 mg water content) and ±0,5 mg for

(20 mg water content)

Display via PC : Absolute value of water content of sample in mg

Measuring range : 0.2 mg - 30 mg

Test temperatures : 50°C up to 210 °C in 1°C-steps

Ambient air temperature : in °C in 0.1 °C steps

Ambient air pressure : 700 - 1100 mbar in steps of 0.1 mbar

Relative air humidity : 0 - 100 % in steps of 0.1 %

Power supply : 230 V, 50 Hz

Dimensions : 285 mm (H) x 170 mm (W) x 250 mm (D)

Interface : Bi-directional RS 232 interface

Required PC : Windows 98, XP; Pentium 233 or equivalent; RAM:128 MB,

serial interface RS 232

Page 8: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 7

Vacuum Oven with Pump

Technical Specifications

Capacity (Vacuum Pump) : 50 Litres/min

Heating Load : 2 KW

Power Supply : 230V / 50 Hz AC

Size (inner Chamber) : 12 inch Diameter X 15 inch Depth

Construction : Inside – SS 304, Outside - MS Body with Powder

coating with proper insulation. Vacuum pump with

0.5 HP, double stage, shutter with glass window

and heat resistant (silicone) Gasket.

Working range : 0 to 200°C

Temperature Accuracy : ± 2°C

Temperature controller : PID action digital temperature controller

Page 9: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 8: Hydrostatic Pressure testing System for Pipes

S. No. 8 (a)

Hydrostatic Pressure Tester

Technical Specifications

Variable Working pressure Range Digital type : 0-10MPa Display Regulating accuracy : 0.001MPa Testing station : 5 Stations (for Acceptance & Type test) : [2 for acceptance (0-10 MPa) & : 3 for Type test (0-4 MPa)] This instrument must be suitable to carryout Hydrostatic Pressure Test as per: IS 4984 & 4985 Suitable pump has to be provided to developing water pressure as required for Acceptance and Type test separately. Fine Filter has to be provided for water inlet Timer has to be provided to record the starting time, Time of failure, over pressure time, pressure compensating time etc.

Page 10: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 8 (b)

Hot Water bath

Technical Specifications

Size (LXWXH) (Inner Dimension) : 1m X 1m X 1.2m Temperature control system : Ambient to 100⁰C

Accuracy : ± 1⁰C (Throughout the bath)

Should conform to the requirements of standard IS 4984.

In built water circulation pump. Well insulated SS body and top cover. Inside the water bath SS stand / grid should be provided.

Page 11: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 8 (c)

Low temperature Cabinet

Technical Specifications

Chamber size inside : 750 X 500 x 900 mm (LxWxH) No of Racks : 3 Racks Chamber material : SS 304 Door : Top opening single door with proper insulation

Temperature range : -10°c to 30°c

Accuracy : ± 0.1⁰C

Page 12: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 8 (d)

End fittings for Pipes

Technical Specifications Sizes (mm) : 12,16,20, 25, 32, 40, 50, 63, 75, 90, 110 - Each 2 sets

125, 140, 160, 180, 200, 225, 250, 280,300 & 355 - Each one set

Fitting must be suitable for UPVC/HDPE Pipes as per IS 4985/IS4984 respectively Should be light weight End Fitting Material : Non rusting metal ‘O’ ring material should be silicone Bolts/ Nuts/ Allen screws/ connection Nipples must be stainless steel / copper / brass.

Page 13: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 9

Constant Temperature Oil Bath for Reversion Test

Technical Specifications

Temperature Range : Ambient to 200°C

Temperature Controller : PID, LED Display

Temperature Accuracy : ± 1°C

Internal Dimensions (LxWxH) : 600 mm x 600 mm x 450 mm

Body Construction

: Inner SS & Outer MS Powder

Coated, double wall

construction with good quality

insulation with lid.

Conform to the standards :

: IS : 4985, IS : 12235,

IS:13592, IS:15328, IS:4984

with latest amendments.

