optimascale.com OP-900 SERIES INDICATOR USER’S MANUAL (OP-900A, OP-900B Series) 9.9.19

OP-900 SERIES INDICATOR USER’S MANUAL

  • Upload
    others

  • View
    9

  • Download
    0

Embed Size (px)

Citation preview

Page 1: OP-900 SERIES INDICATOR USER’S MANUAL

optimascale.com

OP-900 SERIES INDICATORUSER’S MANUAL(OP-900A, OP-900B Series) 9.9.19

Page 2: OP-900 SERIES INDICATOR USER’S MANUAL

TABLE OF CONTENTS

Safety Precautions 1

Preparations and Set Up 1

Features 2

Specifications 3

Power Supply 4

Displays 5

Display and Key Descriptions 6

Operating Instructions 7-9

Calibration 10

Indicator Parameter Settings 12-16

Helpful Definitions 17

Connectors 18-19

RS232 Serial Output Format 20-23

4-20 mA Analog Output 24

Relay Output 25

Troubleshooting 26

Q&A 27

Contact Us 28

Page 3: OP-900 SERIES INDICATOR USER’S MANUAL

1

SAFETY PRECAUTIONS For safe operation of the weighing indicator, please follow these instructions: ● Calibration inspection and maintenance of the indicator are prohibited by non-pro-

fessional staff ● Please ensure that the indicator rests on a stable surface ● The indicator is a piece of static sensitive equipment; Please cut off power during

electrical connections ● Touching the internal components by hand is prohibited ● DO NOT exceed the rated load limit of the unit ● DO NOT step on the unit ● DO NOT jump on the scale ● DO NOT use this product if any of the components are cracked ● DO NOT use for purposes other then weight taking ● To avoid damaging the battery do not keep charger plugged in once battery is fully

charged ● Make sure the weight is not over the Max capacity as it could damage the load cell

inside ● Material that has a static electric charge could influence the weighing. Discharge

the static electricity of the samples, if possible. Another solution to the problem is to wipe both sides of the pan and the top of the case with an anti-static agent

Please take anti-static prevention measuresAny accumulated charge on the body of the human operator should be discharged first before opening the protective container with ESDS devices inside. The discharge can be accomplished by: ● Putting a hand on a grounded surface or, ideally, by wearing a grounded Anti-static

Wrist Strap and an Anti-static Mat

PREPARATION & SET UP ● Plug into a wall outlet to avoid interference with other wirings ● Turn on the indicator while there is no load ● Calibration may be required before weighing when the scale is initially installed or

moved from a location

Page 4: OP-900 SERIES INDICATOR USER’S MANUAL

2

FEATURES ● LED or LCD display option ● Multiple weighing units: kg/lb (g/oz/lb:oz offered on LCD versions) ● Gross/Tare/Pre-Set Tare/Zero ● Multiple Hold functions ● Count weighing ● Accumulation weighing ● Overload / Underload indication ● Connects to multiple printers ● Splash proof keyboard and display ● Connects to a Remote Display/Scoreboard ● Power saving mode ● Ability to modify gravity based on different geographical locations ● RS232 output ● NTEP approved for 5,000 divisions ● Relay output (optional) and 4-20mA analog output (optional) ● Can connect to a PC or printer for data logging (optional) ● Wireless capability (optional) ● Rechargeable battery (optional)

Indicator Model OptionsThe OP-900 series consists of the OP-900A (LED) series and OP-900B (LCD) series. Options for rechargeable battery, stainless steel enclosure, washdown enclosure, computer connection, relay output and analog output can be added.Please contact [email protected] for a list of possible options.

Technical Parameters

● Accuracy class: 5000 e ● Resolution - Display: 30,000 ; ADC: 2,000,000 ● Zero stability error: TK0 < 0.1μV//K ● Span stability error: TKspn < ± 6 ppm//K ● Sensitivity (internal): 0.3 μV / d ● Input voltage: -30 to +30mV DC ● Excitation circuit: 5 VDC, 4 wire connection, 6 load cell of 350ohm max ● AC power: AC 100-250V (use only the included 9V adapter supplied) ● Operation temperature: -10 °C ~ +40 °C ● Operation humidity: ≤90%RH ● Storage temperature: -40 °C ~ +70 °C (32-104°F)

Page 5: OP-900 SERIES INDICATOR USER’S MANUAL

3

SPECIFICATIONSFIGURE 1: INDICATOR MEASUREMENTS

Page 6: OP-900 SERIES INDICATOR USER’S MANUAL

4

POWER SUPPLY

AC Adapter

If the indicator is powered by an adapter, plug the adapter directly into the “DC” pin located at the bottom of the indicator. We recommend to plug into a wall outlet to avoid interference with other wirings.

Battery (Optional)

If you have an indicator with the rechargeable battery option, please charge the internal battery fully before first time use. A 110 to 220V AC adapter should be provided with your indicator. Please use only the AC adapter provided to prevent dam-age to your indicator. To keep the battery in best condition, fully discharge the battery every month by leaving the indicator on until the indicator powers off, and then recharge fully. If the battery is not going to be used for a long period of time it is recommended to remove it to avoid leakage.

