Upload
others
View
0
Download
0
Embed Size (px)
Citation preview
UvA-DARE is a service provided by the library of the University of Amsterdam (http://dare.uva.nl)
UvA-DARE (Digital Academic Repository)
Studies on cardiac pacing : emphasis on pacemaker sensors and cardiac resynchronizationtherapy
Yılmaz, A.
Link to publication
Citation for published version (APA):Yılmaz, A. (2005). Studies on cardiac pacing : emphasis on pacemaker sensors and cardiac resynchronizationtherapy. Amsterdam: Amsterdam University Press.
General rightsIt is not permitted to download or to forward/distribute the text or part of it without the consent of the author(s) and/or copyright holder(s),other than for strictly personal, individual use, unless the work is under an open content license (like Creative Commons).
Disclaimer/Complaints regulationsIf you believe that digital publication of certain material infringes any of your rights or (privacy) interests, please let the Library know, statingyour reasons. In case of a legitimate complaint, the Library will make the material inaccessible and/or remove it from the website. Please Askthe Library: https://uba.uva.nl/en/contact, or a letter to: Library of the University of Amsterdam, Secretariat, Singel 425, 1012 WP Amsterdam,The Netherlands. You will be contacted as soon as possible.
Download date: 25 Feb 2020
Individuall optimization of pacing sensorss improves exercise capacity withoutt influencing quality of life
Aytenn Erol-Yilmaz MD. Tim A, Schrama, Jutta Schroeder Tanka*, MD PhD, Jann G. Tijssen PhD, Arthur A. Wilde MD PhD, and Raymond Tukkie MD PhD
Fromm the department of Clinical and Experimental Cardiology, Academic Medical Centerr and the department of Cardiology. Sint Lucas-Andreas hospital*, Amsterdam,
Thee Netherlands.
PACEPACE 2005;28:17-24
Abstract t
Introductio n n
Programmablee pacemaker sensor features are frequently used in default setting.
Limitedd data are available about the effect of sensor optimization on exercise
capacityy and quality of life (QOL). Influence of individual optimization of sensors
onn QOL and exercise tolerance was investigated in a randomized, single blind study
inn patients with WIR. DDDR or AAIR pacemakers.
Materiall and methods
Patientss with >75% pacing were randomized to optimized sensor settings (OSS) or
defaultt sensor setting {DSS). Standardized optimization was performed using three
differentt exercise tests. QOL questionnaires {OOL-q: Hacettepe, Karolinska and
RAND-36)) were used for evaluation of the sensor optimization. One month before
andd after optimization, exercise capacity using CAEP and the three QOL-q were
assessed. .
Results s
Fifty-fourr patients (26 men. 28 female) with a mean age of 65 16 years were
enrolledd in the study. In each group (OSS and DSS) 27 patients were included. One
monthh after sensor optimization the achieved maximal HR and METS were signifi-
cantlyy higher in OSS compared to DSS (124 28 vs. 108 20 bpm, p=0.036; 7.3 4
vs.. 4.9 4 METS, p = 0.045). Highest HRand METS were achieved in patients with
pacemakerss with accessible sensor algorithms. In patients with automatic slope
settingss (33 %). exercise capacity did not improve after sensor optimization. QOL
didd not improve in OSS compared to DSS.
Conclusions s
Afterr 1 month of individual optimization of rate response pacemakers, exercise
capacityy was improved and maximum heart rate increased, although QOL remained
unchanged.. Accessible pacemaker sensor algorithms are mandatory for individual
optimization. .
6 6
Chapterr 4
Introductio n n
Thee normal heart adapts its rate in response to the body's changing metabolic
demandss and is therefore chronotropically competent. Earlier generation of pace-
makerss provided constant rate and were unable to maintain or restore chronotropic
competence.. Since the first permanent pacemaker implantation in 1958, there have
beenn tremendous advances in pacemaker technology with the availability of com-
plex,, multiprogrammable, dual chamber and rate adaptive pacemakers which could
meett the haemodynamic needs of an individual patient.1"3 Rate adaptive pacing
improvess cardiac output, exercise tolerance and a sense of well-being compared to
fixedd rate pacing.4
Normall sinus node function is used as the golden standard for the development of
thee ultimate pacemaker sensor. Sinus node behavior can be described with several
methods:: exercise testing protocols, the mathematical model according to Wilkoff
andd Holter monitoring in daily life. These methods together with evaluation of the
obtainedd quality of lif e (QOL) associated with a given sensor can all be used to
assesss the efficacy of a pacemaker sensor. The American college of cardiology
guideliness of exercise testing used for evaluating rate adaptive pacing advices
adjustmentt of the exercise test for the subjects need and however several exercise
testss (CAEP, LITE, Bruce, six minute walk test, Kaltenbach step test, stair climbing)
aree described, consensus lacks about the ideal exercise test protocol for pacemaker
sensorr optimization. 8"10' I4"19
Programmablee pacemaker sensor features are frequently used in default sensor
settingg (DSS) in daily practice, although rate responsive pacing has been reported to
improvee the physical capacity and QOL compared to fixed rate pacing. It is un-
knownn whether individual optimization of rate response is a necessary factor to
improvee QOL and exercise capacity with the existing sophisticated sensor technolo-
gyy compared to the default sensor setting (DSS) of the manufacturer.8 Therefore,
wee tested the effect of optimized sensor setting (OSS) with a standardized individu-
all optimization protocol and DSS on exercise capacity and QOL in patients with
AAIR,, W1R and DDDR pacemakers in a prospective single blind randomized trial.
