Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Embed Size (px)

Citation preview

Page 1: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Volume I

Companion Presentation

Frank R. MielePegasus Lectures, Inc.

Ultrasound Physics & Instrumentation4th Edition

Page 2: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

License Agreement

This presentation is the sole property of Pegasus Lectures, Inc.

No part of this presentation may be copied or used for any purpose other than as part of the partnership program as described in the license agreement.

Materials within this presentation may not be used in any part or form outside of the partnership program. Failure to follow the license agreement is a violation of

Federal Copyright Law.

All Copyright Laws Apply.

Page 3: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Volume I Outline

Chapter 1: Mathematics

Chapter 2: Waves

Chapter 3: Attenuation

Chapter 4: Pulsed Wave

Chapter 5: Transducers

Level 1

Level 2

Chapter 6: System Operation

Page 4: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Chapter 5: Transducers - Level 2Level 2 focuses on the evolution of transducers, specific types of transducers, advantages and disadvantages of each type of transducer, and a review of resolution.

Page 5: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Calculating the Focal Depth (NZL)The natural focal depth (also referred to as the Near Zone Length (NZL)) can be calculated using the following equation:

2

4

DNZL

By substituting for the wavelength and assuming the propagation velocity of 1540 m/sec, this expression is approximated as:

2

0( ) ( )( )

6

D mm f MHzNZL mm

Page 6: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Basic Beam Characteristics

Fig. 18: (Pg 252)

By the approximated equation, it is now possible to calculate in your head the approximate natural focal depth.

Natural Focus

Fraunhoefer ZoneFresnel Zone

D/2 D

PZ

T C

ryst

al

D1

NZL = D2 • f0

6

2 • Near Zone Length

Page 7: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Effect of Diameter on Focal Depth

Fig. 19: (Pg 254)

In this example we see that doubling the diameter increases the focal depth by a factor of four.

D1

D2

NZL2

NZL1

D2/2 D2 D1D1/2

D1=2•D2

NZL1=22•NZL2=4•NZL2

Page 8: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Effect of Frequency on Focal Depth

Fig. 20: (Pg 254)

As suggested by the equation, a higher frequency produces a deeper focus; however, by design the focus is usually controlled by the diameter since higher frequency attenuates much faster.

D

NZL

DD/2

PZ

T C

ryst

alP

ZT

Cry

stal

Transmit Frequency = 2 • f0

D

NZL

DD/2

Transmit Frequency = 2 • f0

Page 9: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Depth of Field

Fig. 3: (Pg 236)

When a beam converges and diverges quickly, it has a very shallow depth of field. This yields a very good focus at one depth but poor focus in the relative near field and far field.

Shallow Depth of

Field

Broad Depth of

Field

Page 10: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Transducer and Imaging DimensionsThere are many different names used for the axial and lateral dimensions of the image, as listed below.

axial, depth, range, longitudinal

lateral, azimuthal, side-by-side, transverse

elevation

Fig. 22: (Pg 258)

Page 11: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Pedof (Blind Transducer)

Fig. 23: (Pg 259)

The pedof transducer is still in use today in both cardiac and vascular studies. The clear disadvantage is the inability to produce an image. The unexpected advantage is that these transducers usually are the most sensitive for Doppler.

Page 12: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Two “Pencil Probes”

(Pg 259)

The transducer on the left is a 5 MHz transducer used for vascular applications. The transducer on the right is a 1.9 MHz transducer used for cardiac applications.

Page 13: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Limitation of Pencil Probes: No Image

The greatest limitation to the pencil probe is the inability to create an image. The desire to create images lead to two parallel development paths:

sequencing: used to produce vascular images

mechanical steering: used to produce cardiac images

Page 14: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Sequencing

Fig. 24: (Pg 261)

Sequencing was performed with large arrays in a linear format (multiple elements in a straight line). By turning on and off switches, groups of elements were activated over time (in a sequence) to scan across the patient (as visualized in the animation on the next slide).

