UNOPS eSourcing v2017.1 Section II: Schedule of Requirements eSourcing reference: RFQ/2020/14947 Bidders are required to complete the Comparative Data Tables included in Section II: Schedule of Requirements to demonstrate compliance with UNOPS requirements and inserted below. Bidders are NOT allowed to make any change in the “UNOPS requirements” columns of the Comparative Data Tables. Such changes might disqualify your quotation. Technical specifications for goods – Comparative Data Table LOT 1 - LAB Ite m No UNOPS minimum technical requirements Qty. Is quotation compliant? Bidder to complete Details of goods offered. Bidder to complete 1 Micro centrifuge: Purpose: to separate suspended particles from a liquid or to separate liquids of different densities Heavy duty and compatible design 11 Yes No Insert details of goods offered, including specifications and brand/model offered if applicable ROTATIONAL SPEED To reach a maximum of 13000 rpm or more ,With the ability to choose less rotational speed , for DNA-RNA Extraction the minimum working rpm needed is up to 13000 rpm. Number of tubes (type) 24 * ( 1.5 - 2) ml ROTORS Vertical TIMER : Range, min 0 to ≥60 with increment ≤ 1min BRAKING TIME, sec variable time (speed) 1

RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

Section II: Schedule of Requirements

eSourcing reference: RFQ/2020/14947

Bidders are required to complete the Comparative Data Tables included in Section II: Schedule of Requirements to demonstrate compliance with UNOPS requirements and inserted below. Bidders are NOT allowed to make any change in the “UNOPS requirements” columns of the Comparative Data Tables. Such changes might disqualify your quotation.

Technical specifications for goods – Comparative Data Table

LOT 1 - LAB

Item No UNOPS minimum technical requirements Qty.

Is quotation compliant? Bidder

to complete

Details of goods offered. Bidder to complete

1

Micro centrifuge: Purpose: to separate suspended particles from a liquid or to separate liquids of different densities

Heavy duty and compatible design

11

☐ Yes ☐ No

Insert details of goods offered, including specifications and brand/model offered if applicable

ROTATIONAL SPEED

To reach a maximum of 13000 rpm or more ,With the ability to choose less rotational speed , for DNA-RNA Extraction the minimum working rpm needed is up to 13000 rpm.

Number of tubes (type) 24 * ( 1.5 - 2) ml

ROTORS Vertical

TIMER : Range, min 0 to ≥60 with increment ≤ 1min

BRAKING TIME, sec variable time (speed)

BRUSHLESS Yes

NOISE LEVEL, dB <70

LID INTERLOCK Yes

ALERT INDICATORS

Audible imbalance, leakage current, lid open, end of run

LINE POWER,VAC :220-240VAC , 50-60Hz

Environmental requirements

The equipment suitable for working in the climate conditions of Iraq in

1

Page 2: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

terms of temperature & humidity

Automatic breakingair vent on the coverLED display and a keyboardTachometer to indicate the speed in rpm should be availableInner safety lids that fit over the rotor or tube covers that contain the sample material should be available

Gaskets with high quality material for tight seal

Bowl drains

The centrifuge should allow calibration by phototachometer manner with lid closed

Should use at least an 8 cm rotor radius50 μL sample, and a 75 mm capillary tube, inner protective cover over the rotor assembly

2

Centrifuge

16

Spins 3ml-15ml test tubes at 0-6000 rpm

Minimum 16 tubes

Easy cleanout bowl

0 - 6000 RPM

Digital programmable control and readout

Variable speed and time settings

Lid safety lock

high g-forces for fast separation

Rubber feet to prevent slipping

Impact-resistant lid

Zero-rpm interlocking lid

Wide mouth for easy clean-up

Metal rotor and tube shieldsBrushless motor Angled and (swinging )tube holder as alternativeFast breakingProgrammable time and speed Rubber feet to prevent slipping POWER REQUIREMENTS :220v ,50 Hz

3 Water Purification System 4

Purpose: is used for producing purified water to cover laboratory needs from distilled water.

2

Page 3: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

Configuration: Bench top or wall mounting

External material: epoxy – polyester powder coated heavy duty material

Internal material: Stainless steel or Equivalent

The unit consists of Boiler & condenser· Microprocessor controlled· Distilled water capacity (20 lit/hr. )· Storage tank capacity : > 20· Alarms: visual or audial

Environmental requirements: the equipment suitable for working in the climate conditions of Iraq in terms of temperature & humidity

POWER REQUIREMENTS: 220-240V, 50-60Hz

4 Photometric Microplate Reader/Washer (ELISA) 2

Application: ELISAs assay , Protein Quantitation , Nucleic acid analysis , Cell Viability, Proliferation, and Cytotoxicity, Kinetic assay , Enzyme assay , evaluate the quantity of antigen or antibodies associated with infectious viral diseases (e.g., HIV, measles, herpes, rubella) , other nonviral applications (e.g., mycoplasma pneumoniae) , other to be specified .

A - Photometric micro plate readers: PHOTOMETRIC METHOD: yes , manufacturer dependent on OPTICAL SYSTEM : measurement and reference channels required , multiple channel preferred

photometric accuracy : < 2 %measurement range, ABS : between 0 - 3,0, Or better wavelength range, nm : between 400 – 700,Or betterprecision: ±1%resolution, od : ≤ 0.01light source : Tungsten halogen or xenon flashphotodetector : photomultiplier tube(PMT) preferred and other to be specifiedFILTER TYPE :according to manufacturercompatible micro plates : flat or round or bottom , 96 -well platesIsothermal reading chamber : optionalprogrammable assays : preferredAnalysis: Reading speed, sec/plate : ≤

3

Page 4: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

15/96-wellComputer interface :yesPrinter : yes (external or built-in)Bar-code reader: OptionalRobotics compatible : yesSelf-calibration :yesQuality control :yes B- Microplate Washers:Configuration: full plate washer or other to be specifiedautomatedwashing parameters : wash volume/time, soak time and otheruser programmable: yesPrograms:

- user definable: yes- number of cycles : ≥ 8- maximum soak time, sec: ≤

3.600 compatible plates: well plates other to be specifiedcapacity, number of plates: ≥ 1Wash head : multiple interchangeable wash heads of varying configurationsfluid reservoir :

capacity, l

wash ≤ 4rinse < 4waste ≤ 4liquid-level sensing Yesprecision, % cv ≤ 3minimum residual volume µl / well ≤ 5maximum dispense volume ml /well ≤ 3 power requirement( for system)

220/240 VAC, 50/60 HZ

Environmental requirements: The equipment suitable for working in the climate conditions of Iraq in terms of temperature & humidity

5 Portable Glucose Analyzer 2

APPLICATION

measure blood glucose concentration using a reagent strip and a drop of capillary blood from a finger puncture

PATIENT TYPE Adult , paediatric, neonate

METHODOLOGY

Electrochemical or Reflectance photometry( preferred) and other

REAGENT STRIPS YesSPECIMEN TYPE Capillary whole

4

Page 5: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

Minimum sample volume, µL

blood, venous blood, arterial blood, neonatal blood

≥ 0.3 MEASUREMENT RANGE, mg/dL 20- 500 or better

ALTERNATE-SITE TESTING

Forearm, palm, upper-arm

TEST TIME, sec Between 3 to 60METER CALIBRATION Yes

LOT-SPECIFIC CALIBRATION Yes

READOUT LCD

ALERT INDICATORS• High/low levels• Other

Out of measuring range, out of temperature rangeHypoglycemia, hyperglycemia, low battery, under-dose detection

AUTOMATIC SHUTOFF After 2 -5 min

BATTERY Yes

Life, no. of tests ≥ 1,000

Environmental conditions

The equipment suitable for working in the climate conditions of Iraq in terms of temperature & humidity

6 Immunoassay Analyzer, Automated 1Application It is used in hospital

and clinical laboratories to run automated biochemical tests to detect the presence and concentration of substances in the samples. There are many types of tests that immunoassay analyzers can perform, including testing for cancer markers, diagnosing infectious diseases (viral or bacterial toxin determinations), cardiac analysis, therapeutic drug monitoring, allergy testing, Endocrine hormone testing and

5

Page 6: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

protein.

