Owner’s Manual -For- Stabil-Air Cushions Dynamic Air Flow Cushion Users: Read instructions before using product and save this manual for future reference

Owner’s Manual -For- Stabil-Air Cushions€¦ · provide skin/soft tissue protection, positioning, and an environment to facilitate wound healing. Stabil-Air (2” or 5 cm) Dual

  • Upload
    others

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Owner’s Manual -For- Stabil-Air Cushions€¦ · provide skin/soft tissue protection, positioning, and an environment to facilitate wound healing. Stabil-Air (2” or 5 cm) Dual

14

Manufacturer:StarCushionProducts,Inc.

5CommerceDriveFreeburg,IL62243

Phone:(618)539-7070TollFree:(888)277-7827

Fax:(618)[email protected]

TheQualityManagementSystemofStarCushionProducts,Inc.iscertifiedto

ISO9001:2008andISO13485:2003

EtacSupplyCenterABLånggatan12

SE33424AnderstorpSweden

Owner’sManual

-For-Stabil-AirCushions

DynamicAirFlowCushion

Users:Readinstructionsbeforeusingproductandsavethismanualforfuturereference

Page 2: Owner’s Manual -For- Stabil-Air Cushions€¦ · provide skin/soft tissue protection, positioning, and an environment to facilitate wound healing. Stabil-Air (2” or 5 cm) Dual

AboutthismanualThismanualcontainsinformationontheoperationandmaintenanceofyourStarCushion.Youmustfullyreadtheseinstructionsbeforeuse.StarCushionProducts,Inc.isnotresponsibleforanynegativeresultsiftheinstructionsarenotfollowed.IfyouhaveanyadditionalquestionswithintheUSA,pleasecontactStarCushion.Internationally,pleasecontactyournationaldistributor.

TableofContentsPage1:AboutthisManual/TableofContents

Page2:IntendedUse

Page3-4:CustomerResponsibilities/Cautions

Page5:ProductSpecifications

Page6:WhatisIncluded

Page7-8:InstructionsforUse

Page9:CoverUseandRemoval

Page10-11:CleaningandDisinfectionInstructions

Page12:Warranty/Returns

Page13:Troubleshooting/Disposal

Page14:ContactInformation

ProductsinthismanualmaybecoveredbyoneormoreU.S.andforeignpatentsandtrademarks.Forfurtherdetails,pleaserefertowww.starcushion.com.StarCushionProductshasapolicyofcontinualproductimprovement,andthusmayamendthisdocument.Foracurrentversionofthisdocument,pleasevisitwww.starcushion.com.

1

IntendedUse:EachStarCushionproductthismanualisanairadjustable,cellulardesignwheelchairsupportsurfacewithcontouredfoaminternaltotheadjustableaircells.Thereisnoweightlimittothecushionssolongasthecushionisproperlysizedtotheindividual.StarCushionProducts,Inc.recommendsevaluationbyaclinicianorhealthcareprofessionalwhoisexperiencedinseating,positioningandmobilitytodeterminewhetherthecushionisappropriatefortheindividual,andtodeterminewhetherasolidseatplatformisrecommendedifusingthecushiononasling-seatwheelchair.Eachofthefollowingproductsutilizesdynamicairflowtechnologyandisintendedtoconformtoanindividual’sseatedshapeuponimmersiontoprovideskin/softtissueprotectionandanenvironmenttofacilitatewoundhealing.Stabil-Air(2”or5cm)SingleValveStabil-Air(3”or7.6cm)SingleValveStabil-Air(4”or10cm)SingleValveStabil-Air(5”or12.7cm)SingleValveEachofthefollowingproductsutilizesdynamicairflowtechnologyandisintendedtoconformtoanindividual’sseatedshapeuponimmersiontoprovideskin/softtissueprotection,positioning,andanenvironmenttofacilitatewoundhealing.Stabil-Air(2”or5cm)DualValveStabil-Air(2”or5cm)FourValveStabil-Air(3”or7.6cm)DualValveStabil-Air(3”or7.6cm)FourValveStabil-Air(4”or10cm)DualValveStabil-Air(4”or10cm)FourValveStabil-Air(5”or12.7cm)DualValveStabil-Air(5”or12.7cm)FourValve

2

1

Page 3: Owner’s Manual -For- Stabil-Air Cushions€¦ · provide skin/soft tissue protection, positioning, and an environment to facilitate wound healing. Stabil-Air (2” or 5 cm) Dual

AboutthismanualThismanualcontainsinformationontheoperationandmaintenanceofyourStarCushion.Youmustfullyreadtheseinstructionsbeforeuse.StarCushionProducts,Inc.isnotresponsibleforanynegativeresultsiftheinstructionsarenotfollowed.IfyouhaveanyadditionalquestionswithintheUSA,pleasecontactStarCushion.Internationally,pleasecontactyournationaldistributor.

TableofContentsPage1:AboutthisManual/TableofContents

Page2:IntendedUse

Page3-4:CustomerResponsibilities/Cautions

Page5:ProductSpecifications

Page6:WhatisIncluded

Page7-8:InstructionsforUse

Page9:CoverUseandRemoval

Page10-11:CleaningandDisinfectionInstructions

Page12:Warranty/Returns

Page13:Troubleshooting/Disposal

Page14:ContactInformation

ProductsinthismanualmaybecoveredbyoneormoreU.S.andforeignpatentsandtrademarks.Forfurtherdetails,pleaserefertowww.starcushion.com.StarCushionProductshasapolicyofcontinualproductimprovement,andthusmayamendthisdocument.Foracurrentversionofthisdocument,pleasevisitwww.starcushion.com.

1

IntendedUse:EachStarCushionproductthismanualisanairadjustable,cellulardesignwheelchairsupportsurfacewithcontouredfoaminternaltotheadjustableaircells.Thereisnoweightlimittothecushionssolongasthecushionisproperlysizedtotheindividual.StarCushionProducts,Inc.recommendsevaluationbyaclinicianorhealthcareprofessionalwhoisexperiencedinseating,positioningandmobilitytodeterminewhetherthecushionisappropriatefortheindividual,andtodeterminewhetherasolidseatplatformisrecommendedifusingthecushiononasling-seatwheelchair.Eachofthefollowingproductsutilizesdynamicairflowtechnologyandisintendedtoconformtoanindividual’sseatedshapeuponimmersiontoprovideskin/softtissueprotectionandanenvironmenttofacilitatewoundhealing.Stabil-Air(2”or5cm)SingleValveStabil-Air(3”or7.6cm)SingleValveStabil-Air(4”or10cm)SingleValveStabil-Air(5”or12.7cm)SingleValveEachofthefollowingproductsutilizesdynamicairflowtechnologyandisintendedtoconformtoanindividual’sseatedshapeuponimmersiontoprovideskin/softtissueprotection,positioning,andanenvironmenttofacilitatewoundhealing.Stabil-Air(2”or5cm)DualValveStabil-Air(2”or5cm)FourValveStabil-Air(3”or7.6cm)DualValveStabil-Air(3”or7.6cm)FourValveStabil-Air(4”or10cm)DualValveStabil-Air(4”or10cm)FourValveStabil-Air(5”or12.7cm)DualValveStabil-Air(5”or12.7cm)FourValve

2

2

Page 4: Owner’s Manual -For- Stabil-Air Cushions€¦ · provide skin/soft tissue protection, positioning, and an environment to facilitate wound healing. Stabil-Air (2” or 5 cm) Dual

3

CustomerResponsibilities:

- Carefullyreadandfollowalltheinstructionsprovidedinthismanual

- Followaregularroutineinexaminingandadjustingthecushionforproperinflationandimmersion,recommendedatleastdaily.

- Ifyouareunabletoperformanyofthetasksdescribedinthismanual,seekassistance

- Ifyouneedfurtherassistancewithunderstandingthetasksdescribedinthismanual,seekassistancefromaclinicianwhoisexperiencedinseating,positioningandmobility

Cautions: Checkinflation:Atleastonceperday. CheckSkin:Checkskinfrequently,atleastonceperday.Skinandsofttissuebreakdowncanoccurduetomanyreasons,whichmayvarybyindividual.Redness,bruising,ordiscoloredskinmayindicatethebeginningofsofttissueorskinbreakdown.Ifthisoccurs,consultwithaclinicianimmediately.

CushionandCoverMatching:Thecushionandcovermustbecompatiblesizesandmustbeusedasdirectedinthismanual(InstructionsforUse).Iftheyarenot,thebenefitsprovidedbythecushionmaybedecreasedoreliminated,whichincreasesrisktotheskinand/orsofttissue.Further,itmayincreasetheriskofinstabilityandfalling.

CushionandCoverOrientation:AStarCushionmustbeusedwiththeaircellsfacingup.Ifthecushionisnotfacingup,orthecoverisnotusedcorrectly,itmayreduceoreliminatethecushions’benefitsandcouldincreaserisktotheskinandothersofttissue.

CushionandUserOrientation:TheStabil-AirCushion,duetoitscontouredinternalfoam,cannotbeturnedwhiletheuserisseated.Thefrontofthecushion(markedfrontonboththecushionandthecover)mustbefacingforwardwhiletheuserisseated,andtherearofthecushion(markedbackonboththecushionandthecover)mustbeintherearwhiletheuserisseated.Ifnotpositionedtotheuserproperly,itwillcreatepeakpressureareas,whichwillreduceoreliminatetheproduct’sbenefits,andincreasetheriskofskinbreakdowntotheskinandothersofttissue.

3

Under-inflation:DONOTUSEanunder-inflatedcushion.Insufficientinflationorlossofcushionairmaycauselackofbodysupport.Itisimportantthatyouareimmersedintothecushion,butnotrestingonthebaseofthecushionorsolelyontheinternalfoambolsters.Sittingonthebase(inabottomoutsituation)orrestedsolelyontheinternalfoambolsterscanresultinpeakpressureareas.Prolongedexposuretopeakpressuresimpedesbloodflowandreducesessentialnourishmenttotheskin,increasingtheriskofpressureulcerdevelopment.

Over-inflation:DONOTUSEanover-inflatedcushion.Ifthecushionremainsover-inflatedfordailyuse,thebodywillnotimmerseintothecells,whichmayreduceoreliminatetheproduct’sbenefits,andincreaseriskofbreakdowntotheskinandothersofttissue.

Climate:Ifacushionhasbeenintemperatureslessthan32°F/0°Candexhibitsunusualstiffness,allowthecushiontowarmtoroomtemperature.Then,kneadthecushionwithyourhands,orrollupthecushionandthenunrollit(withthevalveopen)untiltheneopreneissoftandpliableagain.Repeatproperadjustmentinstructionsbeforeusing.Avoidextremeheatandkeepawayfromheatsources,hotashes,oropenflames.

Pressure:Changesinaltitudemayrequireyoutoreadjustyourcushionforproperinflation.Checkinflationeachtimeyouchangealtitude.

Puncture:Inordertopreventyourcushionfrombeingpunctured,keepyourcushionawayfromsharpobjects.

Handling:DONOTattempttocarryyourcushionbythevalve(s)orpullyourcushionbythevalve(s).Carryyourcushionbythecover’shandle,orliftunderneaththebaseofthecushiontocarry.

Obstructions:DONOTplaceanyobstructionsbetweentheuserandthecushion,asthismayreduceoreliminatetheproduct’sbenefitsandincreasetheriskofskinorsofttissuebreakdown.

Instability/FallHazard:DONOTuseyourcushioninconjunctionwith,orplaceontopofanyotherproductsormaterials,exceptaslistedintheIntendedUsesection.Thismayresultininstabilityand/orfalls.

4

Page 5: Owner’s Manual -For- Stabil-Air Cushions€¦ · provide skin/soft tissue protection, positioning, and an environment to facilitate wound healing. Stabil-Air (2” or 5 cm) Dual

4

CustomerResponsibilities:

- Carefullyreadandfollowalltheinstructionsprovidedinthismanual

- Followaregularroutineinexaminingandadjustingthecushionforproperinflationandimmersion,recommendedatleastdaily.

- Ifyouareunabletoperformanyofthetasksdescribedinthismanual,seekassistance

- Ifyouneedfurtherassistancewithunderstandingthetasksdescribedinthismanual,seekassistancefromaclinicianwhoisexperiencedinseating,positioningandmobility

Cautions: Checkinflation:Atleastonceperday. CheckSkin:Checkskinfrequently,atleastonceperday.Skinandsofttissuebreakdowncanoccurduetomanyreasons,whichmayvarybyindividual.Redness,bruising,ordiscoloredskinmayindicatethebeginningofsofttissueorskinbreakdown.Ifthisoccurs,consultwithaclinicianimmediately.

CushionandCoverMatching:Thecushionandcovermustbecompatiblesizesandmustbeusedasdirectedinthismanual(InstructionsforUse).Iftheyarenot,thebenefitsprovidedbythecushionmaybedecreasedoreliminated,whichincreasesrisktotheskinand/orsofttissue.Further,itmayincreasetheriskofinstabilityandfalling.

CushionandCoverOrientation:AStarCushionmustbeusedwiththeaircellsfacingup.Ifthecushionisnotfacingup,orthecoverisnotusedcorrectly,itmayreduceoreliminatethecushions’benefitsandcouldincreaserisktotheskinandothersofttissue.

CushionandUserOrientation:TheStabil-AirCushion,duetoitscontouredinternalfoam,cannotbeturnedwhiletheuserisseated.Thefrontofthecushion(markedfrontonboththecushionandthecover)mustbefacingforwardwhiletheuserisseated,andtherearofthecushion(markedbackonboththecushionandthecover)mustbeintherearwhiletheuserisseated.Ifnotpositionedtotheuserproperly,itwillcreatepeakpressureareas,whichwillreduceoreliminatetheproduct’sbenefits,andincreasetheriskofskinbreakdowntotheskinandothersofttissue.

3

Under-inflation:DONOTUSEanunder-inflatedcushion.Insufficientinflationorlossofcushionairmaycauselackofbodysupport.Itisimportantthatyouareimmersedintothecushion,butnotrestingonthebaseofthecushionorsolelyontheinternalfoambolsters.Sittingonthebase(inabottomoutsituation)orrestedsolelyontheinternalfoambolsterscanresultinpeakpressureareas.Prolongedexposuretopeakpressuresimpedesbloodflowandreducesessentialnourishmenttotheskin,increasingtheriskofpressureulcerdevelopment.

Over-inflation:DONOTUSEanover-inflatedcushion.Ifthecushionremainsover-inflatedfordailyuse,thebodywillnotimmerseintothecells,whichmayreduceoreliminatetheproduct’sbenefits,andincreaseriskofbreakdowntotheskinandothersofttissue.

Climate:Ifacushionhasbeenintemperatureslessthan32°F/0°Candexhibitsunusualstiffness,allowthecushiontowarmtoroomtemperature.Then,kneadthecushionwithyourhands,orrollupthecushionandthenunrollit(withthevalveopen)untiltheneopreneissoftandpliableagain.Repeatproperadjustmentinstructionsbeforeusing.Avoidextremeheatandkeepawayfromheatsources,hotashes,oropenflames.

Pressure:Changesinaltitudemayrequireyoutoreadjustyourcushionforproperinflation.Checkinflationeachtimeyouchangealtitude.

Puncture:Inordertopreventyourcushionfrombeingpunctured,keepyourcushionawayfromsharpobjects.

Handling:DONOTattempttocarryyourcushionbythevalve(s)orpullyourcushionbythevalve(s).Carryyourcushionbythecover’shandle,orliftunderneaththebaseofthecushiontocarry.

Obstructions:DONOTplaceanyobstructionsbetweentheuserandthecushion,asthismayreduceoreliminatetheproduct’sbenefitsandincreasetheriskofskinorsofttissuebreakdown.

Instability/FallHazard:DONOTuseyourcushioninconjunctionwith,orplaceontopofanyotherproductsormaterials,exceptaslistedintheIntendedUsesection.Thismayresultininstabilityand/orfalls.

4

Page 6: Owner’s Manual -For- Stabil-Air Cushions€¦ · provide skin/soft tissue protection, positioning, and an environment to facilitate wound healing. Stabil-Air (2” or 5 cm) Dual

ProductSpecifications:Thecushionsaddressedinthismanualaremadeoflatex-freeneoprenerubberwithaurethanebaseandelectroplatedcoppervalves.Thesecushions,whenusedandmaintainedproperly(includingminorrepairs),areexpectedtolast5years.

Cushion Height AverageWeight*Stabil-Air(single,dual,andfourvalve) 2”(5cm) 2.6lbs(1.18kg)Stabil-Air(single,dual,andfourvalve) 3”(7.6cm) 3.4lbs(1.54kg)Stabil-Air(single,dual,andfourvalve) 4”(10cm) 4.1lbs(1.86kg)Stabil-Air(single,dual,andfourvalve) 5”(12.7cm) 4.6lbs(2.09kg)*Note:Theaverageweightsarebasedon9x9cellscushions.Cells:2”(5cm),3”(7.6cm),and4”

(10cm)cushionsWidth/Depth

(inches)Width/Depth

(cm)6 11.75” 30cm7 13.5” 34cm8 15.5” 39cm9 17.25” 44cm10 19.25” 49cm11 21” 53cm12 22.75” 58cm13 24.75” 63cm14 26.5” 67cm

Cells:5”(12.7cm)cushions Width/Depth

(inches)Width/Depth

(cm)6 14” 36cm7 16” 41cm8 18” 46cm9 20” 51cm10 22” 56cm11 24” 61cm12 26” 66cm13 28” 71cm14 30” 76cm

Note:Abovechartsarenotlimited5

Whatisincluded:A–Cover:Coversaremadeofnylon,spandex,polyester,andanon-skid materialtoaidinpreventingthecushionfrommovingonthe surfaceB–Aircells:eachaircellisconnectedtoeachotherandwillallowairtobe distributedevenlyacrossthewholecushionC–InflationValve:supportstheflowofairintothecushionD-HandPump:dualactionpumptoinflatetheaircellsthroughthe inflationvalveE-RepairKit:Minorpuncturestoyourcushioncanbeeasilyrepairedwith therepairkit.Therepairkitcontains3patches,sandpaper,3 toothpicks,andaninstructioncard.F–Owner’sManual

6

5

Page 7: Owner’s Manual -For- Stabil-Air Cushions€¦ · provide skin/soft tissue protection, positioning, and an environment to facilitate wound healing. Stabil-Air (2” or 5 cm) Dual

ProductSpecifications:Thecushionsaddressedinthismanualaremadeoflatex-freeneoprenerubberwithaurethanebaseandelectroplatedcoppervalves.Thesecushions,whenusedandmaintainedproperly(includingminorrepairs),areexpectedtolast5years.

Cushion Height AverageWeight*Stabil-Air(single,dual,andfourvalve) 2”(5cm) 2.6lbs(1.18kg)Stabil-Air(single,dual,andfourvalve) 3”(7.6cm) 3.4lbs(1.54kg)Stabil-Air(single,dual,andfourvalve) 4”(10cm) 4.1lbs(1.86kg)Stabil-Air(single,dual,andfourvalve) 5”(12.7cm) 4.6lbs(2.09kg)*Note:Theaverageweightsarebasedon9x9cellscushions.Cells:2”(5cm),3”(7.6cm),and4”

(10cm)cushionsWidth/Depth

(inches)Width/Depth

(cm)6 11.75” 30cm7 13.5” 34cm8 15.5” 39cm9 17.25” 44cm10 19.25” 49cm11 21” 53cm12 22.75” 58cm13 24.75” 63cm14 26.5” 67cm

Cells:5”(12.7cm)cushions Width/Depth

(inches)Width/Depth

(cm)6 14” 36cm7 16” 41cm8 18” 46cm9 20” 51cm10 22” 56cm11 24” 61cm12 26” 66cm13 28” 71cm14 30” 76cm

Note:Abovechartsarenotlimited5

Whatisincluded:A–Cover:Coversaremadeofnylon,spandex,polyester,andanon-skid materialtoaidinpreventingthecushionfrommovingonthe surfaceB–Aircells:eachaircellisconnectedtoeachotherandwillallowairtobe distributedevenlyacrossthewholecushionC–InflationValve:supportstheflowofairintothecushionD-HandPump:dualactionpumptoinflatetheaircellsthroughthe inflationvalveE-RepairKit:Minorpuncturestoyourcushioncanbeeasilyrepairedwith therepairkit.Therepairkitcontains3patches,sandpaper,3 toothpicks,andaninstructioncard.F–Owner’sManual

6

6

Page 8: Owner’s Manual -For- Stabil-Air Cushions€¦ · provide skin/soft tissue protection, positioning, and an environment to facilitate wound healing. Stabil-Air (2” or 5 cm) Dual

InstructionsforUse:PleasereadtheinstructionscarefullyandfollowguidelinestoproperlyadjustyourStarCushion.Failuretocomplywiththeinstructionsmaycauseimproperinflationofthisdynamicairflowcushion,resultinginalossoftherapeuticvalue.InflatingtheCushion:

1. Productmustbeusedwiththeaircellsfacingup.2. Locatethecushions’airvalve(orvalves)andopenitbyturning

thesilvertipcounterclockwise.Donotapplypressurewhenopeningandclosingvalve.Thevalvemaintainsatightsealwithlittleeffort.

3. Attachtherubberhoseendoftheblackhandpumptotheopenvalve.Insurethepumphoseisplacedtightlyoverthevalve.

4. Withthepumpattached,pumptomoderateinflation.Note:Moderateinflationisachievedwhenthecenterofthecushionbeginstoarch.

PersonalizingtheInflationofYourCushion:

5. Removepumpfromcushionvalveandquicklyclosevalvetipbylightlyturningclockwise.

6. Withthevalveclosed,noairescapes.

7

7. Nowplaceyourcushiononthechairwiththeinflationvalvefacingthefront.Onasinglevalvecushion,thevalvewillbelocatedonthefront,leftcornerwhensittingonthecushion.TheStabil-Aircushionisalsomarked“Front”and“Back.”

Itisimportanttoalwayshavethevalveinthefront,leftcorner.DONOTturnyourcushion,astheinternalfoamiscontouredsoitfitstoyourbody(onlywhilefacingforward).Turningthecushioncancauseunwantedpressure.

8. Sitonthecushionandinthechairinapositionthatprovidesappropriatesittingposture.Ensurethatyourarmsandlegsarepositionedinanappropriateandcomfortableposition.

9. Nowplaceyourhandbetweenthecushionandyourboniestprominence,usuallyyourischials.

10. Openthevalveandbeginslowlylettingairescapefromthecushion.Asairescapes,useyourfingersasaguidetoestimatedistancebetweenyourboniestprominenceandthebaseofthecushion.

11. Continuereleasingairuntilyouarepositionedapproximately¾”(2cm)offofthebaseofthecushion.(seeimage)Atthispoint,gentlyclosethevalvebyturningtheoutersilvertipclockwisetotheclosedposition.

8

7

Page 9: Owner’s Manual -For- Stabil-Air Cushions€¦ · provide skin/soft tissue protection, positioning, and an environment to facilitate wound healing. Stabil-Air (2” or 5 cm) Dual

InstructionsforUse:PleasereadtheinstructionscarefullyandfollowguidelinestoproperlyadjustyourStarCushion.Failuretocomplywiththeinstructionsmaycauseimproperinflationofthisdynamicairflowcushion,resultinginalossoftherapeuticvalue.InflatingtheCushion:

1. Productmustbeusedwiththeaircellsfacingup.2. Locatethecushions’airvalve(orvalves)andopenitbyturning

thesilvertipcounterclockwise.Donotapplypressurewhenopeningandclosingvalve.Thevalvemaintainsatightsealwithlittleeffort.

3. Attachtherubberhoseendoftheblackhandpumptotheopenvalve.Insurethepumphoseisplacedtightlyoverthevalve.

4. Withthepumpattached,pumptomoderateinflation.Note:Moderateinflationisachievedwhenthecenterofthecushionbeginstoarch.

PersonalizingtheInflationofYourCushion:

5. Removepumpfromcushionvalveandquicklyclosevalvetipbylightlyturningclockwise.

6. Withthevalveclosed,noairescapes.

7

7. Nowplaceyourcushiononthechairwiththeinflationvalvefacingthefront.Onasinglevalvecushion,thevalvewillbelocatedonthefront,leftcornerwhensittingonthecushion.TheStabil-Aircushionisalsomarked“Front”and“Back.”

Itisimportanttoalwayshavethevalveinthefront,leftcorner.DONOTturnyourcushion,astheinternalfoamiscontouredsoitfitstoyourbody(onlywhilefacingforward).Turningthecushioncancauseunwantedpressure.

8. Sitonthecushionandinthechairinapositionthatprovidesappropriatesittingposture.Ensurethatyourarmsandlegsarepositionedinanappropriateandcomfortableposition.

9. Nowplaceyourhandbetweenthecushionandyourboniestprominence,usuallyyourischials.

10. Openthevalveandbeginslowlylettingairescapefromthecushion.Asairescapes,useyourfingersasaguidetoestimatedistancebetweenyourboniestprominenceandthebaseofthecushion.

11. Continuereleasingairuntilyouarepositionedapproximately¾”(2cm)offofthebaseofthecushion.(seeimage)Atthispoint,gentlyclosethevalvebyturningtheoutersilvertipclockwisetotheclosedposition.

88

Page 10: Owner’s Manual -For- Stabil-Air Cushions€¦ · provide skin/soft tissue protection, positioning, and an environment to facilitate wound healing. Stabil-Air (2” or 5 cm) Dual

CoverUseandRemoval:EveryStarCushioncomeswithafittedcovertohelppreventthecushioncellsfromexpandingbeyondthelisteddimensionsofthecushionandtoaidinslideboardtransfers.Thebottomofourcoversaremadeofnon-skidmaterial,withaddedVelcrostrips,tohelpkeepthecushionmoresecureinthechair.Tocoveryourcushion:

1.Setthecoveronaflatsurfacewiththenon-skidmaterialfacingupwardandtheVelcroflapfacingyou.2.OpentheVelcroflapandinsertyourcushionintothecoverwiththeaircellsdown(ifasingleordualvalvecushion,thevalveswillgointothecoverfirst).3.ClosetheVelcroflapandturnthecushionoversocellsarefacingup.4.Makesurethevalve(s)comeoutofthecorner(s)ofthecoverforinflation/deflation.Yourcushionisnowreadyforuseaslongasithasbeenproperlyadjusted(seeInstructionsforUseinthismanual).5.Placethecoveredcushiononthechair.Thenon-skidmaterialwillbeonthebottom,withallaircellscovered,andthevalve(s)willbeshowingthroughthecorner(s).Pleasereferencethelabeloncushiontoverifythecushionispositionedonthechaircorrectly.

Toremovethecover:UndoVelcroflapandremovethecushionfromthecover.ReattachVelcrostripsbeforecleaningorinspectingthecoverfordamage.Replaceifnecessary.

9

CoverCleaningandDisinfectingInstructions:

Beforewashing,closecoverwithVelcrostrip!ToCleantheCover:Removethecover.Machinewashinwarmwater(max.140*For60*C)withamilddetergent,withlikecolors.NoBleach.Rinseincoldwaterandmachinedryonlowtemperatureorlinedry.ToDisinfecttheCover:Thecovercanbemachinewashedinhotwater(60°C)todisinfect.Tumbledrylow.StarCushionProducts,Inc.isnotresponsibleforanydamagethatmayoccurfromyourwashingmachineordryer.

CushionCleaningandDisinfectingInstructions:Beforecleaning/disinfecting,removethecushioncoverandwashseparately.Cleaningistheremovalofsoil,dirtandorganicmatterfromtheproduct.Disinfectingistheeliminationofpathogenicmicroorganisms.Cleaningmustprecededisinfection.

ToCleantheCushion:Closethecushionvalve(s).Usingawashcloth,sponge,and/orsoftbristlebrush,handwashtheStarCushioninlukewarmwaterwithaliquiddishwashingdetergent,laundrydetergent,soap,orgeneralpurposehousekeepingcleaningproducts.(DONOTUSEautomaticdishwasherdetergentswhicharetoocausticforpersonalsafety)Ifnecessary,dilutefollowingtheproductlabeldirectionsforsurfacecleaning.Makesuretothoroughlyscruballoftheaircellsandspacesbetween.Rinsecompletelywithfreshwater.Allowthecushiontodryforapproximately15minutes.Ifnotcompletelydry,useatoweltospotdryyourcushion.DONOTPLACEYOURSTARCUSHIONIN ACLOTHESWASHERORDRYER,ORHEATABOVE100*FARENHEIT(37*C).

9

Page 11: Owner’s Manual -For- Stabil-Air Cushions€¦ · provide skin/soft tissue protection, positioning, and an environment to facilitate wound healing. Stabil-Air (2” or 5 cm) Dual

CoverUseandRemoval:EveryStarCushioncomeswithafittedcovertohelppreventthecushioncellsfromexpandingbeyondthelisteddimensionsofthecushionandtoaidinslideboardtransfers.Thebottomofourcoversaremadeofnon-skidmaterial,withaddedVelcrostrips,tohelpkeepthecushionmoresecureinthechair.Tocoveryourcushion:

1.Setthecoveronaflatsurfacewiththenon-skidmaterialfacingupwardandtheVelcroflapfacingyou.2.OpentheVelcroflapandinsertyourcushionintothecoverwiththeaircellsdown(ifasingleordualvalvecushion,thevalveswillgointothecoverfirst).3.ClosetheVelcroflapandturnthecushionoversocellsarefacingup.4.Makesurethevalve(s)comeoutofthecorner(s)ofthecoverforinflation/deflation.Yourcushionisnowreadyforuseaslongasithasbeenproperlyadjusted(seeInstructionsforUseinthismanual).5.Placethecoveredcushiononthechair.Thenon-skidmaterialwillbeonthebottom,withallaircellscovered,andthevalve(s)willbeshowingthroughthecorner(s).Pleasereferencethelabeloncushiontoverifythecushionispositionedonthechaircorrectly.

Toremovethecover:UndoVelcroflapandremovethecushionfromthecover.ReattachVelcrostripsbeforecleaningorinspectingthecoverfordamage.Replaceifnecessary.

9

CoverCleaningandDisinfectingInstructions:

Beforewashing,closecoverwithVelcrostrip!ToCleantheCover:Removethecover.Machinewashinwarmwater(max.140*For60*C)withamilddetergent,withlikecolors.NoBleach.Rinseincoldwaterandmachinedryonlowtemperatureorlinedry.ToDisinfecttheCover:Thecovercanbemachinewashedinhotwater(60°C)todisinfect.Tumbledrylow.StarCushionProducts,Inc.isnotresponsibleforanydamagethatmayoccurfromyourwashingmachineordryer.

CushionCleaningandDisinfectingInstructions:Beforecleaning/disinfecting,removethecushioncoverandwashseparately.Cleaningistheremovalofsoil,dirtandorganicmatterfromtheproduct.Disinfectingistheeliminationofpathogenicmicroorganisms.Cleaningmustprecededisinfection.

ToCleantheCushion:Closethecushionvalve(s).Usingawashcloth,sponge,and/orsoftbristlebrush,handwashtheStarCushioninlukewarmwaterwithaliquiddishwashingdetergent,laundrydetergent,soap,orgeneralpurposehousekeepingcleaningproducts.(DONOTUSEautomaticdishwasherdetergentswhicharetoocausticforpersonalsafety)Ifnecessary,dilutefollowingtheproductlabeldirectionsforsurfacecleaning.Makesuretothoroughlyscruballoftheaircellsandspacesbetween.Rinsecompletelywithfreshwater.Allowthecushiontodryforapproximately15minutes.Ifnotcompletelydry,useatoweltospotdryyourcushion.DONOTPLACEYOURSTARCUSHIONINACLOTHESWASHERORDRYER,ORHEATABOVE100*FARENHEIT(37*C).

10

10

Page 12: Owner’s Manual -For- Stabil-Air Cushions€¦ · provide skin/soft tissue protection, positioning, and an environment to facilitate wound healing. Stabil-Air (2” or 5 cm) Dual

10

ToDisinfecttheCushion:Periodiccleaningisallthatisrecommendedunlessthereisaknowninfectionoranopenwound.Ifdisinfectionisdesired,followtheabovecleaninginstructionsusingasolutionof1parthouseholdliquidchlorinebleachto9partswarmwater.Soaktheproductfor10minutesinthesolution,rinsecompletely,anddrythoroughly.DoNotUseUndilutedBleach!Note:Mostgermicidaldisinfectantsaresafeifusedfollowingthedisinfectantmanufacturer’sdilutiondirections.DONOT:

- machinewashormachinedrythecushion.- useabrasivessuchassteelwoolorscouringpads.- usecleaningproductscontainingpetroleumororganicsolvents.- allowwaterorcleaningsolutiontoentertheproduct(makesure

thevalve(s)areclosedtightlybeforestarting).- exposetheproducttoultravioletlightorozonegascleaning

methods.- useoil-basedlotions,lanolin,orphenolic-baseddisinfectantson

yourproduct,astheymaycompromisetheintegrityofthematerial.

ToSterilizetheCushion:FollowCleaningandDisinfectionguidelinesabove.StarCushion’sproductsarenotpackagedinasterileconditionnoraretheyintendedorrequiredtobesterilizedpriortousage.Hightemperaturescanaccelerateagingoftheproductandmaycausedamageorfailure.Ifinstitutionprotocoldemandssterilization,openthevalve(s)andusethelowesttemperaturepossiblefortheshortesttimepossible.Neitherarerecommended,butgassterilizationispreferabletosteamautoclave.Inanyprocessusingpressure,ensureallvalvesareopentopreventdamagetotheproduct.StarCushionProducts,Inc.highlydiscouragessterilizationmethodsusingtemperaturesgreaterthan180°F(82°C)forair-adjustableproducts.

Deodorizing:Todeodorizeproductsmixonetablespoonofbakingsodaperquartofwater(25gm/l)in10partswater.Soakforseveralhours,rinseandletdry.Thismayneedtoberepeatedseveraltimes.Cleanproductfrequentlyoruseanincontinentcoveringtoavoidstrongodorsandstaining.

11

Warranty:LimitedTwoYearWarrantyonCushion

LimitedThirtyDayWarrantyonCushionCover

StarCushionwarrantsyourcushionforaperiodoftwoyearsfromtheoriginaldateofpurchase,andthecushionexteriorcoverforthirtydaysfromtheoriginaldateofpurchasetobefreeofdefectsinmaterialand/orworkmanship.

WhatisNOTcoveredbythecushionorcushioncoverwarranty:1. Necessaryreplacementorrepairthathasbeencausedbyimproper

usage,abuse,negligence,orpuncturesfromoutsideobjects.2. Dailywearandtear,ordamagethatisaresultofdisregardingthespecific

directionscontainedinthisinstructionmanual.

Warrantorisnotresponsibleforconsequentialdamagesorincidentaldamages.

StarCushionrequiresproofoforiginalpurchasedate,suchasyouroriginalsalessliptoestablishwarrantyperiod.

Warrantyserviceisavailablebyreturningthecushionand/orcushioncover,atyourexpenseto:

StarCushionProducts,Inc. Attn:Repairs 5CommerceDrive Freeburg,IL62243

Somestatesdonotallowlimitationsonhowlonganimpliedwarrantylasts,sotheabovelimitationmaynotapplytoyou.Allimpliedwarrantieswithrespecttothisproductarelimitedtotwoyearsfromthedateofpurchaseforthecushionandthirty(30)daysforthecushioncover.Thiswarrantygivesyouspecificlegalrightsandyoumayalsohaveotherrightsthatmayvaryfromstatetostate.InternationalWarranty:Pleasecontactyourlocaldistributor.Alistofdistributorscanbefoundatwww.starcushion.comorifyoucannotfindadistributorcontactStarCushion(seeContactInformation).

Returns:Allproductsreturnedmustbenewandunused,requirepriorauthorizationfromStarCushionProducts,Inc.andaresubjecttoarestockingcharge.BeforereturningyourproductintheU.S.A.,contactourWarrantyServiceDepartment

11

Page 13: Owner’s Manual -For- Stabil-Air Cushions€¦ · provide skin/soft tissue protection, positioning, and an environment to facilitate wound healing. Stabil-Air (2” or 5 cm) Dual

10

ToDisinfecttheCushion:Periodiccleaningisallthatisrecommendedunlessthereisaknowninfectionoranopenwound.Ifdisinfectionisdesired,followtheabovecleaninginstructionsusingasolutionof1parthouseholdliquidchlorinebleachto9partswarmwater.Soaktheproductfor10minutesinthesolution,rinsecompletely,anddrythoroughly.DoNotUseUndilutedBleach!Note:Mostgermicidaldisinfectantsaresafeifusedfollowingthedisinfectantmanufacturer’sdilutiondirections.DONOT:

- machinewashormachinedrythecushion.- useabrasivessuchassteelwoolorscouringpads.- usecleaningproductscontainingpetroleumororganicsolvents.- allowwaterorcleaningsolutiontoentertheproduct(makesure

thevalve(s)areclosedtightlybeforestarting).- exposetheproducttoultravioletlightorozonegascleaning

methods.- useoil-basedlotions,lanolin,orphenolic-baseddisinfectantson

yourproduct,astheymaycompromisetheintegrityofthematerial.

ToSterilizetheCushion:FollowCleaningandDisinfectionguidelinesabove.StarCushion’sproductsarenotpackagedinasterileconditionnoraretheyintendedorrequiredtobesterilizedpriortousage.Hightemperaturescanaccelerateagingoftheproductandmaycausedamageorfailure.Ifinstitutionprotocoldemandssterilization,openthevalve(s)andusethelowesttemperaturepossiblefortheshortesttimepossible.Neitherarerecommended,butgassterilizationispreferabletosteamautoclave.Inanyprocessusingpressure,ensureallvalvesareopentopreventdamagetotheproduct.StarCushionProducts,Inc.highlydiscouragessterilizationmethodsusingtemperaturesgreaterthan180°F(82°C)forair-adjustableproducts.

Deodorizing:Todeodorizeproductsmixonetablespoonofbakingsodaperquartofwater(25gm/l)in10partswater.Soakforseveralhours,rinseandletdry.Thismayneedtoberepeatedseveraltimes.Cleanproductfrequentlyoruseanincontinentcoveringtoavoidstrongodorsandstaining.

11

Warranty:LimitedTwoYearWarrantyonCushion

LimitedThirtyDayWarrantyonCushionCover

StarCushionwarrantsyourcushionforaperiodoftwoyearsfromtheoriginaldateofpurchase,andthecushionexteriorcoverforthirtydaysfromtheoriginaldateofpurchasetobefreeofdefectsinmaterialand/orworkmanship.

WhatisNOTcoveredbythecushionorcushioncoverwarranty:1. Necessaryreplacementorrepairthathasbeencausedbyimproper

usage,abuse,negligence,orpuncturesfromoutsideobjects.2. Dailywearandtear,ordamagethatisaresultofdisregardingthespecific

directionscontainedinthisinstructionmanual.

Warrantorisnotresponsibleforconsequentialdamagesorincidentaldamages.

StarCushionrequiresproofoforiginalpurchasedate,suchasyouroriginalsalessliptoestablishwarrantyperiod.

Warrantyserviceisavailablebyreturningthecushionand/orcushioncover,atyourexpenseto:

StarCushionProducts,Inc. Attn:Repairs 5CommerceDrive Freeburg,IL62243

Somestatesdonotallowlimitationsonhowlonganimpliedwarrantylasts,sotheabovelimitationmaynotapplytoyou.Allimpliedwarrantieswithrespecttothisproductarelimitedtotwoyearsfromthedateofpurchaseforthecushionandthirty(30)daysforthecushioncover.Thiswarrantygivesyouspecificlegalrightsandyoumayalsohaveotherrightsthatmayvaryfromstatetostate.InternationalWarranty:Pleasecontactyourlocaldistributor.Alistofdistributorscanbefoundatwww.starcushion.comorifyoucannotfindadistributorcontactStarCushion(seeContactInformation).

Returns:Allproductsreturnedmustbenewandunused,requirepriorauthorizationfromStarCushionProducts,Inc.andaresubjecttoarestockingcharge.BeforereturningyourproductintheU.S.A.,contactourWarrantyServiceDepartment

12

Page 14: Owner’s Manual -For- Stabil-Air Cushions€¦ · provide skin/soft tissue protection, positioning, and an environment to facilitate wound healing. Stabil-Air (2” or 5 cm) Dual

12

Troubleshooting:NotHoldingAir:

1. Openvalve(s)byturningcounterclockwiseandaddairtothecushionusingyourblackhandpump.

2. Closevalve(s)tightlybyturningvalvesclockwise.Ifstillnotholdingair

3. Visiblychecktoseeifthereareanyholes.Ifnoholesarevisible,immersecushion(withthevalvestightlyclosed)insinkortubofwaterandlookforairbubbles.

- Ifpin-holesizedholesarefound,usetherepairkitthatcamewithyourproductandfollowtherepairkitinstructions.

- Iflargeholesorotherleaksarefound,seeWarrantySection.

Uncomfortable/Unstable:

- Makesurecushionisnotover-inflated.(seeInstructionsforUse).- Makesuretheaircellsandthecover’sstretchtoparefacing“UP”.- Makesurecushionisnottoobigortoosmallforthewheelchair.

Allowtheusertositonthecushionforaminimumofonehourtogetusetothetextureoftheproduct’ssurfaceandimmersionintotheaircells.

Disposal:Therearenoknownenvironmentalhazardsassociatedwiththecomponentsoftheproductsinthismanualwhenproperlyusedanddisposed.Disposeoftheproductand/orcomponentsinaccordancewiththeapplicableregulationsinyourjurisdiction.DONOTINCINERATE.

13

ContactInformation:IntheU.S.A:StarCushionProducts,Inc.CustomerCareandWarrantyService:Phone:888-277-7827Fax:618-539-7073www.starcushion.comCustomerCare/WarrantyService:[email protected]:Ifdirectedtoreturnanyitems,pleasesendto:StarCushionProducts,Inc.5CommerceDriveFreeburg,IL62243OutsidetheU.S.A.:StarCushionInternational:intl@starcushion.comOrvisitwww.starcushion.comforalistofStarCushionInternationaldistributors.

13

Page 15: Owner’s Manual -For- Stabil-Air Cushions€¦ · provide skin/soft tissue protection, positioning, and an environment to facilitate wound healing. Stabil-Air (2” or 5 cm) Dual

12

Troubleshooting:NotHoldingAir:

1. Openvalve(s)byturningcounterclockwiseandaddairtothecushionusingyourblackhandpump.

2. Closevalve(s)tightlybyturningvalvesclockwise.Ifstillnotholdingair

3. Visiblychecktoseeifthereareanyholes.Ifnoholesarevisible,immersecushion(withthevalvestightlyclosed)insinkortubofwaterandlookforairbubbles.

- Ifpin-holesizedholesarefound,usetherepairkitthatcamewithyourproductandfollowtherepairkitinstructions.

- Iflargeholesorotherleaksarefound,seeWarrantySection.

Uncomfortable/Unstable:

- Makesurecushionisnotover-inflated.(seeInstructionsforUse).- Makesuretheaircellsandthecover’sstretchtoparefacing“UP”.- Makesurecushionisnottoobigortoosmallforthewheelchair.

Allowtheusertositonthecushionforaminimumofonehourtogetusetothetextureoftheproduct’ssurfaceandimmersionintotheaircells.

Disposal:Therearenoknownenvironmentalhazardsassociatedwiththecomponentsoftheproductsinthismanualwhenproperlyusedanddisposed.Disposeoftheproductand/orcomponentsinaccordancewiththeapplicableregulationsinyourjurisdiction.DONOTINCINERATE.

13

ContactInformation:IntheU.S.A:StarCushionProducts,Inc.CustomerCareandWarrantyService:Phone:888-277-7827Fax:618-539-7073www.starcushion.comCustomerCare/WarrantyService:[email protected]:Ifdirectedtoreturnanyitems,pleasesendto:StarCushionProducts,Inc.5CommerceDriveFreeburg,IL62243OutsidetheU.S.A.:StarCushionInternational:intl@starcushion.comOrvisitwww.starcushion.comforalistofStarCushionInternationaldistributors.

14

Page 16: Owner’s Manual -For- Stabil-Air Cushions€¦ · provide skin/soft tissue protection, positioning, and an environment to facilitate wound healing. Stabil-Air (2” or 5 cm) Dual

Manufacturer:StarCushionProducts,Inc.

5CommerceDriveFreeburg,IL62243

Phone:(618)539-7070TollFree:(888)277-7827

Fax:(618)[email protected]

TheQualityManagementSystemofStarCushionProducts,Inc.iscertifiedto

ISO9001:2008andISO13485:2003

EtacSupplyCenterABLånggatan12

SE33424AnderstorpSweden

PleaseaddCElogoandDocument#QF71-04-1lastedited19May2016

Tolooklikethelastmanualyoudid

Owner’sManual

-For-Stabil-AirCushions

DynamicAirFlowCushion

Users:Readinstructionsbeforeusingproductandsavethismanualforfuturereference

Document #QF71-04-1 last edited 19 May 2016

14

Manufacturer:StarCushionProducts,Inc.

5CommerceDriveFreeburg,IL62243

Phone:(618)539-7070TollFree:(888)277-7827

Fax:(618)[email protected]

TheQualityManagementSystemofStarCushionProducts,Inc.iscertifiedto

ISO9001:2008andISO13485:2003

EtacSupplyCenterABLånggatan12

SE33424AnderstorpSweden

Owner’sManual

-For-Stabil-AirCushions

DynamicAirFlowCushion

Users:Readinstructionsbeforeusingproductandsavethismanualforfuturereference