Upload
buituyen
View
216
Download
3
Embed Size (px)
Citation preview
Review of Literature
5
Review of Literature
2.1. Introduction
Diabetes mellitus is a chronic metabolic disorder characterized by rise in blood glucose
level called “hyperglycaemia” (Vinik and Flemmer, 2001). Monitoring the blood glucose
level is required for diabetic patients, to maintain it within a normal range, and thereby
reducing the risk of severe complications such as kidney failure, blindness and nerve
damage (Genuth et al., 1998). The worldwide prevalence of diabetes in 2000 was
approximately 2.8% and is estimated to grow by 4.4% by the end of 2030. This translates
to a projected rise of diabetes from 171 million in 2000 to well over 350 million in 2030
(Sultanpur et al., 2010). Hence, there is an urgent need for improved treatment of
hyperglycaemia and other risk factors associated with this metabolic syndrome. If left
untreated or improperly managed, diabetes can result in a variety of complications,
including heart disease, kidney disease, eye disease, neuropathy, gangrene, gastroparesis,
blindness, erectile dysfunction, impotence and nerve damage. Poor blood glucose control
also increases the risk of short-term complications of surgery such as poor wound
healing. Historically, measurement of glucose levels has been the method, universally
used to diagnose diabetes. Laboratory methods such as fasting plasma glucose (FPG) or
2-h plasma glucose (2HPG) level have been used for this purpose. However, this
approach still suffers from the same problems and difficulties associated with glucose
biosensors such as the need for fasting, biological variability and the effects of acute
perturbations (e.g., stress- or illness related) on glucose levels (Sheikholeslam et al.,
2011). Also, blood glucose reflects a fluctuating glucose level, and it is affected by daily
diet and requires frequent measurements. Currently, portable-sized glucose measurement
devices are being designed for self use, low costs, and quick results. Some of them adapt
invasive measurement methods for blood extraction, i.e., use a piercing component of a
microneedle or a lancet with a disposable strip for sensing and a handheld fixture for
readout (Yang et al., 1998; Wang and Zhang, 2001; Mukerjee et al., 2004).
The glycated hemoglobin (HbA1C) level, defined as the ratio between HbA1C
Review of Literature
6
concentration and total Hb concentration, is considered to be a very useful diagnostic
marker for diabetic patient in addition to the measurement of the glucose level (John,
2003). Also, the use of this approach solves many of the problems associated with FPG
or 2HPG methods based on glucose measurements such as no need for fasting,
substantially less biological variability and relative insensitivity of HbA1C levels to acute
perturbations (Sheikholeslam et al., 2011). Compared with blood glucose, HbA1C level is
a more stable diagnostic index and is recommended as a baseline test for diabetes. HbA1C
is the stable glucose adduct to the N-terminal group of the β-chain of HbA0. Since it
reflects the average blood glucose level over the preceding 2-3 months and is not affected
by the daily fluctuation of the glucose level, HbA1C level provides a more accurate index
for diagnosis and long term control of the disease. HbA1C level correlates linearly with
mean blood glucose (MBG). 1% change of HbA1C level reflects a fluctuation of MBG
concentration by about 2 mM (MBG (mmol/L) = (1.98.HbA1C)-4.29). The clinical
reference range of HbA1C level is 5-20%, with 4.0-5.7% considered as normal. Since the
lifetime of Hb in blood is approximately 2-3 months, the HbA1C level provides a good
indication of the glucose level over this period of time. Consequently, the measurement
of HbA1C level is important for the long-term control of the glycemic state in the diabetic
patients (Jeppsson et al., 2002).
Normal adult Hb consists primarily of Hbs A (90-95%), A2 (2-3%), F (0.5%), Ala
(1.6%), Alb (0.8%) and A1c (3-6%). HbA1C are the minor Hb molecules separable by
chromatographic techniques into three major components: Ala, Alb, and A1c.
Hemoglobin Al refers to a combination of these three components (Bunn et al., 1979).
Diabetes Control and Complication Trial (DCCT), has demonstrated that 10% stable
reduction in HbA1C determines a 35% risk reduction for retinopathy, a 25-44% risk
reduction for nephropathy and a 30% risk reduction for neuropathy (Calisti and Tognetti,
2005).
Underlying principle
In the normal 120-day lifespan of the red blood cell (RBC), glucose molecules react with
Hb, forming HbA1C. In individuals with poorly controlled diabetes, the quantity of
Review of Literature
7
HbA1C is much higher than in healthy people.
Once a Hb molecule is glycated, it remains that way. A buildup of HbA1C within
the RBC, therefore, reflects the average level of glucose to which the cell has been
exposed during its life-cycle. Measuring HbA1C assesses the effectiveness of therapy by
monitoring long-term serum glucose regulation. The HbA1C level is proportional to
average blood glucose concentration over the previous four weeks to three months. Some
researchers state that the major proportion of its value is related to a rather shorter period
of two to four weeks (HbA1C fact sheet, http://www.med.umich.edu/mdrtc/cores/
ChemCore/hemoa1c.htm.).
The 2010 American Diabetes Association Standards of Medical Care in Diabetes
added the A1c ≥ 48 mmol/l (≥ 6.5%) as another criterion for the diagnosis of diabetes
(Executive Summary, 2010).
Glycation is a non-enzymatic reaction between free aldehyde group of glucose
and free amino groups of proteins (here β chain of Hb). A labile aldiminic adduct (Schiff
base) forms at first, then, through a molecular rearrangement, a stable ketoaminic product
slowly accumulates (Fig. 1). In Hb, the preferential glycation site is the amino-terminal
valine of the β chain of the globin (about 60% of glycosylated globin). Other sites are:
lysine 66 and 17 of the β chain, valine 1 of the α chain. The term HbA1C refers to the Hb
fraction of the glucose bound stably (ketoamine) to beta terminal valines.
2.2. Fructosyl amino acid oxidase
Fructosyl amino acid oxidases (FAOs) (fructosyl amine: oxygen oxidoreductases or
fructosylamine: oxygen oxidoreductases, EC 1.5.3.X) are a group of enzymes that
catalyses the oxidative deglycation of fructosyl amino-acids, which are formed by non-
enzymatic glycation reaction, in which biological amines (i.e., free amino acids, amino
acid residues of proteins, and amino groups of nucleic acids) react with reducing sugars
at an early stage of the Maillard reaction and are the model compounds for glycated
proteins, yielding corresponding amino acids, glucosone and H2O2 (Horiuchi et al., 1989;
Horiuchi and Kurokawa, 1990, Sakai et al., 1995; Yoshida et al., 1995; Takahashi et al.,
Review of Literature
8
Fig. 1 Chemical reactions involved in glycation of hemoglobin
1997a). The resultant Amadori compounds repeat dehydration and condensation to form
stable compound called advanced glycation end products (AGEs) (Ulrich and Cerami,
2001). Glycation affects the function of the proteins in vivo, and protein cross-links with
advanced glycation end product cause the development of diabetic complication and
aging (Schmidt et al., 1994; Singh et al., 2001; Goldin et al., 2006). Thus, it is important
to investigate the intervention against the Maillard reaction. Most FAOs show this type of
oxidation and require FAD as cofactor, but the enzyme found in Pseudomonas sp. has a
somewhat different reaction mechanism, utilizing Cu2+
(Saxena et al., 1996). This
enzyme produces fructosamine residue from fructosyl-amino acid, indicating that C-N
bonds suffered from the cleavage are different in these two types of enzymes. The first
FAOs was isolated by Horiuchi et al. (1989) from Corynebacterium sp. 2-4-1.
2.2.1. Sources of FAO
2.2.1.1. Fungi
The presence of fungal FAOs has been reported most extensively, in strains of various
genera Achaetomiella, Achaetomium, Acremonium, Aspergillus, Chaetomium,
Coniochaeta, Eupenicillium, Fusarium, Gelasinospora, Gibberella, Microascus,
Fructosyl valine
or ketomine
Aldimine
(labile)
Glucose
β-val-NH2 +
H O
C
H-C-OH
HO-C-H
H-C-OH
H-C-OH
H-C-OH
H
CH(CH3)2
CH2NHCH
C=O
HO-C-H
H-C-OH
H-C-OH
CH2OH
CO2H
H N-Val-β
C
H-C-OH
HO-C-H
H-C-OH
H-C-OH
H-C-OH
H
Review of Literature
9
Penicillium, Thielavia and Ulocladium (Staniford et al., 1993; Yoshida et al., 1995;
Takahashi et al., 1997a, b; Sakai et al., 1999, Fujiwara et al., 2006; Hirokawa et al.,
2003). Akazawa et al., 2004 studied the application of fungal FAOs in the clinical
diagnosis of diabetes mellitus (Yoshida et al., 1996), since the amounts of glycated
proteins such as Hb and albumin in blood reflect the level of blood glucose, and
fructosyl-amino acids are model compounds for the glycated proteins. They also found
that FAO is a peroxisomal enzyme and widely distributed in filamentous fungi. Thus, it
was concluded that FAO plays an important role in fungal cells. However, the definite
physiological role of the enzyme is still unknown.
2.2.1.2. Bacteria
FAO has also been found in the bacterial genera Arthrobacter, Corynebacterium,
Klebsiella, and Pseudomonas (Ferri et al., 2004; Ferri et al., 2005; Gerhardinger et al.,
1994; Saxena et al., 1996).
2.2.1.3. Yeast
FAO occurred in Yeast genera Debaryomyces and Pichia (Sode et al., 2001a; Staniford
et al., 1993). There is no report for the existence of FAO in mammalian organisms.