CONFIGURATION Bench top and floor type

Connectable to LAS track Yes

Processing modes Continuous, random access ,stat

THROUGHPUT, max tests/hr ≥ 100SAMPLE CAPACITY ≥ 50

SAMPLE TYPESerum, plasma ,urine other should be specified

SAMPLE SIZE, µL < 250 Smallest is preferred

METHOD USED

Chemiluminescence immunoassay or Electrochemiluminescence (ECL) or fluroscence immunoassay(FIA )

TEST MENU

Anemia , Diabetes ,Cardiac markers, Endocrine function , Immunoglobulin, Toxoplasma, Thyroid markers ,Viruses, Tumor marker or others

REAGENTS, TYPE

Refrigerated on board

Liquid, ready-to-use or other type

YesSYSTEM FEATURES:

-Direct tube sampling-Liquid-level sensing-Clot detection-Auto dilution-Auto calibration-Auto wash

All Features should be available

DATA MANAGEMENTDisplay LCD , ≥ 15"Results stored YESComputer interface YESPrinter YESBar-code reader YESLINE POWER, VAC, Hz

220~240 VAC, 50/60 Hz

Backup power preferred Battery or UPS

Environmental requirements

The equipment suitable for working in the climate conditions of Iraq in terms of temperature & humidity

6

Page 7: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

7

Laboratory Incubator

1

Purpose: used for growing microorganisms and cells that require oxygen for growth (e.g., bacteriology, virology, cultures) in hospital lab, specialist health center lab. & lab. of Primary health care unit.Configuration: tabletop or Floor TypeInner chamber stainless-steel and out chamber heavy duty materialType door : glass or / & solid doorKey Door lock :yesCapacity , L : 120 – 200 LNumber of Shelves : ≥ 2 , stainless steelTemp is controlled by thermostatTemperature range , °C : 5 to 50 °CTemperature Uniformity +/- 0.5°CAccuracy : ± 1 °CIllumination : internal lightControl system : microprocessor controlled temperature regulation display : digitalAlarms: audio & visual, Temperature, power failureThe heated air circulates through the chamber by fansAll the control switches & Pilot lamps are fitted on the front panelPower requirements: 220-240V, 50-60HzEnvironmental requirements: the equipment suitable for working in the climate conditions of Iraq in terms of temperature & humidity

8 Light Microscope 10Purpose : to provide magnification and resolution of specimens under examinationHeavy duty and compatible designOBSERVATION TUBES Type binocular

Eyepieces 10x,15x, wide field, high-eye point

Interpupillary distance adjustment

mm : ≥( 55 -75), Adjustable

NOSEPIECE CONFIGURATION Quadruple or more

Objectives Magnification

(4x,10x,40x,100x) ,other should be specified

Objectives Type

Achromatic, plan achromatic, plan apochromatic ,or others

TOTAL MAGNIFICATION ≥(40x-1000x)CONTRAST METHOD:Brightfield: yesDarkfield: yes Phase: optionalILLUMINATION:

7

Page 8: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

Condenser type: Abbe, achromatic &/ or other typeLight source: LED (preferred),halogen, or other STAND:Focusing mechanism: Moving stage, rack and pinionCoarse, fine adjustments: Coaxial, low mounted, ≥ 1µ incrementsSTAGE Type: Mechanical or betterTension adjustment CoaxialStage Motion X-YNumerical aperture 0.9 to 1.4Dust cover Yes

Environmental requirements

the equipment suitable for working in the climate conditions of Iraq in terms of temperature & humidityPOWER SUPPLY: 220-240V,50-60Hz

9 Spectrophotometer: UV VISIBLE SPECTROPHOTOMETER

7

Purpose: To determine the absorbance or transmission of characteristic wavelengths of radiant energy by a chemical species in solutionCONFIGURATION: Double beam

LAMP TYPE(S) Halogen (Vis), deuterium (UV)

WAVELENGTH, nm Range : 190 -1,100

Accuracy: 0.3 -1MONOCHROMATOR Holographic grating

PATH LENGTH, mm 1-100STRAY LIGHT Minimum PHOTODETECTOR Silicon diode

Scan speed variable to at least 300

Accuracy ±<1%Reproducibility <0.5% @ 0-2 ANoise at 0 A <0.0002 ALinearity ±0.003 A/hrZERO DRIFT <0.0002 A/hrAuto Compensation yes

TEST TUBES different volume test tube holder

Display Color LCD DisplayPRINTER Ink jet, LaserJet

User programmable

Temperature control

8

Page 9: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

PC-controlled Windows-based software to provide spectrum processing and storage, complex data calculation routines, Absorbance, factor, concentration, transmittance, scanning, DNA ratio, kinetics, calibration curves, linear and nonlinear, performance validation, multi method/result storage capability

Protein assay, DNA/RNA purity, concentration by factor method and enzyme kinetics, bacterial density, biocalculations

multiwavelength, peak picking photometryWavelength scanning, spectral manipulation, time course plot, superimposed plotQuantitative analysis, wavelength scan, time scan, fixed wavelength, baseline correction, absorbance/percent T, peak picking, peak height, peak area, peak width, derivativesMulticell/wavelength measure, peak seek, derivative, autozero, reaction kinetics with plot, histogram plot, gel scanning, curve fit and storage, formula entry, for multi derivativesComputer System interfaceMicrocell holder, test-tube holder

10

Laboratory Steam Sterilizer (30 L)(Laboratory autoclave)

12

Type Steam sterilizerTable Top or on Trolley Operating temperature Max. 134 ± 5℃ ℃Pressure indicatorHeater: ElectricChamber material: Stainless steelChamber capacity: Approx. 30 LControl: Automatic control from sterilization to dryingNumber of instrument tray: Max. loadable number of offered itemTray RackSafety device: Automatic Door Looking system

Power requirement AC220V, 50Hz, single phase

11 Laboratory Water Bath 6Applications: sample thawing, bacteriological examinations, warming reagentsHeavy duty and compatible designMaterial: seamless stainless steelExternal: stainless steel, powder coated Temperature range: (10 C above ambient to 100)On off switch and display by thermometer graded for temperatureAccuracy: 0.2ºC Temperature preset buttons

9

Page 10: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

Circulation Pump to mix waterMicroprocessor controlledHeater: 1000- 2000 wattsSafety Device for overheating protectionBath Volume: 20 litersWith stainless steel gabled coverThe exterior should remain cool, glass wool insulatorAlarm indicator for overheat limits, low water cutoffPower Requirements 220-240 V, 50/60 Hz

12 Laboratory Oven 50 L 11Temperature range 50°C ~ 300°CTemperature control: Thermostatically controlledInside material: Double /Triple walled Construction inner chamber made of Stainless steel with supports on the three sides for two/three adjustable perforated stainless steel shelves

Outside material: - Mild steel sheet painted/powder coated- Insulated stainless steel door with locking and rear zinc-plated steel

Capacity > 50 Itr.

Safety Alarm

- Yes (Audible and visual) - Optional automatically shut down protection circuit

Air circulating fanDigital timer

Power supply

- 220/240 VAC, 50/60 Hz - Over current & over heated protection circuit - Three Pin UK type Plug - Voltage corrector/stabilizer circuit of (Input 160-260 V and output 220-240 V and 50 Hz)

Other specification - On/off rocker switch with indicator - Adjustable wire mesh nickel plated trays/aluminum trays- Digital indicator cum controller - The gap between the two walls filled with glass wool insulation

10

Page 11: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

- Easy-to-clean interior - Forced air circulation by quiet air turbine/Fan to ensure uniform temperature - Fitted with load indicator and safety thermostat take over indicator lamp. LCD/LED Indicator - Output available for data acquisition

Environmental requirements

The equipment suitable for working in the climate conditions of Iraq in terms of temperature & humidity

13

Chloride Meter

11

Easy, intuitive user interface with touch screen and multi-language prompts, which lead the operator through the entire measurement process. Factory calibrated so that manual calibration by lab personnel is not required. It can be equipped with an integrated printer and the results are sent in document-ready format or a PC for data transfer can be connected via USB .

Sample volume Approx. 55 µl, or less

Duration of measurement Approx. 30

secondsReproducibility ± 2 digits (20 µl) at

100 mmol/lMeasurement display 0 - 999 mmol/l

Measurement range 10 - 299 mmol/l

Resolution 1 mmol/l over the entire measurement range

Integrated stirring Magnetic stir bar PTFE, cylindrical

Power source 220 V (± 30V), 50Hz, 40 VA,

The measurement principle is based on color metric impulse titration as an absolute measurement method. Through automated sample recognition, the measurement proceeds fully automatically after the sample is inserted without the need for manual confirmation by the user.

14 Dry Chemistry Analyzer (Strip) 13Application: used for measuring

11

Page 12: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

quantitative results of many parameters from whole blood, plasma or serum within short time.

Parameters:

Bilirubin, Cholesterol, Creatinine, Glucose, Hemoglobin, Cholesterol, Triglycerides Uric Acid, Urea, Alkaline Phosphates, or other

Detecting method : Reflectance photometry, or other

Light source: LED

Light detector : photodiode or better

Sample type whole blood, plasma or serum.

Sample volume ≥ 25 μlMeasuring time: less than 5 min.Calibration: AutomaticDisplay: yes , LCD type Memory: ≥ 40 results with

( time ,date , ID)Built in PrinterData interfaceNoise level: < 50 dBA

UPS YES, power back up for ≥ 10 min

Environmental conditions:

The equipment suitable for work in the climate conditions in Iraq in term of temperature & humidity.

POWER REQUIREMENTS: 220-240V, 50-60Hz

15 Hemoglobin Meter 7

PURPOSE: For hemoglobin estimation

Detector: Silicone photodiode other should be specified

Measuring: Transmission/Absorption

Photometric accuracy: ± 0.002

Cuvette types: should accept different type of cuvettes

Display: hemoglobin level in g/dl or g/l, date and time

Measuring range:   0 - 25g/dl Hb, or better

Dilution ratio: should be stated Sample volume: minimum sample volume should be stated

12

Page 13: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

CalibrationQuality control with control cuvette

Sample material : Venus ,arterial or capillary

Environmental requirements

The equipment suitable for working in the climate conditions of Iraq in terms of temperature & humidity

Open system Power supply: 220V,50HzSample compartment: should be prepared for the use with cuvettes, integrated cover for the protection of the sample compartment against ambient light and dirt

16

Bilirubinometer

7

Application To measure the concentration of bilirubin in the blood

Where used Stat labs (hospital) ,amedical clinics

CONFIGURATION Benchtop

METHOD USED Spectrophotometry (Dual wave lengths)

Patient Neonate, optional (adult , pediatric)

Bilirubin measured Total, optional dirct SAMPLE TYPE Serum or whole

blood or plasma SAMPLE VOLUME, µL

≤60

Measurement time, sec

≤5

Calibration Automatic(preferred) Computer interface Optional features

(computer interface)USB port

Output Units: µmol /L or mg / dLRange, mg /dL: 0-30 or other rang corering this range

Display, type LED or LCD power 220 VAC ,50/60 HZ

Environmental requirements

The equipment suitable for working in the climate conditions of Iraq in terms of temperature & humidity

LOT 2 – DENTAL Item UNOPS minimum technical requirements Qty. Is quotation Details of goods offered.

13

Page 14: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

No compliant? Bidder to complete Bidder to complete

1

Dental Unit

5

Pneumatic dental unit with automatic separator.Gearbox motor or hydraulic for dental chairMinimum height of dental chair (35 ~ 38cm).Maximum height of dental chair (75 ~ 80 cm)Adjustable headrest.Light projector with armDentist tray:Hand piece hanging type.Control panel (with X-ray viewer).3-way syringe.3-module Midwest (4-hole).Ultrasonic scalar.2 turbines (Midwest 4 hole, standard head, stainless steel body, ceramic bearing, push button chuck & autoclavable) one with fiber optic and small head and other turbine standard head without fiber optic.Air motor (forward & reverse control) Contra angle (1:1 direct drive, push button chuck, ball bearing, autoclavable, external water spray, non optic). Assistant tray:Control panel.3-way syringe module.LED curing light.Dry-line suction & saliva ejector & sucker & separatorSpittoon section:Spittoon.Cup fillerDental stool.Assistance stool. Distilled water tank.Air compressor with sound proof (cabinet).Air-suction motor.Foot controller.Input power: 220/240 VAC, 50/60 Hz single phase.External voltage regulator.External disposable filter for waterIntra-Oral Camera Environmental Requirements: The supplier confirms that: The equipment suitable for work in the climate conditions in Iraq in terms of temperature & humidity.

2 Digital Dental Radiography 3Type Dental X-Ray DigitalApplication X-ray imaging for

teeth and periodontium

X-RAY TUBE ANODE

- Type : Stationary - Minimum total filtration, mm Al:≤

14

Page 15: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

2.5- Heat capacity, HU: ≥ 6,000- Focal spot size, mm: ≤ 0.7

X-RAY GENERATOR

- Type: High-frequency- kV range: ≥ 60- mA range: Range ≤ 2 to ≤ 8

CONTROL PANEL According to manufacture (Digital)

MOUNTING Stand or wallIMAGING METHOD

Time to process and display image, sec Bit depth IMAGE SIZES, mmPixel size, µm

- Detector type : CCD or CMOS (complementary metal oxide semiconductors), Wire or wireless, water proof

≤ 5

≥ 12≥ 30 X 20≥ 15

IMAGE PROCESSING FEATURES

Standard image manipulation tools

NETWORKING YesPRINTER Laser COMPUTER SYSTEM

- up to date system - Monitor size, type : ≥ 17" LCD- Resolution, pixels : ≥ 1024 x 769- Storage : ≥ 500 GB hard disk drive

POWER REQUIREMENTSSMART UPS

220/240 VAC , 50/60Hz, Three pin plug (G type)≥ 15 Minute for Whole System .

Environmental Requirements

The supplier confirm that: The equipment suitable for work in the climate conditions in Iraq in terms of temperature & humidity.

3 Dental Ultrasonic Scaler 2Optic or Non-Optic, headpiece with suitable length of cordLCD display preferableSterilization caseHand piece standWide selection of scaling tips, Priced separatelyFrequency: 27KHz -33KHz

15

Page 16: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

Output power: 3-30WContents: control unit, foot control, bottle of water, and AC power cordPower supply: AC (220 -240)v (50-60)Hz

4

Dental Unit Autoclave

2

Application: sterilization of solid instruments, hollow instruments, packed instruments & porous loads. Microprocessor controlled.Self-evaluations on the process of sterilization, automatic failure detection report. Instant vaporizer outside the chamber.Inner Chamber capacity: (18~23) liter, stainless steel.Integrated printer (optional).Safety locking systemLCD displayWater supply system: reservoir (≥3.5 L distilled water and ≥ 5L waste reservoir).Possibility of serial connection to a PC, with cycle data recordingTray holder contains ≥2 tray.Vacuum pump performing two stage (pre & post vacuum).Thermal drying system to ensure sterilizing dental instruments without spotsFull water alarm system.Input power: 220/240 VAC, 50/60 Hz single phase. Accessories: tray extraction handle.Replacing the device filters covered in the warranty.Environmental Requirements :The supplier confirm that: The equipment suitable for work in the climate conditions in Iraq in terms of temperature & humidity.

5

Light Cure

1

Application: Solidify all brands of resin materials of light cure.Heavy duty & compatible design.Light source: LED.Light probe: autoclavable.Adjustable working time.Spectrum range: (400 ~ 500nm).Protection fuse.Acoustic signal: acoustical after 10, 20, 40, 60 sec etc.Input power: 220/240 VAC, 50/60 Hz single phase. Accessories: - Eye-protecting cover or optic light

guide.- Additional light probe.

