1520MegOhmMeter
Users Manual
i
Table of Contents
Title Page
Unpacking the Meter......................................................................................... 1Safety Information and Symbols ....................................................................... 3Key functions .................................................................................................... 6
Display .......................................................................................................... 8Using the Meter ................................................................................................ 10
Connecting to the Circuit Under Test ............................................................ 10Auto-Shut-Off ................................................................................................ 12Measuring Insulation Resistance .................................................................. 12Measuring Low Resistance ........................................................................... 16Measuring Resistance................................................................................... 18Measuring Voltage ........................................................................................ 18Checking the Battery..................................................................................... 19
Maintaining the Meter ....................................................................................... 20Cleaning........................................................................................................ 20Replacing and Disposing of the Batteries...................................................... 21Testing and Replacing the Fuse.................................................................... 24
Replacement Parts and Optional Accessories .................................................. 27Principle of Measurement for Resistance.......................................................... 28Service Centers ................................................................................................ 28Specifications.................................................................................................... 29
ii
ii
List of Tables
Table Title Page
1. Key and Switch Descriptions..................................................................... 62. Display Description ................................................................................... 8
List of Figures
Figure Title Page
1. Shipment Content ..................................................................................... 22. Display ...................................................................................................... 93. Connecting to the Circuit Under Test ........................................................ 114. Display Example ....................................................................................... 195. Replacing the Batteries. ............................................................................ 236. Replacing the Fuse ................................................................................... 26
1
Fluke 1520
Unpacking the MeterThe Fluke Model 1520 MegOhmMeter (hereafter, “the meter”) is a handheld instrumentdesigned primarily to make resistance/insulation resistance measurements.The MegOhmMeter includes the following items, see Figure 1:• 2 test leads, red and black, 1.5 m• 2 test probes, red and black• 2 alligator clips, red and black• Hand strap• Carrying case• CD ROM (Not Pictured)
Fluke 1520Users Manual
2
acf02f.eps
Figure 1. Shipment Content
MegOhmMeterSafety Information and Symbols
3
Safety Information and SymbolsA Caution identifies conditions and actions that may damage the Meter. A WWarningidentifies conditions and actions that pose hazard(s) to the user. International symbols onthe Meter and in the manual are explained below.
Y Risk of electric shock J Earth
W See manual I Fuse
TEquipment protected by double orreinforced Insulation D AC or DC
M Battery Recycling information
Conforms to CSA C22.2 No 1010.1-92 +Amendment 2 1997, UL 3111 andANSI/ISA SP82.01 1994
VDE Conforms to VDEEN61010 (Pending)
P Conforms to EU directives t Conforms to UL 3111.1
CAT III
OVERVOLTAGE (Installation) CATEGORY ,,,, Pollution Degree 2 per IEC1010-1refers to the level of Impulse Withstand Voltage protection provided. Equipment ofOVERVOLTAGE CATEGORY ,,, is equipment in fixed installations (e.g., electricitymeter and primary over-current protection equipment.
Fluke 1520Users Manual
4
W Safety Information
Use of instrument in a manner not specified by the manufacturer may impairsafety features/protection provided by the equipment. Read the following safetyinformation carefully before using or servicing the instrument. To avoid electricshock or fire, do the following:• Avoid working alone.• Inspect the test leads for damaged insulation or exposed metal. Check test
lead continuity. Damaged leads must be replaced. Do not use the Meter if itlooks damaged.
• Be careful when working above 30 V ac rms, 42 V ac peak and60 V dc. Such voltages pose a shock hazard.
• When using the probes, keep your fingers away from probe contacts. Keepyour fingers behind the finger guards on the probes.
• Measurements can be adversely affected by impedances of additionaloperating circuits connected in parallel or by transient currents.
• Verify operation prior to measuring hazardous voltages (voltages above 30 Vac rms, 42 V ac peak and 60 V dc).
MegOhmMeterSafety Information and Symbols
5
W Safety Information (cont.)• Place test leads in proper input terminals.• Disconnect the live test lead before disconnecting the neutral test lead.• Do not use the Meter if the battery indicator (G ) shows a battery empty
condition.• Use only Fluke recommended batteries and fuse.• Do not use the Meter with any parts or cover removed.• Do not use the Meter around explosive gas, vapor or dust.• Disconnect the test leads from power sources and from the Meter before
changing the batteries or fuse.• Do not use the Meter in a wet environment.• Use only Fluke specified test leads.
Fluke 1520Users Manual
6
Key functions
Table 1. Key and Switch Descriptions
Rotary switchTo select a measurement function.
