Upload
drpraveen1986
View
3.302
Download
5
Embed Size (px)
Citation preview
PACEMAKER BASICSPACEMAKER BASICS
Dr. R. Praveen Babu
Vijaya Hospital
Chennai
Implantable pulse generator (IPG)
Lead wire(s)
Implantable Pacemaker Systems Implantable Pacemaker Systems Contain the Following Components:Contain the Following Components:
Pulse generator: power source or battery
Leads or wires
Cathode (negative electrode)
Anode (positive electrode)
Body tissue
IPG
Lead
Anode
Cathode
Pacemaker Components Combine with Pacemaker Components Combine with Body Tissue to Form a Complete CircuitBody Tissue to Form a Complete Circuit
Contains a battery that provides the energy for sending electrical impulses to the heart
Houses the circuitry that controls pacemaker operations
Circuitry
Battery
The Pulse Generator:The Pulse Generator:
Deliver electrical impulses from the pulse generator to the heart
Sense cardiac depolarization
Lead
Leads Are Insulated Wires That:Leads Are Insulated Wires That:
Begins in the pulse generator
Flows through the lead and the cathode (–)
Stimulates the heart
Returns to the anode (+)
During Pacing, the Impulse:During Pacing, the Impulse:
Impulse onset*
Flows through the tip electrode (cathode)
Stimulates the heart
Returns through body fluid and tissue to the IPG (anode)
A Unipolar Pacing System Contains a Lead with Only One A Unipolar Pacing System Contains a Lead with Only One Electrode Within the Heart; In This System, the Impulse:Electrode Within the Heart; In This System, the Impulse:
Cathode
Anode
-
+
Anode
Flows through the tip electrode located at the end of the lead wire
Stimulates the heart
Returns to the ring electrode above the lead tip
A Bipolar Pacing System Contains a Lead with Two A Bipolar Pacing System Contains a Lead with Two Electrodes Within the Heart. In This System, the Impulse:Electrodes Within the Heart. In This System, the Impulse:
Cathode
Stimulate cardiac depolarization
Sense intrinsic cardiac function
Respond to increased metabolic demand by providing rate responsive pacing
Provide diagnostic information stored by the pacemaker
Most Pacemakers Perform Four Functions:Most Pacemakers Perform Four Functions:
10
CaptureDepolarization of atria and/or ventricles in
response to a pacing stimulus
11
SensingAbility of device to detect intrinsic cardiac
activityUndersensing: failure to senseOversensing: too sensitive to activity
Rate Responsive PacingRate Responsive Pacing
When the need for oxygenated blood increases, the pacemaker ensures that the heart rate increases to provide additional cardiac output
Adjusting Heart Rate to Activity
Normal Heart Rate
Rate Responsive PacingFixed-Rate Pacing
Daily Activities
A Variety of Rate Response Sensors ExistA Variety of Rate Response Sensors Exist
Those most accepted in the market place are:
– Activity sensors that detect physical movement and increase the rate according to the level of activity
– Minute ventilation sensors that measure the change in respiration rate and tidal volume via transthoracic impedance readings
14
Types
1. Asynchronous/Fixed Rate
2. Synchronous/Demand
3. Single/Dual ChamberSequential (A & V)
4. Programmable/nonprogrammable
Single-Chamber SystemSingle-Chamber System
The pacing lead is implanted in the atrium or ventricle, depending on the chamber to be paced and sensed
DisadvantagesDisadvantagesAdvantagesAdvantages
Advantages and Disadvantages of Advantages and Disadvantages of Single-Chamber Pacing SystemsSingle-Chamber Pacing Systems
Implantation of a single lead
Single ventricular lead does not provide AV synchrony
Single atrial lead does not provide ventricular backup if A-to-V conduction is lost
One lead implanted in the atrium
One lead implanted in the ventricle
Dual-Chamber Systems Have Two Leads:Dual-Chamber Systems Have Two Leads:
Benefits of Dual Chamber PacingBenefits of Dual Chamber Pacing
Provides AV synchrony
Lower incidence of atrial fibrillation
Lower risk of systemic embolism and stroke
Lower incidence of new congestive heart failure
Lower mortality and higher survival rates
Benefits of Dual-Chamber PacingBenefits of Dual-Chamber Pacing
Study ResultsHigano et al. 1990
Gallik et al. 1994
Santini et al. 1991
Rosenqvist et al. 1991
Sulke et al. 1992
Improved cardiac index during low levelexercise (where most patient activity occurs)
Increase in LV filling
30% increase in resting cardiac output
Decrease in pulmonary wedge pressure
Increase in resting cardiac output
Increase in resting cardiac output, especiallyin patients with poor LV function
