LAPORAN
PROGRAM UJI BANDING (PROFICIENCY TEST) ANTAR LABORATORIUM
SKEMA KHUSUS SEMESTER II-2016
BAHAN UJI:
Air Limbah III Batch 3 (ALDS III-3)
Ref. No. 006/UR/BMD/2016
PARAMETER UJI:
Zat Padat Tersuspensi (TSS), Kebutuhan Oksigen Kimiawi (COD), Lemak & Minyak,
dan Derajat Keasaman (pH, 25oC)
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Page 1 of 21
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Ref. No. 006/UR/BMD/2016
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]
www.ujiprofisiensi.com
KATA PENGANTAR
Puji syukur kiranya kita panjatkan kepada Alloh SWT yang berkenan melimpahkan karunia-Nya
sehingga ‘Program Uji Banding Antar Laboratorium (Proficiency Test) Skema Khusus Semester
II-2016’ untuk bahan uji Air Limbah III Batch 3 (ALDS III-3) dengan parameter uji zat padat
tersuspensi (TSS), kebutuhan oksigen kimiawi (COD), lemak & minyak, dan derajat keasaman
(pH, 25oC) telah berhasil dilaksanakan dan diselesaikan sampai dengan diterbitkannya laporan
ini.
Program uji banding antar laboratorium merupakan salah satu instrumen yang sangat penting
sebagai sarana jaminan mutu eksternal (external quality control) bagi laboratorium pengujian
dan kalibrasi sesuai dengan persyaratan standard SNI ISO/IEC 17025:2008. Dengan mengikuti
program uji banding, laboratorium dapat melakukan evaluasi kinerja pengujian yang dilakukan
dalam rangka melakukan peningkatan secara berkesinambungan (continuous improvement)
untuk memberikan layanan terbaik kepada pelanggan.
Program skema khusus kali ini diikuti oleh 9 (sembilan) laboratorium dengan 5 (lima)
diantaranya telah terakreditasi SNI ISO/IEC 17025:2008 oleh Komite Akreditas Nasional (KAN).
Semua parameter uji dapat diolah dengan baik menggunakan Robust Statistic Algoritma A
Mean sesuai ISO 13528: 2015(E). Dalam program uji banding ini masih digunakan nilai
konsensus (consensus value) dari laboratorium peserta dalam melakukan evaluasi, sehingga
tidak memuaskan dalam ketertelusuran (traceability) secara metrologi. Akan tetapi BMD
Laboratory meyakini bahwa hal tersebut tidak mengurangi manfaat dari pelaksanaan program
uji banding ini.
BMD Laboratory sebagai provider uji banding antar laboratorium menyampaikan apresiasi
setinggi-tingginya kepada seluruh pihak yang telah membantu dan menyukseskan kegiatan ini.
Khususnya kepada seluruh laboratorium peserta atas partisipasi aktifnya mengikuti program uji
Page 2 of 21
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Ref. No. 006/UR/BMD/2016
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]
www.ujiprofisiensi.com
banding Air Limbah III Batch 3 (ALDS III-3). Ucapan terimakasih juga disampaikan kepada
tim ahli yang telah bekerja keras mengolah data dan menyusun laporan ini. Akhirnya,
BMD Laboratory menyampaikan permohonan maaf sekiranya terdapat kekurangan dalam
penyelenggaraan program uji banding antar laboratorium. Kritik, saran, dan masukan yang
membangun sangat kami harapkan untuk menghadirkan program uji banding yang semakin
baik ke depannya. Semoga laporan ini dapat dimanfaatkan oleh laboratorium peserta dan
pihak-pihak terkait.
Tangerang Selatan, 26 Oktober 2016
Ir. Renaldo Moontri
CEO BMD Group
(Selaku Manajer Puncak BMD Laboratory)
Page 3 of 21
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Ref. No. 006/UR/BMD/2016
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]
www.ujiprofisiensi.com
DAFTAR ISI
KATA PENGANTAR ....................................................................................................................... 1
DAFTAR ISI ................................................................................................................................... 3
I. PENDAHULUAN .................................................................................................................... 4
II. PELAKSANAAN ..................................................................................................................... 4
III. STATISTIKA PENGOLAHAN DATA ......................................................................................... 5
III.1 Penilaian Laboratorium ............................................................................................... 5
III.2 Penetapan Nilai Evaluasi Uji Banding (xpt) .................................................................. 6
III.3 Penetapan Simpangan Baku untuk Evaluasi (σpt) ....................................................... 6
IV. HASIL UJI BANDING ANTAR LABORATORIUM ...................................................................... 6
IV.1 Parameter Zat Padat Tersuspensi (TSS) ...................................................................... 9
IV.2 Parameter Kebutuhan Oksigen Kimiawi (COD) ........................................................ 10
IV.3 Parameter Lemak dan Minyak .................................................................................. 12
IV.4 Parameter Derajat Keasaman (pH, 25oC) ................................................................. 14
V. PENUTUP ........................................................................................................................... 15
VI. REFERENSI .......................................................................................................................... 16
Lampiran A Rekapitulasi Hasil z-score Program Uji Banding .................................................... 17
Lampiran B Rekapitulasi Bias Nilai Peserta Terhadap Nilai Konsensus .................................... 18
Lampiran C Evaluasi Distribusi Normal Data Program Uji Banding .......................................... 20
Page 4 of 21
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Ref. No. 006/UR/BMD/2016
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]
www.ujiprofisiensi.com
I. PENDAHULUAN
Air limbah industri diatur persyaratan baku mutunya sesuai Keputusan Menteri Negara
Lingkungan Hidup KEP-51/MENLH/10/1995. Persyaratan baku mutu untuk parameter zat
padat tersuspensi (TSS), kebutuhan oksigen kimiawi (COD), lemak dan minyak dapat dilihat
seperti pada Tabel 1.