Material ss and Methods
Patientt population
Fifty-fourr consecutive patients with >75% pacemaker sensor driven HR with mean
agee of 65 16 years (range 27 to 89 years), New York Heart Association (NYHA) class
I-- II. stable medication, stable psychosocial conditions and commercially available
pacemakerss (Medtronic Minneapolis USA. Guidant Minnesota USA, Vitatron Arn-
hemm The Netherlands) were included at least 3 months after pacemaker implanta-
tion.. Pacemaker type and pacing indication are described in table 1.
Fromm our existing institutional database of exercise testing in healthy controls (HO
withoutt co-morbidity and medical therapy , we selected nineteen age and exercise
typee matched individuals (mean age 66.5 6). All patients gave written informed
Tablee 1. Demographics and clinical variables
Patients s
Numberr of patients Agee (year) Sexx Imale/femalel [%) BMII ikg/m2} NYHAA class \l-4) LVEFF (normal/reduced. %) History y
CoronaryCoronary artery disease (%' HeartHeart valve insufficiency (%) Atria!Atria! fibrillation {%)
Medication n DigoxwDigoxw (%l ^-blocker^-blocker (%) CalciumCalcium antagonist {%} AmiodaroneAmiodarone {%)
Facemakerr type VitatronVitatron \%) MedtronicMedtronic (!fci GuidantGuidant i%1
Pacemakerr indication SSSW SSSW AVV nodal block (%i AFAF with AV nodal disease [%) AFAF with His bundle ablation iV Other Other
DSS S
27 7 066 16
59/41 1 299 6
1.77 0.9 89/11 1
4 4 6 6 42 2
4 4 10 0 6 6 2 2
41 1 33 3 26 6
15 5 15 5 11 1 52 2 7 7
OSS S
27 7 655 15 37/63 3 266 4 1.88 9 89/11 1
6 6 12 2 58 8
4 4 14 4 10 0 2 2
60 0 29 9 11 1
22 2 15 5 18 8 42 2
3 3
Tablee 1. DSS= default sensor setting: OSS= optimal sensor setting: BMI - body mass index: NYHAA = New York Heart Association: LVEF= left ventricular ejection fraction: SSS~sick sinus syndrome.. AV = atrioventricular: AF = atrial fibrillation.
78 8
Chapterr 4
consentt and the ethics committee of our institutio n approved the study.
Randomizationn protocol and study design
I nn thi s s ingle bl in d prospect ive randomized trial , pat ients were firs t randomized to
OSSS or DSS. Primar y endpoint of the study is exercise capacity. Patients in OSS were
thenn randomized to one of three exercise protocols used for opt imizat ion of the
sensor.. The three exercise tests were the six minut e hall walk test (6-HWT) , six
m inu t ee hall walk test wi t h stair c l imbin g (6-HWT+SC) and the chronotropi c assess-
mentt exercise protocol (CAEP) according to Wilkoff. 11
Al ll pat ients underwent at basel ine a CAEP exercise test wi t h posture change (PC)
andd sui tcase lift in g (SL) wi t h their pacemakers programmed in default sett ing. The
pacemakerss of the pat ients in OSS group were individual l y opt imized 1 month after
basel ine.. The CAEP test was repeated after 2 mon ths in OSS and DSS groups.
Qual i t yy of l i f e quest ionnaires (QOL-q) were assessed in all pat ients, 1 month before
andd after sensor opt imizat ion (see figur e 1).
