The Development of The Development of Laue Monochromator Laue Monochromator at X17B3 National Synchrotron Light at X17B3 National Synchrotron Light Source Source in 2006 in 2006 3. (111) diffraction intensities by severer of non- bending Si crystals, with only having Si-Laue-1 and the photo diode at position [2], as shown as at Fig. 1.(0.6mm thickness and at 30KeV energy): ========================================================================== The number of Si Total reading of Crystals in the stack at the photo diode-2 back ground Net intensity --------------------------------------------------------------------------------- ------------------------------------------------ 1 21.97K 13.5K 8.5K --------------------------------------------------------------------------------- ------------------------------------------------ 2 26.7K 14.6K 12.0K --------------------------------------------------------------------------------- ------------------------------------------------ 3 29.5K 9.6K 19.9K --------------------------------------------------------------------------------- ------------------------------------------------ 4 18.3K 9.6K 8.7K --------------------------------------------------------------------------------- ------------------------------------------------ National Synchrotron Light Source Brookhaven National Laboratory Upton, NY 11973 COnsortium for Materials Properties Research in Earth Sciences Stony Brook University Stony Brook, NY. 11794-2100 Tel: (631) 632-8213 Fax: (631) 632-8140 Q. Guo , J. Hu Z. Zhong Abstract A prototype 4-crystal sagittal-Laue monochromator was designed, assembled, and tested at X17B3 beam line successfully in 2005. The x- ray flux provided by such four-crystal monochromator system is one order of magnitude less than that provided by two-crystal sagittal- focusing Laue monochromators, currently in use at X17C and X17B3 beamlines. The thickness of the Si crystals was selected to be 0.6 mm to provide design energy 30keV. In order to increase the x-ray flux, we optimized the thickness of Si Laue crystals in 2006. The experimental results show that the three pieces of 0.6 mm crystals provided the strongest diffraction intensity, compared with 1, 2 and 4 pieces of 0.6mm Si crystals. Moreover, the results also show that it is possible to align multiple crystals, cut from the same wafer, to provide monochromatic energy from the 111 diffraction. Such arrangement is also compatible with two-crystal or 4-crystal Laue monochromators. *Introduction of the status of monochromators at X17C and X17B3 in 2005 : 1. 2- Si crystal Laue monochromators have been used at X17C and X17B3 successfully by users. 2. The prototype 4-Si crystals Laue monochromator, which was designed and assembler by Z. Zhong, has tested at X17B3 beam line successfully. The thickness of Si crystal was selected as 0.6mm as optimized thick at 30keV x-ray energy. 3. The weakness of flux for 4-Si Laue mono system is the key problem in comparing with 2-Si Laue monochromators, which are used at X17C and X17B3. (See “the report of 4-Laure monochromator testing in 2005” for the details) *The problems of 2-Laue monochromator at X17B3 beamline : 1. The offset of mono beam from white beam at X17C is about 20mm. Therefore it will take one or two days to change it from white beam to mono beam for the diffraction system realignments at X17C hutch. However, it is even not practicable at X17B3 beamline since the laser heating system. It is not easy to realign the laser system in-situ with x-ray. It might be taking another week to do the switches between white and mono x-ray beams as laser heating system in. 2. To lift the whole diffraction optical table at X17B3 hutch 20mm up would fix the mono beam 20mm offset problem. However, there still are some arguments for this: how is about the lifting accurate and how is about the even level at the new position of the optical table? 