On OP-900A ● When the Battery is low the battery indicator light flashes red ● During charging the red light will stay lit ● The light will turn green once fully charged

ON OP-900B ● symbol will indicate battery’s charge ● symbol indicates that the battery needs to be charged

Page 7: OP-900 SERIES INDICATOR USER’S MANUAL

5

OP-900A (LED)

OP 900B (LCD)

Page 8: OP-900 SERIES INDICATOR USER’S MANUAL

6

DISPLAY AND KEY DESCRIPTION ON/OFF Powers the Indicator On or Off if held for 2 secondsHOLD Holds the weight (5 Hold options; can be changed in parameter settings)TOTAL 1. Accumulates weights

2. Works with “Print” to perform the accumulation function and check the accumulation result

UNITS Shifts between weighing unitsCOUNT Use the scale to count product based on a sample weight TARE 1. Resets the scale to zero when there is something on the scale

(ex. Tare out the weight of a pallet to weigh only the product on it)2. Clears the tare to see the gross weight (pallet + product)

ZERO Zero’s the scalePRINT Print data

The scale is at zero

The scale is stable

Gross Shows you are in Gross weight mode (includes tare); default modeNet Shows you are in Net weight mode (weight without tared weight)pcs Shows you are in Counting modetotal Shows you are in Accumulation modehold Shows you are in Hold modelb The weight is shown in pounds kg The weight is shown in kilogramsbattery Flashes red = low battery, Solid red = charging, Green = fully chargedOver Flashes when weight is higher than set alarm parameter Accept Flashes when weight is within the set alarm parametersUnder Flashes when weight is lower than set alarm parameter

Power

BackSave and Exit

Arrow keys

Return/Enter

Page 9: OP-900 SERIES INDICATOR USER’S MANUAL

7

OPERATING INSTRUCTIONSPower On

● Turn on the power by pressing the power button for 2 seconds. Once on, the scale will flash the voltage and then begin to auto-check and count down from 0-9 sequentially before entering the weighing mode Note: Anything on the scale before powering on will automatically be tared out.

Zeroing ● The zero function is used only when the scale is empty and is not at gross zero due

to material build up ● Pressing the ZERO key will reset your scale to 0 ● Depending on what your manual zero range parameter is set to, you can zero out

any number within your set selection, after that you will receive an error and will need to tare out the weight

Unit Selection ● To switch between measuring units ((kg/g/lb/oz/lb:oz) press the UNITS key

Tare Function ● The Tare function is used when you only wish to see the current change in weight,

not the entire amount of weight that is on the scale ● When the indicator is in gross mode (gross light is shown) pressing the TARE key

will Tare the current weight on the scale and enter the net mode (net light shown) ● For example if you are using a container add the container to the scale, press tare

and the display will show the tare symbol and reset back to 0 ● Add your product to the scale to weigh without the weight of the container ● To exit Tare mode press the TARE key again to enter gross mode and you will see

the total weight of the container and the productNote: If you remove the container the scale will show the minus weight of the container

To use a pre-set tare weight ● Press and hold the TARE key for 2 seconds ● Input the tare weight using the arrow keys ● Press print key to confirm

Page 10: OP-900 SERIES INDICATOR USER’S MANUAL

8

Counting Function ● The counting function is used to count a high volume of identical parts. You can do

this by setting a sample and then either adding to the sample or taking away from the sample to count the number of objects on the scale

● In weighing mode: Put a sample weight on the scale (Sample options are 5, 10, 20, 50, 100, 200, 500)

● Then press the COUNT key to go to the counting mode (“pcs” will light up) ● It will then display the sample number (ex. PCS 0) which you can change using

the up or down arrows ● Press the PRINT key to confirm your sample number ● The scale is now ready to start counting, load your product on the scale and the

indicator will show the quantity ● To exit counting mode press the COUNT key ● If you want to count a different product hold the PRINT and COUNT key together

and the sample pieces will reset back to zero

Accumulation ● The accumulation function is used to add multiple weights and total them together ● In weighing mode load the first weight, once stable press the ACCUM key to enter

the accumulation mode. The “total” indicator/light will display ● The screen will show “n001” to indicate the first weight has been saved ● Remove the first weight and press the ZERO key to stabilize the scale ● Add the second weight to the scale ● Once stable press the ACCUM key to add the weight to the accumulated total ● The screen will show “n002” to indicate the second weight has been saved ● Repeat previous steps until all desired weights have been added to the total

(you can accumulate up to 999 different weights) ● When you are done and want to display the accumulated total, press the ACCUM

and PRINT key together. The accumulated number “n002” (the number of weights you are adding together) will flash on the display followed by the total

● The total will display by flashing between 2 sets of numbers ● There are 8 digits in total, the display will flash 4 at a time, the first 4 on the left and

the last 4 on the right. For example if the first 4 digits are “0012” and the last 4 digits are “3456” the actual weight is 001234.56 or 1234.56 lbs/kg

● If you want to print the accumulated total, hold the PRINT key for one second while the last 4 digits of the total are shown