2.3. Applications of FAOs
FAOs have been explored for use in diabetes diagnosis, detergents and food processing.
The application for examination of diabetes was first realized around the time, enzyme
was discovered (Tatsuo and Yoshiko, 1987), and now FAO-based assays have become a
“gold standard” diagnostic tool (Hirsch and Brownlee, 2005) for determining the levels
of HbA1C, an important indicator for diabetes patients (Kobold et al., 1997; Hirokawa et
al., 2004; Nakamura et al., 2007). In this scheme, FAO reacts with fructosyl-valine (FV),
which is released from the N terminus of HbA1C upon protease treatment. The resulting
H2O2 is measured by peroxidase reaction. Several research groups have also developed
electrode-based biosensor systems which are capable of measuring FV, based on the
FAO from Pichia sp. N1-1 (Fang et al., 2009; Ogawa et al., 2002; Tsugawa et al., 2000;
Tsugawa et al., 2001) or Arthrobacter sp. FV1-1 (Sakaguchi et al., 2003).
Review of Literature
10
Detergent additive is a second promising area of application for FAOs, as
evidenced by a recently filed patent application (O'Connell et al., 2008), which relies on
the ability of FAOs to enzymatically degrade amadori products and also release in situ
hydrogen peroxide, which in itself is a bleaching agent. A preliminary performance test
of an engineered amadoriase II on American gravy stains has already been demonstrated
(Zheng et al., 2009).
While the Maillard reaction also exerts an important influence on the food
properties, the commercial use of FAOs in the food processing and quality control
remains to be seen (Kato et al., 1998; Ishida et al., 2002; Yagi et al., 1997). Lastly, as
amadori compounds and advanced glycation end products (AGEs) have been implicated
in aging and several diseases, the therapeutic use of this class of enzymes should be
worth exploring. For all these applications, FAOs that can react directly with lager
glycated proteins instead of smaller glycated amino acids or short peptides would be of
greater value, and therefore it is interesting to engineer and expand the limited
accessibility of the substrate binding sites of FAOs.
2.4. Various methods for HbA1C determination
A number of chromatographic methods are currently available for the determination of
HbA1C level such as electrophoresis/electroendosmosis (Hageman and Kuehn, 1977), ion
exchange chromatography (Goldstein et al., 1986), high performance liquid
chromatography (HPLC) (Brunnekreeft and Eidhof, 1993; Ellis et al., 1984; Turpeinen et
al., 1995), boronate affinity chromatography (John, 1997), immunoassay (Turner et al.,
1999) and liquid chromatography associated in tandem with mass spectroscopy (LC-
MS/MS) (Jeppsson et al., 2002) prior to fluorometric (Gallop et al., 1981) and
colorimetric (Fluckiger and Winterhalter, 1976) quantification of HbA1C.
2.4.1. Capillary electrophoresis
Principle: Basically, two possibilities exist for separation of HbA1C in capillary
electrophoresis (CE) according to charge-to-mass ratio. Firstly analysis as cations in
acidic buffers of pH below pI of Hb, which is approximately 7.0. Separation of
Review of Literature
11
hemoglobins A1C and A0 occurs due to a charge difference coming from elimination of
one positively charged amino group in the HbA1C molecule by attachment of glucose
moiety. Secondly, Hb analysis as anions in alkaline conditions with selectivity to HbA1C
induced by a cis-diol interaction of its glucose unit with a borate anion from background
electrolyte (BGE) (Fig. 2).
CE includes method for separating Hb derivatives and Hb variants making use of
a dynamic coating technique that allowed rapid separation of Hb variants and derivatives.
CE was performed on a Beckman P/ACE System 5000. Before sample injection, the
capillary is first rinsed with initiator solution (containing a polycation, albumin, pH 4.5),
followed by buffer solution containing a polyanion (chondroitin sulfate, pH 4.5), at the
same pressure. Sample is injected, followed by an injection with buffer solution to rinse
the outside of the capillary. The capillary is then transferred to another vial containing
buffer solution, in which the electrophoresis is performed. Negatively charged molecules
(chondroitin sulfate, pH 4.5) in the buffer solution bind to Hb. Electrophoresis is
performed with the negative electrode at the detector site. Detection is executed with a
UV/VIS absorbance detector at 415 nm. Peak integration for peak area measurement is
performed by a Beckman System Gold chromatography data system. Stability and
coefficient of variance of the method were found to be 20 days and 1.4 to 3.7%. No
interferences were observed from carbamylated or acetylated Hb or from the labile
HbA1C fraction. A good linear correlation coefficient with r=0.98 was found with Bio-
Rex 70 HPLC (Doelman et al., 1997).
A CE method was also developed for measurement of HbA1C by exploring the
potential of cis-diol interactions for separation of HbA1C (Koval et al., 2011). First, a CE
analysis was performed with a sodium tetraborate buffer (pH 9.3) as background
electrolyte in a neutrally coated capillary (Shen and Smith, 2000). HbA1C was separated
from HbA0 due to specific interactions of borate anions with the cis-diol pattern in the
saccharide moiety of glycohemoglobin.
Merit: Easy experimental setup.
Review of Literature
12
Demerit: Not much selective because of some interfering species among the multitude of
known Hb variants.
Fig. 2 Schematic representation of determination of HbA1C by capillary electrophoresis
2.4.2. HPLC method
Principle: HPLC typically utilizes different types of stationary phases, a pump that
moves the mobile phase(s) and analyte through the column, and a detector to provide a
characteristic retention time for the analyte. For HbA1C, automated HPLC employs a
weak cation exchange column. The support material is silica with carboxymethyl
functional groups. The analyzer forms a stepwise gradient composed of three phosphate
buffers of increasing ionic strength. The system is optimized for quantitation of HbA1C,
while HbA2 coelutes with HbA2. Quantitation is based on comparison of the absorbance
values at 415 and 690 nm. Before analysis, the samples are incubated at 37oC for 30 min.,
to remove the labile fraction (Fig. 3).
Review of Literature
13
International Federation of Clinical Chemistry and Laboratory Medicine (IFCC)
Working Group developed HPLC, as a standard reference method for HbA1C. Different
values for HbA1C can be obtained when the same blood samples are measured,
depending on the chromatographic system, e.g. the kind of resin, lotto-lot variation of
resins, column size, buffer composition and elution times. The peak considered to be
HbA1C may contain variable proportion of substances which have the same elution
behaviour as HbA1C but are not HbA1C according to definition, since they lack
immunological activity or do not bind during affinity chromatography.
In an automated HPLC assay 45-min. in acetate buffer (pH 5.5) is required to
eliminate labile HbA1C. The chromatographic column contained “polyCAT” (a weak
cation-exchanger, polyaspartic acid linked to silica). Run time was 14 min per sample
(Ellis et al., 1984).
HbA1C and total Hb were simultaneous determined by isotope dilution and
HPLC-inductively coupled plasma mass spectrometer (ICP-MS). HbA1C was separated
by cation exchange chromatography, employing a Mono S column. Then, the detection
was carried out by measuring the Fe contained in the heme-group of the protein, using an
ICP-MS as the Fe selective detector. Thus, speciation of glycated and non-glycated
haemoglobin was investigated with detection limits around 1 mg mL-1
of protein (Busto
et al., 2008).
Merits: It provides good precision and long-term stability and results correlated very
well with by other ion-exchange method. Also, it gas high resolution, sensitivity,
reproducibility, accuracy and automation.
Demerits: Historically HbA1C is defined as a certain peak in an HPLC system, but this is
no longer scientifically acceptable, as it lacks specificity. Different values for HbA1C in
the same blood were obtained depending on the conditions used, because the HbA1C
peaks contain different kinds and amounts of substances that are not HbA1C. Also, HPLC
require special instrumentation and training and result in patterns that are relatively
complex.
Review of Literature
14
Fig. 3 Schematic representation of determination of HbA1C by HPLC
2.4.3. Capillary isoelectric focusing
Principle: Capillary isoelectric focusing (CIEF) separates the analytes according to their
pI differences. The pI of HbA1C is slightly shifted to the acidic direction with respect to
HbA0 due to one amino group blocked by the glucose moiety. This subtle pI difference
underscores the need for a very well-tuned CIEF method to separate both species.
CIEF methodology for the analysis of HbA1C, in dimethylpolysiloxane coated
fused-silica capillaries was carried out, using a narrow pH ampholyte mixture. In the
focusing procedure, a 0.100 mol L-1
phosphoric acid solution was used as analyte and a
0.040 mol L-1
NaOH solution was used as catholyte. Chemical mobilization was used
which allowed the complete baseline resolution of the HbA1C in less than 8 min.
Correlation coefficient was found to be 0.872 (n=31) with standard cation-exchanger
columns (Jager and Tavares, 2003).
Another CIEF method, exploited a difference in pI values of HbA0 and HbA1C,
performed with a range of carrier ampholytes commercially available like Servalyt pH 6-
Review of Literature
15
8, Biolyte pH 6-8 carrier ampholytes spiked with a narrow pH cut of pH 7.2 prepared by
preparative fractionation of Servalyt pH 4-9 carrier ampholytes (Koval et al., 2011). They
adopted optimization of experimental parameters in CIEF by Beckman Coulter (Mack et
al., 2009). The relative standard deviation (RSD) values of migration time in the interday
study were initially 4.3-5.7% before dropping to 0.1-0.3% for pI after electropherogram
normalization with the aid of pI markers.
