80
Operating instructions Washer-Disinfector PG 8527 / PG 8528 To avoid the risk of personal injury or damage to the appliance, it is essential that you read these operating instructions before installation and use. M.-Nr. 10 317 051 en - CA

Washer-Disinfector PG 8527 / PG 8528 - Miele...Washer-Disinfector PG 8527 / PG 8528 To avoid the risk of personal injury or damage to the appliance, it is essential that you read these

  • Upload
    others

  • View
    13

  • Download
    0

Embed Size (px)

Citation preview

Operating instructions

Washer-Disinfector

PG 8527 / PG 8528

To avoid the risk of personal injury or

damage to the appliance,

it is essential that

you read these operating instructions

before installation and use. M.-Nr. 10 317 051

en - CA

2

Safety Instructions and Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

Correct Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

Using Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

Symbols on the Washer-Disinfector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

Disposing of Your Old Washer-Disinfector . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

User profiles. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Guide to the Washer-Disinfector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

Operating Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

Display Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

Example of a selection list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

Example of a program sequence display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Context menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

Change the display language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

Change the operating level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

Start delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

Change the heating type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

Current faults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

Loading packages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

Automatic mobile unit recognition. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

Protein testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

How to load the washer-disinfector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

Surgical instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

Ophthalmology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

Anesthesia instruments (AN) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

Baby bottles. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33

OR shoes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

Laboratory equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35

Chemical processes and technology. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

Adding liquid detergents and neutralizers . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

Dispensing systems. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

Message "Fill DOS [X] container" . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

Adding liquid detergents and neutralizers . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

Message "Check dispensing system [X]" . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

Priming the dispensing system. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

Contents

3

Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

Main Switch - On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

Switching On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

Starting a program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

Program sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

End of program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

Turning off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

Cancelling a program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

Interrupting a program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49

Barcode scanner (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

Barcode scan functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

Barcode scan with conveyor belt connected. . . . . . . . . . . . . . . . . . . . . . . . . . 51

User scan Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51

User scan Cancel load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51

User scan Unload . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

Skip user scan for unloading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

Scan load. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53

Cancel load scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

Skip load scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

Using loading packages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

Loading packages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56

Add loading package . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57

Display loading package . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58

Delete loading package . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59

Move loading package . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60

Process documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61

Maintenance of the washer-disinfector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62

Routine maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62

Process validation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62

Routine checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62

Cleaning the filters in the wash cabinet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63

Cleaning the flat filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63

Cleaning the filter in front of the circulation pumps. . . . . . . . . . . . . . . . . . . . . 64

Cleaning the spray arms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65

Cleaning the control panel and glass door (Optional) . . . . . . . . . . . . . . . . . . . . . 66

Cleaning the front of the washer-disinfector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66

Contents

4

Cleaning the wash cabinet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66

Cleaning the door seal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66

Mobile units, baskets and inserts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67

Built-in printer (Option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68

Replacing the print paper. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68

Replacing the ribbon cartridge. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68

Customer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69

Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70

Plumbing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71

Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72

Optional equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73

OXIVARIO and ORTHOVARIO Optional Processes . . . . . . . . . . . . . . . . . . . . . 74

Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74

Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74

Safety instructions and warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75

Connecting the H2O2 container(s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75

Optional equipment DK + WRG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76

Functional description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76

Intended application. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76

SAFETY INSTRUCTIONS AND WARNINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76

Applicable standard programs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77

Disposing of an old appliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78

Contents

5

This washer-disinfector complies

with all applicable safety laws and

regulations. However, inappropriate

use can lead to personal injury and

property damage.

Read the operating instructions

carefully before using this

washer-disinfector. This will help

keep you safe and avoid damage to

the washer-disinfector.

Keep these instructions in a safe,

accessible location.

Correct Use

� This washer-disinfector is not

intended for any purpose that is not

indicated in the operating instructions.

Any other use, including conversions or

modifications, is not permitted and may

be dangerous.

The cleaning and disinfection

processes described here apply only to

those instruments, medical products,

and labware that are specifically

designed by the manufacturer to be

re-used. Always comply with the

product and instrument manufacturer's

instructions.

Miele will not be held liable for damage

caused by incorrect use of any kind.

� This washer-disinfector is intended

exclusively for stationary use in interior

locations.

� This washer-disinfector is equipped

with a special lighting system due to

special requirements (e.g. In terms of

temperature, humidity, chemical

resistance, abrasion resistance and

vibration). This special lighting system

may only be used for the intended use.

It is not suitable for room illumination.

The replacement must be carried out

by a Miele Service technician only.

To avoid possible injury, please

follow these instructions!

� This washer-disinfector should only

be installed, maintained, and repaired

by a Miele service technician. For best

compliance with Good Laboratory

Practice guidelines and regulations

governing the use of medical products,

a valid Miele service and repair

contract is recommended. Work

performed by unqualified persons can

place the user at considerable risk of

harm.

� Do NOT install the

washer-disinfector in an area where the

danger of explosion or freezing may be

present.

� The electrical safety of this

washer-disinfector can only be

guaranteed when it has been properly

grounded in compliance with

applicable safety regulations. It is

imperative that this basic safety

requirement be met. In case of any

doubt, have the electrical system

checked by a qualified technician.

The manufacturer will not be held liable

for damages resulting from an

inadequate grounding system (for

example, electric shock).

Safety Instructions and Warnings

6

� A damaged or leaking

washer-disinfector is is a hazard to your

safety. Immediately turn off the

washer-disinfector at the power switch

and contact the Miele Service

Department.

� Washer-disinfector operators must

be trained on a regular basis. Untrained

personnel must not be allowed access

to the washer-disinfector or its controls.

� Always exercise caution when

handling the processing chemicals for

this washer-disinfector (detergent,

neutralizer, saponifier, rinsing agent,

etc.). These products can cause

irritation and burning.

Comply with the applicable safety

regulations! Use protective eyewear

and gloves. Always follow the

manufacturer's safety

recommendations when using process

chemicals (see material safety data

sheets).

� The washer-disinfector is designed

for use with water and designated

processing chemicals only. The

washer-disinfector must not be

operated with organic solvents or

flammable liquids. This may cause an

explosion or damage to rubber and

plastic components, which in turn

allows liquid substances to leak out.

� The water in the wash chamber is

NOT safe to drink.

� Take care when sorting items with

sharp, pointed ends. Position them in

the washer-disinfector so that you will

not hurt yourself or create a danger for

others.

� When operating this

washer-disinfector, beware of the high

temperatures involved. If you bypass

the electrical lock to open the door,

there is the danger of scalding and

heat or chemical burns. If disinfectants

have been used, there is also the

danger of inhaling toxic vapour.

� The door seal and the operation of

the steam condenser should be

regularly checked, as volatile toxic or

chemical substances (e.g. aldehyde in

the disinfection agent) may be created

in the wash water during preparation.

Opening the door of the

washer-disinfector during a program

interruption therefore poses a major

risk.

� If you are exposed to toxic vapours

or processing chemicals, consult the

chemical manufacturer's material safety

data sheets for emergency procedures.

� Always allow mobile units, modular

baskets, inserts, and loads to cool

before removing. Be sure to empty any

water from concave items into the wash

chamber before removing.

� After drying with the drying unit,

leave the door open at first, to allow the

items and inserts to cool down.

� If the boiler has been programmed

to "BOILER READY", always beware of

hot water and steam when opening the

door! Looking from the unclean side,

the inlet pipe is located beneath the

right basket guide rail.

� Steam heating is permissible up to a

pressure of 1000 kPa (10 bar). This

corresponds to a water steam

temperature of 179°C.

Safety Instructions and Warnings

7

� Do NOT hose down the

washer-disinfector or the immediate

vicinity, e.g., with a water hose or

pressure washer.

� Before servicing the

washer-disinfector, disconnect it from

the power supply.

The following points should be

observed, to assist in maintaining

quality standards when processing

medical devices and critical labware,

to protect patients, and to avoid

damage to property!

� If the washer-disinfector is being

used for disinfection ordered by the

authorities, the steam condenser and

its connections to the rinsing chamber

and outlet discharge must be

disinfected before any repair or

exchange.

� Only authorized persons may

interrupt a program, and only in

exceptional cases.

� It is the responsibility of the operator

to routinely check that the required

cleaning and disinfection standards for

medical devices are being met.

Therefore, process results have to be

regularly tested and documented, both

thermo-electrically and through

inspection. For thermochemical

processes, additional testing is

required using chemical or biological

indicators.

� For thermal disinfection, the

appropriate temperatures and holding

times, as required by microbiological

and public health standards and

guidelines, must be used to achieve the

required degree of infection control.

� The processing of medical devices

is carried out by means of thermal

disinfection.

The disinfection of other items that are

not heat-resistant (e.g. OR shoes) can

be carried out using the CHEM-DESIN

program with the addition of a chemical

disinfectant.

The disinfection parameters are based

on the recommendation of the

disinfectant manufacturer. Always

observe their specifications for

handling, applications and efficacy.

These chemo-thermal processes are

not suitable for processing medical

devices.

� Under certain circumstances, the

process chemicals can cause damage

to the washer-disinfector. It is

recommended to follow the

recommendations of the process

chemical manufacturers. In the event of

damages or concerns about possible

material incompatibilities, please

consult Miele.

� Do not use any abrasive process

chemicals with the washer-disinfector.

They increase the wear on machine

components (e.g. spray arm supports).

If such agents are used for manual

pre-treatment of containers or

instruments, they must be completely

removed before placing the items in the

washer-disinfector!

Safety Instructions and Warnings

8

� Pre-treating (e. g. with cleaning

agents or disinfectants), some types of

soiling and the interaction of certain

processing chemicals can cause

foaming. Suds can have an adverse

effect on the cleaning and disinfecting

results obtained. Pre-treatment agents

need to be completely removed before

placing the items in the

washer-disinfector.

� The process must be adjusted so

that no foam escapes the wash

chamber. Escaping foam jeopardizes

the safe operation of the

washer-disinfector.

� The processing must be checked

regularly by the operator in order to

detect any foaming.

� To avoid damage to the

washer-disinfector and the associated

accessories by the effect of process

chemicals, introduced contamination or

their interaction, the instructions in the

chapter "Chemical Process

Technology" must be followed.