Page 14: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 10

Falling Weight Impact Tester for Pipe

Technical Specifications Standards : IS 4985, IS 12818, IS 13592, IS 12235 Falling weight testing machine : Main frame with guide rail or tube fixed in a true vertical

Position Release Mechanism : Release mechanism with free fall from variable height up

to at least 2 Meter with an accuracy o ± 10 mm Max. Fall Height : 2050mm ± 10 mm With calibrated scale (preferable) Mass of the Striker (kgs) : 0,25, 0.5,0.8,1.0,1.25,1.6,2.0,2.5,3.2,4.0,5.0,6.3 Specimen holder for pipes : V Block with adjustable height. :’V’ Block base size 300 x 300 mm : ‘V’Block base thickness 25 mm : ‘V’ Angle : 120⁰

Page 15: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 11

Environmental Stress Cracking resistance Tester

Technical Specifications

Constant temperature bath with S S lid.

: Inside-SS, outside –MS with heat insulation Lid: S S.

Heating System : Heater with digital temperature controller (PID) Temperature Thermometer : RT to 100°C. Accuracy : ±0.10°C To maintain uniformity in temperature throughout the bath

: Stirrer should be provide.

Blanking die for cutting specimen

: 38±2.5 mm by 13±0.8mm with ejector

Nicking Jig of lever arm with dial gauge

: 0-10 mm

Notch depth adjustment : 0.30, 0.40, 0.50, 0.65 mm with notch length 19 mm

Bending cum transferring device : 1 No.

Specimen Holder : S S 10 Nos. Test Tubes (Glass) : 32 mm dia and 200 mm length – 10 Nos Corks fit to the glass tubes : 10 Nos. Interchangeable notching blades : 5 Nos. Thermometer : 1)50°C (20 to 60°C range)

2)100°C (60 to 105°C) Conform to : ASTM D 1693

Page 16: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 12

Free Falling Dart Impact Tester with Built- In Vacuum System, for Plastics

Films Testing

Technical Specifications

Clamp : Inside diameter of 127 mm, vacuum operated clamps with

rubber gasket on both side. (Lower/Upper)

Dart holding device : Electromagnetic releasing device

Masses : SS masses of diameter 30 mm with a centre hole of

diameter 6.6 mm and height varying from 2.5 to 25 mm

(5g,10g,15g,60g - each 4 Nos and 90g - 2 Nos)

Dart : Hemispherical with diameter of 38.1 mm made out of

aluminum

Aluminum Shaft : Length of 115 mm and diameter of 6.4 mm. Tip of length

12.7 mm is made out of steel

Weight of dart including

shaft : 100.0 gm

Reference standard : ASTM D 1709, Method A & B, IS 2508.

S. No. 13

Izod and Charpy Impact tester

Technical Specifications Model : Digital Model Confirms to Standard : ASTM D 256 & ASTM D 6110, ISO 180 & ISO 179

Page 17: CBPST, Kochi Testing Technical Specifications.pdf

Pendulum Impact Energy : 0-25 Joules Least Count of Indication : 0.1 Joules Release Angle of Pendulum : 150⁰ Striking velocity range of Pendulum : 2.5 – 3.5 m/s. Accessories : Standard detachable weights corresponding to (2J, 4J, 7J,

15J,and 25J)to obtain the desired test range. Standard fixtures for fitting Izod /Charpy test and other standard accessories.

Auto Notch Cutter & Sample : Suitable for notching plastics specimens for Izod & Mounting accessories Charpy impact tests. As per ASTM D 256, D 6110 and ISO

180, ISO 179 Power : 230V, 50Hz, Single phase Range of 4 seaters : 0-2.71J Resolution 0.02J 0-5.42J, Resolution 0.05J 0-10.84J Resolution 0.10J 0-21.68J Resolution 0.20J 0-25.0 Resolution 0.20J

Page 18: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 14

Ultrasonicator

Technical Specifications Frequency : 20 kHz Processing volume : 200 µL to 2000 mL Probe Tip diameter : Max. 25.4 mm Probe material : Ti-Al Alloy, autoclave Power output : 0 – 300 Watts Timer : 1 second to 24 hrs Pulser : 1 sec. To 99 hrs Temperature controller : 0 - 100°C