TUsed for the Insulation Resistance and Low Resistance test functions.Press and hold this button until the main reading is stable.
L
Locks the test in Insulation Resistance or Low Resistance function.To lock: press and hold TEST, then press LOCK and simultaneously releaseboth buttons. The icon V LOCK appears on the display.• Insulation Resistance- this mode continuously applies the test voltage
to the circuit to be tested. The beeper sounds every 2 seconds toremind you that you are in the Lock mode.
• Low Resistance- this mode continuously applies the test current to thecircuit to be tested.
To unlock: press LOCK or TEST again.
MegOhmMeterKey functions
7
Table 1 (continued)
NResistance Beeper functionTurns the Beeper function ON and OFF. When ON, the R icon appears onthe display and the Meter beeps at short circuit.
Z
Low Resistance functionTurns the test lead resistance compensation ON.
The X icon appears on the display. To compensate, touch the probe tipstogether, then press and hold ZERO until the Meter beeps. The main displayindicates 0.00.
CBacklight buttonTurns the display backlight ON and OFF.
Fluke 1520Users Manual
8
DisplayTable 2 and Figure 2 describe the display.
Table 2. Display Description
1 Voltage applied to the probes in Insulation Resistance function.
2 Low Resistance/Resistance function indicator.
3 Resistance reading held from the last measurement in Insulation Resistance or LowResistance function.
4 Beeper symbol shows if beeper function is turned on in Resistance function.
5 Zero symbol is on if test leads are zeroed out.
6 Main reading display for all functions.
7 Analog bar graph displays resistance on a logarithmic scale and voltage on a linearscale. The value always tracks the main reading.
8 High voltage warning symbol flashes if voltage ≥ 30 V ac or dc is present on the probes.
9 Battery life indicator. Displays briefly when the Meter is first switched to a function.Displays the amount of battery voltage left in increments of 25 %.
10 Lock symbol is on if the TEST mode is locked in Insulation or Low Resistance functions.
MegOhmMeterKey functions
9
1 32
67
910
8
45
acf12f.eps
Figure 2. Display
Fluke 1520Users Manual
10
Using the Meter
Connecting to the Circuit Under TestFigure 3 shows the proper connections.
W Warning
To prevent electric shock when performing resistance tests, remove allpower from the circuit to be measured.
W Warning
To prevent electric shock, first connect the test leads to the Meter inputsbefore you make connection to the circuit under test.
MegOhmMeterUsing the Meter
11
acf04f.eps
Figure 3. Connecting to the Circuit Under Test
Fluke 1520Users Manual
12
Auto-Shut-OffThe Meter automatically turns off after 15 minutes of non-use. In Lo Ohms mode, the meterturns off after 5 minutes of non-use. To turn the meter back on, turn the rotary switch toOFF, then to the desired function.
Measuring Insulation ResistanceW Warning
• Measuring insulation resistance requires the application of potentiallydangerous voltages to the circuit. This may include exposed bondedmetalwork.
• Before proceeding, ensure that the installation is correctly wired and nopersonnel are endangered by any tests.
MegOhmMeterUsing the Meter
13
To measure insulation resistance, do the following:
1. Select the test voltage.
2. Connect the probes to the circuit to be measured.The meter automatically detects if the circuit is energized, and displays the detected voltage.
W Warning
A repetitive beep and the flashing high voltage symbol (Y)warn the user if voltage more than 30 V ac or dc is present.If the voltage is > 30 V ac or dc, remove the voltage from thecircuit under test before proceeding with the next step.
acf05f.eps
3. Push and hold the TEST button. The upper left display shows the test voltage appliedto the circuit under test. The main display shows the resistance.
Note
The display shows - - - - until the TEST button is pushed.
Fluke 1520Users Manual
14
The Meter beeps when the reading is stable. The upper right display shows the sameresistance reading as the main display.When resistance is higher than the maximum display range, the meter reacts asfollows:• If the 250V range is selected, the display reads >1000 MΩ.• If the 500V range, is selected, the display reads >2000 MΩ.• If the 1000V range, is selected, the display reads >4000 MΩ.
4. While keeping the probes on the test points, release the TEST button.The circuit now discharges through the Meter, while the main reading shows thedecreasing voltage.Keep the probes touched to the test points until the circuit is completely discharged(the main display shows - - - -).The upper right display holds the resistance reading until a new test is started or adifferent function is selected.
MegOhmMeterUsing the Meter
15
Using the LOCK Function to Measure Insulation ResistanceThe LOCK function holds the test voltage on the probes. Use LOCK to make long durationmeasurements without having to push and hold the TEST button.To use the LOCK function:1. Press the TEST button, then press the LOCK button, and then release both
simultaneously.