� Even when a chemical additive (e.g.

a cleaning agent) is recommended, the

washer-disinfector manufacturer takes

no responsibility for the effect of such

chemicals on the labware/instruments

involved.

Note that formulation changes, storage

conditions, etc., that are not disclosed

by the manufacturer may adversely

affect the cleaning results obtained.

� When using process chemicals,

always follow the instructions opf the

chemicals manufacturer. To avoid

material damage and the possibility of

violent chemical reactions (e.g.

hydrogen gas reaction), always comply

with the manufacturer's intended

application(s) for a given cleaning

agent.

� For critical applications, where very

high processing requirements have to

be met, it is strongly recommended that

all process-related factors (processing

chemicals, water quality, etc.) be

discussed in advance with Miele.

� In addition, for applications that

demand especially stringent cleaning

and rinsing results, (e.g., chemical

analysis) the operator must ensure that

quality control occurs on a regular

basis, to meet the standards involved.

� Only use mobile units, modular

baskets, and inserts as intended by the

manufacturer.

Hollow items must be positioned for full

exposure to rinse water.

� Empty all liquids from containers

and hollow items before loading them

into the washer-disinfector.

� Ensure that no acid or solvent

residues, especially hydrochloric acid

or chloride solutions, get inside the

wash chamber. Never place corroding

ferrous materials in the

washer-disinfector. Only trace amounts

of any organic solvents should be

present in any soiling.

Safety Instructions and Warnings

9

� To avoid corrosive damage, make

sure the stainless steel housing does

not come into contact with

solutions/steam containing hydrochloric

acid.

� After any work on the water supply

system, the water lines to the

washer-disinfector must be bled.

Otherwise some components of the

washer-disinfector may be damaged.

� Please be sure to follow the

installation instructions in this manual

and in the installation manual.

Using Accessories

� Whatever the application, only Miele

accessories should be connected to

this washer-disinfector. Consult Miele

for details on the type of equipment to

use.

� Only Miele mobile units, modular

baskets, and inserts should be used. If

Miele accessories are altered in any

way, or mobile units and inserts from

another manufacturer are used, Miele

cannot assure the cleaning and

disinfection results. Damage or injury

caused in this way are excluded from

the warranty.

� Only use process chemicals that are

approved by their manufacturer for the

application involved. Any negative

effects on labware or instruments and

the washer-disinfector itself are the

liability of the chemical manufacturer.

Symbols on the Washer-Disinfector

Caution:

Comply with the Operating

Instructions!

Caution:

Danger of electric shock!

Always comply with

the instructions in

Chapter "Option

DK + WRG"

(Option steam condenser with heat

recovery)!

Disposing of Your Old

Washer-Disinfector

� Any old washer-disinfector may be

contaminated by blood or other

pathogens, bodily fluids, etc., and

therefore must be decontaminated

before disposal.

For safety and environmental

protection, be sure to remove all

process chemical residue. (Comply

with safety regulations and Material

Safety Data Sheets, and wear

protective eyewear and gloves.)

Remove or inactivate the door latch, to

prevent children from locking

themselves inside. The

washer-disinfector can now be sent for

disposal through the proper resource

recovery facilities. For equipment with a

tank system, be sure to drain the water

from the tank before disposal.

The manufacturer will not be held

liable for damage caused by failure

to comply with the Safety

Instructions.

Safety Instructions and Warnings

10

This Miele washer-disinfector is

designed to clean, rinse, disinfect and

dry re-usable medical devices,

laboratory devices and medical and

laboratory accessories.

Whatever items are being processed,

always comply with the particular

manufacturer's instructions (EN ISO

17664).

Typical applications include:

– surgical instruments,

– endoscopic instruments,

– intensive care and anaesthesiology

equipment,

– baby bottles and nipples,

– supply and disposal containers,

– OR shoes,

or

– laboratory equipment for research

and production,

– laboratory equipment for analytical

and preparation applications,

– microbiology and biotechnology

labware.

The type of laboratory glassware can

include anything from beakers, bootles,

cenrifuge tubes, Erlenmeyer flasks,

funnles, pipettes, Petri dishes and

round bottom flasks.

In this operating manual, the general

term labware/instruments is used to

cover a range of equipment and items.

For standardized results, it is preferable

to machine-wash lab/surgical

instruments and equipment.

When disinfection is required to protect

staff and/or patients, the preference is

for thermal disinfection, e. g., with the

DESIN vario TD process.

– An exception are

temperature-sensitive OR shoes which

can be processed using the

CHEM-DESIN program. –

According to the A0 concept described

in EN ISO 15883-1, thermal disinfection

occurs at 80°C (+ 5°C, - 0°C) with

10 min. holding time (A0 600), or at

90°C (+ 5°C, - 0°C) with 5 min. holding

time (A0 3000), depending on the

disinfection result required. The

efficacy standard A0 3000 includes the

inactivation of HBV.

Regional and/or other official directives

like e. g., CSA Z314.8 Decontamination

of reusable Medical Devices are

applicable.

The cleaning parameters should always

be optimally matched to the type of

soiling and labware/instruments

involved.

The processing chemicals used also

should be matched to the particular

cleaning needs / analysis / analytical

methods involved.

Intended Use

11

For re-usable items to be safe for

continued use, the cleaning result must

meet certain minimum standards for

disinfection, sterilization and residue

levels.

Re-usable medical products are best

cleaned with the DESIN vario TD

process or, as appropriate, with the

ORTHOVARIO or OXIVARIO processes.

Also important for acceptable cleaning

results is the use of suitable loading

accessories (mobile units, modules,

inserts). See the section on

"Applications" for examples.

The washer-disinfector is designed to

rinse with tap water or processed water

(e.g., distilled water, pure water,

deionized water, demineralized water)

depending on the quality required for

your application(s). The water quality

selected is especially important for

labware intended for analytical uses.

Under EN ISO 15883, this

washer-disinfector is suitable for

process validation.

User profiles

Daily-routine operators

For daily routine tasks, the operating

personnel must be instructed in the

basic functions and the loading of the

washer-disinfector and should receive

regular training. Basic knowledge of the

machine-based processing of medical

devices is required.

Daily routine work takes place on

operating levels A and C.

Daily-routine supervisors

Advanced knowledge of the

machine-based processing of medical

devices is required for more complex

tasks such as program interruption or

program cancellation.

These tasks take place on operating

level B.

Service tasks and administration

Additional expertise is required for

changing the processes or adjusting

the washer-disinfector, e. g. the

accessories that are used or the site

conditions.

Validations require special expertise in

the machine-based processing of

medical products, in the processing

technology and the applicable

standards and legislation.

Service tasks and validations are

carried out on operating level D.

Intended Use

12

Unclean side

�Housing unit for Drying System DS

and/or Steam Condenser SC

� "Profitronic" electronic control system

(see Programming Manual)

�Multiport

�Main switch "ON/OFF"

�Barcode scanner mount (optional)

�Service panel

�Docking station for MF 27/28 transfer

cart

Lift door (closed)

Guide to the Washer-Disinfector

13

Unclean side

�Lift door (open)

�Connection for transfer cart and

modules

�Lower spray arm

�Filter combination

�Container for dispensing systems

DOS 1 - DOS 4

Guide to the Washer-Disinfector

14

Clean side (PG 8528 only)

�Lift door (closed)

�Service panel with docking station

for transfer cart MF 27/28

�Multiport

�Barcode scanner mount (optional)

�Printer (optional)

� "Door" button

�Display

Guide to the Washer-Disinfector

15

Operating Controls

� (ON/OFF) � Button

Turns washer-disinfector on and off

�Button � (Door opener)

Opens and closes door

�Display

If the washer-disinfector is not in use, the display

automatically switches off after approx. 10 min.

To turn the display back on, touch any button.

Fault messages are shown in the display, together with a

corresponding number.

See the Programming Manual for a table of all the fault

numbers.

Guide to the Washer-Disinfector

16

�Customer Service PC Interface �

Test and transfer point for Customer Service

�Button � (Start button)

Starts the program

�Button �C (Clear button)

– return to the previous level

– discard entries in input fields

– Cancel the program

� and Buttons � (Selection buttons)

– Shift position of marking in display

– changed marked value

– scroll

– Retrieve context menu

– display user-defined operating parameters during

program operation

OK Button

- confirm marked menu item or selected value,

- acknowledge fault messages,

- acknowledge dialogue messages,

- Access A0graph during program operation,

- Viewing the temperature curve during the course of the

program sequence.

Guide to the Washer-Disinfector

17

Display Operation

Depending on the operating level, the following choices are

available in the display:

– a program / the program overview

– the Programming menu

– the Settings menu �

In the Settings � menu, you can adjust the Profitronic

control system of the washer-disinfector to suit various

requirements. For additional information, please see the

Programming Manual.

Example of a selection list

Programme overview B

CHANGE-TA-FINE

DES-VAR-TD

DES-VAR-TD-AN �Use the � buttons to move through the menu items.

The marking in the display can be moved using the �

buttons.

Use the OK button to confirm messages or settings and

change to the next menu or another menu level.

To select a menu item, mark it using the � selection

buttons, then confirm with the OK button.

A maximum of three menu items or selections can be shown

in the display. A scroll bar at the right of the display shows

that additional menu items or selections are available. Use

the � selection buttons to display them.

A dashed line marks the end of the list. The last entry on the

list is above the line, the first entry is below it.

Guide to the Washer-Disinfector

18

Example of a program sequence display

DES-VAR-TD A

Main wash 55,2 °C

Time remaining 0:35 h

While a program is running, the following appears in the

display:

– the program name

– the operating level

– the program segment

– the water temperature

(or the air temperature in the program segment "Dry")

– the estimated time remaining or the program time

completed so far

– any fault messages or notices

Guide to the Washer-Disinfector

19

Certain functions are available through the context menu at

any time, regardless of the operational level:

– Select display language

– Change operating level,

– Enter Start Delay

– In convertible washer-disinfectors:

change the heating type,

– Create loading packages using the barcode scanner.

In operating level D, current fault messages are also

displayed (if applicable).

� To enter the context menu, simultaneously press the �

buttons for at least 3 seconds.

