Scanning Electrochemical Microscopy (SECM) Informational Meeting
June 6th 2018 SEAS Research Equipment Assistance Program (REAP)
“Host” PI’s: D. Esposito and Y. Yang
Superuser: Anna Dorfi
Meeting Outline
2
I. What is SECM, and what can it be used for?
II. Imaging Modes with 3 Examples
III. CHI 920D – Capabilities and limitations
IV. Training, Fees, & other logistics
What is SECM?
3
SECM: a scanning probe microscopy (SPM) technique that utilizes an electrochemical probe (“ultramicroelectrode” (UME) or tip ) to measure the physical and/or chemical properties of an interface.
• Largely credited to Bard and coworkers in 1989.
• Can generate images with quantitative information about spatial variation in:
-chemical reaction rates -topology -conductivity -surface coverages in addition to being used for.. -surface patterning -measuring rxn rate constants -High-throughput screening -much more….. Solid sample
or substrate
d= tip/substrate separation distance
a= tip radius
2a
O R
e-
e- active metal
insulating sheath
Common SECM Imaging Mode: “Feedback Mode”
4
Psuedo steady state diffusion limiting current:
“negative feedback” “positive feedback” Tip current at infinite distance from sample
d d
d= Tip/substrate separation distance
a= probe radius D=diffusion coef. of O c=concentration of O
(-) (-)
(+)
Example #1: Imaging a catalytic disc electrode
5
Optical Image of 150 μm Pt disc.
150 μm
Imaging Conditions: Electrolyte: 1 mM H2SO4 in 0.1 M Na2SO4 , UME Radius: ≈1.5 µm, Scan Height = ≈ 3 µm. Tip potential: 0.6 V vs Ag|AgCl, Substrate Potential: -0.7 V vs Ag|AgCl.
5
Inert Si++ substrate
1 mM H2SO4 electrolyte 2H+ H2
e-
e-
SECM image of Pt disc electrode at right.
inert substrate
catalytic Pt disc
Side-view of constant-height SECM imaging in feedback mode using the H+/H2 redox couple.
catalytic disc
probe scan path
EUME=+0.6 V Ag|AgCl
ESubstrate=-0.7 V Ag|AgCl (-)
(+)
SECM Approach Curve
6
SECM Approach curves can be used for quantitative determination of local reaction kinetics.
Normalized rate constant
k=heterogeneous electron transfer rate constant.
Approach curves for different normalized reaction rate constants.[1]
Ref.: N. Ritzert, H. Abruna, Langmuir, 29 (2013), 1683-1693.
7 [1] Polcari et. al. Chem. Rev. 2016, 116, 13234−13278
What is SECM used for?
List of fields and applications of SECM.[1]
Key Advantages • High resolution imaging
technique. • Quantitative information
about local reaction kinetics.
• Unique chemical imaging modalities.
• Capable of In situ imaging in the liquid environment.
Generic SECM Instrument
8
Schematic of an SECM setup
Polcari et. al. Chem. Rev. 2016, 116, 13234−13278
Cell
SECM: Probes
9
• SECM spatial resolution is strongly dependent on probe dimensions.
• Many studies use “ultra-microelectrodes” (UMEs) with probe diameter <25 μm
• Nanoelectrodes are also possible, with nanoprobes down to ~ 3 nm used for high resolution SECM imaging.
• Materials for electroactive metal:
• Platinum (74%), carbon (10.8%), gold (9.5%), mercury (2.1%), and silver (1.7%)
Polcari et. al. Chem. Rev. 2016, 116, 13234−13278
~ 4 cm
CH Instruments 10 um diam. Pt UME for SECM imaging.
Two Common Amperometric Imaging Modes
10
I. Tip Generation, Substrate Collection
(TG/SC mode)
II. Substrate Generation, Tip Collection
(SG/TC mode)
Ref.: P. Sun, M. Mirkin, “SECM in the 21st Century”, Phys. Chem. Chem. Phys., 2007, 9, 802–823.
CO,b CR,b
11 http://www.chinstruments.com/chi900.shtml
Lots of possibilities!!
Example #2: Monitoring Biochemical Processes
12 Ref.: Takahashi, Y.; Miyamoto, T.; Shiku, H.; Asano, R.; Yasukawa, T.; Kumagai, I.; Matsue, T. Electrochemical Detection of Epidermal Growth Factor Receptors on a Single Living Cell Surface by Scanning Electrochemical Microscopy. Anal. Chem. 2009, 81, 2785−2790
SECM is also used in imaging and studying the uptake or release of chemical species from a surface, including processes in biological cells.
epidermal growth factor receptor
Microelectrode
SGTC mode
SECM imaging of EGF-triggered endocytosis, in which EGF induces entrapment of EGFR in the cell membrane.
13
Example #3: Measuring local Ion Transport through membranes
Ref: M. Shen, S. Amemiya, et al., “Quantitative Imaging of Ion Transport through Single Nanopores by High Resolution SECM”, JACS, 134 (2012), 9856-9859.
TEM image of porous nanocrystalline (pnc) silicon membrane.
SECM image of ion flux through pnc Si membrane. The electroactive species was tetrabutyl-ammonium (TBA), which diffused through the nanopores and was detected at a Pt electrode within a nanopipette probe.
14
Example #4: Charge transfer kinetics at vdW materials
[1] A. Guell, P. Unwin, et al. ACS Nano 2015, 9, 3558-3571
AFM image of exfoliated graphene on Si /SiO2 substrate[1]
SECCM image measured with 5 mM Ru(NH3)6
3+ at E=-0.46 V Ag|AgCl[1]
SECM Resources
15
1. Good introduction: Bard & Faulkner. “Scanning Probe Techniques”, in Electrochemical Methods. 2nd Ed., Section 16.4, pp. 669-677.