W WarningIn this mode a potentially dangerous voltage is continuously applied to theprobes.In this mode, if the probes are disconnected from the circuit, the Metercannot discharge any potentially dangerous capacitive voltages left on thecircuit.
CautionIn this mode the Meter cannot indicate if the circuit is live.Ensure that the circuit is de-energized before connecting the test probes inthis mode.
2. Press LOCK or TEST to disengage the lock function.
Fluke 1520Users Manual
16
Measuring Low ResistanceTo measure low resistance:1. Zero out the test lead resistance (see Table 1).2. Connect the probes to the circuit to be measured.
If voltage is present on the probes, the voltage is displayed.W Warning
A repetitive beep and the flashing high voltagesymbol (Y) warn the user if voltage more than 30 V acor dc is present. If the voltage is > 30 V ac or dc,remove the voltage from the circuit under test beforeproceeding with the next step.
acf06f.eps
3. Press and hold the TEST button. A single beep indicates a stable reading. Theresistance shows on the main display. If the resistance is higher than 40 Ω, >40 Ω isdisplayed.
4. Release the TEST button. The upper right display holds the resistance reading until anew test is started or a different function is selected.
MegOhmMeterUsing the Meter
17
5. Exchange the red (+) and black (-) probes on the circuit and repeat steps 3 and 4 toreverse the polarity of the test current. The reading should be the same as theprevious. This test is useful to detect corroded connections, which can cause differentreadings for both polarities.
Using the LOCK Function to Measure ResistanceThe LOCK function is used to continuously supply the test current to the circuit to be tested.This is useful for making several measurements in succession without having to push andhold the TEST button for each measurement.1. Press the TEST button, then press the LOCK button, then release both simultaneously.2. You can now probe the desired test locations in succession.3. Press LOCK or TEST to disengage the lock function.
CautionThe Meter cannot indicate if the circuit is live in this mode.Ensure that the circuit is de-energized before connecting the test probes inthis mode or the fuse may blow.
Fluke 1520Users Manual
18
Measuring Resistance1. Connect the probes to the circuit to be measured.
Measure voltage first to ensure that no hazardous voltage ispresent, then switch to Ohms.
2. If the resistance is approximately 30 Ω or less, the Meterbeeps. To turn off the beeper, press the beeper button. If theresistance is higher than 4000 Ω, >4000 Ω is displayed.
acf07f.eps
Measuring Voltage1. Connect the probes to the circuit to be measured.2. If the voltage is higher than 660 V, >660 V is displayed.
W Warning
The Meter indicates either ac or dc voltage. If thevoltage being measured has both an ac and dccomponent, the Meter displays only the largestcomponent of the measured signal.
acf08f.eps
MegOhmMeterUsing the Meter
19
Checking the BatteryThis function tests the battery under simulated load perEN61557. Disconnect all test leads from any circuit.
acf09f.eps
Full
Empty
Empty100% - Full 0% -
acf13f.eps
Figure 4. Display Example
Fluke 1520Users Manual
20
Maintaining the MeterThis section provides basic maintenance information, including fuse and batteryreplacement instructions.
CautionDo not attempt to repair or service your Meter unless you are qualified to doso and have the relevant calibration, performance test, and serviceinformation.
CleaningClean only with soap and water. Remove any residue afterwards.Periodically wipe the case with a damp cloth and mild detergent.Do not use abrasives or solvents.
MegOhmMeterMaintaining the Meter
21
Replacing and Disposing of the BatteriesW Warning
To avoid electric shock, disconnect the test leads from the inputs beforeopening the Meter for battery replacement.
To avoid false readings, which could lead to possible electric shock orpersonal injury, replace the batteries as soon as the battery empty indicatorG appears.
Note
This Meter contains alkaline batteries. Do not dispose of these batteries with other solidwaste. Used batteries should be disposed of by a qualified recycler or hazardousmaterials handler. Contact your authorized Fluke Service center for recyclinginformation.
The Meter uses four alkaline C cell batteries (supplied). To replace the batteries, do thefollowing (see Figure 5):1. Turn the rotary switch to the OFF position.2. Disconnect test leads from any power source.3. Remove the holster.
Fluke 1520Users Manual
22
4. Place the Meter face down on a nonabrasive surface and loosen the two screws with aflat-blade screwdriver.
5. Lift the battery access lid away from the Meter.6. Replace the C cells as shown in Figure 5. Observe the battery polarity shown in the
battery compartment.7. Secure the battery access lid back in position with the two screws.
