ION570 User’s Guide D21M069

PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Embed Size (px)

Citation preview

Page 1: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Contents 1

ION570User’s Guide

D21M069

Page 2: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

2 Contents

D21M069 • Printed by Radiometer Analytical SAS • 2004-03C

Page 3: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Contents 3

ContentsContentsContentsContentsContents

1. Introduction ........................................................................ 5

Read me! ........................................................................................... 5

2. Getting started .................................................................... 11

Keyboard ............................................................................................ 11

Basic principles ................................................................................. 12

Main window ...................................................................................... 14

Reagents window ............................................................................... 15

Electrodes window ............................................................................. 17

Cell window ........................................................................................ 19

System configuration ......................................................................... 21

3. Programming guidelines ................................................... 29

Create an electrode ............................................................................ 30

Edit electrode screen ......................................................................... 32

Programming electrode parameters .................................................... 35

Create a reagent ................................................................................ 36

Edit reagent screen ............................................................................ 37

Create a method ................................................................................ 38

Edit method screen ............................................................................ 39

Programming method parameters ...................................................... 41

4. Running analyses ............................................................... 43

Run electrode calibration .................................................................... 44

Enter the reagent titre ........................................................................ 44

Run method ....................................................................................... 45

Run direct measurement .................................................................... 45

Run manual dosing ............................................................................ 46

Page 4: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

4 Contents

5. Viewing data ........................................................................ 47

Sample results ................................................................................... 48

Reagent titre entries ........................................................................... 49

Electrode calibration results ............................................................... 50

6. Printing data ....................................................................... 51

Manual printouts ................................................................................ 52

Automatic printouts ............................................................................ 54

7. Connection of peripherals ................................................. 55

8. General information ........................................................... 57

Cleaning ............................................................................................. 57

Transporting the instrument ................................................................ 57

Servicing ............................................................................................ 57

Page 5: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Introduction 5

1. Intr1. Intr1. Intr1. Intr1. Introductionoductionoductionoductionoduction

Read me!Read me!Read me!Read me!Read me!The interface of the Ion Analyser has been specially designed to clearlyguide you through every step of the programming and running of theanalyses, whether you are a supervisor or a routine user.

An important part of this interface is to check and control the presence ofdifferent elements necessary to run the defined application.

Working in Supervisor modeWorking in Supervisor modeWorking in Supervisor modeWorking in Supervisor modeWorking in Supervisor mode

When programming in “SUPERVISOR” mode, it is recommended towork in stages. These stages should be carried out in the order describedbelow:

To program a methodTo program a methodTo program a methodTo program a methodTo program a method

1. Define your electrode(s)Identify electrodes (including temperature sensors) to be used for theanalysis.

Electrodes can be created from the following lists, Catalogue, Other orCopy from. When creating the electrode, define if electrode calibration isrequired (or not), if yes specify the “periodicity” of the calibrations and thepH standards to be used. Refer to chapter 3.

2. Define reagentIdentify reagents to be used for the analysis.

Reagents can be created from the following lists, Catalogue, Other, Copyfrom.

3. Create a new method or Edit a pre-programmed oneCreate the measurement method to be used for the analyses. Enter theparameters required to calculate the results, refer to chapter 3.

When you have finished programming, select the method or pre-programmed application, refer to Short-Form Reminder no. 3.

If a sample changer is to be used, define the sample changer in theConfiguration menu before selecting Working mode = SACsequence in the main window.

Page 6: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

6 Chapter 1

4. Check icons

The following icons indicate the exact state of your working system.

Sunny icon:

Everything is OK. Run the method or sequence.

Sunny icons are required to run the method.

Cloudy icon:

Action required within 12 or 24 hours. For exampleelectrode calibration and/or reagent bottle exchange.

Stormy icon:

Electrode calibration date elapsed or electrode(s)/reagents(s) not installed.

Question mark:

It is a programming error, reagent and/or electrode is/arenot defined in the selected method. Revise the methodprogramming.

Refer to Reagent and Electrode windows.

When Cloudy/Stormy/Question mark icons appear, Press 1“Check”. The Ion Analyser will automatically guide you through theoperations necessary to solve the errors encountered.

5. Running methodsTo run a method or sequence, refer to chapter 4.

Page 7: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Introduction 7

Working in Routine modeWorking in Routine modeWorking in Routine modeWorking in Routine modeWorking in Routine mode

In “ROUTINE” mode you are guided at every step, thanks to the clear-text messages and the icons present on the large graphic display.

Access methodsAccess methodsAccess methodsAccess methodsAccess methods

A Routine operator has access to all the displays for checking purposes.

Running methodsRunning methodsRunning methodsRunning methodsRunning methods

When working in “ROUTINE” mode, it is necessary to install yourmeasurement system according to the selected method or sequence,prior to running a method or sequence.

1. Select the method or sequenceRefer to chapter 4.

Page 8: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

8 Chapter 1

2. Check icons

The following icons indicate the exact state of your working system.

Sunny icon:

Everything is OK. Run the method or sequence.

Sunny icons are required to run the method.

Cloudy icon:

Action required within 12 or 24 hours. For exampleelectrode calibration and/or reagent bottle exchange.

Stormy icon:

Electrode calibration date elapsed or electrode(s)/reagents(s) not installed.

Question mark:

It is a programming error, reagent and/or electrode is/arenot defined in the selected method. Revise the methodprogramming.

