Physics Laboratory. UNITED STATES DEPARTMENT OF COMMERCE National Institute of Standards and Technology Gaithersburg, MD 208-0001. Radiochromic Film Dosimetry. Christopher G. Soares Ionizing Radiation Division National Institute of Standards and Technology. Important reference. - PowerPoint PPT Presentation

Citation preview

Page 1: Radiochromic Film Dosimetry

Radiochromic Film DosimetryRadiochromic Film Dosimetry

Christopher G. SoaresIonizing Radiation Division

National Institute of Standards and Technology

Physics LaboratoryPhysics Laboratory

UNITED STATES DEPARTMENT OF COMMERCENational Institute of Standards and TechnologyGaithersburg, MD 208-0001

Page 2: Radiochromic Film Dosimetry

Important reference

Task Group 55 RecommendationsMedical Physics, Vol. 25 (11),

2093 - 2115, Nov. 1998- Introduction- Review of Film Dosimetry - Review of Medical Applications- Review of Scanning / Densitometer - Summary & Future Directions

Page 3: Radiochromic Film Dosimetry

IntroductionTwo Dimensional Radiation Dosimeters

A. Silver halide radiographic films• Advantages:

– High spatial resolution– Relative dose measurement

• Disadvantages:– Energy dependence– Room light sensitivity – Wet chemical processing– Not tissue equivalent– Not an absolute dosimeter

Page 4: Radiochromic Film Dosimetry

Introduction (Cont.)Two Dimensional Radiation Dosimeters

B. Radiochromic films• Advantages:

– High spatial resolution– Energy dependence (??)– Tissue equivalent– Not sensitive to room light – No chemical processing

• Disadvantages:– Uniformity– Environmental factors– Not an absolute dosimeter

Page 5: Radiochromic Film Dosimetry

Introduction (Cont.)

• William McLaughlin, NIST– Since 1965:

Organic free-radical imaging medium can combine photopolymerization with leuco dyes that produce color upon irradiation.

• Dave Lewis, International Specialty Products (ISP)

– Since 1980s: Radiochromic films based on polydiacetylene has been introduced for medical applications

Page 6: Radiochromic Film Dosimetry

Radiochromic Films

Radiochromic film consists of a single or double layer

of thin nylon-base with a transparent coating

of radiation-sensitive organic microcrystals monomers.

Page 7: Radiochromic Film Dosimetry

Radiochromic Films (Cont.)• HD-810 (Formerly: DM-1260)

– 20 cm x 25 cm x 0. 1 mm – ISP & Nucl. Assoc. (Model No. 37-040)– Dose Range: 50-2500 Gy

• MD-55-1 (no longer commercially available)– 12.5 cm x 12.5 cm x 0.08 mm– ISP & Nucl. Assoc. (Model No. 37-041)– Dose Range: 10-100 Gy

• MD-55-2 (NMD-55)– 12.5 cm x 12.5 cm x 0.25 mm– ISP & Nucl Assoc. (Model No. 37-041)– Dose Range: 3-100 Gy

67 m

67 m

25 m

100 m

67 m

7 m

15 m

15 m

15 m

45 m

45 m

Polyester

Emulsion

Glue

Page 8: Radiochromic Film Dosimetry

Yellow polyester - 150 microns

Yellow polyester - 150 microns

Active layer - 18 microns

Active layer - 15 microns

Yellow polyester - 150 microns

Opaque white polyester - 150 microns

XR Type R

XR Type T

New GAFCHROMIC MediaNew GAFCHROMIC MediaClear polyester – 150 microns

Active layer - 40 microns

Clear polyester – 150 microns

HS

Page 9: Radiochromic Film Dosimetry

Radiochromic Films (Cont.)

Color of the radiochromic films turns to shade of blue upon irradiation.

Darkness of the film increases by increasing the absorbed dose.

Page 10: Radiochromic Film Dosimetry

Factors to be considered in Selecting 2D

Radiation DosimetersSensitivityUniformity

Dynamic RangeSpatial Resolution

Energy / LET DependenceTissue Equivalence

Dose Rate DependenceLight SensitivitySignal Stability

Environmental Stability: Temp, Humidity, ...

Page 11: Radiochromic Film Dosimetry

RCF: Characteristics• Effective Z: 6.0 - 6.5• Sensor material has similar electron

stopping power as water & muscle• Sensor material has similar mass-energy

absorption coefficients as water and muscle for h > 100 keV

• For secondary electron 0.1 to 1.0 MeV and h 0.1 to 1.33 MeV : ~2% of water and muscle

Page 12: Radiochromic Film Dosimetry

Radiochromic Film Absorbance SpectraRadiochromic Film Absorbance Spectra

Page 13: Radiochromic Film Dosimetry

RCF: Color Stability

• At certain wavelengths of the main absorption band (~ 670 nm), absorption is stable the 1st four hrs.