3. How about the Si crystal optimized thickness is? We should optimize the total thickness of 4 Si crystals as one piece, or should optimize each of the 4 crystals individually? X Y (011) (011) Si-Laue-1 Si-Laue-2 White X-ray Mono beam (111) diffraction Imaging Plate Fluorescent Plate Photo Diode White beam stop [1] [2] FIG. 1. Laue Monochromator testing setup at X17B3 beam line in 2006 ~2000mm Slits Slits The 2-Laue monochromator is using at X17C from 2003 *The 2-Laue monochromator is using at X17B3 with crystal holder modified for mono beam offset shorter The prototype 4-Laue monochromator was used at X17B3 in 2005 The 2-Laue monochromator was tested in X17B3 hutch with the shortest mono beam offset, which results are reported in this poster. 4. The mono beam intensities, measured by photodiode, with numerous non-bending crystals in Si-Laue-1 and only one Si crystal in Si-Laue-2, as shown as at Fig.1. (0.6mm thick and at 30KeV energy) ========================================================================= The number of Si Total reading of Crystals in the stack at the photo diode-(1) back ground Net intensity --------------------------------------------------------------------------------- ----------------------------------------------- 1 15.6K 7.5K 8.1K --------------------------------------------------------------------------------- ---------------------------------------------- 2 19.2K 7.6K 11.6K --------------------------------------------------------------------------------- ---------------------------------------------- 3 21.8K 7.4K 14.4K ========================================================================= * Discussion and Conclusions: 1. The thickness should optimize individually for the Si crystals in the Laue Monochromator system to maximum output flux at different X-ray energies. 2. The results shown that it is possible with multi-pieces Si crystals to have unique energy on the [111] diffraction in the Laue monochromator. 3. Obviously, the uniform of Si wafer thickness is very important for the mono beam produced by multi-pieces crystals. In additional, the crystal cutting is another key skill to have a unique energy too. 4. The dE/E will be affected by the Si crystal thickness. It should be determined by the rocking curves experimentally in next testing experiments. 5. The Laue diffraction flux could be doubled if using the optimized crystal thickness, and consequently, the explore time could reduced to several minutes from couple of ten minutes; and with on-line reader IP will improve the x-ray beam time efficiency a lot. And it also means that 4-Laue Monochromator should be useable routinely for general users. 6. The NSLS original designs of X17 x-ray beam size are limited about 3mm height at vertical direction. If the offset of Laue monochromator can be reduced to smaller than the limiting of 3mm, we would have a chance to solve the mono beam offset problems, and can keep the white beam and mono beam at the same level. 7. The crystal bending in the asymmetric Laue monochromators has two advantages: increasing brightness by the focusing effects and increasing the mono beam diffraction efficiency by the non-perfect crystallization. 8. Moreover, the crystals bending at the multi-pieces crystal will be another challenge at the Laue monochromators. We should test it in next step. References: [1] Z. Zhong, et. al., Acta. Cryst. A 59 (2003) 1-6 [2] J. Hu, et. al., Eos. Trans. AGU, 84(46), Fall Meet. Abstract V42A-0334, 2003. [3] Q. Guo, et. al., COMPRES Annual Meeting, 2004. [4] Q. Guo, J. Hu. “The report of X17B3 4-laue monochromator, 2005”. Unpublished letter. 5. The mono beams testing data on the Image Plate: Fig. 2-1: IP image with Si-Laue-1 has only one crystal Fig. 2-2: IP image with Si-Laue-1 has only two crystals Fig. 2-3: IP image with Si-Laue-1 has only three crystals Fig. 2-4: IP image with Si-Laue-1 has four crystals Fig. 2-5: IP image with Si-Laue-1 has four crystals and Si-Laue-2 has only one crystal at left; while IP image with Si-Laue-1 has four crystals and no Si-Laue-2 at the right. Fig. 2-1 0.6mm Si 1 (Si-Laue-1) + 0 (Si- Laue-2) Fig. 2-2 0.6mm Si 2 (Si-Laue-1) + 0 (Si- Laue-2) Fig. 2-3 0.6mm Si 3 (Si-Laue-1) + 0 (Si- Laue-2) Fig. 2-4 0.6mm Si 4 (Si-Laue-1) + 0 (Si-Laue-2) Fig. 2-5 ---- 0.6mm Si ---- 4(Si-Laue- 1)+1(Si-Laue-2) ; 4(Si-Laue-1)+0(Si- Laue-2) Local contact person: Quanzhong Guo. (631) 344-5343, [email protected] Welcome comments