Refer to Reagent and Electrode windows.

When Cloudy/Stormy/Question mark icons appear, press 1 to runthe “Check” command.

Depending on the icons displayed, the Ion Analyser willautomatically guide you through the steps necessary to run theanalysis, see following page.

Page 9: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Introduction 9

a. Connect/install the electrode(s)

Connect/install electrodes and temperature sensors, Refer tochapter 3.

b. Install reagents(s)

Check that the burette is installed, if not, the Ion Analyser willautomatically guide you to the Install burette procedure.Press 3 to start. Refer to Short Form Reminder no. 2.

c. Calibrate electrode(s)

Run the electrode calibration, refer to chapter 4.

d. Run the method or the sequence, when Sunny icons aredisplayed,

Refer to Short Form Reminder no. 6 and 7.

Page 10: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

10 Chapter 1

Page 11: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Getting started 11

2. Getting star2. Getting star2. Getting star2. Getting star2. Getting startedtedtedtedted

KeyboardKeyboardKeyboardKeyboardKeyboard

1. Alphanumeric keypad to enter data and parameters on the sameprinciple as mobile telephones. These keys can also be used forquick access to the different menus; refer to the display on thefollowing page.

2. Del: deletes the character on which the cursor is positioned.

3. Esc: returns to the previous screen. During analysis, this key allowsyou to enter the method, electrode and reagent menu.

4. Print: prints the data concerning the screen displayed.

5. Stop: stops an analysis or a burette function. Press this key for 3seconds in the Main window to gain access to the setupparameters.

6. RIGHT, LEFT, UP, DOWN arrow keys are used to move to differentoptions within the menus.

7. : confirms a data entry, a message or a function asked for by theuser.

1

2

3

4

567

Page 12: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

12 Chapter 2

Basic principlesBasic principlesBasic principlesBasic principlesBasic principlesWhen the instrument is switched on, the Main window is displayed.When the instrument is switched on for the first time the screen will beas follows:

Title bar

Selected line

Menu options.

Press thecorresponding numberfor quick access to thedifferent menus. Forexample, press 4 toenter the Method library

Method tab Reagent tab Electrode tab Cell tab

If required, you can adjust the contrast of the display by:

• pressing 0 to increase the brightness,

• pressing 7 to decrease the brightness.

The title bar in the menu indicates the instrument name and the actualtime. You will be shown how to personalise the name (Station 1) andadjust the time further on in the manual.

The RIGHT and LEFT arrow keys allow you to move from one tab to theother and enter the Reagent, Electrode and Cell menus.

• Work your way through the 4 tabs, then back to the Main window.

The UP and DOWN keys allow you to select a line. To enter an option,select the line, and press . You can also press the correspondingnumerical key.

Page 13: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Getting started 13

1. Press 4 or select the line Method library and press to enter theMethod library screen. Press Esc to return to the Main window.

2. Select the line Working mode and press . The followingoptions are available:

Method: to run a single method.

Sequence: to create or run a sequence of methods. Beakers aremanually changed between two method runs.

SAC Method: to run a single method to be performed using asample changer.

SAC Sequence: to create or run a sequence of methods to beperformed using a sample changer.

Note: the working mode selected will have no effect on the type ofmethod you wish to create.

Do not forget to define the sample changer in the Configurationmenu before selecting SAC Method or SAC Sequence.

3 Press to confirm your selection.

Page 14: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

14 Chapter 2

Main windowMain windowMain windowMain windowMain windowWhen the instrument is switched on the Main window is displayed.

1. Title bar: indicates the instruments name and the current time.

2. Working mode: Select the way in which you want to work. Choosefrom Method, SAC Method, Sequence, SAC Sequence ; refer toprevious page.

3. Check or Run: check or run the selected method /sequence. Themethod can be run when 2 sunny icons are displayed in theReagents and Electrodes tabs. If cloudy/stormy icons aredisplayed, activate the "Check" command. The Ion Analyserwill automatically guide you through the necessary operationsrequired to solve the problem(s).

4. Sequence/sample stack: if SAC Sequence or Sac Methodhas been selected in step 2 above. Program the sample stack.Select or edit the method sequence.

5. Select method: select method to be used for the analyses.

6. Method library: supervisor use only: create, edit, reset and deletemethods to correspond to your specific needs.Or Display methods - routine use only : display the mainparameters of the selected method.

7. GLP-Archives: access GLP tables and visualise the stored methodsample results.

8. Method tab: Animated icon indicates when a method/sequence isrunning.

8

1

2

34567

Page 15: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Getting started 15

Reagents windowReagents windowReagents windowReagents windowReagents windowUse the RIGHT arrow key to move to this window.

1

2

45

7

9

3

8

1. Title bar: indicates the name of the window.

2. Enter titre: Enter the concentration of the titrant (titre).

3. Install reagents: install or replace reagents in a method orsequence.

4. Check reagents: verify the identification parameters of the reagentsused in the working method or sequence.

5. Reagent library- supervisor use only: create, edit, reset and deletereagents to correspond to your specific needs.

6. Manual dosing: delivers a reagent at a given speed and measuresa signal at a connected electrode.

7. GLP-Archives: access GLP tables and visualise the reagentconcentrations entered.

8. Burette functions: fill, empty, flush, rinse and replace burette.Global flush of all installed burettes simultaneously.

9. Reagent tab: Reagent status icon indicates the state of the reagentsystem. Four types of icons can be displayed.

6

Page 16: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

16 Chapter 2

Sunny icon. Everything is just right!

Cloudy icon. The expiry date of one of the reagents in thesystem will elapse in less than one week.

Stormy icon. The expiry date of one of the reagents in thesystem has elapsed.

At least one of the reagents present in the system has notbeen installed.

The reagent titre of one of the reagents in the system hasnot been entered (ISE methods only).

Check the sequence or method, (press 1 in the Mainwindow). The instrument prompts you to do the necessaryoperations. You are guided step by step.

Question mark. There is a problem in the editing of thereagent system. You need to be in Supervisor mode to solvethe problem. Check the sequence/methodparameters of the reagent or the electrode.

Check the sequence or method, (press 1 in the Mainwindow). The instrument indicates the possible errors andprompts you to correct them until ? disappears.

Page 17: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Getting started 17

Electrodes windowElectrodes windowElectrodes windowElectrodes windowElectrodes windowUse the RIGHT arrow key to move to this window.

1

2

4567

8

3

1. Title bar: indicates the name of the window.

2. Calibrate electrode: run a calibration or a calibration sequenceusing the installed electrodes.

3. Install electrodes: connect, disconnect or replace electrodes in amethod or sequence.

4. Check electrodes: check the identification parameters of theelectrodes used in the method or sequence.

5. Electrode library - Supervisor use only: create, edit, reset anddelete electrodes stored in the instrument.

6. Display measurement: displays mV and/or pH and/or concentrationand/or conductivity and/or temperature at a connected electrode ofthe electrode system.

7. GLP-Archives: access GLP tables and visualise the storedelectrode calibration results.

8. Electrode tab: Electrode status icon indicates the state of theelectrode system. Four types of icons can be displayed.

Page 18: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

18 Chapter 2

Sunny icon. The calibration has been performed on all theelectrodes present in the system. Everything is just right!

Cloudy icon. The electrode calibration of one of theelectrodes present in the system should be performed within12 or 24 hours.

Stormy icon. The calibration date has elapsed for one of theelectrodes present in the system.

If acceptance limits have been set for the calibration: atleast one calibration result lies outside the programmedacceptance limits.

At least one of the electrodes present in the system has notbeen installed.

Check the sequence or method, (press 1 in the Mainwindow). The instrument prompts you to do the necessaryoperations. You are guided step by step.

Question mark. There is a problem in the editing of theelectrode system. You need to be in Supervisor mode tosolve the problem. Check the sequence/methodparameters of the reagent or the electrode.

Check the sequence or method, (press 1 in the Mainwindow). The instrument indicates the possible errors andprompts you to correct them until ? disappears.

Page 19: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Getting started 19

Cell windowCell windowCell windowCell windowCell windowUse the RIGHT arrow key to move to this window.

1

2

5

3

1. Title bar: indicates the name of the window.

2. Instrument: to select the stirrer of the Ion Analyser or an ABU52connected to the Ion Analyser.

3. Internal stirring: command stirrer On/Off.

4. Speed: select the internal stirring speed, from 100 to 1100 rpm bysteps of 50 rpm.

5. Cell tab: animated icon indicates when the magnetic stirrer orpropeller is operating.

Apply internal stirringApply internal stirringApply internal stirringApply internal stirringApply internal stirring

1. For Instrument, select the stirrer of the Ion Analyser or thestirrer of an ABU52 connected to the on Analyser.

2. Select Internal stirring = ON,

3. Press in the field Speed and select a stirring speed.

4

Page 20: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

20 Chapter 2

Apply external stirringApply external stirringApply external stirringApply external stirringApply external stirring

1. Connect the Stirring Propeller, part no. 847-731, to the Ion Analyseror ABU52 Propeller socket. Line 3 is automatically replaced byExternal stirring.

2. For Instrument, select the stirrer of the Ion Analyser or thestirrer of an ABU52 connected to the Ion Analyser.

3. Select External stirring = ON.

4. Adjust stirring by turning the stirring propeller knob. You can consultthe corresponding table between the position (1 to 9) and the stirringspeed by pressing in the field Speed setting.

Page 21: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Getting started 21

System configurationSystem configurationSystem configurationSystem configurationSystem configurationProceed as follows to configurate your workstation

1. Press Stop for 3 seconds in the Main window to enter the Setupmenu.

2. Supervisor code:Entering a Supervisor code enables you to differentiatebetween the 2 operator modes: Routine and Supervisor:In Routine mode, the user is able to select and run methods. InSupervisor mode, the user can create, edit, select and runmethods. A Supervisor code is also used to protect yourparameters from any unwanted changes.

• Continue without entering a Supervisor code.

3. Press 1 to enter the Configuration menu.

The position of the bar indicates the first screen in theConfiguration menu. Use the RIGHT arrow keyto move to the next screen.

Horizontalscroll bar

Use the UP andDOWN arrowkeys to selectthe parameter.

Page 22: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

22 Chapter 2

4. Press the RIGHT arrow key.

The position of the bar indicates thelast screen in theConfiguration menu.

5. Press the UP arrow key.

6. Press the LEFT arrow key to return to the first screen in theConfiguration menu.

Page 23: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Getting started 23

Choosing the languageChoosing the languageChoosing the languageChoosing the languageChoosing the language

Choose your language for displays and printouts in the following way:

1. Use the UP and DOWN arrow keys to select the Language line.

2. Press to change a parameter as indicated at the bottom of thescreen.

3. Select the language.

Esc allows you to leave the screen without changing the language.

Page 24: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

24 Chapter 2

Setting the date and timeSetting the date and timeSetting the date and timeSetting the date and timeSetting the date and time

The current date and time are entered in the following displays:

1. Select Time.

Enter the hours (from 00 to 23).

2. RIGHT arrow key.

Enter the minutes (from 00 to 59).

3. RIGHT arrow key.

Enter the seconds (from 00 to 59).

The LEFT arrow key allows you to return to the previous screen tomodify an entered value.

4. Press to confirm (as indicated on the screen).

Page 25: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Getting started 25

5. Select Date.

Enter the day (from 00 to 31).

6. Press RIGHT arrow key.

7. Press RIGHT arrow key.

Enter the year (from 2000 to 2069).

8. Press .

Use the UP/DOWN keys to select themonth.

Page 26: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

26 Chapter 2

Customising the workstationCustomising the workstationCustomising the workstationCustomising the workstationCustomising the workstation

You can assign a name to your workstation, which will be permanentlydisplayed in the title bar of the Main window. Typing the name willallow you to get used to using the instruments keypad.

1. Before leaving the Configuration menu:

• Select PC keyboard,

• select English (US).

This allows you to use a QWERTY keyboard.

2. Press Esc to return to the SETUP menu.

3. Press 3 (Customise).

4. Press to select the Station parameter.

5. To replace "ION570" by "Chem.lab-1", proceed as follows:

• Press 7 until the letter "C" appears, then release the key. Thecursor moves to the next position.

Page 27: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Getting started 27

• Press 9 until the letter "h" appears.

• Continue until you have entered (em.lab-1).

• To correct a typing error, proceed as follows:

Press the LEFT arrow key to position the cursor on the letter"E". Press 7 to enter the letter "A".

The letter E has been inserted between the letter "A" and theletter "B".

Press Del to delete the "E".

• Press to confirm the entry.

7ABC

Page 28: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

28 Chapter 2

6. You can also use a PC keyboard to enter alphanumeric characters.

• Connect the PC keyboard to the 6-pin plug situated on the righthand side of the instrument.

• Select the line Radiometer Analytical. This line is usedto enter information concerning the workplace, user(s) name(s),location, address etc.

• Enter the text using the PC keyboard (maximum of 32 characterscan be used).If the characters shown on the display do not correspond to theones typed on the keyboard, redefine your keyboard type. To dothis, press Esc then 1 and select PC keyboard.

• Press to confirm.

You have now finished the getting started section. Press Esc to leave theCustomise screen, then press 5 to quit the SETUP menu.

Page 29: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Programming guidelines 29

3. Pr3. Pr3. Pr3. Pr3. Programming guidelinesogramming guidelinesogramming guidelinesogramming guidelinesogramming guidelinesOnly the Supervisor is allowed to program the menus, Electrode, Reagent,and Method.

To select the Supervisor mode, press Stop for 3 seconds in the Mainwindow and enter the Supervisor code. Press 5 (Exit) and select Return inmode = Supervisor.

IMPORTANT For first time users, it is recommended to program the instrument asfollows:

1. Create the electrode(s) to be used to perform the measurements.

2. Create the reagent(s) to be used during the analysis.

Then finally,

3. Create the method(s), which will consequently use the electrode(s)and reagent(s) created in the first two steps of programming.

Once you have finished programming, make sure that NO questionmarks "?" are displayed in the Reagent and Electrode tabs!

Page 30: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

30 Chapter 3

Create an electrodeCreate an electrodeCreate an electrodeCreate an electrodeCreate an electrode1. Press 4 Electrode Library, then 1 New electrode.

2. Press in the Function field and select the function of theelectrode, refer to the table below:

Electrode type Select function

pH single pH

pH combined w/wo temp. sensor pH

Single metal/redox mV (i=0)

Combined metal/redox w/wo temp. sensor mV (i=0)

ISE single ISE or mV (i=0)

ISE combined w/o temp. sensor ISE or mV (i=0)

Conductivity cell w/o temp. sensor Conductivity

Reference single Reference

Temperature sensor T°C

Ground metal Ground

Double metal mV (i > 0)

3. Press in the ID field.

4. Select Other in the From field.The option From = Catalogue allows you to create anelectrode from a list of Radiometer Analytical electrodes.

5. Enter the identification (ID) of the electrode, (up to 16 characters).

6. Press 1 twice.

Page 31: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Programming guidelines 31

7. Select the electrode Type, for electrodes with pH, mV or ISEfunctions:

• Single pH - pH electrode (no reference part),

• Combined pH - pH electrode with a reference part (with orwithout temperature sensor),

• Single metal/redox - mV (i=0) function electrode with a refer-ence part,

• Combined metal/redox - mV (i=0) function electrode with areference part (with or without temperature sensor),

• Single ISE - ISE or mV (i=0) function ion selective electrodewith no reference part,

• Combined ISE - ISE or mV (i=0) function ion selectiveelectrode with a reference part (with or without temperaturesensor).

8. When a combined electrode or a conductivity cell has beendefined, the instrument will prompt you to specify if it has a built-intemperature sensor or not.

When a single electrode has been defined, the instrument willprompt you to define a reference electrode in the Editelectrode screen.

9. When a combined or a reference electrode has been defined,enter the potential (in mV) of the reference element versus theStandard Hydrogen Electrode (SHE).This parameter enables to run calibration with automatic bufferrecognition when working with different reference electrode systems.

10. When a combined pH or a single pH electrodes has been defined,enter the internal pH of the electrode (pH int).

11. For ISE electrodes only; enter the ion valency and molar weight(in g/mol).

12. For conductivity cells only; enter the cable resistance (in Ω) andcapacitance (in pF).

13. Press 1 to confirm.

14. Edit the electrode parameters (see next page).

Page 32: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

32 Chapter 3

Edit electrode screenEdit electrode screenEdit electrode screenEdit electrode screenEdit electrode screen1. Press 4 Electrode library.

2. Press and select the electrode to be edited from the list.

3. Press 2 Edit electrode.

4. Edit the necessary parameters.

a. Electrode address: indicates which socket the electrode is con-nected to:

. ION / E1, E2, Ref, Temp, GND, Pt-Pt or EC (socket of the IonAnalyser),

. ABU1 / E1, E2, Ref, Temp, GND or Pt/Pt (socket of an ABU52connected to the Local socket of the Ion Analyser),

. ABU2 / E1, E2, Ref, Temp, GND or Pt/Pt (socket of an ABU52connected to the Slave socket of another ABU52 connected to theLocal socket of the Ion Analyser).

Page 33: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Programming guidelines 33

b. If you want a message to be displayed once a week upon starting amethod or a sequence of method using this electrode, selectMaintenance = Yes and enter the message (3 lines of 32characters maximum). This message can remind you to check or toclean an electrode.

c. For a pH electrode or a conductivity cell, select the calibrationmode:

- Calibration request = No: the electrode parameteredition is completed.

- Calibration request = Fixed : for a pH electrode,calibration with automatic recognition of buffers among a list ofpredefined values. For a conductivity cell, calibration with astandard that belongs to a list of predefined values.The buffer/standards are chosen during method edition.

- Calibration request = Free: the buffer/standardvalues are entered FREEly by the user. Use this option tocalibrate pH electrode or conductivity cells with buffers/standards that do not belong to the predefined list.

For an ISE electrode, select the calibration mode:

- Calibration = No: the electrode parameter edition iscompleted.

- Calibration = Manual : calibration using 1 to 9standards of known concentration.The standard concentrationsare entered during method edition. This method requires 1 to 9calibration beakers.

- Calibration = Automatic : 1 to 4-point calibrationmethod where the standards are automatically prepared by theIon Analyser using an automatic addition method. The standardconcentrations and an intial volume of standard are enteredduring method edition. This simple to use method requires onlyone calibration beaker.

Page 34: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

34 Chapter 3

d. Enter the Periodicity.

Indicates the maximum period of time between two calibrations. Ifthe period of time is exceeded:

- pH measurements can no longer be performed using this pHelectrode,

- concentration measurement using a Direct ISE method can nolonger be performed using this ISE electrode.

- conductivity measurements can no longer be performed usingthis conductivity cell.

In all cases, the electrode must be calibrated.

e. Enter the Number of cycles.

Enter the number of times the calibration should be repeated. Thisis the number of beakers to be prepared for a given standard. Foreach standard, the result will be calculated from the mean measuredvalue.

f. Enter the Number of buffers orNumber of solutions.

Enter the number of buffers (1 to 5 ) to be used for the pH electrodecalibration or enter the number of standards (1 to 9 in Manual modeor 1 to 4 in Auto mode) to be used for an ISE calibration.

g. Specify if a temperature sensor is in use.

If a temperature sensor is in use, select Probe.

If the temperature is entered manually via the numeric keypad,select Entered.

If the calibration is to be performed at 25°C, select Fixed 25°C.

If you have selected Temperature = Probe, a temperaturesensor must have been selected in the Parameters screen.

Page 35: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Programming guidelines 35

Programming electrode parametersProgramming electrode parametersProgramming electrode parametersProgramming electrode parametersProgramming electrode parameters1. If Calibrate request = yes, enter the electrode definition

parameters and go to the last Edit electrode screen.

2. Press 1:

To enter Calibration parameters: general parametersused during the calibration,

3. Press 2:

To enter Calibration solutions , Calibrationpoints or Standard parameters concerning the standardsused for the calibration.

4. Press 3:

To enter the Results parameters concerning acceptation criteriathat you can set on the results.

5. Press 4:

To enter Printouts parameters defining the calibration report tobe printed.

Page 36: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

36 Chapter 3

Create a reagentCreate a reagentCreate a reagentCreate a reagentCreate a reagent1. Press 4 Reagent library then 1 New reagent.

2. Press in the New reagent screen.

3. Select From = Other.The option From = Catalogue allows you to create areagent from a list of Radiometer Analytical reagents.

4. Press in the ID field and enter the reagent name (up to 16characters).

5. Enter the "approximate" titre of the reagent (5 characters) in theTarget titre field.

6. Enter the units (mM, M, mN, N) indicated on the reagent bottle. Usethe following: mM = mmol/l, M = mol/l, mN = meq/l, N = eq/l.

For info: once the units have been confirmed, they are added to thereagent ID and target titre.