Temperature Accuracy : 1°C

Display : 145 mm LCD

Page 19: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 15

Compact Microprocessor-Controlled Bench-Top Laboratory Fermenter-

Bioreactor

Technical Specifications Power Supply : 230 V /50-60 Hz Operating Temperature : 0 – 60°C

Relative humidity : 0 – 100% Display : Backlit LCD Dimension : 22 cm (W) x 38 cm (H) x 40 cm (D) Working volume : 35 mL to 5 L Vessel Material : Pyrex borosilicate glass, 6 or 9 threaded necks Mixing : Mixer with the speed of 0 – 1200 rpm. Temperature control : +5°C to 60°C Temperature accuracy : 0.1°C Precision : +/- 0.2°C Thermo sensing : High precision Pt 100 built in the glass electrode bulb pH control : Sterilizable glass electrode, range 0 – 13 pH Resolution : 0.01 pH with automatic temperature compensation Pump : High quality pumps with flow rates 0 – 10 Litre/hr. Pump connection : Upto 4 pumps with progressive proportional flow rate control O2 measuring regulation : 0 – 25 mg/lt. Probe: PEEK Membrane: Glass reinforced PTFE Resolution: 0.1 mg/ml Gas input : Max. air pressure 0.2 MPa Airflow meter : Range: 0 – 5 lt./min. Resolution: 0.1 lt./min. Reproducibility: ± 0.5% Remote control & Data Processing : Software for fermenters

Page 20: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 16

Autoclave

Technical Specifications

Chamber Size : 500 mm (W) x 500 mm (L) x 950mm (H) Electrical Power : 8 KW Or Sufficient wattage of industrial immersion type water heater to generate steam within a reasonable period of time on 3 phase 440V/ 50 Hz AC supply Working pressure : 1.2 to 2.2 Kg/sq.cm Working Temperature : at 121°C

Material of Construction : Inner chamber, Jacket, Door : SS 316.(5mm-10mm) Outer Chamber: SS 304 (Insulated properly)

Steam Generator : Non corrosive SS / Chromium plated Brass Heater Plate : Brass / Stainless Steel Pipe Line : Complete with SS Stand : Stainless Steel / High quality non corrosive steel Safety devices : Pressure switch and safety valve, self locking of door when chamber is under pressure

Vacuum breaker for jacket Steam generator with gauge glass valves and Low water protection with audio visual indicator.

Features : The equipment should have ISI mark on quality. Temperature, Pressure and Vacuum gauges : Steam

traps, vacuum driers, water level indicator on steam generator

Page 21: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 17

Incubator

Technical Specifications

Volume : 85 Litres Temperature Range : 5 - 60°C Temperature Accuracy : ± 0.5°C

Working space : Width – Min. 45.0 cm Depth – Min. 40.0 cm Height – Min. 55.0 cm External material : MS with powder coating with proper insulation No. of SS perforated Shelves : 02 Voltage Stabilizer : 1 kVA Chamber internal material : SS 304

Page 22: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 18

Optical Microscope with Hot Stage

Technical Specifications

Microscope Mains Operations : 220 Volts,50Hzs,Single Phase AC supply. Screen : 360° Rotatable Graduated Screen Light Control : Variable Controlled Light Intensity Meter Objective : 5x, 10x, 20x & 40x Magnification : 125x to 1000x.

Hot stage Temperature range : 50 - 400°C Tempearture accuracy : 0.1°C Heating rate : Min. 0.1°C / hr and Max. 100°C / min.

Temperature control method : PID controller Temperature sensor : 100 Ω Platinum RTD Min. Objective Working Distance : 6.1 mm Sample Area : 22 mm x 25 mm Chamber height : 5.0 mm Sample Viewing Aperture : 5 – 7 mm diameter

Page 23: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 19

FTIR with Microscope

Technical Specifications

Wave number : Minimum 7800-350 cmˉ¹ Signal to Noise Ratio : 6000:1 Wave number Accuracy : 0.02 cmˉ¹ Resolution : Minimum 0.5 cmˉ¹ Detector : FR-DTGS or LiTaO3 or MCT array detector with KBr

Beam splitter in addition to micro ATR objective FTIR Software : Window based for data manipulation and analysis and

