This is an official electronic document and secured by BMD Laboratory
©BMD Laboratory 2015
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: [email protected]
LAPORAN
PROGRAM UJI BANDING (PROFICIENCY TEST) ANTAR LABORATORIUM
SKEMA KHUSUS SEMESTER I-2015
Bahan Uji:
AIR MINUM DALAM KEMASAN I (AMDK I)
Ref. No. 005/UR/BMD/2015
PARAMETER UJI:
Nitrat, Sulfat, dan Klorida
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Page i of iii
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: [email protected]
www.ujiprofisiensi.com
Ref. No. 005/UR/BMD/2015
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
KATA PENGANTAR
Puji syukur kiranya kita panjatkan kepada Alloh SWT yang berkenan melimpahkan
karunia-Nya sehingga ‘Program Uji Banding Antar Laboratorium (Proficiency Test) Skema
Khusus Semester I-2015’ untuk bahan uji Air Minum dalam Kemasan I (AMDK I) dengan
parameter uji nitrat (NO3-), sulfat (SO4
2-), dan klorida (Cl
-) telah berhasil dilaksanakan dan
diselesaikan sampai dengan diterbitkannya laporan ini.
Program uji banding antar laboratorium merupakan salah satu instrumen yang sangat
penting sebagai sarana jaminan mutu eksternal (external quality control) bagi laboratorium
pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ISO/IEC 17025:2008. Dengan
mengikuti program uji banding, laboratorium dapat melakukan evaluasi kinerja pengujian yang
dilakukan dalam rangka melakukan peningkatan secara berkesinambungan (continuous
improvement) untuk memberikan layanan terbaik kepada pelanggan.
Program skema khusus kali ini diikuti oleh 11 (sebelas) laboratorium dimana 10 (sepuluh)
laboratorium diantaranya terakreditasi SNI ISO/IEC 17025:2008 oleh Komite Akreditas Nasional
(KAN). Seluruh data dapat diolah dengan baik menggunakan Robust Statistic Algoritma A Mean
sesuai ISO 13528:2005(E) dengan terlebih dahulu dilakukan evaluasi data menggunakan
perangkat lunak Distribution AnalyzerTM
untuk memastikan data telah memenuhi distribusi
normal.
Dalam program uji banding ini masih digunakan nilai konsensus (consensus value) dari
laboratorium peserta dalam melakukan evaluasi, sehingga tidak memuaskan dalam
ketertelusuran (traceability) secara metrologi. Akan tetapi BMD Laboratory meyakini bahwa hal
tersebut tidak mengurangi manfaat dari pelaksanaan program uji banding ini. Untuk itu BMD
Laboratory akan terus mengupayakan perbaikan demi perbaikan guna menghadirkan program
uji banding yang semakin bermanfaat bagi laboratorium-laboratorium di Indonesia.
BMD Laboratory sebagai provider uji banding antar laboratorium menyampaikan apresiasi
setinggi-tingginya kepada seluruh pihak yang telah membantu dan menyukseskan kegiatan ini.
Page ii of iii
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: [email protected]
www.ujiprofisiensi.com
Ref. No. 005/UR/BMD/2015
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Khususnya kepada seluruh laboratorium peserta atas partisipasi aktifnya mengikuti program uji
banding AMDK I. Ucapan terimakasih juga disampaikan kepada tim ahli yang telah bekerja keras
mengolah data dan menyusun laporan ini. Akhirnya, BMD Laboratory menyampaikan
permohonan maaf sekiranya terdapat kekurangan dalam penyelenggaraan program uji banding
antar laboratorium. Kritik, saran, dan masukan yang membangun sangat kami harapkan untuk
menghadirkan program uji banding yang semakin baik ke depannya. Semoga laporan ini dapat
dimanfaatkan oleh laboratorium peserta dan pihak-pihak terkait.