Decreased incidence of mitral and tricuspidvalve regurgitation
21
Examples
VVIV: Ventricle is the paced chamberV: Ventricle is the sensed chamberI: Inhibited response to a sensed signal
Thus, a synchronous generator that paces and senses in the ventricle
Inhibited if a sinus or escape beat occursCalled a “demand” pacer
22
ExamplesDVI
D: Both atrium and ventricle are pacedV: Ventricle is sensedI: Response is inhibited to a sensed
ventricular signal
Examples
DDDRA
Dual chamber, adaptive-rate pacing with multisite atrial pacing (i.e., biatrial pacing, more than one pacing site in one atrium,or both features)
23
Stimulation ProcessStimulation Process
Time (Milliseconds)100 200 300 400 500
Phase 2
Phase 1
Phase 3
Phase 4
Tran
smem
bran
e Po
tent
ial
(Mill
ivol
ts)
-50
0
50
-100
Phas
e 0
Threshold
Stimulation ThresholdStimulation Threshold
The minimum electrical stimulus needed to consistently capture the heart outside of the heart’s refractory period
VVI / 60
Capture Non-Capture
Amplitude
Pulse width
Two Settings Are Used to Ensure Capture:Two Settings Are Used to Ensure Capture:
Amplitude is the Amount of Voltage Amplitude is the Amount of Voltage Delivered to the Heart By the PacemakerDelivered to the Heart By the Pacemaker
Amplitude reflects the strength or height of the impulse:
– The amplitude of the impulse must be large enough to cause depolarization ( i.e., to “capture” the heart)
– The amplitude of the impulse must be sufficient to provide an appropriate pacing safety margin
Pulse Width Is the Time (Duration) Pulse Width Is the Time (Duration) of the Pacing Pulseof the Pacing Pulse
Pulse width is expressed in milliseconds (ms)
The pulse width is just the length of time each pacing pulse is delivered & must be long enough for depolarization to disperse to the surrounding tissue
5 V
0.5 ms 0.25 ms 1.0 ms
The Strength-Duration CurveThe Strength-Duration Curve
The strength-duration curve illustrates the relationship of amplitude and pulse width
– Values on or above the curve will result in capture
DurationPulse Width (ms)
.50
1.0
1.5
2.0
.25St
imul
atio
n Th
resh
old
(Vol
ts)
0.5 1.0 1.5
Capture
Strength-Duration Curve Strength-Duration Curve
SDCSDC
Rheobase- (the lowest point on the curve) by definition is the lowest voltage that results in myocardial depolarization at infinitely long pulse duration
Chronaxie(pulse duration time ) by definition, the chronaxie is the threshold pulse duration at twice the rheobase voltage
The ideal pulse duration should be greater than the chronaxie time
Cannot overcome high threshold exit block by increasing the pulse duration, If the voltage output remains less than the rheobase
SDCSDC
Lead impedance
Amplitude and pulse width setting
Percentage paced vs. intrinsic events
Rate responsive modes programmed “ON”
Factors That Affect Battery Factors That Affect Battery Longevity Include:Longevity Include:
ImpedanceImpedance
The opposition to current flow
In a pacing system, impedance is:
– Measured in ohms
– Represented by the letter “R” (for numerical values)
– The measurement of the sum of all resistance to the flow of current
Impedance Changes Affect Pacemaker Impedance Changes Affect Pacemaker Function and Battery LongevityFunction and Battery Longevity
High impedance reading reduces battery current drain and increases longevity
Low impedance reading increases battery current drain and decreases longevity
Impedance reading values range from 300 to 1,500
– High impedance leads will show impedance reading values greater than 1,500 ohms
ImpedanceImpedance
Factors that can influence impedance
– Resistance of the conductor coils
– Tissue between anode and cathode
– The electrode/myocardial interface
– Size of the electrode’s surface area
– Size and shape of the tip electrode
Ohm’s Law is a Fundamental Ohm’s Law is a Fundamental Principle of Pacing That:Principle of Pacing That:
VV
II RRV = I X RV = I X RI = V / RI = V / RR = V / IR = V / I
Describes the relationship between voltage, current, and resistance
xx
If you reduce the voltage by half, the current is also cut in half
If you reduce the impedance by half, the current doubles
If the impedance increases, the current decreases
When Using Ohm’s Law When Using Ohm’s Law You Will Find That:You Will Find That:
Voltage, Current, and Impedance Voltage, Current, and Impedance Are InterdependentAre Interdependent
The interrelationship of the three components can be likened to the flow of water through a hose
– Voltage represents the force with which . . .