MegOhmMeterMaintaining the Meter
23
acf10f.eps
Figure 5. Replacing the Batteries.
Fluke 1520Users Manual
24
Testing and Replacing the FuseW Warning
To avoid electric shock, disconnect the test leads from the inputs beforeopening the Meter for fuse replacement.To prevent personal injury or damage to the Meter, install ONLY Flukespecified fuse identified in the “Replacement Parts and OptionalAccessories” section.
Before replacing the fuse, test it as described in the next section.
Testing the FuseUse the following procedure to test the internal fuse of the Meter.1. Turn the rotary switch to the Lo e Function.2. Connect the test leads to the input terminals and short them
together. Press TEST.3. The display should indicate approximately 0.5 Ω. If the
display reads >40 Ω, replace the fuse as described next andtest again. acf06f.eps
MegOhmMeterMaintaining the Meter
25
Replacing the FuseW Warning
To avoid electric shock, personal injury or damage to the Meter, use ONLYthe specified fuse, and in accordance with the following procedure.
If the previous fuse test indicates that the fuse is defective (resistance > 40 Ω), replace thefuse as follows:
1. Turn the rotary switch to the OFF position.2. Disconnect test leads from any power source.3. Follow Steps 2 - 5 to remove the battery access lid as described under "Replacing and
Disposing of the Batteries".4. Unscrew the bottom cover as shown in Figure 6.5. Remove the fuse as shown in Figure 6.6. Replace with a new fuse.7. Place the bottom cover on and secure the screws.8. Insert the batteries. Observe the battery polarity shown in the battery compartment.9. Secure the battery access lid.10. Test the fuse as described under "Testing the Fuse".
Fluke 1520Users Manual
26
acf11f.eps
Figure 6. Replacing the Fuse
MegOhmMeterReplacement Parts and Optional Accessories
27
Replacement Parts and Optional AccessoriesReplacement Part Part Number
Battery 1.5 V Alkaline Size C 423582Test Lead set 669058Test Probe, 1 kV, lantern tip, red 803459Test Probe, 1 kV, lantern tip, black 803467Alligator Clip, red 803434Alligator Clip, black 803442Carrying Case 603115Holster 670643Hand Strap 669069W Fuse, 6 mm x 32 mm (0.25 x 1.25 inch), 0.5 A, 660 V, Fast Acting,50 A Minimum Interrupt rating
ORW Fuse, 6 mm x 32 mm (0.25 x 1.25 inch), 0.5 A, 750 V, Fast Acting
NA
1556096
Optional Accessories Part Number
Harness 669074Tool Pack 669903W For safety, use exact replacement only.
Fluke 1520Users Manual
28
Principle of Measurement for ResistanceThe Meter measures resistance by inducing a current in the circuit under test, measuringthe resultant potential across the circuit, and then calculating the resistance of the circuit.The resistance is then established by performing the following calculation:
I
VR = (Ohm’s Law)
Service CentersTo locate an authorized service center, visit us on the World Wide Web:
www.fluke.com or call Fluke using any of the phone numbers listed below:
USA: 1-888-99-FLUKE (1-888-993-5853)Canada: 1-800-36-FLUKE (1-800-363-5853)Europe: +31 402-678-200Japan: +81-3-3434-0181Singapore: +65-738-5655Anywhere in the world: +1-425-446-5500
MegOhmMeterSpecifications
29
SpecificationsEnvironmental
Operating Temperature -10 °C to +50 °CStorage Temperature -40 °C to +70 °CTemperature Coefficient 0.10 x (specified accuracy)/ °C
(< 18 °C or > 28 °C)Relative Humidity:Non condensing (< 10 °C)90 % RH (10 °C to 30 °C)75 % RH (30 °C to 40 °C)45 % RH (40 °C to 50 °C)(Without Condensation)Dust/water resistance IP42 per IEC 529Operating Altitude 2000 mStorage Altitude 12000 m
Fluke 1520Users Manual
30
Mechanical SpecificationsSize 23,4 x 10 x 6,4 cm
(9.2 x 3.9 x 2.5 in)Weight 1 kg (2.2 lbs.)Drop requirement Per IEC 1010-1Shock andVibration
Conforms to MIL-PREF-28800F class 3 & 4
Safety SpecificationsElectrical Safety Meets all requirements of EN61010-1, 1995 and EN61557, 1997MaximumOperating Voltage
600 V ac or dc between any terminal and earth ground
Protection Levels CAT ,,,, 600 V, Pollution Degree 2 per EN61010-1Electromagnetic Compatibility (EMC)
Immunity &Emmissions
EN 61326-1(Unspecified for Field strength in excess of 2V/m)
ESD EN61000-4-2 Satisfies Criteria B
MegOhmMeterSpecifications
31
Electrical SpecificationsBattery C Size 1.5 V alkaline, ANSI/NEDA-14A, IEC-
LR14 (4 pieces)Fuse 6 mm x 32 mm (0.25 x 1.25 inch), 0.5 A 660 V,