Tangerang Selatan, 05 Juni 2015
Ir. Renaldo Moontri
CEO BMD Group
(Selaku Manajer Puncak BMD Laboratory)
Page iii of iii
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: [email protected]
www.ujiprofisiensi.com
Ref. No. 005/UR/BMD/2015
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
DAFTAR ISI
KATA PENGANTAR ......................................................................................................................... i
DAFTAR ISI ................................................................................................................................... iii
I. PENDAHULUAN .................................................................................................................... 1
II. PELAKSANAAN ..................................................................................................................... 1
III. STATISTIKA PENGOLAHAN DATA ......................................................................................... 2
III.1 Penilaian Laboratorium ............................................................................................... 2
III.2 Penetapan Nilai Evaluasi Uji Banding .......................................................................... 2
III.3 Penetapan SDPA .......................................................................................................... 2
IV. HASIL UJI BANDING ANTAR LABORATORIUM ...................................................................... 3
IV.1 Parameter Nitrat (NO3-) ............................................................................................. 4
IV.2 Parameter Sulfat (SO42-
) ............................................................................................. 6
IV.3 Parameter Klorida (Cl-) ............................................................................................... 8
V. PENUTUP ........................................................................................................................... 10
VI. REFERENSI .......................................................................................................................... 11
Annex A Rekapitulasi Hasil z-score Program Uji Banding ......................................................... 12
Annex B Rekapitulasi Bias Nilai Peserta Terhadap Nilai Konsensus ......................................... 13
Annex C Evaluasi Distribusi Normal Data Program Uji Banding ............................................... 15
Page 1 of 17
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 005/UR/BMD/2015
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: [email protected]
www.ujiprofisiensi.com
I. PENDAHULUAN
Salah satu kebutuhan dasar manusia adalah air, terutama yang digunakan untuk minum.
Kualitas kesehatan manusia salah satunya sangat ditentukan oleh kualitas air yang
diminumnya. Oleh karena itu, air minum dalam kemasan (AMDK) merupakan salah satu
komoditi yang diatur persyaratan mutunya sesuai SNI 01-3553-2006. Persyaratan mutu
untuk parameter nitrat (NO3-), sulfat (SO4
2-), dan klorida (Cl-) seperti pada Tabel 1.
Tabel 1. Persyaratan mutu air minum dalam kemasan (air mineral)[1]
No. Parameter Satuan Persyaratan
1. Nitrat (NO3-) mg/L Maks. 45 mg/L
2. Sulfat (SO42-) mg/L Maks. 200 mg/L
3. Klorida (Cl-) mg/L Maks. 250 mg/L
Salah satu jaminan mutu bagi laboratorium pengujian dan kalibrasi sesuai persyaratan
dalam SNI ISO/IEC 17025:2008 adalah partisipasi dalam uji banding antar laboratorium atau
uji profisiensi[2] sebagai jaminan mutu eksternal (external quality control). Untuk itu BMD
Laboratory menyelenggarakan program uji banding antar laboratorium (proficiency test)
untuk bahan uji air minum dalam kemasan I (AMDK I) dengan tujuan sebagai berikut:
1. Sebagai sarana jaminan mutu eksternal (external quality control) bagi laboratorium
peserta khususnya laboratorium yang menerapkan standard SNI ISO/IEC 17025:2008.
2. Membantu laboratorium dalam mengevaluasi unjuk kerja pengujian kadar nitrat (NO3-),
sulfat (SO42-), dan klorida (Cl-) untuk bahan uji air minum dalam kemasan.