7. Press 1 twice to create the reagent.

8. Edit the reagent parameters (see next page).

Page 37: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Programming guidelines 37

Edit reagent screenEdit reagent screenEdit reagent screenEdit reagent screenEdit reagent screen1. Press 4 Reagent library.

2. Press and select the reagent to be edited from the list.

3. Press 2 Edit reagent.

4. Edit the necessary parameters.

a. Modify the units (mM, M, mN, N) indicated on the reagent bottle. Usethe following: mM = mmol/l, M = mol/l, mN = meq/l, N = eq/l.

b. Address. Indicates the location of the burette containing the reagent:

. ION/1,

. ABU1/1 or ABU1/2 (position 1 or 2 of an ABU52 connected to theLocal socket of the Ion Analyser),

. ABU2/1 or ABU2/2 (position 1 or 2 of an ABU52 connected to theSlave socket of another ABU52 connected to the Local socket of theIon Analyser).

Page 38: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

38 Chapter 3

Create a methodCreate a methodCreate a methodCreate a methodCreate a method1. Press and select Method.

2. Press 4 Method library then 1 New method.

3. Press in the ID field.

4. Enter a method name (up to 16 characters).

5. Press 1.

6. Edit the method parameters (see the next page).

Page 39: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Programming guidelines 39

Edit method screenEdit method screenEdit method screenEdit method screenEdit method screen1. Press 4 Method library.

2. Press , select the method to be edited from the list then press 2Edit method.

3. Edit the necessary parameters.

a. Select the method mode. The modes available are:

• Measurement (pH or mV, ISE direct, ISE standard addition orconductivity)

• Coupled (methods chained in a same beaker)

• Addition (1 to 3 additions of reagent. No measurements. Additionsare performed simultaneously or successively)

Page 40: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

40 Chapter 3

b. Select the measurement mode: pH, mV, mV(i>0), ISE direct, ISEstandard addition, conductivityor EC/pH (pH and conductivitymeasurements performed simultaneously).

c. Select the way the temperature is to be measured.

• Probe: temperature sensor is required during measurements.

• Entered: temperature is to be entered manually using thekeypad.

• Fixed 25°C: measurements are assumed to be performedat 25°C.

d. Enter the Number of tests, i.e. the number of times themethod is to be repeated.

e. If you want a message to be displayed upon starting the method,select Notification = Yes then enter the message (3lines of 32 characters maximum).

f. Select the way the cell grounding of the measuring electrode will beperformed:

• Reference: reference electrode connected to the Refsocket on the measurement system.

• Metal: metal electrode connected to the GND socket on themeasurement system. For measurements in very resistantmedium.

• Other: an electrode which does not belong to the electrodesystem.

g. QC sample: the concentration of a QC sample is known accuratelyand the composition is close as possible to that of the samples tobe analysed. QC measurements enable you to perform qualitycontrol tests on the method used. Refer to the Reference Manual formore information concerning this option. Select Yes for QCsample to perform measurements on QC samples.

Page 41: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Programming guidelines 41

Programming method parametersProgramming method parametersProgramming method parametersProgramming method parametersProgramming method parameters1. Go to the last Edit method screen.

2. Press 1:To enter the Method parameters: general parametersconcerning the electrode and reagent (when required) used by themethod.

3. Press 2:To enter the Sample parameters: concerning the preparation ofthe sample to be analysed (only available for an ISE standardaddition method).

4. Press 3:

To enter the Results parameters defining the results and theacceptation criteria that you can set on the results.

5. Press 4:

To enter the Printouts: parameters defining the analysis reportto be printed.

6. Press 5:

To enter the QC data (only available when QC sample = Yes):quality control parameters required when using QC samples.

Page 42: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

42 Chapter 3

Page 43: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Running analyses 43

4. Running analyses4. Running analyses4. Running analyses4. Running analyses4. Running analyses

IMPORTANT Before pressing the Run key, check the following points listed below:

1. Select the method by pressing 3 in the main window.

2. If Question marks "?" are present in the electrode and/or reagenttabs. Programming error due to missing electrode/reagentparameters. Contact your supervisor.

3. Connect/Check electrode(s). Refer to Short-Form Reminder no. 1.

When installing/checking an electrode system, do not forget toconnect all the installed electrodes to the ION570 or ABU52sockets and to install these electrodes on the ION570 (Location =ION) or the SAC80 or SAC90 Sample Changer (Location = SAC).

It is important to immerse, in the same measuring beaker, all theinstalled electrodes having the same location.

For example: conductivity and pH measurements.If your installed electrode system comprises either a conductivitycell or a double platinum electrode with a combined pH electrode, theelectrodes must remain permanently immersed in the same beaker,in order to measure the pH correctly. This is because it is theconductivity cell (or the double platinum electrode) that provides theconnection to the ground and not the reference electrode.This so-called " 3-electrode or differential measurement configuration" will stop the conductivity cell altering the measurement at the pHelectrode.When using this configuration for an electrode calibration, theconductivity cell and the pH electrode must always be immersed inthe same beaker.

Important: Never remove an electrode or an electrode systemwithout running the Installation/checking procedure first. Theinstrument needs to know its configuration at any time.

For more information, refer to the ION570 Reference Manual,keyword index: "Electrode connection - Important".

