MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide

MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

  • Upload
    others

  • View
    4

  • Download
    0

Embed Size (px)

Citation preview

Page 1: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 1

Aeroqual Dust Profiler User Guide

Page 2: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 2

Aeroqual Dust Profiler User Guide

Table of Contents

User Guide Revision History ............................................................................................................... 4

1. Description ..................................................................................................................................... 5 1.1 Optical Particle Counter .......................................................................................................... 6

1.1.1 Electrical connections ..................................................................................................... 6 1.1.3. Inlet Heater ...................................................................................................................... 6 1.1.4. Inbuilt Filters .................................................................................................................... 6 1.1.2 Data Outputs ................................................................................................................... 7

1.2. Pump Module .......................................................................................................................... 7 1.3. Power Module ......................................................................................................................... 8 1.4. Auxiliary Module (Optional) ..................................................................................................... 8 1.5. Crevis PLC (Optional) ........................................................................................................... 10 1.6. Cellular IP GPRS modem (Optional) .................................................................................... 12 1.7. Electrical Connections ........................................................................................................... 12

2. Assembly and Installation .......................................................................................................... 13 2.1. Assembly of heated inlet ....................................................................................................... 13 2.2. Connect Mains Power ........................................................................................................... 14 2.3. Sample Collection Filter (Optional) ....................................................................................... 15

3. Connecting to the Dust Profiler ................................................................................................. 16 3.1. Initial Connection using Aeroqual Connect ........................................................................... 16 3.2. Connection Options ............................................................................................................... 17

3.2.1. Client Mode ................................................................................................................... 17 3.2.2. Cellular Network Connection ......................................................................................... 18

3.3. Aeroqual Cloud ..................................................................................................................... 19 3.4. Connection Scenarios ........................................................................................................... 21

4. Using Aeroqual Connect and Cloud ......................................................................................... 22 4.1. Journal ................................................................................................................................... 22 4.2. Manage Data ......................................................................................................................... 22

4.2.1. Charts ............................................................................................................................ 22 4.2.2. Table ............................................................................................................................. 23 4.2.3. Download Data .............................................................................................................. 23 4.2.4. Manual Sync .................................................................................................................. 24 4.2.5. Auto Export (Optional Extra) ......................................................................................... 24

4.3. Calibration and Service ......................................................................................................... 24 4.3.1. Gain and Offset ............................................................................................................. 24

4.4. Diagnostics and Advanced .................................................................................................... 25 4.4.1. Diagnostics .................................................................................................................... 25 4.4.2. Module Settings............................................................................................................. 25

4.5. Configure Instrument ............................................................................................................. 26 4.5.1. Settings ......................................................................................................................... 26 4.5.2. Configuring the System ................................................................................................. 26 4.5.3. Configuring the Sensors ................................................................................................ 27 4.5.4. Configuring the PLC Output .......................................................................................... 27 4.5.5. PLC Output .................................................................................................................... 27 4.5.6. Sensor List .................................................................................................................... 28 4.5.7. Data filters ..................................................................................................................... 28 4.5.8. Alerts (Optional Extra) ................................................................................................... 29 4.5.9. About ............................................................................................................................. 29

Page 3: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 3

Aeroqual Dust Profiler User Guide

5. Maintenance................................................................................................................................. 30 5.1. Maintenance Schedule .......................................................................................................... 30 5.2. Maintenance Procedures ...................................................................................................... 30

5.2.1. Sample Flow Check and Adjustment ............................................................................ 30 5.2.2. Sheath Flow Check ....................................................................................................... 31 5.2.3. Leak Check ................................................................................................................... 31 5.2.4. Manual Zero Air Check ................................................................................................. 31 5.2.5. Filter Changes ............................................................................................................... 32 5.2.6. Inlet Cleaning ................................................................................................................ 32

6. Mass Calibration ......................................................................................................................... 33

7. Troubleshooting .......................................................................................................................... 34

8. Specifications .............................................................................................................................. 36

9. Schematic Diagrams ................................................................................................................... 37 9.1. Pneumatics ............................................................................................................................ 37 9.2. Electrical ................................................................................................................................ 37 9.3. Communication ..................................................................................................................... 38

10. Third Party Sensors .................................................................................................................... 39 10.1. Met One MSO ....................................................................................................................... 39 10.2. Vaisala Weather Transmitter WXT520 ................................................................................. 39 10.3. Gill WindSonic ....................................................................................................................... 40 10.4. Cirrus MK427 Noise Sensor .................................................................................................. 40 10.5. Novalynx 240-200SZ Silicon Pyranometer ........................................................................... 40

11. Mounting and Site Positioning Guidelines ............................................................................... 41 1.1. Mounting ................................................................................................................................ 41 1.2. Inlet height ............................................................................................................................. 41 1.3. Measurement Interference .................................................................................................... 42 1.4. Safety .................................................................................................................................... 42

12. Appendix ...................................................................................................................................... 43 12.1. Warranty ................................................................................................................................ 43 12.2. Technical Support ................................................................................................................. 43 12.3. Declarations .......................................................................................................................... 43

12.4. Instrument Performance & Test Report………………………...……………………………….44

12.4. Certificates………………………………………………………………………….……………….45

Page 4: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 4

Aeroqual Dust Profiler User Guide

User Guide Revision History Document Number: MRK-D-0021

Product Version: 1.3

Description: User Guide for Dust Profiler

Date Revision number Description of change Affected pages

May 2015 V1.0 Initial Release All

Page 5: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 5

Aeroqual Dust Profiler User Guide

1. Description Aeroqual’s Dust Profiler is a user friendly instrument designed to provide continuous particulate

measurement of PM10, PM2.5, PM1, and TSP. It uses an optical particle counter to count and classify

particulate matter according to size. The instrument then converts the size distribution into a mass

measurement.

The instrument includes an embedded PC and two powerful software systems, Aeroqual Connect

and Aeroqual Cloud. There are numerous additional options available, including wireless

communications, SMS and email notifications, alarm functionality and the integration of factory tested

weather sensors. The instrument is packaged in a robust, lockable enclosure and is light enough for

one person site installations. The ease of installation, as well as the extensive range of possible add-

ons, enables the unit to be used in a wide range of applications.

Note: 12 VDC Power Supply mounted externally at base

The key components of the Dust Profiler are described below.

Pump Module

(Pneumatics)

Pump Module

(Communications)

Sample

Collection Filter

(Optional) Din Rail

Terminals and

PSU termination

Fuse Holder

Sample air filter

(DQ)

Sheath air filter

(BQ) Modem

(Optional)

Optical Particle

Counter

Embedded PC

Page 6: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 6

Aeroqual Dust Profiler User Guide

1.1 Optical Particle Counter

Aeroqual uses an OEM optical particle counter from Met One

Instruments. The particle counter uses scattered light to

measure and count particles. Light from a laser diode is

collimated to illuminate the aerosol sample flow. When a

particle is present it scatters the incident laser light which is

detected using a 60o solid angle elliptical mirror at right angles

to the laser beam. The amount of scattered light is converted

to a voltage pulse and the amplitude of the pulse is calibrated

to a particle diameter. The particles are thus assigned on the

basis of size to one of eight channels.

1.1.1 Electrical connections

The connections to the optical particle counter are at the bottom of the unit. There is also a LED which

turns red if there is a fault condition (see Troubleshooting section). The optical unit also requires an

earth wire to be connected between the housing and the 0VDC line on the power bus. Please check

that this is fitted if the unit has been replaced.