Merits: Outstanding separation power, due to the pH gradient being flattened by a
narrow pH cut of carrier ampholytes. Efficient, expressive, economic (no sophisticated
equipment required), easy (clear, one-dimensional separation principle), fast, sensitive.
Separation in CIEF does not require denaturation of proteins, thus any kind of subsequent
investigations, such as activity staining (e.g. to find separated enzymes) or antibody
detection, is not hindered.
Demerits: More experimentally demanding.
2.4.4. Boronate affinity chromatography
Principle: It is based on use of a “biological interaction” for the separation and analysis
of specific analytes within a sample. For HbA1C, boronate affinity chromatography is a
glycation specific method based on boronate binding to the unique cis-diol configuration
formed by stable glucose attachments to Hb. This method thus measures all four stable
species, altogether. The combined measure of only the four stable species has been
referred to as "Total HbA1C" or by some as "True HbA1C ". Since only two fractions are
present in these methods (glycated and non-glycated), the glycated portion is compared to
the total and results are expressed as % HbA1C (Fig. 4).
The technique was first used for the determination of glycohemoglobin (Mallia et
al., 1981). A low-performance agarose gel was used as the support and absorbance
detection was recorded at 414 nm used to quantify the retained and non-retained Hb
fractions in human hemolysate samples. Elution was performed by passing through the
column a soluble diol-containing agent (i.e., sorbitol) that displaced the retained
glycohemoglobin from the column. After the initial report by Mallia et al., similar low-
Review of Literature
16
performance methods were reported or evaluated by other groups (Gould et al., 1982;
Klenk et al., 1982; Fluckiger et al., 1984; Johnson and Baker, 1988). The same approach
has since been adapted for use in HPAC and HPLC-based systems (Gould et al., 1982;
Hjerten and Li, 1990; Kitagawa and Treat-Clemens, 1991; Singhal and DeSilva, 1992).
Merits: Analytical advantages of this method are low susceptibility to variations in
temperature and pH. It is the only interference-free method for quantifying HbA1C. This
method is specific for all HbA1C species regardless of charge. It provides very good
precision and accuracy.
Demerit: Boronate affinity does not measure A1a1, A1a2, A1b or the labile fraction.
Fig. 4 Chemical reactions involved in determination of HbA1C by affinity
chromatography
2.4.5. Electrospray ionisation mass spectrometry (ESI-MS)
Principle: It is based on formation of polyprotonated or polydeprotonated ions of intact
analyte molecules that are produced from a fine spray of an aqueous solution of the
analyte, assisted by a strong electrical field at atmospheric pressure. The ions have low
internal energy and are not prone to fragmentation. The measurement of mass-to-charge
values of these multiple charged ions gives the inherent molecular mass of the analyte
molecules. The ion currents produced by ESI are dependent on the analyte concentration
(Fig. 5).
Review of Literature
17
A method was developed for quantification of HbA1C, based on the specific N-
terminal residue of the Hb β-chains. Enzymatic cleavage of the intact Hb molecule with
endoproteinase Glu-C was optimized to obtain the β -N-terminal hexapeptides of HbA1C
and HbA0. These peptides were first separated by reversed-phase HPLC and quantitated
by ESI-MS. The stability, reproducibility and repeatability of the total analytical system
were found to be 1 month with coefficient of variance 2.0%, 1.2%, and 2.5% (n=3) for
250 uses. This demonstrates a very good reproducibility of the system. This method
distinguished the smaller N-terminal parts of the β-chains of HbA1C from those of the
HbA0 molecules, thus avoiding the heterogeneity created by modifications of other
glycation sites of the Hb molecule. With these peptides and hyphenated separation
techniques, it was possible to overcome the insufficient resolution of currently used
protein separation systems for HbA1C (Kobold et al., 1997).
Merit: Relatively new mass spectrometric technique suited for the analysis of polar
biomolecules of low and high molecular mass. ESI-MS is ideally suited for an on-line
coupling with HPLC, and the combination of these techniques might become a powerful
instrument for the development of new reference methods for a variety of analytes.
Demerit: High cost of ESI-MS equipment.
Fig. 5 Schematic representation of determination of HbA1C by Electrospray ionization
mass spectrometry
Review of Literature
18
2.4.6. Colorimetric method
Principle: Colorimetric approach of detection of HbA1C is based on conversion of
hexose moiety of HbA1C to 5-hydroxymethylfurfural (HMF) by heating at 100oC in the
presence of a weak acid (oxalic acid). HMF then reacts with 2-thiobarbituric acid (TBA),
and the resulting color is measured colorimetrically.
In the colorimetric method, glucose moiety of HbA1C is converted to 5-
hydroxymethylfurfural (5-HMF) by heating with oxalic acid for 60 min. in an autoclave
at 124 kPa (18 lb/in.2). The adduct thus formed by reaction of TBA with HMF was
measured photometrically and results were expressed either as nanomoles of HMF or as
fructose equivalents. Within-assay and between-assay coefficients of variation were <2%
and <3%, respectively. The method showed a very good correlation coefficient with
r=0.98 (n=50) of the present method and as analyzed by liquid chromatography (Parker et
al., 1981).
In another colorimetric assay for, HbA1C was formed from HbA by the chemical
condensation of a molecule of glucose specifically with the NH2-terminal of the chain of
HbA (Bunn et al., 1975). In this method, the carbohydrate moiety was cleaved by acid
hydrolysis to yield 5-HMF, which is subsequently complexed with TBA. The resulting
chromogen has a maximal absorbance at 443 nm. The normal mean (±SD) for HbA1C
was found to be 5.85% ± 0.79 for males (70) and 5.88% ± 0.79 for females (n=30). A
good correlation coefficient with r=0.92 was found with microcolumn test kit. This
colorimetric method was able to meet many criteria of an ideal laboratory test
(Karunanayake and Chandrasekharan, 1985).
Merits: More rapid and reproducible than other methods and correlates well with
standard liquid-chromatographic method. This method is cheap and not laborious. No
specialized equipments and expertise is required and can even be automated, if required.
Demerits: Most manual methods for quantitating HbA1C based on this approach require
very long incubation period (approx. 3-16 h) for the production of HMF. The yield of 5-
HMF is low and depends critically on the protein concentration and results may vary.
Review of Literature
19
2.4.7. Fluorometry
Principle: The method is based upon periodate oxidation of the carbohydrate moieties
present on both the - and Ԑ -amino groups of globin. The formaldehyde product is
measured as the fluorescent 3,5-diacetyl-1,4-dihydrolutidine formed from the
condensation of formaldehyde with acetylacetone and ammonia.
A sensitive fluorometric method was based on periodate oxidation and by
measurement of the formaldehyde produced, for the measurement of HbA1C. Thus, it was
necessary to remove all periodate sensitive formaldehyde yielding compounds. This was
achieved by preparing globin with acid tetrahydrofuran (THF), as both the hemin and the
acetone usually used for the preparation of globin were found to interfere with the
fluorometric determination of formaldehyde (Gallop et al., 1981). The formaldehyde
liberated was measured spectrofluorometrically with the DDL assay (Nash, 1953). The
linearity of glycoglobin was found to be in the range, 1 to 5 mg. The method was
rigorously designed to assay HbA1C levels and to give a direct measure of the number of
glycogroups per mg of Hb. It requires only 1 mg of protein and may also be used to
determine the extent of the nonenzymatic glycosylation of other proteins. Coefficient of
variance and correlation coefficient were found to be 1.3-3.8% and 0.95 respectively.
2.4.8. Turbidimetric Immunoassay
Principle: The method utilizes the interaction of antigen and antibody to directly
determine HbA1C in whole blood.
A light-scattering competitive immunoassay for HbA1C was developed and
compared to an automated clinical chemistry analyzer (Vucic et al., 1998). The method is
based on a combination of the latex-enhanced competitive turbidimetric immunoassay for
HbA1C and colorimetric determination of total Hb within the same sample. After
hemolysis combined with enzymatic cleavage, HbA1C and total Hb were determined by
immunoassay and colorimetry, respectively, and final result was calculated as a
percentage from the two measured parameters ratio. The enzymatic cleavage led to
release of -N-terminal fragments for reaction with a fixed amount of latex-bound
Review of Literature
20
monoclonal antibodies. Antibody excess was then removed by agglutination with a
synthetic polyvalent complex of -N-terminal fragments. The reaction was monitored by
turbidimetry with light-scattering signal (absorbance at 540 nm) inversely related to the
amount of HbA1C in the sample. Total Hb was measured in the same hemolysate by the
alkaline hematin method (absorbance at 410 nm). The within and between batch
coefficient of variance, correlation coefficient with standard ion-exchange
chromatography (n=117) and stability of assay were found to be <5%, r=0.989 and 2
months respectively.
A turbidimetric immunoassay was developed in which mouse antihuman A1C
monoclonal antibodies were added to whole blood, latex A1C antihuman A1C antibody
complex was formed. Agglutination occured when goat antimouse IgG polyclonal
antibody interacts with the monoclonal antibody. The amount of agglutination measured
as absorbance was proportional to the amount of A1C absorbed onto the surface of latex
particles. Multipoint calibration was used for preparing calibration curves. The method
was compared with standard HPLC method on Biorad variant with r=0.986 (n=15) and
an intraclass correlation value of 0.993 was obtained. Stability of the assay was found to
be 15 days (Lakshmy and Gupta, 2009).