Page 15: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Sequencing (Animation)

(Pg 261)

Page 16: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Linear Switched Array

Fig. 25: (Pg 261)

Linear switched arrays are now obsolete, but the fundamental principle of sequencing is still used today in phased array linear transducers.

Page 17: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Mechanical Steering

Fig. 26: (Pg 263)

Mechanical steering was produced by mounting a single crystal on a motor. By “wobbling” the motor, the crystal was pointed in different directions over time, creating the ability to produce an image. (as visualized by the animation of the next slide)

Page 18: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Mechanical Steering (Animation)

(Pg 263)

Page 19: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Mechanical Sector Scan

Fig. 27: (Pg 264)

Sector images were produced for cardiac scans so as to provide rib access.

Eleva

tion

Axial

Lat

eral

Page 20: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Examples of Mechanical Transducers

Although the original design of mechanical transducers facilitated cardiac imaging, mechanical transducers were also designed for other applications, taking on a variety of form factors such as the endovaginal transducer shown below.

Page 21: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Limitation of Mechanical Transducers

Although there are many limitations to mechanical transducers, one of the largest issues was the fact that there was a “fixed” focus. In other words, there was no ability to vary the focus. This limitation lead to the design of the mechanically steered annular array transducer.

Page 22: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Mechanical Annular Array

Fig. 28: (Pg 265)

Annular arrays allow for a variable focus in both the lateral and elevation planes. As the name suggests, by creating an array of concentric element rings, the transducer diameter can be varied, varying the focal depth (as visualized in the animation of the next slide).

Activate All Rings Activate Inner Rings Activate Center Disc

Page 23: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Mechanical Annular Array (Animation)

(Pg 265)

Page 24: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Example of an Annular Array

(Pg 266)

Annular array transducers still were steered mechanically, and as such, did not appear significantly different than other mechanical transducers. However, the manufacturing of annular arrays was much more difficult and expensive. Furthermore, the complexity of the system increased to allow for control of multiple elements and to receive signals from more than one channel.

Page 25: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Limitations of Mechanical Transducers

Even with variable focus from annular arrays, the limitations to mechanical steering were significant and motivated the design of a new family of transducers which used electronic steering.

Page 26: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Electronic Control of Array Transducers

To overcome the many limitations of mechanical steering and fixed focus, electronic steering and focusing with arrays of elements was created.

Electronic control is produced by using small time delays (phase delays) between the excitation pulses which drive each element. By changing the delay profile (pattern of delays to a group of elements) different transmit steering angles and varying transmit focuses can be achieved. By also applying varying delay profiles for the received signals, receive steering and receive focus can be achieved.

Page 27: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Steering by PhasingBy using tiny time delay between the excitation pulses to each of the transducer elements, electronic beam steering can be achieved.

Fig. 30: (Pg 268)P

ha

se

De

lay

Pro

file

Beam Directio

n

Page 28: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Electronic Steering (Animation)

(Pg 269)

Page 29: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Receive Delay Profile

Fig. 31: (Pg 270)

Notice that the distance is different from the “red dot” labeled “X” to each of the elements labeled 1 through 8. As a result of the varying distances, the signal from the red dot arrives a little earlier at element 8 than element 7, which is earlier than element 6, etc. Therefore, for the signal to add up correctly from each of the individual elements, a delay must be applied with the greatest delay applied to element 8 and the least delay applied to element 1.

1 2 3 4 5 6 7 8

X

Page 30: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Focusing by PhasingCompare the two delay profiles and resulting wavelets from each transducer element.

No Focusing or Steering

Electronic Focusing

Fig. 32: (Pg 270)

Fig. 33: (Pg 271)

Page 31: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Electronic Focusing (Animation)

(Pg 271)

Page 32: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Simultaneous Steering and Focusing

Fig. 34: (Pg 272)

Steering and focusing simultaneously is obviously greatly desired. Quite simply, the steering delay profile is added to the focusing delay profile to achieve a steered and focused beam. This approach works for both the transmitted and receive beams.