Patients s
Baseline: : Groupp OSS (n = 271
CAEPP + PC+ SL in default
Groupp DSS (n = 27)
22 months
Afterr I month optimization using ++ QOL
Afterr 2 months: QOL +
t t
HWT T
t t
HWTT + SC
CAEPP + PC+ SL CAEP++ PC+ SL
Figuree 1. Randomization and flowchart of the study. OSS- optimalized sensor setting: DSS-defaultt sensor setting: CAEP- chronotropic assessment exercise protocol; PC- posture change:: SL- suitcase lifting: QOlq = quahty of lif e questionnaires: 6-HWT= 6 minute hall walk lest: SC== staircase ascent and descent.
79 9
Sensorr optimization
Alll sensor optimizations were done by one investigator (A-E.Y). The optimized
pacemakerr sensor setting was individually determined by 3 parameters: 1) detailed
analysiss of HR curve obtained with the exercise test after 1 month (onset of exer-
cise,, slope, total exercise time, time to peak HR, maximal HR). One of the three
exercisee tests (depending on randomization) were used for optimization (CAEP, 6-
HWTT or 6-HWT+SC). HR curves of 19 age matched HC (see figure 2) and the availa-
blee literature about normal HR during exercise were used as reference.12 2) upper
ratee limi t was programmed according to Astrand (220 - age)13 and 3) development of
complaints,, Threshold, lower rate limi t (LRL), upper rate limi t (URL), slope and
sensorr specific settings (sensor blending in dual sensor systems) were adjusted to
obtainn the predicted optimal sensor setting.
HR R
180 0 - * -- 6-HWT
P P
CAEP-HC C
- * -- 6-HWT-HC
i—i—i—i—n—i— rr r r t i i i—i—r~i—i—i—r—i—i—i—n—!—I—i—i—i—i—i—i— r
11 4 7 10 13 16 19 22 25 28 31 34 Minutes s
Figuree 2. Changes in heart rate ImeanV HR-heart rate; 6-HWT= 6 minute hall walk test: CAEP = chronotropicc assessment exercise protocol. HC = healthy controls.
SO O
Chapterr 4
Exercisee test protocols
PosturePosture change and suitcase lifting
Alll patients were examined in the same conditions (before noon, uniform room
temperaturee and footgear (patients own shoes)). After instrumentation, patients
restedd supine on an examination table with one pillow for 5 minutes. They then
elevatedd to the sitting position and immediately to the standing position for 2
minutes.. The patients raised a standard suitcase (9 kg weight and measuring 47 cm
xx 37 cm x 15 cm) from the floor onto the examination couch (a height of 100 cm)
usingg their preferred arm (left, right, both).
ChronotropicChronotropic assessment of exercise protocol
Alll patients underwent a symptom limited treadmill test using the CAEP protocol
accordingaccording to Wilkoff .
Onee MET equals 35 ml oxygen uptake/kg body weight/min. representing the
approximatee metabolic cost to stand quietly. In his protocol oxygen consumption
andd carbon dioxide production was not performed, and thus metabolic workload
(METSS ) was not directly measured during exercise. Rather, metabolic levels during
eachh stage of exercise were estimated using treadmill grade and speed.
SixSix minute hall walk test
Afterr 5 minutes of rest patients were brought to the parcour. A parcour of 100 m
wass created by attaching stickers each meter in an oval form showing the walk
distance.. Patients were instructed to walk or run for 6 minutes at the parcour after
hearingg the start sign. After the symptom limited 6-HWT, the patients were brought
backk to the test room for a recovery period of 10 minutes.
StaircaseStaircase descent and ascent
First,, the patients descended 5 flights of stairs as rapidly as possible (82 steps, with
totall horizontal distance of 41.5 m and vertical distance of 15 m). After 2 minutes of
rest,, the patients then ascended the same flights as fast as they could.
81 1
Qualityy of lif e
Too compensate for the limited pacemaker patient specific OOL-q, we used 3 OOL-q.
Thee Rand-36 consisting 9 domains (physical functioning, social functioning, role
functioningg physical problems, role functioning emotional problems, mental health,
energy,, domains bodily pain, general health perception and change in general
health)) was used as a generic core module. The Rand-36 was evaluated using the
scoree system from 0 to 100 % in each domain (0% =low OOL to 100% = high QOL). H
Thee second questionnaire assessed was the Hacettepe QOL-q. which is a pacemaker
patientt specific questionnaire with 8 domains (general well-being, physical activity
andd symptoms, sleeping dysfunction, appetite, sexual functioning, cognitive
functioning,, social participation and work performance). The scores of each domain
aree between 3-50 points. (3 = low QOL and 50 = high OOL).