4. Connect/Check reagent(s). Refer to Short-Form Reminder no. 2.

Page 44: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

44 Chapter 4

5. Run the analysis when Sunny icons are visible in the electrode andreagent tabs.

6. Run an electrode calibration when a Stormy icon appears in theelectrode tab.

7. Enter the reagent titer when a Stormy icon appears in the reagenttab.

8. If you are unable to display the "Run" command due to the presenceof Cloudy/Stormy icons in the Reagent and Electrode windows,activate the "Check" command. The Ion Analyser willautomatically guide you through the necessary operations requiredto solve the problem(s).

Run electrode calibrationRun electrode calibrationRun electrode calibrationRun electrode calibrationRun electrode calibrationRefer to Short-Form Reminder no. 4.

1. Select the method.

2. Install the electrode group.

3. Enter the Electrodes window.

4. Select 1 Calibrate electrode. Press to select theelectrode to be calibrated from the proposed list.

5. Press 1 to start the calibration.

Enter the reagent titreEnter the reagent titreEnter the reagent titreEnter the reagent titreEnter the reagent titreThe entry of a reagent titre is only required when this reagent is used as astandard in an ISE standard addition method.

1. Select the method using the reagent you want to enter the titer.

2. Enter the Reagents window.

3. Select 1 Enter titre. Press and select thereagent from the proposed list.

4. Press 1, enter user ID, press 1 to confirm.

5. Enter the titre value in the Titre line.

6. Enter the reagent calibration certificate number (up to 10characters).

7. Press 1 to confirm.

Page 45: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Running analyses 45

Run methodRun methodRun methodRun methodRun methodRefer to Short-Form Reminder no. 5.

1. Select the method in the Method library.

2. Install the electrode then the reagent groups to be used to performthe analysis.

3. Display the Main window.

4. Press 1 to start the analysis.

Run direct measurementRun direct measurementRun direct measurementRun direct measurementRun direct measurementRefer to Short-Form Reminder no. 7.

1. Select the method.

2. Install the electrode group to be used to perform the measurement.

3. Prepare the sample solution and place on the sample stand.

4. Display the Electrodes window.

5. Press 5 to start the measurement.

6. Select the electrode in the list of connected electrodes. You canstop or start stirring by pressing 1.

Page 46: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

46 Chapter 4

Run manual dosingRun manual dosingRun manual dosingRun manual dosingRun manual dosingRefer to Short-Form Reminder no. 9.

1. Select a method.

2. Install the electrode group to be used to perform the measurement(if measurements are to be done).

3. Install the reagent group to be used to perform the manual dosing.

4. Prepare the sample solution and place on the sample stand.

5. Display the Reagents window and press 5.

6. Select the reagent in the list of installed reagents and select theelectrode in the list of connected electrodes (if measurements are tobe done).

7. Enter an increment size, increment speed and maximum volume tobe dosed.

8. Select the type of measurement to be performed (mV, pH or nomeasurement).

9. Press 1 to start.

Page 47: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Viewing data 47

5. V5. V5. V5. V5. Viewing dataiewing dataiewing dataiewing dataiewing data

How to access?How to access?How to access?How to access?How to access?

The GLP-Archives (Good Laboratory Practice) command is available ineach window:

• Main: press 5 to access,

• Reagents: press 6 to access,

Electrodes: press 6 to access.

The GLP-Archives window displays the method, reagent orelectrode ID with the expiry date (reagent) and the installation date(electrode).

Press and enter the resultnumber from where you want tostart visualising the results.Example: result no. 9 out of 26results stored for the methodselected.

Press 1 to consult.The following pages give asummary of the informationavailable in the Main,Electrodes andReagents windows.

Page 48: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

48 Chapter 5

Sample rSample rSample rSample rSample resultsesultsesultsesultsesultsThe last 200 sample results are saved in the archives.Once you have selected the method and the result (see previous page),the following data are displayed:

Result no. 9 / 26 Result no. 10 / 26

1

Scroll the results.The results can be: Test results (Rx) or Average ± standard deviation

calculated on several tests of the same sample. "ION" or "SAC" means that anION or a SAC method has been run. "CPL, ION" or "CPL, SAC" means that the

method run belongs to an ION or SAC coupled method.

Mean and standard deviationcalculated on three R1accepted test results.

1

R1 = result of test no. 3 (testaccepted).

26 isthenumberofresultssavedfor theselectedmethod.

Pt-PtE1/E2EC

Page 49: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Viewing data 49

Reagent titre entriesReagent titre entriesReagent titre entriesReagent titre entriesReagent titre entriesThe last 100 reagent titre entries are saved in the archives.Once you have selected the reagent then the result (see first page of thischapter), the following data are displayed:

4 is thenumberof resultssavedfor theselectedreagent.

Result no. 1 / 4 Result no. 2 / 4

Scroll the results.The results can be: reagent titre entered by the operator or the reagent titre

currently used.

Double metal electrodeOther electrode typesConductivity cells

Page 50: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

50 Chapter 5

1

Result no. 13 / 19 Result no. 14 / 19

1

Electrode calibration resultsElectrode calibration resultsElectrode calibration resultsElectrode calibration resultsElectrode calibration resultsThe last 100 electrode calibration results are saved in the archives.Once you have selected the electrode then the result (see first page ofthis chapter), the following data are displayed (for a pH calibration):

Mean calculated on 3 cyclesperformed and accepted. Eachcycle has been run using 2 buffersolutions.

Result of cycle no. 3 (cycleaccepted). This cycle has beenrun using 2 buffer solutions.

19 isthenumberofresultssavedfor theselectedelectrode.