The context menu appears in the display:

Context menu B

Scanned loads

Change display language �

Change operating level �Use the � buttons to move through the menu items.

Context menu

20

Change the display language

You can use the context menu to change the language used

in the display.

This setting is stored until the washer-disinfector is switched

off by the � button or the main switch.

The language set in Settings / Language �does not

change. This means that all reports and print-outs will

continue to be done in the system language.

� To select the menu itemChange display language use �

then activate it with OK.

The flag � after the text Change display language is to

help you identify an unfamiliar language.

In this case, continue to change the menu item until you

see the flag �.

� Select the desired language and confirm with OK.

The selected language appears in the display.

� To leave the context menu, use�C.

Context menu

21

Change the operating level

Four operating levels are available for selection in the

electronic control unit of the washer-disinfector,

Operating

levels

Authorized access for

A and B Select from a list of approved programs.

C Automatic program assignment by mobile unit

coding.

D Open program selection through the Program

overview, Programming and Settings � menu.

The context menu can be used to switch between operating

levels.

To help prevent unauthorized access to the settings, a code

is required to change operating levels.

To change the codes for existing users, or to enter new

users, you must be a registered administrator. The

Administrator is registered by Miele Customer Service.

� Simultaneously press the � buttons for at least 3

seconds.

The context menu appears in the display:

Context menu B

Scanned loads

Change display language �

Change operating level �

Context menu

22

� Select the menu item Change operating level, then press

OK to switch to the submenu.

Select operator B

Operator ABC

Administrator

Miele Service

� Select the user group and confirm with OK.

� Enter the required code.

Change operating level B

0 � � �Enter code

The � buttons change the numbers, the OK button

confirms the set value and changes to the next number to be

entered.

The control unit switches to the selected operating level.

� To leave the context menu, use �C.

If an incorrect entry is made, Code invalid will appear in the

display.

� The message can be cancelled by using OK.

You can then resume the operating level change.

Context menu

23

Start delay

The start time for a program can be pre-selected.

� To use this feature, go to the Delay start menu item and

confirm.

� To activate the Start Delay option, go to menu item

Activate and select Yes .

Delay start B

Activate Yes

Delay start

� To use this feature, go to Delay start and confirm.

The start time menu will appear.

Delay start B

17:48Enter time

� Use the numbers to enter the desired start time, then

confirm.

� Keep pressing the �C button until the Main menu appears

in the display.

� To select a program, go to Programme overview.

Programme overview B

CHANGE-TA-FINE

DES-VAR-TD

DES-VAR-TD-AN �� Select a program and use the � button to start.

Context menu

24

The selected program, the preset start time, the current time

of day, and the time remaining until the program starts are

shown.

DES-VAR-TD B

Startzeit 17:15 15:20

Time remaining 01:55 h

The selected program automatically starts at the preset start

time.

The automatic program start can be cancelled.

� To do so, press the �C button.

You will be asked to confirm:

Cancel start time?

ja No

� To cancel the program start, select Yes and confirm.

� The automatic program start is cancelled. The display

reverts to the program overview. The Delay start function

is de-activated.

If you select No, the Delay Start option will continue.

� Once the program has run, the Delay start function is

automatically de-activated.

� The Delay start function can also be de-activated

independent of a program start. To do so, go to the Delay

start menu, and set the Activate parameter to No.

Context menu

25

Change the heating type

In convertible washer-disinfectors, you can use the context

menu to change the heating type - e. g. from electric to steam

heating.

� Select the menu item Steam/Electric, then use OK to

activate.

� Change the heating type as desired, e. g., to

Steam+electric, then confirm the selection with OK.

The new heating type is saved.

� To leave the context menu, use�C.

Current faults

In operating level D the current fault messages can be read

under this menu item.

� Select the menu item and confirm using OK.

The current fault messages are displayed.

� Use OK to scroll through the fault messages.

� To leave the context menu, use�C.

Loading packages

If a barcode scanner is connected to the washer-disinfector

and load scan is activated, loading packages can be created

with this menu item in the Context menu.

All stages in connection with the barcode scanner are

described in the "Barcode scanner" section.

Context menu

26

The automatic mobile unit recognition

feature assigns a program place to a

mobile unit. For this to work, the mobile

units must be coded with a magnetic

strip (with a bit combination).

In operating level C, each coded

mobile unit is assigned to the

corresponding program place.

When a coded unit is inserted into the

machine and the washer-disinfector

door closes, the automatic mobile unit

recognition selects the assigned

program.

See "Mobile unit coding" in the

washer-disinfector Programming

Manual for information on how to

code a mobile unit and assign

programs.

Make sure that no metallic parts

or pieces are stuck to the magnetic

strip or its underside.

Attached metallic objects can cause

the coding to be misread or be

unreadable.

Mobile units with side coupling are

coded with tracks that have Bit 6 set

to I.

Mobile units without side coupling

must be coded with tracks that do

NOT contain Bit 6.

The magnetic strips in

washer-disinfectors PG 8527 /

PG 8528 must contain black

magnets.

Automatic mobile unit recognition

27

The washer-disinfector can be fitted

with various mobile units, adapted to

the type and shape of the items that

need to be cleaned and disinfected;

they can be equipped with a wide

variety of modules and inserts.

Mobile units, modules and inserts

should be properly selected to

match the application involved.

On the following pages you will find

loading examples and additional

information.

Before EVERY program start,

conduct the following visual

inspection:

– Are the items properly sorted, loaded

and connected in the

washer-disinfector?

– Are the spray arms clean and can

they rotate freely?

– Is the flat filter free of debris (remove

any coarse material and clean the

filter as needed)?

– Is the flat filter horizontally aligned

with the wash chamber and is the

seal flush with the wash chamber?

– Is the adapter properly connected to

the water supply for the spray arms

and nozzles?

– Are the process chemical dispensers

full (detergent, neutralizer, etc.)?

After EVERY program completion,

conduct the following visual

inspection:

– Inspect the labware or instruments

for cleanliness.

– Have any hollow items been

dislodged from their proper position

on the nozzles?

Any items that were dislodged

from the adapters during a program

cycle must be washed again.

– Are the cavities and channels in

hollow items cleaned through?

– Are the nozzles and connections

firmly attached to the mobile

units/inserts?

Protein testing

After washing, items used in medical

applications should be checked at

random for proteins.

Application

28

How to load the

washer-disinfector

Select mobile units, baskets and

inserts appropriate for the

application.

– Always load items in such a way that

all surfaces will be exposed to the

cleaning water. This ensures that the

items will be properly cleaned.

– Do not nest any items inside other

items. Keep each item separate.

– Hollow items must be arranged for

complete inside exposure to the

cleaning water.

– Instruments with long, narrow hollows

must be loaded so that they can be

completely flushed with washwater

and rinsewater.

– Hollow items should be inverted and

placed in suitable mobile units,

modules and inserts so that water

can flow in and out unrestricted.

Protective netting can be used to

avoid damage to sensitive

instruments or glass breakage.

– Lightweight items can be secured

with cover nets (e.g. A 6), and small

parts can be placed in a small item

mesh tray, so that they do not block

the spray arms or become attached

to the magnetic strip on the automat-

ic mobile unit recognition.

– Deep based items should be

arranged at an angle, so that water

can run off easily.

– Tall, narrow hollow items should be

placed in the center of the mobile

unit for best water coverage.

– Mobile units with an adapter must be

properly connected.

– Modules must be properly

connected to the mobile unit.

– The spray arms must not be blocked

by items that are too tall or hang

through the baskets. Gently rotate

the arms by hand to check mobility.

– To avoid corrosion, use only

unblemished instruments made of

stainless steel.

– Heat-sensitive instruments should be

cleaned and disinfected with a

thermochemical process only.

NEVER process disposable

instruments in the

washer-disinfector.

Prepare items before loading

Empty all items before loading into the

machine. Comply with all applicable

infection-control regulations.

Ensure that no acid or solvent

residues, especially hydrochloric

acid or chlorides, get inside the

washer-disinfector.

Instrument sorting

Whenever possible, instruments should

be reprocessed in washer-disinfectors

without pre-soaking.

Application

29

Surgical instruments

The wait-time before processing

surgical instruments should be as short

as possible and should not exceed two

hours.

The instruments should be processed

using the program DES-VAR-TD.

For surgical instruments with a longer

wait time, use the OXIVARIO or

ORTHOVARIO process (see section on

"Optional features").

Surgical instruments, even those for

minimally invasive procedures, should

be thermally disinfected for best

results.

Whenever possible, demineralized

water should be used for rinsing

(conductivity ~ 15 �S/cm) for a spotless

result and to avoid corrosion. The use

of tap water with more than 100 mg

chloride/L entails the risk of corrosion.

When processing instruments for

minimally invasive procedures, e.g.,

minimal invasive surgery an intensive

internal cleaning result is imperative.

Only the programs DES-VAR-TD and

OXIVARIO are suitable for thorough

cleaning. Always observe the specific

loading instructions, and all applicable

procedures, including for the use of

appropriate detergents for these

sensitive instruments.

For rinsing, use demineralized water

with a conductivity of ~15 �S/cm.

Very narrow-lumen instruments must

be pre-cleaned manually where

necessary. Follow the instrument

manufacturer's instructions!

Most rigid sterilization containers can

be processed thermally with the

CONTAINER program. For anodized

alumium containers, demineralized

water must be used for cleaning and

rinsing.

They should not be cleaned at 93 °C

mit and 10 min holding time with an

alkaline detergent.

When processing surgical container

with a disinfection standard A0=600 the

program CONTAINER-600 must later

be programmed by Miele.

The mobile units for surgical

instruments and sterilization

containers come with their own

operating instructions.

Application

30

Ophthalmology

Using the E 929/1 mobile unit,

ophthalmological surgical instrument

sets can be cleaned and disinfected

automatically.

For rinsing, use demineralized water

with a conductivity of ~15 �S/cm

(microsiemens per centimetre).

In addition, the rinse water must have a

low endotoxin and pyrogen content!

The upper level is equipped with

separate connections for hollow

instruments, e.g. rinse and suction

grips and cannulae.

Silicone holders and stoppers fitted on

the mesh tray fix and secure the

instruments to the inlet hose of the

mobile unit.