Fromm the final questionnaire, the Karolinska OOL-q only A domains (chest pain,
palpitations,, dizziness, dyspnoea) were used as complementary to the other two
questionnaires.. The 4 domains of the Karolinska QOL-q were evaluated using Visual
Analogg Scales and required patients to place a mark along a line of 10 cm in length
fromm a minimum of 0 (no complaints) to a maximum of 10 (maximal grade of
complaints).. The results were expressed as a percentage of the distance from the
discretee minimum point to the position of the mark divided by the length of the
line.155 In the Rand-36 and the Hacettepe QOL-q a high score means a high OOL and
inn the Karolinksa QOL-q a high score means a low QOL.
Measurements s
Beatt to beat HR was recorded during the physical tests with the Polar advantage
systemm using electrodes mounted in a belt (Polar Electro OY, Kempele, Finland).
Duringg the CAEP exercise test patients were also continuously monitored by 12-lead
electrocardiographicc recordings. HR at rest, time to peak HR, maximal achieved HR.
HRR at 10 minutes recovery period, exercise duration and METS were measured, The
OOL-qq was assessed after 1 and 2 months of inclusion.
82 2
Chapterr 4
Statisticall analysis
Powerr analysis was performed on the primary endpoint of the study, exercise
capacityy in METS. Assuming that the common standard deviation is 35 and a mean
differencee in achieved METS is 3 (the difference between mean DSS l and OSS, JJL2
inn METS ), we calculated the power of 80 % using a two group t-test with a 0.05 two-
sidedd significance level. A sample size of 23 patients in each group DSS vs. OSS had
aa power of 80% to detect a difference in means of 3. Parametric data was analysed
usingg the Student's t-test while nonparametric data was analysed using the Mann-
Whitneyy U test. All data are expressed as mean SD. A P value <0,05 is considered
statisticallyy significant.
Results s
Fifty-fourr patients (26 men, 28 female) with a mean age of 65 16 years (range 27
too 89 years) were enrolled in the study. In each group (OSS and DSS) 27 patients
weree included. Demographics and baseline characteristics between both groups
weree not significantly different (see table 1).
Fivee patients (1 in OSS and 4 patients in DSS) withdrew from the study. The patient
inn group OSS stopped because of diagnosis of a pulmonary tumour. In group DSS,
threee patients stopped after programming to DSS because of symptoms of heart
failure.. The forth patient withdrew for psychosocial reasons.
Pacemakerr and sensor type
Thee majority of patients (58% in OSS group vs. 39% in DSS group) had pacemakers
equippedd with dual sensor systems of Vitatron (QT and activity. Vitatron. Arnhem
Thee Netherlands). Activity sensors were the most frequent used sensors in the
singlee sensor systems, The OSS group had 31% of activity sensors of Medtronic
(Medtronic.. Minneapolis USA), while the DSS group had 35% Medtronic and 9%
Guidantt (Guidant. Minnesota USA) activity sensors. Accelerometers of Guidant
(Guidant,, Minnesota USA) were used 17% in the DSS group and 11% in OSS.
83 3
Tciblee 2. Pacemaker optimization settings
Mode e Baseline e
WIRR i%! WII i%* DDDRR i%i ODDD i%l AAIRR i%̂ AAII i%l
4 4
26 6 9 9 4 4 7 7
fticemakerfticemaker settings changed Baseline
Afterr optimization
59 9 0 0 35 5 0 0 6 6 0 0
Afterr optimization Frequency of change (%)
LRMbprrO O URLL ibpm) THH ilow, medium/low. medium,, medium/high, high' S l o pee
Sensor r
60 0 1211 4.4f medium m
I I
default t QT=ACT T
611 2.7
14544 14.31 low,, medium/low
moree agressive OKACT.. ACT
12 2 92 2 11 1
15 5 31 1
Tablee 2. LRL = lower rate limit ; URL = upper rate limit : TH^threshold: QT= OT sensor: ACT = activityy sensor: \*) dependent of the manufacturer: response factor, response time or acceleration time:: blending OT=ACT-> 50:50. QT<ACT-> 25:75: t URL after optimization is significantly different,, p <0.001.