2. More detailed overview: F. Fan, J. Mauzeroll, et al. “Chapter 12: Scanning Electrochemical Microscopy”, in Handbook of Electrochemistry, Elsevier, 2007, 471-540, XII-XIII.
3. SECM Review Articles: • P. Sun, M. Mirkin, “SECM in the 21st Century”, Phys. Chem. Chem. Phys., 2007, 9, 802–823.
• C. Zoski, “Advances in SECM”, J. Electrochem. Soc., 163 (4) H3088-H3100 (2016)
• SECM for energy applications: P. Bertoncello, Energy & Environ. Sci., 3 (2010), 1620-1633.
• SECM for Biological applications: S. Amemiya, et al., Analytical and Bioanalytical Chemistry 386 (2006), 458–471.
16
Reference Electrode
Probe (tip)
Counter Electrode
Sample connection
[1] http://www.chinstruments.com/chi900.shtml
CHI 9200 SECM
Cell
Z-positioner
Optical (air) table
SECM sample cells
CHI 9200 SECM- Specifications & Limitations
17
• Scanning capabilities w/ closed loop positioners:
• Minimum X,Y,Z step size: 1.5 nm
• Maximum X,Y,Z scan lengths: 50 mm
• CHI bipotentiostat (2 working electrodes)
• Measured current accuracy down to 0.1 pA (min. res. 0.3 fA)
• Impedance analyzer for EIS and impedance mapping.
• Compliance voltage: ±13 V
• Sample, electrolyte, & scanning restrictions:
• Roughness >> desired resolution is very challenging
• Samples must be stable in electrolyte (unless studying corrosion!)
• Electrolytes must be air-stable & safe for use in open lab
• Max scan speed is typically a couple probe radii per second
Logistics Training
• Users: Ph.D. and postdocs. Exceptional UG or MS may be considered. • Attend general demo / overview of instrument (email
[email protected] ) • Read through SOP/manual and take short quiz. • 1-on-1 training with certified user. • Final certification by super user (use instrument to generate an image), who
will email Ariel Sanchez ([email protected] ) to grant lab access. • Submit SECM new user form to [email protected] before independent
use.
Usage / Reservation • Instrument must be reserved through a google calendar. • Manual sign-in on notebook next to instrument. • SOPs, forms, analysis files, etc. available on shared google drive folder.
User fees • Initial user fee rate of 30 $/hr. (Based on recovering SECM cost over 10 yrs) • Pi’s will be emailed by ChemE admin. staff once every 3 mo.’s to ask for
approval of a list of charges incurred by group members.
Logistics : Supplies
• Probes / electrodes • Miniature Ag|AgCl reference electrodes & Pt counterelectrodes provided. • 10 um diam. Pt UMEs: supplied for training & general use. (~$120 /ea.) • 1 um or 5 um diam. Pt UMEs: Available through Sensolytics. (~ $215/ea.) • 12 um, 25 um diam. Au UMEs: available through CH Instruments. • Nanoscale probes: You can use our pipette puller (no fee).
• Chemicals • Two standard redox couples will be kept in stock (FFC, Ferrocenemethanol),
with 18 MΩ DI water available in the lab. • Any other chemicals must be approved by the super user; no chemical
storage is available within the lab (take-in / take-out policy). • Access to fume hood will be available.
• Test cells • Standard teflon test cell is freely available • Custom cell may be necessary (different sample size, etc.)
[1] https://www.sensolytics.de/en/products/microelectrodes/58-ultramicroelectrodes-platinum.html [2] http://www.chinstruments.com/accessories.shtml
Probes for SECM
20
• Fabrication of UMEs or nanoelectrodes.
Laser heated pipette during UME fabrication.
Source: Sutter Instruments
Laser pipette puller (Sutter P-2000) located in the Esposito Lab.
• Nanoelectrodes made with a laser pipette puller have been demonstrated with probe diameter down to ≈6 nm.[1] (not easy!)
• See review article on nanoelectrodes.[2]
[1] Sun, Mirkin, et al., Angew. Chemie Int. Ed., 53, (2014), 14120-14123. [2] Cox, Zhang, et al., Annual Review of Analytical Chemistry, 5 (2012), 253-272.
Closing Thoughts…….
21
• SECM has a lot of powerful / unique capabilities that are of interest to a wide range of scientific fields and applications
• but……there is a learning curve (have patience!)
• Knowledge of electrochemistry helps (but not required).
• Plan 1-2 session to understand basic e-chem properties.
• Not all projects/samples are equal…
• Smooth sample, microscopic features, simple chemistry…
• Sample roughness, nanoscopic features, exotic chemistry, advanced imaging modes….. have patience!
• Keep in mind for future proposals
•Spread the word to others!
22
Questions? Suggestions?
23
•Extra Slides
Background:
24
Scanning Electrochemical Microscopy:
• A scanning probe technique that examines:
- Surface Reactivity, processes at solid/liquid or liquid/liquid interfaces, Redox reactions and kinetics involving active species, etc.
Polcari et. al. Chem. Rev. 2016, 116, 13234−13278
25
SECM: Procedures
1. Characterize the reaction of interest in your system and mode of operation (i.e. determination of applied potentials, etc.) 2. Perform an approach curve - Sets distance for scanning measurements 3. Set 2D area for measurement of sample - Set scan step size, etc. 4. Analyze signal
Inert
Electroactive features
High Low