The lower level of the mobile unit is

fitted with E 441/1 inserts or E 142

mesh trays to prepare instruments

without a hollow body.

In washer-disinfectors used to

process small-calibre

ophthalmological instruments,

plastic mesh must not be used.

The mobile units for

ophthalmological surgical instrument

sets come with their own instructions

for use.

Mobile unit E 929 (for

ophthalmological surgical

instruments) must only be used in

washer-disinfectors that do NOT

have a dispensing system with a

flow rate of 465 mL/min installed.

Ophtalmological instruments

must not be treated in

washers-disinfectors with steam

condenser with heat recovery

(DK+WRG option)!

Application

31

Anesthesia instruments (AN)

Anaesthetic instruments should be

thermally disinfected using the program

DES-VAR-TD-AN.

If no sterilization will occur after

washing, germ growth will need to

be inhibited during storage.

Because this requires thorough

drying, a sufficient amount of drying

time must be selected.

The units for anaesthesia

instruments come with their own

instructions for use.

Application

32

Baby bottles

Baby bottles in E 135 containers can be

placed in the two lower levels of mobile

unit E 935/1 and then cleaned and

disinfected in this washer-disinfector. In

addition, E 364 containers with

wide-necked teats and E 458

containers with screw-cap teats can be

placed in the upper level of the mobile

unit.

– Only use baby bottles with

dishwasher-safe volume markings.

– Bottles that are stored for longer than

4 hours prior to processing must be

filled with water to prevent food

residues from drying on the surface.

If no sterilization will occur after

washing, germ growth will need to

be inhibited during storage.

Because this requires thorough

drying, a sufficient amount of drying

time must be selected.

The inserts for baby bottles and

nipples come with their own

instructions for use.

Loading in E 935/1 mobile unit

Lower level with 4 x E 135 containers

Upper level with 3 x E 135 containers

The containers must be arranged so

that the two blocked off corners are

in the positions marked with

crosses, as shown. If bottles were to

be placed in these positions, water

would not be able to access the

inside of the bottles properly and

cleaning results would be impaired.

Application

33

OR shoes

OR shoes made of

temperature-sensitive material and/or

insoles must be thermochemically

cleaned and disinfected at 60°C

(program CHEM-DISIN).

A thermal disinfection program

(SHOE-TD-75/2) can be used as long

as the manufacturer has confirmed the

thermal stability of the product(s).

For information on the efficacy of

thermochemical disinfection, contact

the disinfectant manufacturer directly.

OR shoes should only be cleaned

and disinfected in a

washer-disinfector that has been

installed for that application.

If OR shoes are processed with

other applications in a

washer-disinfector, the operator

must assess the risks associated

with this.

Mobile unit E 975/1 with a suitable

insert can be used for this purpose,

e.g.

– Insert E 930 for up to shoe size 43

(European), size 10 (US) can be

placed in the lower level of mobile

unit E 975/1,

– Insert E 931 for up to shoe size 48

(European), size 15 (US).

When OR shoes are cleaned, large

quantities of lint can accumulate. For

this reason, be sure to check the

wash cabinet filters frequently and

clean as needed (see the “Cleaning

the wash cabinet filters” section in

“Maintenance of the

washer-disinfector”).

Application

34

Laboratory equipment

Wide-necked labware, e. g., beakers,

wide-necked Erlenmeyer flasks and

Petri dishes, or cylindrical items, e. g.,

test tubes, can be cleaned inside and

out by the rotating spray arms. For

proper cleaning, load these items in

whole, half or quarter inserts and place

them in an empty lower basket or upper

basket with spray arm.

For narrow-necked equipment, e. g.,

narrow-necked Erlenmeyer flasks,

round-bottomed flasks, measuring

cylinders and pipettes, you will need to

use injector baskets or injector

modules.

This section only covers basic

information on how to prepare and load

labware.

How to load the machine

– Petri dishes and similar items should

be loaded in a suitable insert, with

the soiled side facing center.

– Load pipettes with the tip facing

down.

– Quarter baskets should be arranged

as far from the centre as possible.

– Do not let glass items protrude near

the door opening.

Prepare items before loading

� Empty all glassware before loading

into the machine. Comply with all

applicable infection-control

regulations.

Ensure that no acid or solvent

residues, especially hydrochloric

acid or chlorides, get inside the

machine.

� Remove all agar residue from Petri

dishes.

� Remove all blood residue and clots

from test tubes.

� Remove all stoppers, corks, labels,

sealing wax residue, etc.

� Small parts, such as stoppers and

plugs, should be secured in suitable

small parts baskets.

E 940 Injector basket

For use with narrow-necked laboratory

equipment on two levels (can also be

used without upper level).

Application

35

Program Intended application

LAB-STANDARD A simple, short program for light soiling and low rinse

requirements:

– for a wide range of soiling types,

– not suitable for denatured residues such as proteins,

– not suitable for acid-soluble residues such as metallic salts

and amines.

LAB-UNIVERSAL General program for light to moderate soiling with moderate

rinse requirements:

– for removing organic residues such as proteins and some

oils and greases,

– for some inorganic residues (e.g. pH 7 water-soluble

metallic salts),

– for preparation applications and for analytical applications.

LAB-INTENSIVE Program for moderate to heavy soiling with moderate to high

rinse requirements:

– for removing organic residues such as proteins, cell and

tissue cultures, and some oils and greases,

– for some inorganic residues (e.g. pH 7 water-soluble

metallic salts),

– for preparation applications and analytical applications.

ORGANIC Program for moderate to heavy soiling with moderate rinse

requirements:

– for removing organic residues such as oils, greases, wax

and agar,

– not suitable for acid-soluble residues such as metallic salts

and amines.

INORGANIC Program for light to moderate soiling, with moderate to high

rinse requirements:

– for removing inorganic residues,

– for analytical applications and water analytics,

– for aqueous culture media with acid-soluble metallic salts

such as Ca2+, Mg

2+ etc.

Application

36

General notes

Problem How to resolve it

If seals, hoses and plastics in the

washer-disinfector suffer damage they

will not be water-tight and will not

function correctly.

– Establish the cause of the damage

and rectify it.

See information regarding "Chemical

agents“, "Soiling“ and "Reaction

between chemical agents and soiling“.

Heavy foaming during a program

affects cleaning and rinsing results.

Foam escaping from the wash cabinet

can cause damage to the

washer-disinfector.

Cleaning processes cannot be

regulated and validated where there

has been a build-up of foam.

– Establish the cause of the foam and

rectify it.

– Check the process used regularly to

monitor foaming levels.

See information regarding "Chemical

agents“, "Soiling“ and "Reaction

between chemical agents and soiling“.

Corrosion to stainless steel in the wash

cabinet and to accessories can give

them a different appearance:

– Rust (red marks / discolouration),

– black marks / discolouration,

– white marks / discolouration (etched

surface).

Corrosive pitting can lead to the

washer-disinfector not being

water-tight. Depending on application

corrosion can influence cleaning and

rinsing results (laboratory analysis) or

cause corrosion to stainless steel items

in the cabinet.

– Establish the cause of the corrosion

and rectify it.

See information regarding "Chemical

agents“, "Soiling“ and "Reaction

between chemical agents and soiling“.

Chemical processes and technology

37

Chemical agents

Problem How to resolve it

The ingredients in chemical agents

have a strong influence on the longevity

and functionality (throughput) of the

dispensing system. The dispensing

system (hoses and pumps) should be

set up for a particular type of chemical

agent.

General types:

– Alkaline to pH neutral products,

– acidic to pH neutral products,

– hydrogen peroxide.

– Follow the instructions and

recommendations of the process

chemical manufacturers.

– Carry out a regular visual check of

the dispensing system for any

damage.

– Regularly check the flow rate of the

dispensing system.

Chemical agents can damage

elastomers and plastics in the

washer-disinfector and accessories.

– Follow the instructions and

recommendations of the process

chemical manufacturers.

– Carry out a regular visual check of

any accessible elastomers and

plastics for damage.

Hydrogen peroxide can release large

amounts of oxygen.

– Only use approved processes such

as OXIVARIO or OXIVARIO PLUS.

– The wash temperature must be less

than 70 °C when using hydrogen

peroxide.

– Please contact Miele for advise.

The following chemical agents can

cause large amounts of foam to build

up:

– Cleaning agents and rinsing agents

containing tensides.

Foam can occur:

– in the program block in which the

chemical agent is dispensed,

– in the following program block if it

has been spilt,

– in the following program with rinsing

agent if it has been spilt.

– Process parameters in the wash

program, such as dispensing

temperature, dosage concentration

etc. must be set to ensure the whole

process is foam free or very low

foaming.

– Please observe chemical agent

manufacturer's instructions.

Chemical processes and technology

38

Chemical agents

Problem How to resolve it

De-foaming agents, especially silicone

based ones can cause the following:

– Deposits to build up in the cabinet,

– deposits to build up on the load,

– damage to elastomers and plastics

in the washer-disinfector,

– damage to certain plastics (e.g.

polycarbonate and plexiglass) in the

load being processed.

– De-foaming agents should be used

in exceptional cases only, for

instance when absolutely essential

for the process.

– The wash cabinet and accessories

should be periodically cleaned

without a load and without

de-foaming agent using the

ORGANICA program where available

or an optional "Organica" program

defined by Miele Service.

– Please contact Miele for advice.

Soiling

Problem How to resolve it

The following substances can damage

elastomers (hoses and seals) and

plastics in the washer-disinfector:

– Oil, wax, aromatic and unsaturated

hydrocarbons,

– emollients,

– cosmetics, hygiene and care

products such as creams (analytical

applications).

– If possible, retrofit the

washer-disinfector with oil resistant

elastomeres, provided by Miele.

– Depending on usage wipe the lower

door seal on the washer-disinfector

periodically with a lint-free cloth or

sponge.

Clean the wash cabinet and

accessories without a load using the

ORGANICA program where available

or an optional "Organica" program

defined by Miele Service.

– Prepare the load

using the OIL program where

available or an optional "Oil" program

defined by Miele Service that

dispenses cleaning agents

containing tensides.