– Current (water) is delivered through . . .
– A hose, or lead, where each component represents the total impedance:
The nozzle, representing the electrode
The tubing, representing the lead wire
Voltage and Current FlowVoltage and Current Flow
Spigot (voltage) turned up(high current drain)
Spigot (voltage) turned low(low current drain)
Resistance and Current FlowResistance and Current Flow
“Normal” resistance
“Low” resistance
“High” resistance Low current flow
High current flow
Electrode Design May Also Impact Electrode Design May Also Impact Stimulation ThresholdsStimulation Thresholds
Lead maturation process
Lead Maturation ProcessLead Maturation Process
Fibrotic “capsule” develops around the electrode following lead implantation
Lead Maturation Process
3 phases
1. A/c phase, where thresholds immediately following implant are low
2. Peaking phase- thresholds rise and reach their highest point(1wk) ,followed by a ↓ in the threshold over the next 6 to 8 wks as the tissue reaction subsides
3. C/c phase- thresholds at a level higher than that at implantation but less than the peak threshold
Trauma to cells surrounding the electrode→ edema and subsequent development of a fibrotic capsule.
Inexcitable capsule ↓ the current at the electrode interface, requiring more energy to capture the heart.
Steroid Eluting LeadsSteroid Eluting Leads
Steroid eluting leads reduce the inflammatory process and thus exhibit little to no acute stimulation threshold peaking and low chronic thresholds
Porous, platinized tipfor steroid elution
Silicone rubber plugcontaining steroid
Tines forstablefixation
Lead Maturation ProcessLead Maturation ProcessEffect of Steroid on Stimulation Thresholds
Pulse Width = 0.5 msec
03 6
Implant Time (Weeks)
Textured Metal Electrode
Smooth Metal Electrode
1
2
3
4
5
Steroid-Eluting Electrode
0 1 2 4 5 7 8 9 10 11 12
Vol
ts
A Pacemaker Must Be Able to Sense A Pacemaker Must Be Able to Sense and Respond to Cardiac Rhythmsand Respond to Cardiac Rhythms
Accurate sensing enables the pacemaker to determine whether or not the heart has created a beat on its own
The pacemaker is usually programmed to respond with a pacing impulse only when the heart fails to produce an intrinsic beat
Accurate Sensing...Accurate Sensing...
