7
Krypton® Multi For measuring disinfectants, pH and temperature - optional ORP and 5th measuring input. subjects to alterations - version 07/2015 Online measuring board for measuring disinfectants, pH and temperature - optional ORP and 5th selectable measuring input Worldwide access by internal webserver (LAN or WLAN) Krypton® Multi Applications Disinfection Process Water Drinking Water / Beverages Description The measuring system Krypton® Multi includes all you need for disinfection and pH measurement: instrument, sensors, assembly and wiring. The system is used for measuring and control in industrial desinfectants applications like in breweries, water works or cooling towers. The integrated modular assembly Argon® Stabiflow ensures a constant flow of approx. 30 l/h. As a consequence the system provides stable, precise and reliable measuring values. Flow fluctuations not any longer shown up in your measuring curves as long as you guarantee a water inlet above 35 l/h. The entry level version of Neon® Multi is equipped with 3 measurements: disinfection, pH, and temperature. The disinfection measuring input can be configured to measure free Chlorine, Chlorine dioxide, Ozone or Hydrogen peroxide via menu. The measuring range is adjustable. An ORP measurement can be added and even a 5th measuring input for total Chlorine. The entry level version of Neon® Multi offers 6 digital inputs and 8 potential free CO relays as control or as alarm relay. The instruments offers PID as well as 3 point control function with and without position feedback. Neon® Multi is expandable through software upgrades or add on modules. It is possible to add up to five analog outputs, Modbus, automatic sensor cleaning ASR® and datalogger. The complete system is pressure resistant up to 6 bar (at 20°C) and brine resistant. 1 / 7

Multivariable liquid analysis system

Embed Size (px)

Citation preview

Page 1: Multivariable liquid analysis system

Krypton® MultiFor measuring disinfectants, pH and temperature - optionalORP and 5th measuring input.

subjects to alterations - version 07/2015

Online measuring board for measuring disinfectants, pH and temperature - optional ORPand 5th selectable measuring inputWorldwide access by internal webserver (LAN or WLAN)

Krypton® Multi

Applications

Disinfection

Process Water

Drinking Water / Beverages

Description

The measuring system Krypton® Multi includes all you need for disinfection and pH measurement: instrument,

sensors, assembly and wiring. The system is used for measuring and control in industrial desinfectants applications

like in breweries, water works or cooling towers. The integrated modular assembly Argon® Stabiflow ensures a

constant flow of approx. 30 l/h. As a consequence the system provides stable, precise and reliable measuring

values. Flow fluctuations not any longer shown up in your measuring curves as long as you guarantee a water inlet

above 35 l/h. The entry level version of Neon® Multi is equipped with 3 measurements: disinfection, pH, and

temperature. The disinfection measuring input can be configured to measure free Chlorine, Chlorine dioxide, Ozone

or Hydrogen peroxide via menu. The measuring range is adjustable. An ORP measurement can be added and even

a 5th measuring input for total Chlorine. The entry level version of Neon® Multi offers 6 digital inputs and 8 potential

free CO relays as control or as alarm relay. The instruments offers PID as well as 3 point control function with and

without position feedback. Neon® Multi is expandable through software upgrades or add on modules. It is possible

to add up to five analog outputs, Modbus, automatic sensor cleaning ASR® and datalogger. The complete system is

pressure resistant up to 6 bar (at 20°C) and brine resistant.

1 / 7

Page 2: Multivariable liquid analysis system

Krypton® MultiFor measuring disinfectants, pH and temperature - optionalORP and 5th measuring input.

subjects to alterations - version 07/2015

Particular characteristics

85.. 265 V AC power supply90x50 mm touch displayReal time clockPressure proofed to 6 bar (at 20°C without TCl)Precise measurement by a constant flowUp to 3 separate PID controller for disinfectant measurement, pH and the 5th measuringinput (Total Chlorine)disinfection measurement or pH: 3-point-controller with or without position feedbackDosage checkUp to 4 potential-free alarm relaisLower and upper limit value with delay for each measuring parameter adjustable andassignable to the alarm relayAlerts selectable for alarm relayEventlog and event helpSafety by autolock function4 user level by password functionIndividual user codes are possibleIntuitive operation with touch screen and graphical menuFunctional upgrade by activation code or with modulesSD card function: firmware update, configuration and diagnosis fileTest menu for wiring checkScreensaver: Auto-shut-off of LCD backlightAutomativ pH-compensation for free Chlorine measurementAutomatic or manual temperature compensationDisplay of organic Chlorine by combination of total Chlorine and free ChlorinemeasurementOption: 5th measuring input fot total ChlorineOption: ASR® automatic sensor cleaningOption: 5 mA outputs 0/4 to 20 mA adjustable with fault currentOption: RS 485 Modbus RTUOption: data log function - storage of all measuring data up to one year and trendview ofstored data