1.1.3. Inlet Heater

The Dust Profiler uses a 12 V heater on the sample inlet tube to reduce the humidity of sampled air to

prevent particle growth and fogging of the particle counter optics in high RH conditions. The inlet

should always be heated.

1.1.4. Inbuilt Filters

The Particle Counter engine requires two filters which are installed externally. They are designed to

provide protection to the pump and optical window. Both filters should be replaced at specified

intervals (See Maintenance Section 5). The BQ filter is the sheath air filter and the DQ filter is the

sample air filter.

Not used

RS232 serial

LED Green = Normal Red = Fault

12 VDC in

Not used Air out

Safety: This sensor is considered a Class I laser product. Class I laser products are not

considered to be hazardous. There are no user serviceable parts inside the cover of

the sensor. The device contains a laser operating at 670 nm which is visible to the eye

and can cause damage to the eye if directly exposed. Only trained service personnel

should attempt servicing or repair of the sensor.

Page 7: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 7

Aeroqual Dust Profiler User Guide

1.1.2 Data Outputs

The Dust Profiler is configured to display and log particle mass by default. However, there is also an

option to configure the instrument to record particle counts. Refer to Section 4.5 for more information

on how to change the active sensors.

Sensor name Definition Range Units

Default configuration

PM1 Particle mass below 1 µm 200 µg/m3

PM2.5 Particle mass below 2.5 µm 2000 µg/m3

PM10 Particle mass below 10 µm 5000 µg/m3

TSP Total suspended particle mass 5000 µg/m3

Optional configuration

8PC0.3 Number of particles with diameter larger than 0.3 µm 0-1000000 particles/L

8PC0.5 Number of particles with diameter larger than 0.5 µm 0-1000000 particles/L

8PC0.7 Number of particles with diameter larger than 0.7 µm 0-1000000 particles/L

8PC1.0 Number of particles with diameter larger than 1.0 µm 0-1000000 particles/L

8PC2.0 Number of particles with diameter larger than 2.0 µm 0-1000000 particles/L

8PC3.0 Number of particles with diameter larger than 3.0 µm 0-1000000 particles/L

8PC5.0 Number of particles with diameter larger than 5.0 µm 0-1000000 particles/L

8PC10 Number of particles with diameter larger than 10 µm 0-1000000 particles/L

1.2. Pump Module

The Pump Module is split into two sections:

1. The electronics

2. The pump and pneumatics

The functionality of the electronics module can be seen in the following diagrams.

Status LED

12VDC Power

Connectors

RJ45 Connectors for

RS485 bus

Connection to

Optical Particle

Counter

Connection to

Inlet Heater

Connection

to sample

pump

Page 8: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 8

Aeroqual Dust Profiler User Guide

The pump and pneumatics are easy to access for servicing and replacement:

1.3. Power Module The Dust Profiler internal power requires regulated 12 VDC to

maintain a constant pump speed. An external mains to

12VDC power supply (Meanwell HLG80-12A 60W 12V) is

fitted on the outside of the enclosure and this will allow

connection to mains supply.

Note: If you wish to run the Dust Profiler on solar power

you will need to use a DC-DC convertor in order to supply a regulated 12 VDC for the Dust

Profiler

1.4. Auxiliary Module (Optional)

The auxiliary module acts as an interface between third party sensors and the Dust Profiler

communication bus. It is configured with different operating modes which can be selected by using

the dipswitches located on the side of the module. Aeroqual has integrated a number of third party

sensors and is able to supply the auxiliary module preconfigured for your sensors.

Programming

Port

Programming Dip

Switch

Exhaust

Sample In

Purge Out

Sample

Pump

Page 9: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 9

Aeroqual Dust Profiler User Guide

The tables below list the various functions:

Firmware: AUX_MODULE_01.

Use for: Analogue inputs, Vaisala WXT520 weather, Gill Windsonic wind, Cirrus MK:427 noise.

1 2 3 4 Function

OFF OFF OFF OFF Default - standard Auxiliary module with AN1, AN2, Freq

ON OFF OFF OFF Vaisala WXT520 with RS232 communication + AN1, AN2, Freq

OFF ON OFF OFF Vaisala WXT520 with RS232 communication + Cirrus MK427 Noise

ON ON OFF OFF Wind Sonic with RS232 communication + AN1, AN2, Freq

OFF OFF ON OFF Wind Sonic with RS232 communication + Cirrus MK427 Noise

ON OFF ON OFF Cirrus MK427 Noise module only

Firmware: AUX_MODULE_02.

Use for: Analogue inputs, Met One MSO weather, Met One 034b wind, Cirrus MK:427 noise.

1 2 3 4 Function

OFF OFF OFF OFF Default - standard Auxiliary module with AN1, AN2, Freq

ON OFF ON OFF Cirrus MK427 Noise module only

OFF ON ON OFF Met One MSO with RS232 communication + Cirrus MK427 Noise

ON ON ON OFF Met One MSO with RS232 communication + AN1, AN2, Freq

OFF OFF OFF ON Met One 034B analogue module + Cirrus MK427 Noise

The third party sensor needs to be correctly wired into the auxiliary module for it to function correctly.

The image and table below provide further information on wiring the auxiliary module.

Pin 12

Pin 1

Status LED

12VDC Power

Connectors

RJ45 Connectors

for RS485 bus

Wiring of Aux Module:

PIN 1: GND

PIN 2: 12V FUSED

PIN 3: RESERVED

PIN 4: RESERVED

PIN 5: RX

PIN 6: TX

PIN 7: 0-5V IN

PIN 8: 4-20mA IN

PIN 9: FREQ IN

PIN 10: AGND

PIN 11: METONE 034B PWR

PIN 12: CIRRUS ACTUATOR

Page 10: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 10

Aeroqual Dust Profiler User Guide

Example of wiring:

Wind Sonic (Pin 1) GND,

SIGNAL GND (Pin 2) 12V (Pin 5) RX (Pin 6) TX

Vaisala

(Pin 1) GND for

operating, data &

heating

(Pin 2) 12V for

operating & heating (Pin 5) RX (Pin 6) TX

Met One MSO

(Pin 1) GND,

SIGNAL,

COMMON, SHIELD

(Pin 2) 12V (Pin 5) RX (Pin 6) TX

Met One 034B (Pin 1) GND (Pin 11) METONE

034B PWR (Pin7) WD (Pin 9) WS

Cirrus MK:427 (Pin 1) GND,

ACTUATOR GND

(Pin 2) 12V, LOOP

IN (Pin 8) LOOP OUT

(Pin 12)

ACTUATOR IN

A programming port is also exposed through the side of the module to allow custom programs to be

loaded into the module.

Note 1: Aeroqual can supply a standard programming tool for approved technicians to

reprogram the auxiliary module to the specified requirements.

Note 2: The above wiring connections are subject to change, please refer to the third party

sensor manuals for the latest wiring instructions.

Note 3: If a Novalynx 240-200SZ Pyranometer is purchased the Auxilliary module will be

preconfigured for this sensor. Pin 2 (12V) and Pin 8 (4-20mA IN) need to be wired in.

1.5. Crevis PLC (Optional)

The Dust Profiler can be configured with a Crevis controller and

Crevis output cards. These output cards are designed to convert

the Dust Profiler’s digital values to analogue outputs. The cards

offer 4 x 4-20mA and 2 x relay outputs which can be user

configured in Aeroqual Connect or Aeroqual Cloud.

The Crevis controller type is Modbus NA-9171 and the mode type

is Modbus RTU. The outputs cards are ST-4214 for the 4-20mA

outputs and ST-2747 for the relay outputs. Further information

and specification on both these cards can be found on the USB

stick provided with the instrument.

The Crevis is powered by the Dust Profiler 12V power supply. As

a result the maximum load is 300ohms. If a higher maximum load

is required an external 24V power supply can be used to power

the Crevis. This will allow for a maximum load of at least

500ohms.

Wiring instructions can be found below. There are several glands

located at the base of the Dust Profiler which can be used to wire an external device.

12V Power In

Page 11: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 11

Aeroqual Dust Profiler User Guide

Wiring the 4-20mA output card

Wiring the relay output card

Note 1: The AC power option will not work on the Dust Profiler. The maximum voltage is

24VDC

Note 2: The default relay setting is Normally Open

Note 3: The default 4-20mA setting is to report 4mA. If this option is disabled it will continue to

read 4mA.

Note 4: Please see Section 4.5.5. for information on configuring the set points and channels.

Page 12: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 12

Aeroqual Dust Profiler User Guide

1.6. Cellular IP GPRS modem (Optional)

The Moxa G3111 Modem is a compact quad-band GPRS IP gateway

which allows connection of the Dust Profiler to a cellular network. The

modem is preconfigured to work alongside the embedded PC installed in

the instrument and is set up with a virtual server to allow direct connection

to Aeroqual Connect. If required, a DynDNS can also be created for direct

connection to Aeroqual Connect on a public dynamic IP address.

Note: See Section 3.2.2. for more information on configuring the

modem.

1.7. Electrical Connections

RS485 Bus

The two wire RS485 bus connections are made using 20 cm CAT5 cables between the sensor

modules.

12 VDC Power Bus

All modules inside the Dust Profiler operate from the 12VDC power. The power is supplied by a daisy

chain of black and red cables. To turn off the 12VDC power to the Dust Profiler, release the fuse

holder.

12 VDC Power in

12 DC power is connected into the instrument at the din rail terminals.

Page 13: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 13

Aeroqual Dust Profiler User Guide

2. Assembly and Installation

2.1. Assembly of heated inlet

Parts List:

A. Inlet Tube/Heater including power cable

B. TSP Inlet

i. Connect parts A and B

ii. Open door of enclosure and remove protective tape from the optical engine

iii. Insert Inlet Tube Assembly through base mount and fix the three mounting screws

Note 1: Make sure the O-Ring is fitted before connecting the inlet tube assembly. The O-Ring

should be lubricated to reduce wear and tear. Lubricant is provided in the Dust Monitor

Service Kit.

Note 2: Ensure the power cable is fed through the inlet hole when connecting

B A

Tape

Mounting

Screws

Page 14: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 14

Aeroqual Dust Profiler User Guide

iv. Connect power to Inlet Tube/Heaterinside the enclosure

2.2. Connect Mains Power

Locate the power connector terminal which is situated outside the enclosure at the bottom.

The power supply will already be wired to the connector however the cable to connect to

mains power needs to be wired.

Caution: The high voltage mains supply must be wired by a certified electrician in

compliance with local electrical regulations.

3. Reconnect into the connector

ensuring the plugs are fitted securely

and the nut gland tightened.

Plug

Nut gland

Connector

Terminal block

Neutral

(Blue)

Earth

(Yellow/Green)

Live

(Brown)

2. Feed the power cable through the nut gland

plugs and connector and wire the cable to the

terminal block according to the diagram.

Note: To remove the terminal block from inside the connector unclick it from the casing by

turning it slightly then pull out

1. Unscrew the nut

gland to loosen the

terminal block and pull

connector apart

Power cable

from heated inlet

Page 15: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 15

Aeroqual Dust Profiler User Guide

2.3. Sample Collection Filter (Optional)

The Dust Profiler can be fitted with a 47mm collection filter holder if requested. It is used to collect

particulate samples for composition analysis.

Note: If the Dust Profiler is shipped with a sample collection filter holder, it

will not contain a filter on arrival. The filter will be packed separately in the

accessories box.

If you wish to use the sample collection filter:

1. Unscrew the collection filter housing, remove the black gasket and insert

a 47 mm filter.

Note: When the filter holder is not being used make sure the gasket is

fitted to reduce potential for leaks.

2. Tighten the collection filter housing using the wrench provided.

3. Check and adjust the sample flow to 1.0 LPM (See Section 5.2.1.)

The sample collection filter must be changed at regular intervals to avoid the filter blocking and

overloading the pump.

Instrument type Elevated PM levels Typical PM levels

Dust Profiler 2 weeks 4 weeks

Note 1: These are recommended maximum timeframes

Note 2: See Section 9.1 for flow diagram

Page 16: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 16

Aeroqual Dust Profiler User Guide

3. Connecting to the Dust Profiler

3.1. Initial Connection using Aeroqual Connect

Aeroqual Connect is the standard user interface which comes with the Dust Profiler and allows the

user to be connected directly to the instrument. It can be accessed via WIFI, Ethernet or a cellular

modem. The on-board computer has an Ethernet output for direct wired connection. Alternatively,

connection can be made over WIFI by connecting to the Dust Profiler network (Access point mode), a

local network (Client mode) or via a modem.

The default setting of the instrument when first purchased will be to connect to Aeroqual Connect via

access point mode. Access point mode allows connection to the Dust Profiler via the Dust Profiler

network signal which will show up on a laptop, tablet or smart phone within range. This type of

connection will most commonly be used when working on the instrument in the field. However, it will

also be used to initially connect when the instrument first arrives.

Connecting by this mode will mean the user’s device is connected via WIFI directly to the Dust

Profiler. There will be no internet connection on the user device when using this option. To connect

you will need to select the Dust Profiler WIFI network and enter the password.

Note: The default user name and passwords can be found on Page 1 of the log book.

Once a WIFI connection has been made Aeroqual Connect can be accessed via the PC browser by

entering the IP address 10.10.0.1 in the address bar. The login screen below will then be displayed.

Page 17: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 17

Aeroqual Dust Profiler User Guide

1. An Aeroqual Connect login name and password will be provided when Aeroqual Connect is

first purchased. If you do not know your login details please contact [email protected].

2. Once you have logged in the homepage will be displayed.

Note 1: Different applications will appear depending on the user accessibility configured for

your account.

Once the first connection has been made (via access point mode) the WIFI communications can be

configured according to how you would like the system to respond. This can be done via the settings

tab under the Configure Instrument application.

3.2. Connection Options

3.2.1. Client Mode

Client mode allows connection to the Dust Profiler via a local network (e.g. your office network). If

there is a local WIFI or LAN network available at the Dust Profiler installation site then the instrument

can be set up to connect to this network. This might be the case if the Dust Profiler is installed on the

perimeter of an industrial site and the WIFI network extends out to where the unit is installed.

Alternatively the Dust Profiler can be connected to the local network via an Ethernet cable from the

instrument to the local router using the Ethernet output on the base of the Dust Profiler.

When the Dust Profiler is connected to your local network (such as your office WIFI network), the local

network will automatically assign a local IP address to the instrument. The IP address assigned to the

instrument can be found in the Configure Instrument application when logged in via Aeroqual Cloud.

Entering this IP address into the web browser will open Aeroqual Connect for that specific instrument,

assuming you are within the local network.