II. PELAKSANAAN
Berikut adalah jadual pelaksanaan program uji banding antar laboratorium (proficiency test)
yang diselenggarakan oleh BMD Laboratory untuk bahan uji air minum dalam kemasan I
(AMDK I):
Page 2 of 17
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 005/UR/BMD/2015
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: [email protected]
www.ujiprofisiensi.com
a. Penutupan pendaftaran: 30 Maret 2015
b. Pendistribusian bahan uji: 13 s.d. 30 April 2015
c. Pelaporan hasil analisis oleh laboratorium peserta: Maks. 21 Mei 2015
d. Pendistribusian laporan program uji banding oleh BMD Laboratory: Maks. 08 Juni 2015
III. STATISTIKA PENGOLAHAN DATA
III.1 Penilaian Laboratorium
Penilaian terhadap unjuk kerja laboratorium menggunakan z-score dengan tiga kriteria
penilaian sebagai berikut:
� Untuk |z-score|≤2,0 dikategorikan memuaskan diberi lambang ‘OK’
� Untuk 2,0<|z-score|<3,0 dikategorikan meragukan diberi lambang ‘$’
� Untuk |z-score|≥3,0 dikategorikan kurang memuaskan diberi lambang ‘$$’
III.2 Penetapan Nilai Evaluasi Uji Banding
Penetapan nilai evaluasi uji banding menggunakan dua pendekatan statistika yaitu:
� Robust Statistic Algoritma A Mean sesuai ISO 13528:2005(E)[3]. Nilai tersebut
merupakan nilai konsensus dari peserta (consensus value)
� Seleksi data menggunakan perangkat lunak Distribution AnalyzerTM, Taylor
Enterprises, Inc.[4]
III.3 Penetapan SDPA
SDPA (simpangan baku yang digunakan dalam evaluasi) ditetapkan menggunakan tiga
pendekatan statistika yaitu:
� Robust Statistic Algoritma A Mean sesuai ISO 13528:2005(E)
� Simpangan Baku Horwitz[5]
� SDPA yang ditetapkan (Assigned SDPA)
Page 3 of 17
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 005/UR/BMD/2015
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: [email protected]
www.ujiprofisiensi.com
IV. HASIL UJI BANDING ANTAR LABORATORIUM
Sebanyak 11 (sebelas) laboratorium telah berpartisipasi dalam program uji banding antar
laboratorium (proficiency test) yang diselenggarakan oleh BMD Laboratory untuk bahan uji
air minum dalam kemasan I (AMDK I) dengan parameter uji nitrat (NO3-), sulfat (SO4
2-), dan
klorida (Cl-). Adapun 10 (sepuluh) laboratorium diantaranya merupakan laboratorium
terakreditasi SNI ISO/IEC 17025:2008 oleh Komite Akreditasi Nasional (KAN). Seluruh data
hasil uji banding dapat diolah dengan baik menggunakan Robust Statistic Algoritma A
Mean sesuai ISO 13528: 2005(E). Selain itu dilakukan juga evaluasi data menggunakan
perangkat lunak Distribution AnalyzerTM guna memastikan bahwa data yang diolah
mengikuti pola distribusi normal pada tingkat kepercayaan 95%. Nilai evaluasi uji banding
(assigned value) merupakan nilai konsensus (consensus value) dari laboratorium peserta
sehingga nilai tersebut bukan merupakan nilai acuan yang tertelusur secara metrologi ke
sistem satuan internasional (SI).
Tabel 2. Rekapitulasi Hasil Pengujian Laboratorium Peserta
No. Kode Lab. Nitrat (NO3
-) Sulfat (SO4
2-) Klorida (Cl
-)
mg.L-1
mg.L-1
mg.L-1
1. BMD-201-15 9.84 ± 0.06 4.68 ± 0.43 7.12 ± 0.81
Metode SNI 01-3554-2006 SNI 01-3554-2006 SNI 01-3554-2006
2. BMD-202-15 - 5.64 ± 0.44 16.19 ± 0.24
Metode - SNI 6989.20:2009 SNI 6989.19:2009
3. BMD-204-15 3.46 ± 0.20 3.89 ± 2.43 6.92 ± 0.22
Metode SNI 06-6989.79-2011 SNI 6989.20:2009 SNI 6989.19:2009
4. BMD-205-15 2.44 4.50 11.41
Metode SNI 6989.79:2011 SNI 06-6989.20:2009 SNI 06-6989.19:2009
Page 4 of 17
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 005/UR/BMD/2015
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: [email protected]
www.ujiprofisiensi.com
IV.1 Parameter Nitrat (NO3-)
Sebanyak 10 (sepuluh) laporan hasil pengujian untuk parameter nitrat (NO3-) diterima oleh
BMD Laboratory dan 5 (lima) diantaranya melaporkan nilai estimasi ketidakpastian
diperluas dengan k=2 pada tingkat kepercayaan 95%. Dari hasil pengolahan data secara
statistika diperoleh sebanyak 5 (lima) laboratorium masuk dalam kategori ‘memuaskan’ (|z-
score|≤2,0), 2 (dua) laboratorium masuk dalam kategori ‘meragukan’ (2,0<|z-score|<3,0),
dan 3 (tiga) laboratorium masuk dalam kategori ‘kurang memuaskan’ (|z-score|≥3,0). Hasil
pengolahan data disajikan pada Tabel 3 dan Gambar 1.
