M.D. Nov 27th 2002
M0' analysis workshop 1
Monitoring system : lessons from M0’Monitoring system : lessons from M0’
Contents : Monitoring electronics Laser APD electronics Conclusions & requirements for 2003
M.D. Nov 27th 2002
M0' analysis workshop 2
Monitoring system descriptionMonitoring system description
M.D. Nov 27th 2002
M0' analysis workshop 3
MEM stability : Use self measurement of Vinj (stabilized at 10-4 / year)
Stability better than 50 V / 500 mV over 1 month
Monitoring electronicsMonitoring electronics
M.D. Nov 27th 2002
M0' analysis workshop 4
FE-Asic stability : Use electronics injection pulse
Rather good stability over 2 months (1/1000 drift)
Monitoring electronicsMonitoring electronics
2002/08/08
M.D. Nov 27th 2002
M0' analysis workshop 5
PN stability : Compare PN0 with PN2 (level1 fan-out) :
Not stable at all (2.5 % over 2 months) !
Monitoring electronicsMonitoring electronics
2002/08/08
M.D. Nov 27th 2002
M0' analysis workshop 6
PN stability : Look in archives : Accelerated aging
in oven in 200160-80 degrees
PN2 should have never been put on M0’ !Zero quality control on PN diodes at this time.
Monitoring electronicsMonitoring electronics
2001/03/06
PN0 PN2
PN3PN1
M.D. Nov 27th 2002
M0' analysis workshop 7
Energy stability :
Huge variation with time (short term and long term)“Raison d’être” of monitoring systemOK
LaserLaser
M.D. Nov 27th 2002
M0' analysis workshop 8
Pulse width stability :
Huge variation with time (short term and long term)
Huge effect on APD/PN0 Need to be monitored !
LaserLaser
d(f
pp
a)/d
Nanogreen laserParticles
Quantronix lasernominal
Quantronix laser2002
M.D. Nov 27th 2002
M0' analysis workshop 9
Electronics stability : Small dependence on T
No problem with present regulation Dependence with LV :
No significant change in gain up to nowPedestals affected
Regular pedestal runs Dependence with VFE mode :
Pedestal affected Free vs Forced mode Hard to monitor for hidden gains (G9, G5 and G1 in frre
mode) Dependence with laser energy
Pedestals affected Compute dynamic pedestals
Dependence with laser pulse widthConvolution with FPPA shaping time
Need to monitor laser pulse width event by event
APD lineAPD line
M.D. Nov 27th 2002
M0' analysis workshop 10
Electronics stability : Dependence with HV :
A priori big dependence of APD gain with HVNeed to checked stability of APD response with present HV
system On good channels :
VFE electronicsVFE electronics
= 1.4/1000
M.D. Nov 27th 2002
M0' analysis workshop 11
Back to crystal 1142 : Calibration drift by 5-6/1000 in 2 weeks :
Compare with electron runs
APD lineAPD line
Calibration runs @ 120 GeV
M.D. Nov 27th 2002
M0' analysis workshop 12
ConclusionsConclusions
Monitoring system : Need to implement an efficient quality control on PN diodes Electronics very stable but :
Need to clarify pedestal puzzle (fppa 200x) Need to monitor laser pulse width event by event
Requirement for 2003 Residual instabilities seems physical
Parasitic irradiationsHV fluctuation ?Red laser required for next year
To-do :Reprocess irradiation data with all the correction
Should be able to get S/R curves with PN diodes as only reference
Fine study of HV data