Ensures that undersensing will not occur –the pacemaker will not miss P or R waves that should have been sensed
Ensures that oversensing will not occur – the pacemaker will not mistake extra-cardiac activity for intrinsic cardiac events
Provides for proper timing of the pacing pulse – an appropriately sensed event resets the timing sequence of the pacemaker
Sensitivity – The Greater the Number, the Sensitivity – The Greater the Number, the LessLess Sensitive the Device to Intracardiac EventsSensitive the Device to Intracardiac Events
Accurate Sensing Requires That Accurate Sensing Requires That Extraneous Signals Be Filtered OutExtraneous Signals Be Filtered OutSensing amplifiers use filters that allow appropriate
sensing of P waves and R waves and reject inappropriate signals
Unwanted signals most commonly sensed are:
– T waves
– Far-field events (R waves sensed by the atrial channel)
– Skeletal myopotentials (e.g., pectoral muscle myopotentials)
Pacemaker TimingPacemaker Timing
Pacing Cycle : Time between two consecutive events in the ventricles (ventricular only pacing) or the atria (dual chamber pacing)
Timing Interval : Any portion of the Pacing Cycle that is significant to pacemaker operation e.g. AV Interval, Ventricular Refractory period
Single-Chamber TimingSingle-Chamber Timing
Single Chamber Timing TerminologySingle Chamber Timing Terminology
Lower rate
Refractory period
Blanking period
Upper rate
Lower Rate IntervalLower Rate Interval
Lower Rate Interval
VP VP VVI / 60
Defines the lowest rate the pacemaker will pace
Refractory PeriodRefractory Period
Lower Rate Interval
VP VP VVI / 60
Interval initiated by a paced or sensed event
Designed to prevent inhibition by cardiac or non-cardiac events
Refractory Period
Blanking PeriodBlanking Period
Lower Rate Interval
VP VP VVI / 60
The first portion of the refractory period
Pacemaker is “blind” to any activity
Designed to prevent oversensing pacing stimulus
Blanking PeriodRefractory Period
Upper Sensor Rate IntervalUpper Sensor Rate Interval
Lower Rate Interval
VP VP VVIR / 60 / 120
Defines the shortest interval (highest rate) the pacemaker can pace as dictated by the sensor (AAIR, VVIR modes)
Blanking PeriodRefractory Period
Upper Sensor Rate Interval
Single Chamber Mode ExamplesSingle Chamber Mode Examples
VOO ModeVOO Mode
Blanking Period
VP VP
Lower Rate Interval
VOO / 60
Asynchronous pacing delivers output regardless of intrinsic activity
VVI ModeVVI Mode
Lower Rate Interval
VP VSBlanking/Refractory
VP
{
VVI / 60
Pacing inhibited with intrinsic activity
VVIR VVIR
VP VP
Refractory/Blanking
Lower Rate
Upper Rate Interval(Maximum Sensor Rate)
VVIR / 60/120Rate Responsive Pacing at the Upper Sensor Rate
Pacing at the sensor-indicated rate
Dual-Chamber TimingDual-Chamber Timing
Rate = 60 bpm / 1000 msA-A = 1000 ms
APVP
APVP
V-AAV V-AAV
Atrial Pace, Ventricular Pace (AP/VP)
Four “Faces” of Dual Chamber PacingFour “Faces” of Dual Chamber Pacing
Rate = 60 ppm / 1000 msA-A = 1000 ms
AP VS
AP VS
V-AAV V-AAV
Atrial Pace, Ventricular Sense (AP/VS)
Four “Faces” of Dual Chamber PacingFour “Faces” of Dual Chamber Pacing
ASVP
ASVP
Rate (sinus driven) = 70 bpm / 857 msA-A = 857 ms
Atrial Sense, Ventricular Pace (AS/ VP)
V-AAV AV V-A
Four “Faces” of Dual Chamber PacingFour “Faces” of Dual Chamber Pacing
Rate (sinus driven) = 70 bpm / 857 msSpontaneous conduction at 150 msA-A = 857 ms
ASVS
ASVS
V-AAV AV V-A
Atrial Sense, Ventricular Sense (AS/VS)
Four “Faces” of Dual Chamber PacingFour “Faces” of Dual Chamber Pacing
Dual Chamber Timing ParametersDual Chamber Timing Parameters
Lower rate
AV and VA intervals
Upper rate intervals
Refractory periods
Blanking periods
Lower Rate Interval
APVP
APVP
Lower Rate Lower Rate
The lowest rate the pacemaker will pace the atrium in the absence of intrinsic atrial events
DDD 60 / 120
APVP
ASVP
PAV SAV
200 ms 170 ms
Lower Rate Interval
AV IntervalsAV Intervals
Initiated by a paced or non-refractory sensed atrial event
– Separately programmable AV intervals – SAV /PAV
DDD 60 / 120
Lower Rate Interval
APVP
APVP
AV Interval VA Interval
Atrial Escape Interval (V-A Interval)
The interval initiated by a paced or sensed ventricular event to the next atrial event
DDD 60 / 120PAV 200 ms; V-A 800 ms
200 ms 800 ms
Atrial Escape Interval (V-A Interval)Atrial Escape Interval (V-A Interval)
Lower rate interval- AV interval=V-A interval
The V-A interval is the longest period that may elapse after a ventricular event before the atrium must be paced in the absence of atrial activity.