Kode Lab. Nitrat (NO3
-) Sulfat (SO4
2-) Klordia (Cl
-)
mg.L-1 mg.L
-1 mg.L-1
5. BMD-206-15 12.10 5.96 7.40
Metode Spectroquant Spektrofotometri Spectroquant
6. BMD-207-15 11.04 3.96 5.94
Metode Spektrofotometri Spektrofotometri Titrimetri
7. BMD-208-15 2.02 3.77 ± 0.29 10.00 ± 0.24
Metode SNI 06-2480-1991 SNI 6989.20:2009 SNI 6989.19:2009
8. BMD-209-15 10.21 ± 1.021 3.08 ± 0.30 7.09 ± 0.7
Metode Ion Kromatografi Ion Kromatografi Ion Kromatografi
9. BMD-210-15 3.19 11.75 7.35
Metode Spektrofotometri Spektrofotometri Titrimetri
10. BMD-213-15 9.81 ± 0.09 4.60 ± 0.45 8.01 ± 0.69
Metode SNI 01-3554-2006 SNI 6989.20-2009 SNI 6989.19-2009
11. BMD-215-15 2.26 ± 0.07 3.20 ± 0.63 9.41 ± 0.44
Metode SNI 01-3554-2006 (2.8) SNI 01-3554-2006 (2.11) SNI 01-3554-2006 (2.12)
No.
Page 5 of 17
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 005/UR/BMD/2015
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: [email protected]
www.ujiprofisiensi.com
Tabel 3. Hasil Pengolahan Data Parameter Nitrat (NO3-)
No. Kode Peserta Nilai Laporan
Peserta (mg/L)
U, k=2
(mg/L)
Nilai Evaluasi Uji
Banding (mg/L)
SDPA
(mg/L) z-score Kategori
1. BMD-201-15 9.84 0.06 8.85 1.95 0.51 OK
2. BMD-204-15 3.46 0.20 8.85 1.95 -2.77 $
3. BMD-205-15 2.44 - 8.85 1.95 -3.29 $$
4. BMD-206-15 12.10 - 8.85 1.95 1.67 OK
5. BMD-207-15 11.04 - 8.85 1.95 1.12 OK
6. BMD-208-15 2.02 - 8.85 1.95 -3.51 $$
7. BMD-209-15 10.21 1.021 8.85 1.95 0.70 OK
8. BMD-210-15 3.19 - 8.85 1.95 -2.91 $
9. BMD-213-15 9.81 0.09 8.85 1.95 0.49 OK
10. BMD-215-15 2.26 0.07 8.85 1.95 -3.39 $$
Keterangan:
Nilai evaluasi uji banding merupakan nilai konsensus (consensus value) laboratorium
peserta yang diperoleh dari pengolahan data secara statistika menggunakan Robust
Statistic Algoritma A Mean sesuai ISO 13528:2005(E) dengan sebelumnya dilakukan analisis
menggunakan perangkat lunak Distribution AnalyzerTM (lihat Anex C). SDPA adalah
simpangan baku yang ditetapkan sebesar 22% dari nilai evaluasi uji banding (assigned
value), merupakan rata-rata dari simpangan baku Horwitz (SDHorwitz) dan simpangan baku
Algoritma A Mean. U yaitu ketidakpastian diperluas dengan faktor cakupan (k)=2 pada
tingkat kepercayaan 95%.
Page 6 of 17
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 005/UR/BMD/2015
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: [email protected]
www.ujiprofisiensi.com
0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00
10.00 11.00 12.00 13.00 14.00 15.00 16.00 17.00 18.00 19.00 20.00 21.00 22.00
BM
D-2
08-1
5
BM
D-2
15-1
5
BM
D-2
05-1
5
BM
D-2
10-1
5
BM
D-2
04-1
5
BM
D-2
13-1
5
BM
D-2
01-1
5
BM
D-2
09-1
5
BM
D-2
07-1
5
BM
D-2
06-1
5
NO
3-(m
g/L
)
Lab. Code
Cons. value
+2SDPA
+3SDPA
-2SDPA
-3SDPA
Cons. value = 8.85 mg/LAlgoritma A Mean ISO 13528:2005(E)SDPA = 1.95 mg/L (22% Assigned Value)
©BMD Laboratory
Gambar 1. Grafik Hasil Uji Banding Parameter Nitrat (NO3-)
IV.2 Parameter Sulfat (SO42-
)
Sebanyak 11 (sebelas) laporan hasil pengujian untuk parameter sulfat (SO42-) diterima oleh
BMD Laboratory dan 7 (tujuh) diantaranya melaporkan nilai estimasi ketidakpastian
diperluas dengan k=2 pada tingkat kepercayaan 95%. Dari hasil pengolahan data secara
statistika diperoleh sebanyak 7 (tujuh) laboratorium masuk dalam kategori ‘memuaskan’
(|z-score|≤2,0), 2 (dua) laboratorium masuk dalam kategori ‘meragukan’ (2,0<|z-
score|<3,0), dan 2 (dua) laboratorium masuk dalam kategori ‘kurang memuaskan’ (|z-
score|≥3,0). Hasil pengolahan data disajikan pada Tabel 4 dan Gambar 2.
Page 7 of 17
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 005/UR/BMD/2015
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: [email protected]
www.ujiprofisiensi.com
Tabel 4. Hasil Pengolahan Data Parameter Sulfat (SO42-)
No. Kode Peserta Nilai Laporan
Peserta (mg/L)
U, k=2
(mg/L)
Nilai Evaluasi Uji
Banding (mg/L)
SDPA
(mg/L) z-score Kategori
1. BMD-201-15 4.68 0.43 4.23 0.54 0.82 OK
2. BMD-202-15 5.64 0.44 4.23 0.54 2.58 $
3. BMD-204-15 3.89 2.43 4.23 0.54 -0.63 OK
4. BMD-205-15 4.50 - 4.23 0.54 0.49 OK
5. BMD-206-15 5.96 - 4.23 0.54 3.17 $$
6. BMD-207-15 3.96 - 4.23 0.54 -0.50 OK
7. BMD-208-15 3.77 0.29 4.23 0.54 -0.85 OK
8. BMD-209-15 3.08 0.30 4.23 0.54 -2.11 $
9. BMD-210-15 11.75 - 4.23 0.54 13.80 $$
10. BMD-213-15 4.60 0.45 4.23 0.54 0.68 OK
11. BMD-215-15 3.20 0.63 4.23 0.54 -1.89 OK
Keterangan:
Nilai evaluasi uji banding merupakan nilai konsensus (consensus value) laboratorium
peserta yang diperoleh dari pengolahan data secara statistika menggunakan Robust
Statistic Algoritma A Mean sesuai ISO 13528:2005(E) dengan sebelumnya dilakukan analisis
menggunakan perangkat lunak Distribution AnalyzerTM (lihat Anex C), sedangkan SDPA
adalah simpangan baku Horwitz (SDHorwitz). U yaitu ketidakpastian diperluas dengan faktor
cakupan (k)=2 pada tingkat kepercayaan 95%.
Page 8 of 17
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 005/UR/BMD/2015
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: [email protected]
www.ujiprofisiensi.com
0.00
1.00
2.00
3.00
4.00
5.00
6.00
7.00
8.00
9.00
10.00
11.00
12.00
13.00
BM
D-2
09-1
5
BM
D-2
15-1
5
BM
D-2
08-1
5
BM
D-2
04-1
5
BM
D-2
07-1
5
BM
D-2
05-1
5
BM
D-2
13-1
5
BM
D-2
01-1
5
BM
D-2
02-1
5
BM
D-2
06-1
5
BM
D-2
10-1
5
SO
42
-(m
g/L
)
Lab. Code
Cons. value
+2SDPA
+3SDPA
-2SDPA
-3SDPA
Cons. value = 4.23 mg/LAlgoritma A Mean ISO 13528:2005(E)SDPA = 0.54 mg/L (Horwitz Function)
©BMD Laboratory
Gambar 2. Grafik Hasil Uji Banding Parameter Sulfat (SO42-)
IV.3 Parameter Klorida (Cl-)
Sebanyak 11 (sebelas) laporan hasil pengujian untuk parameter klorida (Cl-) diterima oleh
BMD Laboratory dan 7 (tujuh) diantaranya melaporkan nilai estimasi ketidakpastian
diperluas dengan k=2 pada tingkat kepercayaan 95%. Dari hasil pengolahan data secara
statistika diperoleh sebanyak 7 (tujuh) laboratorium masuk dalam kategori ‘memuaskan’
(|z-score|≤2,0), 2 (dua) laboratorium masuk dalam kategori ‘meragukan’ (2,0<|z-
score|<3,0), dan 2 (dua) laboratorium masuk dalam kategori ‘kurang memuaskan’ (|z-
score|≥3,0). Hasil pengolahan data disajikan pada Tabel 5 dan Gambar 3.