Polymer data library. Atmosphere compensation : Automatic Atmospheric compensation Validation Standards : Automatic Validation including certified standards Instrument Calibration : Traceable standard material Humidity sensors : Physical as well as electronic humidity sensor Sample holder : Holders suitable for powder / solid samples should be

provided Accessories : Automatic optimization and calibration, Suitable work station : GC compatible : Search facility in library for plastics and additives Microscope : Microscope should have visible and infrared mode

Fully automated with motorised sample stage Integral CCD video camera and video capture card for PC Auto aperture, One touch auto focus and mapping Auto marker and auto illumination Auto view and Auto correction Simultaneous variable magnification, ATR Mapping IR image display as pixels images displays Upgradeability to IR imaging,

Signal to noise ratio for microscope : 6000:1 mrs Polarizer : Visible and IR polarizer Auto mode : Transmission, Reflectance and micro ATR

Page 24: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 20

Ultra Violet Visible Double Beam spectrophotometer

Technical Specifications Wavelength : 190-1100 nm Spectral Bandwidth : 1 nm (190 -1100 nm) Wavelength Display : 0.1 nm increments Wavelength Setting : 0.1 nm increments Wavelength accuracy : ± 0.1 nm at 656.1 nm D2

± 0.3 nm (190 – 110 nm) Scanning speed : Fast / medium / slow Wavelength repeatability : ± 0.1 nm Stray light : < 0.02% at 220 nm and 340 nm

< 1.0% at 198 nm Photometric System : Double beam Photometric range : Absorbance : 0-4 to 4 : Transmittance : 0 to 400% Photometric accuracy : ± 0.002 abs (0.5 abs)

± 0.004 abs (1.0 abs) ± 0.006 abs (2.0 abs)

Photometric repeatability : < ± 0.001 abs (0.5 abs) < ± 0.001 abs (1 abs) < ± 0.003 abs (2.0 abs)

Baseline stability : < 0.0003 abs/hr at 700 mm(one hr after light source turned ON)

Baseline flatness : Within ± 0.0006 abs (190 nm – 1100 nm, one hr after light source turned)

Noise Level : Within 0.00005 abs RMS value (at 700 nm) Printers : DPU,ESC/P, PCL Printers, USB I/F windows –

complaint printers are available with USB memory and PC software

Memory : USB memory saved as text and UVPC file Performance of PC : USB Memory + UV probe WinXP Detector : Dual silicon photodiode Power : AC 230 V / 50 HZ Sample holders : Square cell with stopper and cylindrical cell

made of fused silica or quartz with temperature controller

Software : Analysis software should be provided

Page 25: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 21

Universal Testing Machine

Technical Specifications

Maximum Load Capacity : 50 KN (Microprocessor controlled) Cross head Travel distance : Min 1000 mm Horizontal daylight : Min. 500mm Test Speed : Variable between 0.05 mm / min to 1000 mm / min. Load cells : 100 N, 1 KN, 5 KN & 50 KN Load Accuracy : ≤ 0.5 %, Fixtures : Tensile, compression, flexural, and shear fixtures. Tensile grips : Suitable to Plastic films (rubber faced grips), Rigid plastics (self lock winch grip, opening up to 12mm) woven sacks (50mm width), rubber, fibre/filament. Power : 230 Volts, 50 HZ, single phase. Conform to : ASTM, ISO, DIN, BS and IS standards. Extensometer : Contact Strain guage, Output : Software attached & data storage for sample test

methods Grips : Neumatic and Manual Accuracy for Cross head speed : ± 0.1 mm Safety lock provisions Software : Suitable for all functions such as max. load; strain,

modulus, etc

Calibration certificate for load cells traceable to NIST should be provided

Page 26: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 22 High Temperature Gel Permeation Chromatograph

Technical Specifications

GPC Pump

Flow rate Range: 0.001 to 5 ml / min.

Flow precision: <0.07 % RSD

Operating pressure: Upto ~ 5800 psi

Flow rate accuracy: ± 1%

Leak sensors Provided. CAN, LAN, HP-1B, RS 232, APG

remote – start are interfaced to communicated the modules with

computer

Vacuum solvent degassing module is Provided.

GPC injector

Capacity: 30 vials

Operating temperature: 220°C.