Scroll the results.The results can be: cycle performed on a standard, mean calculated on several

cycles performed on the same standard or current calibration data. "ION" or"SAC" means that an ION or a SAC method has been run.

Page 51: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Printing data 51

6. Printing data6. Printing data6. Printing data6. Printing data6. Printing data

IMPORTANT

Before printing, it is necessary to perform the following:

1. Enter the Setup window: press Stop 3 seconds in the Mainwindow.

2. Select the Printer

Press 1.In Configuration, select Printer = 80 columns andFormat = Listing or Page by Page.

3. Enter User ID (if required)

In Configuration, select User ID = Yes. You will beprompted to enter a user ID at the start of a run method. This ID willappear on the printouts.

4. Customise the printout (if required)

In Configuration, press 3.In Customise, enter the name of your workstation (max. 4 linesof 32 characters). This information will appear as a header at thestart of the printout.

5. For automatic printout - select a condensed or detailed printout

In the Printouts screen of the Edit method/electrode, selectDetailed = High to obtain a full detailed printout.Select Low for condensed printout.

Page 52: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

52 Chapter 6

Manual printoutsManual printoutsManual printoutsManual printoutsManual printouts

Method libraryMethod libraryMethod libraryMethod libraryMethod library

Press Print, in the Main window to give you an overview of the methodsavailable in the method list.

Reagent libraryReagent libraryReagent libraryReagent libraryReagent library

Press Print, in the Reagent window to give you an overview of thereagents available in the reagent list.

Electrode libraryElectrode libraryElectrode libraryElectrode libraryElectrode library

Press Print, in the Electrode window to give you an overview of theelectrodes available in the electrode list.

Page 53: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Printing data 53

Edit method dataEdit method dataEdit method dataEdit method dataEdit method data

In the Main window, press 4 Method library then Print to give youan overview of the parameters of the current programmed method. Theseare the parameters entered in the Edit method screen (press 2 toaccess).

Edit reagent dataEdit reagent dataEdit reagent dataEdit reagent dataEdit reagent data

Press 4 Reagent library in the Reagent window then Print togive you an overview of the parameters of the current programmedreagent. These are the parameters entered in the Edit reagentscreen (press 2 to access).

Edit electrode dataEdit electrode dataEdit electrode dataEdit electrode dataEdit electrode data

Press 4 Electrode library in the Electrode window thenPrint to give you an overview of the parameters of the currentprogrammed electrode. These are the parameters entered in the Editelectrode screen (press 2 to access).

Page 54: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

54 Chapter 6

Header: information enteredin the Customise screenof the Setup window.

Title of report:entered inPrintoutsscreen duringmethod editing

Analysis ID: User ID and SampleID entered at the start of themeasurement (if option selectedduring editing)

Calibration data: of the electrodeused

Result: obtained at the end ofthe analysis

Foot note: automatically appearsat the end of all printouts

Automatic printoutsAutomatic printoutsAutomatic printoutsAutomatic printoutsAutomatic printoutsThe Print key is inactive during a sample measurement or anelectrode calibration.

The results obtained during a "Run" are printed automatically.

Depending on the option selected for Detailed in the Edit method/electrode - Printout window, you will obtain different types of printouts.

Example of a printout for a pH measurement method

Page 55: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

Connection of peripherals 55

7. Connection of peripherals7. Connection of peripherals7. Connection of peripherals7. Connection of peripherals7. Connection of peripherals

Double metal electrodes

Temperature sensors

Reference electrodes

Single metal electrodes for grounding only

PC, cable A95X501Printer, cables A95P201/A95X506

Combined pH electrodesw/wo temperature sensor

Single pH electrodes

Single or combined metalelectrodes (Redoxmeasurements)

Single or combined ionselective electrodes

Not usedSample Changer, cableA95A202

Line cord, 230 V (A95S001),115 V (A95S002)

ABU52, cable A95X501

Conductivity cell

Page 56: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

56 Chapter 7

Page 57: PDF Contents gb - Frank's Hospital Workshopfrankshospitalworkshop.com/equipment/documents/laboratory...Contents 3 Contents 1. Introduction 5 Read me! 5 2. Getting started 11

General information 57

8. General infor8. General infor8. General infor8. General infor8. General informationmationmationmationmation

CleaningCleaningCleaningCleaningCleaningThe instrument requires minimum maintenance. The exterior surface canbe cleaned with tepid water and wiped dry with a soft cloth. Never useanother solvent unless you have first consulted yourRadiometer Analytical representative

TTTTTransporransporransporransporransporting the instrumentting the instrumentting the instrumentting the instrumentting the instrumentAlways use the packaging supplied by the manufacturer.

IMPORTANT!

Remove the metal rod before transporting the instrument.

Never pick-up or carry the instrument by the metal rod.

ServicingServicingServicingServicingServicingDO NOT ATTEMPT TO SERVICE THIS PRODUCT YOURSELF,except as noted in the Reference Manual. For servicing, please contactyour Radiometer Analytical service representative at the address givenbelow:

RADIOMETER ANALYTICAL SAS72, rue d'Alsace69627 Villeurbanne CEDEX - France

Tel.: +33 (0) 4 78 03 38 38Fax: +33 (0) 4 78 03 38 27E-mail: [email protected]

or your local service representative:

_________________________________

_________________________________

_________________________________

_________________________________

_________________________________