Sensorr optimization
Ratee response was programmed on in all patients. The LRL was changed in 12% of
thee patients but increased not significant. The URL was programmed higher in 92%
off patients, from 121 4.4 to 145.4 14.3 bpm, (p<0.001). The threshold was
changedd in 11% of the patients: from medium to medium/ low in 4% vs. medium to
loww in 7 % and the slope settings were changed in 15 % of the patients. The re-
sponsee factor vs. response time vs. acceleration of the slope was programmed in a
moree aggressive setting: from factor 8 to 14 in 7 % vs. 30 to 10 seconds in 4 % vs.
standardd to fast in 4 %. In 33% of patients, the slope settings could not be adjusted
duee to the autoslope setting inVitatro n pacemakers (Arnhem, the Netherlands),
despitee the need for a more aggressive slope setting. Sensor blending was changed in
311 %. In 12% the default sensor blending setting (QT = ACT) of the Vitatro n pacemak-
erss (Arnhem, the Netherlands) were programmed to activity only and in 19 % to
OT<ACT ,, see table 2.
84 4
Chapterr 4
Tablee 3. Summary of HR during rest, exercise and recovery
Restingg rate ibpm.1-Timee to peak HR ymui' Programmedd maximum Achievedd maximum HR 100 mill recovery ibpm.1
METS S
Exercisee duration mini. M.tximall HR during SI. Posturee change
HRR \bpm l
^bpm.1 1
DSS S
Baseline e
vn-27^ ^ Default t
677 14
100 5 1211 4
1099 21 744 10 66 4 111 5
711 12 722 11
22 months
li ii = 231 Default t
633 29 100 7
I21 4" 11 OS 20 #
666 6 4.99 4"
100 5 688 10
688 + 6
Baseline e
inn = 27̂ Default t
711 14
n 5 121 4
1144 15 777 10
7 4
5 5 711 13 700 10
inn = Q̂ CAEP P
700 13 111 +. 0
1477 + I4f I I SS 9 699 15
66 + 4 111 5
7 6 + 17 7
766 5
OSS S
22 months
inn = 81 6-HWT T
666 + 7 100 4
1422 12 S 128++ 34
65 7 7.88 4*
122 5 9 9
67 6
m-Ql l &-HWTT + SC
666 9 1 1 +5 5
1477 18+ * 1233 29 #
62 8
7.99 4* 133 * 4 68 8 67 7
P-value e
ns s ns s
t<0 .001 1 -*1== 0.036
ns s
0.045* * ns s ns s nn s
Tablee 3. HR= heart rate; bpm = beats per minute: min = minutes; DSS = default sensor setting. OSSS = optimalized sensor setting: CAEP= chronotropic assessment exercise protocol; 6-HWT = 6 minute halll walk test; SC= staircase ascent and descent; SL = suitcase lifting; #-11= the achieved maximal heart ratee is significantly higher in the total group of OSS compared to DSS, p = 0.036; * = the achieved METS wass significantly improved in the subgroups 6-HWT and 6-HWT +SC compared to These three exercise testss iCAEP. 6-HWT and ó-HWT +SO were used for optimization. DSS. p = 0.045.
HR R
140 0
1300 -
120 0
110 0
100 0
90 0
80 0
70 0
60 0 50 0
40 0
-*-DSS-B B -- 2 months
Rest t Exercise e Recovery y ~ii i r r T r ~!! T
33 6 9 12 15 18 21 24 27 30 33 36 Minutes s
Figuree 3. Changes in heart rate tinea nV HR = heart rate. DSS-B = default sensor setting at baseline.
S5 S5
HR R 160 0
140 0
120 0
100 0
SO O
60 0
40 0
-è^^ OSS-B 22 months
Rest t Exercise e Recovery y
- x -- -x--x-
11 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 Minutes s
Figuree 4. Changes in heart rate (mean). HR-heart rate: OSS-B- optimalized sensor setting at baseline.. *P = 0.036.
Exercisee test protocols
PosturePosture change, suitcase lifting
Thee achieved maximal HR during posture change and suitcase liftin g were compara-
blee in OSS and DSS at baseline and after 2 months (see table3). One patient could
nott lif t the suitcase.
chronotropicchronotropic assessment exercise protocol
Thee achieved maximal HR and METS after sensor optimization were significantly
higherr compared to DSS (124 28 vs. 108 20. p = 0.036: 7.3 4 vs. 4.9 4 METS,
p=0.045). .
Highestt increase in maximal HR, METS and exercise duration was achieved in the
subgroupss whereby the sensor was adjusted with 6-HWT (128 34 vs. 7.8 4 vs.12
86 6
Chapterr 4
5)and6-HWT +SC (123 29 vs. 7.9 4 vs. 13 4) compared to the subgroup
CAEPP (118 19 vs. 6 4 vs. 11 5Ï-
Otherr measured parameters between the groups and subgroups like resting rate,
timee to peak HR, HR at 10 minute recovery, maximal HR during suitcase liftin g and
posturee change were comparable between the two CAEP tests (see table 3, figure 3
andd figure 4).
Qualityy of lif e
Thee achieved scores from all the three QOL-q (RAND-3Ó, Hacettepe and Karolinska)
onee month after optimization compared to DSS were not significantly different.
Alsoo scores attained within the subgroups of the three QOL-q were not significantly
different. .
Discussion n
Exercise e
Ourr study shows that individual optimization of pacemaker sensors results in
improvedd exercise capacity. Also, after sensor optimization the reason for stopping
thee exercise test shifted to concomitant restrictions (e.g. physical activity level of
thee elderly pacemaker patients) rather than cardiovascular limitations. The in-
creasedd METS and HR indicate that patients have an improved exercise capacity
afterr sensor optimization
Thee highest benefit from optimization in achieved METS and maximal HR was
obtainedd with the 6-minute HWK and 6-minute HWK+ SC. In the CAEP subgroup,
theree were 33% more pacemakers with automatic slope settings, which could not be
individuallyy adjusted. Other possible factors, which could influence the exercise
capacityy between the groups. like medical history and medical therapy, were not
significantlyy different. These results underline the importance of individually
adjustmentt of pacemaker sensors and the necessity of accessible pacemaker
algorithms. .
Too date, limited studies investigated the effect of sensor adjustment and to the best
87 7
off our knowledge: our study is the first and largest randomized controlled study,
comparingg DSS with OSS in detail. Previous studies mainly compared different
sensorss and were not primarily designed to evaluate individual adjustment,5
Sulkee et al. showed that appropriate programming of sensors is crucial in rate
responsivee pacing. & In 20 patients the effects on exercise and QOL of appropriate,
overr and under programming of the sensor were assessed. In contrast to our study,
onlyy activity sensors were evaluated and the sensors were optimized according to
thee manufacturer's instructions without detailed insight in these instructions. In
anotherr study, Klonis et al. investigated whether automatic algorithms for sensor
optimizationn could reduce clinical follow up time compared to manual adjustment.
Usingg the activity level of patients (more active, the same, or less active) they
concludedd that automatic adjustment is less time consuming than manual adjust-
ment,, This study supports the current trend in increasing automaticity in pacemak-
ers,, although it is disputable if such a rough guide as three levels of activity can be
usedd as a guide for sensor optimization. In general, sensor optimization methods
aree poorly described, mostly referring to the manufacturers advice.
Inn agreement with previous reports, exercise duration was not significantly im-
provedd after sensor optimization.8 2D This can partly be explained by the category of
patientss in need of permanent pacing (elderly and relatively sedentary). In these
patients,, exercise is often limited by loss of muscle strength and mass rather than
cardiopulmonaryy capacity. This loss of muscle strength is particularly apparent
whenn exercise testing is performed on a bicycle or treadmill. The CAEP exercise test
consistt nonlinear characteristics. The first 10 minutes it requires a low metabolic
workload,, beyond which it increases abruptly. Patients with preserved functional
capacity,, capable of exercising for more than 10 minutes, may quit before reaching
maximall 02 uptake, mainly because of excessive increments in workload near the
endd of the test, thus, being limited by mechanical rather than metabolic barri-
ers.12211 In our study there is a trend to increase of exercise duration with a mean of
11 5 minutes. This increased exercise duration results in a higher achieved speed
andd steeper slope on the CAEP. And this longer exercise duration is achieved due to
thee higher achieved heart rate. The symptom limited 6 minute hall walk test resem-
bless closer daily activities and less time consuming compared to the treadmill or
bicyclee exercise tests. 22
Thee healthy controls achieved higher HR than patients despite individual optimiza-
tion,, probably because current sensors are still hypochonotropic and physicians feel
88 8
Chapterr 4
reluctancee to program such high upper rates due to concomitant heart disease. The
maximall HR in the default setting is generally too low, well below the age predicted
maximall HR of the mean age of a typical pacemaker patient of 70 year, according to
thee Astrand formula (220-age). An inappropriately low programmed maximum sensor
ratee and failure to reach maximum exercise will result in sensor indicated pacing rates
duringg exercise testing below the calculated expected HR. The clinician may attempt
too compensate for this by programming more aggressive rate response parameters
whichh may result in excessive rate response behavior of the pacemaker during the
patient'ss normal ambulatory activities.0
Qualityy of lif e
Qualityy of life did not improve with individual sensor optimization. Specific sub-
groupss that derived benefit were not observed, including stratified to device indica-
tion,, age or specific complaints. Again, only the study of Sulke et al. assessed the
effectt of rate responsive pacing on QOL. Appropriate, over and under programming
off the rate response were evaluated in both dual and single chamber activity sensor
ratee adaptive pacemakers with visual analog scales and specific activity question-
naires.. Symptoms were least in the appropriate programmed pacemakers after 2
weekss of follow up. The found differences between both studies could be explained
duee to the greater programming steps in Sulke's study (rate response off, appropri-
atee programming and aggressive programming). Another explanation for the failure
too improve QOL after sensor optimization could be the relatively good baseline
functionall capacity, because patients with relative preserved functional capacity at
enrollmentt show the lowest improvement in health related values.25
Alsoo in large pacemaker trials, improvements in QOL were minimal. The PAcemaker
Selectionn in the Elderly study (PASE, n=407) showed only in the subgroup with sick
sinuss syndrome a moderate improvement in QOL in patients with dual chamber
pacingg as opposed to ventricular pacing, whereas in the Canadian Trial Of Physiolog-
icc Pacing investigating (CTOPP, n = 172II only the global well being score was better
inn the physiologic vs. WIR mode.