Tabel 1. Persyaratan baku mutu air limbah industri[1]
No. Parameter Satuan Persyaratan (Gol. II)
1. TSS mg/L Maks. 400
2. COD mg/L Maks. 300
3. Lemak dan Minyak mg/L Maks. 60
Salah satu jaminan mutu bagi laboratorium pengujian dan kalibrasi sesuai persyaratan
dalam SNI ISO/IEC 17025:2008 adalah partisipasi dalam uji banding antar laboratorium atau
uji profisiensi[2] sebagai jaminan mutu eksternal (external quality control). Untuk itu BMD
Laboratory menyelenggarakan program uji banding antar laboratorium (proficiency test)
untuk bahan uji Air Limbah III Batch 3 (ALDS III-3) dengan tujuan sebagai berikut:
1. Sebagai sarana jaminan mutu eksternal (external quality control) bagi laboratorium
peserta khususnya laboratorium yang menerapkan standard SNI ISO/IEC 17025:2008.
2. Membantu laboratorium dalam mengevaluasi unjuk kerja pengujian kadar zat padat
tersuspensi (TSS), kebutuhan oksigen kimiawi (COD), lemak & minyak, dan derajat
keasaman (pH, 25oC) untuk bahan uji air limbah.
II. PELAKSANAAN
Berikut adalah jadual pelaksanaan program uji banding antar laboratorium (proficiency test)
yang diselenggarakan oleh BMD Laboratory untuk bahan uji Air Limbah III Batch 3 (ALDS III-
3) untuk parameter uji zat padat tersuspensi (TSS), kebutuhan oksigen kimiawi (COD), lemak
& minyak, dan derajat keasaman (pH, 25oC):
Page 5 of 21
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Ref. No. 006/UR/BMD/2016
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]
www.ujiprofisiensi.com
a. Penutupan pendaftaran: 28 Juli 2016
b. Pendistribusian bahan uji: 11 s.d. 25 Agustus 2016
c. Pelaporan hasil analisis oleh laboratorium peserta: Maks. 08 September 2016
d. Pendistribusian bahan uji khusus untuk parameter lemak dan minyak*): 15 s.d. 29
September 2016
e. Pelaporan hasil analisis oleh laboratorium peserta (khusus untuk parameter lemak dan
minyak)*): Maks. 13 Oktober 2016
f. Pendistribusian laporan program uji banding oleh BMD Laboratory: Maks. 27 Oktober
2016
*)Khusus untuk parameter lemak dan minyak dilakukan pendistribusian ulang bahan uji
dikarenakan terdapat kesalahan preparasi bahan uji oleh beberapa laboratorium
peserta.