GPC soft ware

PC controlled soft ware to control and monitor all functions

including pump, detector, Oven, autosampler and solvent vapour

sensors

Detector

a. UV-Visible Variable Wavelength detector

Detector type: Monochromator (UV detector)

Wavelength range: 190 -0740 nm

Wavelength accuracy: ± 3 nm

Sources: Deutirium

Noise: < 20 x 10-5

AU

Drift: 1.5 x 10-4

AU/h

Time constants: 0.1 / 0.2 / 0.5 / 1 / 2 / 5 / 10

Path length: 10 mm

b. Refractive index detector

RI range: 1 – 1.75 RIU

Cell volume: 10µl

Sensitivity: < 2 x 10-7

RIU / mV

Noise: < 3 x 10-8

RIU / hr.

Page 27: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 23

CHN Analyzer

Technical Specifications

Operating Principle: Based on classical technique of Pregi & Dumas. The system includes patented Pressurised Oxygen Injection System (TurboFlash Technology) for perfecting combustion in widest variety of samples matrices. Operating Mode: simultaneous (CHNS); (CNS); (S); (O); (CHN); (CN); (N). Full field up gradability from one operating mode to another through full modularity. Detection Principle: Separation of combustion product on GC Column followed by detection using Supersensitive Thermal Conductivity detector Instrument Features: Oven temperature range : From 40°c to 190°C Furnace Temperature : Up to 1100°C for all operating mode Combustion temperature : Up to 1800 °C Oxygen injection (volume) : Volume set via S/W Oxygen injection (pressure) : Set Independently from Carrier velocity via S/W Gas leak test : Automatic by keypad push button or via S/W Gas flow rate in stand-by : Zero flow in Gas Save Gas save in stand-by : Automatically selectable by keypad or S/W Catalyst & TCD protection : Automatic isolation with gas save routine Exhausted reactor change : Bayonet-lock action from outside of the instrument Sample introduction : 80 sample via electrically driven AutoSampler Sample Type : Solid, Viscous and liquid SW Instrument Control : Fully automatic control of all parameters and results reporting by user selected format. Set up guide for Windows XP/ME SW : Real time chromatogram display. Storing of an unlimited methods. Ability to reprocess individual runs and batches of samples. AutoRun & AutoCal. AutoReady & Wake up routine. Automatic Leak test. Export on Excel Calibration : Traditional K-factor or linear regression Safety : Full fail-safe operation. Sensors for high temperature compartment and transformer. Dual redundant control valving for Oxygen delivery circuit Shut-off safety interlock for users accessing the hot compartment

Page 28: CBPST, Kochi Testing Technical Specifications.pdf

Instruments Performances: Practical Working Range 0,01% up to 100% Precision: Using Cystine as Reference Std Material and Sartorius Mod CP2P, MicroBalance, weighing 5-6 standard within 0,5 mg up to 1,5 for linear, regression calibration the typical results obtained are From 0,20% or up to 0,30% depending on Element % content Speed of Analysis : CHNS/CNS = 5 min; CHN = 3 minutes; for Oxygen = 2 min

S. No. 24

Horizontal Laminar air flow Specification

Technical Specifications Air Velocity : Between 0.4 to 0.5 mls +-20% HEPA Efficiency Pre-filter : 99.997% at 0.3 microns with DOP test Washable with an arrestance up to 90% at 5 microns Less than 65 dB (A) Sound Level : Lux 900 - 1300 Lux Light Intensity Cleanliness : Standard 209 E Class 100 Federal Power Supply : 220 V Single phase 50 Hz

Others:

S. S Table Top.

U. V Lights attachment

Front Acrylic Door

Gas/ Vacuum Cock

Castor wheels at the Bottom

Construction:

The cabinets are fabricated out of thick board duly sunmica clad or mild steel cabinet duly epoxy coated.

The work table is made of thick board which is sunmica clad at top or S. S Table top will be provided as

optional accessories. All joints are sealed with silastic sealant. Side panels are provided of thick

transparent acrylic sheets. The unit is fitted with pre-filter and is made to pass through highly effective

HEPA (High Efficiency Particular Air) filter having efficiency rating as high as 99.997% with hot, DOP,

thus retaining all air-borne particles of size : 0.3 micron and larger.