The V-A interval is also commonly referred to as the atrial escape interval
DDDR 60 / 120A-A = 500 ms
APVP
APVP
Upper Activity Rate Limit
Lower Rate Limit
V-APAV V-APAV
Upper Activity (Sensor) RateUpper Activity (Sensor) Rate
In rate responsive modes, the Upper Activity Rate provides the limit for sensor-indicated pacing
ASVP
ASVP
DDDR 60 / 100 (upper tracking rate) Sinus rate: 100 bpm
Lower Rate Interval {
Upper Tracking Rate Limit
Upper Tracking RateUpper Tracking Rate
SAV SAVVA VA
The maximum rate the ventricle can be paced in response to sensed atrial events
Post Ventricular Atrial Refractory Period (PVARP)
Refractory PeriodsRefractory PeriodsVRP and PVARP are initiated by sensed or paced
ventricular events
– The VRP is intended to prevent self-inhibition such as sensing of T-waves
– The PVARP is intended primarily to prevent sensing of retrograde P waves
AP
VPVentricular Refractory Period (VRP)
A-V Interval(Atrial Refractory)
Post-Ventricular Atrial Refractory PeriodPost-Ventricular Atrial Refractory Period
PVARP is initiated by a ventricular event(sensed/paced), but it makes the atrial channel refractory
PVARP is programmable (typical settings around 250-275 ms)
Benefits of PVARP
– Prevents atrial channel from responding to premature atrial contractions, retrograde P-waves, and far-field ventricular signals
– Can be programmed to help minimize risk of pacemaker-mediated tachycardias
Blanking PeriodsBlanking PeriodsFirst portion of the refractory period-sensing is disabled
AP
VPAP
Post Ventricular Atrial Blanking (PVAB)
Post Atrial Ventricular Blanking
Ventricular Blanking (Nonprogrammable)
Atrial Blanking (Nonprogrammable)
PVARP and PVABPVARP and PVAB
The PVAB is the post-ventricular atrial blanking period during which time no signals are “seen” by the pacemaker’s atrial channel
It is followed by the PVARP, during which time the pacemaker might “see” and even count atrial events but will not respond to them
PVAB-independently programmable
– Typical value around 100 ms
What is happening here?
DDD / 60 / 120 / 310
PVARP
Wenckebach Operation
Upper Tracking Rate
Lower Rate Interval {
AS AS AR APVPVP VP
TARPSAV PAV PVARP SAV PVARP
P Wave Blocked (unsensed or unused)
• Prolongs the SAV until upper rate limit expires– Produces gradual change in tracking rate ratio
TARP TARP
Wenckebach
• Occurs when the intrinsic atrial rate lies between the UTR and the TARP rate
• Results in gradual prolonging of the AV interval until one atrial intrinsic event occurs during the TARP and is not tracked
What is happening here?
DDD / 60 / 120 / 310
• Every other P wave falls into refractory and does not restart the timing interval
Upper Tracking Limit
Lower Rate Interval {
{P Wave Blocked
AS AS
VPVPARAR
Sinus rate = 133 bpm (450 ms)PVARP = 300 ms SAV = 200 ms
TARP=500 ms
AV PVARP AV PVARPTARP TARP
2:1 Block
PVARP
Upper Tracking Rate
Lower Rate Interval
{No SAV started for events sensed in the TARP
AS AS
VPVP
SAV = 200 msPVARP = 300 ms
Thus TARP = 500 ms (120 ppm)
DDDLR = 60 ppm (1000 ms)
UTR = 100 bpm (600 ms) SAVTARP
PVARP
Total Atrial Refractory Period (TARP)• Sum of the AV Interval and PVARP• defines the highest rate that the pacemaker will
track atrial events before 2:1 block occurs
SAV
Fixed Block or 2:1 Block
• Occurs whenever the intrinsic atrial rate exceeds the TARP rate
• Every other atrial event falls in the TARP when the atrial rate exceeds the TARP rate
• Results in block of atrial intrinsic events in fixed ratios
PACEMAKER MODE?PACEMAKER MODE?
PACEMAKER MODE?PACEMAKER MODE?
PACEMAKER MODE?PACEMAKER MODE?
WHAT IS HAPPENING HERE?WHAT IS HAPPENING HERE?
WHAT IS HAPPENING HERE?WHAT IS HAPPENING HERE?
THANK YOUTHANK YOU