LOT 3 – IMAGERY

16

Page 17: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

Item No UNOPS minimum technical requirements Qty.

Is quotation compliant? Bidder

to complete

Details of goods offered. Bidder to complete

1 Mobile X-Ray Unit Digital 5APPLICATION Radiographic imaging

of patients who cannot be transported to radiology department

SYSTEM TYPE Description

Mobile digital radiographic systemIntegrated digital (not upgraded from analog to digital)i.e. (not retrofit)

TYPE OF POWER SOURCE BatteryCan be directly powered by AC line

 Yes Yes

X-RAY TUBE ANODE

Heat capacity, HU Cooling, HU/min Maximum output, kVp Range Increments mAs range

Rotating

≥ 200,000≥ 20,000≥ 120 ≥ (40-120) Yes ≥ 300

X-RAY GENERATOR Power rating, kW

High-frequency generator ≥ 30

ALUMINUM FILTER, mm

Yes

SID RANGE, cm According to manufacturer

INDICATOR METERS

Type kVp mAs

 

Digital Digital Digital

TUBE MOVEMENT

Horizontal, cm

Vertical, cm

Rotation :

Z-axis, °X-axis, °

Applicable for the need of different radiological examination Yes (specified by the manufacturer)Yes (specified by the manufacturer) 

Yes (specified by the manufacturer)Yes (specified by the manufacturer)

HANDHELD CONTROL Cord length, m

Yes

Wire or wireless DIGITAL  DR

17

Page 18: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

DETECTOR Type Pixel size Wireless Size, cm Size of built-in display monitor Type of network connectionProcessing (analyses) toolsFilm storage capacity

 CSI≤ 160 µm Yes ≥ 35X43≥ 15" color screen LCD medical monitorDICOM 3.0Specified by the manufacturer ≥ 120 GB

PATIENT-CENTERING/ COLLIMATOR LIGHT

Yes

Power supply Line voltage

- Medical Approved power supply board - 220/240 VAC , 50/60 Hz single phase,- Three pin UK type plug

2 Portable Ultrasound 3Scanning mode  Convex/Linear/Micro-

convexDisplay modes  B,B+B,B+M,M,4BScanning depth 240mmGray scales  256Cine-loop  ≥ 500 frames Image permanent storage 

≥ 64 frames

Image conversion up/down, left/rightLocal zoom 2 times both in real-

time and frozenMeasurement Distance, circumferenc

e, area, volume, , heart rate

OB measurement BPD, CRL, FL, AD, HC,AC, EDD, GA, FW

Probe connector  ≥ 3Monitor ≥ 15 inch Wide application  Abdominal, OB/GYN, 

urology, Doppler, cardiology, small parts examination, …etc

Probes Standard configuration approx 2-5 MHz convex probe ,transvaginal 4-8 MHz Optional configuration optional probe ( 5-10MHz Linear,Probe)

Power Supply - 220/240 VAC, 50/60 Hz - Over current & over heated protection circuit

18

Page 19: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

- Three Pin G type Plug- Voltage corrector/stabilizer circuit of (Input 160-260 V and output 220-240 V and 50 Hz)-UPS (Rechargeable battery) of suitable rating with voltage regulation and spike protection for ≥ 15 minutes back up

LOT 4 - MED

Item No UNOPS minimum technical requirements Qty.

Is quotation compliant?

Bidder to complete

Details of goods offered. Bidder to

complete

1

Otoscope

5

Offering clear, live and still images of the ear canal and tympanic membraneChrome Plate handleStandard, heavy duty rechargeable batteries.Reusable and disposables specula different sizesSoft pouch or hard Plastic caseRheostat for adjustment of illuminationFiber optic for cool, obstruction free light transmissionMagnification z(2.5-≥3) X.Bulb: LED typeEnvironmental requirements: the equipment suitable for work in the climate conditions in Iraq in terms of temperature & humidity

2 Fetal Heart Detector, Ultrasonic 8TYPE/CONFIGURATION

Handheld

APPLICATIONS Detect fetal heartbeat measure FHR

FHR DISPLAY RANGE, bpm

≤ 60 - ≥199

PROBEFrequency, MHzOutput, mW/cm²

≥ 12 to 3< 20

OUTPUTSSpeakerHeadphones

YESOptional

DisplayDisplay variables

LCD , LED or better FHR

CONTROLS Volume, power on/off

POWER REQUIREMENTS Rechargeable

19

Page 20: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

Battery, type

Operating time, hrLow-battery indicator

preferred or Alkaline or dry cell ≥ 2YES

Accessories Ac adapter for rechargeable battery CaseGel (start up qty.)

Environmental requirements:

The supplier should be confirmed that: The Equipment is suitable for work in climate conditions of Iraq in terms of temperature and humidity.

3 General Surgical Operating Table 2Application General TABLE POSITIONS(Degrees from horizontal)Trendelenburg

Reverse Trendelenburg

Lateral tilt

≥25

≥25

≥18Vertical range, cm (in) Height adjustment

with mattress(750-980) mm

TABLE SECTIONS Head section (removable), Back section, seat section, Split Leg section (Removable)

Back section (+85 Degree, -45 Degree)

Foot/leg section (+20 to -90 )Head section VariableACCESSORY SIDERAIL

Yes

PERINEAL CUTOUT YesMAXIMUM PATIENT CAPACITY, kg (lb)

Max. dynamic load capacity ≥ 200

CONTROLS

Type

Manual override

Wired remote control and Foot controlElectric or Electrohydraulic or Electromechanical Yes

Radiolucent tabletopC-arm accessible

YesYes2 arm board 2D

20

Page 21: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

ACCESSORIES Left & Right with attachment clamp, pair of shoulder support with attachment clamp , Body straps, Anesthesia screen

DIMENSIONS, H x W x L, cm (in)Fully extended

(width , Length 4 section) W 500mm or less , L 2000mm or more

Column housing Stainless steelCaster lock Yes POWER REQUIREMENTSBattery

Power input 220-240VAC, 50Hz.YES

Environmental requirement

The supplier should be confirmed that: The Equipment is suitable for work in climate conditions of Iraq in terms of temperature and humidity.

4 Infant Incubator 11Application Used for providing

a closed controlled environmental that warms an infant by circulating heated air over the infant skin.

Operating modes Air temperature control.

Skin temperature control (servo controlled)

MOBILE Yes, castors( size≥ 10 cm) with breaks

TEMPERATURE CONTROL RANGE, °C:Air (Increment: 0.1°)

Skin (Increment: 0.1°)

Min:(20-25°) Max: (37- 39°)°

Min:(33-35)°

Max:(37-38)°

TEMPERATURE DISPLAY (air & skin)

Digital or LCD(LED)

ALARMS High air temp

Low air temp High skin

21

Page 22: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

temp Low skin

temp Fan failure Sensor

failure Power

failureBACKUP THERMOSTAT

YES

HEATER POWER INDICATOR

YES

HAND PORTS ≥4TUBING PORTS ≥5O2 SOURCE

Inlet Ports

Controllers

Alarm

≥1

servo controller preferred ( 21%-60%)

YES

Supplemental HumidityAdjustable Water tank for humidity control

YESYES (Digital controller)Shall be removable and sterilizeable

Internal Noise, dB ≤ 60Double Wall YESMattress Water proof Trendelenburg + 12 degreeAccessories IV pole.

Drawers in-bed

scale integrated

x-ray tray Disposable

temperature sensors (25 pcs)

Options: Battery up for alarms & Trends to 1 hr.

Variable-height stand.

Environmental requirements

The equipment suitable for work in the climate conditions in Iraq in terms of temperature & humidity.

22

Page 23: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

Disinfection: According to manufacturer The manufacturer should specify the types of disinfectant suitable to the incubator. Should be not affect the incubator coating & environment.

5 Infant Scale 6Material & Design 1)Ergonomic tray

provides secure

support with a

robust plastic tray

for easy and

hygienic cleaning

2)Secure and non-

slip positioning with

rubber feet

Power supply - 220/240 VAC,

50/60 Hz

- Over current

protection circuit

- Three Pin UK type

Plug

- Voltage

corrector/stabilizer

circuit of (Input

160-260 V

and output 220-

240 V and 50 Hz)

- Can work using

( alkaline or

chargeable)

battery(s)

- Battery Life > 200

hours for not

chargeable type of

batteries

Other specification 1) simple and

convenient 1-key

23

Page 24: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

operation

2) Automatic Data

Hold Function

Batteries included

Weighing range >10KgWeighing accuracy 0.01 KgDisplay Digitaldimensions(W x D x H) mm

To be mentioned

6 Infusion Pump 2Application Accurately deliver

liquids through intravenous (IV) or epidural routes for therapeutic and/or diagnostic purposes

CONFIGURATION Number of channels

 ≥1 

DISPLAY Data displayed

 Alarms, drug name, pumping status, volume infused, VTB

PUMP CAPABILITIES Flow range, mL/hr Increments, mL KVO rate, mL/hr Accuracy, % VTBI Selector, mL Automatic piggybacking Fluid resistant Front-panel lockout

 

≤ 1- ≥999 0.1-100 mL/hr 1-≥3±5 Yes(1-9,999) Yes( if number of channels > 1).Yes Yes

Type of system OPEN SYSTEM, the pump must be able to operate using standard (non-dedicated) IV set from multiple manufacturers

IV SET Free-flow protection Air-trapping capabilityNeedleless IV connection

  YesYesYes

ALARMS & INDICATORS Occlusion upstream Occlusion downstream Detection mechanism Pressure, psi Real-time display Air in line System malfunction Set loaded improperlyDoor open

 

Yes Yes Yes ≤15 YesYes Yes YesYes

24

Page 25: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

Infusion complete Low battery Depleted battery Clinical advisory messages

Yes Yes Yes Yes

AUDIBLE ALARM Volume control Momentary silence

 YesYes  

DOSE ERROR REDUCTION SYSTEM (smart technology)Defaults to DERS on startupLibrary sizeNo. of care areasNo. of drug entities per care areaDATA PORT

Yes

Yes

≥ 100≥10≥100Yes

EVENT LOG Events stored

Yes , by InterfaceKey presses, program settings, alarms, volume infused, dose limit warnings

*Power supply Line voltage BatteryLife, hr @ flow of 100 mL/hrRecharge time, hr

Medical Approved power supply board 220/240VAC,50/60Hz, Three pin (G type) plug YES≥5<8

7 Mobile Surgical Light 4Application Surgical lights

illuminate the surgical site for optimal visualization of small, low-contrast objects at varying depths in incisions and body cavities

LIGHT SOURCE, TYPE Number of light heads

Number of LED s

Volts

Service Life, hr

LED Single

Should be specified

According to LED type

≥ 30,000

COLOR TEMPERATURE, K Adjustable

4,000-5,000 Yes

COLOR RENDERING INDEX

R9 value

≥ 94≥ 90 (higher preferred)

25

Page 26: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

FIELD SIZE, cm Diameter Depth

≥16 ≥ 80

FOCAL LENGTH, cm ≥ 65

ILLUMINATION LEVEL, maximum lux at 1 m

≥ 100,000

MOUNTING mobile floor stand

CONTROLS Dimmer Focus Field size On sterile handle

 YesAdjustableAdjustableYes 2 for each unit

REFLECTOR MATERIAL

According to manufacturer

Light head diameter, cm

To be mentioned

Power supply - 220/240 VAC, 50/60 Hz- Three pin G type plug- ≥ 1 hours back up battery(s) with automatic charging and change over circuits

8 Phototherapy, Free Stand 8

TYPE FreestandingANGLE ADJUST, ° AdjustableIRRADIANCE @ 420 to 480 nm, µW/cm²/nm

400-550 (any value between)

FAULT PROTECTION Fuse or breakerLIGHT SOURCES LEDType: LED Number needed: To be specified by manufacturerExpected life, hr ≥ 8,000HOUR METER YesPROCEDURE TIMER PROCEDURE

TIMERHeight adjust Yes

≥ (107- 150) or wider range

Radiometer Optional Others The unit should be

mobile with swivel castors fixed to a base to be accommodated beneath incubator /cot.Accessories: glasses for eye protection & UV filter cover.

Power & Environmental

Input power: 220/240 VAC, 50

26

Page 27: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

Requirements Hz single phase, minimum length of the wire (3m). Three Pin G type Plug, Voltage stabilizing and over current protection circuit.The supplier should be confirmed that: The Equipment is suitable for work in climate

9

Sphygmomanometer

49

Stationary blood pressure measurement device..The precision pressure gauge is recessed in a premium, impact-resistant, is easy to clean the ABS housing.The easy to use spiral hose, inflation bulb with precision adjustable threaded discharge valve and optimized readability on the oversize scale through anti-glare glass guarantee ultimate functionality.

10

StethoscopeAcoustic sensitivity and durability.Traditional bell.

44The device has conventional combination of two-sided chest piece. One side of the device has tunable diaphragm and on the other side it hasThe headset is anatomically designed. black tube colors.

11 Surgical Suction Unit 14Application: suctioning to remove blood and irrigating fluids that accumulate in the operative field and obstruct the surgeon's view, used in OR, Wards , ICU, and others.CONFIGURATION: Mobile on cart with castorsPUMP TYPE: Piston, or Diaphragm, Others to be specifiedNumber of pumps: ≥ 1VACUUM LIMIT, mm Hg, (cmH2O): ≥ 500 , (679.75)Free airflow, L/min ≥25VACUUM CONTROL YESCOLLECTION CANISTER(S)

Reusable &Disposable

Number ≥2Capacity limit line, mL ≥2000Overflow protection YESFILTER TYPE Bacterial,

hydrophobicSTANDARD ACCESSORIES(Included in Warranty)

Collection bottle.

Filters. Cart.

27

Page 28: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

Optional **footswitch

Electrical Ratings - Power Supply -220/240VAC- Rated Frequency 50/60Hz

Environmental The supplier should be confirmed that: The Equipment is suitable for work in climate conditions of Iraq in terms of temperature and humidity.

12

Ophthalmoscope

4

Application Diagnosis of eye

Component Interchangeable handleOphthalmoscope head

Battery (dry cell or rechargeable )

Power requirement Ac Adapter be 220-240VAC, 50Hz.For rechargeable battery

Corrective power range

From- 20 to +20 or wider range

Aperture: 5 apertures or moreFilter red-free , others to

be specified Light source LED

color rendering index > 95

Number of Lenses ≥ 18Environmental requirements:

The supplier should be confirmed that: The Equipment is suitable for work in climate conditions of Iraq in terms of temperature and humidity.

13 Digital Scale 11ElectronicLCD displaycolumn typewith BMI calculationkg (496.04 lb) \0.1 kg (0.22 lb) or morewaist-level digital physician scales LCD display with Body Mass Index

28

Page 29: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

calculation powered by 6 AA batteries or equivalent or optional AC adapter (neither included). Height rod available.500 lb x 0.2 lb / 225 kg x 0.1 kg capacity. or more

14

Oxygen Cylinder

5

The Steel "M" medical oxygen cylinder back up units or to use to supply oxygen filling system. The Steel "M" cylinder can supply up to 3540 liters of oxygen. The cylinder comes standard with a CGA540 valve and can be conveniently stored using our STANDM60T01-A adjustable cylinder stand.

15 Surgical Curettage Set 3All should be stainless steel – each set should contain the followings: DRESSING FORCEPS, 25 CM 1 pc. setTOWEL FORCEPS, BACKHAUS, SHARP, 9 CM 4 pcs. setSPONGE FORCEPS, BOZEMANN, S-SHAPE, 26 CM 1 pc. setSPONGE FORCEPS, RAMPLEY, STR., 24.5 CM 2 pcs. setCATHETER F. WOMEN, CVD., 14 CH, 15 CM 1 pc. setVAGINAL SPECULUM, SIMS, 65X25/70X30 MM 1 pc. setVAGINAL SPECULUM, SIMS, 70X30/70X35 MM 1 pc. setVAGINAL SPECULUM, AUVARD, 75X43 MM 1 pc. setUTERINE DILATOR, HEGAR, DOUBLE, 3/4 MM 1 pc. setUTERINE DILATOR, HEGAR, DOUBLE, 5/6 MM 1 pc. setUTERINE DILATOR, HEGAR,

1 pc. set

29

Page 30: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

DOUBLE, 7/8 MMUTERINE DILATOR, HEGAR, DOUBLE, 9/10 MM 1 pc. setUTERINE DILATOR, HEGAR, DOUBLE, 11/12 MM 1 pc. setUTERINE DILATOR, HEGAR, DOUBLE, 13/14 MM 1 pc. setUTERINE DILATOR, HEGAR, DOUBLE, 15/16 MM 1 pc. setUTERINE DILATOR, HEGAR, DOUBLE, 17/18 MM 1 pc. setUTERINE PROBE, SIMS, MALLEABLE, 33 CM 1 pc. setTENAC. FCPS., SCHROEDER, 1X1 T., 25.5 CM 2 pcs. setUTER. CURETTE, 6 MM, SH., RIG., 30.5 CM 1 pc. setUTER. CURETTE, 11 MM, SH., RIG., 31 CM 1 pc. setUTER. CURETTE, 14 MM, SH., RIG., 31.5 CM 1 pc. setUTER. CURETTE, 6 MM, BL., RIG., 30.5 CM 1 pc. setUTER. CURETTE, 11 MM, BL., RIG., 31 CM 1 pc. setUTER. CURETTE, 14 MM, BL., RIG., 31 CM 1 pc. setOVUM FORCEPS, WINTER, CVD., N. 2, 28 CM 1 pc. setTissue forceps 2 pcs. setRetractor, small 1 pc. setRetractor, medium 1 pc. setScalp handler, 3, 7 and 3L one each

1 set set

30

Page 31: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

Sterilizing Case 1 pc. set

16

Lead Barrier 3mm Lead 1x2 M

9

Small Window X-ray Mobile BarrierDimensions (80*190CM).(W*L) ,Or better Thickness of 2mm or better Powder coated steel frames, precision TIG welded for years of trouble-free service completely enclosing the leaded glass panel for increased durability and safetyLead glass or lead acrylic options available3 mm lead equivalent opaque base covered in “Sintra” plastic for easy cleaning and durabilityModular parts can be easily replaced, including the glass if needed.Stainless steel or coated metal finish. Swivel casters.Window OptionsLead: 2 mm Pb equivalencyAcrylic: .50 mm Pb equivalency

LOT 5 – MEDICAL FURNITURE

Item No UNOPS minimum technical requirements Qty.

Is quotation compliant? Bidder

to complete

Details of goods offered. Bidder to complete

1

Bedside Cabinet

5

With metallic drawer and cupboard.Upper board made of plywood with antis hock edges in ABS plastic.Feet with non-slip ends.Made of epoxy coated sheet steelDIMENSIONSApprox. 390 x 480 x 880 mm

2

Examination Couch With Steps

23

Backrest adjustable by self-locking system with bilateral controlThick foam with high quality vinyl fabric upholstery, at least 65 mm thickPaper roll holder with one rollWelded tube reinforcementTips of pads adjustableTubular stainless steel frame Ø 30 mm or better Non slippery, two level stepladder made up of stainless steel Or coated metal with a tubular frameApproximate dimensions (L x W): 1800 x 600 mm

3 Drug Trolley 4Transport, multi-function2 detachable 2 pressed trays or more, with dimensions of minimum 600 x 400 x 60 mm.Two drawers or more

31

Page 32: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

Handle to facilitate movementsø80 mm casters or more, with bumpers (two with brakes).Slide removable basketMade of stainless steel Minimum Dimensions: 950 x 420 x 830 mm

4

Stainless Steel Paravane

22

Folding 3-section screen panel for hospital useConstructed of stainless steelWashable curtains (White color fabric)Each screen panel folds independentlyWith rolling wheels Dimensions: - Hight: 180 cm approx.- Width (for each section): 60 cm approx.- Total width: 180 cm approx

5 Delivery Bed 2Application: used for birthing cases. Table’s upper flat surface has ≥3 sections.Frame: powder coated or stainless steel bed frame.Mattress: dynamic surfaces (multi-layer plywood, foam & leather).Height, back rest, Trendelenburg, table’s head & foot parts are controlled by controller buttons within the side rail.Type: electro-hydraulic.Dimensions:L×W: Pease specify H: (≥60 cm) ,(+.-)10cm4 Castors(size ≥12.5 cm) with 2 breaks.Patient weight: (≥175kg).Drain pan: yes.Number of motors: (≥2).Bumpers: yes.Input power: 220/240 VAC, 50/60 Hz single phase.Movements:Adjustable backrest.Adjustable leg section.Adjustable Trendelenburg & reverse TrendelenburgAdjustable height.Adjustable foot section.Standard accessoriesLeg rest (knee crutches).Foot rest.Side rails.Lifting pole.IV pole (2 or 4 hooks).Labour / squatting bar.Monitor tray.Arm rest.Battery backup.Optional accessories:Removable headboardEnvironmental requirements: the equipment

32

Page 33: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

suitable for work in the climate conditions in Iraq in terms of temperature & humidity.

6

Phlebotomy Chair

4

Functions movement: up/down, back section turn, and leg section turn.castors :NOMaterial: high quality (strength).Easy to clean seamless cushions, Antimicrobial fabricAllows right or left-sided proceduresChair dimension : To be mentioned Arm board swing out : 0~≥ 45oArm board height adjustment : ≥ 70mmArm board slide ≥80 mmArm board dimension : To be mentioned Capacity(loading):≥140 kg

7

IV Pole

11

The minimum features of the intravenous pole shall be:Material of IV pole: stainless steel (grade: 304)Number of IV hooks: 4Minimum payload capacity of each hook: 2 kgMaterial of hooks: stainless steel (grade: 304)Height: shall be with adjustable rod height (approx.. 120 to 200 cm)IV pole base: with 5 anti-static swivelling castorsIV pole base design: the base shall be with low center of gravity to ensure stability

8

Wheelchair

5

Aluminium footrestFolding manual wheelchair.Varnished steel. Max load at least 120 Kg.Manual wheelchair in varnished steel with removable rear wheels quick-release 12" PU.Removable footrests.Leather upholstery and flip-open armrests. Wheels for narrow passages. Head rest with stands

9

Stretcher Stainless Steel

4

Pneumatic back raiseHi-lo adjustment by torque release handleAluminium or steel frameVinyl leather mattressRemovable stretcher aluminium plate platformOxygen cylinder holderFold-down safety side railsIV PoleDimensions : 550W x 1905L x 666~1016mm H150mm caster with total locking system

10 Dressing Trolley 28

33

Page 34: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

Transport, multi-functionTwo detachable 2 pressed trays with dimensions of minimum 600 x 400 x 60 mm. ≥ Two drawers. Handle to facilitate movements. ≥ ø80 mm casters, with bumpers (two with brakes). Slide removable basket Made of stainless steel Minimum Dimensions:800 x 420 x 830 mm

LOT 6 – MONITORING

Item No UNOPS minimum technical requirements Qty.

Is quotation compliant? Bidder

to complete

Details of goods offered. Bidder to complete

1 ECG 21MEDICAL APPLICATION

Detect the electrical signals associated with cardiac activity and produce an ECG, a graphic record of the voltage versus time to diagnose and assist in treating some types of heart disease and arrhythmias.