Chemical processes and technology

39

Soiling

Problem How to resolve it

The following substances can lead to a

heavy build-up of foam during washing

and rinsing:

– Some disinfecting agents and

cleaning detergents,

– reagents for analysis e.g. for

microtiter plates,

– cosmetics, hygiene and care

products such as shampoos and

creams (analytical applications),

– active foaming agents such as

tensides.

– Thoroughly rinse items in water

beforehand.

– Select a cleaning program with at

least one short pre-rinse in cold or

hot water.

– Depending on application use

de-foaming agents that do not

contain silicone oils.

The following substances cause

corrosion to stainless steel in the wash

cabinet and on accessories:

– Hydrochloric acid,

– other substances containing

chlorides such as sodium chloride

etc.,

– concentrated sulphuric acid,

– chromic acid,

– Particles of iron and swarf.

– Thoroughly rinse items in water

beforehand.

– Place items drip dried into mobile

units, baskets and inserts and then

place these in the wash cabinet.

Chemical processes and technology

40

Reaction between chemical agents and soiling

Problem How to resolve it

Natural oils and fats can be emulsified

with alkaline chemical agents. This can

lead to a heavy build-up of foam.

– Use the OIL program where available

or an optional "Oil" program defined

by Miele Service.

– This special program dispenses

cleaning agents containing tensides

(pH neutral) in the pre-rinse.

– Depending on application use

de-foaming agents that do not

contain silicone oils.

Soiling containing high protein levels

such as blood can cause a heavy

build-up of foam when processed with

alkaline chemical agents.

– Select a cleaning program with at

least one short pre-rinse in cold

water.

Non-precious metals such aluminium,

magnesium and zinc can release

hydrogen when processed with very

acidic or alkaline chemical agents

(oxyhydrogen reaction).

– Please observe chemical agent

manufacturer's instructions.

Chemical processes and technology

41

Use only detergents and

neutralizers specially designed for

use in washers-disinfectors, and

observe the manufacturer's

recommendations for use! ALWAYS

refer to the information regarding

toxicologically harmless residual

amounts.

The washer-disinfector can be fitted

with a maximum of 5 internal

dispensing systems. They are

colour-coded, to match the dispensing

pumps to the corresponding siphon

tubes.

Dispensing system Colour coding

1 blue

2 white

3 red

4 green

5 yellow

Depending on the application(s)

chosen for this washer-disinfector, the

appropriate amounts of detergent and

neutralizer are dispensed through these

systems.

The washer-disinfector comes with two

internal dispensing systems as stan-

dard equipment:

– Dispensing system DOS 1 for use

with alkaline products, e. g., liquid

detergents. The dosing rate is

200 mL/min.

– Dispensing system DOS 3 for use

with acidic products, e. g.,

neutralizers. The dosing rate is

105 mL/min.

Up to three additional dispensing

systems can be installed optionally:

Dosing rate Product(s)

465 mL/min alkaline

200 mL/min alkaline

105 mL/min acidic

Instead of the internal dispensing

system, up to five external dispensing

systems can be connected to the

washer-disinfector.

For special applications, the external

dispensing systems can also be

activated, in addition to the internal

systems.

If a dispensing system is to

dispense different process

chemicals, the change of chemicals

must be carried out by the Miele

Service Department.

For all special information about

the OXIVARIO and ORTHOVARIO

processes and how to connect the

container with H2O2 solution, see the

section on "Optional Equipment -

OXIVARIO and ORTHOVARIO".

Adding liquid detergents and neutralizers

42

Optional equipment - OXIVARIO

For the OXIVARIO process, dispensing

systems 2 and 5 are needed, to

dispense a hydrogen peroxide solution

(H2O2 solution).

The dispensing systems for H2O2

solution are colour-coded black.

The dosing rate for system 2 is 105

mL/min. Dispensing system 5 runs the

H2O2 solution through the

washer-disinfector.

The OXIVARIO option makes

dispensing system DOS 5

unavailable for dispensing any other

processing chemical.

Optional equipment - ORTHOVARIO

For the ORTHOVARIO process, the

washer-disinfector must be fitted with

an additional dispensing pump in

dispensing system DOS 4 (add-on

equipment):

– Dispensing system DOS 4 (green) for

dispensing a special tenside

detergent. The dosing rate is

105 mL/min.

The ORTHOVARIO option makes the

DOS 4 dispensing system

unavailable for dispensing chemical

disinfectants.

Dispensing systems

Four 10-L containers of processing

chemicals can be placed behind the

service panel in the washer-disinfector.

Any additional containers will need to

be set up outside the

washer-disinfector.

Use caution when handling

processing chemicals! Some of

these products are corrosive or

cause irritation.

Always follow all applicable safety

regulations and observe the

manufacturer's material safety data

sheets!

Use protective eyewear and gloves.

Message "Fill DOS [X] container"

� Refill or replace the container

indicated on the display.

[x] Instead of an X, the number of the dispensing

system involved will appear.

Fill the container(s) when you see a

prompt on the display, e. g., Fill

DOS1 container. This will prevent

containers from becoming

completely empty and needing to be

primed (cleared of air).

Adding liquid detergents and

neutralizers

� Open the left and right latches of the

service panel.

Adding liquid detergents and neutralizers

43

� Then lift it up and out at the bottom.

� Remove the container from the

washer-disinfector.

� Unscrew the siphon tube and remove

it.

� Fill the container with the required

product.

� Insert the siphon tube into the

container opening and screw it into

place.

Once the containers have been filled,

the message clears from the display.

� Place the container in the

washer-disinfector.

� Insert, close and lock the service

flap.

Message "Check dispensing system

[X]"

The currently running program pauses.

� Check the container(s) and

dispensing hoses indicated on the

display.

[x] Instead of an X, the number of the affected

dispensing system will appear.

� Refill or replace the container(s)

shown in the display.

� Use the appropriate service program

to prime the affected dispensing

system (clear it of air).

Containers and dispensing systems

that are not in use can be shut off, to

avoid fault messages in the display

(see the Programming Manual,

"Machine function / Query

container").

Priming the dispensing system

Whenever a container has been

allowed to completely empty, it must be

primed (cleared of air) after refilling.

� Select the corresponding service

program, e. g., DOS1-FILL.

� Press the � start button.

Adding liquid detergents and neutralizers

44

Main Switch - On/Off

The main switch disconnects the user

side of the washer-disinfector from the

power supply.

� Turn the main switch to I ON.

Once the start procedure is completed,

the washer-disinfector is ready to

operate.

The following message may appear: "

"New power frequency 60Hz Flow

rates automatically adjusted" or "New

power frequency 50Hz Flow rates

automatically adjusted".

During the start procedure, the

Profitronic detected a different power

frequency and adjusted the flow rate

of the dispensing pump accordingly.

Before operation, you will need to

check whether the change in power

frequency was caused by a fault in

the power supply.

If so, incorrect amounts of

processing chemicals would be

dispensed during the program.

The fault message can only be

confirmed in operating level D and in

the service mode.

Switching On

� Hold down the � button for at least

1.5 sec.

The actuating time for the � button

can be set up to a maximum of 10

sec. (see for the section on "System

functions /Actuating time ON/OFF

Button" in the ).

Depending on the set operating level,

the following will appear in the display:

Operating

levels

Displayed

A and B A list of approved

programs.

C Instructions for Automatic

Mobile Unit Recognition.

D Select from:

– Program Overview

– Programming

– Settings �.

If the washer-disinfector is not in use,

the display automatically switches off

after approx. 10 min.

To turn the display back on, touch

any button.

Operation

45

Opening and closing the door

� Press the � button.

The door raises or lowers.

To open the door on the clean side

(PG 8528), the following must also

apply:

– Programs with a process statement

must have been completed without

fault in accordance with the program

parameters,

or

– the interlock function has been

activated, i. e., door mechanism is

set to one of the following

parameters:

- interlock,

- door clean + interlock,

- door unclean + interlock.

(see Programming Manual under

"Machine functions / Door

mechanism").

If the pre-heater has been

programmed for heating, always

beware of hot water or steam exiting

the open door! Looking from the

unclean side, the inlet pipe is

located beneath the right basket

guide rail.

Starting a program

You can find detailed and important

information regarding the Miele

default programs in the program

overview chart in the accompanying

Programming Manual.

Any program or dosing changes

in connection with the preparation of

medical devices must be

documented.

In such cases, the

washer-disinfector results (cleaning,

disinfection) will need to be

re-validated.

If a barcode scanner is connected to

the washer-disinfector and the user

or load scan is activated, the

selected program cannot be started

until scanning is completed.

All work steps in connection with the

barcode scanner are described in

the "Barcode scanner" section.

Operation

46

Option Steam condenser with

heat recovery (DK+WRG):

After extended standstill times such

as weekends, start the FILL BOILER

utility before starting operation.

For hygienic reasons, the AD water

from the steam condenser and the

boiler is replaced by fresh, cold AD

water.

Operating levels A and B

A list of all selectable programs

appears in the display.

� Select the desired program with �,

then confirm with OK.

� Press the � start button.

The program runs.

Operating level D

� Select the menu item Programme

overview and confirm with OK.

� Select the desired program with �,

then confirm with OK.

� Press the � start button.

The program runs.

Operating level C

Make sure that no metallic parts

or pieces are stuck to the magnetic

strip or its underside.

Attached metallic objects can cause

the coding to be misread or be

unreadable.

Before pressing the start button

to begin the program, it is essential

to check that the program required

for this mobile unit is the one

shown in the display!

Otherwise, inadequate cleaning and

disinfection could result!

Always make sure that mobile units

with automatic recognition (AWK)

are properly coded.

Option Steam condenser with heat

recovery (DK+WRG)

Always comply with the instructions

in Chapter "Option Steam condenser

with heat recovery (DK+WRG)".

� Push the coded mobile unit into

place in the washer-disinfector.

� Close the door � and press the

start button �.

The program runs.

Operation

47

Program sequence

Once started, the program proceeds

automatically.

As it proceeds, the program steps are

shown on the display.

Detailed information on the program

sequences is available in the

Programming Manual.

Do not change the printer paper

roll or ribbon cartridge when a

program is running.

End of program

� Once the program ends, you will see

Programme finished in the display.

Turning off

� Press the � button.

If the machine is to be left off for

several days, be sure to turn the main

switch to 0 OFF.

Cancelling a program

A program can only be interrupted

or cancelled in operating levels B

and D.