The Development of Laue Monochromator at X17B3 National Synchrotron Light Source in 2006

Embed Size (px)

DESCRIPTION

The Development of Laue Monochromator at X17B3 National Synchrotron Light Source in 2006. References:. [1] Z. Zhong, et. al., Acta. Cryst . A 59 (2003) 1-6 [2] J. Hu, et. al. , Eos. Trans. AGU, 84 (46), Fall Meet. Abstract V42A-0334, 2003. - PowerPoint PPT Presentation

Citation preview

Page 1: The Development of  Laue Monochromator  at X17B3 National Synchrotron Light Source  in 2006

The Development of The Development of Laue Monochromator Laue Monochromator at X17B3 at X17B3 National Synchrotron Light Source National Synchrotron Light Source in 2006in 2006

3. (111) diffraction intensities by severer of non-bending Si crystals, with only having Si-Laue-1 and the photo diode at position [2], as shown as at Fig. 1.(0.6mm thickness and at 30KeV energy):

==========================================================================The number of Si Total reading ofCrystals in the stack at the photo diode-2 back ground Net intensity---------------------------------------------------------------------------------------------------------------------------------

1 21.97K 13.5K 8.5K--------------------------------------------------------------------------------------------------------------------------------- 2 26.7K 14.6K 12.0K---------------------------------------------------------------------------------------------------------------------------------

3 29.5K 9.6K 19.9K--------------------------------------------------------------------------------------------------------------------------------- 4 18.3K 9.6K 8.7K---------------------------------------------------------------------------------------------------------------------------------

National Synchrotron Light Source Brookhaven National Laboratory Upton, NY 11973

COnsortium for Materials

Properties Research in Earth SciencesStony Brook University

Stony Brook, NY. 11794-2100Tel:  (631) 632-8213Fax: (631) 632-8140

Q. Guo , J. Hu Z. Zhong

AbstractA prototype 4-crystal sagittal-Laue monochromator was designed, assembled, and tested at X17B3 beam line successfully in 2005. The x-ray flux provided by such four-crystal monochromator system is one order of magnitude less than that provided by two-crystal sagittal-focusing Laue monochromators, currently in use at X17C and X17B3 beamlines. The thickness of the Si crystals was selected to be 0.6 mm to provide design energy 30keV. In order to increase the x-ray flux, we optimized the thickness of Si Laue crystals in 2006. The experimental results show that the three pieces of 0.6 mm crystals provided the strongest diffraction intensity, compared with 1, 2 and 4 pieces of 0.6mm Si crystals. Moreover, the results also show that it is possible to align multiple crystals, cut from the same wafer, to provide monochromatic energy from the 111 diffraction. Such arrangement is also compatible with two-crystal or 4-crystal Laue monochromators.

*Introduction of the status of monochromators at X17C and X17B3 in 2005:

1. 2- Si crystal Laue monochromators have been used at X17C and X17B3 successfully by users.2. The prototype 4-Si crystals Laue monochromator, which was designed and assembler by Z. Zhong, has tested at X17B3 beam line successfully. The thickness of Si crystal was selected as 0.6mm as optimized thick at 30keV x-ray energy.3. The weakness of flux for 4-Si Laue mono system is the key problem in comparing with 2-Si Laue monochromators, which are used at X17C and X17B3. (See “the report of 4-Laure monochromator testing in 2005” for the details)

*The problems of 2-Laue monochromator at X17B3 beamline:

1. The offset of mono beam from white beam at X17C is about 20mm. Therefore it will take one or two days to change it from white beam to mono beam for the diffraction system realignments at X17C hutch. However, it is even not practicable at X17B3 beamline since the laser heating system. It is not easy to realign the laser system in-situ with x-ray. It might be taking another week to do the switches between white and mono x-ray beams as laser heating system in. 2. To lift the whole diffraction optical table at X17B3 hutch 20mm up would fix the mono beam 20mm offset problem. However, there still are some arguments for this: how is about the lifting accurate and how is about the even level at the new position of the optical table? 3. How about the Si crystal optimized thickness is? We should optimize the total thickness of 4 Si crystals as one piece, or should optimize each of the 4 crystals individually?