III. STATISTIKA PENGOLAHAN DATA
III.1 Penilaian Laboratorium
Penilaian terhadap unjuk kerja laboratorium menggunakan z-score yang dihitung
berdasarkan persamaan di bawah ini:
( - )i pt
i
pt
x xz
σ
=
zi = z-score laboratorium peserta
xi = nilai yang dilaporkan peserta
xpt = nilai evaluasi uji banding
σpt = simpangan baku untuk evaluasi
Interpretasi dari z-score (sesuai ISO/IEC 17043:2010, B.4.1.1)[3] adalah sebagai berikut:
� Untuk |z-score|≤2,0 dikategorikan ‘diterima’ (acceptable) diberi lambang ‘OK’
Page 6 of 21
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Ref. No. 006/UR/BMD/2016
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]
www.ujiprofisiensi.com
� Untuk 2,0<|z-score|<3,0 dikategorikan ‘diperingatkan’ (warning signal) diberi
lambang ‘$’
� Untuk |z-score|≥3,0 dikategorikan ‘tidak diterima’ (unacceptable) diberi lambang
‘$$’
III.2 Penetapan Nilai Evaluasi Uji Banding (xpt)
Penetapan nilai evaluasi uji banding menggunakan dua pendekatan statistika yaitu:
� Robust Statistic Algorithm A sesuai ISO 13528:2015(E) Annex C[4]. Nilai tersebut
merupakan nilai konsensus dari peserta (consensus value)
� Seleksi data menggunakan perangkat lunak Distribution AnalyzerTM, Taylor
Enterprises, Inc.[5]
III.3 Penetapan Simpangan Baku untuk Evaluasi (σpt)
Simpangan baku yang digunakan dalam evaluasi (standard deviation for proficiency
assessment) ditetapkan menggunakan tiga pendekatan statistika yaitu:
� Robust Statistic Algorithm A sesuai ISO 13528:2015(E) Annex C atau SD Robust
� Simpangan Baku Horwitz atau SD Horwitz[6]
� Simpangan baku yang ditetapkan (Assigned σpt)
IV. HASIL UJI BANDING ANTAR LABORATORIUM
Sebanyak 9 (sembilan) laboratorium telah berpartisipasi dalam program uji banding antar
laboratorium (proficiency test) yang diselenggarakan oleh BMD Laboratory untuk bahan uji
Air Limbah III Batch 3 (ALDS III-3) dengan parameter uji zat padat tersuspensi (TSS),
kebutuhan oksigen kimiawai (COD), lemak & minyak, dan derajat keasaman (pH, 25oC).
Adapun 5 (lima) laboratorium diantaranya merupakan laboratorium terakreditasi SNI
ISO/IEC 17025:2008 oleh Komite Akreditasi Nasional (KAN). Kode laboratorium peserta dan
status akreditasi dapat dilihat seperti pada Tabel 2.
Page 7 of 21
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Ref. No. 006/UR/BMD/2016
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]
www.ujiprofisiensi.com
Tabel 2. Kode Laboratorium Peserta dan Status Akreditasi
Rekapitulasi hasil pengujian oleh laboratorium peserta disajikan seperti pada Tabel 3. Data
hasil uji banding dengan parameter uji zat padat tersuspensi (TSS), kebutuhan oksigen
kimiawai (COD), lemak & minyak dan derajat keasaman (pH, 25oC) dapat diolah dengan baik
menggunakan Robust Statistic Algorithm A sesuai ISO 13528:2015(E). Selain itu, untuk
parameter uji zat padat tersuspensi (TSS), kebutuhan oksigen kimiawai (COD) dan derajat
keasaman (pH, 25oC) juga dilakukan evaluasi data menggunakan perangkat lunak
Distribution AnalyzerTM guna memastikan bahwa data yang diolah mengikuti pola distribusi
normal pada tingkat kepercayaan 95%. Adapun untuk parameter uji lemak dan minyak,
evaluasi distribusi normal tidak dapat dilakukan karena hanya terdapat 7 (tujuh) laporan
hasil analisis, dimana untuk evaluasi distribusi normal menggunakan perangkat lunak
Distribution AnalyzerTM diperlukan minimal 8 (delapan) data. Nilai evaluasi uji banding
(assigned value) merupakan nilai konsensus (consensus value) dari laboratorium peserta
No. Kode Laboratorium Peserta Status Akreditasi
1. BMD-0801-16B3 Tidak
2. BMD-0802-16B3 Ya
3. BMD-0803-16B3 Ya
4. BMD-0804-16B3 Tidak
5. BMD-0805-16B3 Ya
6. BMD-0806-16B3 Tidak
7. BMD-0807-16B3 Tidak
8. BMD-0808-16B3 Ya
9. BMD-0809-16B3 Ya
Page 8 of 21
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Ref. No. 006/UR/BMD/2016
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]
www.ujiprofisiensi.com
sehingga nilai tersebut bukan merupakan nilai acuan yang tertelusur secara metrologi ke
sistem satuan internasional (SI)
Tabel 3. Rekapitulasi Hasil Pengujian Laboratorium Peserta
No. Kode Lab. TSS COD Lemak & Minyak pH (25
oC)
mg.L-1
mg.L-1
mg.L-1
-
1. BMD-0801-16B3 61.0±2.0 73.6±3.2 254.33 7.88±0.047
Metoda SNI 06-6989.3-
2004 SNI 6989.73-2009 Gravimetri
SNI 06-6989.11-2009
2.
BMD-0802-16B3 124.00 73.32 171.7 7.65
Metoda APHA 22
nd
(2012) : 2540-Solid D
APHA 22nd
(2012) : 5220 C
APHA 22nd
(2012) : 5220-O&G-B
APHA 22nd
(2012) : 4500-
H+B
3. BMD-0803-16B3 73.00±0.3607 106.06±12.592 186.60±0.0591 7.91±0.0415
Metoda SNI 06-6989.3-
2004 SNI 6989.2-2009
SNI 06-6989.10-2011
SNI 06-6989.23-2005
4. BMD-0804-16B3 76 69.88 378.40 7.99
Metoda SNI 06-6989.3-
2004 SNI 06-6989.73-
2009 SNI 6989.10-2011
SNI 06-6989.11-2004
5. BMD-0805-16B3 73±0.7 110.52 70±0.1 7.27
Metoda SNI 06-6989.3-
2004 SNI 6989.2-2009 SNI 6989.10-2011
SNI 06-6989.11-2004
6. BMD-0806-16B3 93.35 - - 7.8285
Metoda SNI 06-6989.3-
2004 - -
SNI 06-6989.11-2004
7. BMD-0807-16B3 65.833 133.714 553.5 7.6
Metoda SNI 06-6989.3-
2004 SNI 6989.2-2009
SNI 06-6989.10-2004
SNI 06-6989.11-2004
8. BMD-0808-16B3 88.00±10.22 59.25±0.04 - 7.24
Metoda SNI 06-6989.3-
2004 APHA 21
st (2005) :
5220 -
SNI 06-6989.11-2004
9. BMD-0809-16B3 64±0.20 74.44±1.78 278±11.71 7.03±0.29
Metoda SNI 06-6989.3-
2004 SNI 06-6989.15-
2004 SNI 6989.10-2011
SNI 06-6989.11-2004
Page 9 of 21
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Ref. No. 006/UR/BMD/2016
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]
www.ujiprofisiensi.com
IV.1 Parameter Zat Padat Tersuspensi (TSS)
Sebanyak 9 (sembilan) laporan hasil pengujian untuk parameter zat padat tersuspensi (TSS)
diterima oleh BMD Laboratory dan 5 (lima) diantaranya disertai dengan laporan nilai
estimasi ketidakpastian diperluas dengan k=2 pada tingkat kepercayaan 95%. Dari hasil
pengolahan data secara statistika diperoleh sebanyak 6 (enam) data yang dilaporkan
laboratorium masuk dalam kategori ‘diterima’ (|z-score|≤2,0), 2 (dua) data yang dilaporkan
laboratorium masuk dalam kategori ‘diperingatkan’ (2,0<|z-score|<3,0), dan 1 (satu) data
yang dilaporkan laboratorium masuk dalam kategori ‘tidak diterima’ (|z-score|≥3,0). Hasil
pengolahan data disajikan pada Tabel 4 dan Gambar 1.
Tabel 4. Hasil Pengolahan Data Parameter Zat Padat Tersuspensi (TSS)
No. Kode Peserta Nilai Laporan
Peserta (mg/L)
U, k=2
(mg/L)
Nilai Evaluasi Uji
Banding (mg/L)
σpt (mg/L)
z-score Kategori
1. BMD-0801-16B3 61.0 2.0 76.32 6.36 -2.41 $
2. BMD-0802-16B3 124.00 - 76.32 6.36 7.50 $$
3. BMD-0803-16B3 73.00 0.3607 76.32 6.36 -0.52 OK
4. BMD-0804-16B3 76 - 76.32 6.36 -0.05 OK
5. BMD-0805-16B3 73 0.7 76.32 6.36 -0.52 OK
6. BMD-0806-16B3 93.35 - 76.32 6.36 2.68 $
7. BMD-0807-16B3 65.833 - 76.32 6.36 -1.65 OK
8. BMD-0808-16B3 88.00 10.22 76.32 6.36 1.84 OK
9. BMD-0809-16B3 64 0.20 76.32 6.36 -1.94 OK
Keterangan:
Nilai evaluasi uji banding (xpt) merupakan nilai konsensus (consensus value) laboratorium
peserta yang diperoleh dari pengolahan data secara statistika menggunakan Robust
Statistic Algorithm A sesuai ISO 13528:2015(E) dengan sebelumnya dilakukan analisis
menggunakan perangkat lunak Distribution AnalyzerTM (lihat Lampiran C). σpt adalah
Page 10 of 21
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Ref. No. 006/UR/BMD/2016
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]
www.ujiprofisiensi.com
40.00
45.00
50.00
55.00
60.00
65.00
70.00
75.00
80.00
85.00
90.00
95.00
100.00
105.00
110.00
115.00
120.00
125.00
BM
D-0
801-
16B
3
BM
D-0
809-
16B
3
BM
D-0
807-
16B
3
BM
D-0
803-
16B
3
BM
D-0
805-
16B
3
BM
D-0
804-
16B
3
BM
D-0
808-
16B
3
BM
D-0
806-
16B
3
BM
D-0
802-
16B
3
TS
S (m
g/L
)
Lab. Code
Cons. value = 76.32 mg/LAlgorithm A - ISO 13528:2015(E) Annex C
σpt = 6.36 mg/L (SD Horwitz)
SD Robust = 13.96 mg/L
©BMD Laboratory
Cons. value (X)
X+3σpt
X+2σpt
X-2σpt
X-3σpt
simpangan baku yang diperoleh dari persamaan Horwitz (SD Horwitz). U yaitu
ketidakpastian diperluas dengan faktor cakupan (k)=2 pada tingkat kepercayaan 95%.