Page 9 of 17
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 005/UR/BMD/2015
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: [email protected]
www.ujiprofisiensi.com
Tabel. 5 Hasil Pengolahan Data Parameter Klorida (Cl-)
No. Kode Peserta Nilai Laporan
Peserta (mg/L)
U, k=2
(mg/L)
Nilai Evaluasi Uji
Banding (mg/L)
SDPA
(mg/L) z-score Kategori
1. BMD-201-15 7.12 0.81 7.58 0.89 -0.52 OK
2. BMD-202-15 16.19 0.24 7.58 0.89 9.63 $$
3. BMD-204-15 6.92 0.22 7.58 0.89 -0.74 OK
4. BMD-205-15 11.41 - 7.58 0.89 4.28 $$
5. BMD-206-15 7.40 - 7.58 0.89 -0.20 OK
6. BMD-207-15 5.94 - 7.58 0.89 -1.84 OK
7. BMD-208-15 10.00 0.24 7.58 0.89 2.70 $
8. BMD-209-15 7.09 0.7 7.58 0.89 -0.55 OK
9. BMD-210-15 7.35 - 7.58 0.89 -0.26 OK
10. BMD-213-15 8.01 0.69 7.58 0.89 0.48 OK
11. BMD-215-15 9.41 0.44 7.58 0.89 2.04 $
Keterangan:
Nilai evaluasi uji banding merupakan nilai konsensus (consensus value) laboratorium
peserta yang diperoleh dari pengolahan data secara statistika menggunakan Robust
Statistic Algoritma A Mean sesuai ISO 13528:2005(E) dengan sebelumnya dilakukan analisis
menggunakan perangkat lunak Distribution AnalyzerTM (lihat Anex C), sedangkan SDPA
adalah simpangan baku Horwitz (SDHorwitz). U yaitu ketidakpastian diperluas dengan faktor
cakupan (k)=2 pada tingkat kepercayaan 95%.
Page 10 of 17
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 005/UR/BMD/2015
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: [email protected]
www.ujiprofisiensi.com
2.00
3.00
4.00
5.00
6.00
7.00
8.00
9.00
10.00
11.00
12.00
13.00
14.00
15.00
16.00
17.00
18.00
BM
D-2
07-1
5
BM
D-2
04-1
5
BM
D-2
09-1
5
BM
D-2
01-1
5
BM
D-2
10-1
5
BM
D-2
06-1
5
BM
D-2
13-1
5
BM
D-2
15-1
5
BM
D-2
08-1
5
BM
D-2
05-1
5
BM
D-2
02-1
5
Cl-
(mg
/L)
Lab. Code
Cons. value
+2SDPA
+3SDPA
-2SDPA
-3SDPA
Cons. value = 7.58 mg/LAlgoritma A Mean ISO 13528:2005(E)SDPA = 0.89 mg/L (Horwitz Function)
©BMD Laboratory
Gambar 3. Grafik Hasil Uji Banding Parameter Klorida (Cl-)
V. PENUTUP
Program uji banding antar laboratorium merupakan salah satu sarana jaminan mutu
eksternal bagi laboratorium pengujian dan kalibrasi. Hasil dari program uji banding ini
diharapkan dapat digunakan sebagai sarana evaluasi bagi laboratorium untuk meningkatkan
kinerja dalam kegiatan pengujian yang dilakukan.