In operating levels B or D

� Press the �C button.

The program is interrupted.

The following prompts appear in the

display:

Cancel programme (OK)

or continue (Clear)?

In cases involving mandatory

decontamination, disinfectant must

be added to the contaminated water

before it can be released into the

public sewer system. The door on

the unclean side can be opened for

this purpose.

� Confirm program cancellation with

the OK button.

Programme cancelled,

Water drainage appears in the

display.

After the water has drained away, the

program list is displayed again.

Operation

48

Interrupting a program

A program can only be interrupted

or cancelled in operating levels B

and D.

If the door must be opened for urgent

reasons, e. g., if items are moving

around too much, or the cleaning

performance needs to be checked

(deliberate intervention):

If a program with process statement

is interrupted and then continued,

make sure to check the display

upon program completion. If the

display reads Process parameters

not achieved, the door was opened

after process parameter monitoring

began, and so the process

parameter was not achieved. The

program may need to be repeated.

In operating levels B or D

� Press the �C button.

The program is interrupted.

The following prompts appear in the

display:

Cancel programme (OK)

or continue (Clear)?

� Open the door �

Note PG 8528:

Which door of the washer-disinfector

can be opened depends on which

setting has been programmed under

"Machine function/Door mechanism"

(see the Programming Manual).

The door on the unclean side of the

washer-disinfector can always be

opened, no matter which parameter

has been set.

The door on the clean side can only be

opened if one of the following

parameters are set:

- Interlock,

- Door clean + Interlock,

- Door unclean + Interlock.

When using the

washer-disinfector as a medical

device (as per EN ISO 15883,

CAN/CSA Z.15883), the door

interlock function should be off.

Caution! Items in the machine

may be hot. There is danger of

burning or scalding.

In programs with thermochemical

disinfection, steam containing high

levels of disinfectant can escape!

� Arrange items securely. Take all

necessary infection control measures

and wear gloves.

� Close the door �.

� Press the �C button.

The program continues.

Operation

49

Barcode scanners can be connected to the

washer-disinfector via the multiport on the unclean side and,

if available, on the clean side as well.

The configuration of the serial interface for the barcode

scanner is described in the section "PC / Print functions /

Interface configuration" in the Programming Manual.

Barcode scan functions

All barcode scanner functions are available at every

operating level:

– Scan loading operator

– Scan unloading operator

– Scan the load

The scanned barcodes are used to record the users and/or

loads in the wash protocol.

User scan and load scan can be activated independently of

each other. Activation is described in the section "System

functions / Barcode" in the Programming Manual.

Barcode scanner (optional)

50

Barcode scan with conveyor belt connected.

If a conveyor belt is connected to the washer-disinfector, only

loading packages that have already been scanned are

processed on operating level C.

The sequence of loading packages in the Profitronic

must match that of the mobile units on the feeder belt.

At the end of the program, the message Process parameters

achieved appears and the completed program package is

deleted.

If the program is cancelled or if an error occurs during the

program run (Process parameters not achieved), the last

processed loading package is not deleted. When restarted

after correcting the error, the program is run again.

If processing of the loading package does not need to be

repeated after cancellation of a program, it must be

deleted manually.

User scan Load

When Scan loading operator is activated, the following

message is displayed after program selection (immediately in

the case of automatic mobile unit recognition in operating

level C):

Scan operator

Once the user barcode has been scanned, the program can

be started.

When Scan the load is also activated, you are prompted to

scan the load after the user scan.

User scan Cancel load

User scan Load can be cancelled using the �C key. The

program overview is displayed again.

Barcode scanner (optional)

51

User scan Unload

If Scan unloading operator is activated, you will be

prompted to scan the user barcode once a program has

ended:

Programme finished

Scan operator

The door cannot be opened until the user barcode has been

scanned. Depending on the settings in the Menu Machine

functions / Automatic door the door will open

automatically.

After cancellation of a program with the �C button

If Scan unloading operator is activated, the following

message is displayed after cancellation of the program:

Programme cancelled

Water drainage

After the water has drained, the Scan user prompt will

appear:

Scan operator

Skip user scan for unloading

In operating levels B and D, you can skip Scan unloading

operator by pressing the OK key.

At the end of the wash protocol, "User scan skipped" is

printed.

Barcode scanner (optional)

52

Scan load

When load scanning is switched on, the following message

prompts the operators to scan the load:

Scan the load A

� Scan the load

� Scan the first barcode of the load.

In the Scan the load menu the barcode numbers of the

scanned load are listed.

Scan the load A

0987664

� Scan further barcodes if applicable.

If a scanned barcode already exists, the following query

appears:

Load already scanned. Accept?

A maximum of 99 load barcodes can be scanned. After that

the following message appears:

Max. number of entries reached

The last scanned barcode can be deleted from the list with

�C. To prevent unintentional removal from the list you will be

prompted to confirm the deletion:

Delete last entry?

Barcode scanner (optional)

53

Cancel load scan

If no load was scanned, the load scan can be cancelled

using the �C key. The program overview is displayed again.

If load scans already exist, they must be deleted prior to

cancellation. To do this, answer the query Delete last

entry? with Yes for each scanned load.

Skip load scan

In operating levels B and D, you can skip the load scan by

pressing the OK button.

If you answer the query Skip load scan? with No, the

program overview will be displayed.

If you answer the query with Yes, the program can be started

by pressing the h key.

At the beginning of the wash protocol, "User scan skipped"

is printed.

Finish load scan

� After all the load barcodes have been scanned, finish the

process by pressing OK.

The �key flashes and the program can be started.

Using loading packages

When load scan is on, the Profitronic checks, before the

program is started, that at least one loading package has

been saved.

Barcode scanner (optional)

54

If this is the case, the following message is displayed:

A

Accept scanned loads?

Yes No

� Acknowledge the message by pressing Yes.

The list with all the loading packages is displayed:

Select scanned loads A

1_14.02.2007_08:23

2_14.02.2007_09:11

� Select a loading package and confirm by pressing OK.

The �key flashes and the program can be started.

At the end of the program, the message Process parameters

achieved appears and the completed program package is

deleted.

The last processed loading package is not deleted:

– if the program is cancelled or if an error occurs during the

program (Process parameters not achieved),

– with programs in whose program header the parameter

Report process from is not defined.

The next time the program is started the loading package will

be processed again. It can be manually deleted or moved on

the list.

If none of the saved loading packages are to be used:

� Select No and confirm.

� The message Scan the load is displayed.

� Scan the load manually as described under "Scan load".

Barcode scanner (optional)

55

Loading packages

Loading packages can be created to optimize the work flow

when loading the washer-disinfector. To do this, the load

scan must be activated – see the "System functions /

Barcode" section in the Programming Manual).

Context menu B

Scanned loads

Change display language �

Change operating level �After the context menu has been opened and the menu item

Scanned loads confirmed, the following functions are

available:

Add scanned loads

Display scanned loads

Delete scanned loads

Move scanned loads

Barcode scanner (optional)

56

Add loading package

To create a loading package, open the context menu, then

select and confirm the menu item Scanned loads.

Scanned loads B

Move scanned loads

Add scanned loads

Display scanned loads �� Open and confirm the menu item Add scanned loads.

The following message appears on the display:

Scan the load

� Scan the individual components of the load.

� Press OK to confirm the complete loading package.

Repeat the steps for further loading packages. The loading

packages are stored in a list (see "Display loading

packages"). The last scanned loading package is added at

the end of the list.

A maximum of 50 loading packages can be stored.

Loading packages must be processed or deleted before

more loading packages can be added.

Barcode scanner (optional)

57

Display loading package

To display the loading packages, open the context menu,

then select and confirm the menu item Scanned loads.

� Open and confirm the menu item Display scanned loads.

The list with all the loading packages is displayed: Each

loading package is identified by a number, the date and the

time.

Display scanned loads A

1_14.02.2007_08:23

2_14.02.2007_09:11

� Select and confirm the desired loading package.

The components of the loading package are listed in the

Scan the load menu .

Scan the load A

0987664

0983558

0991724

� Press �C to return to the Display scanned loads menu.

� Press �C to return to the Scanned loads menu.

Barcode scanner (optional)

58

Delete loading package

To delete a loading package, open the context menu, then

select and confirm the menu item Scanned loads.

� Open and confirm the menu item Delete scanned loads.

All the existing loading packages are displayed with number,

date and time.

Delete scanned loads A

1_14.02.2007_08:23

2_14.02.2007_09:11

� Select and confirm the desired loading package.

You will be asked to confirm:

Delete scanned loads?

� Confirm the request.

The loading package is deleted.

� Press �C to return to the Scanned loads menu.

Barcode scanner (optional)

59

Move loading package

To move a loading package, open the context menu, then

select and confirm the menu item Scanned loads.

� Open and confirm the menu item Move scanned loads.

All the existing loading packages are displayed with number,

date and time.

Move scanned loads A

1_14.02.2007_08:23

2_14.02.2007_09:11

3_14.02.2007_10:31

� Select the desired loading package.

� Move the loading package using � to the desired

position and confirm.

The list will be saved in the new sequence.

� Press �C to return to the Scanned loads menu.

Barcode scanner (optional)

60

The washer-disinfector provides the

option of documenting the cleaning

processes (process documentation).

External process documentation

software or an internal or external

printer can be used for process

documentation.

Network connection

A 5 m long interface cable is included

for connecting the washer-disinfector

with suitable network components (e.g.

hub, switch).

It is connected to the Profitronic

Ethernet interface and is rolled up in the

control cabinet of the

washer-disinfector.

Only Miele Technical Service may

connect the cable to another

interface.

Serial interface

The cable may also be used to connect

the washer-disinfector to an external

device via the Profitronic RS 232

interface. An RJ45 / Sub-D adapter

plug is included.

For instructions on how to configure

the interface, refer to the section on

PC/Print functions in the

Programming Manual.

If connecting a printer or PC, please

note the following:

– Only use an industry-standard PC

(IEC 60950).

– Select a printer or PC that is suited to

the

installation location

– The maximum length for an extension

cable from the serial interface to the

printer or PC is 10 m and from the

Ethernet interface 100 m.

– A crossover cable is required for

direct connection between the

washer-disinfector and a PC.