Gambar 1. Grafik Hasil Uji Banding Parameter Zat Padat Tersuspensi (TSS)
IV.2 Parameter Kebutuhan Oksigen Kimiawi (COD)
Sebanyak 8 (delapan) laporan hasil pengujian untuk parameter kebutuhan oksigen kimiawi
(COD) diterima oleh BMD Laboratory dan 4 (empat) diantaranya disertai dengan laporan
nilai estimasi ketidakpastian diperluas dengan k=2 pada tingkat kepercayaan 95%. Dari hasil
pengolahan data secara statistika diperoleh sebanyak 5 (lima) data yang dilaporkan
laboratorium masuk dalam kategori ‘diterima’ (|z-score|≤2,0) dan 3 (tiga) data yang
dilaporkan laboratorium masuk dalam kategori ‘tidak diterima’ (|z-score|≥3,0). Hasil
pengolahan data disajikan pada Tabel 5 dan Gambar 2.
Page 11 of 21
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Ref. No. 006/UR/BMD/2016
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]
www.ujiprofisiensi.com
Tabel 5. Hasil Pengolahan Data Parameter Kebutuhan Oksigen Kimiawi (COD)
No. Kode Peserta Nilai Laporan
Peserta (mg/L)
U, k=2
(mg/L)
Nilai Evaluasi Uji
Banding (mg/L)
σpt
(mg/L) z-score Kategori
1. BMD-0801-16B3 73.6 3.2 70.10 5.92 0.59 OK
2. BMD-0802-16B3 73.32 - 70.10 5.92 0.54 OK
3. BMD-0803-16B3 106.06 12.592 70.10 5.92 6.08 $$
4. BMD-0804-16B3 69.88 - 70.10 5.92 -0.04 OK
5. BMD-0805-16B3 110.52 - 70.10 5.92 6.83 $$
6. BMD-0807-16B3 133.714 - 70.10 5.92 10.75 $$
7. BMD-0808-16B3 59.25 0.04 70.10 5.92 -1.83 OK
8. BMD-0809-16B3 74.44 1.78 70.10 5.92 0.73 OK
Keterangan:
Nilai evaluasi uji banding (xpt) merupakan nilai konsensus (consensus value) laboratorium
peserta yang diperoleh dari pengolahan data secara statistika menggunakan Robust
Statistic Algorithm A sesuai ISO 13528:2015(E) dengan sebelumnya dilakukan analisis
menggunakan perangkat lunak Distribution AnalyzerTM (lihat Lampiran C). σpt adalah
simpangan baku yang diperoleh dari persamaan Horwitz (SD Horwitz). U yaitu
ketidakpastian diperluas dengan faktor cakupan (k)=2 pada tingkat kepercayaan 95%.
Page 12 of 21
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Ref. No. 006/UR/BMD/2016
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]
www.ujiprofisiensi.com
20.00
25.00
30.00
35.00
40.00
45.00
50.00
55.00
60.00
65.00
70.00
75.00
80.00
85.00
90.00
95.00
100.00
105.00
110.00
115.00
120.00
BM
D-0
808-
16B
3
BM
D-0
804-
16B
3
BM
D-0
802-
16B
3
BM
D-0
801-
16B
3
BM
D-0
809-
16B
3
BM
D-0
803-
16B
3
BM
D-0
805-
16B
3
BM
D-0
807-
16B
3
CO
D (m
g/L
)
Lab. Code
Cons. value = 70.10 mg/L
Algorithm A - ISO 13528:2015(E) Annex C
σpt = 5.92 mg/L (SD Horwitz)
SD Robust = 15.61 mg/L
©BMD Laboratory
Cons. value (X)
X+3σpt
X+2σpt
X-2σpt
X-3σpt
Gambar 2. Grafik Hasil Uji Banding Parameter Kebutuhan Oksigen Kimiawi (COD)
IV.3 Parameter Lemak dan Minyak
Sebanyak 7 (tujuh) laporan hasil pengujian untuk parameter lemak dan minyak diterima
oleh BMD Laboratory dan 2 (dua) diantaranya disertai dengan laporan nilai estimasi
ketidakpastian diperluas dengan k=2 pada tingkat kepercayaan 95%. Dari hasil pengolahan
data secara statistika diperoleh sebanyak 4 (empat) data yang dilaporkan laboratorium
masuk dalam kategori ‘diterima’ (|z-score|≤2,0) dan 3 (tiga) data yang dilaporkan
laboratorium masuk dalam kategori ‘tidak diterima’ (|z-score|≥3,0). Hasil pengolahan data
disajikan pada Tabel 6 dan Gambar 3.
Page 13 of 21
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Ref. No. 006/UR/BMD/2016
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]
www.ujiprofisiensi.com
Tabel 6. Hasil Pengolahan Data Parameter Lemak dan Minyak
No. Kode Peserta Nilai Laporan
Peserta (mg/L)
U, k=2
(mg/L)
Nilai Evaluasi Uji
Banding (mg/L)
σpt
(mg/L) z-score Kategori
1. BMD-0801-16B3 254.33 - 221.93 28.85 1.12 OK
2. BMD-0802-16B3 171.7 - 221.93 28.85 -1.74 OK
3. BMD-0803-16B3 186.60 - 221.93 28.85 -1.22 OK
4. BMD-0804-16B3 378.40 - 221.93 28.85 5.42 $$
5. BMD-0805-16B3 70 0.10 221.93 28.85 -5.27 $$
6. BMD-0807-16B3 553.5 - 221.93 28.85 11.49 $$
7. BMD-0809-16B3 278 11.71 221.93 28.85 1.94 OK
Keterangan:
Nilai evaluasi uji banding (xpt) merupakan nilai konsensus (consensus value) laboratorium
peserta yang diperoleh dari pengolahan data secara statistika menggunakan Robust
Statistic Algorithm A sesuai ISO 13528:2015(E). σpt adalah simpangan baku yang ditetapkan
sebesar 13% dari nilai evaluasi uji banding (xpt). Kriteria penetapan 13% ini berdasarkan dari
pengalaman penyelenggaraan uji banding sebelumnya untuk parameter lemak dan minyak.
Berdasarkan pengalaman penyelenggaraan uji banding parameter lemak dan minyak, BMD
Laboratory menetapkan σpt terbesar yang dapat digunakan untuk evaluasi peserta adalah
20% dari nilai evaluasi uji banding (xpt). U yaitu ketidakpastian diperluas dengan faktor
cakupan (k)=2 pada tingkat kepercayaan 95%.
Page 14 of 21
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Ref. No. 006/UR/BMD/2016
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]
www.ujiprofisiensi.com
10.00
60.00
110.00
160.00
210.00
260.00
310.00
360.00
410.00
460.00
510.00
560.00
BM
D-0
805-
16B
3
BM
D-0
802-
16B
3
BM
D-0
803-
16B
3
BM
D-0
801-
16B
3
BM
D-0
809-
16B
3
BM
D-0
804-
16B
3
BM
D-0
807-
16B
3
Le
ma
k&
Min
ya
k (m
g/L
)
Lab. Code ©BMD Laboratory
Cons. value = 221.93 mg/LAlgorithm A - ISO 13528:2015(E) Annex C
σpt = 28.85 mg/L (13% Cons. value)
SD Horwitz = 15.75 mg/LSD Robust = 96.13 mg/L
Cons. value (X)
X+3σptX+2σpt
X-2σpt
X-3σpt
Gambar 3. Grafik Hasil Uji Banding Parameter Lemak dan Minyak
IV.4 Parameter Derajat Keasaman (pH, 25oC)
Sebanyak 9 (sembilan) laporan hasil pengujian untuk parameter derajat keasaman (pH,
25oC) diterima oleh BMD Laboratory dan 3 (tiga) diantaranya disertai dengan laporan nilai
estimasi ketidakpastian diperluas dengan k=2 pada tingkat kepercayaan 95%. Dari hasil
pengolahan data secara statistika diperoleh sebanyak 9 (sembilan) data yang dilaporkan
laboratorium masuk dalam kategori ‘diterima’ (|z-score|≤2,0). Hasil pengolahan data
disajikan pada Tabel 7 dan Gambar 4.