Inn general, there is a lack of validated pacemaker patient specified quality of life
QOL-q.244 To date, only two questionnaires were specifically developed for pacemak-
err patients (Hacettepe Qol-q2\ Aquarel QOL-q). This complicates comparing
studiess using different QOL-q's and methods for assessing QOL (transtelephonic,
89 9
selff administered at hospital by researcher /patient). Therefore, it is advisable to
interprett QOL data in pacemaker patients cautiously.15 A The recent developed
Aquarell Qol-q is an accurate QOL-q for study design as our present study, therefore
wee suggest to use this OOL-q in future studies. We could not use this QOL-q for our
studyy because it was not available.
Inn this study we showed an improvement in exercise capacity without improvement
inn QOL. The reverse is shown in cardiac resynchronization therapy studies with
onlyy minor changes in exercise capacity yet major changes in QOL. The PATH-CHF
studyy there was only a minor improvement of 60 meters after cardiac resynchroniza-
tionn therapy while QOL improved clearly.2' These disconnect between exercise
capacityy and QOL can be explained possibly due to several factors. The pacemaker
implantationn has already a large impact and therefore it is more difficult to obtain
moree additional improvement. In this study we compared in hospital tests with
evaluationn of QOL based on the activities at home. As described above the in- hospital
testss are artificial, therefore it would be theoretically and scientifically better when
wee compared the sensor function at home using holter registration with QOL.
Conclusions s
Afterr 1 month of individual optimization of rate response pacemakers, exercise
capacityy and maximum HR was improved, although QOL remained unchanged.
Accessiblee pacemaker sensor algorithms are mandatory for individual optimization,
althoughh automatic features are indisputably important in the pacemaker sensor
development. .
Developmentt of sensor algorithms which create the possibility for monitoring
sensorr behavior in detail at home is a great challenge to overcome the disadvantages
off hospital exercise tests for sensor optimization.
Acknowledgments s
Thee authors thank Wandena Ramsoekh. pacemaker technician for the technical
supportt and Michael Kortz, MD (department of Cardiology, Flevohospital. Almere.
Thee Netherlands) for helping with the inclusion of patients.
Q0 0
References s
Chapterr 4
1.. Lass J, Kaik J, Meigas K, Hinrikus H, Blinowska A. Evaluation of the quality of rate adaptation algorithmss for cardiac pacing. Europace. 2001:3:221-8.
2.2. Sack S. [Modern pacemaker therapy]. Herz. 2001;26:1. 3.. Elmqvist R. Review of early pacemaker development. Pacing Clin Electtophysiol. 1978:1:535-6. 4.. Leung SK. Lau CP. Tang MO. Cardiac output is a sensitive indicator of difference in exercise
performancee between single and dual sensor pacemakers. Pacing Clin Electrophysiol. 1998;21:35-41. .
5-- Zegelman M. Cieslinski G, Kreuzer J. Rate response during submaximal exercise: comparison off three different sensors. Pacing Clin Electrophysiol. 1988:11:1888-95.
6.. Freedman RA, Hopper DL, Mah J. Hummel ). Wilkoff BL. Assessment of pacemaker chrono-tropicc response: implementation of the Wilkoff mathematical model. Pacing Clin Electrophys-iol.iol. 2001:24:1748-54.