Further information about suitable

printers and software solutions may

be obtained from Miele Sales or

Miele Customer Service.

Process documentation

61

Routine maintenance

After every 2000 hours of operation,

or at least once a year Miele Service

will need to perform routine

maintenance on the

washer-disinfector.

Maintenance covers the following

points:

– electrical integrity,

– door mechanics, door seals

– connectors and fittings in the wash

comparment,

– water intake and drainage,

– internal and external dispensing

systems,

– spray arms,

– filter combination,

– sump with drain pump and

non-return valve,

– steam condenser,

– all mobile units, modules, and inserts

if applicable:

– the drying unit.

During maintenance, the following

function checks are run:

– a program test run,

– electrothermal measurement,

– seal testing,

– all safety-related measurement

systems (fault report)

– the safety features.

Process validation

It is the responsibility of the operator to

routinely check that the required

cleaning and disinfection processes

are functioning properly.

In some countries, this is also required

by national legislation, regulations, or

recommendations.

Operators are also required to conduct

such checks under international stan-

dard EN ISO 15883, ; CAN/CSA

Z15883.

Routine checks

Before each day's use, the operator

must conduct a series of routine

checks. A routine checklist is supplied

with the washer-disinfector.

The following items must be checked:

– the filters in the wash cabinet,

– the spray arms in the

washer-disinfector and on the

baskets and inserts,

– the wash cabinet and the door seals,

– the mobile units, modules, and

inserts.

Maintenance of the washer-disinfector

62

Cleaning the filters in the wash

cabinet

The flat filter in the bottom of the wash

cabinet and the filters in front of the

circulation pumps prevent the

introduction of contaminant particles

into the circulation system.

The washer must not be

operated without the filters in place.

Risk of injury due to glass

splinters, needles, etc. which are

retained by the flat filter.

Cleaning the flat filter

Retained contaminant particles may

clog the flat filter. For this reason, the

filter requires cleaning at least once

every day.

If heavy contamination is visible after a

program cycle, the flat filter requires

cleaning before the program is started

again.

If there is any visible residue on the

flat filter after manual cleaning, it can

be removed by cleaning in the

washer-disinfector.

The exact process depends on the

mobile unit. Please contact Miele

Technical Service if you have any

questions.

� Use the handles to remove the flat

filter.

Do not pull at the ventilation pipes

as they could bend and thus lose

some of their functionality.

� Use running water (and a brush, if

necessary) to rinse off deposits and

contaminants.

� Blow compressed air through the flat

filter from the back against the

direction of flow.

� Replace the flat filter.

Insert the flat filter horizontally and

ensure that the gasket is flush with

the washing cabinet.

� Manually turn the lower machine

spray arm to check its free rotation.

Maintenance of the washer-disinfector

63

Cleaning the filter in front of the

circulation pumps.

Two filters for the protection of the

circulation pumps are installed below

the flat filter.

These filters must be checked daily and

cleaned, if necessary.

� Remove the flat filter.

� Pull the filters straight up from their

brackets and clean under running

water.

� Replace the filters.

� Replace the flat filter.

Check that all filters have been

properly returned.

� Manually turn the lower machine

spray arm to check its free rotation.

Maintenance of the washer-disinfector

64

Cleaning the spray arms

The spray arms can become clogged.

Check them daily and clean if needed.

� Use a sharp pointed object to push

any particles into the spray arm jets,

then rinse thoroughly under running

water.

To do so, remove the spray arms as

follows:

� Remove the mobile unit.

Washer-disinfector spray arms

� Loosen the nut with a wrench and

pull the spray arm either upward or

downward.

Spray arms on mobile unit / inserts

� Use a wrench to loosen the nut on

the spray arm guide, then pull the

spray arm downward.

If you notice wear and tear on the

bearing liners or the gasket, please

contact Miele Service.

� After cleaning, put the spray arms

back in place and tighten the nut.

Make sure the nut is in the proper

position.

After replacing the spray arms,

check to make sure that they rotate

freely.

Make sure that no metallic parts

or pieces become attached to the

magnets at the ends of the spray

arms.

Attached metallic objects can cause

the spray arm speed (RPM) to be

miscalculated.

Maintenance of the washer-disinfector

65

Cleaning the control panel and

glass door (Optional)

� Press � to switch off the

washer-disinfector.

� Clean the control panel and the glass

door with a damp cloth or a standard

glass cleaner.

To disinfect, use an approved

cleaning agent.

Do NOT use abrasive or

all-purpose cleaners.

The chemicals contained in these

products can seriously damage the

surface of the glass.

Cleaning the front of the

washer-disinfector

� To clean the stainless steel exterior,

use a damp cloth and dish liquid or a

non-abrasive stainless steel cleaner.

� To help prevent re-soiling

(fingerprints, etc.), a stainless steel

conditioner can be used after

cleaning.

Do NOT use thinner or

ammonium-based cleaners.

They can damage the surface.

NEVER hose down the

washer-disinfector or the immediate

vicinity, e.g., with a water hose or

pressure washer.

Cleaning the wash cabinet

The wash cabinet is mostly

self-cleaning.

If you notice a build-up of deposits,

please contact Miele Service.

Cleaning the door seal

To remove soiling, clean the door seal

regularly with a damp cloth.

Damaged or leaky door seals should

be replaced by Miele Service.

Maintenance of the washer-disinfector

66

Mobile units, baskets and

inserts

To ensure that the mobile units, baskets

and inserts function properly, they must

be checked daily. A checklist is

supplied with the washer-disinfector.

Check the following points:

– Are the rollers in proper condition

and are they securely attached to the

mobile unit / insert?

– Are the counter-nuts on the guide

rails firmly secured?

– Is the mobile unit connection

correctly set and firmly screwed on?

– For mobile units in the modular

system, are the caps in the module

connections working properly?

– Are all spray jets, spray sleeves and

hose adapters securely attached to

the mobile unit / insert?

– Are all spray jets, spray sleeves and

hose adapters unclogged so that

water can flow through?

– Are all caps, covers and fasteners

securely attached to the spray

sleeves?

– Is the magnetic strip of the Automatic

Mobile Unit Recognition free of

adhering metallic items?

– Are the screws in the magnet rails for

Automatic Mobile Unit Recognition

tightly secured?

If there are spray arms:

– Do the spray arms rotate freely?

– Are the spray jets clogged? See the

section on "Cleaning and Care /

Cleaning the spray arms".

– Are the magnets in place on the

spray arm ends?

– Are the magnets on the spray arms

free of attached metal parts?

Mobile units, modules and inserts

must also undergo routine

maintenance within the framework of

periodical maintenance of the

washer-disinfector every 2000

operating hours or at least once

per year.

See Maintenance of the

washer-disinfector.

Maintenance of the washer-disinfector

67

Built-in printer (Option)

Replacing the print paper

When the red indicator light behind the

front panel on the printer comes on,

replace the paper roll. Check the print

paper regularly.

To do this:

� Open the front panel of the printer by

the upper edge, and pull down.

� Remove the empty paper roll and

spindle together from the holder, fit

the new roll onto the spindle, and

install them back into the printer.

� Guide the paper up and over the

paper transport roller (slit behind the

ribbon cartridge). Press the grey

paper transport button until the paper

re-emerges above the ribbon

cartridge.

� Guide the paper through the slit in

the front panel. Shut the panel.

Replacement paper rolls (58 mm wide /

outer diameter approx. 50 mm) can be

obtained from Miele Service.

Replacing the ribbon cartridge

� Open the front panel of the printer by

the upper edge, and pull down.

� The ribbon cartridge is above the

paper roll. Pull it forward and out of

the holder and replace it with a new

one. The paper must be guided

between the ribbon and the cartridge

housing.

� On the right, turn the small wheel for

manual ribbon transport clockwise

until the ribbon is taut.

� Guide the paper through the slit in

the front panel. Shut the panel.

Replacement ribbon cartridges can be

obtained from Miele Service.

Maintenance of the washer-disinfector

68

Repairs should only be carried

out by Miele Service.

Work performed by unqualified

persons can place the user at

considerable risk of harm.

To help avoid unnecessary service

calls, please be sure to check first

whether a fault message is due to

operator error.

See the "Messages" section in the

Programming Manual for an

overview of the various fault

messages that can appear in the

display.

If a fault occurs and cannot be

corrected using the "Operating

Instructions" and "Programming

Manual", please contact Miele Service.

For contact information see last page

of the operating instructions.

When contacting the service

department, you will need to quote the

model and serial number of your

washer-disinfector. This information can

be found on the data plate (see

"Electrical Connection").

Customer Service

69

All electrical work must be carried

out by a qualified electrician in

accordance with local and national

safety regulations.

– Installation, repairs and other work

by unqualified persons could be

dangerous. The manufacturer can

not be held responsible for

unauthorized work.

– For hard-wired machines connection

should be made via a suitable

isolator, with an on-off switch which

is easily accessible for servicing and

maintenance work after the machine

has been installed. An electrical

safety test must be carried out after

installation and after any servicing

work. When switched off there must

be an all-pole contact gap of at least

3 mm in the mains connection

switch.

– For extra safety it is advisable to

protect the machine with a 30 mA

residual current device (RCD).

– Check that the phase rotation is

correct when connecting the

machine.

� For technical data see the data plate

or wiring diagram supplied.

The washer-disinfector must only be

operated with the voltage, frequency

and fusing shown on the data plate.

The data plate is on the cover plate

behind the service panel of the unclean

side.

The wiring diagram is secured to the

inner side of the service panel.

This machine can be converted in

accordance with the conversion

diagram and wiring diagram supplied.

Further notes on electrical

connection are given on the

installation diagram supplied with

the machine.

WARNING:

THIS APPLIANCE MUST BE

GROUNDED

Electrical connection

70

Water connection

This washer-disinfector must be

connected to the water supply in

accordance with all national and

local plumbing codes. Please refer

to the installation diagram supplied

with the machine.

Water in the

washer-disinfector is not suitable

for drinking.

If the water supply has a high iron

content there is a danger of corrosion

occurring on items being processed in

the machine, as well as in the machine

itself.

If the chloride content of the water

exceeds 100 mg/l the risk of corrosion

to items being processed in the

machine will be further increased.