Using a dynamic machine, the blower and motor assembly is statically and dynamically balanced.

Blowers are mounted on vibration isolation mounts to minimize vibration. The working is illuminated by

fluorescent lightings fitted to the unit. Height of the working table provides comfortable “ Sit Down”

working position for the operation to work on 220/230 volts AC. supply.

Page 29: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 25

Water vapour Transmission Tester

Technical Specifications

Measuring range : 0.03 – 10000 g/m2 /day

Measuring accuracy : 0.001 g/m2 /day

Sample surface area : 50cm2

Minimum sample size : 100 x 100mm

Thickness of sample : 0.01 – 12mm

Voltage settings : 230 V AC / 50 Hz

Power consumption : Max. 300 VA

Environmental temperature control : 5-40°C

Temperature control accuracy : ± 0.5°C

Communication interface : RS 232 – Interface

Warning : Hot surface (over 70°c)

Equilibrium Detector : 0 – 99.9%

Reproducibility : High permeability samples : better

than 1%

Low permeability samples better than

5%

Humidity measuring range : 40 – 95% RH in measuring chamber

(std : 90%)

Humidity accuracy : ± 1%

Keyboard : Alphanumeric

Data storage by Lithium battery : Typically >10 years

Typical time consumption for a simple

measurement

Range 1g / m2 / 24 hrs

20g / m2 / 24 hrs

>1000 g/m2/ 24 hrs

:

:

:

:

< 1 hour

< 10 minutes

< 5 minutes

Conforms to : ASTM E96, ASTM D 1653, ISO

2528, DIN 53122

Page 30: CBPST, Kochi Testing Technical Specifications.pdf

S. No. 26

Gas Permeability Tester

Technical Specifications

Measuring range : 0.02- 50000 cc/m² day

Measurement accuracy : 0.01 cc/ m² day

Temperature range : 15⁰C - 45⁰C (std: 23⁰C)

Temperature accuracy : ± 0.01⁰C

Test pressure : 0.1 pa

Suitable for : O2, CO2, N2 gas operation

Air pressure : 0.1 MPa – 0.8 MPa

Specimen size : Diameter - 80 mm, Area: 50 cm²

Sample : 3 (extension up to more pieces)

Carrier gas interface (inlet) : 1/8 inch

Sample thickness : ≤ 2 mm

Power : AC 220 V / 50 HZ

Humidity control : 0%, 35-90%, 100% RH

Humidity control accuracy : ± 3% RH

Interface : RS-232 interface

Data collection & storage : Windows based software suitable for data

collection, storage and analysis.

Confirms to Standard : ASTM D 1434, IS :13360, ISO 2556

Page 31: CBPST, Kochi Testing Technical Specifications.pdf

PROCESSING MACHINERY

EXTRUSION BLOWN FILM UNIT WITH CARONA TREATER

EXTRUSION BLOWN FILM UNIT WITH CARONA TREATER

S. No Items Unit Specifications Model :

1. Screw Diameter mm 30-40

2. Screw Speed rpm Please specify

3. Screw L/D Ratio - Please specify

4. Extruder Output Kg/hr 10 - 35

5. Film Thickness mm/Micron Please specify

6. Material to be process - All Thermoplastics

7. LFW (Min & Max.) mm Please specify

8. Control System - Microprocessor

9. Synchronisation of Extruder drive with Haul off unit etc

- Please specify

10. Melt temperature sensor & Melt Pressure transducers

- Please specify & quote

11. Die for HMHDPE & LLDPE - Please specify & quote

12. Die gap mm Please specify

13. Corona Treater suitable for above machine

- Please specify & quote

14. Safety

- Appropriate safety features to be provided

15. Machine dimensions (LxWxH) - Please specify

16.

Essential Spares a. Temperature controller- 02 nos. b. Thermocouples- 02 nos. c. Heaters- 01 set d. Synchronisation card -01 nos.

-

Please Specify & Quote

17. Optional a) Pipe/tube attachment b) Pelletizing attachment

- -

Please specify & quote Please specify & quote

Page 32: CBPST, Kochi Testing Technical Specifications.pdf