• During the 1st 24 hrs after irradiation, absorption increases by up to 16%. – 4% thereafter for up to 2 wks

• Greatest increase in absorption occurs at higher temp: ~ 40ºC

Page 14: Radiochromic Film Dosimetry

POST-EXPOSURE DENSITY GROWTHGAFCHROMIC® HS Lot# 31141-3

Normalized to density change at 24 hrs after exposure

60%

80%

100%

120%

0.1 1 10 100 1000

Time, hours

% o

f den

sity

@ 2

4 hr

s

40.1Gy30.2Gy20.2Gy10.1Gy

Page 15: Radiochromic Film Dosimetry

RCF: Effect of Polarized Light

• Klassen et al studied polarity effect for MD-55-2:– Microcrystals in sensitive layers have a

preferred orientation (i.e. flat)– Monomers in microcrystals have a preferred

orientation in the same plane of the film OD would vary with the plane of polarization

of the analyzing light and would change with film orientation

Page 16: Radiochromic Film Dosimetry

RCF: Dose Response• HD-810 is insensitive to light at > 300 nm

– But sensitive to UV at lower • HD-810 needed 50 Gy for 2% precision• MD-55-2 needs 3 GY for 2% precision• Films should be stored in the dark, at temp <

25ºC and relative humidity < 50% to optimize the useful life of the film

• Dependence of the absorption spectrum with dose has been documented

• Fractionated and unfractionated resp. ~ 1%

Page 17: Radiochromic Film Dosimetry

Film Type Wavelength Irradiated Response*Readout with mAu/Gy

HD-810 633 60Co 3

MD-55-2 633 60Co 20

XR-T 633 60Co 30

HS 633 60Co 40

Kodak XV white linac 3000

*Response is defined as net optical density per unit absorbed dose

Comparison of Film SensitivitiesComparison of Film Sensitivities

Page 18: Radiochromic Film Dosimetry

RFC: Energy Response• Muench et al have studied the energy

response of HD-810 as compared with LiF for h (20 to 1710 keV)– Response by ~ 30% as h – Similar but in opposite direction as of TLD– Silver halide film response by ~ 980%

• Chiu-Tsao et al studied the energy response of MD-55-1 using brachytherapy sources– Sensitivity is ~ 40% lower for I-125 than Co-60

Page 19: Radiochromic Film Dosimetry

RCF: Energy Response (Cont.)• McLaughlin et al studied MD-55-2

– Response is ~ 40% lower for h 20 to 40 keV than with Cs-137 and Co-60

• Sayeg et al suggest that the lower response of RCF for h < 100 keV is due to the larger carbon content relative to soft tissue

Page 20: Radiochromic Film Dosimetry

Problem of GAFChromic MediaProblem of GAFChromic MediaIrradiated by Low E photonsIrradiated by Low E photons

excess of C relative to water gives causes an under response

GAFCHROMIC SENSOR

-50

-40

-30

-20

-10

0

10

0.001 0.01 0.1 1 10Energy /MeV

Diff

eren

ce fr

om w

ater

/%

Mass-energy absorption coefficients

Page 21: Radiochromic Film Dosimetry

Fractions by weightFractions by weight H C N

OWater 0.89 -- -- 0.11

GAFChromic sensor 0.09 0.61 0.11 0.19

Page 22: Radiochromic Film Dosimetry

ResultsResults

0

0.2

0.4

0.6

0.8

1

1.2

0 20 40 60 80 100Average Spectrum Energy, keV

Res

pons

e re

lativ

e to

Co-

60

M-series x-ray beam exposures:violet squares: HD-810 no backinggreen squares: HD-810 with PS backingred circles: MD-55-1 with no backingblue circles: MD-55-1 with PS backing

I-125 source exposures:green +'s: HD-810 in SW at 0.5 and 1 cmwhite x's: MD-55-2 in SW at 0.5 and 1 cm

blue line: ratio of mass-energy absorptioncoefficients GAFChromic medium and /water

±10%

Page 23: Radiochromic Film Dosimetry

RCF: Dose Rate Response• Saylor et al show that for MD-55-1:

– There is no dose rate response within an uncertainty of ~ 5% ( one SD)

• McLaughlin et al studied MD-55-2 for:– Dose = 20, 40, 60 Gy– Dose rate = 0.08 to 80 Gy/min– No dose rate response within an uncertainty of ~ 5%

( one SD)– But at 60 Gy there is ~ 10% higher response at

lower dose rate

Page 24: Radiochromic Film Dosimetry

RCF: Environmental Factors• Humidity (HD-810) effect for 6 to 94% < ± 2%• Temperature effect (MD-55-1 & MD-55-2) for 10 to