X

Y

(011)

(011)

Si-Laue-1

Si-Laue-2

White X-ray

Mono beam

(111) diffraction

Imaging Plate

Fluorescent Plate

Photo DiodeWhite beam stop

[1]

[2]

FIG. 1. Laue Monochromator testing setup at X17B3 beam line in 2006

~2000mm

Slits

Slits

•The 2-Laue monochromator is using at X17C from 2003

*The 2-Laue monochromator is using at X17B3 with crystal holder modified for mono beam offset shorter

•The prototype 4-Laue monochromator was used at X17B3 in 2005

•The 2-Laue monochromator was tested in X17B3 hutch with theshortest mono beam offset, which results are reported in this poster.

4. The mono beam intensities, measured by photodiode, with numerous non-bending crystals in Si-Laue-1 and only one Si crystal in Si-Laue-2, as shown as at Fig.1. (0.6mm thick and at 30KeV energy)

=========================================================================The number of Si Total reading ofCrystals in the stack at the photo diode-(1) back ground Net intensity-------------------------------------------------------------------------------------------------------------------------------- 1 15.6K 7.5K 8.1K------------------------------------------------------------------------------------------------------------------------------- 2 19.2K 7.6K 11.6K------------------------------------------------------------------------------------------------------------------------------- 3 21.8K 7.4K 14.4K=========================================================================

* Discussion and Conclusions:

1. The thickness should optimize individually for the Si crystals in the Laue Monochromator system to maximum output flux at different X-ray energies. 2. The results shown that it is possible with multi-pieces Si crystals to have unique energy on the [111] diffraction in the Laue monochromator. 3. Obviously, the uniform of Si wafer thickness is very important for the mono beam produced by multi-pieces crystals. In additional, the crystal cutting is another key skill to have a unique energy too. 4. The dE/E will be affected by the Si crystal thickness. It should be determined by the rocking curves experimentally in next testing experiments. 5. The Laue diffraction flux could be doubled if using the optimized crystal thickness, and consequently, the explore time could reduced to several minutes from couple of ten minutes; and with on-line reader IP will improve the x-ray beam time efficiency a lot. And it also means that 4-Laue Monochromator should be useable routinely for general users.6. The NSLS original designs of X17 x-ray beam size are limited about 3mm height at vertical direction. If the offset of Laue monochromator can be reduced to smaller than the limiting of 3mm, we would have a chance to solve the mono beam offset problems, and can keep the white beam and mono beam at the same level.7. The crystal bending in the asymmetric Laue monochromators has two advantages: increasing brightness by the focusing effects and increasing the mono beam diffraction efficiency by the non-perfect crystallization. 8. Moreover, the crystals bending at the multi-pieces crystal will be another challenge at the Laue monochromators. We should test it in next step.

References:

[1] Z. Zhong, et. al., Acta. Cryst. A 59 (2003) 1-6

[2] J. Hu, et. al., Eos. Trans. AGU, 84(46), Fall Meet. Abstract V42A-0334, 2003.

[3] Q. Guo, et. al., COMPRES Annual Meeting, 2004.

[4] Q. Guo, J. Hu. “The report of X17B3 4-laue monochromator, 2005”. Unpublished letter.

5. The mono beams testing data on the Image Plate:

Fig. 2-1: IP image with Si-Laue-1 has only one crystalFig. 2-2: IP image with Si-Laue-1 has only two crystalsFig. 2-3: IP image with Si-Laue-1 has only three crystalsFig. 2-4: IP image with Si-Laue-1 has four crystalsFig. 2-5: IP image with Si-Laue-1 has four crystals and Si-Laue-2 has only one crystal at left; while IP image with Si-Laue-1 has four crystals and no Si-Laue-2 at the right.

Fig. 2-1 0.6mm Si1 (Si-Laue-1) + 0 (Si-Laue-2)

Fig. 2-2 0.6mm Si 2(Si-Laue-1) + 0(Si-Laue-2)

Fig. 2-3 0.6mm Si 3(Si-Laue-1) + 0(Si-Laue-2)

Fig. 2-4 0.6mm Si4 (Si-Laue-1) + 0 (Si-Laue-2)

Fig. 2-5 ---- 0.6mm Si ---- 4(Si-Laue-1)+1(Si-Laue-2) ; 4(Si-Laue-1)+0(Si-Laue-2)

Local contact person: Quanzhong Guo.

(631) 344-5343, [email protected] Welcome comments