● To exit accumulation mode, wait for the last 4 digits to the right of the screen to appear, and then press and hold the ACCUM key for one second

● “CLr n” will be displayed, asking you if you want to keep the data? ● If NO you do not want to clear the accumulated total, then keep “CLr n”. ● If YES you do want to clear the total, then use the arrow key to change to “CLr y”

● Finally, press the PRINT key to select exit accumulation mode

Page 11: OP-900 SERIES INDICATOR USER’S MANUAL

9

HoldThere are 4 different hold functions you can choose from in the C11 parameter1. Peak Hold: Grabs the highest weight (for materials testing, ie. tension and pulling force)

● Press the HOLD key then add weight to the scale ● The indicator will show the highest weight it recorded and hold it on the screen until

a higher weight is placed on the scale2. Manual Hold: Grabs the current weight and holds it so it will not change/fluctuate

● While weighing, press HOLD and the indicator will hold the current weight on the screen until HOLD is pressed again

3. Auto Hold: If the weight on the scale is above 20d (20 x division) and is stable, the indicator will hold that weight on the screen for 3 seconds then go back to general weighing

● Pressing the hold key is unnecessary, holding is done automatically when the scale is stable

4. Average Hold: Used for animal weighing, the indicator will display the average weight sampled from 3 seconds

● Add livestock to scale and press HOLD ● Indicator screen will show “LOC” for 3 seconds, then display the average weight

from those 3 seconds ● Press HOLD again to exit holding mode

Print ● If the indicator is connected to a printer and the weight on the scale is stable press the

PRINT key to print the current weight ● In accumulation mode hold the PRINT key for one second when the last 4 digits of the

total weight are shown to print the total weightNote: In tare mode the printer can not print if negative weight is shown

Page 12: OP-900 SERIES INDICATOR USER’S MANUAL

10

CALIBRATION PROCEDURE1. Turn on the scale by holding ON/OFF for 2 seconds. 2. Press HOLD and PRINT together to access the setup menu.3. If done correctly, the display should now show C01.4. Press PRINT to access the C1 channel. The display should show [C1 #].5. Press ZERO to choose which unit you want to calibrate in (1 = kg, 2 = lb).6. Press PRINT to set the value. The display will now show C02. 7. Press PRINT to access the C2 channel. The display should show [C2 #].8. Press ZERO to change the setting to the decimal places desired (The C2 channel is

used to adjust the decimal point on the scale. A value of 1 means there is one digit behind the decimal point.)

9. Press PRINT to set the value. The display will now show C03.10. Press PRINT to access the C3 channel. The display should show [C3 #].11. Press ZERO to cycle through the values until the desired graduation appears.

(The C3 channel adjusts the divisions on the scale. A value of 1 selected and C2 set to 1, the scale will read in 0.1 lb. increments.)

12. Press PRINT to set the value. The display will now show C04.13. Press PRINT to access the C4 channel. The display will show [######].14. Enter in the maximum capacity you want to use for this scale by using UNIT and

COUNT to move the cursor left and right, and TARE and ZERO to move the values down and up. (The C4 channel is used to enter in the max capacity of the scale; Make sure this doesn’t exceed the max capacity of the scale; Max capacity divided by the increment set in C02 and C03 above cannot exceed 5000.)

15. Press PRINT to set the value. The display will now show C05.16. Press PRINT to access the C5 channel. The display should show [C5 0].17. The C5 channel calibrates zero on the scale. Make sure the scale is empty.18. Press ZERO to change the value to 1.19. Press PRINT . The display will count down from 10-1 while the scale is calibrating zero. When the display shows 0 the zero calibration is complete. OP-900 offers 2 calibration methods, Single Point which uses one weight to calibrate or Linear Calibration, which uses multiple (2-7) weights for a more accurate calibration. To Calibrate using only 1 calibration weight (Single Point Calibration)

20. Press PRINT to continue. The display will now show C06.21. Press PRINT to access the C06 channel. The display will show [C6 0].22. The C6 channel is used to calibrate the scale with a known weight. Press ZERO to

set the value of C6 to [C6 1]. Press PRINT . The display will flash SPAN, and then show [######].

Page 13: OP-900 SERIES INDICATOR USER’S MANUAL

11

23. Enter the calibration weight value you will use (at least 10% of max capacity you set in C04 by using UNIT and COUNT to move the cursor left and right, and TARE and ZERO move the values down and up.

24. Place the calibration weight you have on the empty scale and press PRINT .25. The scale will count down from 10 to 0. Once 0 has been reached, the display will

show CALEnd.26. Press PRINT to continue. The display will now show C07.27. Press ACCUM to save and exit the setup menu.28. The scale has now been calibrated. The display will show the value of the calibration

weight on the scale.29. If the scale does not show the value of the calibration weight, check that the feet on

the platform are not screwed in too tightly, and verify that the platform is level.30. Unload the scale; the display should read 000000.31. If the scale does not display 000000, check that the feet on the platform are not

screwed in too tightly, and verify that the platform is level.