2 / 7

Page 3: Multivariable liquid analysis system

Krypton® MultiFor measuring disinfectants, pH and temperature - optionalORP and 5th measuring input.

subjects to alterations - version 07/2015

Interface diagram

Technical data

Measuring parameter

Free Chlorine, Chlorine dioxide,

Ozone

0.. 1000 µg/l

0.. 5.00 mg/l

0.. 10.00 mg/l

Free Chlorine, Chlorine dioxide 0.. 20.00 mg/l

pH-value 0.. 14.00 pH

ORP -1500.. 1500 mV

Temperature 0.. 50 °C

Hydrogen Peroxide 0.. 30 mg/l

Total Chlorine 0.. 10 mg/l

3 / 7

Page 4: Multivariable liquid analysis system

Krypton® MultiFor measuring disinfectants, pH and temperature - optionalORP and 5th measuring input.

subjects to alterations - version 07/2015

Input characteristics

Accuracy +/-2% of measuring range limit

Temperature measuring range 0.. 50°C

Temperature coefficient 0,0.. 8,0 %/K adjustable

Digital input Flow control, external controller stop, 2x level control, activation 2nd control

parameter set

Ambient conditions pH-range 6.. 8 pH (free Chlorine) 6.. 9 pH

(Chlorine dioxide, Ozone, Hydrogen

peroxide) total Chlorine depending on

sensor

Flow >20 l/h, constant

Min. conductivity > 150 S/cm

Temperature Depending on sensor, assembly and

reference measurement

Output characteristics

Alarm relay up to 4 potentialfree CO, max. 250 V; 2 A, 550 VA

Output signal Optional: 5 x 0/4.. 20 mA (scaleable, galvanically isolated)

Load 500 Ohm

Registration range Scaleable within the measuring range

Storage media Accessory: SD card up to 2 GB - Industry standard

Baud rate 19200 kbs (Modbus)

Data format 8 bit

Power supply

Line voltage 85.. 265 V AC / DC, 50.. 60 Hz

Power consumption 10 VA

Ambient conditions

Temperature Storage -20.. +65°C, exception sensor 0..

+30°C

Operation 0.. +50°C

Humidity max. 90% rH at 40°C (non-condensing)

Protection class Wall mounted IP 65

Controller

Control mode on/off controller (adjustable hysteresis) P/PI/PID controller (pulse-pause,

pulse-frequency or continous output) 3-point controller with or without position

feedback, coagulation dosing optional)

Relay 8 relays, each a potential-free CO contact, max. 250 V, 2A, 550 VA

Start delay 0.. 200 sec till controller active

Controller stop Digital input

4 / 7

Page 5: Multivariable liquid analysis system

Krypton® MultiFor measuring disinfectants, pH and temperature - optionalORP and 5th measuring input.

subjects to alterations - version 07/2015

Design configuration

Material Board: PVC, assembly: PMMA, cocks: PP, PVC, instrument: ABS, sensors:

glass

Dimensions 400 x 700 mm

Connection cable inlet: 2x M16, 2x M12 + optional: 2x M12 und 1x M25 plug-in terminal:

rigid/ flexible 0,14 - 1,5 mm² relays / power supply - starr/flexibel 0,2-1 /

0,2-1,5 mm² distribution block 0,5-2,5/ 0,5-2,5 mm²

Mechanical drawing

5 / 7

Page 6: Multivariable liquid analysis system

Krypton® MultiFor measuring disinfectants, pH and temperature - optionalORP and 5th measuring input.

subjects to alterations - version 07/2015

Order information

Choose the components you need and that´s how your "assembly version" is designed. We will have to technically

inspect and approve a free combination of individual key features.

6 / 7

Page 7: Multivariable liquid analysis system

Krypton® MultiFor measuring disinfectants, pH and temperature - optionalORP and 5th measuring input.

subjects to alterations - version 07/2015

Accessorieshand-held units

Radon Des- pH The Radon photometer is used for measuring of

Chlorine, Chlorine dioxide, Ozone, Hydrogen peroxide

and pH. Accurate and consistent results are obtained

quickly. Delivery contents is everything neede to

measure Chlorine, Chlorine dioxide and Ozone. The

delivery content conains evrything needed for

measuring Chlorine, Chlorine dioxide and Ozone. For

measurement of Hydrogen peroxide you need a

H2O2-addon and for the pH measurement the

reagent phenolred.

accessories

Test Plug DES With the test plug you can check the function and

connection of our disinfection measurement - not for

Total Chlorine measurement (not TCl).

Powered by TCPDF (www.tcpdf.org)