Focus Profile Steer Profile

Page 33: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Electronic Steering and Focusing (Animation)

(Pg 272)

Page 34: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Sector Format

Fig. 35: (Pg 273)

A sector scan is produced by phasing. For each beam, a new phase delay profile is applied to steer both the transmitted and receive beam in a different direction. The sector format is acquired over time.

(Sector functionality is further explained and demonstrated in a few later slides and through an animation.)

Eleva

tion

Axial

Lat

eral

Page 35: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Sector Formatted Cardiac Image

(Pg 273)

Page 36: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Sector Transducers

(Pg 273)

The most often recognized form of a sector transducer format is the transducer designed primarily for cardiac imaging, The sector format is very useful for access through the ribs. The “fanning” out of the beams produces a broader far-field while the narrow near-field if the direct consequence of having to get between the ribs which would otherwise produce significant shadowing.

Page 37: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

TEE (sector format)

(Pg 273)

A transesophageal transducer also produces a sector formatted image.

Page 38: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Creating a Sector Scan

Fig. 36: (Pg 274)

A sector scan is created by phasing. A phase delay profile is produced to steer the beam in the desired direction and receive the resulting echoes. The delay profile is then changed to steer in a different direction, and the process repeated until the desired region is scanned (as visualized in the animation of the next slide).

Time 1 Time 2 Time 3

Time mid Time n

Page 39: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Sector Scan (Animation)

(Pg 274)

Page 40: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Varying Angles with Sector Formats

Fig. 37: (Pg 275)

Notice that for a “straight” vessel, the angle formed between the steered beam and the flow direction varies across the entire sector image. In this example, the angle on the left side of the image is less than 90 degrees. In the middle of the image, the angle equals 90 degrees. On the left side of the image, the angle is greater than 90 degrees.

Page 41: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Phased Array Linear Transducers

Fig. 38: (Pg 276)

For vascular applications, sector transducers are clearly suboptimal since there is such a narrow near field image. To overcome this drawback, phased array linear transducers were produced. These transducers can be used by sequencing alone (like the earlier switched linear arrays) or can in a more complex manner using both sequencing and phasing.

Page 42: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Unsteered Linear Image

(Pg 276)

Page 43: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Intraoperative Phased Linear Array

(Pg 276)

There are many different forms of linear transducers (including dimensions, number of elements, handle design, and operating frequency range) depending on the specific application. The transducer pictured here is an example of an intra-operative linear array. These transducers typically have significantly fewer elements than the larger arrays used for more “conventional” vascular imaging and tend to be relatively high frequency.

Page 44: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Linear Array Transducers

(Pg 276)

These two transducers are of the form most commonly seen for vascular applications such as cerebrovascular, arterial, and venous scans. Usually these transducers are designed to span a range of frequencies (broad bandwidth) for both easier and more challenging patients.

Page 45: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Creating an Unsteered Linear Scan

Fig. 39: (Pg 277)

Unsteered linear images are produced by sequencing. As shown earlier, sequencing is a method by which a group of elements are activated with a flat delay profile, producing a beam that transmits straight ahead. Once the echoes are received, another group of elements laterally displaced are activated, producing a parallel beam. This process repeats until the desired scan region is complete (as visualized in the animation of the next slide).

Page 46: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Unsteered Linear Scan Animation

(Pg 277)

Page 47: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Creating a Steered Linear Image

Fig. 40: (Pg 278)

Steered linear images are produced by sequencing and phasing simultaneously. The phasing is used to steer each beam to the desired angle and the sequencing is used to “traverse” across the patient. Notice that the delay profiles applied to each group of elements during the time intervals, T1, T2, T3, etc. is always the same. The result is that all of the beams of the image are parallel (as visualized in the animation on the slide after the next slide).