PATIENT TYPE Adult, paediatricType Open System CONFIGURATIONCART

PortableYes with ECG cable holder

LEADSSwitchingSensitivity, mm/mV

Calibration signalFrequency range, Hz

Diagnostic

FilteredCMRR @ 60 Hz, dBLeads-off indicator

 Automatic, optional manual ≥ 3 values ex. (5, 10, 20)Automatic, optional manual Range ≥ (0.05-150)HzYes ≥ 100Yes

RECORDERNo. of channelsRecordingmethod

Printer

Paper size

Lead marker

Timing marker

12 leadsThermal array

Integrated

According to manufacturer Yes

Yes

34

Page 35: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

Event marker

Chart speed, mm/sec

Channels acquired simultaneouslyChannels printed simultaneously

According to manufacturer

Minimum ≤ 25, Maximum ≥ 50(steps should specified by the manufacturer)

≥ 12≥ 12

NO. WAVEFORMS DISPLAYED

≥ 12

NO. WAVEFORMS STORED

Specified by the manufacturer

ECG TRANSMISSION TO ECG DATA MANAGEMENT SYSTEM

Yes

PROGRAMS Adult resting ECG, other specified by the manufacturer

USER INTERFACE Keyboard or Touchscreen

STORAGENormal, no. ECGs

 

≥ 50PATIENT DATA Name, sex, age,

others (specified by the manufacturer)

MEASUREMENTS Arrhythmia ID

Other printouts

Yes HR, PR,QT and others YesOptional (Specified by the manufacturer)

4.8.1 ST segment analysis

Yes

INTERPRETATIONAnalyzed waveformAnalysis window (sec)

YesSpecified by the manufacturer≥10

PREVIEW SCREEN YesDISPLAY TYPE ≥ 7 in. LCD or TFT

or betterVISUAL INDICATORS

Specified by the manufacturer

DEFIBRILLATOR OVERLOAD PROTECTION

Yes

Reusable vElectrodes for startups

At least 5 sets

BATTERY OPERATION

According to manufacturer (device should be

35

Page 36: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

Battery type (No.) able to work on direct AC line)Specified by the manufacturer

Power supply Line voltage

Medical Approved power supply board - 220/240 VAC , 50/60 Hz single phase, Three pin G type plug

2 Defibrillator- Monitor With AED 4Acuity Area OR. , ICU. PICU ,

CCU , wardsEnergy selection

-External: (≤ 2 -≥200)J BiphasicPaediatric / Neonatal: (2-20) J.-Internal: MAX. 50 J

Paddles (adult + paediatric)

ExternalInternal

Paddle controls Charge, discharge, Others to be specified

Synchronizer Yes AED mode Yes with disposable

pads start up qty for at least 100 patients

CPR measurement and feedback

ECG MONITOR - Color LCD or better- Screen, (in): ≥ 6- Sweep speed, mm/sec: ≤ 50 selectable.- Lead configuration: ≥ 3. Through-the-paddles monitoring- HR display- HR alarms- Lead-fault indicator

Reusable ElectrodesDisposable electrodes

Adult and paediatric , start-up Qty (should be specified)start-up Qty for at least 200 patient

Self-test YesExternal pacemaker - Pacing mode:

demand and fixed.- Pacing rate (ppm): ≤(40-210)

ECG RecorderAnnotation

- Thermal printer.Time, date, lead, gain, heart rate, operating mode.

36

Page 37: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

Other Monitored Parameters

- SpO2- NIBP

Cart YesPower 220-240VAC,

50Hz,G type Plug.Battery PowerOperating time, hrFull Recharge time, hr.

Rechargeable≥ 2 continuous ECG monitoring or ≥ 100 discharges.≤ 4>

protection Against harmful ingress of water:IP24, IP21Against electrical shock: Type Class Iequipment

Environmental requirements:

The equipment suitable for work in the climate conditions in Iraq in terms of temperature & humidity.

Acuity Area OR. , ICU. PICU , CCU , wards

Energy selection

-External: (≤ 2 -≥200)J BiphasicPaediatric / Neonatal: (2-20) J.-Internal: MAX. 50 J

LOT 7 – VENT Equ.

Item No UNOPS minimum technical requirements Qty.

Is quotation compliant? Bidder

to complete

Details of goods offered. Bidder to complete

1 Anaesthesia Machine 3

MEDICAL APPLICATION

Dispense a mixture of gases and vapors and vary the proportions to control a patient's level of consciousness and/or analgesia during surgical procedure.

PATIENT TYPE Adult, pediatric

Pipeline gas inlet

3 (N2O, O2, air), including hoses with probes to outlets, probe to be compatible with the type of outlets in the hospital

Gas cylinder yokes (O2, N2O, air),

37

Page 38: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

including hoses to connect to backup cylinders with regulators

Vaporizers agentsNumberInterlock Agent level indicator

Isoflurane, sevoflurane≥ 2Yes Yes

O2-fail safe Yes Hypoxic mixture fail-safe Yes

Automatic ventilatorBellows size or otherVentilation modes

Tidal volume (cc)

Minute volume (L/min)Frequency (Bpm)Inspiratory flow (L/min)Pressure limit (cm H2O)PEEP (cm H2O)

Yes -Adult, pediatric-Manual / spontaneous, VCV, PCV, SIMV or pressure support & pressure control volume guaranted-Minimum (≤20) and for neonatal option (≤5 by pressure control or volume control), Maximum (≥1400)-≤ (0-100)

-≤ (2-100)-≤ (0-200)

-≤ 125 (adjustable)

-≤ (0-50)System checks Yes Scavenging system Passive

Breathing circuit

Startup kit with CO2 absorption canister (soda lime or barium lime)

Monitors Air way pressureWhere measured High pressure alarmSubatmospheric pressure alarmContinuing pressure alarmLow pressure/apnea Expiratory volume/flowType of sensorWhere measured Apnea alarmReverse-flow alarmHigh/low minute volumeHigh/low flowCapnograph O2 concentration

Yes Inspiratory limb or expiratory limbYesYes

Yes

Yes Yes According to manufacture Expiratory and/or inspiratory limbYes Yes Yes Yes Yes Yes

38

Page 39: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

Type of sensorResponse time (sec)CO2 concentrationApnea alarm

According to manufacture≤ 35Yes Yes

N2O Agent monitorsType of agentAuto IDAgent concentration alarm

Yes Yes Isoflurane, sevofluraneYesYes

Display NOType size

- ≥1- color LCD or better≥12ʺ

User interface According to manufacture

Event logTrack alarmsTime and date stamped

Yes Yes Yes

Accessories

-Reusable patient circuits for adult, pediatric and neonates-Reusable anaesthesia masks for adult, pediatric and neonates -Reusable re-breathing bags for adult, pediatric and neonates -Reusable flow and oxygen sensor-Disposable microbial filters with heat & moisture exchangers (HME) for adult, pediatric and neonates-Test lungs -Soda Lime-Articulating patient circuit holder/arm

Power supply Input power

Back up battery

220/240VAC,50/60Hz, Three pin (G type) plug, voltage stabilizing and over current protection circuit

Yes, ≥ 0.5 (hr)2 CPAP , Neonate 5

Application - Deliver air or a mixture of air and oxygen (O2)

39

Page 40: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

through a nasal mask or nasal tubing to the patient’s airways- Non – invasive

MODES CPAP &Optional Bi-level CPAP (other modes should be specified)

MASKConfigurationMaterialNo.Size

Nasal prongs , nasal mask (neonatal size)Any except (PVC), latex freeStart-up kit (Qty: should be specified )neonate

CONTROLSInspiratoryPressure, cm H2OExpiratory

Pressure, cm H2ORate, bpmTrigger sensitivity

Min ≤ 6MAX. ≤40

according to operating modes.