50C:– Response varies with dose as well as of analysis– At ~ 50C there is an erratic variation in response– At > 60C the blue dye changes to red and may cause significant

change in sensitivity• UV exposure with > 300 nm(sunlight or continuous

white fluorescent lights) colors the film– Needs to be stored in opaque container

• Shipping and handling may cause damaged locations– Color of the film turns from clear to milky white– Need to be ~ 1.5 mm away from cut edge

Page 25: Radiochromic Film Dosimetry

RCF: Uniformity• Non-uniformity due to local fluctuations (spikes):

– Small scale: film grain size, spatial and signal resolution of scanner, pixel size, electronic noise, …

– Relative response is compared with the mean response in the ROI

• Non-uniformity due to regional variations:– Large scale: non-uniformity in film emulsion layer(s),

systematic scanner problem(s)– Difference (or ratios) of max-min response in ROI

Page 26: Radiochromic Film Dosimetry

RCF: Uniformity (Cont.)• Acceptable tolerances for film uniformity varies

with application• Meigooni et al studied regional variation for

MD-55-1 & MD-55-2 along two central orthogonal directions:– Longitudinal ( to coating application): ~ 4%– Transverse ( to coating application): ~ 15%

• 5 institution studies of local fluctuations:– Dose > 20 Gy: ~ ± 3%– Dose < 10 Gy: ~ ± 5%

Page 27: Radiochromic Film Dosimetry

RCF: Double-Exposure Technique

Zhu et al suggest using a matrix of correction factors from relative film response

ODnet( i,j) = [OD2(i,j) - OD1(i,j)] / f(i,j)

f(i,j) = OD1(i,j) / <OD1(i,j)>

Page 28: Radiochromic Film Dosimetry

RCF: Calibration & Sensitivity• Calibration:

– Large well-characterized uniform radiation field• suggest: 40 x 40 cm, d 5 cm

• Calibration curve:– Relation between dose and response

• Sensitivity:– Average change in response per unit dose

• calculate over most linear portion of the calibration curve – Depends on: used for readout, scanner, film batch,

beam quality, readout time, T, H, ...

Page 29: Radiochromic Film Dosimetry

RCF: Medical Applications• Ophthalmic Applicator Dosimetry

– Sayeg et al used Sr-90 eye applicator• He-Ne scanning laser densitometer

– NIST extrapolation ionization chamber(Agreement ~ 6%)

– Soares mapped surface dose• LKB scanning laser densitometer

– Soares found response to e same as to Co-60 (5% at 95% confidence level)

Page 30: Radiochromic Film Dosimetry

RCF: Medical Applications (Cont.)• Muench et al studied brachytherapy sources

– Ir-192 (370 Gbq) – Compared with TLDs– Kodak X-Omat films

• Farahani et al studied tissue-metal interfaces• Bjarngard et al and McLaughlin et al studied small

fields for stereotactic (Agreement with calculation ~ 2%)

• Galvin et al studied penumbra region(Good agreement with silver halide film)

Page 31: Radiochromic Film Dosimetry

RCF: Medical Applications (Cont.)

• Soares et al studied hot particles in nuclear reactor

• Van Hoek et al studied inactivation of proteins using hundreds of kGy doses

(Agreement with Fricke dosimeter ~ 1%)

• Vatnitsky et al studied proton beams• Linear response (10 - 100 Gy) for h, e, p (except

at Bragg peak region: 5-10% in response)

Page 32: Radiochromic Film Dosimetry

RCF: Medical Applications Intravascular Brachytherapy

• Soares et al studied dosimetry of 90Sr/Y seed sources

• Mourtada et al studied dosimetry of 32P wire sources

• Duggan et al studied 32P coated stents for intravascular application

Page 33: Radiochromic Film Dosimetry

Cylindrical Film GeometryCylindrical Film Geometry

End View Side View

SourceFilm

Page 34: Radiochromic Film Dosimetry

Cylinder Film of a Uniform Cylinder Film of a Uniform 9090Sr/Y SeedSr/Y Seed

Page 35: Radiochromic Film Dosimetry

A Series of A150 Blocks for Depth Dose DeterminationsA Series of A150 Blocks for Depth Dose Determinations

Page 36: Radiochromic Film Dosimetry

Film Irradiation GeometryFilm Irradiation Geometry

End View Side View

Film

SourceA150 plastic

A150 plastic

Page 37: Radiochromic Film Dosimetry

Generation of Depth Dose Profiles from Film:Generation of Depth Dose Profiles from Film:Multiple Exposures for Each PlaneMultiple Exposures for Each Plane