2.4.9. Automated HbA1C immunoassay
In the method, whole blood is hemolyzed using an automatically onboard analyzer. The
results of imprecision studies showed comparable performance of both procedures,
coefficients of variation being slightly higher with the automated procedure (2.2-2.7%
versus 1.7-2.1% in within-run experiments, and 2.4-3.5% versus 2.1-3.0% in between-run
experiments). Comparison of results in 100 fresh samples showed acceptable correlation
between the two procedures (R2=0.94, y=0.98x+0.43). Sedimentation of whole blood in
sample tubes prior to automatic hemolysis did not alter the results (Schneider et al.,
2002).
A heterogeneous affinity-matrix based immunoassay was developed for the
determination of HbA1C. The method is based on an immunoenzymometric assay
(IEMA), where the glycated pentapeptide Val-His-Leu-Thr-Pro (VHLTP), as HbA1C
Review of Literature
21
analogon was immobilized either to the surface of a microtiter plate by adsorption or to
an amino-modified cellulose membrane by covalent linkage. The immobilized analogon
competed together with the HbA1C in the sample for the antigen binding sites of the anti-
HbA1C antibodies. Glucose oxidase-labeled antibodies were used to indicate the antigen
antibody reaction indirectly and enzyme activity was detected optically. A linear range
was obtained between 5-50 % HbA1C (Stöllner et al., 2001).
Demerit: The automated procedures developed for the autoanalyzer require a manual
extraction into a suitable solvent and a preliminary purification of the extract, before it is
introduced into the instrument. Thus, the autoanalyzer procedures are only partly
mechanized.
2.4.10. Enzymatic Assays
The method is based on structural difference of HbA1C and HbA0. Lysed blood samples
are subjected to proteolytic digestion. Glycated valines are released and serve as substrate
for fructosyl valine oxidase. The produced hydrogen peroxide is measured using a
horseradish peroxidase-catalyzed reaction with a chromogen.
2.5. Biosensors
A biosensor is a system of two transducers, biochemical and physical, in intimate contact
or in close proximity with each other that relates the concentration of an analyte to a
measurable signal. It uses specific biochemical reactions mediated by isolated enzymes,
immunosystems, tissues, organelles or whole cells to detect chemical compounds usually
by electrical, thermal or optical signals. Biosensors are used for analysis of bio-material
samples to gain an understanding of their bio-composition, structure and function and
also provide selective identification of toxic chemical compounds at ultratrace levels in
industrial products, chemical substances, environmental samples (e.g., air, soil, and
water) or biological systems (e.g., bacteria, virus, or tissue components) for biomedical
diagnosis.
The history of biosensors started in 1962, with the development of enzyme
electrodes by scientist Leland C. Clark (Clark and Lyons, 1962), who is known as the
Review of Literature
22
father of Biosensors. He carried out this by using an enzyme transducer. Clark’s oxygen
electrode was the enzyme glucose oxidase, entrapped in dialysis membrane. The method
of detection was based on the decrease in oxygen concentration that is proportional to
glucose concentration. Clark's ideas became commercial reality in 1975 with the
successful re-launch (first launch 1973) of the Yellow Springs Instrument Company
(Ohio) glucose analyser based on the amperometric detection of hydrogen peroxide.
Since then, the designs and applications of biosensors in analytical chemistry have
developed rapidly for the last thirty years and communities from various fields such as
very large scale integration (VLSI), physics, chemistry, and material science have come
together to develop more sophisticated, reliable, and mature biosensing devices.
Applications for these devices are in the fields of medicine, agriculture, biotechnology as
well as the military and bioterrorism detection and prevention.
Fig. 6 List of possible biological and sensor elements in biosensor
Review of Literature
23
2.5.1. Basic principle of biosensors
A biosensor is generally defined as an analytical device which converts a biological
response into a quantifiable and processable signal (Lowe, 1984). Fig. 7 shows
schematically the parts comprising a typical biosensor: a) bioreceptors that specifically
bind to the analyte; b) an interface architecture where a specific biological event takes
place and gives rise to a signal picked up by c) the transducer element; the transducer
signal (which could be anything from the in-coupling angle of a laser beam to the current
produced at an electrode) is converted to an electronic signal and amplified by a detector
circuit using the appropriate reference and sent for processing by, e.g., d) computer
software to be converted to a meaningful physical parameter describing the process being
investigated; finally, the resulting quantity has to be presented through e) an interface to
the human operator. Biosensors can be applied to a large variety of samples including
body fluids, food samples, cell cultures and be used to analyze environmental samples.
Fig. 7 Mechanism of action of a biosensor
2.5.2. Classification of biosensor
2.5.2.1. Classification of biosensor based on evolution
A. First generation biosensors are those in which the normal product of the reaction
diffuses to the transducer and causes the electrical response. It works by means of the
direct detection of electroactive species that are enzymatically produced or consumed
Review of Literature
24
(Dicks et al., 1989). A good example is the traditional glucose sensor, which can detect
the hydrogen peroxide produced in the oxidation process catalyzed by a glucose oxidase
enzyme.
B. Second generation biosensors involve specific 'mediators' between the reaction and
the transducer in order to generate improved response. It makes use of an artificial
electron carrier, or mediator, in place of oxygen, which shuttles the electron involved in
the redox process from the active center of the enzyme to the electrode or vice versa.
Both mediator and enzymatic substrate must be in the analytical solution.
C. Third generation biosensors are those in which the reaction itself causes the
response and no product or mediator diffusion is directly involved. Interest about third
generation biosensors grew a few years ago when it was found that direct
electrochemistry with redox proteins is possible without using a co-substrate (Armstrong
et al., 1987). The foremost work in this field used mercury electrodes, onto which
enzyme adsorption was strong and lead to molecule denaturation.
2.5.2.2. Classification of biosensor based on transducer element involved
A. Electrochemical transducers: In this type of sensor, sensing molecules are either
coated onto or covalently bonded to a probe surface. The sensing molecules react
specifically with analyte, sparking an electrical signal proportional to its concentration.
The underlying principle for this class of biosensors is that many chemical reactions
produce or consume ions or electrons, causing some change in the electrical properties of
the solution that can be sensed out and used as a measuring parameter. These can be
classified based on the measuring electrical parameters as conductometric,
amperometric, and potentiometric (Fig. 8). Amperometric and potentiometric
transducers are the most commonly used electrochemical transducers.
(i) Amperometric. Amperometric is a high sensitivity biosensor that can detect
electroactive species present in biological test samples. The potential between the two
electrodes is set and the current produced by the oxidation or reduction of electroactive
species is measured. The height of the peak current is directly proportional to the concen-
Review of Literature
25
Fig. 8 Types of different biosensors
-tration of the electroactive material. Most electrodes are made of metals like platinum,
gold, sliver, and stainless steel, or carbon-based materials that are inert at the potentials at
which the electrochemical reaction takes place. This mode is also known as voltammetric
(Fig. 9).
(ii) Potentiometric. These involve the measurement of the emf (potential) of an
electrochemical cell at very low current. The emf is proportional to the logarithm of the
concentration of the substance being determined. The working principle relies on the fact
that when a ramp voltage is applied to an electrode in solution, a current flow occurs
because of electrochemical reactions. The voltage at which these reactions occur,
indicates a particular reaction and particular species (Fig. 10).
(iii) Conductometric. Most electrochemical reactions involve a change in the
composition of the solution resulting in production of ions or electrons. This will
normally result in a change in the electrical conductivity of the solution, which can be
measured electrically. Conductance measurements have relatively low sensitivity. The
electric field is generated using a sinusoidal voltage (ac), which helps in minimizing
undesirable effects such as Faradaic processes, double layer charging, and concentration
polarization.
Review of Literature
26
(iv) FET-based sensors. Miniaturization can sometimes be achieved by constructing one
of the above types of electrochemical transducers on a silicon chip-based field-effect
transistor (FET). FET is a potentiometric devices based on the measurement of potential
at an insulator-electrolyte interface. The metal gate of a FET can be substituted by an ion
selective membrane to make a pH transducer (pH ISFET). ISFETs having an ion-
sensitive surface. The surface electrical potential changes when the ions and the
semiconductor interact. This change in the potential can be subsequently measured. The
Fig. 9 Schematic representation of an amperometric biosensor
Enzymes
Potential Response
Analyte
Signal
Amplifier
Biocompatible layer
Transducer
Review of Literature
27
Fig. 10 Schematic representation of potentiometric biosensor
ISFET can be constructed by covering the sensor electrode with a polymer layer. This
polymer layer is selectively permeable to analyte ions. The ions diffuse through the
polymer layer, causing a change in the FET surface potential. This method has mainly
been used with potentiometric sensors, but could also be used with voltammetric or
conductometric sensors. This type of biosensor is primarily used for pH detection (Fig.
11).
B. Optical Sensors: In optical biosensors, the output transduced signal that is measured
is light. The biosensor can be made based on optical diffraction or electro-
chemiluminescence. Since they are non-electrical, optical biosensors have the advantages
of lending themselves to in vivo applications and allowing multiple analytes to be
detected by using different monitoring wavelengths. The versatility of fiber optics probes
is due to their capacity to transmit signals that reports on changes in wavelength, wave
propagation, time, intensity, distribution of the spectrum, or polarity of the light. In
n source n drain gate
Biocatalytic
membrane Ion selective
membrane
H+ sensitive
membrane
p silicon
Encapsulant
Reference
electrode
+
-
Review of Literature
28
general, acquisition of the signal from these devices is accomplished through flexible
cables, which can transmit light to the biological component.