Page 11 of 17
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 005/UR/BMD/2015
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: [email protected]
www.ujiprofisiensi.com
VI. REFERENSI
[1] SNI 01-3553-2006
[2] SNI ISO/IEC 17025:2008
[3] ISO 13528:2005(E)
[4] Distribution AnalyzerTM, Copyright © by Taylor Enterprises, Inc. USA
[5] M. Thompson (2004) The Amazing Horwitz Function, AMC Technical Brief No.17, Royal
Society of Chemistry
Page 12 of 17
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 005/UR/BMD/2015
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: [email protected]
www.ujiprofisiensi.com
Annex A
Rekapitulasi Hasil z-score Program Uji Banding
No. Kode Peserta z-score
Nitrat (NO3-) Sulfat (SO4
2-) Klorida (Cl
-)
1. BMD-201-15 0.51 OK 0.82 OK -0.52 OK
2. BMD-202-15 - - 2.58 $ 9.63 $$
3. BMD-204-15 -2.77 $ -0.63 OK -0.74 OK
4. BMD-205-15 -3.29 $$ 0.49 OK 4.28 $$
5. BMD-206-15 1.67 OK 3.17 $$ -0.20 OK
6. BMD-207-15 1.12 OK -0.50 OK -1.84 OK
7. BMD-208-15 -3.51 $$ -0.85 OK 2.70 $
8. BMD-209-15 0.70 OK -2.11 $ -0.55 OK
9. BMD-210-15 -2.91 $ 13.80 $$ -0.26 OK
10. BMD-213-15 0.49 OK 0.68 OK 0.48 OK
11. BMD-215-15 -3.39 $$ -1.89 OK 2.04 $
Keterangan:
OK : Kategori ‘memuaskan’
$ : Kategori ‘meragukan’
$$ : Kategori ‘kurang memuaskan’
Page 13 of 17
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 005/UR/BMD/2015
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: [email protected]
www.ujiprofisiensi.com
Annex B
Rekapitulasi Bias Nilai Peserta Terhadap Nilai Konsensus
-8.00
-6.00
-4.00
-2.00
0.00
2.00
4.00
b (
mg
/L)
Lab. Code
Parameter Nitrat (NO3-)
-2.00-1.000.001.002.003.004.005.006.007.008.00
b (
mg
/L)
Lab. Code
Parameter Sulfat(SO42-)
Page 14 of 17
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 005/UR/BMD/2015
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: [email protected]
www.ujiprofisiensi.com
-4.00
-2.00
0.00
2.00
4.00
6.00
8.00
10.00
b (
mg
/L)
Lab. Code
Parameter Klorida (Cl-)
Page 15 of 17
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 005/UR/BMD/2015
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: [email protected]
www.ujiprofisiensi.com
No Kode PesertaNilai Laporan
Peserta (mg/L)
1 BMD-201-15 9.842 BMD-204-15 3.463 BMD-205-15 2.444 BMD-206-15 12.105 BMD-207-15 11.046 BMD-208-15 2.02
7 BMD-209-15 10.218 BMD-210-15 3.199 BMD-213-15 9.81
10 BMD-215-15 2.26
Sample Size = 10Average = 6.637Standard Deviation = 4.248Skewness = 0.05Excess Kurtosis = -2.32
Test of Fit: p-value = 0.0033 (SK All) Decision = Fail (SK Spec) Decision = No specs
2.0200 12.10007.0600
No Kode PesertaNilai Laporan
Peserta (mg/L)
1 BMD-201-15 9.842 BMD-204-15 3.463 BMD-205-15 2.444 BMD-206-15 12.105 BMD-207-15 11.047 BMD-209-15 10.218 BMD-210-15 3.199 BMD-213-15 9.81
Sample Size = 8Average = 7.761Standard Deviation = 3.996Skewness = -0.53Excess Kurtosis = -2.05
Test of Fit: p-value = 0.9604 (SK All) Decision = Pass (SK Spec) Decision = No specs
With 95% confidence more than 99% of the values are between 2.4248 and infinity
2.4400 12.10007.2700
Annex C
Evaluasi Distribusi Normal Data Program Uji Banding
Normal Distribution Analysis
(Parameter NO3-)
Powered by:
Distribution AnalyzerTM
©Taylor Enterprises, Inc.
5510 Fairmont Rd., Suite A
Libertyville, IL 60048
USA
www.variation.com
Normal Distribution Analysis
(Parameter NO3-)
Powered by:
Distribution AnalyzerTM
©Taylor Enterprises, Inc.