Example at depth of 0.44 mmExample at depth of 0.44 mm

1 min 3 min 10 min 100 min1 min 3 min 10 min 100 min

Page 38: Radiochromic Film Dosimetry

Densitometer Densitometer ParametersParameters

• light source

• light detection

• response linearity

• response accuracy

• response reproducibility

• response stability

• spatial resolution

• positional accuracy

• bed geometry

• acquisition time

• control software

• environmental factors

Page 39: Radiochromic Film Dosimetry

Light SourcesLight Sources

• wavelength• FWHM• intensity• size• uniformity• match to light detector• match to film being

read

• Examples:• HeNe laser, 633 nm, nearly 0 FWHM• 1 mW, 50μm diameter spot• Filtered white light• LED light bed

Page 40: Radiochromic Film Dosimetry

Types of Densitometers• Moving:• single light source• single detector• sample and/or light source-detector moved

• Imaging:• uniform backlit bed• imaging device• no movement

• Hybrids:• combination of the above

Page 41: Radiochromic Film Dosimetry

Example of aExample of a Moving SystemMoving System

Page 42: Radiochromic Film Dosimetry

ExampleExampleof anof an

ImagingImagingSystemSystem

Page 43: Radiochromic Film Dosimetry

LightLight DetectorsDetectors

sensitivity

spectral efficiency

linearity

signal resolution

Types:

PMT

CCD

Page 44: Radiochromic Film Dosimetry

Response LinearityResponse Linearity

calibration curve using film

calibrated neutral density filters

comparison with spectrophotometer

Page 45: Radiochromic Film Dosimetry

Spectrophotometer vs.Spectrophotometer vs.Scanning DensitometerScanning Densitometer

Page 46: Radiochromic Film Dosimetry

Response ReproducibilityResponse Reproducibility

Page 47: Radiochromic Film Dosimetry

Spatial ResolutionSpatial Resolution

Moving Systems:

light source size

space between readings

Imaging Systems:

pixel size

dead area

Both:

light diffusion in sample

stray light

Page 48: Radiochromic Film Dosimetry

Five-Seed Five-Seed 192192Ir Ir Train at 0.5 mmTrain at 0.5 mm

Page 49: Radiochromic Film Dosimetry

Positional AccuracyPositional Accuracy

Moving Systems:stepping accuracybi-directional motion

Imaging Systems:regularity of imaging grid

Page 50: Radiochromic Film Dosimetry

Dimensional AccuracyDimensional Accuracy

Page 51: Radiochromic Film Dosimetry

A problemA problemwithwith

steppingstepping

Page 52: Radiochromic Film Dosimetry

Reader Bed ParametersReader Bed ParametersMoving Systems:maximum pixel size

transmission uniformity

Imaging Systems:light source uniformity

Both:film positioning

size

Page 53: Radiochromic Film Dosimetry

Scanning Densitometer Control GraphScanning Densitometer Control Graph

Page 54: Radiochromic Film Dosimetry

Net Reading of the Reference FilterNet Reading of the Reference Filter

Page 55: Radiochromic Film Dosimetry

Data Acquisition TimeData Acquisition TimeMoving Systems:

time to step between positionsBoth:

time to make measurement

data transfer time to host computer

Page 56: Radiochromic Film Dosimetry

SoftwareSoftwareflexibility

access to data

color images

iso-density contour plots

conversion to dose via stored calibration

Page 57: Radiochromic Film Dosimetry

Environmental FactorsEnvironmental Factors

temperature during readout

setup/interior lighting

Page 58: Radiochromic Film Dosimetry

Radiochromic Film Calibration ExposuresRadiochromic Film Calibration Exposures

Page 59: Radiochromic Film Dosimetry
Page 60: Radiochromic Film Dosimetry

Summary of Recommended Procedures:Summary of Recommended Procedures:Film HandlingFilm Handling

• visually inspect films prior to use• handle films with care, avoiding dust,

fingerprints or over bending• store film in dry, dark environment• avoid prolonged exposure to UV light

Page 61: Radiochromic Film Dosimetry

Summary of Recommended Procedures:Summary of Recommended Procedures:Film UseFilm Use

• note model and lot numbers• note film orientation and allignment• note emulsion side (check by wetting corner)• control time between irradiation & readout (at

least 24 h)• check film uniformity

Page 62: Radiochromic Film Dosimetry

Summary of Recommended Procedures:Summary of Recommended Procedures:ReadersReaders

• select densitometer with sufficient signal resolution

• determine maximum measurable OD• measure OD at red wavelengths for

maximum sensitivity

Page 63: Radiochromic Film Dosimetry

Summary of Recommended Procedures:Summary of Recommended Procedures:CalibrationCalibration

• use a large well-characterized field, ideally of the same quality as unknown field

• obtain response vs. dose over range of interest (extrapolation is dangerous)

Page 64: Radiochromic Film Dosimetry

Future DirectionsFuture Directions

• greater sensitivity• greater uniformity• better reading software• better photon energy dependence• 3D imaging with gels