Fig. 11 Schematic outline of biosensors based on ion-selective field effect transistor
C. Piezoelectric Sensors: Guilbault and Montalvo (1969) were the first to detail a
potentiometric enzyme electrode. These devices involve the generation of electric
currents from a vibrating crystal. In this mode, sensing molecules are attached to a
piezoelectric surface-a mass to frequency transducer-in which interactions between the
analyte and the sensing molecules set up mechanical vibrations that can be translated into
an electrical signal proportional to the amount of the analyte. Example of such a sensor is
quartz crystal micro or nano balance (Fig. 12).
D. Acoustic. Electroacoustic devices used in biosensors are based on the detection of a
change of mass density, elastic, viscoelastic, electric, or dielectric properties of a
membrane made of chemically interactive materials in contact with a piezoelectric
material. Bulk acoustic wave (BAW) and surface acoustic wave (SAW) propagation
transducers are commonly used. In the first, a crystal resonator, usually quartz, is
connected to an amplifier to form an oscillator whose resonant frequency is a function of
the properties of two membranes attached to it. The latter is based on the propagation of
n+
n+
P
Sapphire
Au
Immobilized
enzyme
membrane
Inactivated
enzyme
membrane
Si3N4
SiO2
P
Review of Literature
29
Fig. 12 Schematic representation of piezoelectric biosensor
SAWs along a layer of a substrate covered by the membrane whose properties affect the
propagation loss and phase velocity of the wave.
E. Thermal Sensors. Thermal detection biosensors are based on absorption or
production of heat that in turn changes the temperature of the medium in which the
reaction takes place. When the analyte comes in contact with the enzyme, the heat
reaction of the enzyme is measured and calibrated against the analyte concentration. The
total heat produced or absorbed is proportional to the molar enthalpy and the total number
of molecules in the reaction.
2.5.2.3. Basic Principles of HbA1C biosensors
HbA1C is formed through the nonenzymatic glycation of the terminal valine of β chain in
Hb. This HbA1C can be digested to small glycated peptide FV that can be further
oxidized by the enzyme FAO. Enzymatic assay of HbA1C is based on the oxidation of FV
(as a model compound). The general strategy used for electrochemical sensing of HbA1C
is by measuring oxygen consumption or the amount of hydrogen peroxide produced by
Oscillator
Piezoelectric crystal
21258967.998 Hz
Frequency counter
Review of Literature
30
the enzyme reaction. The enzymatically produced H2O2 can be monitored by an
electrochemical sensor. Therefore, the HbA1C level can then be determined using this
biosensor by measuring the output current changes from the enzymatic oxidation reaction
producing H2O2. The reactions involved were as shown in Fig. 13.
2.6. Classifications of HbA1C biosensors
2.6.1. Biosensors based on Fructosyl Valine (FV)
The current electrochemical measurements of FV need to be improved in terms of the
sensitivity and the repeatability of the detection. Ideally a FV biosensor should operate at
a lower potential, ambient temperature and would require a small sample volume in order
to minimize the potential interference, improving the operational convenience and
patient’s compliance. Various biosensors based on FV detection have been reported and
extensively studied; in order to construct a biosensor, which could lay a successful foun-
dation for future HbA1C biosensing. Principle of FV based biosensor is as given below:
Fru-Val-His-Leu-Thr-Pro-Glu-Glu-Lys-ser… Protease Fructosyl valyl histidine + amino
(N-terminal residue of β-chain in HbA1C) and/or Fructosyl valine acid
(His, Leu, Thr)
Fructosyl valyl histidine and/or + O2 + H2O FAO Valyl histidine + D-glucosone
Fructosyl valine or valine +H2O2
H2O2 High potential 2H+ + O2 + 2e
2e- through nanoparticles/immobilization support working electrode
2.6.1.1. Amperometric FV biosensors
2.6.1.1.1. Platinum (Pt) electrode coated with membrane
A novel amperometric FV sensor was prepared by coating the surface of a Pt electrode
with a membrane. The partially-purified fructosyl amino acid oxidase (FAO) (0.05 units)
was casted onto the membrane and placed onto a 3 mm Pt electrode. This membrane was
successfully employed as an immobilization support for FAO (from marine yeast). This
FAO membrane catalyzed FV producing hydrogen peroxide, which was then detected
Review of Literature
31
Fig. 13 Scheme of electrochemical reactions involved in HbA1C biosensor based on FAO
(Where MNPs = Metal nanoparticles)
31
MNPs
FAO
Protease Fru-Val-His-Leu-Thr-Pro-Glu-Glu-Lys-ser…
(N-terminal residue of β-chain in HbA1c)
Fructosyl valyl histidine + amino acid
or Fructosyl valine (His, Leu, Thr)
valyl histidine + D-glucosone +H2O2
2H+ + O2 + 2e
Gold electrode
Amperometric
response
Review of Literature
32
electrochemically in a stirred 10 ml testing solution.
The biosensor had following characteristics:
Optimum pH: 7.0-7.5; Optimum temperature: 30-45ºC; Linearity: 0.05-1.8 mM;
Working potential: 0.6 V (vs Ag/AgCl reference electrode); Detection limit: 0.05 mM
Sensitivity: 0.42 µAmM-1
cm-2
(Tsugawa et al., 2000).
2.6.1.1.2. Polyvinylimidazole polymer onto carbon paste electrode
Polyvinylimidazole (PVI), a synthetic polymer was used as a catalyst for fabrication of an
amperometric FV sensor. Molecular imprinting technique was used for oxidative
cleavage of FV. A mixture of PVI and carbon paste was applied on the electrode. The
constructed electrode was then immersed in the phosphate buffer electrolyte containing
1-methoxyphenazinemethosulphate (m-PMS) as mediator.
The biosensor had following characteristics:
Linearity: 20-700 mM; Working potential: +0.1 V (vs Ag/AgCl electrode); Detection
limit: 20 mM; Sensitivity: 0.135 µAmM
-1cm
-1 (Sode et. al., 2001b).
Since FV is an expensive reagent, it is the limiting factor for its utilization as the
template for sensor fabrication. Proteolytic digestion of HbA1C for production of FV also
leads to the formation of another fructosylamine compound (fructosyl lysine denoted as
Fru-ε-lys) which is the proteolytic product of digestion of glycated albumin in the blood
and can interfere with the detection of FV. So, Sode et al., 2003 developed a sensor for
better selectivity for FV over fructosyl lysine and used methyl valine (m-val), which is a
cheaper analogue of expensive FV as the template. Also, they used the positively charged
functional monomer allylamine to improve the selectivity of the sensor toward FV. Both
the sensitivity and selectivity (FV/Fru-ε-lys) decreased from 135 nA/mM to 95 nA/mM
and 1.8 to 1.6, respectively. However, with the introduction of alkylamine as the
functional monomer, the selectivity increased to 1.9, while a sensitivity of 95 nA/mM
could be maintained. Thus, with these two modifications, selectivity increased slightly,
while the sensitivity decreased in exchange for a more inexpensive template (m-val).
2.6.1.1.3. Prussian Blue (PB) modified enzyme electrode
Review of Literature
33
Low potential amperometric enzyme sensor for fructosyl-valine, a model compound of
HbA1C, was constructed using a FAO and Prussian blue film as the artificial peroxidase.
These results obtained indicated that the PB-based FAO electrode may avoid the inherent
problems of hydrogen peroxide measurement at the high potential and also have
advantage in its simple construct of the system compared with the two enzyme
peroxidase-FAO-ferrocene system. The biosensor had following characteristics:
Linear range: 0.1-0.3 mM; Sensitivity: 0.42 µAmM-1
cm-2
; Applied potential: -0.05 V
(Tsugawa et al., 2001).
2.6.1.1.4. Poly(vinylalcohol)-Stylbazole (PVA–SbQ) membrane on oxygen electrode
An amperometric FV biosensor based on FAO immobilized onto silk fibroin membrane
mounted over an oxygen electrode in flow injection analysis (FIA) was constructed.
Source of FAO: Pichia sp. N1-1 strain, Type of immobilization: Cross-linking
The biosensor had following characteristics:
Optimum pH: 7.0; Optimum temperature: 25ºC, Linearity: 0.2-10.0 mM;
Working potential: 0.6 V; Detection limit: 0.2 mM; Sensitivity: 0.0046 µAmM−1
cm−2
;
Application: The biosensor was used for determination of blood HbA1C; Operational
stability: The sensor was capable of detecting 120 consecutive sample injections over 20
h (Ogawa et al., 2002).
This FIA enzyme biosensor was claimed to be the first sensor towards fructosyl
dipeptide, though with insufficient sensitivity.
2.6.1.1.5. Screen-printed iridium modified carbon electrode
A methodology was devised for determination of FV by iridium modified screen-printed
carbon electrodes. It was single-use, disposable FV amperometric biosensor. A single-use
amperometric FV biosensor, incorporating a three-electrode configuration was fabricated
on a polyester substrate using low cost screen printing (thick-film) technology. Both the
working and counter electrodes were prepared by screen-printing commercial carbon ink.
Source of FAO: Recombinant Escherichia coli, Type of immobilization: Adsorption
The biosensor had following characteristics:
Review of Literature
34
Optimum pH: 7.0; Optimum temperature: 25ºC; Linearity: 0.2-2.0 mM; Working
potential: 0.25 V; Response time: 120s; Detection limit: 0.2 mM; Sensitivity: 21.5
μAmM−1
cm−2
; Application: The biosensor was used for determination of blood-HbA1C
(Fang et al., 2009).
Merits: Cost-effective and disposable, so routine measurement of HbA1C level could be
significantly easier for the diabetic patient management.