Page 15 of 21
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Ref. No. 006/UR/BMD/2016
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]
www.ujiprofisiensi.com
Tabel 7. Hasil Pengolahan Data Parameter Derajat Keasaman (pH, 25oC)
No. Kode Peserta Nilai Laporan
Peserta U, k=2
Nilai Evaluasi Uji
Banding σpt z-score Kategori
1. BMD-0801-16B3 7.88 0.047 7.60 0.38 0.72 OK
2. BMD-0802-16B3 7.65 - 7.60 0.38 0.12 OK
3. BMD-0803-16B3 7.91 0.0415 7.60 0.38 0.80 OK
4. BMD-0804-16B3 7.99 - 7.60 0.38 1.01 OK
5. BMD-0805-16B3 7.27 - 7.60 0.38 -0.88 OK
6. BMD-0806-16B3 7.8285 - 7.60 0.38 0.59 OK
7. BMD-0807-16B3 7.6 - 7.60 0.38 -0.01 OK
8. BMD-0808-16B3 7.24 - 7.60 0.38 -0.96 OK
9. BMD-0809-16B3 7.03 0.29 7.60 0.38 -1.51 OK
Keterangan:
Nilai evaluasi uji banding (xpt) merupakan nilai konsensus (consensus value) laboratorium
peserta yang diperoleh dari pengolahan data secara statistika menggunakan Robust
Statistic Algoritma A sesuai ISO 13528:2015(E) dengan sebelumnya dilakukan analisis
menggunakan perangkat lunak Distribution AnalyzerTM (lihat Lampiran C). Adapun σpt
adalah simpangan baku yang diperoleh dari Robust Statistic Algorithm A sesuai ISO
13528:2015(E) atau SD Robust. U yaitu ketidakpastian diperluas dengan faktor cakupan
(k)=2 pada tingkat kepercayaan 95%.
V. PENUTUP
Program uji banding antar laboratorium merupakan salah satu sarana jaminan mutu
eksternal bagi laboratorium pengujian dan kalibrasi. Hasil dari program uji banding ini
diharapkan dapat digunakan sebagai sarana evaluasi bagi laboratorium untuk meningkatkan
kinerja dalam kegiatan pengujian yang dilakukan.
Page 16 of 21
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Ref. No. 006/UR/BMD/2016
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]
www.ujiprofisiensi.com
5.00
5.50
6.00
6.50
7.00
7.50
8.00
8.50
9.00
9.50
10.00
BM
D-0
809-
16B
3
BM
D-0
808-
16B
3
BM
D-0
805-
16B
3
BM
D-0
807-
16B
3
BM
D-0
802-
16B
3
BM
D-0
806-
16B
3
BM
D-0
801-
16B
3
BM
D-0
803-
16B
3
BM
D-0
804-
16B
3
pH
(25
oC
)
Cons. value = 7.60Algorithm A - ISO 13528:2015(E) Annex C
σpt = 0.38 (SD Robust)
©BMD Laboratory
Cons. value (X)
X+3σpt
X+2σpt
X-2σpt
X-3σpt
Gambar 4. Grafik Hasil Uji Banding Parameter Derajat Keasaman (pH, 25oC)
VI. REFERENSI
[1] Keputusan Menteri Negara Lingkungan Hidup Nomor: KEP-51/MENLH/10/1995
[2] SNI ISO/IEC 17025:2008
[3] ISO/IEC 17043:2010
[4] ISO 13528:2005(E)
[5] Distribution AnalyzerTM, Copyright © by Taylor Enterprises, Inc. USA
[6] M. Thompson (2004) The Amazing Horwitz Function, AMC Technical Brief No.17, Royal
Society of Chemistry
Page 17 of 21
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Ref. No. 006/UR/BMD/2016
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]
www.ujiprofisiensi.com
Lampiran A
Rekapitulasi Hasil z-score Program Uji Banding
No. Kode Peserta z-score
TSS COD Lemak&Minyak pH (25oC)
1. BMD-0801-16B3 -2.41 $ 0.59 OK 1.12 OK 0.72 OK
2. BMD-0802-16B3 7.50 $$ 0.54 OK -1.74 OK 0.12 OK
3. BMD-0803-16B3 -0.52 OK 6.08 $$ -1.22 OK 0.80 OK
4. BMD-0804-16B3 -0.05 OK -0.04 OK 5.42 $$ 1.01 OK
5. BMD-0805-16B3 -0.52 OK 6.83 $$ -5.27 $$ -0.88 OK
6. BMD-0806-16B3 2.68 $ - - - - 0.59 OK
7. BMD-0807-16B3 -1.65 OK 10.75 $$ 11.49 $$ -0.01 OK
8. BMD-0808-16B3 1.84 OK -1.83 OK - - -0.96 OK
9. BMD-0809-16B3 -1.94 OK 0.73 OK 1.94 OK -1.51 OK
Keterangan:
OK : Kategori ‘diterima’atau acceptable
$ : Kategori ‘diperingatkan’ atau warning signal
$$ : Kategori ‘tidak diterima’ atau unacceptable
Page 18 of 21
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Ref. No. 006/UR/BMD/2016
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]
www.ujiprofisiensi.com