If the local network allows outside access (e.g. via VPN), connection to the Dust Profiler can be made

from anywhere in the world. In the more common scenario where the local network does not permit

outside access you may still connect to the Dust Profiler using Aeroqual Cloud through the internet

because most local networks permit internet access.

To configure the Dust Profiler to use client mode the instrument needs to be configured via the

settings tab under the Configure Instrument application.

1. The WIFI mode needs to be set to Client

2. The WIFI SSID (local network name) needs to be entered (e.g. your office WIFI name)

3. The WIFI password needs to be entered (e.g. the password for your office WIFI)

Note 1: Before changing the WIFI settings it recommended that you have another connection

to the instrument established – typically by connecting the instrument to your LAN using the

Ethernet connection and using Connect with the LAN IP address.

Note 2: Once WIFI client mode is set up Aeroqual Connect will automatically disconnect from

WIFI access point mode. Connection will then need to be made via Aeroqual Cloud (See

section 4.5) to see the local IP address assigned to the instrument.

Note 3: If the local WIFI network password changes or you enter it incorrectly, you could be

locked out of the Dust Profiler. If this happens you will need to re-set the Dust Profiler back to

factory default settings, which puts the Dust Profiler back in to Access Point mode with a

default password. You can then reconfigure the settings correctly. Contact

[email protected] for information on how to perform this process.

Page 18: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 18

Aeroqual Dust Profiler User Guide

3.2.2. Cellular Network Connection

If the Dust Profiler has an IP modem installed in the instrument a connection can be made via a

cellular network. This option is useful when remote access to the Dust Profiler is required but there is

no local network signal at the instrument installation site so client mode cannot be used. Aeroqual will

preconfigure the modem and load a default APN (Internet). This may work without having to be

changed. If set up is still required, only the APN, User name and Password needs to be modified.

Note1: A local data telecommunications contract and SIM card is required.

Note 2: The Dust Profiler uses approximately 20 megabytes per sensor per day. This is an

estimate for a Cloud connection and can vary wildly depending on the cellular provider.

Note 3: Using Aeroqual Connect via the modem will incur higher data usage.

1. The IP modem will need to be set up by plugging directly from the modem Ethernet port to a

laptop Ethernet port and browsing to 192.168.127.254.

2. In the menu select Network Settings GSM GPRS Settings and configure settings as per

picture below. Contact your SIM card provider to obtain full ATD and APN settings. Click

Submit button.

3. Unplug MOXA unit from the network and plug it into Dust Profiler embedded PC. Restart Dust

Profiler embedded PC and modem.

4. In the Configure Instrument application the Ethernet Mode setting needs to be set to Auto

(DHCP). WIFI mode should still be set to Access Point so a direct connection can still be

made.

5. If you do not wish to use Aeroqual Cloud, a DynDNS service will need to be set up to create a

permanent web address for the Dust Profiler.

6. A Dyn account can be created at https://account.dyn.com/cart/?via=upgrade-now

7. Log in to your account and add a new host name:

Page 19: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 19

Aeroqual Dust Profiler User Guide

8. Go back to MOXA modem configuration interface, and choose System Management - > Misc.

Network Settings - > DDNS. Select enable and DynDns.org. Input the host name created in

Dyn account, user name and password as well for the Dyn account. Then click Submit button,

Save and Restart. Now the modem should be ready to be connected via host name.

9. Enter the set up web address into the web browser to access Aeroqual Connect.

Note 1: No data is stored on the DynDNS server. It only acts as an access point to

communicate with the Dust Profiler.

Note 2: In some countries the telecommunications provider will block incoming requests to

the modem. If this is the case DynDns will not work with regular SIM cards. Contact your

telecommunications company about using an open APN or use a private APN.

3.3. Aeroqual Cloud When using Aeroqual Cloud, the user is connected to the Aeroqual Cloud server. The Dust Profiler

instrument is also connected to the Cloud server. The user connects to Cloud over the internet; the

Dust Profiler is connected to the Cloud server using WIFI, Ethernet or Cellular (modem required).

Aeroqual Cloud is a website.

Aeroqual Cloud allows management of either a single Dust Profiler unit or a network of Dust Profiler

units by using any internet-enabled browser.

Page 20: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 20

Aeroqual Dust Profiler User Guide

Aeroqual Cloud has all of the control features of Aeroqual Connect. The additional benefits of having

Aeroqual Cloud include:

View and manage a network of instruments

Remotely update instrument firmware

Advanced user access management

Enjoy faster browsing speeds

Minimise data usage costs

Back up data so it can never be lost

Access data even when the instrument is offline

Send SMS/text and email alerts (optional)

Auto export data via FTP to another server (optional)

Auto export a data file via email (optional)

As a user, you simply access Aeroqual cloud through the internet; Aeroqual Cloud is a website which

provides total control over your Dust Profiler. You will need to go to http://cloud.aeroqual.com and

enter your login name and password.

Note 1: If you do not know your login name and password please contact

[email protected]

As well as having access to all the features available in Aeroqual Connect, Aeroqual Cloud can be

used to manage a network of Dust Profilers. Depending on the user accessibility configured to your

account, Dust Profiler units can be managed by Organisations and Networks. This feature is available

in the Administration application.

The Dust Profiler must have internet access for you to see it and control it using the Cloud website.

For the Dust Profiler to have internet access it must be either in Client mode, as described above, and

be on a network which permits internet access (most networks do permit internet access). A modem

is not required in this situation.

Or if a local network is not available, the Dust Profiler must have an IP modem installed and a SIM

card and data contract with a local telecommunications provider.

Organisations Networks Instruments

Page 21: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 21

Aeroqual Dust Profiler User Guide

3.4. Connection Scenarios

There are a number of ways to connect to the Dust Profiler instrument using Aeroqual Connect or

Aeroqual Cloud. To help clarify which connection should be used please refer to the tables below

which shows the 4 most common scenarios.

Dust Profiler

Location

User

Location Recommended Connection

Office Office

The Dust Profiler should be connected to the office WIFI network

using client mode. You can then connect using Aeroqual Cloud.

Aeroqual technicians will also be able to connect using Aeroqual

Cloud if support is required. A wired Ethernet connection can also be

made from the Dust Profiler direct to the local router.

In the field and

modem is installed Office

If the Dust Profiler is on a 3G network permanently you can connect

using Aeroqual Cloud from your office at any time.

In the field and

modem is installed At site

If you are in the field with the Dust Profiler, you can connect directly

to the instrument via Aeroqual Connect using the Dust Profiler as an

access point

In the field and no

modem installed At site

If you are in the field with the Dust Profiler and there is no 3G

available you can connect as an access point while in the field. Use

manual sync to upload data to Aeroqual Cloud when back in the

office and WIFI is available.

If you are not using Aeroqual Cloud there are 3 options:

Dust Profiler

Location User Location Recommended Connection

In the field and modem

is installed Office

If the Dust Profiler is on a 3G network permanently use a

DYDNS service to connect directly to the instrument via

Aeroqual Connect.

In the field and modem

is installed At site

If you are in the field with the Dust Profiler, you can connect

directly to the instrument via Aeroqual Connect using the Dust

Profiler as an access point

In the field At site If there is no 3G available you can only connect to the Dust

Profiler using WIFI access point whilst on site.