Demerits: It is single use, interference by presence of potential interferents, since FV is
an expensive reagent, it is the limiting factor for its utilization as the template for sensor
fabrication.
2.6.1.1.6. Non-enzymatic biosensor based on glassy carbon paste electrode (GCPE)
GCPE serving as the working electrode was obtained by dropping the glassy carbon
microparticle carbon paste onto the indium-doped tin oxide (ITO) substrate using a
baking process. A coiled Pt wire and Ag/AgCl electrode (in saturated KCl) were used as
the counter electrode and the reference electrode, respectively. This GCPE was
characterized and reported to have higher sensitivity on FV and lower background current
compared with conventional glassy carbon electrodes.
Optimum pH: 7.4; Optimum temperature: 35ºC; Linearity: 0-1 mM; Working
potential: 1 V; Response time: 40s; Detection limit: 0.05 mM; Sensitivity: 5.26
AmM-1
cm-2
; Square correlation coefficient (R2): 0.999; Interference: No interference
on current response from D-fructose, D-glucose and L-valine.
Merits: Relatively low oxidation potential required for detection of FV without the need
of an enzyme, low cost and the ease of fabrication of the materials involved. These
HbA1C and glucose sensor array were expected to be used for simultaneous point-of-care
measurements, thereby gathering more accurate patient information (Chien and Chou,
2010).
2.6.1.1.7. Non-enzymatic biosensor based on ferrocene boronic acid (FcBA) bound
glassy carbon paste electrode (GCPE)
A novel non-enzymatic amperometric method using a GCPE for fast monitoring of FV
Review of Literature
35
was developed. The method had advantages in terms of: the relatively low oxidation
potential required for detection in 3 mM FcBA, low cost, and the ease of fabrication of
the materials involved. FcBA was employed as useful tool for the capture of Fru-Val
through the interaction of the boronic acid-diols to form a complex and for the signal
transfer by iron ion oxidation. In conclusion, our novel method shows that it is possible to
generate an easily detectable signal, using a very low oxidation potential, to detect Fru-
Val without the need for an enzyme.
Optimum pH: 7.4; Optimum temperature: 25ºC; Linearity: 0.5-4.0 mM; Working
potential: 0.1 V; Response time: 10s; Detection limit: 0.5 mM; Sensitivity: 5.9
AmM-1
cm-2
; Application: The biosensor was used for determination of blood HbA1C
(Chien and Chou, 2011).
Merits: These type of biosensors are amenable to miniaturization and have compatible
instrumental sensitivity.
2.6.1.1.8. Core-shell magnetic bionanoparticles (MNPs) modified gold electrode
An electrochemical biosensor for HbA1C detection by monitoring FV level based on
core-shell MNPs modified gold electrode has been reported by us recently (Fig. 14). The
biosensor had following characteristics:
Working voltage (V): 250 mV; Response time: 4s; Optimum pH: 7.5; Optimum
temperature: 35
oC; Linear range: 0.1-2.0 mM; Detection limit (mM): 0.1 mM;
Stability: 90 days; Reusability: 250 times (Chawla and Pundir, 2011).
2.6.1.2. Potentiometric FV biosensors
2.6.1.2.1. Poly-aminophenylboronic acid (p-APBA) polymer on conductive ITO
electrode
Molecular imprinting technique was used to fabricate a potentiometric FV biosensor.
Molecular imprints of FV in a poly-aminophenylboronic acid (p-APBA) polymer on
conductive ITO electrodes were made. Electrochemical characterization of the fabricated
biosensor was carried out by comparing the open circuit potential (Eoc) of the ITO
carrying the molecular imprinted polymer (MIP) with that measured on a non-imprinted
control in 10 ml of 0.1 M phosphate buffer (pH 7.0) with a standard Ag/AgCl reference
Review of Literature
36
Fig. 14 Chemical sequence of electropolymerization of magnetic bionanoparticles onto
gold electrode and chemical reaction of immobilization of FAO on modified Au electrode
electrode to assess the affinity of the FV imprints for FV, D-fructose, D-glucose and L-
valine. The affinity of the imprinted electrode for FV was higher than for the others. The
measurement of open circuit potentials (ΔEoc) showed that imprinted p-APBA was able
to demonstrate selectivity for the carbohydrate component of the Amadori compound that
is characteristically present in diabetic patients.
Optimum pH: 7.0; Optimum temperature: 25oC; Optimum potential: 5 mV;
Response time: 1000s (Chuang et. al., 2009).
2.6.2. Biosensors based on HbA1C
Other types of HbA1c biosensors detect HbA1c directly. Different methods and
techniques have been used for these HbA1C biosensor. One of their potential advantages
was that there was no need for two time-consuming preliminary steps to release FV from
HbA1C by a protease (one of the main drawbacks with FV point-of-care (POC)
instruments).
Review of Literature
37
2.6.2.1. Amperometric HbA1C biosensors
2.6.2.1.1. Haptoglobin onto 1,1'-carbonyldiimidazole (CDI)-activated cellulose
membranes
An amperometric immunosensor for HbA1C determination, utilizing CDI-activated
cellulose membrane immobilized haptoglobin as affinity matrix with a Pt-working
electrode was described. The HbA1C assay was carried out in a two-step procedure
including the selective Hb enrichment on the sensor surface and the specific HbA1C
detection by a glucose oxidase (GOx) labeled anti- HbA1C antibody.
Enzyme linked immunosorbent assay (ELISA) studies confirmed the advantage of
a sandwich-type format with haptoglobin as capture molecule for selective Hb binding
over the direct adsorption method. Results by the sandwich immunoassay showed a linear
correlation within the clinically relevant range 5-20% (cofficient of variation < 3). The
immobilization procedure of haptoglobin onto CDI-activated cellulose membranes was
optimized for sensor application.
Linearity: 7.8-39 nM (HbA1C linear range: 0 and 25%); Working potential: 0.6 V;
Precision: Within-batch CV 0.9-3.3% for the haptoglobin based assay and 2.2-6.9% for
the anti-Hb antibody based assay (Stöllner et al., 2002).
2.6.2.1.2. Ferroceneboronic acid (FcBA)/zirconium dioxide nanoparticles
(ZrO2NPs)-modified pyrolytic graphite electrode (PGE)
The principle is based on the electrochemical measurement of FcBA that has been
specifically bound to the glycated N-terminus. Hb was immobilized on a ZrO2NPs
modified PGE in the presence of didodecyl dimethyl ammonium bromide (DDAB). The
incubation of this sensor in FcBA solution leads to the formation of an FcBA-modified
surface due to the affinity interaction between boronate and the glycated sites of the Hb.
The PGE is used for protein (total Hb) immobilization and DDAB accelerates electron
transfer between Hb and the electrode. Purified hemolysed erythrocytes from real human
blood sample were mixed with the suspension of ZrO2NPs in the DDAB solution and
then applied to the electrode surface for total Hb immobilization. Square wave
Review of Literature
38
voltammetry (SWV), was used instead of cyclic voltammetry (CV) since the chemically
modified sensor with bound Hb exhibited a relatively large charging current and higher
sensitivity for the Fc label.
Optimum pH: 8.0; Optimum temperature: 22ºC; Working potential: 0.299 V versus
Ag/AgCl (1M KCl); Linearity: HbA1C from 6.8 to 14.0%; Incubation time: 30 min.;
Reproducibility: 12.7% for 10.2% HbA1C samples at the different total Hb
concentrations (n=3) (Liu et al., 2006a).
Demerits: The major drawback of this method was that deviations between the results
obtained with the present two methods range from 10.7 to 31%. The requirement for the
separate determination of the total Hb content also made this an inconvenient aspect of
this method. Further improvement of the method is needed in terms of measuring time
and precision.
2.6.2.1.3. 3-AminoPhenylboronic acid (APBA) modified graphene oxide (GO) onto
glassy carbon electrode
APBA modified GO was synthesized and was applied for electrochemical detection of
HbA1C through modified glassy carbon electrode. The compound (GO-APBA) was
synthesized by forming an amide linkage between the amino group (-NH2) of APBA and
the carboxylic group (-COOH) of GO. The compound was characterized using IR
spectroscopy. Detection of HbA1C was carried out using Electrochemical Impedance
Spectroscopic (EIS) measurements. The other characteristics of the biosensor were:
Optimum pH: 8.0; Optimum potential: 0.2 V (Krishna et al., 2011).
Table 1 provides a comparison of analytical properties of amperometric FV and HbA1C
biosensors.
2.6.2.2. Potentiometric HbA1C biosensors
2.6.2.2.1. Integrated chip and a micro extended gate electrode array
A micro potentiometric immunosensor based on an integrated chip and a micro extended
gate electrode array was reported, which can detect Hb and HbA1C simultaneously. It is
free labeling, simple and convenient. The integrated chip based on standard complement-
Review of Literature
39
Table 1. A comparison of analytical properties of amperometric FV and HbA1C biosensors
39
S.
No.