MIN ≤ 4 MAX≤ 25≥ (1-30) optional according to operation modeAutomatic or adjustable

MONITORED PARAMETERSRespiratory rateFlow rangeLeakPEEP / or PIPOther

Optional according to available modesYesYesYesShould be specified

PC COMPATIBLE YesSoftware Yes

EQUIPMENT ALARMS:

Power failureFaultMask alarm or pressureLow battery

Preferred

YesYes Yes

DISPLAY TYPE LED or LCD or better

NOISE LEVEL, dBA ≤ 45Air/O2 supply YESHumidifier YESCompressor or turbine

YES

Oxygen analyzer YESAIR FILTER Yes (changing filter

40

Page 41: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

the responsibility of the supplier during the warranty and maintenance period).

TUBING Length, cm (in)Material

Yes ( start-up Qty should be specified )≥ 180Any except (PVC)

ACCESSORIES Complete manufacturer line of accessories (Should be specified )Bonnets for single-use ( start-up qty should be specified )

BATTERYRechargeableOperating time, hrCharger

Yes≥ 1 hrYes

Power supply - Medical approved power supply board- 220/240 VAC, 50/60 Hz, Three pin G type plug

Environmental requirements

Manufacturer confirms that the equipment suitable for work in the climate conditions in Iraq in terms of temperature & humidity.

3 Nebulizer 20Method of Nebulization 

Ultrasonic Vibration

Ultrasonic frequency 1.7MHz+-10%Maximum pressure  200kPa

Frequency ≥ 2.5MHZ Oscillation rate

Continuous working time ≥ 4 hours

Operating Humidity & Temperature Range 

10°C- 40°C | 30%-95%RH

Storage Humidity & Temperature Range 

0°C - 70°C | 30%-95%RH

Particle Size diameter

0.5-5.5 (MMAD-3.8 Microns)

Max nebulizing ≥ 4ml/minDroplet diameter 1~5 μm

Capacity of cups Small cup ≥ 100mlLarge cup ≥350 ml

Mechanical timing Step less adjustable 0-60min

Water tank ≥ 300 mL observed by water level line

Accessories portable battery pack, pediatric

41

Page 42: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

mask, adult mask, AC adapter, nozzle, air filter and mouthpiece, closed water chamber system, timer and air flow adjuster, 9-12 VDC DC adapter, carrying bag

Power supply - 200/240 VAC 50/60 Hz , 9-12VDC- Typical solid state, over – current circuit protection

Battery Rechargeable ≥ 3 hours backup battery

Other specification Silent operation(totally noiseless), compact design and one hand use makesTreatment times are 50% less than a motor driven nebulizerSafe – Low battery indicatorOverheat protectionDistilled water or drinking water can be used

4

O2 Regulator, Low Pressure

10

Application:For use on MEDICAL oxygen cylinders

Have a 20-micron, sintered-bronze inlet filter.All flows are tested to accuracy within +/- 15 percent tolerances.Brass components in high-pressure areas and relief valve zones.Standard Liter Flow Options: 0.5, 1, 2, 3, 4, 5, 6, 8, 10, 12, and 15Inlet Pressure: approx. (200-3000) PSI

1. General requirements:1. All medical equipment shall be 100% new and from current manufacturer production lines. It must NOT

be refurbished. It must NOT be discontinued at the time of signing the Contract.

2. All medical equipment must be CE, FDA, and/or TUV certified (or equivalent). Quality assurance test certificates shall be submitted to UNOPS, whenever requested.

2. Full functionality

42

Page 43: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

1. The offered medical equipment shall include a full set of compatible parts, components, accessories, software, hardware, start-up consumables and howsoever required to put the medical equipment into fully operational condition to the satisfaction of UNOPS.

2. Any parts, components, adapters, software, hardware etc, that are NOT mentioned in the technical requirements or technical offer, but required to put the equipment into fully operational condition shall be deemed part of the awarded medical equipment and must be provided by the Supplier at no additional cost.

3. All electrically operated equipment must be designed to run on the standard AC power (voltages and frequencies) in Iraq. Equipment operated at 110V (with and without transformers) shall NOT be acceptable.

3. Pre-installation requirements:1. All medical equipment, or its component, that will be permanently built-in or mounted on floor, wall or

ceiling shall include all needed fixtures, supports, arms, mounting parts/interfaces, finishing seals, and howsoever required to mount the unit and complete the installation.

2. All ceiling suspended equipment must include the design, supply and installation of the interface steel structure between the false-ceiling and concrete ceiling.

4. Packaging, Shipping, Storage and Delivery:1. The equipment package shall be well labeled, with instructions for handling, lifting, etc. The equipment

package shall be labeled with the Supplier name, Manufacturer, model number, and date of manufacture.

2. The equipment shall be packaged in a way to withstand handling, loading, unloading, temperature, humidity and other extremes likely to be encountered during shipping and transport to Iraq.

3. The Supplier shall be responsible for shipping and delivery of the equipment within the time frame stipulated in the Invitation to Bid.

4. The supplier is responsible for delivering the medical equipment to the specific locations as mentioned in the Invitation To Bid. The Supplier is responsible for loading, unloading, rigging and inside delivery of the medical equipment to its final destination room inside the building.

5. Installation 1. The Supplier shall assemble, mount, install, configure, calibrate, test and commission the medical

equipment as per published manufacturer’s instructions, applicable international & local standards, to the satisfaction of UNOPS.

2. Any materials or work found to be defective, not in conformity to the requirements herein, or damaged shall be removed immediately, and new materials or work substituted without delay.

6. Installation inspection1. Installation Inspection shall demonstrate proper and safe installation and operation of the medical

equipment as per the published manufacturer’s specifications, applicable standards, and to the satisfaction of UNOPS. Whenever deemed necessary by UNOPS, the Supplier shall provide calibrated testing equipment, analyzers etc to verify proper function/performance of the equipment as per the published manufacturer’s specifications.

43

Page 44: RFQ for goods/services€¦ · Web viewYes, ≥ 0.5 (hr) 2 CPAP , Neonate 5 Application - Deliver air or a mixture of air and oxygen (O2) through a nasal mask or nasal tubing to the

UNOPS eSourcing v2017.1

2. The Supplier shall provide 2 original user manuals and 2 original technical service manuals. Soft copy of user and service manuals (on CD) shall be also provided. Technical service manuals shall include spare parts lists, electronic circuits schematic diagrams, and detailed troubleshooting guides (where applicable).

7. Warranty and Maintenance:1. The Supplier shall provide full warranty, for the period stipulated in the Invitation to Bid. The warranty

shall cover free maintenance/labor and free spare parts throughout the warranty period.

2. All maintenance and services must be provided by qualified and certified staff.

3. The Supplier shall conduct scheduled Preventive Maintenance (PM) according to the manufacturer recommendations and applicable standards.

4. The Supplier shall fix malfunctioning equipment within a maximum of 4 days from the date of notification.

5. The Supplier shall replace or repair all defective equipment and software, and shall correct any defects -without charges for parts or labor- both during and after regular working hours, during the warranty period.

6. The Supplier shall provide -without charges for parts or labor- an equivalent replacement of any defective part that cannot be repaired or corrected to the satisfaction of UNOPS/MOH during the warranty period.

7. The Supplier must guarantee an uptime of 95% calculated yearly during warranty, unless otherwise requested by MOH to follow current protocols applied by MOH.

8. All warranties and rights shall be transferrable from UNOPS to the Ministry Of Health (MOH) upon transfer of ownership, the date of which shall be agreed between UNOPS and the MOH.

Delivery requirements –– Comparative Data Table

UNOPS RequirementsIs quotation

compliant? Bidder to complete

Details Bidder to complete

Delivery timeBidder shall deliver the goods 4 - 6 months after Contract signature.

☐ Yes ☐ No Insert details

Delivery place and Incoterms rules

DAP Incoterms rules as per Incoterms 2020.as per Annex I distribution plan

☐ Yes ☐ No Insert details

Consignee details TBC ☐ Yes ☐ No Insert details

UNOPS Right to vary requirements

At the time the Contract is awarded, UNOPS reserves the right to vary the quantity of the goods and associated services specified above, provided this does not exceed +/- 35%, without any change in the unit prices or other terms and conditions of the RFQ.

☐ Yes ☐ No Insert details