5510 Fairmont Rd., Suite A
Libertyville, IL 60048
USA
www.variation.com
BMD-208-15; BMD-215-15
were rejected
Page 16 of 17
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 005/UR/BMD/2015
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: [email protected]
www.ujiprofisiensi.com
No Kode PesertaNilai Laporan
Peserta (mg/L)
1 BMD-201-15 4.682 BMD-202-15 5.643 BMD-204-15 3.894 BMD-205-15 4.505 BMD-206-15 5.966 BMD-207-15 3.967 BMD-208-15 3.77
8 BMD-209-15 3.089 BMD-210-15 11.75
10 BMD-213-15 4.6011 BMD-215-15 3.20
Sample Size = 11Average = 5.003Standard Deviation = 2.412Skewness = 2.53Excess Kurtosis = 7.23
Test of Fit: p-value = 0.0034 (SK All) Decision = Fail (SK Spec) Decision = No specs
3.0800 11.75007.4150
No Kode PesertaNilai Laporan
Peserta (mg/L)
1 BMD-201-15 4.682 BMD-202-15 5.643 BMD-204-15 3.894 BMD-205-15 4.505 BMD-206-15 5.966 BMD-207-15 3.967 BMD-208-15 3.77
8 BMD-209-15 3.0810 BMD-213-15 4.6011 BMD-215-15 3.20
Sample Size = 10Average = 4.328Standard Deviation = 0.948Skewness = 0.49Excess Kurtosis = -0.47
Test of Fit: p-value = 0.7439 (SK All) Decision = Pass (SK Spec) Decision = No specs
With 95% confidence more than 99% of the values are between 0.1275 and 8.5285
3.0800 5.96004.5200
Normal Distribution Analysis
(Parameter SO42-
)
Powered by:
Distribution AnalyzerTM
©Taylor Enterprises, Inc.
5510 Fairmont Rd., Suite A
Libertyville, IL 60048
USA
www.variation.com
Normal Distribution Analysis
(Parameter SO42-
)
Powered by:
Distribution AnalyzerTM
©Taylor Enterprises, Inc.
5510 Fairmont Rd., Suite A
Libertyville, IL 60048
USA
www.variation.com
BMD-210-15 was rejected
Page 17 of 17
FR.06 Laporan Program Uji Banding (Final Report)
Ed./Rev./Date: 01/02/16-04-2015
Ref. No. 005/UR/BMD/2015
BMD Laboratory–Provider of Proficiency Testing Program
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316
Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email: [email protected]
www.ujiprofisiensi.com
No Kode PesertaNilai Laporan
Peserta (mg/L)
1 BMD-201-15 7.122 BMD-202-15 16.193 BMD-204-15 6.924 BMD-205-15 11.415 BMD-206-15 7.406 BMD-207-15 5.947 BMD-208-15 10.008 BMD-209-15 7.099 BMD-210-15 7.35
10 BMD-213-15 8.0111 BMD-215-15 9.41
Sample Size = 11Average = 8.804Standard Deviation = 2.922Skewness = 1.84Excess Kurtosis = 3.70
Test of Fit: p-value = 0.0085 (SK All) Decision = Fail (SK Spec) Decision = No specs
5.9400 16.190011.0650
No Kode PesertaNilai Laporan
Peserta (mg/L)
1 BMD-201-15 7.123 BMD-204-15 6.924 BMD-205-15 11.415 BMD-206-15 7.406 BMD-207-15 5.947 BMD-208-15 10.00
8 BMD-209-15 7.099 BMD-210-15 7.35
10 BMD-213-15 8.0111 BMD-215-15 9.41
Sample Size = 10Average = 8.065Standard Deviation = 1.679Skewness = 0.99Excess Kurtosis = 0.24
Test of Fit: p-value = 0.2731 (SK All) Decision = Pass (SK Spec) Decision = No specs
With 95% confidence more than 99% of the values are between 0.6223 and 15.5077
5.9400 11.41008.6750
Normal Distribution Analysis
(Parameter Cl-)
Powered by:
Distribution AnalyzerTM
©Taylor Enterprises, Inc.
5510 Fairmont Rd., Suite A
Libertyville, IL 60048
USA
www.variation.com
Normal Distribution Analysis
(Parameter Cl-)
Powered by:
Distribution AnalyzerTM
©Taylor Enterprises, Inc.
5510 Fairmont Rd., Suite A
Libertyville, IL 60048
USA
www.variation.com
BMD-202-15 was rejected
©BMD Laboratory 2015