Fast Acting, 50 A Minimum Interrupt ratingInsulation Resistance
Display Ranges 4.000 MΩ, 40.00 MΩ, 400.0 MΩ, 4000 MΩMeasurement Range (per EN61557-2) 0.250 MΩ to 4000 MΩResolution 0.001 MΩ to 10 MΩ
2 % + 2 counts0.250 MΩ to 100.0 MΩ
Accuracy
10 % + 2 counts + 1 %/1000 MΩ100.0 MΩ to 4000 MΩ
Analog Bar Graph 0 to ∞
Fluke 1520Users Manual
32
Test Voltages 250 V, 500 V, 1000 VAccuracy +20 %, -0 %Nominal Current 1 mANumber of measurements 5,000Circuitry Protection test inhibited if ≥ 30 V ac or dc at inputs
Lo ΩDisplay Range 40.00 ΩMeasurement Range 0.10 Ω to 40.00 ΩAccuracy 2 % + 2 countsResolution 0.01 ΩAnalog Bar Graph 0 to 100 ΩOpen Circuit Voltage 6 V typicalShort Circuit Current 200 mA minimum, (0.2 to 2.0 Ω for >20%
battery capacity remaining).Test Leads Zero 2 ΩNumber of measurements 5,000Circuitry Protection test inhibited if ≥ 30 V ac or dc at inputs
MegOhmMeterSpecifications
33
Voltage
Range 600 V, dc, 50/60 Hz
Resolution 1 V
Accuracy 2 % + 2 counts
Analog Bar Graph 0 to 1000 V
Resistance
Range 4000 ΩAccuracy 2 % + 2 counts
Resolution 1 ΩAnalog Bar Graph 0 to 10 kΩBeeper On at ≈30 Ω or less
Fluke 1520Users Manual
34
Limited Warranty & Limitation of LiabilityEach Fluke product is warranted to be free from defects in material and workmanship under normal useand service. The warranty period is three years and begins on the date of shipment. Parts, productrepairs and services are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries or to anyproduct which, in Fluke’s opinion, has been misused, altered, neglected, contaminated, or damaged byaccident or abnormal conditions of operation or handling. Fluke warrants that software will operatesubstantially in accordance with its functional specifications for 90 days and that it has been properlyrecorded on non-defective media.Fluke authorized resellers shall extend this warranty on new and unused products to end-usercustomers only but have no authority to extend a greater or different warranty on behalf of Fluke.Warranty support is available if product is purchased through a Fluke authorized sales outlet or Buyerhas paid the applicable international price. Fluke reserves the right to invoice Buyer for importation costsof repair/replacement parts when product purchased in one country is submitted for repair in anothercountry.Fluke’s warranty obligation is limited, at Fluke’s option, to refund of the purchase price, free of chargerepair, or replacement of a defective product which is returned to a Fluke authorized service centerwithin the warranty period.To obtain warranty service, contact your nearest Fluke authorized servicecenter to obtain return authorization information, then send the product to that service center, with adescription of the difficulty, postage and insurance prepaid (FOB Destination), to the nearest Flukeauthorized service center. Fluke assumes no risk for damage in transit. Following warranty repair, theproduct will be returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines that thefailure was caused by neglect, misuse, contamination, alteration, accident or abnormal condition ofoperation or handling, including overvoltage failures caused by use outside of the product’s specifiedrating, or normal wear and tear of mechanical components, Fluke will provide an estimate of repair costsand obtain authorization before commencing the work. Following repair, the product will be returned to
MegOhmMeterSpecifications
35
the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportationcharges (FOB Shipping Point).THIS WARRANTY IS BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHERWARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OFMERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FORANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSSOF DATA, ARISING FROM ANY CAUSE OR THEORY.Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion orlimitation of incidental or consequential damages, the limitations and exclusions of this warranty may notapply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court or otherdecision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of anyother provision.
Fluke CorporationP.O. Box 9090Everett, WA 98206-9090U.S.A.
Fluke Europe B.V.P.O. Box 11865602 BD EindhovenThe Netherlands
11/99
Fluke 1520Users Manual
36