To Calibrate using only multiple calibration weights (Linear Calibration)

1. Press PRINT to continue. The display will now show C06.2. Press PRINT to access the C06 channel. The display will show [C6 0].3. The C6 channel is used to calibrate the scale with a known weight. Press ZERO

to set the value of C6 to [C6 2]. Press PRINT . The display will flash SPAN, and then show [LnE 2].

4. Press ZERO to enter the number of weights you want to use (2-7); the more you use, the more accurate the calibration will be. (an example of 2 will be used)

5. Press PRINT to set your value. The screen will flash [dbno01] then [000100]6. Enter the lowest calibration weight value you will use (at least 10% of max

capacity you set in C04) by using UNIT and COUNT to move the cursor left and right, and TARE and ZERO move the values down and up.

7. Place the calibration weight you have on the empty scale and press PRINT .8. The scale will count down from 10 to 0. Once 0 has been reached, the display will

show he screen will flash [dbno02] then [000200]9. Enter the next calibration weight value you will use by using UNIT and COUNT

to move the cursor left and right, and TARE and ZERO move the values down and up.

10. Place the calibration weight you have on the empty scale and press PRINT .11. The scale will count down from 10 to 0. Once 0 has been reached, the display will

show he screen will show CALEnd.12. Then follow steps 26-31 on the single point calibration instructions

Page 14: OP-900 SERIES INDICATOR USER’S MANUAL

12

INDICATOR PARAMETER SETTINGSThe parameter settings menu has a calibration section (C01 to C07 explained above) and a parameter settings section (C08 and up). To access the calibration section the seal switch (located at one corner of the PCB) must beOFF. This will allow access to all C01 and up settings. If the seal switch is ON, then only C08 and upcan be accessed by the user. If you break the official seal by opening the back of the indicator to access the seal switch, you may need to have the indicator recertified. Be sure to adjust the seal switch back to the original setting after calibration/configuration has been performed.

To enter calibration/parameter settings, follow the procedure below:1. Make sure the unit is set to either kg or lb2. Press and hold the HOLD and PRINT key at the same time for 2 seconds3. Navigate through the settings (C01 to C45) as shown in the table 4 below by using the arrow

keys and return keys as labeled under each indicator button4. Press the PRINT key to enter/edit the parameter settingPress the ACCUM key to save and exit settings at any time

Table 1. Indicator Parameter Settings

Function Parameter Settings/Options

Weighing Unit C01

1 = kg2 = lb3 = gram4 = ozNote: for calibration only kg or lb are allowed

Decimal Setting C02

0 = no decimal1 = 0.02 = 0.003 = 0.0004 = 0.0000

Graduation Setting(readability of the least significant digit)

C03

options: 1/2/4/10/20/50 Example with no decimal places (ie. C02=0)1 = 1 lb2 = 2 lb5 = 5 lb10 = 10 lb20 = 20 lb50 = 50 lb

Maximum Capacity C04 set max capacity ex. 100kg = 0100.00

Zero Calibration C050 = zero calibration not needed1 = set the zero calibration (Please ensure scale is empty and the stable light is on)

Calibration C06

0 = calibration not needed1 = Ready to calibrate with one calibration weight2 = Ready to calibrate using multiple calibration weights (Linear)3 = Sensitivity Output

Page 15: OP-900 SERIES INDICATOR USER’S MANUAL

13

Function Parameter Settings/Options

Restore DefaultSettings C07

0 = do not restore1 = restore to default settings

Warning Tone C080 = turn off warning tone1 = turn on warning tone

Automatic Power Off C09

0 = turn off auto power off10 = power off automatically if no change within 10 minutes30 = power off automatically if no change within 30 minutes60 = power off automatically if no change within 60 minutes

Power Saving Mode C10

LED Version OP900A:0 = turn off power saving setting3 = turn off display if no change within 3 minutes5 = turn off display if no change within 5 minutesLCD Version OP900B:0 = turn off the backlight 1 = backlight only when the weight changes or keyboard is pressed2 = constant backlight

Hold Function C11

0 = turn off hold function1 = Peak hold - Grabs the highest weight2 = Manual hold - Grabs the current weight3 = Auto hold - Automatically holds data when stable4 = Average hold - for animal weighing, averages the weight from a sample of 3 seconds5 = Auto Average hold - Average hold without the need to press the hold key

Unit Conversion C12 * See table 2

Upper Limit Alarm C13 Set upper limit within the max. capacity

Lower Limit Alarm C14 Set lower limit within the max. capacity

Inner Code Display C15 check the inner code (raw data)

Set Date C16 Set date from left to right: year/month/day

Set Time C17 Set the time from left to right: hour/minute/second

Communication Setting C18

Set the serial interface data output method:0 = Turn off serial interface data output1 = Continuous sending mode, for remote display2 = Print to paper thermal ticket printer3 = Command request mode, for computer.4 = PC continuous sending mode, for computer5 = PC/remote display, continuous sending mode6 = Print to adhesive label thermal printer7 = Print to Zebra/large adhesive label thermal printer8 = Reserved

Baud Rate C19

0 = 1200 (for OP-910 remote display)1 = 24002 = 48003 = 9600 (for all printers and OP-910X, OP-910XL)4 = 14400

Page 16: OP-900 SERIES INDICATOR USER’S MANUAL

14

Function Parameter Settings/Options

ManualZero Range C20

0 = turn off manually zero setting1 = ±1% max capacity2 = ±2% max capacity4 = ±4% max capacity10 = ±10% max capacity20 = ±20% max capacity100 = ±100% max capacity

Initial Zero Range C21

0 = no initial zero setting1 = ±1% max capacity2 = ±2% max capacity5 = ±5% max capacity10 = ±10% max capacity20 = ±20% max capacity100 = ±100% max capacity

Zero Tracking C22

0= turn off zero tracking0.5 = ±0.5d d = division1.0 = ±1.0d2.0 = ±2.0d3.0 = ±3.0d4.0 = ±4.0d5.0 = ±5.0dNote: the zero tracking range can not be bigger than manual zero range

Zero Tracking Time C23

0 = turn off zero tracking time1 = 1 second2 = 2 seconds3 = 3 seconds

Overload Range C2400 = turn off overload range01-99d = overload range setting d = division

Negative Display C25

0 = -9d10 = -10% max. capacity20 = -20% max. capacity50 = -50% max. capacity100 = -100% max. capacity

Standstill Time C260 = quick1 = medium2 = slow

Standstill Range C27

1 = 1d d = division2 = 2d 5 = 5d 10 = 10d

Digital Filter (for filtering moving weight such as animals)

C28

0 = turn off dynamic filter1 = 1 digital filter strength2 = 2 digital filter strength3 = 3 digital filter strength4 = 4 digital filter strength5 = 5 digital filter strength6 = 6 digital filter strengthNote: The higher the number, the higher the filter strength

Page 17: OP-900 SERIES INDICATOR USER’S MANUAL

15

Function Parameter Settings/Options

Noise Filter C29

0 = turn off noise filter1 = 1 digital filter strength2 = 2 digital filter strength3 = 3 digital filter strength

Print Time and Date C30

0 = yy.mm.dd1 = mm.dd.yy2 = dd.mm.yy3 = yy.mm.dd

Analog Output Setting C310 = 0 - 5V ouput1 = 4 - 20mA output

Calibrate Current C32 4 - 20mA current

Relay Output Setting C33

0 = turn off relay output1 = turn on relay output function 12 = turn on relay output function 23 = Reserved menu

Gravity of Calibration Location C36 9.7000 - 9.9999

Gravity of Destination C37 9.7000 - 9.9999

Version No. C38

Print Mode C410 = auto mode1 = gross mode2 = tare mode

Print Carriage Return C42 0 - 9 (How much space between print outs)

Space Print C43 0 - 9 (Where the data prints on the paper: 0 = left ; 9 = right)

Date Print C440 = do not print the date1 = print the date

Time Print C450 = do not print the time1 = print the time

Baud Rade for 2nd RS232 C48

0 = 1200 (for remote display)1 = 24002 = 48003 = 9600

Table 2. Unit Conversion Parameter Settings

Parameter Settings Units AvailableC01= 3 & C12= 0 gram onlyC01= 4 & C12= 0 oz onlyC01= 1 & C12= 0 kg onlyC01= 1 & C12= 1 kg/lbC01= 1 & C12= 2 kg/lb/ozC01= 1 & C12= 3 kg/lb/lb:oz/ozC01= 1 & C12= 4 kg onlyC01= 2 & C12= 0 lb onlyC01= 2 & C12= 1 lb/kgC01= 2 & C12= 2 kg/lb/ozC01= 2 & C12= 3 kg/lb/lb:oz/ozC01= 3 & C12= 4 lb only

Page 18: OP-900 SERIES INDICATOR USER’S MANUAL

16

Table 3. Default Parameter Settings

Function Parameter Default Setting

Weighing Unit C01 1

Decimal Setting C02 0

Graduation Setting C03 1

Maximum Capacity C04 1000

Zero Calibration C05 0

Calibration C06 0

Restore Default C07 0

Warning Tone C08 1

Automatic Power Off

C09 0

Power Saving Mode C10 0

Hold Function C11 0

Unit Conversion C12 1

Upper Limit Alarm C13 000000

Lower Limit Alarm C14 000000

Inner Code Display C15

Set Date C16

Set Time C17

Communication Setting C18 0

Baud Rate C19 3 (9600)

Manual Zero Range C20 10

Initial Zero Range C21 10

Zero Tracking C22 0.5

Zero Tracking Time C23 1

Overload Range C24 9Negative Display C25 10

Standstill Time C26 1

Standstill Range C27 2

Digital Filter C28 0

Noise Filter C29 2

Print Time and Date C30 0

Analog Output Setting C31 1

Calibrate Current C32 4

Relay Output Setting C33 1

Multi-connection add. C34 0

Wireless Communica-tion

C35 6

Gravity of Calibration Location

C36 9.7936

Gravity of Destination C37 9.7936

Page 19: OP-900 SERIES INDICATOR USER’S MANUAL

17

HELPFUL DEFINITIONSDivision: The amount of increments a scale offers. How accurate the scale can be

Capacity: the maximum amount the scale can contain

Initial Zero Range: The percentage of weight allowed on the scale when indicator is powered on that will automatically zero. example: If initial zero range is set to 10% of the max. capacity and your max. capacity is 100lbs, you can place up to 10lbs of weight on the scale and when the indicator is powered on, it will automatically zero out the weight.

Manual Zero Range: The percentage of weight allowed on the scale where the indicator will let you manually zero (anything above this percent will be tared)

Zero Tracking Range: A subset to the manual zero range; if the weight on the scale is not stable, the zero tracking range still allows you to zero within a set division of the scale

Zero Tracking Time: A subset to the zero tracking range, it is the time allowed for the scale to fall within the zero tracking range tolerance and still qualify to be zero’d

Overload Range: Weight allowance that is out of the set calibrated range. Adds a tolerance to the calibrated max. capacity without having to recalibrate. example: If your scale has a max. capacity of 1000lbs with a division of 1 and you set the overload range to 60, you can add 1060lbs of weight to the scale without it displaying an error code

Negative Display: How far you can go in the negative direction before displaying an error code

Standstill Time: How fast the scale will stabilize

Standstill Range: How much the scale can fluctuate before being determined stable

Digital Filter: For filtering moving weight, such as animals, It changes how sensitive the scale is to variations in movement.

Noise Filter: A filter for how susceptible the scale is to general variations

Baud Rate: The rate at which information is transferred in a communication channel. example: In the serial port context, “9600 baud” means that the serial port is capable of transferring a maximum of 9600 bits per second.

Page 20: OP-900 SERIES INDICATOR USER’S MANUAL

18

CONNECTORS

Connecting load cells to the indicator ● The indicator can connect with 6 load cells of 350Ω at most ● 4 wire or 6 wire load cell connections are both okay ● Please contact us directly if you have other special needs for your application ● There are two connection methods between the load cell and indicator

Quick Disconnect as shown below:

FIGURE 2: QUICK DISCONNECT CONNECTION DIAGRAM

Hardwire (Using Inner Terminal Block Connection:Note: Make sure you follow all the anti-static rules to avoid damage to your indicator

● Excitation voltage: 5V DC ● Largest output current: 120 mA ● Excitation circuit: 5 VDC, 4 wire connection, 6 load cell of 350ohm maximum ● Open the back cover of the weighing indicator, and insert signal cable to the

terminal block (see figure 3); Make sure the screw on terminal block is fixed tightly

Page 21: OP-900 SERIES INDICATOR USER’S MANUAL

19

FIGURE 3: INNER TERMINAL BLOCK CONNECTION DIAGRAM

Table 4. Wiring Color Code

Signal Name Color Code Description+Exe/ +EX RED Positive excitation voltage to load cell+IN / +SIG GREEN Positive output signal from load cellHD / SHLD YELLOW/THICK BLACK Shield Wire-IN / -SIG WHITE Negative output signal from load cell-EXC / -EX BLACK Negative excitation voltage to load cell

DB9 Connection (9 pin Serial Connector)The DB9 9 pin serial connector is used for different purposes depending on the indicator model

● Figure 4 shows the pin assignment on the DB9 9 pin connector

FIGURE 4: DB9 SERIAL CONNECTOR PINOUT

There are 3 Output formats to choose from

1. RS232 Serial Output Format (Standard)2. 4-20 mA Analog Output (Optional)3. Relay Output (Optional)

Page 22: OP-900 SERIES INDICATOR USER’S MANUAL

20

RS232 SERIAL OUTPUT FORMAT Follow the pin out of Table 5 below to connect the indicator the RS-232 Serial device

Table 5. DB9 Pin DescriptionDB9 Pin Definition Function2 TXT Transmit Data3 RXD Receive Data5 GND Ground Interface

The serial output format depends on the settings for parameter C18. The serial output consists of a string of ASCII characters. Here is a list of the serial parameters

● 8 data bits ● 1 stop bits ● No parity ● No handshaking

Note: With the RS232 Output option we have data logging software available as seen in Figure 5.

FIGURE 5. DATA LOGGING SOFTWARE

Below are the formats of the serial output ● C18=0 - Turn off serial interface data output ● C18=1 - Continuous sending mode, connect 2nd big display ● C18=2 - Print mode, connect printer ● C18=3 - Command request mode, connect computer. ● C18=4 - PC continuous sending mode, connect computer ● C18=5 - PC/big display, continuous sending mode ● C18=6 - Print to adhesive label printer ● C18=7 - Print to Zebra adhesive label printer

Page 23: OP-900 SERIES INDICATOR USER’S MANUAL

21

Remote Display Continuous Sending Mode (C18=1)

For use with a Scoreboard/Remote Display Note: Baud Rate must be set to 1200 (C19 = 0)

Output Continuous FormatSTX

SWA

SWB

SWC

X X X X X X X X X X X XCR

CKS

1 2 3 4 5 6

State ABits0,1,2

0 1 2 Decimal point position1 0 0 XXXXXX00 1 0 XXXXXXX1 1 0 XXXXX.X0 0 1 XXXX.XX1 0 1 XXX.XXX

Bits3,4 Division0 1 X11 0 X2

State BBitsS functionBits0 gross=0, net=1Bits1 Symbol: positive=0, negative=1Bits2 Overload (or under zero)=1Bits3 dynamic=1Bits4 unit: lb=0, kg=1Bits5 Constant 1Bits6 Constant 0

State CBit2 Bit1 Bit0 unit0 0 0 Kg or lb0 0 1 g0 1 0 t

Bit 3 printing=1Bit 4 Extend display=1Bit 5 Constant 1Bit 6 Constant 0

Page 24: OP-900 SERIES INDICATOR USER’S MANUAL

22

Print Mode (C18 = 2)For printing on a non-adhesive ticket printer. Parameters 16, 17, 30, & 42-45 all effect your ticket print out. Normal weighing ticket printout example:

Date: 05/01/2017Time: 11:30:52Net: 25.6lbTare: 10.3lbGross: 35.9lb

Accumulation weighing ticket printout example:

Date: 05/01/2017Time: 11:30:52

n001 15.4lb

n002 17.2lb

n003 35.6lb

Total 68.2lb

Command Request Mode (C18=3)In this mode, the indicator can recieve ASCII commands as listed below

Command Name FunctionT Tare Save and clear tareZ Zero Zero gross weightP Print Print the weightR G.W/N.W Read gross weight or net weightC Kg/lb Kg/lb conversionG G.W Check gross weight at net weight mode

The R command will trigger the indicator to output the followind data format:

<STX> <POL> XXXXX.XX <SP> <LB/KG> <SP> <GR/NT> <CR> <LF>

StartTransmission

Polarity:<SP> = Positive“-” = Negative

Weight Data Space Units:lb = puntkg = kilogrampcs = pieces*

SpaceGross/Net:GR = GrossNT = Net

CarriageReturn

Line Feed

Page 25: OP-900 SERIES INDICATOR USER’S MANUAL

23

Computer Continuous Sending Mode (C18=4)

, , CR LF

{ { {

S 1 S 2 S 3 Data S 4

S1: weight status, ST=standstill, US=not standstill, OL=overloadS2: weight mode, GS=gross mode, NT=net modeS3: weight of positive and negative, “+” or “-”Data: weight value, including decimal pointS4: “kg” or “lb”CR: carriage returnLF: line feed

PC or Remote Display Continuous Sending Mode (C18=5)

<STX> <POL> XXXXX.XX <L/K> <G/N> <STAT> <CR> <LF>

StartTransmission

Polarity:<SP> = Positive“-” = Negative

Weight Data

Units:lb = puntkg = kilogrampcs = pieces*

Gross/Net:GR = GrossNT = Net

CarriageReturn

Line FeedStatus:<SP> = ValidM = MotionO = Over/under range

Print to Adhesive Label Printers (C18=5) & (C18=6)The following are the printing formats:

Date: 05/01/2017Time: 11:30:52Net: 25.6lbTare: 10.3lbGross: 35.9lb

Date: XX.XX. XX (yy.mm.dd)Time: XX.XX.XX (hh.mm.ss)

NET 6.00kg (net weight) TARE 2.88kg (tare) GROSS 8.88kg (gross weight)

Optima Label Printer Zebra Label Printer

Page 26: OP-900 SERIES INDICATOR USER’S MANUAL

24

4-20 mA ANALOG OUTPUT (OPTIONAL)The 4-20 mA analog output of the OP-900 scale is a voltage sourcing sensor that will output current which is proportional to the calibrated scale’s weight range (i.e. 4 mA = 0 LBS and 20 mA = 10,000 LBS). It is important to note, that this is a sourcing output sensor (i.e. the sensor will source 9VDC with an output current range of 4-20 mA). Unlike many other “loop powered” type 4-20 mA sensors, an external supply voltage should not be connected to the unit’s 4-20 mA circuit, however, the ground connection (pin 6) of the DB-9 connector will need to be connected to the same ground as the data acquisition device which will be responsible for interoperating the 4-20 mA signal. This ground connection is imperative, as both the data acquisition device’s power supply and the scale’s internal 9VDC power supply will need to be on the same ground plane for the output current to be synced and measured correctly. Please reference Figure 6 as a visual clarification on how to connect your scales 4-20 mA output to a data acquisition device.

Below is a list of important notes when using the indicator with the 4-20 mA option ● Resolution: 1/1000 ● Outside Load: 100-350ohms ● Inside connection: load input port pin “1” of J2, ground port pin “GND” of J2 ● Outside connection: load input port pin 1 of DB9, ground port pin 6 of DB9 ● To test the connection, connect a 250 ohm load; Locate a volt meter, and probe

across the 250 ohm load. As the weight input to the indicator varies, the voltage of the volt meter will change accordingly

● Pressing the TARE key will reset to output current to 4mA ● 0-20mA output can be set by setting parameter C31 to 0 ● Please note that this option will disable the RS232 weigh data output (serial ASCII

data) that comes in the standard indicatorCalibration: ● Press PRINT and HOLD key to go into configuration mode ● Go to C32 and press PRINT key ● The display should show [out-4] and output should be at 4mA ● Press the up/down arrow keys and the [out-#] will increase/decrease