Type of
biosensor
Support of
immobilization
Working
voltage
(V)
Response
time (s) Optimum
pH Optimum
temp. (
oC)
Linear
range
(mM)
Detection
limit
(mM)
Sensitivity
(µA
mM-1
cm-2
)
Coefficient
of variance
(%)
Reference
1. Amperometric FAO/membrane/Pt
electrode
0.6
ND
7.0-7.5
30-45
0.05-1.8
0.05
0.42
ND
Tsugawa
et al., 2000
2. Amperometric PVI polymer onto
carbon paste
electrode
0.1
ND
ND
ND
20-700
20
0.135
ND
Sode et.
al., 2001
3. Amperometric PB film/Electrode
-0.05
ND
ND
ND
0.1-0.3
0.1
0.42
ND
Tsugawa
et al., 2001
4. Amperometric FAO/PVA-SbQ on
oxygen electrode
0.6
ND
7.0
25
0.2-10.0
0.2
0.0046
ND
Ogawa
et al., 2002
5. Amperometric Haptoglobin/CDI-
activated cellulose
membranes
0.6
ND
ND
ND
0.78x10-5-
3.9 x10-5
ND
ND
0.9-3.3
Stöllner et
al., 2002
6. Amperometric FcBA/ZrO2NPs/PG
E
0.299
1800
8.0
22
6.8 to
14.0%
ND
ND
12.7
Liu et al.,
2006a
7. Amperometric FAO/iridium
modified carbon
electrode
0.25
120
7.0
25
0.2-2.0
0.2
21.5
ND
Fang
et al., 2009
8. Non-
enzymatic
Glassy carbon
microparticle
carbon paste/ITO
1.0
40
7.4
35
0-1.0
0.05
5.26
ND
Chien and
Chou, 2010
Review of Literature
40
FAO: Fructosyl amino acid oxidase; PVI: Polyvinylimidazole; PB: Prussian blue; PVA-SbQ: Poly(vinylalcohol)-Stylbazole; CDI: 1,1'-
carbonyldiimidazole; FcBA: Ferroceneboronic acid; GCPE: Glassy carbon paste electrode; MNPs: Magnetic bionanoparticles; APBA: 3-
AminoPhenylboronic acid; GO: Graphene oxide; GCE: Glassy carbon electrode; ND: Not detected; ZnONPs: zinc oxide nanoparticles; PPy:
polypyrrole
9. Non-
enzymatic
FcBA/GCPE
0.1
10
7.4
25
0.5-4.0
0.5
5.90
ND
Chien
and Chou,
2011
10. Amperometric FAO/core shell
MNPs/Au electrode
0.25
4
7.5
35
0.1-2.0
0.1
35.72
2.58-5.63
Chawla and
Pundir, 2011
11. Amperometric APBA/GO/GCE
0.2
ND
8.0
ND
ND
ND
ND
ND
Krishna et
al., 2011
12. Amperometric FAO/ZnONPs/PPy/
Au
0.27 2 7.0 35 0.1-3.0 0.05 38.42 1.58- 2.07 Present
40
Review of Literature
41
-ary metal-oxide-semiconductor (CMOS) technology consisted of the ion-sensitive field-
effect transistors (ISFET)/reference field-effect transistor (REFET) sensing device and
signal-processing circuit. The electrode array was fabricated by micro electro mechanical
systems (MEMS) method. A two-layer structured monolayer is formed on the anti-
HbA1C/anti-Hb gold electrode by the mixed self-assembled monolayers (SAMs)
technique. Following characteristics of biosensor were studied:
Optimum pH: 7.4; Optimum temperature: 25oC; HbA1C range: 5-20%; Sensitivity:
0.35 mV/μg ml-1
and 0.13mV/μg ml-1
for HbA1C and Hb, respectively; Response time:
120s; Accuracy: 99.4%; Correlation coefficient: 0.996; RSD: 0.85% (Xue et al., 2010).
2.6.2.2.2. Mixed self assembled monolayers (SAMs) wrapped nano-spheres array
A micro potentiometric immunosensor based on mixed SAMs wrapped nano-spheres
array for the detection of HbA1C level. Nano-spheres array was prepared by wrapping
gold nanoparticle with mixed SAMs on the surface of micro immunosensor. Mixed
SAMs make the nano-gold particles distribute uniformly without aggregation and render
the signal less susceptible to noise. Based on this nano-spheres array, antibody is
covalently immobilized on the immunosensor surface. The micro immunosensor,
consisting of ISFET integrated chip and MEMS electrodes array is applied to measure
HbA1c level by detecting the concentration of HbA1c and Hb simultaneously. The
biosensor had following characteristics:
Linearity: 166.7-570 ng/ml Hb and 50-170.5 ng/ml HbA1C; Recovery: 99.08% and
103.27%; RSD: 5%; Sensitivity: 94.73µV/(ngmL−1
); Interference: No significant
interference by interfering substances like bovine serum albumin, lysis solution and some
potassium ion, chloride ion. Although the HbA1C detected channel and Hb detected
channel were integrated on one electrode array chip, the two channels can get good
results in one drop of simulated blood sample contained Hb and HbA1C, with no
interference each other (Xue et al., 2011a).
2.6.2.2.3. FET sensor chip and a disposable microelectrode-array chip
Field-effect transistor (FET)-based micro potentiometric immunosensor consisting of an
Review of Literature
42
integrated FET sensor chip was fabricated with CMOS processes and a disposable micro
electrode array chip fabricated by MEMS technique. This immunosensor was applied for
simultaneous detection of Hb and HbA1C. The integrated FET sensor chip consisted of
ISFET/REFET sensing devices and signal-processing circuit. Three methods were
researched for immobilizing antibody i.e. mixed SAMs method, seed mediated growth
nano-gold method and nano-gold wrapped with mixed SAMs method on the micro
extended gate electrode array chip. The immunosensor fabricated by nano-gold wrapped
with mixed SAMs method showed the highest sensitivity. Whole blood samples were
detected by this immunosensor. Good consistency and high accuracy were achieved. As
shown from the results, the immunosensor is free labeling, simple and has the potential to
become a portable device for diabetes control. The biosensor had following
characteristics:
Sensitivity (V/ng mL−1
): 189.8 for HbA1C and 40.42 for Hb; Recovery: 102.54; RSD:
2.63
Merits: It could detect Hb and HbA1C at the same time (Xue et al., 2011b).
2.6.2.2.4. Polypyrrole (PPy)-gold nanoparticles (AuNPs) composite
A novel micro-potentiometric HbA1C immunosensor based on electrochemically
synthesized PPy-AuNPs composite. PPy-AuNPs film with AuNPs uniformly distributed
in it was deposited on gold electrode surface by a simple and direct procedure, without
the addition of any nanoparticles or reducing agent. This generic method made it possible
to deposit different polymers on miniaturized electrodes. With the existence of AuNPs,
the antibody immobilization onto the electrode surface was facilitated. Based on an
ISFETs integrated chip, a micro-potentiometric immunosensor for Hb and HbA1C has
been constructed.
The biosensor had following characteristics:
Linearity: 60-180 µg/ml Hb and 4-18µg/ml HbA1C; Sensitivity: 0.20mV µg−1
ml-1
;
Response time: less than 1 min.; Linear dose-response behavior: 125 and 197 µg/ml;
Storage stability: 35.1% at 4oC under dry conditions for 5 days in 8 µg/ml Hb in
phosphate buffer saline (PBS) (pH 7.4); Reproducibility: The variation coefficients were
Review of Literature
43
relatively high. Further optimization is needed to improve the reproducibility for real
sample analysis; Interference: 9.2% decrease in biosensor response by number of
interferents such as immunoglobulin G, α-fetoprotein and BSA (Qu et al., 2009).
2.6.2.2.5. Boronic acid-modified thin film/gold electrode
A boronic acid-modified thin film interface for selective binding of HbA1C followed by
electrochemical biosensing using an enzymatic backfilling assay. A freshly evaporated
gold working electrode for the bottom-up layer formation process. This procedure began
with the formation of an amine-reactive Dithiobis(succinimidyl propionate) (DTSP)
SAM on the gold which was then transferred to a poly(amidoamine) G4 dendrimer
solution. Then 4-formyl-phenylboronic acid (FPBA) was immobilized on the dendrimer
layer selective for HbA1C. FPBA functionalization was confirmed by XPS and cyclic
voltammetry. To carry out the backfilling assay, samples with various ratios of
HbA1C/HbA0 (with normal adult human Hb concentration i.e. 150 mg/ml) in a pH 9.0
bicarbonate buffer were contacted with the functionalized surface to react with FPBA for
1 hour. After rinsing with buffer and PBS, 1 mg/ml activated GOx in PBS was added in
order to bind to the remaining unreacted amine groups on the dendrimer-FPBA layer or
30 min. The response of this electrode sensor was assessed by subjecting it to a
voltammetric scan from 0 to +500 mV vs. Ag/AgCl at a rate of 5 mV/s in PBS in the
presence of 0.1 mM ferrocenemethanol (as mediator) and 10 mM glucose (as substrate).
The anodic current measured at +400 mV was chosen as the sensor signal because of
stable current at this potential in the voltammogram. Although this sensor had the
advantage of signal amplification without the need for pretreatment such as labelling or
use of labeled secondary antibody, if required incubation of the Hb sample and then GOx
solution requires 1 hour and 30 minutes, respectively. In addition, the sensitivity at
HbA1C levels below 5% is not sufficient. The biosensor had following characteristics:
Optimum pH: 7.2; Working potential: 0.4 V versus Ag/AgCl; Potential sweep rate:
50 mV/s; Linearity: HbA1C from 2.5-15% (Song and Yoon, 2009).
Table 2 provides a comparison of analytical properties of potentiometric FV and HbA1C
biosensors.