Aeroqual Connect is the lead system. If both the Aeroqual Connect and Cloud systems are

being run simultaneously, any changes made in Aeroqual Connect will override the

changes made in Aeroqual Cloud.

Page 22: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 22

Aeroqual Dust Profiler User Guide

4. Using Aeroqual Connect and Cloud Aeroqual Connect and Cloud have a range of applications available which allow the user to access

data, assess performance, schedule calibrations and diagnose faults remotely.

4.1. Journal The journal can be accessed by selecting Journal on the menu bar located at the top of the screen. It

is used to log when any changes are made to the Dust Profiler. The journal is searchable by date

using the side bar and filtered via the drop down menu.

The filter allows entries to be searched by:

User Entry – Manually entered journal entries

Calibration – Changes associated to calibration e.g. gain and offset modifications and

notification when service actions start and stop

Configuration – Changes made to the Dust Profiler configuration e.g. sensor added or

communications altered

Module Settings – Changes made to the sensor module settings

System changes – Notifies when the Dust Profiler starts up, shuts down or is rebooted

Note 1: To manually create a journal entry select “Add user entry”, type in the relevant

information and press save.

Note 2: Service mode is initiated by starting “Manual Service Mode” in the Calibration and Service application. A journal entry will be automatically generated when the service is started and stopped.

4.2. Manage Data

The manage data application allows the user to view the data either as a chart or in a table. There is

the option to view the raw or averaged data as well as the ability to filter the data to show specific time

periods. The data can also be exported for further analysis within this application.

4.2.1. Charts

To view the data in graphical format select the “Charts” option.

The charts can be configured to view a specific time period using the “From” and “To” dates

shown at the top of the screen.

An averaging period can also be selected from the drop down menu

Note: The minimum average period is determined by the date range

Page 23: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 23

Aeroqual Dust Profiler User Guide

4.2.2. Table

To view the data in a table select the “Table” option.

The data can be configured to view a specific time period using the “From” and “To” dates

shown at the top of the screen.

The data can also be filtered using the “Averaging Period” drop down menu to display:1

minute

Note 1: The minimum average period is determined by the date range

Note 2: Depending on the administration rights, raw data can also be viewed. This is 1 minute

data without data filtering in place

4.2.3. Download Data

Data can be downloaded by selecting “Download Data”.

A specific data set can be chosen for download by using the “From” and “To” date selection.

The Journal can also be downloaded if required.

Note: The data will be downloaded in csv. format

Page 24: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 24

Aeroqual Dust Profiler User Guide

4.2.4. Manual Sync

If the Dust Profiler is outside of cellular and WIFI range, the Dust Profiler data can always be

downloaded to your device using a WIFI access point (see Section 3.1.) when at the instrument site.

The data will be downloaded to an unreadable file and will include all information stored on the Dust

Profiler on-board PC. The file can then be uploaded onto Aeroqual Cloud, once a WIFI or cellular

network is available, to view the data from the unit, including diagnostics and settings.

To download the data: In Aeroqual Connect select Manual Sync in the Manage Data application.

Here it will prompt you to select a location on your PC to download the data to.

To upload the data: In Aeroqual Cloud select Manual Sync in the Manage Data application. Here it

will prompt you to upload the file which was previously downloaded from Aeroqual Connect.

Note 1: If there is a lot of data to download from Aeroqual Connect multiple manual syncs may

need to be performed. The date in the manual sync function updates showing how much data

has been downloaded.

4.2.5. Auto Export (Optional Extra)

The auto export option allows the data to be automatically exported to either a selected email address

or to an FTP server. To set either of these features up they need to be enabled by selecting the

enable check box under the auto export option.

Email: To enable auto export to a specific email the frequency, format and averaging period needs to

be selected. There is also the option to include the journal in the export. Make sure a correct email

address is entered and save the changes for it to take effect.

FTP: To enable auto export to a FTP server the frequency, format and averaging period needs to be

selected. There is also the option to include the journal in the export. The FTP server, user name and

password also need to be entered correctly. Save the changes for it to take effect.

Note: The journal captures any error messages; these are particularly useful for

troubleshooting.

4.3. Calibration and Service The calibration and service application allows the user to perform manual adjustments to the gain and

offset settings for the particle counter.

4.3.1. Gain and Offset

Selecting the “Gain and Offset” menu will

display the gain and offsets for all the

configured PM sizes in the Dust Profiler.

Between 5 and 50 real time measurements will

also be displayed, as well as the average and

standard deviation for the shown readings.

These numbers will automatically update.

Page 25: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 25

Aeroqual Dust Profiler User Guide

Prior to any manual calibration or servicing activities the “Manual Service Mode” button

should be started. This will ensure the raw data is highlighted and labelled appropriately to

show that a service is being performed. This is particularly helpful to alert when a calibration

has taken place to ensure the readings are disregarded from any future data analysis.

Note 1: Any flagged data is disregarded when calculating averaged readings.

Note 2: Remember to stop the “Manual Service Mode” after the service has been completed to

ensure future reported data is tagged as “Sample” data.

Note 3: The Manual Service Mode will automatically stop after 24 hours if left running.

The calibration parameters can be adjusted manually by clicking the setting that requires

changing, entering the new value and pressing save.

4.4. Diagnostics and Advanced The diagnostics and advanced application allows the user to view the sensor module settings as well

as the sensor diagnostics. This information is useful when fault diagnosing an issue with a module.

The diagnostic data can be exported for further analysis and the sensor module settings can be

updated or modified in this application.

4.4.1. Diagnostics

The diagnostics view will display the diagnostics for the selected parameter. The data can be

paused or exported using the buttons above the table.

40 minutes of data will be displayed at a time and the data will automatically update unless

the pause button is pressed.

4.4.2. Module Settings

The module settings view will show the individual module settings and version number for all the

parameters configured inside the Dust Profiler.

The module settings can be changed by selecting the value that needs changing, entering the

new value and clicking save.

Note: Incorrect modification of module settings may cause irreversible damage. Do not change

settings without contacting Aeroqual or a qualified service agent.

Page 26: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 26

Aeroqual Dust Profiler User Guide

4.5. Configure Instrument The configure instrument application stores all the configuration settings relevant for the specific Dust

Profiler connected. Depending on the user type different features will be accessible. The application

can also be used to view a complete sensor list for all parameters that have been configured on the

Dust Profiler, filter data and set up SMS and email alerts if this function has been purchased.

4.5.1. Settings

The settings tab is split into 5 columns which display different information relevant to the Dust Profiler.

4.5.2. Configuring the System

The serial number is listed here. It cannot be changed and is also shown on the

log book and inside the Dust Profiler enclosure.

Page 27: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 27

Aeroqual Dust Profiler User Guide

The name of the instrument can be set by the user, for example, to reflect the

location or customer that is running the Dust Profiler

The device, software version and time last contact was made with the cloud

server is listed here for reference.

The time zone of the Dust Profiler can be set here to display the correct date for

the region the instrument is installed in.

The default averaging period can be configured here. Data can be configured to

read from 1 minute to daily averages.

Service time out can be set here. This is important if the manual service mode is

left on by accident. The time out value will stop manual service mode after the

time period set.

The restart instrument checkbox will force a restart of the embedded PC. The

checkbox will be cleared automatically after restart.

4.5.3. Configuring the Sensors

The default unit of measurement for PM is µg/m3.

The poll interval and poll time out functions will control how often the embedded

PC polls the modules. This should not be changed.

The active sensors will be displayed here. If a module has been added the

sensor will need to be selected from the drop down menu and added to the active

sensors list. Sensors can also be removed by clicking on the sensor name.

4.5.4. Configuring the PLC Output

If a Crevis is installed in the Dust Profiler the 4-20mA output module and relay

output module options should be check marked.

The baud rate and COM port used will also be shown here. These settings

should not be changed.

Note 1: For information on the communications section please see Section 3.

Note 2: For information on the legacy section please contact Aeroqual.

4.5.5. PLC Output

The 4-20mA and relay output modules can be configured here. There are four 4-20mA output

channels available which can be set to read a user defined parameter and range. The relay output

module has two relay options that can also be user configured. The paramenter, set point and reset

point can be added here. The relays can be configured as either rising or falling edge triggers.

If the set point is exceeded the relay is activated. This can be used to activate a siren or strobe or

other peripheral devices. The relay remains activated until the measured value drops below or

reaches above the reset point.

Rising Edge: Relay is energised when the concentration reaches above the set point and stays on

until the concentration drops below the reset point, which is set at a lower concentration than the set

point.

Falling Edge: Relay is energised when the concentration is below the set point and stays on until the

concentration reaches above the reset point, which is set at a higher concentration than the set point.

The relay will remain off until the concentration drops down below the set point.

Page 28: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 28

Aeroqual Dust Profiler User Guide

Note: Please see Section 1.5 for more information on the Crevis PLC

4.5.6. Sensor List

The sensor list tab displays all the sensors that can be configured into the Dust Profiler. The table is

mainly for reference and shows all the settings configured for each sensor e.g. the units and

conversion factors. If a new sensor list is released it can be imported in this section.

4.5.7. Data filters

The negative number filter can be turned on for specific

parameters in this section. If you do not want the data to

show negative numbers for certain parameters, check the

checkbox of that specific parameter. Any negative

numbers reported will now show as zero.

Note: The negative number filter is NOT applied to

RAW readings. It is applied to all averaged readings.

Page 29: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 29

Aeroqual Dust Profiler User Guide

4.5.8. Alerts (Optional Extra)

SMS and email alerts can be set up in this

section. Alerts can be set to notify when:

1. a concentration exceeds the user

defined set point

2. a sensor fails

3. the Dust Profiler is in operation mode

4. the Dust Profiler goes offline

Note 1: Operation mode allows alerts to be

sent when the mode changes. For example,

if the unit changes from normal mode to zero

calibration, or manual service mode an alert

is triggered. An alert will also be sent when

the unit returns to normal mode.

An email address and/or SMS number needs to be entered into the relevant fields to allow the alerts

to be sent. Multiple alerts can be set up by pressing the Add a new alert button.

Note 2: Phone numbers need to be in the format +64234566 with no spaces. A space is treated

as a separate phone number.

Note 3: SMS alert allowance is 50 per month. If two phone numbers are entered in a single alert

it counts as 2 messages.

4.5.9. About

The about screen shows information regarding your Aeroqual

Cloud license. It shows:

Instrument Type

Version

Licence Type

Last Contact

SMS Alerts Remaining

SMS Monthly Allowance Roll Over Date

Page 30: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 30

Aeroqual Dust Profiler User Guide

5. Maintenance

5.1. Maintenance Schedule

Procedure Section Frequency

Sample Flow Check 5.2.1 Monthly

Sheath Flow Check 5.2.2 Monthly

Manual Zero Air Check 5.2.4 Monthly

Filter Changes 5.2.5 3 to 6 months

Inlet Cleaning 5.2.6 3 Months

Optical sensor factory calibration

Contact Aeroqual 12 Months

Pump Replacement Contact Aeroqual 24 Months

Note 1: Suggested intervals for maintenance procedures may vary with sampling intensity and

environmental conditions

1.7.1.1.1.1.1. Note 2: All checks should be carried out during the initial commissioning of the Dust Profiler

5.2. Maintenance Procedures

5.2.1. Sample Flow Check and Adjustment

A constant 1.0 LPM flow is essential to ensure monitor is measuring correctly. To measure or set the

flow perform the procedure below:

1. Remove the TSP head from the inlet and connect a 0-2.5 LPM volumetric flow meter to the sample

inlet. Ensure the flow meter is on a steady surface and the correct flowmeter port is connected to the

Dust Profiler inlet.

2. Adjust the position of the Sample flow adjustment valve until the flowmeter reads 1.0 LPM.

Note: The inlet should be connected to the TOP of the volumetric flow meter (negative flow)

Afterwards replace the inlet components with care making sure there is no leak.

Page 31: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 31

Aeroqual Dust Profiler User Guide

5.2.2. Sheath Flow Check

The sheath flow is a constant stream of air which ensures the optics remains clean. It is important this

is working correctly to maintain the accuracy of the measurement. To check the sheath flow:

1. Remove the TSP head from the inlet and connect a 0-2.5 LPM volumetric flow meter to the

sample inlet. Ensure the flow meter is on a steady surface and the correct flowmeter port is

connected to the Dust Profiler inlet.

2. Check that the sample flow is 1 LPM, if not adjust Sample flow valve (see above).

3. Pinch the purge tubing to block flow through the purge. This will increase the sample flow. If

the sample flow > 2.0 LPM then the sheath flow is good. If the sample flow < 2.0 LPM then

there is a problem. Change the filters and check for a leak.

5.2.3. Leak Check

If the correct sample or sheath flow cannot be achieved, there may be a leak in the pump module or

engine. First check the entire flow system:

Remove exhaust connection from the base of the enclosure

Remove the TSP head and block the PM inlet.

Connected the pressure end of a flow meter to the exhaust tubing.

If there is no leak the flow should drop to zero.

If the flow does not drop to zero it suggests there is a leak somewhere in the system. To check the

pump module:

Remove the sample connection and cap off the module sample port.

Connect the exhaust port of the module to the pressure end of a flow meter.

The flow should drop to zero.

If there is a leak in the pump module the module will need to be sent back to the Aeroqual factory.

Please contact technical support.

If there is no leak in the pump module it is likely the leak is located in the engine. In this case, the

engine will need to be sent back to the Aeroqual factory. Please contact technical support.

5.2.4. Manual Zero Air Check

A zero air check can be carried out manually. To do this the TSP inlet needs to be removed and the

particle filter (supplied with the instrument) needs to be attached to the monitor. Ensure there is a

good seal around the inlet.

The filter will remove 99.99% of particulates from ambient air. Wait 5 minutes and then check the

readings on the control module. The reported value should be <2 µg/m3.

Page 32: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 32

Aeroqual Dust Profiler User Guide

5.2.5. Filter Changes

There are two green filters installed in the Dust Profiler. They are designed to provide

protection to the pump and optical window. The filters will become dirty and must be

replaced periodically.

To do this, unclip the filters from the connectors and replace. Ensure the filters are

connected the right way in accordance to the flow arrow. These filters can be

purchased from Aeroqual. Please contact technical support for a quotation.

Note: BQ filter – Sheath air filter

DQ filter – Sample air filter

5.2.6. Inlet Cleaning

The TSP inlet can be dismantled by unscrewing the 3 screws and separating the head into two parts.

Use a lint-free cloth wetted with isopropyl alcohol to clean both the inside and outside of the inlet.

Screws

Page 33: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 33

Aeroqual Dust Profiler User Guide

6. Mass Calibration

All particle monitors based on light scattering require a site calibration for optimum accuracy. This is

because the average density and light scattering behaviour of dust particles will vary from site to site.

This adjustment is sometimes known as a “K-factor”. Aeroqual recommends a site calibration be

performed at the time of initial installation and then every 6-12 months. Calibration is performed by co-

locating the Dust Profiler with a Reference Particulate Matter measurement system and adjusting the

Gain of the particle monitor so the average reading matches that of the Reference Instrument. A list of

Reference instruments can be found at: http://www.epa.gov/ttnamti1/files/ambient/criteria/reference-

equivalent-methods-list.pdf

Install the Dust Profiler beside the Reference instrument. Make sure both instruments are setup

correctly. Two procedures are given below depending on whether the reference instrument is a

gravimetric or continuous instrument.

Procedure overview for Gravimetric Reference Instrument

1. Weigh a reference filter and install into the Reference instrument.

2. Leave the Dust Profiler and Reference instrument running for at least 24 hours until a weighable

quantity of particulate matter sample has been collected.

3. Check the sample flows of both instruments to ensure they are within specification.

4. Remove the Reference filter and re-weigh

5. Download the Dust Profiler PM data for the calibration period.

Calculations

1. Reference Filter Weight of PM (ug):

PM(Ref) = (filter weight at finish) – (filter weight at start)

2. Volume of air sampled by reference sampler (m3)

Volume(Ref) = minutes of test * Flow Ref (LPM) / 1000

3. Calculate average reference PM density (ug/m3)

AvPM(Ref) = PM(Ref)/ Volume(Ref)

4. Calculate average PM (ug/m3):

AvPM(Dust Profiler) = sum(PM readings over test period)/number of readings

Procedure overview for a Continuous Reference Instrument

1. Leave the Dust Profiler and Reference instrument running for at least 24 hours

2. Check the sample flows of both instruments to ensure they are within specification.

3. Download the Dust Profiler and Reference instrument data for the calibration period Calculations

1. Calculate average Dust Profiler PM (ug/m3):

AvPM(Dust Profiler) = sum(PM readings over test period)/number of readings

2. Calculate average Reference PM (ug/m3):

AvPM(Ref) = sum(PM readings over test period)/number of readings

Calculation of Dust Profiler Gain

1. Calculate K factor

K factor = AvPM(Ref) / AvPM(Dust Profiler)

2. You should then apply this to the Gain Factor on all the PM channels.

new GAIN = old GAIN x K Factor

Page 34: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 34

Aeroqual Dust Profiler User Guide

7. Troubleshooting

Symptom Possible Cause Fault isolation/Solution

PM values seem incorrect

Sample flow incorrect

Sample and/or purge filters

dirty

Gain factors incorrect

Offset factor incorrect

Check sample flow and adjust to

correct value using flow adjusters

Replace filters

Perform span calibration

Perform zero calibration

Low Sensitivity Laser is old

The laser current can be

measured under the

Diagnostics and Advanced

application. If above 20 mA

then fault

Laser needs replacing.

Send to your authorised distributor or

contact Aeroqual technical support.

Low Sensitivity Dirty optics

Check H0 value for PM in

Module Parameter Table. Ig

above 0.1 then optics are dirty

Optics need cleaning

Send to your authorised distributor or

contact Aeroqual technical support.

Noisy Readings

Laser is old

Send to your authorised distributor or

contact Aeroqual technical support.

Laser needs replacing.

Send to your authorised distributor or

contact Aeroqual technical support.

Negative readings Purge filter new and shedding

particles

Purge flow zero

Purge filter dirty

Run zero cycles until purge filter no

longer shedding particles.

Adjust purge flow

Replace purge filter

Fibre Span has changed

significantly since installation

Dirty Optics

Laser is old

Optics need cleaning

Send to your authorised distributor or

contact Aeroqual technical support.

Page 35: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 35

Aeroqual Dust Profiler User Guide

Laser needs replacing.

Send to your authorised distributor or

contact Aeroqual technical support.

Readings flat Laser failed (check laser

current)

Photo-detector failed

Laser needs replacing.

Send to your authorised distributor or

contact Aeroqual technical support.

Send optical engine to Met One for

replacement detector and calibration.

Page 36: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 36

Aeroqual Dust Profiler User Guide

8. Specifications

Optical Particle Counter

(OPC)

PM1, PM2.5,

PM10 and TSP

PM1 PM2.5

PM10 TSP

200 μg/m3

2000 μg/m3

5000 μg/m3

5000 μg/m3

<±(5 μg/m3 + 15%

of reading) 1.0 LPM <1 μg/m3

Control System Embedded fanless PC, Intel Atom N2600, 1.6GHz, 2GB RAM, 32GB SSD, Ubuntu Linux

Communications Standard: WIFI, Ethernet (LAN)

Optional: Cellular IP GPRS modem

Software

Connect: Runs on embedded PC, accessed via web browser ( IE, Firefox, Chrome, Safari)

Cloud: Runs on secure ‘cloud’ servers, accessed via web browser

Connect / Cloud Features: configuration, diagnostics, journal, calibration and data acquisition,

plus SMS and email alerts (optional), and auto data export via FTP and email (optional)

Data logging 32GB Hard Drive (>20 years data storage)

Averaging period 1 min, 5 min, 10 min, 15 min, 30 min, 1 hr, 2hr, 4 hr, 8 hr, 12 hr, 24 hr

Outputs

RS232 (legacy mode)

2 x Relay (optional)

4 x 4-20mA (optional)

Power requirements 100-260VAC (standard): 15W / 24W*

Regulated 12VDC (if required): 15W / 24W*

Enclosure Lockable IP65 GRP cabinet with integrated aluminium solar shield armour

Inlet: 36cm heated inlet

Dimensions 483H x 330W x 187D mm (including solar shield armour & mounting brackets)

Weight <13 kg*

Environmental operating range -10°C to +50°C

Mounting Pole, tripod and wall mounting brackets included

47mm Sample Filter (Optional) 47mm filter for particle loading analysis

Factory Integrated & Tested

Sensors (Optional)

Gill WindSonic (ultrasonic wind sensor)

Vaisala WXT520 (weather transmitter)

Met One MSO (weather transmitter)

Cirrus MK427 Class 1 (noise sensor)

Novalynx Pyranometer (solar radiation)

Page 37: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 37

Aeroqual Dust Profiler User Guide

9. Schematic Diagrams

9.1. Pneumatics

9.2. Electrical

Page 38: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 38

Aeroqual Dust Profiler User Guide

9.3. Communication

Page 39: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 39

Aeroqual Dust Profiler User Guide

10. Third Party Sensors

10.1. Met One MSO Description: Measures wind speed and direction; air temperature; relative

humidity; and barometric pressure

Specifications: For full details visit the company website www.metone.com

10.2. Vaisala Weather Transmitter WXT520 Description: Measures wind speed and direction; liquid precipitation; barometric

pressure; air temperature; and relative humidity

Specifications: For full details visit the company website www.vaisala.com

Wind Speed

Range 0-50 m/s

Accuracy ±2%

Resolution 0.1 m/s

Wind Direction

Range 0-360°

Accuracy ±5°

Resolution 1°

Air Temperature

Range

Accuracy

-40°C - +60°C

±0.5°C

Resolution 0.1°C

Relative Humidity

Range

Accuracy

Resolution

0-100%

±4%

1%

Barometric Pressure

Range

Accuracy

Resolution

500-1100 mbars

± 2 mbars

0.1 mbar

Wind Speed

Range

Accuracy

0-60 m/s

±3% (0-35 m/s)