Note: X corresponds to the output current. For example if out-12 is displayed, then 12mA should be the output. If not, press the left and right arrow key to adjust/calibrate

FIGURE 6. CONNECTION DIAGRAM

Page 27: OP-900 SERIES INDICATOR USER’S MANUAL

25

RELAY OUTPUT (OPTIONAL) ● The indicator can output 4 signals, which when connected to outside equipment,

can perform an automatic control function and an upper/lower limit alarm function. ● Change parameter setting C33 following Table 6 below:

Table 6: Relay Output Parameter Setting

Output Port Port Definition Function

C33=0Out1 Turn off output function No Output SignalOut2 Turn off output function No Output SignalOut3 Turn off output function No Output SignalOut4 Turn off output function No Output Signal

C33=1Out1 Turn on overload control function Output overload control signalOut2 Turn on compliance control function Output compliance control signalOut3 Turn on under-load control function Output under-load control signalOut4 Turn on stable control function Output stable control signal

C33=2,3 Preserved, no function

Table 7 below shows the DB9 port pinout for the relay output option. Please note that this is optional and only available for the OP-900 indicator ordered with this option.

The relay output option will disable the RS232 weigh data output (serial ASCII data) that comes in the standard indicator.

Table 7: Relay Output Pin Definition

DP9 Pin Definition Port1 Red 1st output signal pin Out16 Red 1st output signal pin Out12 Green 2nd output signal pin Out27 Green 2nd output signal pin Out23 White 3rd output signal pin Out38 White 3rd output signal pin Out34 Black 4th output signal pin Out49 Black 4th output signal pin Out4

Rating:AC 250V 3A DC 30V 3A

1627

3849

Page 28: OP-900 SERIES INDICATOR USER’S MANUAL

26

TROUBLESHOOTINGError Codes

Error Reason Solution

UUUUUU

1. Overload2. Wrong connection with load cell3. Load cell has quality problem

1. Reduce the weight2. Check load cell connection3. Inspect load cell; Check the input/output4. See Q&A section

nnnnnnn

1. Calibration is no good2. Wrong connection with load cell3. Load cell has quality problem

1. Make sure scale is level2. Check load cell connection3. Check load cell input and output resistance4. See Q&A section

ERR1During calibration, weight is not used or the weight is above the max. capacity Use correct weight within the defined range

ERR2During calibration, the weight is below the minimum required weight

The calibration weight minimum is 10% of the max. capacity set in C04. Recommended to use 60%-80% of max. capacity if possible

ERR3During calibration, the input signal is negative

1. Check all wire connections2. Check load cell3. Recalibrate4. PCB replacement needed if steps 1-3 fail

ERR4 During calibration signal is unstable After the platform is stable, start calibration

ERR5 EEPROM Error Change PCB

ERR6 Exceed Zero Range See Q&A section

Page 29: OP-900 SERIES INDICATOR USER’S MANUAL

27

Q&A Q: The scale does not turn onA: Make sure the power cord is plugged in, and that there is power. One easy way to

test this is by connecting another appliance to the same outlet and see if it’s operational

Q: The reading goes negative when a load is appliedA: Try interchanging the Sig+ and Sig- wiring connected to the load cell and/or

junction box (if one is used)Q: How do I resolve ERR6 error?A: Please follow the procedure below:

1) Turn on the indicator and make sure nothing is on the scale, and that the scale is level and not wobbling2) Press and hold the “PRINT and HOLD” key simultaneously for a few of seconds3) The screen will read “C01” 4) Using the arrow keys, change C01 to C20. You have to change the 1st digit from 0 to 2 first before you can change the 2nd digit 1 to a 0.5) Press “PRINT” key to enter C20 parameter6) Change the value of C20 on the right to 100 if possible using the up arrow key. If 100 is not available change to 207) Press “PRINT” key to enter your selection8) The screen will read “C21” now9) Press “PRINT” key to enter C21 parameter10) Change the value on the right of C21 to 100 if available, 20 if not11) Press “PRINT” key to enter your selection12) he screen will read “C22” now13) Press “TOTAL” key to save and exit14) Power the indicator off and then on, and see if this resolves the ERR 6 issue. If not, then following the Q&A answers below for resolving “nnnnnn”and “uuuuuu” errors

Q: How do I resolve “nnnnnn” and “uuuuuu” error?A: 1) Check to see if the cable that runs from the indicator to the junction box is

damaged. If it is, replace the cable.2) Open up the junction box (if available) and check to see if there is any water damage. If so, replace the junction box3) Make sure all the wires on all 5 terminal blocks (5 wires on each terminal block) are not loose. Re-tighten the screws even if the wires seem to be connected4) Recalibrate5) If steps 1-4 do not work, there is a possibility one or more load cells are defective (consult with [email protected] for further instructions)

Page 30: OP-900 SERIES INDICATOR USER’S MANUAL

CONTACT USPlease e-mail [email protected] for any sales related questions.

Please e-mail [email protected] for any support related questions.or call 1(800) 360-9619 ext. 2

Don’t forget to visit our website at:

optimascale.com