Lampiran B
Rekapitulasi Bias Nilai Peserta Terhadap Nilai Konsensus
-20.00
-10.00
0.00
10.00
20.00
30.00
40.00
50.00
60.00
b (
mg
/L)
Lab. Code
Parameter Zat Padat Tersuspensi (TSS)
-20.00
-10.00
0.00
10.00
20.00
30.00
40.00
50.00
60.00
70.00
b (
mg
/L)
Lab. Code
Parameter Kebutuhan Oksigen Kimiawi (COD)
Page 19 of 21
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Ref. No. 006/UR/BMD/2016
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]
www.ujiprofisiensi.com
-200.00
-100.00
0.00
100.00
200.00
300.00
400.00
b (
mg
/L)
Lab. Code
Parameter Lemak dan Minyak
-0.80
-0.60
-0.40
-0.20
0.00
0.20
0.40
0.60
b
Lab. Code
Parameter Derajat Keasaman (pH, 25oC)
Page 20 of 21
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Ref. No. 006/UR/BMD/2016
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]
www.ujiprofisiensi.com
Sample Size = 8Average = 74.272Standard Deviation = 11.412Skewness = 0.71Excess Kurtosis = -0.55
Test of Fit: p-value = 0.5953 (SK All) Decision = Pass (SK Spec) Decision = No specs
With 95% confidence more than 99% of the values are between 18.4619 and 130.0831
61.0000 93.350077.1750
No Kode PesertaNilai Laporan
Peserta (mg/L)
1 BMD-0801-16B3 61.02 BMD-0803-16B3 73.003 BMD-0804-16B3 764 BMD-0805-16B3 735 BMD-0806-16B3 93.356 BMD-0807-16B3 65.8337 BMD-0808-16B3 88.008 BMD-0809-16B3 64
Lampiran C
Evaluasi Distribusi Normal Data Program Uji Banding
BMD-0802-16B3 was rejected
Normal Distribution Analysis
(Parameter TSS)
Powered by:
Distribution AnalyzerTM
(licensed software)
©Taylor Enterprises, Inc.
No Kode PesertaNilai Laporan
Peserta (mg/L)
1 BMD-0801-16B3 61.02 BMD-0802-16B3 124.003 BMD-0803-16B3 73.004 BMD-0804-16B3 765 BMD-0805-16B3 736 BMD-0806-16B3 93.357 BMD-0807-16B3 65.833
8 BMD-0808-16B3 88.009 BMD-0809-16B3 64
Sample Size = 9Average = 79.798Standard Deviation = 19.716Skewness = 1.58Excess Kurtosis = 2.67
Test of Fit: p-value = 0.0425 (SK All) Decision = Fail (SK Spec) Decision = No specs
61.0000 124.000092.5000
Normal Distribution Analysis
(Parameter TSS)
Powered by:
Distribution AnalyzerTM
(licensed software)
©Taylor Enterprises, Inc.
Page 21 of 21
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/03/02-01-2016
Ref. No. 006/UR/BMD/2016
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2493 Email: [email protected]
www.ujiprofisiensi.com
No Kode PesertaNilai Laporan
Peserta
1 BMD-0801-16B3 7.882 BMD-0802-16B3 7.653 BMD-0803-16B3 7.914 BMD-0804-16B3 7.995 BMD-0805-16B3 7.276 BMD-0806-16B3 7.82857 BMD-0807-16B3 7.6
8 BMD-0808-16B3 7.249 BMD-0809-16B3 7.03
Sample Size = 9Average = 7.600Standard Deviation = 0.344Skewness = -0.57Excess Kurtosis = -1.20
Test of Fit: p-value = 0.4175 (SK All) Decision = Pass (SK Spec) Decision = No specs
With 95% confidence more than 99% of the values are between 6.0080 and 9.1920
7.0300 7.99007.5100
Normal Distribution Analysis
(Parameter COD)
Powered by:
Distribution AnalyzerTM
(licensed software)
©Taylor Enterprises, Inc.
No Kode PesertaNilai Laporan
Peserta (mg/L)
1 BMD-0801-16B3 73.62 BMD-0802-16B3 73.323 BMD-0803-16B3 106.064 BMD-0804-16B3 69.885 BMD-0805-16B3 110.526 BMD-0807-16B3 133.714
7 BMD-0808-16B3 59.258 BMD-0809-16B3 74.44
Sample Size = 8Average = 87.597Standard Deviation = 25.865Skewness = 0.88Excess Kurtosis = -0.48
Test of Fit: p-value = 0.4397 (SK All) Decision = Pass (SK Spec) Decision = No specs
With 95% confidence more than 99% of the values are between -38.9008 and 214.0958
59.2500 133.710096.4800
Normal Distribution Analysis
(Parameter pH, 25oC)
Powered by:
Distribution AnalyzerTM
(licensed software)
©Taylor Enterprises, Inc.
©BMD Laboratory 2016