Wind Direction

Range

Accuracy

0-360°

±3°

Liquid Precipitation

Rainfall

Output Resolution

0.1mm/min

Barometric Pressure

Range

Accuracy

600-1100 hPa

±1 hPa @ -52 to +60 °C

Air Temperature

Range

Accuracy @ +20 °C

-52 to +60 °C

±0.3 °C

Relative Humidity

Measurement range

Accuracy

0-100 %RH

±3 %RH (0-90 %RH); ±5 %RH (90-100

%RH)

Page 40: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 40

Aeroqual Dust Profiler User Guide

10.3. Gill WindSonic Description: 2-axis ultrasonic wind sensor, measures wind speed and

direction

Specifications: For full details visit the company website www.gill.co.uk

Wind Speed

Range 0-60 m/s

Accuracy ±2⁰ @ 12 m/s

Resolution 0.01 m/s

Wind Direction

Range 0-359° (no dead band)

Accuracy ±3⁰ @ 12 m/s

Resolution 1 °

10.4. Cirrus MK427 Noise Sensor

Description: Outdoor environmental noise meter:

Specifications: For full details visit the website www.cirrus- environmental.com/

10.5. Novalynx 240-200SZ Silicon Pyranometer Description: Outdoor environmental solar radiation meter:

Specifications: For full details visit the website www.novalynx.com/

Frequency Weighting dB(A) to IEC 61672-1:2002

Automatic Calibration Electrostatic Actuator System

with DC voltage control

Measurement Range 30 – 100dB(A)

Resolution 0.1 dBA

Sensor High stability silicon voltaic detector

Accuracy ± 5% typical under natural daylight conditions

Sensitivity 100 µA per 1000 W/m² typical

Linearity Max deviation of 1% up to 3000 W/m²

Resolution 0.1 W/m2

Temperature

dependence 0.15% per °C max

Operating

temperature -40°C to +65°C (-40°F to +149°F)

Page 41: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 41

Aeroqual Dust Profiler User Guide

11. Mounting and Site Positioning Guidelines

It is important that the positioning of the Dust Profiler is suitable so as to yield data which is

representative of that specific location.

1.1. Mounting The Dust Profiler comes with 2x mounting brackets and 3x U-bolts for 2 inch poles. The same

brackets are used for both pole and wall mounting. To wall mount the Dust Profiler the brackets are

screwed into the wall instead of using the U-bolts.

For pole mounting:

1. Secure top mounting bracket at a suitable height using 2x U-bolts provided

2. Fix lower mounting bracket to the Dust Profiler chassis with screws

3. Hang the Dust Profiler on the top mounting bracket

4. Place U-bolt through lower mounting bracket

Note: Only one U-bolt is used to secure the bottom bracket

5. Add nuts and washers and tighten the bottom U-bolt into position

1.2. Inlet height If PM monitoring is related to human exposure, the sampling inlet height should be positioned

in the “breathing zone”. This is located between 2 and 15 meters above ground level.

If PM monitoring is related to specific emission sources, the position of the sampling inlet can

be more flexible. It is more important that there is no obstruction between the approaching air

from the emission source and the sampling inlet.

If more than one Dust Profiler, or any other particulate instrument, is being used at the site,

the height of the inlets should be uniform.

If the Dust Profiler inlet is the highest point at the site, a lightning rod must be installed to

prevent damage to the unit during electrical storms.

1

.

2. 3.

4. 5.

Page 42: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 42

Aeroqual Dust Profiler User Guide

1.3. Measurement Interference The meteorological conditions of the site should be taken into consideration when positioning

the Dust Profiler. For example, there should be no obstruction to the air flow in the

predominant wind direction. A minimum clear sky angle of 120 degrees is recommended.

The inlet should be at least 1 meter away from any objects that could potentially influence the

airflow characteristics e.g. trees, vertical surfaces or walls.

Avoid overhead high-voltage cables which may cause electrical interference with the

sampling equipment.

Demolition/construction activities and change to normal transport patterns due to road works

etc. can significantly affect the data. Ensure a record of such events is kept to account for

unexpected peaks in concentration.

1.4. Safety The intended data capture rate should be considered when positioning the Dust Profiler. If

data capture above 90% is essential, the unit should be located in an area which has 24 hour

access available.

The positioning should allow for routine maintenance checks to be performed safely by

personnel.

If using a tripod, ensure the tripod legs are bolted to the ground to prevent the unit from

falling. Ensure the Dust Profiler is in a secure location to avoid vandalism or theft.

Page 43: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 43

Aeroqual Dust Profiler User Guide

12. Appendix

12.1. Warranty

Please review the Aeroqual Warranty Policy on our website at www.aeroqual.com

12.2. Technical Support Technical information, service and spare parts are available through your distributor. In addition,

worldwide technical support is available from Aeroqual Ltd. Please contact Aeroqual Technical

Support by email: [email protected]. This will generate an automatic ticket and your enquiry

will be addressed as quickly as possible by a qualified service person.

Aeroqual Limited

109 Valley Road, Mt Eden, Auckland 1024, New Zealand

Phone: +64 9 623 3013

Fax: +64 9 623 3012

Email: [email protected]

12.3. Declarations Copyright Aeroqual Limited. All rights reserved. Reproduction, transfer, distribution or storage of part

or all of the contents of this document in any form without the prior written permission of Aeroqual

Limited is prohibited.

“Aeroqual” is a registered trademark of Aeroqual Limited. Other product and company names

mentioned herein may also be trademarks or trade names.

Aeroqual operates a policy of continuous development. Aeroqual reserves the right to make changes

and improvements to any of the products described in this document without prior notice.

Under no circumstances shall Aeroqual be responsible for any loss of data or income or any special,

incidental, consequential or indirect damages howsoever caused.

The contents of this document are provided "as is". Except as required by applicable law, no

warranties of any kind, either express or implied, including, but not limited to, the implied warranties of

merchantability and fitness for a particular purpose, are made in relation to the accuracy, reliability or

contents of this document.

Aeroqual reserves the right to revise this document or withdraw it at any time without prior notice. The availability of particular products may vary by region. Please check with the Aeroqual dealer nearest to you.

© Aeroqual Limited 2015. All rights reserved.

1. The Aeroqual Dust Profiler complies with EN 61000-6-1:2001 2. The Aeroqual Dust Profiler complies with EN 61000-6-3:2001 3. The Aeroqual Dust Profiler complies with Part 15 of the FCC Rules. Operation is subject to

the following two conditions: (1) these devices may not cause harmful interference, and (2) these devices must accept any interference received, including interference that may cause undesired operation.

Page 44: MRK-D-0021 V1.0 Page | 1 Aeroqual Dust Profiler User Guide(Pneumatics) Pump Module (Communications) Sample Collection Filter (Optional) Din Rail Terminals and Sample air filter (DQ)

MRK-D-0021 V1.0 Page | 44

Aeroqual Dust Profiler User Guide

USE SENSIBLY

Use the Aeroqual DUST PROFILER instrument only as per this user guide.

USE AEROQUAL APPROVED SERVICE

Only approved service personnel must work on this product.

ACCESSORIES

Use only approved accessories. Do not connect incompatible products.

CONNECTING TO OTHER DEVICES

When connecting to any other device, read the appropriate user guide for detailed safety

instructions. Do not connect incompatible products.