8
Life Science Instrumentation Light Sheet Fluorescence Microscope New Generation Alph

Life Science Instrumentation Light Sheet Fluorescence

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Life Science Instrumentation Light Sheet Fluorescence

Life Science Instrumentation

LightSheetFluorescenceMicroscopeNewGenerationAlph

Page 2: Life Science Instrumentation Light Sheet Fluorescence

Alpha3 is a new generation of light sheet fluorescence microscope addressing the needs of high temporal resolution along with spatial high resolution to achieve qualitative and quantitative 3D imaging of fixed or live biological specimens.

From in‐vivo imaging to large cleared samples, the Alpha3 microscope delivers unprecedented image quality while keeping the necessary flexibility and modularity expected for cutting-edge scientific research instruments.

Patented Smart Light Sheet Illuminators 

The light sheet system architecture comprises dual illumination units, each integrating multi-directional light sheet, combined with wide-field microscope detection, offering:

Real-time laser focus sweeping for optimized sharpness over the entire field of view

Removal of stripe artifacts for absorbing or scattering specimens

Ultrafast 3D acquisition with remote focal plane scanning

Flexible imaging system with a broad selection of detection objectives

Modular microscopy system for multimodal optical microscopy: FLIM, FRET, FRAP, photo ablation ...

Cleared mouse adipose tissue vascular network 

Alphα3 features 

Imaging flexibility, from in  vivo imaging to large cleared sample imaging

Macro to micro view imaging, from whole organs at sub-cellular resolution to very small specimens

Compatible with all clearing solutions: aqueous buffers and organic solvents

Multiple mounting accessories to accommodate different sample natures and sizes

Unique live observation using optical eyepieces

Modular Light Sheet Microscope 

Page 3: Life Science Instrumentation Light Sheet Fluorescence

The advanced sharp optical sectioning and smart scanning functions drastically alleviate spatial and temporal resolution constraints for 3D image acquisition in light sheet microscopy.

Sharp Optical Sectioning 

The multidirectional light sheet illuminators perform real time focus sweeping to extend the thinnest focus area over the entire field of view, while improving homogeneity for artefact-free imaging.

 

Alpha3 microscope, with its innovative light sheet illuminators and system architecture, broadens the possibilities of life science imaging, providing new quantitative and qualitative imaging capabilities.

Smart 3D Scanning 

The remote focal plane scanning technology combined with the dynamic illumination of light sheet planes allow for ultrafast Z-stacking and perturbation-free acquisition.

Remote focal plane scanning principle 

Camera 

Detection unit 

Illumination unit Illumination unit 

Tunable lens 

Galvo mirror 

Real time lateral focus sweeping principle 

Camera 

Detection unit 

Illumination unit  Illumination unit 

Tunable lens 

Laser waist 

Rolling shutter 

Cleared mouse brain image without focus sweeping (left) and with active focus sweeping (right) 

Cutting Edge Light Sheet Technology 

Page 4: Life Science Instrumentation Light Sheet Fluorescence

The chamber allows for easy sample insertion and observation with air or immersion objectives. As biological specimens are of different nature and size, multiple sample holders are provided to accommodate a large variety of fixed or live samples.

In Vivo Imaging 

In  vivo imaging may require precise temperature and CO2 control as well as supplies of fresh medium for long term time lapse imaging.

The easily accessible chamber of Alpha3 is pre-equipped to maintain required environmental conditions during specimen imaging. The chamber support is provided with a circulation device that allows temperature control over a range of temperatures (from 15°C to 40°C). Thanks to the QtSPIM software, stacks of image data can be recorded for many hours or days for various applications such as developmental bio processes or stimuli response.

The architecture using a fluorescence microscope stand as a detection unit allows Alpha3 to flexibly adapt to any experiment setting. Complementary imaging capabilities can be easily added for multimodal microscopy.

Alphα3 features 

Multiple accessories provided for sample mounting to accommodate a large variety of specimens

Chamber and sample holders highly resistant to corrosive media or clearing solutions

Low volume chamber minimizes evaporation and use of costly clearing solutions

Easy chamber access, allows addition of various experiment tools

In vivo imaging environmental controls

Flexible Sample Mounting 

Page 5: Life Science Instrumentation Light Sheet Fluorescence

From sample mounting to image acquisition, Alpha3 performs seamless light sheet imaging. The QtSPIM software provides a clear and intuitive interface for collecting X, Y, Z, θ, T, λ images at maximum speed.