The washer-disinfector is designed so

that it may be connected to a water

supply without an extra non-return

valve, unless required by code.

Taps and drain connections should be

situated as close to the

washer-disinfector as possible, and be

easily accessible.

The machine is designed for

connection to cold, hot and DI water.

If there is no hot water supply, the

intake valve for the warm water

connection should be connected to the

cold water supply.

Refer to the Installation Diagram for

more information.

Large area filters, supplied with the

machine, are to be installed in the hose

attached to the water supply.

Do not shorten or damage the

hoses!

Do not remove the filters in the

hoses.

Plumbing

71

Height 250 cm (minimum room height)

Width 115 cm

Depth 87 cm

Weight (net) approx. 550 kg

Operating weight approx. 800 kg

Voltage (volts) see data plate

Connected load (watts) see data plate

Fuse rating see data plate

Air pressure connection 600 kPa (required for steam operation)

Steam connection:

with Electric TA

with Steam TA

250 - 1000 kPa / 140 - 180°C

250 - 600 kPa * / 140 - 164°C

Water pressure (flow rate) 200 - 1000 kPa overpressure

Cold, hot, and deionized (DI) water§

up to max. 70°C

Room temperature 5°C to 40°C

Relative humidity: maximum

linear decreasing to

80 % for temperatures up to 31°C

50 % for temperatures up to 40°C

Altitude above sea level up to 1500 m#

Contamination level P2 (as per IEC/EN 61010-1)

Ingress protection rating (as per IEC

60529)

IP20 (ingress of dust)

Noise level in dB (A), peak LpA for

cleaning and drying cycles

< 70

Certification marks CSA, Medical Class II Device Licence

�mark MPG Guidelines - 93/42/EEC, Class IIb

Manufacturer’s address Miele & Cie. KG, Carl-Miele-Str. 29,

33332 Gütersloh, Germany

* At a steam pressure of 250 - 300 kPa, the drying unit reaches a maximum drying

temperature of 90 - 100°C.

§for AD water connection with Option Steam condenser with heat recovery

(DK+WRG) max. 20 °C

#If the machine is installed above 1500 m, the wash water will boil at a lower

temperature. This may require you to lower the disinfection temperature and

lengthen the holding time (A0 value).

Technical data

72

optional modules:

– Drying unit

– Steam condenser

– Steam condenser with heat recovery (DK+WRG) – not available in all countries.

– Pre-heater

– Conductivity detector

– Report printer

– Barcode scanner connection

– Up to 3 additional internal dispensing systems

– OXIVARIO (after-sale option through Miele Service)

– ORTHOVARIO (after-sale option through Miele Service)

– Glass doors and wash cabinet lighting

– Combination steam/electric wash cabinet heating

Water softener

If your tap water hardness is � 4 gr/gal, you should plan to install a water softener.

The "PG 8597 Aquasoft system" is suitable (up to 40 gr/gal).

Optional equipment

73

Intended Use

The washer-disinfector can be

equipped (or retro-fitted) for the

OXIVARIO process, with two additional

dispensing pumps and an intermediate

reservoir for hydrogen peroxide solution

(H2O2solution). In this case, the DOS 2

dispensing system will dispense the

H2O2solution.

For the ORTHOVARIO process, the

DOS 4 dispensing system must be

equipped (fitted) with a special

dispensing pump for a special tenside

detergent.

The OXIVARIO PLUS, OXIVARIO and

ORTHOVARIO programs are available

for the special processes.

The OXIVARIO process uses active

oxygen release under alkaline

conditions. The detergent used in this

process must be tenside-free, with a

pH value between 11 and 11.5.

The ORTHOVARIO process uses a

material-tolerant tenside detergent in

the first cleaning phase, and active

oxygen release in the second phase,

produced at a pH between 10.0 and

11.0 at a slightly higher temperature of

65°C.

Applications

The OXIVARIO process uses alkaline

cleaning, and is designed to meet the

special demands involved in the

preparation of surgical instruments.

This process is especially suited to the

preparation of OP instruments, i. e.,

from HF (high frequency) surgery,

orthopaedic surgery,

instruments that dry out during long

delays until preparation, and for

instruments affected by antiseptics.

The OXIVARIO PLUS process is

designed to prevent iatrogenic prion

transmission (vCJK).

The cleaning processes are gentle

enough for minimally invasive

instruments, including ophthalmic, as

long as the manufacturer has approved

alkaline cleaning methods for them.

However, this process is not suitable for

anodized aluminium. The materials in

titanium alloys, e.g., implants, may vary

in terms of material tolerance.

Colour-coding can change, resulting in

errors. Consult the manufacturer for

verification.

The ORTHOVARIO program is

designed for the material-tolerant

preparation of alkaline-sensitive

instruments made of aluminum, in

particular for instruments used in

orthopaedics, e.g., for motor systems.

The oxidative effects of this process

make it unsuitable for titanium alloys,

in particular for colour-coded

implants.

The processes clean so

thoroughly that, to prevent

damages, instruments with metal

surfaces sliding across each other

must be carefully attended to with

suitable material immediately

following preparation.

OXIVARIO and ORTHOVARIO Optional Processes

74

Safety instructions and

warnings

These safety instructions and

warnings are in addition to those

found at the beginning of the

Installation and Operating

Instructions!

� The H2O2solution must only be used

in special containers from "Ecolab" or

"Dr. Weigert", with the matching

adapters.

� Always read and comply with the

material safety data sheets from the

manufacturers of the process

chemicals.

� Exercise care when handling

H2O2solution - it is a corrosive

substance!

Always comply with all applicable

safety regulations!

Use protective eyewear and gloves.

� Follow the manufacturer's

instructions for disposal of empty

containers.

� The H2O2solution must never be

mixed with any other processing

chemicals. A violent chemical reaction,

such as a flash fire, may occur.

� Only special tenside detergents

from "Ecolab" or "Dr. Weigert" should be

used.

Connecting the H2O

2

container(s)

The hose line for the H2O2 container is

identified by a black label. It is supplied

without an adapter, because the

extraction systems for each container

differ by manufacturer.

� Connect the appropriate adapter

from the manufacturer to the (black)

hose line.

� Connect the H2O2 container.

� Start the DOS2-FILL service program.

In contrast to the other processing

chemicals, the H2O2 container should

be completely empty before you

exchange it.

� Wait until you see Fill container DOS

2 in the display before installing a

new container with H2O2-, then start

the DOS2-FILL service program.

� If the display shows Check

dispensing system 2, check the

container and dispensing line. The

program automatically pauses for

this.

OXIVARIO and ORTHOVARIO Optional Processes

75

Functional description

Washers-disinfectors with a steam

condenser with heat recovery

(DK+WRG) are identified with this

symbol.

The variant Steam condenser with heat

recovery (DK+WRG) may be used only

with a built-in boiler.

In this variant, the steam condenser is

connected to the AD water and not to

the cold water supply. The AD water

used as cooling medium flows through

the steam condenser, where it is

preheated, into the boiler of the

washer-disinfector.

Thus, heating the boiler water requires

less energy. Compared to a connection

of the steam condenser to the fresh

water supply, this variant also reduces

the overall water consumption.

Due to the modified water conduits in

the washer-disinfector, the water inlet

times increase in purge blocks using

warm AD water.

Intended application

Washers-disinfectors with Steam

condenser with heat recovery

(DK+WRG) may be only used in

medical applications.

For optimum utilization of the heat

recovery system, operate the

washer-disinfector without extended

standstill times.

After extended standstill times

such as weekends, start the FILL

BOILER utility before starting

operation.

For hygienic reasons, the AD water

from the steam condenser and the

boiler is replaced by fresh, cold AD

water.

SAFETY INSTRUCTIONS AND

WARNINGS

These safety instructions and

warnings apply in addition to those

found at the beginning of the

Operating Instructions!

� Ophthalmological instruments must

not be treated in washers-disinfectors

with the Steam condenser with heat

recovery (DK+WRG) option!

� When using the heat recovery

function, do not lower the boiler

temperature of 90 °C. For sensitive

wash ware such as optical instruments

for example, select a process without

thermal shocks. For instance, use warm

water in the last purge block before the

thermal disinfection.

� In the case of very delicate wash

ware, please verify whether it is suited

for the treatment in a process with

potential changes in temperature.

Please follow the instructions provided

by the manufacturers of the wash ware

and the instruments.

Optional equipment DK + WRG

76

Applicable standard programs

Due to the particular requirements to be taken into account in

connection with the Steam condenser with heat recovery

(DK+WRG) option, only the following standard programs can

be used without customization:

Program name Intended application

DES-VAR-TD General cleaning/disinfection process (A0=3000)

DES-VAR-TD-AN Cleaning/disinfection process for anaesthetic

instruments/devices (A0=600)

VAR-TD-NR Cleaning/disinfection process for alkali-sensitive instruments

with a pH-neutral cleaning agent (without neutralization)

(A0=3000)

ORTHOVARIO * Oxidative vario program with aluminum tolerance (A0=3000)

OXIVARIO * Alkali-oxidative vario program for intensive cleaning

(A0=3000)

OXIVARIO PLUS * Recommended for the prevention of vCJD (A0=3000)

* For special washing programs, see Chapter "Options

OXIVARIO / ORTHOVARIO".

All other standard programs must be customized, if

necessary. See the Programming Manual, Chapter "Edit

programs".

If the standard programs require customizing or if new

programs must be created, you must note the

corresponding program requirements; see also the

Programming Manual, Chapter "Option DK with WRG".

When using the automatic mobile unit recognition, the

coding of the mobile units must be assigned only to

programs customized for Steam condenser with heat

recovery (DK+WRG)!

Optional equipment DK + WRG

77

Old electric and electronic equipment

often contains valuable materials.

However, it also may contain harmful

substances that were necessary for

safe use. Improperly disposing of

appliances in your household waste

can be detrimental to health and the

environment. This is why it is important

NEVER to dispose of your old

appliance in the regular household

garbage.

Instead, please contact your local

community waste collection point to

return and recycle electric and

electronic appliances.

If the appliance is being stored for

disposal, please make sure it has been

child-proofed.

Disposing of an old appliance

78

79

M.-Nr. 10 317 051 / 00

0297

Subject to change without notice / Creation date 2016-01-08