Page 48: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Steered Linear Image

Fig. 40: (Pg 278)

This image is actually comprised of two images. The black and white portion (the 2D or B-mode image) is not steered and was produced by sequencing alone. The color image is steered by setting the color box, and was produced by both phasing and sequencing.

(Color is applied – refer to picture in book.)

Page 49: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Steered Linear Image Animation

(Pg 278)

Page 50: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Trapezoidal (format) Scanning

Fig. 41: (Pg 279)

In order to produce a larger field of view, trapezoidal scanning was created. To create the trapezoid format, a group of elements are phased as if a sector transducer to produce the “wings”. Sequencing is then used to produce the unsteered middle part of the image (as visualized in the animation on the next slide).

Page 51: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Trapezoidal Scan Animation

(Pg 279)

Page 52: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Trapezoidal Scanning Example

(Pg 279)

* Color is applied – refer to picture in book.

Page 53: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Curved Linear Phased Array

Fig. 42: (Pg 280)

Page 54: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Curved Linear Image Format

(Pg 280)

Page 55: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Curved Linear Image Format

(Pg 280)

Page 56: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Curved Linear Array Transducers

(Pgs 280 - 281)

As with all phased array format types, curved linear arrays take many different forms as best suits the application. Transducers used on the abdomen are generally “relatively” large whereas probes that are more invasive are for obvious reasons physically much smaller.

Page 57: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Curved Linear Format

(Pg 281)

For a conventional 2-D image using a curved linear array, the scan is produced by sequencing only. Phasing can be used to affect the focus within the image or for Doppler and color Doppler steering. The curvature of the transducer face determines the curvature of the image.

Page 58: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

1.5-D Arrays

Fig 43: (Pg 281)

The 1.5-D array was the first electronic step towards controlling the focus in the elevation direction. Either the center elements alone could be used (shallower elevation focus) or both the center and outer set of elements could be used to make the elevation focus deeper. These were the precursor to the 2-D arrays.

Page 59: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

2-D Arrays

Fig 44: (Pg 282)

Two-dimensional arrays have multiple elements in both the lateral and elevation directions (2 dimensions). By electronically phasing these elements both steering and focusing can be achieved in both the lateral and elevation planes. The ability to steer electronically in the elevational direction allows for 3-D scanning.

Page 60: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

2-D Array Elements

Fig 45: (Pg 282)

This image shows how small the crystals are for the new matrix arrays that are being developed. The arrows in this picture indicate a human hair which is overlaid on the matrix. Notice that there are elements in two directions, giving control in both the elevation and lateral directions.

Page 61: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

2-D Array Posts

Fig. 46: (Pg 283)

This image shows how piezocomposite materials are constructed. These materials are a composite of PZT posts and a polymer. The polymer results in lower acoustic impedances which results in a better efficiency both into and out of the patient.

Page 62: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Lateral Resolution

Fig. 47: (Pg 284)

The lateral resolution of an image is determined by the lateral dimension of the beam. The beam must fit between two structures so as to not result in a combined echo. Therefore, the lateral resolution equals the beamwidth. Since the beamwidth changes with depth, the lateral resolution varies with the beam changes over depth.

Lateral Resolution Lateral Beamwidth

Page 63: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Elevation Resolution

Fig. 48: (Pg 284)

Resolution in the elevation direction is determined by the beam dimension elevationally. Like the lateral resolution, the elevation resolution is different at different depths and is best at the elevation focus.

Elevation Resolution Elevation Beamwidth

Page 64: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Axial Resolution

Fig. 49: (Pg 285)

Resolution in the axial direction is determined by the spatial pulse length. Because of the roundtrip effect, the resolution is actually better than the pulse length by a factor of 2. Recall that smaller numbers are always better for resolution.

S.P.L.Axial Resolution

2

Page 65: Pegasus Lectures, Inc. COPYRIGHT 2006 Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc. Ultrasound Physics & Instrumentation 4 th Edition

Pegasus Lectures, Inc.COPYRIGHT 2006

Notes: