94
Operating Instructions BioProTT™ Flow Measurement System

Operating Instructions · with supplied 24V DC power supply and connect analog or digital ... 7.2.1 Pin Assignment and Cable Colors for ... Read these Operating Instructions carefully

Embed Size (px)

Citation preview

Operating Instructions

BioProTT™

Flow Measurement System

BioProTT™ Flow Measurement System Version 2.0 page 2 of 94

Brief Operating Instructions - Quick Guide BioProTT™ Flow Measurement System

Safety instructions Chapter 2.1 – Page 6

Familiarize yourself with the safety instructions!

Setting up and connecting the flow meter Chapter 6 - Page 18

Connect the BioProTT™ flow meter to power supply, e.g.

with supplied 24V DC power supply and connect analog or

digital interfaces as needed.

Installing flow sensor Chapter 8.1 - Page 43

Connect Clamp-On Transducer to BioProTT™ flow meter

by plugging it into the correct socket.

Please ensure that transducer is proper size for the tubing.

Insert tubing into BioProTT™ Clamp-On Transducer and

close lid. The flow direction is indicated with an arrow.

Prepare Measurement Chapter 8.2 - Page 43

Power the BioProTT™ flow meter on and wait some

seconds for initialization.

Usually a sensor will show a flow value even with no flow

present. Zero this (flow value) offset by pressing the

appropriate button, selecting the appropriate menu option,

sending a digital command or using remote zero.

Note: ensure that

correct calibration table is selected (for sensors with

more than one table programmed)

tubing is filled with medium,

there is sufficient coupling and

the liquid in the tubing is not moving (no flow) before

adjusting the zero offset (flow value ml/min!)

Perform custom calibration if necessary

Perform Measurement Chapter 8.7 - Page 47

The BioProTT™ flow measurement system is now ready

for use.

Please note: If troubleshooting is necessary please see chapter TROUBLESHOOTING.

BioProTT™ Flow Measurement System Version 2.0 page 3 of 94

TABLE OF CONTENTS

1 INTENDED USE AND RESTRICTIONS ........................................................................................... 6

1.1 INTENDED USE ..................................................................................................................................... 6 1.2 USAGE RESTRICTIONS AND LIMITATIONS ............................................................................................ 6

2 GENERAL SAFETY AND IMPORTANT INFORMATION ........................................................... 6

2.1 GENERAL SAFETY INFORMATION ......................................................................................................... 6 2.2 GENERAL IMPORTANT INFORMATION .................................................................................................. 9

3 SYMBOLS, UNITS AND ABBREVIATIONS .................................................................................. 10

3.1 SYMBOLS USED IN THESE OPERATING INSTRUCTIONS ........................................................................ 10 3.2 SYMBOLS ON EXTERNAL POWER SUPPLY UNIT ................................................................................. 10 3.3 SYMBOLS ON FLOW METER, SENSORS AND ON PACKAGING .............................................................. 11 3.4 UNITS ................................................................................................................................................. 13 3.5 DEFINITIONS AND ABBREVIATIONS .................................................................................................... 13

4 DESCRIPTION OF MEASURING PRINCIPLE ............................................................................. 14

5 PACKAGING CONTENTS ................................................................................................................ 15

6 INITIAL SET UP ................................................................................................................................. 18

6.1 IMPORTANT CURRENT LOOP DEFINITIONS ......................................................................................... 18 6.2 POWERING OF BIOPROTT™ FLOW METER ........................................................................................ 20 6.3 POWERING OVER ANALOG SOCKETS ................................................................................................. 21 6.4 CURRENT LOOP(S) POWERED BY FLOW METER ................................................................................. 22 6.5 CURRENT LOOP(S) POWERED BY EXTERNAL SUPPLY, E.G. FROM PCS ............................................... 23

7 INSTALLATION AND OPERATION OF BIOPROTT™ FLOWTRACK ................................... 26

7.1 FRONT PANEL BIOPROTT™ FLOWTRACK ......................................................................................... 26 7.2 REAR PANEL BIOPROTT™ FLOWTRACK ........................................................................................... 27

7.2.1 Pin Assignment and Cable Colors for Round Connector for Digital Interface ....................... 28 7.2.2 Pin Assignment and Cable Colors for Round Connectors for Power, Flow and RSS Interfaces

28 7.2.3 Connecting and Powering up .................................................................................................. 29

7.3 INSTALLATION AND OPERATION OF BIOPROTT™ FLOWTRACK DINRAIL ......................................... 30 7.3.1 DIN Rail Mounting .................................................................................................................. 31 7.3.2 Installation of Flow Sensor Extension Cable (IP50 at panel) ................................................. 31 7.3.3 Front and Rear Panel BioProTT™ FlowTrack DINrail ......................................................... 32 7.3.4 Remote Zero ............................................................................................................................. 33

7.3.4.1 Electrical Connection using BioProTT Power Supply........................................................................ 34 7.3.4.2 Electrical Connection using External Power Supply .......................................................................... 34 7.3.4.3 Remote Zero Pulse Response ............................................................................................................. 35

7.4 INSTALLATION AND OPERATION OF BIOPROTT™ FLOWTRACK PLUS ............................................... 36 7.4.1 Front Panel BioProTT™ FlowTrack plus ............................................................................... 37 7.4.2 Menu Structure of BioProTT™ FlowTrack plus ..................................................................... 38 7.4.3 Rear Panel BioProTT™ FlowTrack plus ................................................................................ 39 7.4.4 Connecting and Powering up .................................................................................................. 39

7.5 MOUNTING OF BIOPROTT™ FLOWTRACK PLUS FOR PANEL MOUNT ................................................ 40 7.5.1 Installation of Flow Meter ....................................................................................................... 40 7.5.2 Installation of Flow Sensor Extension Cable (IP 68 at panel) ................................................ 41 7.5.3 De-Installation of Flow Meter and Extension Cable ............................................................... 42

8 PERFORMING MEASUREMENTS ON TUBING SYSTEMS USING BIOPROTT™ CLAMP-

ON TRANSDUCERS .................................................................................................................................... 43

8.1 INSTALLATION OF THE BIOPROTT™ CLAMP-ON TRANSDUCER ........................................................ 43 8.2 CLAMPING FLOW SENSOR ONTO TUBE ............................................................................................... 43 8.3 MEASUREMENT ACCURACY AND TUBE ARRANGEMENT .................................................................... 44 8.4 RECEIVED SIGNAL STRENGTH (RSS / ACOUSTIC COUPLING)............................................................. 46 8.5 USE OF ACOUSTIC COUPLANT ............................................................................................................ 46 8.6 ZERO ADJUSTMENT............................................................................................................................ 47 8.7 PERFORMING FLOW MEASUREMENTS ................................................................................................ 47

BioProTT™ Flow Measurement System Version 2.0 page 4 of 94

8.7.1 Measurement Using the Analog Interface of BioProTT™ Flow Meters ................................. 47 8.7.1.1 Received Signal Strength (RSS / Acoustic Coupling) on Analog Interface........................................ 47 8.7.1.2 Flow Value on Analog Interface......................................................................................................... 48

8.7.2 Measurement Using the Digital Interface (RS232).................................................................. 52 8.7.3 Measurements Using BioProTT™ FlowTrack plus ................................................................. 56

8.7.3.1 Flow Value and Coupling ................................................................................................................... 56 8.7.3.2 Selection of Calibration Table ............................................................................................................ 57 8.7.3.3 Selection of Calibration Factor ........................................................................................................... 58 8.7.3.4 Adjustment of Brightness ................................................................................................................... 59 8.7.3.5 Totalizing ........................................................................................................................................... 59 8.7.3.6 Status and Error Information .............................................................................................................. 61

8.7.4 Custom Calibration Procedure ................................................................................................ 61

9 STATUS INFORMATION, TROUBLESHOOTING, SUPPORT AND SERVICE...................... 63

9.1 AVAILABLE STATUS CODES/INFORMATION FOR FLOW METER .......................................................... 63 9.2 AVAILABLE ERROR CODES/INFORMATION FOR FLOW METER ........................................................... 64 9.3 TROUBLESHOOTING ........................................................................................................................... 66 9.4 CHECKING CURRENT LOOPS OF FLOW METER ................................................................................... 69

9.4.1 Method 1 – Using Power from Flow Meter and Measuring with Ampere meter ..................... 69 9.4.2 Method 2 – Providing Power Externally in Current Loop and Measuring with Ampere Meter

70 9.4.3 Method 3 – Checking Functionality with Voltmeter ................................................................ 71

9.5 RECOMMENDED PERIODIC CHECKS ................................................................................................... 72 9.6 SUPPORT AND SERVICE ...................................................................................................................... 73

9.6.1 Contact Information for Technical Support ............................................................................. 73 9.6.2 Returns / RMA ......................................................................................................................... 74

10 CLEANING, LOW LEVEL DISINFECTION AND MAINTENANCE ......................................... 75

10.1 CLEANING OF THE BIOPROTT FLOW METER ............................................................................. 75 10.2 CLEANING AND DISINFECTION OF THE CLAMP-ON TRANSDUCER ................................................. 76

11 APPROVAL INFORMATION ........................................................................................................... 77

11.1 EU CE COMPLIANCE ..................................................................................................................... 77 11.2 US FCC COMPLIANCE .................................................................................................................. 77 11.3 CANADA COMPLIANCE .................................................................................................................. 77 11.4 ELECTRICAL SAFETY AND ELECTROMAGNETIC COMPATIBILITY ................................................... 78

12 TECHNICAL DESCRIPTION ........................................................................................................... 79

12.1 TECHNICAL DATA BIOPROTT™ FLOW METER ............................................................................ 79 12.2 TECHNICAL DATA BIOPROTT™ CLAMP-ON TRANSDUCER .......................................................... 81

13 ENVIRONMENTAL PROTECTION AND DISPOSAL ................................................................. 84

14 MANUFACTURER'S DECLARATIONS......................................................................................... 85

14.1 INFORMATION ON ACOUSTIC OUTPUT DATA................................................................................. 85 14.2 EU DECLARATION OF CONFORMITY ............................................................................................. 87 14.3 FCC CERTIFICATE OF COMPLIANCE .............................................................................................. 88

15 ORDERING INFORMATION FOR THE BIOPROTT™ FLOW MEASUREMENT SYSTEM

AND ACCESSORIES ................................................................................................................................... 89

16 RETURN INFORMATION AND DECLARATION OF CONTAMINATION ............................. 91

17 APPENDIX ........................................................................................................................................... 92

17.1 CALIBRATION INFORMATION FOR BIOPROTT™ CLAMP-ON TRANSDUCERS ................................ 92 17.2 CHANGES DUE TO SOFTWARE VERSION V3.0 ................................................................................ 92 17.3 OUTPUT STRINGS OF V2.4 VERSUS V3.0 ....................................................................................... 93

BioProTT™ Flow Measurement System Version 2.0 page 5 of 94

Read these Operating Instructions carefully before starting up the

device!

Although the BioProTT™ flow measurement system represents state-of-the-art

technology, the user may be put at risk if the device is operated incorrectly. You should

therefore carefully read the Operating Instructions before use. Inspect your equipment for

completeness and damage after first unpacking.

These instructions contain important information on the safe handling of the BioProTT™

flow measurement system and accessories. Read these Operating Instructions carefully

before using the device and accessories, and keep them in an easily accessible location.

Familiarize yourself with and observe all warning and safety information.

It is the responsibility of the operator of the device to ensure it is used, inspected and

maintained in accordance with the Operating Instructions. Subsequent revisions or

instructions from the manufacturer must also be taken into account in this regard.

The manufacturer reserves the right to modify technical data without prior notice. Your

local distributor will supply you with current information and updates to these Operating

Instructions.

Please note:

When BioProTT™ is used in this manual, it usually refers to the complete flow

measurement system consisting of flow meter and Clamp-On Transducer.

There are 3 versions of the flow meter:

o BioProTT™ FlowTrack: flow meter without display in standard industry

housing.

o BioProTT™ FlowTrack DINrail: flow meter without display for DIN rail

mounting in cabinets.

o BioProTT™ FlowTrack plus: flow meter with display and enhanced

functionality in standard industry housing as bench top or panel mount

device

The differences of the flow meters (including SW features) will be pointed out in

the appropriate chapters.

This manual covers the following software version(s) of the BioProTT™ flow meter:

o SW version SW 3.0.x.x

The software version can be retrieved using the display or the digital

interface.

o Further information and technical papers can be found on em-tec’s website.

The serial communication (output format) has been changed in V3.0

to support higher flow rates. This needs to be taken into account

when integrating with other software or process control.

BioProTT™ Flow Measurement System Version 2.0 page 6 of 94

1 Intended Use and Restrictions

If this device is not used as intended, the user may be exposed to

risks that were not taken into account during development.

1.1 Intended Use

The BioProTT™ flow measurement system is designed for non-invasive measurement

of volumetric flow of liquids. The measurement is based on the ultrasonic transit time

method. It is usually used in laboratory, bioprocess and industrial processes.

1.2 Usage Restrictions and Limitations

The BioProTT™ flow measurement system is not intended to be used

- In an outdoor environment

- With explosive liquids

- For gases

- As a legal metrological control

2 General Safety and Important Information

2.1 General Safety Information

Warning! The general safety information below must be observed

without fail to ensure safe handling of the device.

Type of danger Safety information

Risk of explosion,

Risk of fire

The BioProTT™ flow measurement system and accessories are not

intended for use in an explosive atmosphere, with explosive agents,

oxidants or flammable agents or gases.

Risk of failure to

safeguard a safety

operation of plant

or equipment

The BioProTT™ flow measurement system and accessories are not

intended for use to control and regulate a safety operation of a plant

or equipment, especially those requiring mandatory surveillance by

law.

BioProTT™ Flow Measurement System Version 2.0 page 7 of 94

Type of danger Safety information

Risk to safety due

to defective

device, applied

parts or

accessories

A defective device, sensors/parts or accessories must not be used

and must be replaced immediately. This applies in particular if the

cable insulation is damaged or if parts have broken off or are bent.

Electric shock Fit mains input plug of power supply before use.

The power supply unit may not be immersed into liquids and is

intended for use in dry locations. However, in the event of the

ingress of liquid, pull out the power plug immediately. Stop using

the device and return it to the factory for inspection; see chapter

RETURNS / RMA.

Risk to electrical

safety, fire or of

increased

electromagnetic

emissions or

lower resistance

of device to

interference

Use of accessories, power supply units, cables (lines) and sensors

other than those specified in these Operating Instructions is not

permitted.

Any external power supply that is not supplied with the system shall

have an output voltage of 24 V DC (±10 %), must not deliver more

as 6.25 A, shall be continuously short circuit protected and provide a

separate switch.

The device must not be used in residential areas.

If longer cable lengths are used than specified in these Operating

Instructions or more than one extension cable in series is used, the

emission of the device could increase or its immunity to interference could

be reduced. Existing cable ferrites must not be removed.

Please ensure that all national requirements with respect to EMC

and electrical safety regulations are met when connecting a power

supply that was NOT originally shipped.

See chapter ELECTRICAL SAFETY AND ELECTROMAGNETIC

COMPATIBILITY; see chapter INITIAL SET UP

Malfunctions,

inaccurate

measured value

due to high-

frequency devices

The BioProTT™ flow measurement system may be influenced by

radio frequency (RF) devices. This includes mobile RF

communication equipment. Use of a RF device in the vicinity of the

BioProTT™ flow measurement system may also cause its

accessories to malfunction which may, among other things, lead to

measured values being provided inaccurately or incorrectly.

Overheating of

the device &

power supply

Prevent the device and power supply from being covered and direct

heat or sun radiation.

Faulty repairs,

Malfunctions

The device must not be opened. Repairs to the device shall only be

carried out by authorized service personnel. Unauthorized opening

or repairs will void manufacturer’s warranty.

Faulty electrical

installation

Comply with the installation requirements; refer to the chapter

ELECTRICAL SAFETY AND ELECTROMAGNETIC

COMPATIBILITY.

BioProTT™ Flow Measurement System Version 2.0 page 8 of 94

Type of danger Safety information

Measuring

inaccuracy,

Risk to electrical

safety

The measuring accuracy and safety-relevant aspects should be

checked at least every 2 years. Please refer to chapter

RECOMMENDED PERIODIC CHECKSFehler! Verweisquelle

konnte nicht gefunden werden. for details.

Risk of

infection/environ-

mental

contamination

due to improper

disposal

The device and accessories must be disposed of in accordance with

the national specifications for waste electrical and electronic

devices.

In case microbial contamination may be present on the device after

being used, we recommend you dispose of these separately in

accordance with the applicable regulations.

In case microbial contaminated, device or accessories shall be

cleaned and disinfected before they are sent to the manufacturer.

Do not return a device/sensor if you are not absolutely certain that

all traces of hazardous substances have been removed, e.g.

substances which have penetrated crevices or diffused through

plastic. Refer to chapter RETURN INFORMATION AND

DECLARATION OF CONTAMINATION for additional

information.

Damage to the

flow

measurement

system by

chemicals

BioProTT™ components may be damaged or deteriorated when

coming in contact with aggressive chemical components as used in

liquid systems, couplants or cleaning solutions.

Therefore contact with chemicals should generally be avoided.

Ensure also that tubing used is compatible with intended fluid.

Please refer to chapter CLEANING, LOW LEVEL

DISINFECTION AND MAINTENANCE for details Table 1: General Safety Information

For more application-specific warnings, refer to the corresponding chapters.

The manufacturer only considers itself to be responsible for the safety, reliability and

usability of the device if:

The device is used in accordance with the instructions for use.

Installation, connection to the electricity supply and process control system,

additions, readjustments, changes or repairs and maintenance are only carried out

by qualified persons who have been authorized and trained. These qualified persons

must have read and understood this manual and must follow them accordingly.

The flow measurement system must be operated by persons authorized and trained

by the plant operator. Compliance with the instructions in this manual is

compulsory.

The manufacturer provides no warranty and assumes no liability in the event of non-

compliance with the above.

BioProTT™ Flow Measurement System Version 2.0 page 9 of 94

2.2 General Important Information

The following symbols are used in these Operating Instructions to highlight specific

information.

Caution! Important information regarding correct handling. To be

carried out and strictly observed!

If this following information is not observed, faults or damage to the

product or its surroundings may occur.

Type of damage Important information

Malfunction or

irreparable

damage

The BioProTT™ flow measurement system is intended to be used

with its own power supply (shipped with the system) or e.g. within

electrical control cabinets with their own power supply.

The DC power supply line shall be shorter than 30 m and shall not

leave the building.

Damage due to

ultrasonic sound

At all device settings of the BioProTT™ flow measurement system

the Clamp-On Transducer generates ultrasonic sound only at a very

low acoustic output level.

The ultrasonic sounds generated are harmless to persons, protein

solutions (e.g. blood, blood cells) and the environment according to

current knowledge, providing the system is used properly.

For more detailed information, see chapter INFORMATION ON

ACOUSTIC OUTPUT DATA.

Damage or mal

function

Unused connector sockets should be covered e.g. with the provided

flexible grip caps to avoid accumulation of dirt or fluids potentially

causing short circuits, corrosion or mal function.

No legal

metrological

control

Device is not designed to be used as a legal metrological control as

i.e. defined in EU Directive 2004/22/EC Measuring Instruments

Table 2: General important information

BioProTT™ Flow Measurement System Version 2.0 page 10 of 94

3 Symbols, Units and Abbreviations

All symbols and abbreviations used are explained in this chapter.

3.1 Symbols used in these Operating Instructions

The following symbols are used in these Operating Instructions to highlight specific

information.

Symbol Meaning

Warning!

This safety symbol precedes critical information that must be strictly

observed in order to prevent injuries and fatal hazards. This warning

symbol is the most important safety symbol.

Caution!

Important information regarding correct handling. Must be performed

and strictly observed!

If this information is not observed, faults or damage to the product or

its surroundings may occur. Table 3: Symbols Operating Instructions

3.2 Symbols on External Power Supply Unit

Symbol Meaning

Alternating current:

Alternating current must be fed to the supplied external power

supply unit. The values for the supply voltage must correspond to

those of the power supply unit: 100-240 VAC / 50-60 Hz.

Friwo Gerätebau GmbH (manufacturer logo)

Protection Class II (protective insulation)

As a safety guarantee, this power supply unit has additional

insulation to prevent dangerous touch voltage in the event of a fault.

IP40 Protected against access by wire

Ta40/B Rated ambient temperature/transformer class

Do not dispose of this device as domestic waste!

Devices must be disposed of as waste devices according to WEEE

directive and national legislation.

CE marking: The external power supply unit satisfies the

requirements of Low Voltage Directive 2006/92/EC and

EMC Directive 2004/108/EC.

UL test mark for recognized components for Canada and the USA

by Underwriter Laboratories Inc.

BioProTT™ Flow Measurement System Version 2.0 page 11 of 94

Symbol Meaning

European Norms Electrical Certification (10 = ID number of VDE)

Direct current

(24 V DC from external power supply unit)

Caution

Risk of electric shock. Dry location use only. Fit mains plug before

use AC/DC adapter.

Further warnings and precautions exist.

(Refer to these instructions for use)

Table 4: Symbols Power Supply

3.3 Symbols on Flow Meter, Sensors and on Packaging

Symbol Meaning

Caution

Further warnings and precautions exist.

(Refer to these instructions for use)

IP65 Protected against dust and high pressure water jets / wash down

from all directions (BioProTT™ FlowTrack, BioProTT

FlowTrack plus, BioProTT™ Clamp-On Transducer).

IP20 Protected against access by finger (BioProTT™ FlowTrack DINrail)

Direct current

(24 V DC from external power supply unit)

Do not dispose of this device as domestic waste!

Waste devices must be disposed of in accordance with WEEE

Directive and national legislation.

Manufacturer:

em-tec GmbH

Lerchenberg 20

86923 Finning

Germany

Year of Manufacturing

Serial number

Order number

BioProTT™ Flow Measurement System Version 2.0 page 12 of 94

Symbol Meaning

CE marking: This device complies with the relevant European

directives. The applicable legislation is given in the EU

DECLARATION OF CONFORMITY.

FCC-Mark: The device satisfies the requirements of the United States

Federal Communication Commission. See US FCC Compliance

Caution, fragile!

Handle with care!

Protect against moisture! Store in a dry place.

Temperature limit during storage

non-condensing

Moisture limit during storage (non-condensing)

Air pressure limit

Storage

Transport

Table 5: Symbols Devices and Packaging

BioProTT™ Flow Measurement System Version 2.0 page 13 of 94

3.4 Units

Unit Meaning

" Tube size in inches, 1 inch ≈ 25.4 mm

ID" x WT"

Tube dimensions in inches :

Outer diameter = internal diameter ID + (2 x wall thickness WT)

Table 6: Units

3.5 Definitions and Abbreviations

Definitions,

Abbreviation

Meaning

AWG American Wire Gauge

BCT BioProTT Clamp-On Transducer

CT Clamp-On Transducer

DAQ Data Acquisition System

EMC Electromagnetic Compatibility

flow meter BioProTT™ FlowTrack, BioProTT FlowTrack plus,

BioProTT FlowTrack DINrail

flow sensor BioProTT Clamp-On Transducer

N/A Not applicable

PCS Process Control System incl. DAQ

ID Inner diameter

OD Outer diameter

Qmax Maximum flow

TT Transit-Time

TTFM Ultrasonic Transit-Time Flow Measurement

RSS Received Signal Strength which corresponds to the acoustic coupling

WT Wall thickness

Table 7: Definitions and Abbreviations

BioProTT™ Flow Measurement System Version 2.0 page 14 of 94

4 Description of Measuring Principle

The physical measurement principles applied in the BioProTT™ flow measurement system

are described in this chapter.

The BioProTT™ flow measurement system uses the ultrasonic transit-time method (TT =

transit-time) which delivers precise flow values in tube systems and pipes.

The ultrasonic converters (piezoceramics) in the flow sensors transmit high-frequency

acoustic signals in and against the flow direction. The time difference between these

signals is proportional to the volumetric flow.

The basic principle is explained below.

Transmitter/

Receiver

Receiver/

Transmitter A

B

Tube

Flow

a

Figure 1: Principle of ultrasonic flow measurement (simplified using only two ceramics)

One piezoceramic (A) is excited by a set of high-frequency vibrations and transmits

ultrasonic waves to a second opposing piezoceramic (B) that acts as a receiver. The

piezoceramics are arranged at a specific angle α in relation to the flowing medium. The

transit time is influenced by the medium. The flow velocity of the medium can be

calculated using the measured transit time differences. The flow in liters per minute is

established using the known cross-section of the tube being scanned by ultrasonic

transmission.

Four ultrasound converters in a crosswise arrangement transmit high-frequency sound

signals alternately in and against the direction of the flow. The transit time for each

impulse is measured; the transit-time difference between the upstream and downstream

movement of the impulses is proportional to the volumetric flow.

BioProTT™ Flow Measurement System Version 2.0 page 15 of 94

5 Packaging Contents

The BioProTT™ flow measurement system is shipped with the following components:

One Flow meter in either of the following variants:

a) BioProTT™ FlowTrack (without display)

b) BioProTT™ FlowTrack plus (with display) as bench top model

c) BioProTT™ FlowTrack plus (with display) for panel mount

d) BioProTT™ FlowTrack DINrail (without display for DIN rail mounting)

Please note: Flexible grip caps are provided for all circular sockets and should

not be removed until a socket is actually used to avoid any dirt accumulation or

any potential short circuits.

One Power supply with country specific adaptor. Currently the following adaptors are

available: EURO, UK, USA/JAPAN, IEC - others on request.

Two pre-configured 2 m long cables with plug on one end and cage clamp terminal

blocks on other end. Included with flow meter variant a), b) and c).

One pre-configured serial cable, 2 m long, for digital serial output with plug on one

end and standard female D-sub connector. Included with flow meter variant a), b) and

c).

BioProTT™ Flow Measurement System Version 2.0 page 16 of 94

One male D-Sub serial to USB adapter for connection to standard USB ports.

Please download current driver from link specified on actual adapter labeling.

The current link is also available on em-tec’s web site.

One pre-configured serial cable, 2 m long, for digital output with ten (10) pole

terminal block on one end and standard female D-sub connector. Included with

BioProTT™ FlowTrack DINrail.

One pre-configured sensor line extension cable (ID12466), 1 m long for flow sensor

for use in process control cabinet. Included with flow meter BioProTT™ FlowTrack

DINrail.

BioProTT™ Flow Measurement System Version 2.0 page 17 of 94

One pre-configured sensor line extension cable (ID12560), 1.1 m long for flow

sensor for use in process control cabinet. Included with flow meter BioProTT™

FlowTrack plus (with display) for panel mount. A watertight protection cap is

included.

One BioProTT™ Clamp-On Transducer

The BioProTT™ Clamp-On Transducer is shipped pre calibrated and ready to be

connected to either of the flow meters. It is included in the shipment if it has been

ordered accordingly.

One Printed Operating Instructions

BioProTT™ Flow Measurement System Version 2.0 page 18 of 94

6 Initial Set Up

In this chapter some basic consideration about integration of the BioProTT flow meters are

given and appropriate schematics of power supply and wiring of current loops are shown.

6.1 Important Current Loop Definitions

Generally, devices in 4-20 mA environments can be classified based on the current source

of the 4-20 mA loop as “active” or “passive”. Passive devices or sensors rely on (external)

loop power, whereas active devices are powering the current loop.

With respect to the 2-/4-wire classification of sensors, as a rule of thumb, one can assume

that 2-wire sensors are usually working in passive mode, whereas 4-wire sensors can

usually support the signal current loop being passive or active.

The BioProTT™ flow measurement system supports passive and active modes of

operation for flow and RSS current loop – see figure below. However, the flow meter

always needs power to operate!

Figure 2: BioProTT™ flow meter supporting “passive” loop configuration (example: flow)

BioProTT™ Flow Measurement System Version 2.0 page 19 of 94

Figure 3: BioProTT™ flow meter supporting “active” loop configuration (example: flow)

The figures below show some of the electrical design of the BioProTT™ flow meters

which is important to properly wire the system. The pin assignment corresponds to the pins

as outlined in the following chapters.

Figure 4: BioProTT™ flow meter schematics with focus on power supply (w/o DIN rail variant)

BioProTT™ Flow Measurement System Version 2.0 page 20 of 94

Figure 5: BioProTT™ FlowTrack DINrail schematics with focus on power supply

6.2 Powering of BioProTT™ Flow Meter

The flow meter works with DC power, which can be supplied with

The power supply unit shipped with the system using the power socket or

With an external 24 V DC (±10%) power supply delivering at least 300 mA and

continuously short circuit protected.

o DC power can be supplied either through PIN 1/3 of the Power, Flow or

RSS socket on BioProTT™ FlowTrack and BioProTT™ FlowTrack plus.

o BioProTT™ FlowTrack DINrail can be supplied by DC power via PIN 1/2

of the terminal block or PIN 1/3 of the Power socket.

BioProTT™ Flow Measurement System Version 2.0 page 21 of 94

o This option is e.g. useful, when connecting it to a process control and

centrally powering the devices on or off. In case power is provided via the

Flow or RSS socket or the terminal block, the on/off button is no longer

“active”, i.e. the flow meter will start up as soon as power is available. In

case of a device error the power supply must be turned off.

When power is supplied via the power socket the on/off switch is used

to power on the flow meter otherwise the on/off switch does not have

an effect.

Ensure that there are no conflicting power sources/voltages applied to

Pin 1 and Pin 3 of either of the Power, Flow or RSS socket, as they are

internally bridged between the sockets.

For the BioProTT™ FlowTrack DINrail, please ensure this similarly

between Pin 1 and Pin3 of the Power socket and Pin 1 and 2 of the

terminal block.

6.3 Powering Over Analog Sockets

Below pin assignments and cable colors for flow meters with round connectors for Power,

Flow and RSS sockets are shown.

Figure 6: BioProTT flow meter rear view with one analog plug connected -example w/ Flow socket

Figure 7: BioProTT flow meter: view of power and analog Flow and RSS round sockets

Pin No Signal Power Signal Flow/RSS Wire color

(in supplied cabling)

1 24 V (24C_in) 24 V (24C_in) Brown

2 Not assigned Current Loop + White

3 Ground in (GND_in) Ground in (GND_in) Blue

4 Not assigned Current Loop - Black Table 8: Pin assignment BioProTT™ flow meter with round connectors

BioProTT™ Flow Measurement System Version 2.0 page 22 of 94

6.4 Current Loop(s) powered by Flow Meter

In this set-up, the flow meter current loop(s) are powered by the flow measurement system,

using wire bridges. From a flow meter perspective the current loop is “active”.

The wire bridge is applied from Pin 1 to Pin 2 of the Flow respective RSS cable or the

respective terminal block connections from BioProTT™ FlowTrack DINrail.

For details see the circuit diagrams below and observe the wiring instructions.

Figure 8: BioProTT flow meter rear view with one analog plug connected -example w/ flow socket

Pin No Flow / RSS Wire Color

(in supplied cabling)

Connection

1 24 V (24C_in) Brown + 24 V DC supply

2 Current Loop + White Connect Pin 1

with Pin 2

3 Ground in (GND_in) Blue Ground (GND) /

Return of PCS

4 Current Loop - Black Send (+) of PCS Table 9: Wiring connections for BioProTT™ flow meter with round connectors and flow meter power

Figure 9: Schematics – showing wiring of BioProTT™ FlowTrack /plus with Flow and RSS current

loop being connected and powered using the flow meter power supply

BioProTT™ Flow Measurement System Version 2.0 page 23 of 94

Figure 10: Schematics – showing the terminal block wiring of BioProTT™ FlowTrack DINrail with

Flow and RSS current loop being powered via flow meter and power supply shipped with unit

6.5 Current Loop(s) powered by External Supply, e.g. from PCS

In this set-up, the flow meter current loop does not have to be powered by the flow

measurement system, as it is already being supplied with an external power supply, e.g.

using either power from a signal processing card of a PCS or separate power supply.

From a flow meter perspective the current loop is “passive”.

For details see the circuit diagrams as follows and observe the wiring instructions.

BioProTT™ Flow Measurement System Version 2.0 page 24 of 94

Figure 11: BioProTT flow meter rear view with one analog plug connected -example w/ flow socket

Pin No Flow / RSS Wire color

(in supplied cabling)

Connection

1 24 V (24C_in) Brown Not connected

2 Current Loop + White Receive (-) of PCS

3 Ground in (GND_in) Blue Not connected

4 Current Loop - Black Send (+) of PCS Table 10: Wiring connections for BioProTT™ flow meter with round connectors and external power

Figure 12: Schematics – showing wiring of BioProTT™ FlowTrack /plus with Flow and RSS current

loop being powered via external power supply while the flow meter is powered with power supply

shipped with unit

BioProTT™ Flow Measurement System Version 2.0 page 25 of 94

Figure 13: Schematics – showing the terminal block wiring of BioProTT™ FlowTrack DINrail with

Flow and RSS current loop being powered via external power supply in PCS while flow meter is

powered with power supply shipped with unit

Figure 14: Schematics – showing the terminal block wiring of BioProTT™ FlowTrack DINrail with

flow and RSS current loop being powered via signal processing cards of PCS while flow meter is

powered with power supply shipped with unit

BioProTT™ Flow Measurement System Version 2.0 page 26 of 94

7 Installation and Operation of BioProTT™ FlowTrack

Compliance with the prescribed operating parameters and safety

information must be ensured prior to putting into service. The following

installation instructions must be strictly observed.

The BioProTT™ FlowTrack can be used as bench top flow meter or mounted with screws

on surfaces.

The BioProTT™ FlowTrack is connected to the mains supply via the corresponding power

supply unit, via the process control or another external power supply meeting the

requirements as outlined in chapter POWERING OF BIOPROTT™ Flow Meter.

The device is being powered on or off by

pushing the switch at the rear of the flow meter when powered through the dedicated

power input

OR alternatively by connecting/disconnecting a power supply via one of the analog

interfaces (Flow or RSS)

A green LED on the BioProTT™ FlowTrack indicates that it is has been powered-on and

initialized.

7.1 Front Panel BioProTT™ FlowTrack

Figure 15: Front view of BioProTT FlowTrack

Components Description

1 Push button

“Zero adjustment”

Push button to adjust offset at zero flow. The zero

adjustment is performed after button is released when

coupling is sufficient.

2 Status LED A green light indicates device has been powered-on. The

system requires a few seconds for powering on and

initialization. Table 11: Front panel BioProTT™ FlowTrack

1 2

BioProTT™ Flow Measurement System Version 2.0 page 27 of 94

7.2 Rear Panel BioProTT™ FlowTrack

Figure 16: Rear view of BioProTT FlowTrack

Components Description

1 On/Off button Push the button to power the device on or off.

Note: this button is only relevant when power is supplied

via the power socket. The button does not affect power

supplied via either Flow or RSS socket.

2 Power 4-pin connecting socket for DC power.

Note: only 2 pins are actually used.

3 Digital out 3-pin connecting socket for the signal digital interface.

4 Flow 4-pin connecting socket for the analog flow signal.

5 RSS 4-pin connecting socket for the analog signal “received

signal strength” (RSS).

6 Flow Sensor 16-pin connecting socket for the

BioProTT™ Clamp-On Transducer connector with push

and pull unlock mechanism. Table 12: Rear panel BioProTT™ FlowTrack.4

2 3 4 5 6 1

BioProTT™ Flow Measurement System Version 2.0 page 28 of 94

7.2.1 Pin Assignment and Cable Colors for Round Connector for Digital

Interface

Please note: the supplied serial cable for the digital interface is completely wired and

ready to use.

Figure 17: Pin assignment of digital interface of BioProTT FlowTrack

Pin No Signal Wire color

(in supplied cabling)

Sub-D connector

1 PC_TXD Brown PIN 3

3 GND Blue PIN 5

4 PC_RXD Black PIN 2 Table 13: Pin assignment for digital interface for BioProTT™ FlowTrack /plus

7.2.2 Pin Assignment and Cable Colors for Round Connectors for Power,

Flow and RSS Interfaces

Please see chapter POWERING OF BIOPROTT™ FLOW METERfor details.

The circular plug design optimally supports the flow measurement

application. Connection to other devices or sensors than intended can lead

to damage.

Please note: unused connector sockets should be covered with the provided flexible grip

caps. For additional protection the use of suitable screw caps for all sockets that are not

connected/in use is recommended.

BioProTT™ Flow Measurement System Version 2.0 page 29 of 94

7.2.3 Connecting and Powering up

The external power supply must not deliver more as 6.25 A at 24 VDC,

shall be continuously short circuit protected and have a separate switch.

Please ensure that all national requirements with respect to EMC and

electrical safety regulations are met for when connecting a power

supply that was NOT originally shipped.

The BioProTT™ flow meter shall be powered by 24 V DC (±10 %).

The power supply shall deliver at least 300 mA.

Please note: the flow meter can be powered with the provided power supply or with

another external power supply meeting the above mentioned specifications. If the power

supply does not meet the specifications above the protection provided by the device may be

impaired.

For installation with provided power supply observe the following steps:

1. Connect power supply

2. Connect Clamp-On Transducer to BioProTT™ flow meter.

3. Connect the 4-20 mA and/or digital serial output to your data acquisition system. The

connectors should be firmly tightened. Unused connector sockets should be covered

with provided flexible grip caps.

4. Power on the device with the switch. Wait some seconds until the system is

completely initialized. The system will automatically start to transfer measurement

values.

Please note: The current loops can either be powered through the process control

system or via the flow meter. The wiring needs to be done as show in chapter

INITIAL SET UP

5. Clamp transducer around tubing (for details see chapter INSTALLATION OF THE

BIOPROTT™ CLAMP-ON TRANSDUCER). An arrow on the sensor lid indicates

the flow direction; make sure that the sensor is oriented in the flow direction

otherwise negative readings occur. Make sure the tubing size, material and conditions

fit calibration data and the sensor lid is closed (locked in) to assure a reliable

measurement.

6. Select corresponding sensor table; the default setting is table 1

7. If not yet done fill tubing with liquid, you will see an increase of the RSS value.

BioProTT™ Flow Measurement System Version 2.0 page 30 of 94

8. Make sure that the liquid is not in motion and press the Zero Adjustment Button.

Pressing the zero offset button while the fluid is in motion will cause errors in the

flow measurement reading, due to wrong flow offset.

9. The flow measurement system is now ready for use.

Please note: when system is powered down, the sensor will reset to factory defaults.

7.3 Installation and Operation of BioProTT™ FlowTrack DINrail

Compliance with the prescribed operating parameters and safety

information must be ensured prior to putting into service. The following

installation instructions must be strictly observed.

The terminal block requires a screwdriver with 2.5 mm blade. The wires

used in conjunction with the terminal block should not exceed 1.5 mm2,

The BioProTT™ FlowTrack DINrail flow meter supports DIN rail mounting, e.g. for use

in cabinets of process controls. From a feature point of view, the device is identical with

the BioProTT™ FlowTrack. The physical connections have been moved and the form

factor has partially been changed to support specific installation environment.

The flow meter is connected to the mains via the process control or another external power

supply meeting the requirements as outlined in chapter POWERING OF BIOPROTT™

Flow Meter.

The device is powered either by

Power socket and pushing the “power on/off” switch at the front of the flow meter

when powered through the dedicated power input.

Please note: this switch is available, but usually not used, i.e. typically the switch is

always in the “on” position.

External power supply via terminal block

A green LED on the BioProTT™ FlowTrack DINrail indicates that it is has been powered-

on. The system requires some seconds for initializing.

BioProTT™ Flow Measurement System Version 2.0 page 31 of 94

7.3.1 DIN Rail Mounting

BioProTT™ FlowTrack DINrail is equipped with a mounting adapter for DIN rail type O,

35 mm x 7.5 mm or compatible. Position the upper hook at the rear on the DIN rail and

push in the lower. Please ensure that it is firmly mounted onto the rail before setting up any

connections.

When removing the device, push down the spring loaded bracket with a minus screwdriver

and pull.

Spring loaded DIN rail adapter

DIN rail type O, 35 mm x 7.5 mm

Figure 18: Flow meter mounting on DIN rail

7.3.2 Installation of Flow Sensor Extension Cable (IP50 at panel)

The female plug (Lemo Type PFG.2B.316.CLLD72) of the flow sensor extension cable

can be mounted in the cabinet wall with a thickness up to 6.5 mm (0.26 inch). The male

plug is inserted in the sensor port of the BioProTT™ FlowTrack DINrail. The cut out panel

is indicated below. The red marks indicate the correct orientation of the connection.

Figure 19: Panel cut out in cabinet

Please use the yellow cap to protect the plug from dust when not needed or use compatible

accessories from Lemo or ODU to cover the plug.

Please ensure that only one extension cable is used for connecting a flow sensor.

BioProTT™ Flow Measurement System Version 2.0 page 32 of 94

7.3.3 Front and Rear Panel BioProTT™ FlowTrack DINrail

Figure 20: Front and rear view of BioProTT FlowTrack DINrail

Components Description

1 Terminal block 10-pin connecting socket for terminal block with PIN

assignment as shown below.

2 Flow Sensor 16-pin connecting socket for the

BioProTT™ Clamp-On Transducer or extension cable.

3 On/Off button Push the button to power the device on or off.

Note: this button is only relevant when power is supplied

via the power socket. The button does not affect power

supplied via terminal block.

4 Status LED A green light indicates device has been powered-on. The

system requires some seconds for initializing.

5 Power 4-pin connecting socket for DC power.

Note: only 2 pins are actually used.

6 Push button

“Zero adjustment”

Push button to adjust offset at zero flow. The zero

adjustment is performed after button is released when

coupling is sufficient.

7 Mounting adapter Adapter plate for DIN rail mounting with spring loaded

lock mechanism. Table 14: Front and rear panel BioProTT™ FlowTrack DINrail

2 3 4 5 6 1 7

BioProTT™ Flow Measurement System Version 2.0 page 33 of 94

The PIN assignment on the terminal adapter block respectively the corresponding plug is

as follows:

PIN Number Meaning Color (in scheme) Remark

1 +24V Black

2 Ground Blue

3 Flow Current Loop + White

4 Flow Current Loop - Brown

5 RSS Current Loop + Yellow

6 RSS Current Loop - Green

7 PC_TXD Brown PIN 3 Sub-D

8 GND Blue PIN 5 Sub-D

9 PC_RXD Black PIN 2 Sub-D

10 Remote Zero + Red Pulse input Table 15: Terminal Block pin assignment for BioProTT™ FlowTrack DINrail

The terminal block connector for BioProTT™ FlowTrack DINrail is prewired and ready to

use for serial communication.

Figure 21: Pre-wired terminal block of BioProTT FlowTrack DINrail

7.3.4 Remote Zero

BioProTT™ FlowTrack DINRail devices with software version SW 3.0 or later have an

integrated “Remote Zero” option to correct the offset by remote control switch. Please see

chapter ZERO ADJUSTMENT for correct zeroing conditions.

The flow offset correction can be performed by pushing the “Zero adjustment” button on

the panel or additionally by generating a direct current voltage pulse by a switch on PIN

10. The digital input for remote zero accepts pulses of 24 V DC ± 10% and has an internal

current consumption of < 25 mA.

BioProTT™ Flow Measurement System Version 2.0 page 34 of 94

7.3.4.1 Electrical Connection using BioProTT Power Supply

Figure 22: Remote Zero using BioProTT Power Supply

When using the BioProTT™ power supply, a switch contact between PIN 1 and PIN 10

can be used to generate the remote zero pulse.

7.3.4.2 Electrical Connection using External Power Supply

Figure 23: Remote Zero using external Power Supply

When using an external power supply a switch contact between +24 V DC of the power

supply and PIN 10 can be used to generate the remote zero pulse. PIN 2 must be connected

to the GND of the power supply.

Please ensure the correct polarity of the electrical connection to avoid

device damage!

BioProTT™ Flow Measurement System Version 2.0 page 35 of 94

7.3.4.3 Remote Zero Pulse Response

Figure 24: Remote Zero Pulse induces Offset correction

Please see chapter ZERO ADJUSTMENT for correct zeroing conditions.

The zero adjustment is triggered on the falling edge of the 24 V DC remote zero pulse with

pulse length of least 500 ms and only accepted by system when coupling is sufficient. The

time interval between two pulses should not be below 2000 ms.

BioProTT™ Flow Measurement System Version 2.0 page 36 of 94

7.4 Installation and Operation of BioProTT™ FlowTrack plus

Compliance with the prescribed operating parameters and safety

information must be ensured prior to set up. The following installation

instructions must be strictly observed.

The BioProTT™ FlowTrack plus is available as bench top device. In addition, the

BioProTT™ FlowTrack plus can also be ordered in a variant for panel mount. The devices

are identical in features and differ only in their mounting options, which are covered in

section MOUNTING OF BIOPROTT™ FLOWTRACK PLUS FOR PANEL MOUNT.

The device is powered on or off by

pushing the switch at the rear of the flow meter when powered through the dedicated

power input

OR alternatively by connecting/disconnecting a DC power supply via one of the

analog interfaces (on/off switch is not active) or the power socket (which engages the

on/off switch)

BioProTT™ Flow Measurement System Version 2.0 page 37 of 94

7.4.1 Front Panel BioProTT™ FlowTrack plus

After power on, the flow meter takes some seconds for initialization.

The device is ready to operate, once the “flow” screen appears.

Please note: the display will be updated every second.

Figure 25: Front view of BioProTT FlowTrack plus

Basic operation of the device:

Button Meaning

The Up- and Down Arrow Buttons are used to select a possible

setting on the current screen or switch to the previous / next

screen.

If a selection of a value is possible, “sel” is displayed on the

screen.

The Enter Button is used

to change to “sel” mode,

to confirm a value or change and exit “sel” mode and

to change from “hold” to “run” status when in totalizer

mode.

Note: Press this button only short time, otherwise “sel” mode

may be left.

The Zero Adjustment Button is used for zero adjustment of the

flow or reset totalizer value(s) and exclamation mark.

Table 16: Front panel elements for BioProTT™ FlowTrack plus

BioProTT™ Flow Measurement System Version 2.0 page 38 of 94

7.4.2 Menu Structure of BioProTT™ FlowTrack plus

Flow chart of menu structure:

Figure 26: Menu structure of BioProTT FlowTrack plus

Zero Flow / Measure Flow

Initializing Device

Start/Stop/Reset Totalizing*

Set Calibration Table*

Set Calibration Factor*

Adjust Brightness

Display Status Information

Note: The message “initializing

device” only appears after

switching on the device.

*Screens can be accessed only if

a valid sensor is connected.

BioProTT™ Flow Measurement System Version 2.0 page 39 of 94

7.4.3 Rear Panel BioProTT™ FlowTrack plus

The rear panel of BioProTT™ FlowTrack plus is identical to the flow meter without

display, i.e. BioProTT™ FlowTrack. Please see chapter REAR PANEL BIOPROTT™

FLOWTRACK.

7.4.4 Connecting and Powering up

The flow meter is operated with the provided power supply.

Please note: the flow meter can alternatively be powered via an external power supply

either via the dedicated power input or either of the analog sockets. For details see chapter

POWERING OF BIOPROTT™ Flow Meter .

1. Connect power supply unit that is shipped with the device

2. Connect Clamp-On Transducer to the BioProTT™ flow meter.

3. Connect the 4-20 mA and/or digital serial output to your data acquisition system.

The connectors should be firmly tightened. Unused connector sockets should be

covered with provided flexible grip caps.

4. Power on the device with the switch. Wait some seconds for initialization, which is

done when the flow screen appears. The system will automatically start to transfer

measurement values.

Please note: The Current loops can either be powered through the process control

system or are powered by the flow meter. The wiring needs to be done as shown in

chapter INITIAL SET UP.

5. Clamp transducer around tubing (for details see chapter INSTALLATION OF THE

BIOPROTT™ CLAMP-ON TRANSDUCER). An arrow on the sensor lid indicates

the flow direction; make sure that sensor is oriented in the flow direction

otherwise negative readings occur. Negative Values will be shown correctly on the

display and in the digital reading. However, the analog interface will output

negative flow values into the range of 1 mA to 4 mA to the extent possible. For

details see chapter: MEASUREMENT USING THE ANALOG INTERFACE OF

BIOPROTT™ FLOW METERS. Make sure the tubing size and material fit and the

sensor lid is closed (locked in).

6. Select the appropriate sensor table

7. If not yet done fill tubing with liquid, you will see an increase of the RSS value on

the screen (or the other corresponding interfaces).

8. Make sure that the liquid is not in motion and press the Zero adjustment button on

BioProTT™ FlowTrack plus when in the flow screen.

9. The flow measurement system is now ready for use.

BioProTT™ Flow Measurement System Version 2.0 page 40 of 94

7.5 Mounting of BioProTT™ FlowTrack plus for Panel Mount

7.5.1 Installation of Flow Meter

The BioProTT™ FlowTrack plus for panel mount with standard industry housing is

designed as device which is mounted in a panel. When mounted it supports IP66 at the

panel.

The flow meter is equipped with larger face plate which can be mounted with 6 M3 (ISO

Metric Coarse Thread DIN 13) screw nuts on threaded bars. See below diagram for exact

location and dimensions of threaded bars as well as the panel cut out (in [inch] and [cm]).

Screws nuts must be tightened with torque value of 0.5 Nm +/-0.1 Nm to reach IP level

mentioned above when mounted in panel.

The intended maximum panel thickness supported is 3mm.

Figure 27: Front and rear view of BioProTT FlowTrack plus in panel

BioProTT™ Flow Measurement System Version 2.0 page 41 of 94

Figure 28: Drill template – values in [mm] and [inch]

7.5.2 Installation of Flow Sensor Extension Cable (IP 68 at panel)

The female socket (ODU Type G82B0C-P16LFG0-000L) of the flow sensor extension

cable can be mounted in the cabinet wall with a thickness up to 3 mm (0.12 inch). In order

to achieve the specified IP rating, the special ODU nutdriver for slotted mounting nut type

702 098 001 000 000 must be used. The maximum torque value is 4.5 Nm.

The male plug is inserted in the sensor port of the BioProTT™ FlowTrack plus. The panel

cut out panel is indicated below. The red marks indicate the correct orientation of the

connection.

BioProTT™ Flow Measurement System Version 2.0 page 42 of 94

Figure 29: Panel cut out in cabinet in [mm]

Please use the provided cap to protect the plug from dust etc. and achieve above IP rating

when no sensor is connected.

Please ensure that only one extension cable is used for connecting a flow sensor.

7.5.3 De-Installation of Flow Meter and Extension Cable

If the flow meter needs to be un-mounted from the panel, please disconnect all electrical

connections and open only the screw nuts accessible from inside the panel which fix the

faceplate of the device to the panel. Do not loosen any of the four cross-head screws on the

front of the device.

For de-installation of the sensor extension cable, please open screw nut of socket

accessible from front of panel.

When removing the device and cable e.g. for servicing please perform the mounting steps

in reverse order. In case of remounting the tightness of the once compressed gasket may be

compromised.

BioProTT™ Flow Measurement System Version 2.0 page 43 of 94

8 Performing Measurements on Tubing Systems using

BioProTT™ Clamp-On Transducers

8.1 Installation of the BioProTT™ Clamp-On Transducer

The Clamp-On transducer and the extension lines contain a push and pull locking

mechanism to avoid unintended disconnection. This permits the simple insertion and

locking. Unlocking of the connector is performed by pulling on the connector housing.

Make sure you are using the tubing, medium and temperature for which the

transducer was calibrated. The stated measurement accuracy cannot be

guaranteed if the measurements are carried out using tubing that differs in

size, wall thickness or material from the tubing on which the factory

calibration was performed. If measuring with at a different temperature or

with a different medium than the factory calibration, measurement errors

may occur unless adjustments are made by performing a custom calibration

on site and adjusting the calibration factor.

The calibration parameters are listed in the calibration report supplied with

the transducer.

Ensure that the sensor surface and tubing are clean for good coupling.

Select a location so that bubbles cannot accumulate, like below the highest

point of the overall tubing line.

8.2 Clamping Flow Sensor onto Tube

The 16-pin circular connector of the BioProTT™ Clamp-On Transducer is connected to

the BioProTT™ flow meter (see e.g. VIEW OF FRONT AND REAR). The transducer is

then clamped to the tube.

To obtain a positive flow value, make sure the arrow is pointing in the flow

direction.

BioProTT™ Flow Measurement System Version 2.0 page 44 of 94

Clamp the transducer onto the tube as follows:

Figure 30: Clamping onto tube

1. Open the spring lock and lift up the cover.

2. Insert the tube into the recess provided.

3. Close the lid by applying gentle pressure until the spring lock snaps into place.

4. Please note: the tubing should fit snugly in the transducer and should not be loose.

If the Clamp-On Transducer is used at significantly different parameters

compared to the calibration, like medium temperatures, this can lead to

inaccurate measurement results.

Please note: Customer-specific calibrations can also be carried out on request. Please

contact our SUPPORT AND SERVICE DEPARTMENT for further details.

8.3 Measurement Accuracy and Tube Arrangement

Accuracy ± 3% of reading ± max. 20 ml/min, for all

transducer sizes up to 1/2 x 1/8”

The accuracy has been confirmed under the following reference conditions:

Liquid temperature calibration temperature ±5°C (± 9° F)

Ambient temperature 20°C ±5°C (68° ± 9° F)

Straight inlet section 10 x ID of tubing, fully developed flow profile

Tubing Raumedic-ECC-Blood Line PVC

Acoustic couplant applied

Calibration factor 1.00

BioProTT™ Flow Measurement System Version 2.0 page 45 of 94

Figure 31: Accuracy graph

Optimized Accuracy:

As the measurement allows a maximal flexibility with respect to positioning the sensor the

given accuracy is a standard under the reference conditions. Certain setups allow higher

levels of accuracy and can be evaluated by a factory calibration certificate on customer

request.

Please also observe the steps below to ensure optimal measurement conditions.

Ensure a straight lead-in and run-on area (“straight inlet”):

To prevent turbulence and the associated measurement inaccuracies, it is

recommended to stretch the tube out straight in the area where it is clamped.

Notes on measurement accuracy:

Place the sensor on straight part of the tubing to avoid measurement in areas of

recirculation, vortices or spin (preferably at least 6 inches before and behind the

sensor.)

Figure 32: Using Clamp-On Transducer on straight part of tubing

Allow the sensor to reach a stable temperature profile before pressing the offset adjust

button; e.g. if the sensor has ambient temperature and the liquid medium a different

temperature, it usually takes a few minutes until a stable temperature profile is

reached.

Use only the specified tubing.

Keep to the specified temperature range of the liquid:

calibration temperature ±5°C (± 9° F).

Make sure that there is sufficient acoustic coupling (see next chapter).

BioProTT™ Flow Measurement System Version 2.0 page 46 of 94

8.4 Received Signal Strength (RSS / Acoustic Coupling)

The received signal strength (RSS) of the ultrasound beam is an indication of the ultrasonic

signal transmission quality through liquid, tube and sensor. In this context the RSS value is

an indication for acoustic coupling. A sufficient acoustic coupling is a pre-requisite for

accurate measurements. Please ensure that the right calibration table with the correct table

and medium is selected.

Sufficient acoustic coupling is reached when the indicated RSS value is greater than

50%. Below 50% the flow measurement result is considered to be inaccurate and is

no longer shown on the different interfaces.

Note: The RSS value is shown on the display or indicated on analog and digital

interfaces when a valid sensor is connected.

8.5 Use of Acoustic Couplant

In order to ensure good repeatability of the flow measurement, acoustic couplant should be

applied to the tube before use. This is especially important when soft tubing (like e.g.

silicone) with a low wall thickness is used. Fitting this tubing into the measurement

channel might lead to concave folding of the tubing wall and misalignment (low coupling,

acoustic focuses) to the active sensor areas. A couplant ensures an easy slip of the tube in

the channel and provides repeatable alignment of the tubing wall.

To ensure optimal measurement em-tec will provide support about the use of couplant for

your specific tubing- sensor system.

Note: It will be shown in the calibration table and, if requested, in the calibration certificate

if a sensor was calibrated with or without the specified couplant. Please make sure that

your measurements are performed accordingly. Otherwise over amplification or too low

coupling might occur.

If deterioration in the coupling is evident e.g. during long-term measurements, this is

normally a sign that the couplant is drying out. To prevent failure of the measurement,

refresh the couplant on time.

Commercially available water based ultrasonic couplants can quickly dry out and lead to

poorer acoustic coupling. If other long time couplants e.g. vaseline are used, please ensure

compatibility with tubing and sensor material and follow manufacturer’s instructions.

To ensure optimal measurement and material compatibility em-tec can provide support to

identify the right couplant for your specific tubing, sensor and process requirements. If

necessary, couplant may also be used during calibration. Couplants distributed by em-tec

are available on request.

BioProTT™ Flow Measurement System Version 2.0 page 47 of 94

8.6 Zero Adjustment

Zero adjustment must be carried out before the start of every measurement to avoid

possible offsets of measured values.

When doing this, the transducer should be clamped at the right location on the tube. The

tube must be filled with fluid and the fluid must be at stationary. Make sure that there are

no air bubbles in the tube.

Also allow sufficient time to adjust for a stable temperature profile.

The system is now ready for a measurement or custom calibration by adjusting the

calibration factor if needed (see chapter CUSTOM CALIBRATION PROCEDURE).

Zero adjustment is also necessary when re-clamping the sensor or changing to a different

sensor table, which adjusts for different calibration parameters, like medium, temperature

or tubing.

Zero adjustment cannot be performed when the coupling is below 50%.

8.7 Performing Flow Measurements

8.7.1 Measurement Using the Analog Interface of BioProTT™ Flow Meters

The flow meter can be connected to data acquisition or process control systems with the

analog interface. The flow meter has two analog interfaces for transmitting the Flow and

RSS (acoustic coupling) values. The values are transmitted within the range of

(0)4-20 mA.

8.7.1.1 Received Signal Strength (RSS / Acoustic Coupling) on Analog Interface

The software maps the RSS value from 0-100% linearly to 4-20 mA. The current output of

for RSS can be interpreted in the following way:

0 mA indicates a broken cable, circuit or power leakage

Currents in range of 0.2 to 0.5 mA indicate an error or invalid sensor or sensor not

yet recognized

4 mA equals 0 % coupling / received signal strength (or no signal)

20 mA equals 100% RSS (excellent signal)

The RSS value should be at least 50 % respectively 12.0 mA. This indicates a good signal

and will ensure accurate flow measurements.

If the tube is too small or the transducer is too large, or there is no liquid or bubbles inside

the tubing this can also explain why the coupling is insufficient. Make sure that the tube

fits the transducer you are using. Also follow the procedure described in chapter

BioProTT™ Flow Measurement System Version 2.0 page 48 of 94

INSTALLATION OF THE BIOPROTT™ CLAMP-ON TRANSDUCER when clamping

the sensor onto the tube.

8.7.1.2 Flow Value on Analog Interface

The analog interface consist of a (0)4-20 mA current loop. Make sure that the voltage

supply for the current loop is 24 V DC.

The current output of the device for flow can be interpreted in the following way:

0 mA indicates a power outage or a broken cable respectively broken circuit

0.2 to 0.5 mA indicate invalid flow conditions e.g. low coupling, or an error

condition like invalid sensor or sensor not yet recognized

1 mA to < 4 mA corresponds to negative flow, with 1 mA corresponding to the

negative flow value applying the same gradient as with the positive flow values. This

means the mapping of the negative flow into a current value is truncated at 1 mA.

The 1 mA value corresponds to a negative flow of 28% of the nominal flow range of

the calibration table.

4 mA equals zero flow, with offset correction properly (sensor offset zeroed)

20 mA corresponds to the 1.5 x maximum nominal flow Qmax of the selected table

of the connected sensor. See the table below for the calculation of the specific

conversion factors that must be applied.

> 20 mA to 23 mA indicate flow out of nominal and extended range

Figure 33 shows the flow rate versus the output current for a sensor with a standard flow

range of 10000 ml/min. The conversion of a flow to a current is scaled to the Qmax in the

sensor table.

BioProTT™ Flow Measurement System Version 2.0 page 49 of 94

Figure 33: Output Current versus Flow

To calculate an output current into the flow value (with adjusted calibration factor

accounted for), use the following equation:

Flow[ml/min]=(Current Output [mA]- 4 mA)*Analog Current Conversion Factor

The analog current conversion factor depends on Qmax of the selected sensor table. This

value is indicated in the calibration report, indicated as range on the display, or given on

the serial port. The Qmax related to analog current conversion factors are indicated in

Table 17.

BioProTT™ Flow Measurement System Version 2.0 page 50 of 94

To calculate an output current into the “original” flow value as the system is measuring it

(without adjusted calibration factor), use the following equation:

Factorn Calibratio

range) flow(max Factor ConversionCurrent Analog4mA)-[mA](Current n]Flow[ml/mi

The current conversion factors and for different Qmax and the maximum reveres flow for

the analog flow output are listed below:

Standard flow

QMAX [ml/min]

for a calibration

factor of 1.00

Analog Current

Conversion Factor

[(ml/min)/mA]

Max. reverse flow

[ml/min] at 1 mA

4000 375 -1125

6000 562.5 -1688

8000 750 -2250

10000 937.5 -2813

20000 1875 -5625

50000 4687.5 -14063

100000 9375 -28125

Table 17: Range and Conversion factors for analog interface of BioProTT™ flow meters

Please note: the conversion factor is a fixed value based on the maximum flow range of a

sensor table when the calibration factor is set to 1.50. It is not dependent on the actual

chosen calibration factor. The conversion factor is calculated as follows:

mA]*[min

[ml]

16

max*5.1

mA 4-[mA]max Current

n]Qmax[ml/miFactor)n Calibratio(max Factor ConversionCurrent Analog

Q

BioProTT™ Flow Measurement System Version 2.0 page 51 of 94

Reliability of flow values at distortions, e.g. bubbles

Note: As Ultrasound is reflected by gaseousness and solid particles the flow reading can be

inaccurate when there are bubbles present in the system. Bubbles can be detected by

observing the coupling value. This value is either fluctuating (small bubbles) or constantly

low during the passage of larger bubbles or foam. To avoid inaccurate readings the flow is

disabled at insufficient coupling values below 50% and is enabled when the coupling

reaches 50% again.

As the flow values are averaged values we recommend ignoring flow values for 1-2

seconds after coupling has increased beyond 50%. This is especially true for the 10 second

mean flow transmitted by the serial interface where it can take up to 10 seconds to average

out the effect. Since the flow measurement system has no insufficient coupling status

storage hold time and no measurement error ignore time, assessment and handling of the

flow values after recovery of a drop in RSS value must be performed by the user according

to his needs.

Considerations when only working with analog interfaces e.g. within a process control cabinet

If only the analog interfaces of the flow meters without display, i.e. either

BioProTT™ FlowTrack or BioProTT™ FlowTrack DINrail, are used to process data, only

the flow and RSS value are available. BioProTT™ FlowTrack DINrail provides in addition

a remote zero option, to zero flow with a 24V DC pulse.

There are no means to

Change the calibration factor on the device. The default factor applied is 1.00.

Select a different calibration table than the first one (default table).

Check the maximum flow Qmax of the sensor. Note: the analog interface is mapped

according to Qmax of selected sensor table.

Zero the flow value via an analog interface of BioProTT™ FlowTrack; zeroing

manually is possible.

The calibration factor at device level is usually of lesser importance as this can actually be

included in the conversion of the analog flow signal into a flow value.

Only the first calibration table of a sensor can be used in such a scenario. The Qmax of the

sensor must be taken into account when converting the analog flow reading.

Remote zeroing of BioProTT™ FlowTrack is not possible in this scenario with a direct

analog input signal. However, based on the specific mapping of flow values into current

values it is possible to correct the analog flow reading accordingly and have a correct

reading for the process control available. This is done by recording the flow reading (or

flow offset) when the user has confirmed the zero condition, i.e. tubing filled, sensor lid

closed and no movement in the tube. This flow offset is then subtracted from the flow

reading on the interface and results in the actual offset corrected flow reading, which the

flow meter would provide when the zero button is pressed. The flow offset can be used

until the next zeroing needs to take place, e.g. because a sensor is changed or a new

process is started.

BioProTT™ Flow Measurement System Version 2.0 page 52 of 94

8.7.2 Measurement Using the Digital Interface (RS232)

Via the asynchronous serial communication port the different BioProTT™ flow meters can

communicate with a host system.

The interface must be initialized to the following parameters:

Baud rate 38400 Baud

Bits 8 data bits

Start bit 1

Stop bit 1

Parity no

No hardware (CTS/RTS) or software (Xon/Xoff) handshake

The data transmission runs without any request from the host system.

The following commands are available via the digital interface. Every command has to be

sent with carriage return <CR>.

The recommended delay between 2 commands should be not less than 1 second.

Command Meaning Remark

I<CR > Idle (stop communication)

R<CR> Restart (restart communication)

Z<CR> Zero Adjustment

S<CR> Sensor and Device Status Information Request

The sensor and device status information consists of an

ASCII data text string with the following information.

sensor serial number (16 bytes)

tube size of the sensor (16 bytes)

blank delimiter

tube type of selected calibration table (16 bytes)

blank delimiter

medium of selected calibration table (8 bytes)

temperature of the selected sensor table (6 bytes)

blank delimiter

number of sensor parameter tables (6 bytes)

Qmax of the selected sensor table (8 bytes)

<CR> and <LF>.

Flow meter serial number (16 bytes)

blank delimiter

SW-Version number (16 bytes)

The data string is finished by <CR> and <LF>.

T1<CR>...T7<CR> Request to set/ use calibration table number 1…7

Requesting a non existing table, will not change table.

Cx.yz<CR> set calibration factor (Values between 0.50 and 1.50

with 2 decimal digits e.g. C1.10<CR>)

Table 18: Commands for digital interface of BioProTT™ flow meters

BioProTT™ Flow Measurement System Version 2.0 page 53 of 94

The digital interface transmits every 100 ms (10 Hz) the following ASCII data string:

ee_ss_aaa_p.pp_fffffff_mmmmmmm_sssssss_tttttt_<CR><LF>

Meaning of the different Bytes:

Byte Meaning Remark

ee Hexadecimal error messages according to separate

table in Chapter Available Error Codes/Information

for Flow Meter

00 means No Error

“_” blank delimiter between the different parameters

(hex 0x20) Underscore is only

used to represent

blank

ss Hexadecimal status information

See Chapter Available Status Codes/Information

for Flow Meter for details

aaa Acoustic coupling / RSS in %; range 0 to 100%

p.pp Calibration factor

ffffff mean flow calculated over 100 ms in ASCII (range:

-999999 ml/min to +999999 ml/min) including

calibration factor

Note *)

mmmmmm mean flow calculated over 1 second in ASCII

(range: -999999 ml/min to +999999 ml/min) with

calibration factor applied

Note *)

ssssss mean flow calculated over 10 seconds in

ASCII(range: -999999 ml/min to +999999 ml/min)

including calibration factor

Note *)

tttttt board temperature in °C (38 := 38 °C)

<CR> 0x0d Carriage Return

<LF> 0x0a Linefeed

Table 19: Output of digital interface of BioProTT™ flow meters

*) Note: Beside numerical values these strings may contain the following characters:

a Minus symbol is used for negative values or to indicate blanked out values

(character hex “0x2D”).

^ indicating overflow condition (character hex: “0x5E”)

v indicating underflow condition (character hex: “0x76”)

Please note: if your own applications are programmed, both error and status byte must be

evaluated for a proper system assessment.

In case of flow distortions, e.g. as bubbles, the mean flow values may be affected.

BioProTT™ Flow Measurement System Version 2.0 page 54 of 94

Ensure that correct serial interface parameters are set and no invalid letters

are sent, as this can possibly lead to hang up of the system which requires

restarting of the system by powering off.

Below are some example output strings to help understanding the interface better. All

strings are terminated with <carriage return> <linefeed> and are shown below between

quotation marks (“). These output strings are transmitted periodically each 10 Hz.

Sensor working correctly, calibration factor 0.99, table 1 “00 00 100 0.99 7195 7193 6897 +41” <CR><LF>

Sensor working correctly, calibration factor 1.00, table 2, with negative flow values “00 04 100 1.00 -3588 -3590 -3589 +43” <CR><LF>

Sensor working correctly, calibration factor 1.34, table 2, with negative flow values “00 04 100 1.34 -4804 -4805 -4805 +43” <CR><LF>

Sensor connected, after stable zeroing “00 40 100 1.00 0 0 0 +43” <CR><LF>

Sensor connected, insufficient acoustic coupling / RSS value “00 24 34 0.99 ------- ------- ------- +43” <CR><LF>

Sensor connected, over temperature error 77°C on PCB inside device (data blanked

out) “1A 41 --- ---- ------- ------- ------- +77” <CR><LF>

Flow out of range condition: Flow exceeds +999999 ml/min; Cal_factor 1.01; “00 40 75 1.01 ^^^^^^^ ^^^^^^^ ^^^^^^^ +29” <CR><LF>

Flow out of range condition: Flow exceeds -999999 ml/min; Cal_factor 1.50;

“00 40 89 1.50 vvvvvvv vvvvvvv vvvvvvv +29” <CR><LF>

Sensor disconnected “00 E0 --- ---- ------- ------- ------- +35” <CR><LF>

BioProTT™ Flow Measurement System Version 2.0 page 55 of 94

Response to serial commands

Changing the calibration table (in sensors with more than one table) from table 1 to

table 2 by command “T2” <CR>. The status bit changes from 44 to 40. Due to

clearance of averaging storage the first 3 strings are blanked.

“00 44 100 1.00 0 0 0 +41” <CR><LF>

“00 44 100 1.00 0 0 0 +41” <CR><LF>

“00 60 0 1.00 ------- ------- ------- +41” <CR><LF>

“00 60 0 1.00 ------- ------- ------- +41” <CR><LF>

“00 60 0 1.00 ------- ------- ------- +41” <CR><LF>

“00 40 100 1.00 2 0 0 +41” <CR><LF>

“00 40 100 1.00 2 43 67 +41” <CR><LF>

“00 40 100 1.00 3 36 67 +41” <CR><LF>

Request for status information “S” <CR> when sensor is connected “83599 3/8" x 3/32" PVC Blood

37 °C 6 10000” <CR><LF>

“59915 V3.0.0.0 ” <CR><LF>

The sensor information above corresponds to the currently selected calibration table

and reads as follows in the first line:

o 83599 = Serial number of Sensor

o 3/8" x 3/32" = Tube sizing

o PVC = Tubing material of selected table

o Blood = Calibration media of selected table

o 37°C = Calibration temperature of selected table

o 6 = Number of Calibration Tables programmed in Sensor

o 10000 = Flow range of selected calibration table (QMax)

The device information in the second line contains:

o 59915 = Serial number of BioProTT device

o V3.0.0.0 = Software version

Request for status information “S” <CR><LF>.when no sensor is connected “-------------------------------- -------- -------- -----

--- ------ --------” <CR><LF>

“59916 V3.0.0.0 ” <CR><LF>

The sensor information in the first line is blanked.

The device information in the second line contains:

o 59915 = Serial number of BioProTT device

o V3.0.0.0 = Software version

BioProTT™ Flow Measurement System Version 2.0 page 56 of 94

8.7.3 Measurements Using BioProTT™ FlowTrack plus

BioProTT™ FlowTrack plus is equipped with a display and supports the following

functions via the graphical user interface (GUI). The screens shown are only examples of

actual possible screen contents.

When using the display and changing settings like calibration table or

calibration factor in parallel via the digital interface the change via the

digital interface will only be visible at the display when the specific screen

is “flashed” i.e. if the user changes to a different screen and back. The new

value will however take immediate effect on the measurements.

8.7.3.1 Flow Value and Coupling

Flow value and coupling are displayed as follows.

Screen Meaning

Flow value and coupling (RSS) are displayed as follows.

The flow value is calculated using the selected calibration table

and calibration factor.

When the flow value is greater than ±999 ml/min, the units

displayed are changed to [l/min].

In order to adjust the “zero” offset, please ensure that there is

sufficient coupling and no flow in the tubing and press the

“zero” button until the flow value is reset to zero.

The current RSS value is shown if a sensor is connected. If

coupling becomes insufficient (<50%), the flow value will be

blanked out. For details see Received Signal Strength (RSS /

Acoustic Coupling).

Table 20: Flow and coupling screen of BioProTT™ FlowTrack plus

BioProTT™ Flow Measurement System Version 2.0 page 57 of 94

8.7.3.2 Selection of Calibration Table

A Clamp-On Transducer can hold up to 7 calibration tables, which can be selected via the

appropriate screen, using the Arrow and Enter buttons.

Screen Meaning

The calibration table screen can be accessed only when a valid

sensor is connected.

The calibration table is an attribute connected to the sensor.

When the device is powered on/off and/or the sensor is changed

or reconnected the table is reset to table one (1).

Selecting a different calibration table does not change the

calibration factor.

When selecting a new table, “sel” appears on the screen. Please

confirm the new selection with the Enter button.

If the calibration table is changed, please zero the system at zero

flow for optimal measurement accuracy. Please also ensure that

the actual system setup and the calibration parameters shown in

the table are matching, otherwise inaccurate readings can occur.

The upper value of the range indicates the Qmax of the selected

table. Table 21: Calibration table screen of BioProTT™ FlowTrack plus

BioProTT™ Flow Measurement System Version 2.0 page 58 of 94

8.7.3.3 Selection of Calibration Factor

It is possible to adjust the calibration factor to accommodate e.g. a different medium or

temperature compared to the calibration settings of the connected Clamp-On Transducer.

The calibration factor can be adjusted in the following range: 0.50 to 1.50

Screen Meaning

The calibration factor screen can only be accessed when a valid

sensor is connected.

While on the calibration screen, the actual flow value is shown.

The default setting of the calibration factor is one (1.00).

The calibration factor is an attribute connected to the sensor.

When the device is powered on/off and/or the sensor is changed

or reconnected the factor is reset to one.

By pressing the Enter button and using the Arrow buttons the

factor can be changed in a range from 0.50 to 1.50 in 0.01 steps.

Pressing the Enter button again will store the value.

Changing the calibration factor will instantly affect the flow

value displayed on the screen and also on the analog and digital

output. Changes can be observed directly at the indicated flow,

below the factor.

In case of insufficient coupling the flow value is blanked out.

Table 22: Calibration factor screen of BioProTT™ FlowTrack plus

The calibration factor alters the flow by the following equation:

Flow value Calibration Factor = Calibration Factor * Flow value (Measured in sensor)

See chapter Custom Calibration Procedure for details.

BioProTT™ Flow Measurement System Version 2.0 page 59 of 94

8.7.3.4 Adjustment of Brightness

It is possible to adjust the display brightness as follows:

Screen Meaning

The default value is 50%. The value can be set to 25, 50, 75 and

100 %.

By pressing the Enter button the menu is activated by indicating

“sel” and the user can select the value by the Arrow Buttons.

Repeated pressing of the Enter button stores the selected value.

Table 23: Brightness screen of BioProTT™ FlowTrack plus

8.7.3.5 Totalizing

There is a built in flow totalizer available via the following screen.

Screen Meaning

The totalizer screen can be accessed only when a valid sensor

is connected. When entering the totalizing screen, initially zero

elapsed time and total flow value are shown.

After the totalizing has been engaged with the Enter button, the

totalizer indicates “Hold” and shows elapsed time and total

flow value in the units shown on the screen.

If the coupling is insufficient or a sensor table has been

selected an exclamation mark appears.

This indicates that the indicated totalized volume can differ

from real total flow due to changes in the liquid or the

calibration.

Table 24: Totalizing screen of BioProTT™ FlowTrack plus

BioProTT™ Flow Measurement System Version 2.0 page 60 of 94

The total flow is displayed in unit [ml] with three digits when the value is below one

liter (< 1 l). When the value is equal or higher than one liter the unit is changed to [l].

In this case the maximum displayed total flow is 999.9 liter whereas the digits for ml

shall be one. When the sum exceeds a value of 999.9 liters the lower left corner

shows 1 m³ and the totalizer will continue operating. The maximum digits for the m³

counter is 3 digits which corresponds to a maximum of 999 m³.

The totalizer has two limits. Either a totalizing time of 999:59:59 hours (approx. 6

weeks) or a totalized flow of 999 m³ 999.9 l. Whenever one of this limits is reached

the Totalizer is stopped and the string “OVR” (Overrun) is shown in the display. The

values for Totalizing Time and Total Flow are hold and shown on the display until

the user resets the totalizer, turns power off or disconnects the sensor.

When entering the totalizer screen the screen shows

o “Hold” when the totalizer is running and

o “Run” when the totalizer has not yet been started or stopped.

o By pressing the Enter Button the user is able to start/stop the totalizer.

Entering other menus of the device while the totalizer is running will not

have an effect on the totalizer.

When the totalizer is running and the Enter Button is pressed the totalizer is stopped.

The display will show the value for the total flow at this point and indicate “Run”

status.

Pressing the enter button again shall resume the totalizing progress starting with the

Totalizing Time and Total Flow before the stop. Additionally the Totalizing Time

value shall stop flashing.

While in the totalizer menu the totalizer is reset by pressing the Zero/Reset button.

The resetting can be performed while the totalizer is in the Hold or Run status. When

the totalizer is reset during run it continues working, starting with a zero value.

In case of insufficient low coupling (< 50%) during the run of the totalizer

an exclamation mark is shown next to the string “Total Flow” (Indication

for the user). During the time period of insufficient coupling the Totalizing

Time is stopped and the Total Flow value not changed (no further

summation).

If the coupling increases back to a “good” signal (value higher or equal to

50 %; e.g. after bubble has passed) the totalizing process resumes with the

values before the coupling drop. The exclamation mark shall stay on the

screen until the zero is pressed.

If a sensor is disconnected during the totalizer run the data are deleted.

BioProTT™ Flow Measurement System Version 2.0 page 61 of 94

8.7.3.6 Status and Error Information

The following information is displayed on the status and error screens.

Screen Meaning

When the system is starting up, the message “Initializing

Device” is displayed on the screen.

When accessing the status information screen, SW version and

serial number (SN) of the BioProTT™ flow meter are displayed.

When a sensor is connected and recognized its serial number

(CT) is displayed too.

When an error is detected, an error screen is shown with the

respective error code and message. An example is shown here.

Please refer to chapter AVAILABLE ERROR

CODES/INFORMATION FOR FLOW METER for detailed

error information. Note error information before turning the

device off.

This information is important for troubleshooting.

Table 25: Status and error screens of BioProTT™ FlowTrack plus

8.7.4 Custom Calibration Procedure

The BioProTT™ provides the possibility to do an on-site calibration adjustment to account

for different media, tubing wall thickness and temperatures. Strong differences in the

acoustic properties of the tubing and medium can lead to insufficient acoustic signal

strength (RSS) and affect the accuracy of the measurement.

The customization can be performed by sending the command “C” followed by the

calibration factor via the serial interface and confirmed with carriage return (see chapter

MEASUREMENT USING THE DIGITAL INTERFACE (RS232)). Additionally the

factor can be set via the display for the BioProTT FlowTrack plus (see chapter

SELECTION OF CALIBRATION FACTOR)

BioProTT™ Flow Measurement System Version 2.0 page 62 of 94

The calibration factor can be adjusted in a range from 0.50 to 1.50 in 0.01 steps with the

default value being 1.00. The internally derived flow value is multiplied with the factor.

The result is used to display the actual flow value on the different interface(s).

In order to do an on-site calibration adjustment you can set up a circulation loop and

compare the actual flow values with the values of a reference device. The reference device

can either be another flow meter or you can perform a timed collection of the flow.

Please note: the timed collection can also have a measurement error

depending on the collection time (measurement) and accuracy of the used

scale; generally the measurement error is diminished by a longer collection

time.

We recommend doing the custom calibration procedure for at least two different flow

rates, ideally in the typical flow range of your application.

With the first flow rate you compare the flow value with the reference device (or timed

collection) and adjust the calibration factor till both values match. With the second flow

rate you can check if both values still match in this range.

Please always ensure (as for any other measurements), that the system had enough time to

reach a stable temperature profile and that zeroing was done properly.

Please note: due to the different physical properties of liquids it cannot be

guaranteed that the calibration factor is valid for the complete flow range of

a sensor given in the datasheet. This can be checked with a reference device

or timed collection.

BioProTT™ Flow Measurement System Version 2.0 page 63 of 94

9 Status Information, Troubleshooting, Support and Service

9.1 Available Status Codes/Information for Flow Meter

The following status information is provided via the different interfaces:

Description Digital

Interface

RS 232

Status

Byte

“ss” *)

Analog

Interface

4-20mA

Flow

Analog

Interface

4-20mA

RSS

Display

**)

MEASURES

See also

chapter

TROUBLE-

SHOOTING

Sensor

disconnected

Bit 7 = 1 < 0.5 mA < 0.5 mA Flow screen

shows “-“

RSS shown

Connect

sensor, check

cable

Set if

uncalibrated

transit time

value between

-3000 and

+ 3000

Bit 6 = 1 Flow value RSS value Flow value

RSS value

None

Insufficient

coupling

(RSS < 50%)

Bit 5 = 1 < 0.5 mA RSS value Flow screen

shows “-“

RSS value

Check

coupling

Selection of

calibration

table

Bit 4,3,2

are set

according

to selected

calibration

table

N/A N/A Selected

table shown

on screen

None

Check if

correct table

is selected

Flow invalid Bit 1 = 1 < 0.5 mA RSS value Flow screen

shows “-“

RSS value

Check

coupling

Temperature

to high

Bit 0 = 1 < 0.5 mA RSS value Flow screen

shows “-“

RSS value

See troubles-

hooting

Table 26: Status codes for BioProTT™ flow meters

*) Bit value is converted into hexadecimal value when transmitted with string over digital

interface

**) BioProTT™ FlowTrack plus only – information on flow screen

BioProTT™ Flow Measurement System Version 2.0 page 64 of 94

9.2 Available Error Codes/Information for Flow Meter

In case of an error, the flow meter will not allow any additional input or

change of screen. However, the device can be powered off and on again

and the sensor can be reconnected. Please observe also the chapter

TROUBLESHOOTING.

If the error continues to appear, the device needs to be serviced, see chapter

SUPPORT AND SERVICE.

The following error information is provided via the different interfaces:

Description Digital

Interface

RS 232 Error

Byte “ee”

Analog

Interface

4-20mA

Flow

Analog

Interface

4-20mA

RSS

Display

(BioProTT™

FlowTrack plus only –

on flow screen)

No error 00 Current

Analog

Flow value

Current

Analog RSS

value

Flow value

RSS value

FLASH error

during start up

test

01 < 0.5mA < 0.5mA Error page w/

01

FLASH error

during runtime

test

02 < 0.5mA < 0.5mA Error page w/

02

RAM error

during runtime

RAM test

08 < 0.5mA < 0.5mA Error page w/

08

Error in task

handling

0F < 0.5mA < 0.5mA Error page w/

0F

EEPROM test

error (Sensor)

19 < 0.5mA < 0.5mA Error page w/

19

Board

temperature

error (> 75°C)

1A < 0.5mA < 0.5mA Error page w/

1A

+15V check

error

1B < 0.5mA < 0.5mA Error page w/

1B

-15V check

error

1C < 0.5mA < 0.5mA Error page w/

1C

EEPROM CRC

error (Sensor)

21 < 0.5mA < 0.5mA Error page w/

21

EEPROM

factor inconsis-

tent (Sensor)

22 < 0.5mA < 0.5mA Error page w/

22

BioProTT™ Flow Measurement System Version 2.0 page 65 of 94

Description Digital

Interface

RS 232 Error

Byte “ee”

Analog

Interface

4-20mA

Flow

Analog

Interface

4-20mA

RSS

Display

(BioProTT™

FlowTrack plus only –

on flow screen)

EEPROM of

the sensor not

consistent

(Sensor)

27 < 0.5mA < 0.5mA Error page w/

27

Complementary

values in

EEPROM not

equal (Sensor)

29 < 0.5mA < 0.5mA Error page w/

29

Illegal sensor

type or sensor

version

2A < 0.5mA < 0.5mA Error page w/

2A

Sensor

parameter not

correct

2B < 0.5mA < 0.5mA Error page w/

0x2B

Wrong CRC in

Burst table

2E < 0.5mA < 0.5mA Error page w/

0x2E Table 27: Overview of error information on different interfaces of BioProTT™ flow meters

BioProTT™ Flow Measurement System Version 2.0 page 66 of 94

9.3 Troubleshooting

If you experience any problems with the flow measurement system, please try the

following solutions:

Problem Meaning / Solution

General

LED is not on

(only BioProTT™

FlowTrack and

BioProTT™

FlowTrack DINrail)

Power supply may be disconnected. Please check correct power

supply see chapter POWERING OF BIOPROTT™ Flow Meter

and disconnect Flow and RSS analog circuits.

If after correct power supply switching on and off leads to the

same behavior device needs to be serviced

Unexpected change

of RSS / acoustic

coupling value

One potential cause could be that calibration parameters do not

match the selected calibration table. Please check and adjust

accordingly.

Check also tubing for deformation, bubbles or debris.

“on”/”off” flow

reading (toggling

values)

A potential cause for a flow reading that toggles between an actual

value (“on”) and no value (“off”) can be disturbances in the flow

like small bubbles.

Please ensure that the medium is bubble free and apply acoustic

couplant. In case of a couplant used, check calibration conditions

and if necessary perform a custom calibration with check for

calibration factor.

Please also see chapter FLOW VALUE and handling of flow

distortions.

No Flow and RSS

reading

No sensor recognized; please check sensor connection incl. sensor

extension cable if deployed.

Note: flow sensors with higher flow rates are only supported on

flow meters with SW V3.0 and higher. Please check Qmax of

sensor tables, e.g. shown on calibration report.

Flow reading with

no flow present

Zero offset again and make sure that the liquid is NOT in motion

during zeroing.

Flow meter gets

very warm

Switch device off and let it cool down. Check if the flow meter is

exposed to an external heat source / direct sun radiation or if it is

covered up so that no heat can be dissipated. Change location

accordingly.

Display (BioProTT™ FlowTrack plus only)

Display is blank or

hangs

Power supply may be disconnected. Please check correct power

supply; see chapter INITIAL SET UP.

If after correct power supply switching on and off leads to the

same behavior device needs to be serviced.

Missing

information/values

in screens

Change and reselect affected page with Up- and Down arrow

buttons. This should lead to a refresh of the displayed information.

BioProTT™ Flow Measurement System Version 2.0 page 67 of 94

Problem Meaning / Solution

Digital Output

Erroneous offset on

flow reading

- The calibration factor might have to be re-adjusted

- Tubing might be out of specification

- Wrong temperature range

- Check offset with no flow in tubing and perform zeroing again

No reading on

digital interface

- Ensure that connector cable is not broken

- Ensure proper fit of plug connector

- Ensure correct adapters (e.g. from RS 232 to USB) and that the

PC program (e.g. LabView) is correctly reading the serial port

connected to the flow meter. In addition the correct drivers must

be installed prior to use of the adapters

- Ensure correct port parameters

- Ensure that the buffer of your serial communication device is

properly refreshed

- Ensure that no NULL MODEM cable is used

- Check if “R” restarts connection (sending data)

- Ensure that only commands are used listed in this document

Analog Output – Flow Output

0 mA A reading of 0 mA could indicate the following:

- broken circuit

- missing power supply for the current loop; usually power is

supplied via process control

See also chapter AVAILABLE ERROR

CODES/INFORMATION FOR FLOW METER

0.2-0.5mA A reading in this range could indicate

- insufficient coupling

- no sensor recognized

- a circuit error

- device error

See also chapter AVAILABLE ERROR

CODES/INFORMATION FOR FLOW METER

> 0.5 to < 1 mA Undefined output

> 20 mA The flow range could exceed 20 mA when the flow value exceeds

the specified range; it is limited to maximum of

approx. 23 mA.

Analog Output – RSS Output

0 mA A reading of 0 mA could indicate the following:

- broken circuit

- missing power supply for the current loop

See also chapter AVAILABLE ERROR

CODES/INFORMATION FOR FLOW METER

BioProTT™ Flow Measurement System Version 2.0 page 68 of 94

Problem Meaning / Solution

0.2-0.5 mA A reading in this range could indicate

- no sensor recognized

- a circuit error

- device error

See also chapter AVAILABLE ERROR

CODES/INFORMATION FOR FLOW METER

> 0.5 to <4 mA Undefined output

Table 28: Troubleshooting for BioProTT™ flow meters

The maximum flow range depends on the size of the Clamp-On Transducer, please check

in the Clamp-on Transducer datasheet whether the given flow range is suitable for your

application.

BioProTT™ Flow Measurement System Version 2.0 page 69 of 94

9.4 Checking Current Loops of Flow Meter

Current loops (Flow or RSS) can be tested with different measurement configurations

described in this chapter. Please make always sure that a Clamp-On Transducer is

connected and actual flow is occurring so that a reading is possible. The checks can be

performed similarly with the DIN rail variant of the flow meter, although not shown below.

Note: the flow meter needs to be powered properly.

9.4.1 Method 1 – Using Power from Flow Meter and Measuring with Ampere

meter

Figure 34: BioProTT flow meter rear view with one analog plug connected -example w/ RSS socket

Pin No Signal Wire colour

(in supplied cabling)

Connection

1 24 V (24C_in) Brown Connect pin 1 with pin 2

2 Current Loop + White [see above]

3 Ground in

(GND_in)

Blue Connect pin 3 to the appropriate

connector of your reading device /

data logger, e.g. “return” or “-“ of

an ampere meter

4 Current Loop - Black Connect pin 4 to appropriate

connector of your reading device /

data logger e.g. “send” or “+” of

an ampere meter Table 29: Wiring for BioProTT™ flow meters with round connectors for loop check w/ ampere meter

Figure 35: Schematics - looking at pins of rear of flow meter of either Flow or RSS socket

BioProTT™ Flow Measurement System Version 2.0 page 70 of 94

Tolerances for current measurement (range of ampere meter 0-20 mA DC) are as follows:

Max 1 % + 5 digits at > 50.000 mA range

Internal resistance < 10 Ohm

9.4.2 Method 2 – Providing Power Externally in Current Loop and Measuring

with Ampere Meter

Pin No Signal Wire colour

(in supplied cabling)

Connection

1 24 V (24C_in) Brown Connect pin 1 with pin 2

2 Current Loop + White Connect pin 2 to “+” of battery /

24V DC power source

3 Ground in

(GND_in)

Blue Connect pin 3 to the appropriate

connector of your reading device /

data logger, e.g. “return” or “-“ of

an ampere meter. Close loop by

connecting also to “-“ of battery

4 Current Loop - Black Connect pin 4 to appropriate

connector of your reading device /

ampere meter e.g. “send” or “+” Table 30: Wiring for BioProTT™ flow meters with round connectors for loop check w/ external power

Figure 36: Schematics - looking at pins of rear of flow meter of either Flow or RSS socket

Tolerances for current measurement (range of ampere meter 0-20 mA DC) are as follows:

Max 1 % + 5 digits at > 50.000 mA range

Internal resistance < 10 Ohm

BioProTT™ Flow Measurement System Version 2.0 page 71 of 94

9.4.3 Method 3 – Checking Functionality with Voltmeter

It is also possible to test the functionality of the current loops (Flow or RSS) with a

voltmeter. In this case, a resistor must be applied. Ohm’s law has to be observed to

calculate the correct current through the resistor. The loop can be powered either externally

or via flow meter as shown in the diagram below.

To avoid damages overvoltage protection of the voltmeter of at least 30 V DC is

recommend. The table indicates examples of conversion of 0-20 mA signal to

voltage.

Figure 37: Schematics - looking at pins of rear of flow meter of either Flow or RSS socket

Tolerances for voltage measurement (range of voltmeter 25 V DC or higher) are as

follows.

Resistor min 100 Ohm, max. 600 Ohm,

Tolerance of resistor min 1 % better 0,1 %

Internal impedance of voltmeter minimal 1 MOhm

Conversion

Voltage

Range

Recommended

Measuring

Resistance

Current to Voltage

Conversion Factor

Zero Point Voltage at

4 mA

0-2 V 100 Ohm 0.1 V per mA 0.4 V

0-5 V 250 Ohm 0.25 V per mA 1.0 V

0-10 V 500 Ohm 0.5 V per mA 2.0 V Table 31: Voltage conversion for BioProTT™ flow meters for loop check w/ voltmeter

When no electrical load or high resistance resistor or is used the voltage between Current

Loop (+) and (-) will be set nearly to the voltage (24 V) of the power supply.

Calculating the flow by the measured voltage

range) flow(max Factor Conversion*1000Resistance

[V]) VoltagePoint Zero-[V] (Voltage n]Flow[ml/mi

Conversion factor (Qmax) see chapter FLOW VALUE

Calculating the acoustic coupling RSS by the measured voltage

100Resistance*0.016

)Voltage[V]Point Zero-[V] (Voltage RSS[%]

BioProTT™ Flow Measurement System Version 2.0 page 72 of 94

9.5 Recommended Periodic Checks

If the recommended measures are not successful, please contact the technical support.

Make a note of the serial number of the relevant device or accessory and also record the

software version.

It is recommended that periodic inspections are carried out at least every 24

months to make sure the device complies with the safety requirements and

delivers the required measuring accuracy, taking into account the valid

international and national standards and regulations. Use of the following

inspection record is recommended by a qualified person who has adequate

training, knowledge, and practical experience to perform these tests.

If the checks reveal faults or limits are exceeded, stop using the device and

contact our Service department (see chapter SUPPORT AND SERVICE)

General

Test organisation,

Name of Tester

Initial test (reference value):

Recurrent test:

Test after repair:

Device BioProTT™

Type

Reference no.

SN

Accessories

Visual Inspection

Visual check for mechanical damages that could impair functions.

Visual inspection Passed? Device & accessories configured according these operating instructions and not changed? Yes No

Power supply cord Yes No Other connection cables, e.g. flow sensor extension cable Yes No Housing Yes No Soiling Yes No Is the type plate legible? Yes No Are the safety relevant markings legible? Yes No Are the accessories used compatible? Yes No

Checking Functionality according the Operating Instructions Switch the device on, connect a flow sensor and check the following points:

Function check Passed? Does the LED light up (BioProTT FlowTrack and BioProTT FlowTrack DINrail) or the

display light up when the device is switched on?

Is the brightness increasable and reducible? (see chapter INITIAL SET UP)

Yes No

Perform a function check in accordance with the Operating Instructions

(see chapter PERFORMING MEASUREMENTS ON TUBING SYSTEMS USING

BIOPROTT™ CLAMP-ON TRANSDUCERS)

Yes No

Clamp-On Transducer: Check coupling value with tube filled:

The acoustic coupling value must be sufficient (see chapter FLOW VALUE AND

COUPLING).

Yes No

BioProTT™ Flow Measurement System Version 2.0 page 73 of 94

9.6 Support and Service

If the recommended measures are not successful, please contact our Service department.

Make a note of the serial number of the relevant device or accessory before you contact our

staff. Please also state the software version.

9.6.1 Contact Information for Technical Support

Technical support in USA / for the region Americas is provided by:

em-tec Flow Technology LP

PO Box 132

Berkshire, NY 13736

USA

Phone: (607) 330-2737

E-mail: [email protected]

Technical support for rest of world is provided by:

em-tec GmbH

Lerchenberg 20

86923 Finning

Germany

Phone: +49 8806 9236 0

E-mail: [email protected]

BioProTT™ Flow Measurement System Version 2.0 page 74 of 94

9.6.2 Returns / RMA

The following procedures must be carried out before a flow meter is returned e.g. for repair

or calibration:

Please request an RMA (Return Material Authorization) number and enclose a

detailed description of the error or problem and name and telephone number/e-mail

address of the contact partner.

Always enclose a fully completed and signed “Declaration of contamination” form.

This will ensure that your order can be handled properly.

o Please note: a copy of the “Declaration of Contamination” can be found in

chapter RETURN INFORMATION AND DECLARATION OF

CONTAMINATION.

Enclose special handling instructions if necessary, for example a technical data sheet

(TDS) or a material safety data sheet (MSDS).

Remove all fluid residues. Pay special attention to the grooves for seals and crevices

which could contain fluid residues. This is particularly important if the fluid is

hazardous to health, e.g. flammable, toxic, caustic, carcinogenic, etc.

Do not return a device if you are not absolutely certain that all traces of

hazardous substances have been removed, e.g. substances which have

penetrated crevices or diffused through plastic.

Costs incurred for waste disposal and injury (caustic burns, etc.) due to inadequate

cleaning will be charged to the owner/operator.

BioProTT™ Flow Measurement System Version 2.0 page 75 of 94

10 Cleaning, Low Level Disinfection and Maintenance

Details on how to clean and, if necessary, disinfect your BioProTT™ flow meter, the

applied parts and the accessories are given in this chapter.

Please observe GENERAL SAFETY AND IMPORTANT INFORMATION.

10.1 Cleaning of the BioProTT Flow Meter

The BioProTT™ FlowTrack and FlowTrack plus devices can be cleaned by wiping it with

a damp cloth and warm water. For low level disinfection an aqueous 70% isopropanol

solution can be used.

The BioProTT™ FlowTrack DINrail device can be cleaned by wiping it with a dry or

lightly damp cloth and warm water. For low level disinfection an aqueous 70% isopropanol

solution can be used. Avoid ingress of any fluids.

When using other cleaning or disinfection agents make sure that they do not

interfere with the housing of the device and have a nearly neutral pH.

Do not use cleaning agents that scratch, are abrasive or are corrosive (e.g.

scouring powder)!

The accessories must only be plugged in when dry. The accessory connectors

(e.g. plug of power supply unit) must not be immersed in liquid.

The concentrations and exposure times of cleaning agents and disinfectants

specified by the manufacturer must be closely observed, including material

compatibility.

BioProTT™ Flow Measurement System Version 2.0 page 76 of 94

10.2 Cleaning and Disinfection of the Clamp-On Transducer

Risk of ingress of fluids!

Make sure that cleaning agents or disinfectants do not enter the

device. The accessory connecting plugs must only be plugged in

when dry.

No machine cleaning or sterilization!

Clamp-On Transducers are not suitable for machine cleaning

methods. Sterilization methods, particularly steam sterilization and

autoclaving, must not be used.

Inspection following every cleaning/disinfection procedure

Check your Clamp-On Transducer for damage and wear following

every cleaning/disinfection procedure. To prevent the ingress of

fluid, check carefully for cracks or fractures and make sure the

insulation has not been damaged.

No ultrasonic cleaning!

Use of an ultrasonic bath is not permissible as this could adversely

affect the function.

Damaged sensors must not be used!

We recommend the use of commercially available mild domestic neutral cleaning agents or

a 70% Isopropanol solution to clean the BioProTT™ Clamp-On Transducers. To prevent

encrustation, heavy soiling and residual ultrasonic couplant should be removed

immediately following each application.

All soiling must be removed in the clamping zone in particular.

The chemical material resistivity of the following surface cleaning and disinfection agents

was proofed:

Trade name Manufacturer Contact

Bacillol® AF, Bacillol® AF Tissues,

Bacillol® Tissues and Bacillol® Wipes

Bode Chemie GmbH www.bode-chemie.com

Bacillol® CR Bode Chemie GmbH www.bode-chemie.com

Before usage please read the manufacturer information and safety data

sheets of the agent carefully and follow the instructions and

precautions.

BioProTT™ Flow Measurement System Version 2.0 page 77 of 94

11 Approval Information

11.1 EU CE Compliance

The BioProTT™ flow measurement system is CE marked and conforms to the

requirements of EU Directive 2004/108 EC (EMC) and 2011/65/EU (RoHS).

Declaration of conformity see chapter EU DECLARATION OF CONFORMITY.

11.2 US FCC Compliance

The BioProTT™ flow measurement system complies with Part 15 of the FCC Rules.

Operation is subject to the following two conditions:

(1) This device may not cause harmful interference, and

(2) This device must accept any interference received, including interference that

may cause undesired operation.

This equipment has been tested and found to comply with the limits for a Class B digital

device, pursuant to part 15 of the FCC Rules. These limits are designed to provide

reasonable protection against harmful interference in a residential installation. This

equipment generates uses and can radiate radio frequency energy and, if not installed and

used in accordance with the instructions, may cause harmful interference to radio

communications. However, there is no guarantee that interference will not occur in a

particular installation. If this equipment does cause harmful interference to radio or

television reception, which can be determined by turning the equipment off and on, the

user is encouraged to try to correct the interference by one or more of the following

measures:

Reorient or relocate the receiving antenna.

Increase the separation between the equipment and receiver.

Connect the equipment into an outlet on a circuit different from that to which the

receiver is connected.

Consult the dealer or an experienced radio/TV technician for help.

A copy of the certificate can be found in chapter FCC CERTIFICATE.

11.3 Canada Compliance

This Class B digital apparatus complies with Canadian ICES-003.

BioProTT™ Flow Measurement System Version 2.0 page 78 of 94

11.4 Electrical Safety and Electromagnetic Compatibility

The BioProTT™ is state-of-the-art technology. Although the requirements of

EN/IEC 61010-1 and EN/IEC 61326-1 were taken into consideration during development

and manufacturing, the user may be at risk if the device is used improperly.

Electrical Installation Requirements

Please follow the general safety information when installing the

BioProTT™ flow measurement system. Please observe any separate relevant safety and

technical information of other electrical components used.

Electromagnetic Compatibility Requirements

The emission and immunity are in accordance with the standards listed in chapter technical

data.

BioProTT™ Flow Measurement System Version 2.0 page 79 of 94

12 Technical Description

This chapter provides information on the technical performance of the

BioProTT™ flow measurement system.

12.1 Technical Data BioProTT™ Flow Meter

BioProTT™ FlowTrack and BioProTT™ FlowTrack plus

Operating principle Ultrasonic Transit-Time Flow Measurement (TTFM)

Device dimensions (HxWxD) 65 mm x 110 mm x 140 mm

Face plate dimensions of panel

mount variant

78 mm x 126 mm x 6 mm

Weight 500 g (bench top variant)

510 g (panel mount variant)

Case material Aluminum

IP code IP65

Panel mount variant: IP66 at panel when mounted

Supply voltage 24 V DC (± 10%) supplied external or by process

control system (PCS), min. 300 mA

Power consumption typically 3 Watt, max 8 Watt

Round connector w/ pins Circular industry 3 and 4 pole connectors M8,

according to IEC/EN 61076-2-104 (2008); Type

Hirschman (ELST 3308RVFM805/ELST

4408RVFM805) or 768/718 series from Binder

Cage clamps (WAGO 222-412) on 4 pole cable.

Flow sensor extension cable

1.1 m, to allow flow sensor connection from the

outside control cabinet.

Included as standard cable with panel mount version

BioProTT™ FlowTrack DINrail

Operating principle Transit-time Ultrasound

Dimensions (HxWxD) 110 mm x 65 mm x 150 mm

Total length with extension

cable plugged in

200 mm

Weight 650 g

Case material Aluminum

IP code IP20

Supply voltage 24 VDC (± 10 %) supplied external or by process

control system (PCS), min. 300 mA

Power consumption typically 3 Watt, max 8 Watt

DIN rail mounting type Top hat section type O according to EN 50022,

size: 35 mm x 7.5 mm (or compatible)

Flow sensor extension cable 1 m, to allow flow sensor connection from the

outside control cabinet

Terminal block connector Type: Phoenix Types

DFK-MC 1,5/10-GF-3,81 (device)

FK-MCP 1,5/10-STF-3,81 (at cable)

BioProTT™ Flow Measurement System Version 2.0 page 80 of 94

Interfaces:

Interface types 1 x digital RS-232 (RxD, TxD, GND)

2 x analog 0 - 20 mA

1 x Remote Zero – Pulse (DINRail only)

RS-232 cable 9-pin D-Sub-cable (F), USB adapter included

Interface data rates digital:

data strings are sent every 0.1 s; containing RSS,

flow values, status and error information and device

temperature

flow values are available in the data string with

following averaging times: 0.1 s, 1 s, 10 s

analog:

output update rate is 0.1 s with the 1s average flow

value being mapped

display: BioProTT™ FlowTrack plus only

the display values are refreshed at least every second,

the flow value show is the 1 s average flow value

Analog interface: nominal 0-20 mA analog with 4 mA being defined as

zero value for flow and acoustic coupling (RSS);

max. current output limited to 23 mA

1 kOhm ±5 % at 24 V, 16-bit resolution, passive or

active mode

Accuracy of digital to analog conversion: 1 %.

Ambient conditions:

Air pressure 70 kPa to 106 kPa

Operating altitude up to 2000 m (6600 feet)

Operating temperature range 10º C to 40º C (50° F to 104° F)

Storage temperature range –20° C to 45° C (-4° F to 113° F)

Transport temperature range –20° C to 55° C (-4° F to 131° F)

Relative air humidity:

Operation 30 % to 75 % (non-condensing)

Storage and transportation 10 % to 96 % (non-condensing)

EMC:

Safety:

Conforms to:

IEC/EN 61326-1

EN 55011, CISPR 11 Class B, Group 1

CISPR 22 Class B

47 CFR 15 / B

Conforms to:

IEC/EN 61010-1 3. Edition

BioProTT™ Flow Measurement System Version 2.0 page 81 of 94

Power supply unit shipped with flow meter:

Type: Friwo FW 7660, GPP 10, part number 1827497

Dimensions (HxWxD) 40 mm x 58 mm x 150 mm w/o primary adapter plug

Input voltage / current: 100 - 240 VAC (± 10 %) / 50-60 Hz, max. 250 mA

Output voltage: 24 VDC (± 5 %), 420 mA

Power consumption: max. 10 VA

Fuse protection: none

Power supply connection:

EMC:

Safety:

Worldwide use with exchangeable primary adapters

Conforms to:

EN 55011, EN 55022 / B,

FCC 47 part 15, EN 61000-3-2,

EN 61000-4-2, EN 61000-4-3,

EN 61000-4-4, EN 61000-4-5,

EN 61000-4-6, EN 61000-4-11

Fulfils Class II for the following applications:

EN 60950 / IEC 60950, UL 60950,

CSA 950 (cUL), VDE, CE label

Requirements for external power supply:

The external power supply intended to connect to the BioProTT™ FlowTrack DINrail

shall deliver minimum 300 mA at 24 V DC (± 10 %),

must deliver always less than 6,25 A at U = 24 VDC (Circuit with energy limitation

according EN/UL/IEC61010-1, section 9.3.),

shall be short circuit protected,

shall provide a separate switch.

12.2 Technical Data BioProTT™ Clamp-On Transducer

BioProTT™ Clamp-On Transducer (up to BCT 1/2 x 1/8” A):

Resolution 1 ml/min

Accuracy* ± 3 % of reading ± max. 20 ml/min

Measurement range Dependent on sensor size and calibration (see

calibration certificate)

Dimensions (HxWxD)

25 mm x 33 mm x 45 mm (ID < 1/2”),

Total weight (incl. cable/plug)

Weight sensor head

52 g (ID < 1/2”), 72 g (ID = 1/2”)

Housing / lid material Epoxy resin, aluminum, stainless steel

Cable length 2.9 m

IP code IP65

Fluid types Water, Buffer Solutions, Cell Culture Media,

Fermentation Media, Non-Aerated Liquids

Tubing type e.g. Silicon, PVC and Polyurethanes

Medium operating temperature 4 to 50 ºC (40 to 122°F)

BioProTT™ Flow Measurement System Version 2.0 page 82 of 94

Supported flow meters BioProTT™ FlowTrack, BioProTT™ FlowTrack

plus, BioProTT™ FlowTrack DINrail

Repeatability

Dependent on tubing size, wall thickness and

material and manufacturing tolerances and sensor

positioning

* Reference conditions

Liquid temperature

Ambient temperature

Straight inlet section

Tubing

Liquid pressure

Calibration temperature ±5 °C (± 9° F)

20° C ±5° C (68° ± 9° F)

10 x ID of tubing, fully developed flow profile

Raumedic-ECC-Blood Line PVC

Open end at tubing outlet

Please note: as the measurement allows a maximal flexibility with respect to

positioning the sensor the given accuracy is a standard under the reference conditions.

Certain setups allow higher levels of accuracy and can be evaluated by a factory

calibration certificate on customer request.

Ambient conditions:

Air pressure 70 kPa to 106 kPa

Operating altitude up to 2000 m (6600 feet)

Operating temperature range 4º C to 50º C (40° F to 122° F)

Storage temperature range –20° C to 45° C (-4°F to 113° F)

Transport temperature range –20° C to 55° C (-4°F to 131° F)

Relative air humidity:

Operation 30 % to 75 % (non-condensing)

Storage and transportation 10 % to 96 % (non-condensing)

Customer-specific calibrations for different flow mediums, tubing materials e.g. Silicone,

PVC, Masterflex® tubing and medium temperatures can be made upon request.

Each sensor can store up to 7 different calibration tables in the sensor plug.

BioProTT™ Clamp-On Transducer Sizes* and Flow Measurement Range

Type Maximum

Flow Range

Qmax**

Tube size in inch Tube size in mm

[l/min] ID x WT [“] OD [“] ID x WT [mm] OD [mm]

BCT 1/8 x 1/16” A up to ± 4 1/8 x 1/16 1/4 3.18 x 1.59 6.35

BCT 11/64 x 3/64” A up to ± 6 11/64 x 3/64 17/64 4.30 x 1.25 6.80

BCT 3/16 x 1/16” A up to ± 6 3/16 x 1/16 5/16 4.76 x 1.59 7.94

BCT 1/4 x 1/16” A up to ± 8 1/4 x 1/16 3/8 6.35 x 1.59 9.53

BCT 1/4 x 3/32” A up to ± 8 1/4 x 3/32 7/16 6.35 x 2.38 11.11

BCT 3/8 x 1/16” A up to ± 10 3/8 x 1/16 1/2 9.53 x 1.59 12.70

BCT 3/8 x 3/32”A up to ± 10 3/8 x 3/32 9/16 9.53 x 2.38 14.29

BCT 3/8 x 1/8”A up to ± 10 3/8 x 1/8 5/8 9.53 x 3.18 15.88

BioProTT™ Flow Measurement System Version 2.0 page 83 of 94

Type Maximum

Flow Range

Qmax**

Tube size in inch Tube size in mm

[l/min] ID x WT [“] OD [“] ID x WT [mm] OD [mm]

BCT 1/2 x 3/32” A up to ± 20 1/2 x 3/32 11/16 12.70 x 2.38 17.46

BCT 1/2 x 1/8” A up to ± 20 1/2 x 1/8 3/4 12.70 x 3.18 19.05

ID = internal diameter, OD = outer diameter, WT = wall thickness

*custom sizes available ** depending on medium and tube size Table 32: BioProTT™ Clamp-On Transducer overview

BioProTT™ Flow Measurement System Version 2.0 page 84 of 94

13 Environmental Protection and Disposal

The BioProTT™ flow measurement system is made of conventional materials (plastic,

metal and electronic parts) that are harmless to humans and the environment providing it is

used correctly according to current knowledge.

The device and its accessories must be disposed of in accordance with the applicable

national provisions for electronic devices.

In accordance with the requirements of EU Directive2012/65/EC Waste Electrical and

Electronic Equipment (WEEE), our customers in the EU are entitled to return to us all

waste devices -- in clean and disinfected condition. The em-tec GmbH WEEE Registration

No is "DE 99135207".

Upon receipt, we repair or dispose these devices properly. For our delivery address see

chapter RETURNS / RMA.

BioProTT™ Flow Measurement System Version 2.0 page 85 of 94

14 Manufacturer's Declarations

14.1 Information on Acoustic Output Data

Ultrasound signals are emitted at a low output power by the BioProTT™ Clamp-On

Transducer. The ultrasound signals generated are harmless to persons, protein solutions

(e.g. blood, blood cells) and the environment according to current knowledge, providing

the BioProTT™ flow meter and the sensors are used properly.

The Thermal Index (TI) and Mechanical Index (MI) is < 1.0 for all settings of the

BioProTT™ flow sensors.

The acoustic measurements have been carried out according to:

- IEC/EN 60601-2-37:2007/2008 Medical Electrical Equipment - Part 2-37:

Particular requirements for the basic safety and essential performance of ultrasonic

medical diagnostic and monitoring equipment

- FDA Guidance 560:2008 Guidance for Industry and FDA Staff - Information for

Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and

Transducers

Acoustic Output Data of Clamp-On Transducer

Manufacturer: em-tec GmbH

Sensor type: Clamp-On Transducer CT 1/4" x 1/16" (2.25 MHz) SN: 2010-49039

Operating mode: PW

Parameter Measured

value

Limit value for 201.12.4.2 Note 1 and 2

Power [mW] 0.3

p- [MPa] 0.11 1

Ispta [mW/cm²] 2.2 100

Iob [mW/cm²] 1.5 20

Table 33: Acoustic output values (table according to IEC/EN 60601-2-37:2007/2008)

Acoustic output values MI ISPTA.3

(mW/cm²)

IPA.3@MI

(W/cm²)

Global maximum values 0.07 0.02 0.25

Associated

acoustic

parameters

Pr.3 (MPa) 0.11

W0 (mW) 0.3 0.3

fc (MHz) 2.25 2.25 2.25

zsp (cm) 0.44 0.44 0.44

Beam

dimensions

X-6 (cm) 0.49 0.49

Y-6 (cm) 0.33 0.33

PD (μsec) 7.98 7.98

PRF (Hz) 1000 1000

EBD Az. (cm) 0.49

Ele. (cm) 0.41

Operating

conditions

Default Default Default

Table 34: Acoustic output values (table for FDA Guideline 560.2008 Track 1 Non-Auto-Scanning

Mode)

BioProTT™ Flow Measurement System Version 2.0 page 86 of 94

Transducer:

Clamp-On Transducer

Test: stationary air Max •T: 27°C

Mode System settings

Type Freq.

[MHz]

Depth

[mm]

B-

Width

[%]

Focus setting Acoustic

output

(power)

Line

Dens T

[° C]

PW 2.25 - - - - 100% < 0.3°

Table 35: Increase in temperature according to IEC/EN 60601-2-37:2007/2008

BioProTT™ Flow Measurement System Version 2.0 page 87 of 94

14.2 EU Declaration of Conformity

BioProTT™ Flow Measurement System Version 2.0 page 88 of 94

14.3 FCC Certificate of Compliance

BioProTT™ Flow Measurement System Version 2.0 page 89 of 94

15 Ordering Information for the BioProTT™ Flow Measurement

System and Accessories

Order Number Description

BioProTT™ Flow Meters

12214 BioProTT™ FlowTrack

Flow meter without display

12271 BioProTT™ FlowTrack plus

Flow meter with display as bench top variant

12565 BioProTT™ FlowTrack plus

Flow meter with display as panel mount variant

12390 BioProTT™ FlowTrack DINrail

Flow meter without display for DIN rail mounting

BioProTT™ Clamp-On Transducers,

Size: ID x WT in inch; one calibration table included, e.g. 20°C, water, PVC; up to 7

calibration tables possible

12386 BioProTT™ Clamp-On Transducer, 1/8” x 1/16”

12345 BioProTT™ Clamp-On Transducer, 11/64” x 3/64”

12344 BioProTT™ Clamp-On Transducer, 3/16” x 1/16”

12341 BioProTT™ Clamp-On Transducer, 1/4" x 1/16”

12342 BioProTT™ Clamp-On Transducer, 1/4" x 3/32”

12346 BioProTT™ Clamp-On Transducer, 3/8“ x 1/16”

12367 BioProTT™ Clamp-On Transducer, 3/8" x 3/32"

12607 BioProTT™ Clamp-On Transducer, 3/8" x 1/8"

12343 BioProTT™ Clamp-On Transducer, 1/2“ x 3/32”

12388 BioProTT™ Clamp-On Transducer, 1/2“ x 1/8”

BioProTT™ Flow Measurement System Version 2.0 page 90 of 94

Order Number Description

BioProTT™ Accessories

On request Factory calibration certificate

On request Ultrasonic Coupling Paste

On request RS 232 to USB adapter

12466 Extension cable 1 (IP50 rating at panel) – standard for

BioProTT™ FlowTrack DINrail.

The cable is compatible to all BioProTT™ flow meters.

12560 Extension cable 2 (IP68 [2 m, 24 h] rating at panel) – standard for

BioProTT™ FlowTrack plus (for panel mount).

The cable is compatible to all BioProTT™ flow meters.

On request Power supply for BioProTT™ flow meter with country specific

adaptors

On request Custom Calibration (additional table with “standard” parameters)

On request Customization of Clamp-On Transducer to fit custom tubing size

Table 36: Overview Order Numbers

BioProTT™ Flow Measurement System Version 2.0 page 91 of 94

16 Return Information and Declaration of Contamination

Because of legal regulations and for the safety of our employees and operating equipment, we need the

"Declaration of Contamination", with your signature, before your order can be handled. Please make

absolutely sure to include it with the shipping documents, or - even better - attach it to the outside of the

packaging.

Type of device, e.g. flow meter, sensor _________________ Serial number _______________

Relevant process data

Temperature / _________ [°C] Pressure / __________ [ Pa ]

Conductivity / _________ [ S ] Viscosity / __________ [mm /s]

Medium and warnings

Medium /

concentration

Identi-

fication /

CAS No

Flammable Toxic Corrosive Harmful/

Irritant

Other* Harmless

Process

Medium

Medium for

process cleaning

Returned

part cleaned with

* explosive; oxidizing; dangerous for the environment; biological risk; radioactive

Please tick should one of the above be applicable, include security sheet and, if necessary, special handling

instructions.

Reason for return:

Company contact data:

Company:

Contact person/department:

Address:

Fax / E-Mail

Reference No:

We hereby certify that the returned parts have been carefully cleaned. To the best of our knowledge they are

free from any residues in dangerous quantities.

_______________________________ __________________________________

(Place, Date) (Company stamp and signature)

BioProTT™ Flow Measurement System Version 2.0 page 92 of 94

17 Appendix

17.1 Calibration Information for BioProTT™ Clamp-On Transducers

Each sensor is shipped with a calibration report containing all relevant calibration data.

If a more comprehensive and enhanced factory calibration certificate is required, please

contact our service department. It offers a special calibration services and for this purpose

and – if required – can also provide a replacement device. Please contact our service

department in advance to clarify the necessary arrangements.

17.2 Changes due to Software Version V3.0

For information for user with devices with former software version the following change

list is given:

Issue/Change Change has impact

on

Software Version

up to V 2.4.

Software Version

V3.0

Extended Flow

range

Accepted sensor

types

Standard sensors up

to 20 l/min

Sensors up to

999 l/min

System behavior

without sensor

BioProTT™

FlowTrack Plus

All screens can be

accessed

Access restriction to

useful screens only

Zero flow limitation Offset correction at

device

None Prohibited when

insufficient

coupling, no sensor

Remote Zero DINrail variant

(Terminal block

Pin10)

No Remote Zero Support of the

Remote Zero feature

Selection of sensor

table

Analog interface No indication on

analog interface

Analog indication by

0% RSS and

disabling flow for

approx. 0.2 seconds

Selection of sensor

table

Display device:

Totalizer

No indication of

change of sensor

table in totalizer

Exclamation mark is

indicated when a

sensor table is (re-)

selected

Extended flow range Serial output string Flow value can

contain up to 6

characters including

prefix

Flow value can

contain up to 7

characters including

prefix

Response String to

„S“

Serial output string Q-Max contains up

to 6 characters

including prefix

Q-Max contains up

to 8 characters

including prefix

Averaging flow Flow value on

analog and serial

interface

No restriction

conditions

Values at

insufficient flow

conditions are not

used for flow

calculation

BioProTT™ Flow Measurement System Version 2.0 page 93 of 94

Issue/Change Change has impact

on

Software Version

up to V 2.4.

Software Version

V3.0

RSS threshold for

insufficient coupling

Flow output due to

insufficient coupling

Hysteresis behavior

of RSS in range 50 to

60%

Fix threshold for

RSS of 50%, no

hysteresis

Additional

indication of invalid

flow values

Indication on digital

output

No indication Overflow indication

in serial string:

“-------“

Overflow condition: Indication on

display and digital

output

No indication Overflow indication

in serial string:

“^^^^^^^”

Underflow

condition:

Indication on display

and digital output

No indication Underflow

indication:

“vvvvvvv“ Table 37: Overview Changes from V2.4 to V3.0

17.3 Output Strings of V2.4 versus V3.0

Due to extend of flow range the string can handle flows of 6 characters with prefix instead

of 5 characters plus prefix.

Version Output String Examples V2.4 (and

lower)

Positive flow: “00 00 100 1.00 -16653 -16655 -16694 +30” <CR><LF>

Negative flow: “00 00 100 1.00 16720 16712 16732 +31” <CR><LF>

V3.0 Negative flow: “00 00 100 1.00 -16109 -16121 -16122 +35” <CR><LF>

Positive flow: “00 00 100 1.00 16072 16061 16054 +34” <CR><LF>

Large negative flow: “00 00 100 1.00 -100107 -100111 -100122 +35” <CR><LF>

Large positive flow: “00 00 100 1.00 100205 100208 100205 +34” <CR><LF>

Due to the extended flow value, the output string starts at the same position but is two

characters longer.

Version Output String Examples (Response on “S” command) V2.4 (and

lower)

“83599 3/8" x 3/32" PVC Water

23 °C 2 +10000” <CR><LF>

“ 59914 V2.4.0.1 ” <CR><LF>

V3.0 “83599 3/8" x 3/32" PVC Water

23 °C 2 10000” <CR><LF>

“ 59916 V3.0.0.0 ” <CR><LF>

BioProTT™ Flow Measurement System Version 2.0 page 94 of 94

Manufacturer:

em-tec GmbH

Lerchenberg 20

86923 Finning

Germany

Tel.: +49 (0) 8806 9236-0

Fax: +49 (0) 8806 9236-50

Internet: www.em-tec.de

E-Mail: [email protected]

© em-tec GmbH

All rights reserved, especially with respect to reproduction, distribution and translation. These

Operating Instructions or any part thereof must not be reproduced, saved, processed, duplicated or

distributed without the written consent of em-tec GmbH.

IFU BioProTT-en Version 2.0 Date: 2015-03-18