Raw image data, along with their metadata, are saved in 16 bits TIFF format (compatible with open source or commercial software for further 3D display and analysis).

QtSPIM controls all image acquisition parameters. When paired with a workstation grade computer in an optimized configuration, the system is able to perform ultra-fast acquisition.

Alphα3 features 

Seamless X,Y,Z,θ,T,λ image data acquisition

Ultra-fast multichannel acquisition via a multi-notch filter, completed by a large selection of emission filters

Wide-field acquisition with sensitive high QE sCMOS camera

Wide choice of detection objectives: LWD 2X to 60X air, dipping lenses, clearing objectives with correction collar for RI matching

QtSpim user interface 

QtSPIM 

Acquisition Software 

Page 6: Life Science Instrumentation Light Sheet Fluorescence

Typical applications of Alpha3 include in toto imaging of small animal models such as whole mouse embryos, morphogenesis and embryogenesis of model organisms: C.elegans, Drosophila, Zebra fish, live imaging of cell cultures, functional imaging of neuronal activity, fluorescence imaging of marine organisms or plant developmental biology.

 

a) In vivo Arabidopsis leaf, false colours membrane staining at 10x; b) Mouse embryo, iDisco clearing; c) Cleared Mouse Embryo, GFP staining at 10x; d) Fixed Stentor, GFP cilia and DAPI poly‐nuclei at 10x; e) Cleared Zebrafish larvae head, vascular and neural staining at 20x;  f) Cubic 2 Mouse brain, TagRFP neurones at 20x; g) Fixed drosophila egg, 3  channels at 20x; h) iDisco Cleared Mouse  Embryo, motor  neurones  and  extensions  in  body  at  2x;  i) Arabidopsis meristeme, GFP  in microtubules  at  40x;  j) Cubic Cleared Zebrafish, nervous system and muscle staining, raw 6 tiles at 20x. 

a  c 

d  e 

g  h 

Applications 

From In Vivo Imaging to Clearing 

Page 7: Life Science Instrumentation Light Sheet Fluorescence

Specifications

2 Impasse de la Noisette91370 Verrieres Le Buisson, FrancePhone: +33 9 54 03 05 43Email: [email protected]: www.phaseview.com

Laser source Laser combiner with up to 4 laser lines: fibered lasers CW / Laser diodes or DPSSWavelength selection from 405 nm to 785nm, output power from 25 to 250mW

Light sheet Dual smart illuminators with fibered connection to laser combinerMulti-directional light sheet with real time focus sweepingChromatic correction 400 nm - 632 nmMinimum light sheet thickness 2 µm, range: 2 µm - 12µm, width: 2 mm - 15 mm

Chamber & sample holders Chamber dimensions : width 21 mm, length 70 mm, height 25 mm; volume < 15 mlChamber highly resistant to various corrosive media, clearing agents, sea water, etc.Sample size from µm to cm rangeMultiple holders for sample mounting: molds, coverslips, glass supportsOptional: temperature and CO2 controls

Volume scanning Motorized Z-stage: range 15 mm, precision 0.1 µm, acquisition speed 40 fpsOptional: motorized XY-stages for tiling: range 15 mm, precision 0.1 µmOptional: ultra fast 3D scanning module (75 images/second)

Detection unit Fluorescence microscope stand comprising 2-position objective slider, eyepieces,video port, motorized filter wheel and multi-notch filter

Detection objectives Large selection of long working distance objectives: air, dipping lenses,clearing objectives with correction collar for RI 1.33 - 1.56 Magnification: from 2x to 60x

Image sensor sCMOS: 2048 x 2048 pixels, format: 13 mm x 13 mm, size: 6.5 µm x 6.5 µm,USB 3.0 / CameraLink Interface

Software QtSPIM software for Z-stacking, XY-tiling and time lapse acquisition,providing easy export of raw images and metadata to open sourceor commercial 3rd party software

PC configuration Desktop i7-7700K 4.2Ghz - 2 x 16Go RAM - GPU GeForce GTX1060-6Go,Hard Disk 4 x 4 To RAID0 - Hard Disk SSD 500Go, QHD Screen 31.5",Windows 10 / 64 bit professional

Dimensions & weight Microscope breadboard format 600 mm W x 600 mm L x 750 H mm, 27 kg

Page 8: Life Science Instrumentation Light Sheet Fluorescence

2 Impasse de la Noisette91370 Verrieres Le Buisson, FrancePhone: +33 9 54 03 05 43Email: [email protected]: www.phaseview.com

LightSheetFluorescenceMicroscopeNewGenerationAlph