TR 73U Datasheet

Embed Size (px)

Citation preview

  • 7/22/2019 TR 73U Datasheet

    1/2

    T&D Corporation

    The TR-73U has all the great features of TandDs TR-72U temperatureand humidity data logger plus a third channel for monitoring and recordingambient pressure. This compact unit has a large 4 digit LCD display forreading current values that can be set to show any single channel, or allthree on a rotating basis. The logger can store up to 8,000 sample triplets,operates on a single AA battery and incorporates a USB interface for rapid

    data downloading. As with all of TandDs data loggers, the TR-73U shipscomplete with software, sensor and cable included.

    Temperature, Humidity and Ambient Pressure

    Three Channel Data Logger

    TR-73U

  • 7/22/2019 TR 73U Datasheet

    2/2

    Software SpecificationsT&D Recorder for Windows

    TR-71U TR-72U TR-73U

    Recording Start (Programmed Start / Immediate Start) Recording Stop

    Main Unit Settings (Recording Interval Recording Method Upper and Lower Limits)

    Downloading of Recorded Data

    Temperature and Humidity Graphs for each Channel

    Zoom in, out and scroll with mouse or keyboard

    Change Channel Colors Turn ON and OFF Channel Display

    Channel Name Recording Interval Number of Readings

    Highest, Lowest and Average Readings Unit of Measurement.

    AB Cursor Dates, Times and Data Readings

    Calculated Difference between Cursor A and B

    8 Channel Simultaneous Display and Processing

    Possible to process mixed data from TR-71U and TR-72U units. (up to 4units.)

    Data List Display Calculation Range Settings Data Maintenance

    Edit Recording Conditions Delete Data by Channel

    Re-order Data by Channel

    Serial Port Auto Find Function

    T&D Common Data File Text File (CSV, etc)

    (Selected Range or File for Entire Period)

    Graphs / Tables

    Microsoft Windows 2000 / XP English

    Microsoft Windows 98 / Me E nglish

    Microsoft WindowsNT 4.0 English **1

    IBM Compatible with higher than Pentium 90MHz

    USB Port (1.0 / 1.1)

    More than 16MB

    More than 4 MB of free space (Data will need more space)

    VGA (640 480) more than 256 colors

    Software

    Compatible Devices

    Communication

    Functions

    Temp/Humidity Graph

    Graph Display

    Data Display

    Number of Channels

    Others

    Other

    File Output

    Printing

    Compatible OS

    PC/CPU

    Memory

    Hard Disk

    Monitor

    **1: USB communication with TR-71U / 72U data loggers cannot be carried out using WindowsNT4.0.

    1, 2, 5, 10, 15, 20, 30 Seconds / 1, 2, 5, 10, 15, 20, 30, 60 Minutes- Total of 15 choices

    8,000 Readings 2 or 3 Channels

    Endless Loop Method (Overwrite from the oldest data when recording capacity is full)

    One Time Method (Stop recording when recording capacity is full)

    Measurements, (Any Single Channel or alternating display of all Channels),

    Recording Status, Battery Life Warning, Amount of Recorded Data, Unit of Measurement

    AA alkaline battery (LR6)

    Approx. 1 year**2

    Activated when battery power is low or when switch is off (Approx. 1 year)

    USB Communication Cable (included)

    Approximately 8 seconds (For full memory)

    H55mm W78mm D18mm

    About 62g (Including one AA battery)

    -10 to 60C, 90% RH (non-condensing)

    TR-3100 External Temperature & Humidity Sensor

    AA alkaline Battery (LR6), User's Manual,

    USB Cable (US-15C ; length 1.5m), Software CD Set

    Pressure

    1 Channel

    750 to 1100 hPa

    ---

    ---

    1.5% hPa

    0.1 hPa

    Atmospheric

    Pressure Sensor

    4 Sec. **1

    Humidity

    1 Channel

    ---

    10 to 95% RH

    ---

    5% RH

    (25C, 50% RH)

    1% RH

    Macromolecular

    ---

    Temperature

    1 Channel

    -10 to 60C

    0 to 50C

    -40 to 110C

    0.3% (0 to 50C)

    0.5% (-40 to -20 / 80 to 110C)

    0.1C

    Thermistor

    12 Sec.

    TR-73U

    Measurement Parameter

    Measurement Channel

    Internal Sensor

    Standard External Sensor

    Optional External Sensor

    Measurement Accuracy

    Display Resolution

    Sensor Type

    Internal Sensor Time Constant

    Recording Interval

    Recording Capacity

    Recording Mode

    LCD Display

    Power

    Battery Life

    Data Back-up

    Communication Interface

    Data Download Time

    Dimensions

    Weight

    Operating Environment

    Standard Sensor (Included)

    Accessories

    included in package

    **1 Pressure time constant is 40 seconds for sample rates of one minute or longer.**2 Battery life of TR-73U is approximately 10 months.

    Product Specifications

    Web Site T&D OnlineProduct information, FAQ and software update downloads.

    http://www.tandd.com/

    TandD US, LLC.Phone: (518) 669-9227

    Fax: (413) 639-9227E-mail: [email protected]

    5652-169 Sasaga Matsumoto City.

    Nagano 399-0033 Japan

    Facsimile(+81)263-26-4281

    E-mail: [email protected]

    T&D Corporation

    System Setup

    Output

    Standard Humidity

    and Temperature Sensor

    (TR-3100)

    Software

    (T&D Recorder for Windows)

    Printout

    Install

    ConnectConnect

    Insert

    AA Alkaline battery

    LR6 x 1

    Text FileUSB Communication Cable

    US-15C

    TR-73U

    Easy to Read Multi-Functional DisplayThe easy to read LCD displays current readings , recording condit ion,

    bat te ry li fe wa rn in g, an d un it o f me as ure me nt .

    (1) Re cording condition:

    ON:Recording in progress. BLINKING:Waiting for programmed start.

    (2) Memory status:

    After every 2000 readings the scale will increment from left to right.

    (3) Data communication.

    ON: USB cable is connected. BLINKING: Data is being transferred.

    (4) ONTIME:When the number of recorded readings reaches 8000, "FULL"will appear in the

    unit's LCD display and recording will automatically stop.

    ENDLESS:When the number of recorded readings reaches 8000, the oldest data reading will

    be ove rwr it ten and rec ord ing wil l c ont inu e.

    (5) When the battery power becomes low, this will appear in the LCD display. If the batterypow er bec ome s e ven low er, SLP wil l a ppe ar and nor mal ope rat ion s w ill sto p. If the

    signal appears, please change the battery as soon as possible.

    (6) The unit of measurement (C, F, %, hPa) for the display will appear.

    (7) Current measurements or operationa l messages such as FULL or SLP will appear.

    (1) (2) (3) (4) (5)

    (6)

    (7)