7.. Lau CP. Leung SK. Guerola M, Crijns HJ. Comparison of continuously recorded sensor and si-nuss rates during daily lif e activities and standardized exercise testing: efficacy of automatical-lyy optimized rate adaptive dual sensor pacing to simulate sinus rhythm. Pacing Clin Electro-physiol.physiol. 1996;19:1672-7.
8.. Sulke N. Dritsas A. Chambers J. Sowton E. Is accurate rate response programming necessary? PacingClinPacingClin Electrophysiol. 1990:13:1031-44.
9-- Strobel JS, Kay GN. Programming of sensor driven pacemakers. Cardiol Clin. 2000:18:157-76. ix. 10.. Kay GN. Quantitation of chronotropic response: comparison of methods for rate- modulating
permanentt pacemakers. J Am Coll Cardiol. 1992:20:1533-41. 11.. Wilkoff BL, Miller RE. Exercise testing for chronotropic assessment, Cardiol Clin. 1992:10:705-
17. . 12.. Ellestad MH. stress testing principles and practice.tirth edition:567-573-13.. Astrand. Aerobic work capacity its relation to age, sex. and other factors, circulation research.
1967;20:1-2111 1-217. 14.. Hays RD, Sherbourne CD, Ma2el RM. The RAND 36-Item Health Survey 1.0. Health Econ.
1993:2:217-27. . 15.. Linde C. How to evaluate quality-of-life in pacemaker patients: problems and pitfalls. Pacing
ClinClin Electrophysiol. 1996:19:391-7. 16.. Klonis D. Zhang X, Patel U. Gulamhusein S, Patel J, Hurwit H. Banish D. Marco D. Automatic
sensorr algorithms expedite pacemaker follow-ups. Pacing Clin Electrophysiol. 2003:26:225-8. 17.. Sulke N, Pipilis A, Bucknall C. Sowton E. Quantitative analysis of contribution of rate re-
sponsee in three different ventricular rate responsive pacemakers during out of hospital activi-ty.. Pacing Clin Electrophysiol. 1990;13:37-44.
18.. Bellamy CM, Roberts DH, Hughes S, Charles RG. Comparative evaluation of rate modulation inn new generation evoked QT and activity sensing pacemakers. Pacing Clin Electrophysiol. 1992:15:993-9. .
19.. Mahaux VA. Verboven Y], Waleffe A, Kulbertus H. Stepwise analysis of the calibration proce-duree of an accelerometer-based pacemaker. PacingClin Electrophysiol. 1994;17:1955-9-
20.. Hasegawa A. Hatori M, Amano M, Iijima T. Adachi H, Yamaguchi E. Fukuda T, Murata K. Nagai R.. Adequacy of pacing rate during exercise in rate responsive ventricular pacing. Pacing Clin Electrophysiol.Electrophysiol. 1997:20:307-12.
21.. Page E. Bonnet JL, Durand C. Comparison of metabolic expenditure during CAEP versus a test adaptedd to aerobic capacity (Harbor test) in elderly healthy individuals. Pacing Clin Electro-physiol.physiol. 2000;23:1772-7.
22.. Hayes DL, Von Feldt L, Higano ST. Standardized informal exercise testing for programming ratee adaptive pacemakers. Pacing Clin Electrophysiol. 1991;14:1772-6.
23.. Lopez-Jimenez F, Goldman L. Orav EJ. Ellenbogen K. Stambler B, Marinchak R, Wilkoff BL, Mangionee CM. Yoon C, Vitale K, Lamas GA. Health values before and after pacemaker implan-tation.. Am Heart J. 2002;144:687-92.
24.. Stofmeel MA. Post MW. Kelder JC. Grobbee DE, van Hemel NM. Quality-of-life of pacemaker patients:: a reappraisal of current instruments. Pacing Clin Electrophysiol. 2000:23:946-52.
91 1
25.. Oto MA, Muderrisoglu H, Ozin MB. Korkma2 ME. Karamehmetoglu A. Oram A, Oram E. Ugur-luu S. Quality of lif e in patients with rate responsive pacemakers: a randomized, cross-over study.. Pacing Clin Electrophysiol. 1991:14:800-6.
26.. Stofmeel MA. Post MW, Kelder JC. Grobbee DE, van Hemel NM. Psychometric properties of Aquarel,, a disease-specific quality of lif e questionnaire for pacemaker patients, ƒ Clin Epide-miolmiol 2001:54:157-65.
27.. Huth C. Friedl A, Klein H. Auricchio A. [Pacing therapies for congestive heart failure consider-ingg the results of the PATH-CHF study]. Z Kaidiol. 2001:90 Suppl 1:10-5.
92 2