Upload
lucy-price
View
25
Download
2
Embed Size (px)
DESCRIPTION
The geometrical tolerances of the SSS At the cold mass construction ( ACCEL) and after cold test of the cryo magnets (CERN WP18) M. Bajko AT MAS 28th of March. 2006. Summary: ECR on the length Angular defect of V line flanges. Tolerance table. - PowerPoint PPT Presentation
Citation preview
28th of March 2006marta bajko, at mas for WGA 77
SSS Tolerances WGA 77
Summary: ECR on the length
Angular defect of V line flanges
The geometrical tolerances of the SSS At the cold mass construction ( ACCEL)
and after cold test of the cryo magnets (CERN WP18)
M. Bajko AT MAS 28th of March. 2006
28th of March 2006marta bajko, at mas for WGA 77
SSS Tolerances WGA 77
1. The Radial tolerance of V1&V2 on the Connection at WP18 are relaxed values as offered by VAC, thank
to experiment gained with BPM/beam-screen/drift tube installation. A ‘grey area’ (up to 0.8mm) can be
accepted with some caution, see minutes WGA72.2. If the tolerance is exceeded, the angle w.r.t. the local
axes of the cold bore tube shall be measured and checked with the same tolerance.
All tolerances at ACCEL are given in the 12 plug co-ordinate system, while at WP18 at CERN is given in
the GA co-ordinate system
r
g Aperture tolerances at WP18 : race-track area, see figure below
Connection side :g = 0.5 mm, r = 0.6 mm
Lyra side : g = 0.5 mm, r = 0.65 mm
A) In case of non-conformity for aperture, a shift or rotation is applied to the whole assembly at installation in order to close it.
Tolerance table
Nominal length to be changed with +2 mm. ECR by MM8 8 1.5
1461.25
4124.25 5585.5
28th of March 2006marta bajko, at mas for WGA 77
SSS Tolerances WGA 77
ECR on length
28th of March 2006marta bajko, at mas for WGA 77
SSS Tolerances WGA 77
Angular defect of the V line flanges on downstream side
28th of March 2006marta bajko, at mas for WGA 77
SSS Tolerances WGA 77
V lines Downstream side
V1 & V2 Lyra side Radial position of center
Ref is CM-B-plan axis for ACCEL / GA at WP18
0.6 0.87 mm
Angle to end cover plane
Ref is B-plan at ACCEL / GA at WP18
± 4 ± 4.51 mrad
Distance from end-cover
On nominal of 64 mm ± 0.5 ± 0.5 mm
Go-nogo gauge confirmed by P. L. VAC 73.45 73.499 mm Min diameter (on drawing sat ACCEL
73.44,73.75) 73.25 73.25 mm
1 If the tolerance is exceeded, the angle w.r.t. the local axes of the cold bore tube shall be measured and checked with the same tolerance.
(see overleaf)
V1 & V2 Lyra side Radial position of center
Ref is CM-B-plan axis for ACCEL / GA at WP18
0.6 0.87 mm
Angle to end cover plane
Ref is B-plan at ACCEL / GA at WP18
± 4 ± 4.51 mrad
Distance from end-cover
On nominal of 64 mm ± 0.5 ± 0.5 mm
Go-nogo gauge confirmed by P. L. VAC 73.45 73.499 mm Min diameter (on drawing sat ACCEL
73.44,73.75) 73.25 73.25 mm
1 If the tolerance is exceeded, the angle w.r.t. the local axes of the cold bore tube shall be measured and checked with the same tolerance.
(see overleaf)
RACCEL= 0.6 mm
RCERN= 0.87 mm
ACCEL= 0.6 mm ( 8 mrad)
CERN= 0.6 mm ( 8 mrad)
LACCEL= ±0.5 mm
LCERN= ±0.5 mm
ACCELmax=CERNmax = 73.25 mm
ACCELmin=73.45 mm
CERNmin = 73.4999 mm
28th of March 2006marta bajko, at mas for WGA 77
SSS Tolerances WGA 77
• There is one repair finished on SSS133: procedure as in dipoles: see doc. Edms 677407
• Good co-orrelation is found between ACCEL and CERN . On 154 cold masses: correlation co-efficent is 0.93. therefore the defect can be identified before the cold mass is cryostated and plane the necessary repairs. The list of cold masses to be repaired is done and given to SSS co-ordination.
• After OCT. 2005 no Nc were detected on magnet coming from from ACCEL . It was requested to do, if necessary ,repairs in Accel (1 case was reported of such repair by. K. Schirm).
ACTIONS IN AT-MAS and ACCEL