Review of Literature
44
2.6.2.3. Piezoelectric sensor
2.6.2.3.1. Ferroceneboronic acid modified piezoelectric quartz crystal
An electrochemical immunoassay based on the multiple affinity labeling of the indicator
antibody with an electro-active tag is presented. Hb is adsorbed to the surfactant-
modified surface of a piezoelectric quartz crystal. Whereas the quartz crystal nanobalance
is used to validate the total Hb binding, the HbA1C on the sensor surface is recognized by
an antibody and quantified electrochemically after the sugar moieties of the antibody
have been labeled in-situ with ferroceneboronic acid. The sensitivity of this sensor is
about threefold higher than the sensitivity of a Hb sensor, where the ferroceneboronic
acid is bound directly to HbA1C. The total Hb content was determined using a mass-
sensitive quartz crystal modified with a surfactant, while the FcBA-bound HbA1C on the
surface was measured using square wave voltammetry. The biosensor had following
characteristics:
Optimum pH: 8.0; Optimum temperature: 25oC; Response time: 300s; Optimum
potential (vs. Ag/AgCl): +200mV; RSD: 5.1% (Halamek et al., 2007).
2.6.2.3.2. 3-aminophenylboronic acid (APBA) modified gold electrode
The biosensor employed a combination of the piezoelectric biosensor for HbA1C and the
flow-through photometric sensor for total haemoglobin (Hb). The modification of gold
electrodes with APBA, as a specific ligand was studied; the chemisorbed conjugate of
APBA with a long-chain thiocompound (11- mercaptoundecanoic acid) provided the best
affinity for HbA1C. The total haemoglobin content was analysed from 50 to 1000 µg/ml
as absorbance of the haemoglobin-cyanide derivative at 540 nm. Thus, Interaction
between immobilised boronic acid and glycated haemoglobin was used as the recognition
reaction on the surface of the piezoelectric biosensor. Only one standard (calibrator)
diluted in various ratio was necessary for calibration and 1 µl of blood was sufficient for
analysis. The full range of HbA1C content (4-15%) in blood was analysed. In the
presence of higher concentrations of HbA1C, the surface became saturated and the
response versus concentration dependency was non-linear.
Review of Literature
45
Table 2. A comparison of analytical properties of potentiometric FV and HbA1C biosensors
Poly-aminophenylboronic acid: p-APBA; Polypyrrole: PPy; AuNPs: Gold nanoparticles Indium tin oxide: ITO; Self Assembled Monolayers:
SAMs; FET: Field effect transistor; ND: Not detected; RSD: Relative standard oxidation
S.
No.
Support of
immobilization
Linear range
(%)
Applied
potential
(mV)
Response
time (s)
Optimum pH Optimum
temperature
(oC)
Sensitivity
(mVµg−1
mL-1
)
RSD (%) Reference
1. p-APBA
polymer on
ITO electrode
ND +5 1000 7.0 25 ND ND Chuang et al.,
2009
2. PPy/AuNPs
composite
4-18 µg mL-1 ND 60 ND ND 0.20 0.20 Qu et al., 2009
3. Boronic acid-
modified thin
film/Au
electrode
2.5-15.0 +400 ND 7.2 ND ND ND Song and Yoon,
2009
4. Integrated chip
and micro
extended gate
electrode array
5.0-20.0 ND 120 7.4 25 0.35 0.85 Xue et al., 2010
5. Mixed SAMs
wrapped nano-
spheres array
0.05-0.171 g
mL-1
ND 120 7.4 ND 0.095x10-3 5.0 Xue et al.,
2011a
6. FET sensor chip
and a disposable
micro electrode-
array chip
5.0-15.0 ND ND ND ND 0.189 2.63 Xue et al.,
2011b
45
Review of Literature
46
Optimum pH: 8.0-9.0; Optimum temperature: 22ºC; Working range: 10-90 µg/ml;
Response time: 15 min/sample with 20 measurements without any user intervention.
The developed method was successfully evaluated on blood samples collected from
diabetics (Pˇribyl and Skl´adal, 2006).
2.6.2.3.3. 3-boronic acid (T3BA) self-assembled monolayer (SAM)/Au electrode
HbA1C was selectively immobilized on a gold electrode modified with T3BA-SAM and
was detected by label-free electrochemical impedance spectroscopy (EIS). The reported
binding mechanism is based on bonding between the sulphur atom of the π-stacked
thiophene SAM and the gold. The binding of T3BA and formation of a SAM on the gold
was confirmed by the use of a quartz crystal microbalance (QCM), atomic force
microscopy (AFM) and EIS experiments. Electrochemical determination was based on
measuring the change in the capability of the gold electrode for electron transfer due to
blocking of the electrode surface by HbA1C after immobilization. The biosensor had
following characteristics:
Optimum pH: 8.5; Concentration range: 0.1-1 µg/mL; Response time: 600s
Drawbacks: The variation of signal with HbA1C concentration also depends on total Hb
concentration. Consequently, the total Hb concentration was also needed to be
determined separately to obtain the HbA1C content. Since this method is not selective for
HbA1C over glycated albumin, glycated albumin must be separated from RBC by centrif-
-ugation (Park et. al., 2008).
2.6.2.4. Biochips for HbA1C detection
2.6.2.4.1. m-amino-phenylboronic acid (m-APBA) modified agarose bead/inter-
digitated array IDA electrodes
A disposable biochip was constructed by using an integrated system consisting of a pair
of interdigitated array (IDA) electrodes, HbA1C binding chamber, blood lysis chamber,
filter, micro-pump and microchannel. Ferricyanide (K3Fe(CN)6) was used as a mediator
so that the electrons released from the oxidation of Fe2+
in Hb were transferred to the
electrode by the ferricyanide/ferrocyanide couple. When Fe2+
in Hb is oxidized to Fe3+
by
Review of Literature
47
an electron delivery medium of ferricyanide (K3Fe3+
(CN)6), the ferricyanide is
deoxidized to a ferrocyanide (K4Fe2+
(CN)6), and current flows as a result of the applied
electric potential to the ferrocyanide. After plasma separation (1) and red blood cell
(RBC) lysis (2), the total Hb stream branches off into two separate streams: in the lower
stream HbA1C was immobilized on a packed agarose bead containing m-amino-
phenylboronic acid (m-APBA) in the binding chamber and releases Hb, while total Hb
flows in the upper stream (3). The ratio of the resulting electrochemical signals from the
lower and upper streams after passing through the IDA electrodes yields the % HbA1C.
Due to the nonhomogeneous distribution of Hb, the instantaneous current varies as a
sample flows through the IDA electrodes. Consequently, the integral of the current over
time was used for measurement of % HbA1C (Son et al., 2006).
2.6.2.4.2. Microfluidic system using Au dielectrophoretic (DEP) electrodes
RBCs were successfully separated and lysed in an automatic manner utilizing the Au
dielectrophoretic (DEP) forces in a microfluidic system composed of a circular
micropump and normally closed valves. The chip was fabricated using MEMS technique.
The electrochemical detection module composed of three electrodes on a glass substrate
for sensing the concentration of HbA1C electrochemically. The three electrodes including
a working (carbon paper), a counter (Pt) and a reference (Ag) electrode were designed
and fabricated using thin-film deposition processes.
Optimum pH: 7.4; Optimum potential: 1000 mV; Linear range: 0 to 2 mg/ml;
Response time: 10 min.; Standard deviation: 1.56% (Huang et al., 2011).
Table 3 provides a comparison of analytical properties of piezoelectric biosensors and
biochips for HbA1C detection.
2.7. Conclusion and Future Outlook
During past few years, various methods for HbA1C determination were reported.
Compared to conventional methods, the biosensing methods are very rapid, selective and
sensitive. However, more efforts are needed to tackle technical difficulties in biosensor to
find better alternatives such as: (1) need of blood pretreatment (2) separate determination
Review of Literature
48
of total Hb (3) leaching out of enzyme or larger biomolecules from electrode (4)
selectivity and detection limit require to be more selective and wide. There is still need of
more reliable, specific, accurate and portable HbA1C biosensors.
Most of the reports have focused on the use of FAO for detection of HbA1C.
These biosensors have the potential to be converted to point-of-care testing (POCT)
devices, which can potentially play an important role in diabetes diagnosis and
management. Miniaturization is expected to have a marked impact on the development
and applications of biosensors. Miniaturization of a biosensor not only reduces the size of
detection device and sample volume, but also integrates all steps of the analytical process
into a single-sensor device. Thus, it results in reduction of both time and cost of analysis.
Moreover, it is expected to lead to further portability for in vivo sensing and in-field
applications. Trials can be made to integrate the laboratory model with chip designing
companies (VLSI) in order to develop a fully automated chip system in a device, to be
used by any patient on his bedside.
Review of Literature
49
Table 3. A comparison of analytical properties of piezoelectric biosensors and biochips for HbA1C detection
FcBA: Ferroceneboronic acid; APBA: 3-aminophenylboronic acid; T3BA: 3-boronic acid; SAM: Self-assembled monolayer; ND: Not detected;
RSD: Relative standard deviation
*******************
S. No. Support of
immobilization
Linear
range (µg
mL-1
)
Applied
potential (mV)
Response time
(s)
Optimum pH Optimum
temperature
(oC)
RSD (%) Reference
1. APBA modified
Au electrode
10-90 ND 900 8.0-9.0 22 ND Pˇribyl and Skl´adal,
2006
2. FcBA modified
piezoelectric
quartz crystal
ND +200 300 8.0 25 5.1 Halamek et al., 2007
3. T3BA/SAM/Au
electrode
0.1-1.0 ND 600 8.5 ND ND Park et. al., 2008
4. Integrated chip
and a micro
extended gate
electrode array
0 to 2000 +1000 600 7.4 ND 1.56 Huang et al., 2011
49
Review of Literature
50