Cite this article: Hidaka S, Kobayashi S,Maesato K, Mochida Y, Ishioka K, et al. (2015) Hydrophilic Polymer-Coated Polysulfone Membrane Improves Endo- thelial Function of Hemodialysis Patients: A Pilot Study. J Clin Nephrol Res 2(2): 1020. Central Journal of Clinical Nephrology and Research *Corresponding author Sumi Hidaka, Department of Nephrology, Immunology, and Vascular Medicine, Kidney Disease and Transplant Center, Shonan Kamakura General Hospital, Okamoto 1370-1, Kamakura, 247- 8533, Tel: 81467461717 ; Fax: 81467478234 ; E-mail: Submitted: 04 August 2015 Accepted: 05 September 2015 Published: 07 September 2015 ISSN: 2379-0652 Copyrighta © 2015 Hidaka et al. OPEN ACCESS Clinical Image Hydrophilic Polymer-Coated Polysulfone Membrane Improves Endothelial Function of Hemodialysis Patients: A Pilot Study Hidaka S 1 *, Kobayashi S 1 , Maesato K 1 , Mochida Y 1 , Ishioka K 1 , Oka M 1 , Moriya H 1 , Ohtake T 1 and Nomura S 2 1 Department of Nephrology, Immunology, and Vascular Medicine, Shonan Kamakura General Hospital, Japan 2 Department of Internal Medicine, Kansai Medical University, Japan ABBREVIATIONS PDMPs: Platelet-Derived MicroParticles; HD: Hemodialysis; FMD: Endothelium-Dependent Vasodilatation; PVP: Polyvinylpyrrolidone; PS: Polysulfone. INTRODUCTION Accelerated atherosclerosis is a major risk factor for long- term survivors receiving maintenance hemodialysis (HD); in particular, cardiovascular disease is a leading cause of morbidity and mortality in HD patients [1,2]. Atherosclerosis results from complex processes accompanied by endothelial dysfunction and inflammation. HD patients have 3 types of risk factors for atherosclerosis: traditional, uremia-related, and dialysis-related [3]. Dialysis-related risk factors result from bio-incompatibility between the blood and medical equipment, including the dialyzer. Therefore, it is extremely important to reduce dialysis- related risk factors to prevent the progression of accelerated atherosclerosis. Platelet-derived microparticles (PDMPs) are released from activated platelets during HD [4]. PDMPs contain platelet granular proteins, such as P-selectin (CD62P), and various platelet surface membrane glycoproteins (GPs), such as GP Ib/IX (CD42) and GP IIb/IIIa (CD41) [5,6]. They possess procoagulant activity themselves. The level of PDMPs is significantly increased in many prothrombotic diseases, including diabetes, hypertension, acute coronary syndromes, peripheral artery disease, and uremia [4,6-8]. When uremic patients undergo HD, shear stress and the contact between blood and non-human materials could be the mechanisms of PDMPs generation [4]. Because PDMPs contribute to the development of thrombotic complications and atherosclerosis, lower levels of PDMPs are desirable Biocompatibility characteristics and solute clearance of HD Abstract Most polysulfone hemodialysis membranes are hydrophilized to improve biocompatibility; however, platelets are still activated by adhering to the membrane surface. We evaluated a newly developed polysulfone membrane in terms of platelet activation and endothelial function. Twenty-four patients underwent dialysis with a traditional polysulfone membrane (n = 12) or a new polysulfone membrane coated with hydrophilic polymer (n = 12) for 3 months. We analyzed the level of platelet- derived microparticles and flow-mediated dilatation of the brachial artery before and after 3 months. Our results showed that treatment with the traditional membranes did not significantly affect the level of platelet-derived microparticles (11.8 + 2.2 to 12.1 + 3.2 U/mL, p = 0.76) or flow-mediated dilatation (2.3 + 1.5% to 2.7 + 0.9%, p = 0.41). However, treatment with the new polysulfone membrane significantly decreased the level of platelet-derived microparticles (16.7 + 8.1 to 15.0 + 6.1 U/mL, p = 0.03) and increased flow-mediated dilatation (3.0 + 1.8% to 4.1 + 1.9%, p = 0.04).A newly developed polysulfone membrane coated with hydrophilic polymer improves the level of platelet-derived microparticles and endothelial function. Keywords Hemodialysis Polysulfone Platelet-derived microparticles Flow-mediated dilatation Endothelial function

Hydrophilic Polymer-Coated Polysulfone … this article: Hidaka S, Kobayashi S,Maesato K, Mochida Y, Ishioka K, et al. (2015) Hydrophilic Polymer-Coated Polysulfone Membrane Improves

  • Upload
    hanhi

  • View
    219

  • Download
    2

Embed Size (px)

Citation preview

Page 1: Hydrophilic Polymer-Coated Polysulfone … this article: Hidaka S, Kobayashi S,Maesato K, Mochida Y, Ishioka K, et al. (2015) Hydrophilic Polymer-Coated Polysulfone Membrane Improves

Cite this article: Hidaka S, Kobayashi S,Maesato K, Mochida Y, Ishioka K, et al. (2015) Hydrophilic Polymer-Coated Polysulfone Membrane Improves Endo-thelial Function of Hemodialysis Patients: A Pilot Study. J Clin Nephrol Res 2(2): 1020.

Central Journal of Clinical Nephrology and Research

*Corresponding authorSumi Hidaka, Department of Nephrology, Immunology, and Vascular Medicine, Kidney Disease and Transplant Center, Shonan Kamakura General Hospital, Okamoto 1370-1, Kamakura, 247-8533, Tel: 81467461717 ; Fax: 81467478234 ; E-mail:

Submitted: 04 August 2015

Accepted: 05 September 2015

Published: 07 September 2015

ISSN: 2379-0652

Copyrighta© 2015 Hidaka et al.

OPEN ACCESS

Clinical Image

Hydrophilic Polymer-Coated Polysulfone Membrane Improves Endothelial Function of Hemodialysis Patients: A Pilot StudyHidaka S1*, Kobayashi S1, Maesato K1, Mochida Y1, Ishioka K1, Oka M1, Moriya H1, Ohtake T1 and Nomura S2

1Department of Nephrology, Immunology, and Vascular Medicine, Shonan Kamakura General Hospital, Japan2Department of Internal Medicine, Kansai Medical University, Japan

ABBREVIATIONSPDMPs: Platelet-Derived MicroParticles; HD: Hemodialysis;

FMD: Endothelium-Dependent Vasodilatation; PVP: Polyvinylpyrrolidone; PS: Polysulfone.

INTRODUCTIONAccelerated atherosclerosis is a major risk factor for long-

term survivors receiving maintenance hemodialysis (HD); in particular, cardiovascular disease is a leading cause of morbidity and mortality in HD patients [1,2]. Atherosclerosis results from complex processes accompanied by endothelial dysfunction and inflammation. HD patients have 3 types of risk factors for atherosclerosis: traditional, uremia-related, and dialysis-related [3]. Dialysis-related risk factors result from bio-incompatibility between the blood and medical equipment, including the dialyzer. Therefore, it is extremely important to reduce dialysis-

related risk factors to prevent the progression of accelerated atherosclerosis.

Platelet-derived microparticles (PDMPs) are released from activated platelets during HD [4]. PDMPs contain platelet granular proteins, such as P-selectin (CD62P), and various platelet surface membrane glycoproteins (GPs), such as GP Ib/IX (CD42) and GP IIb/IIIa (CD41) [5,6]. They possess procoagulant activity themselves. The level of PDMPs is significantly increased in many prothrombotic diseases, including diabetes, hypertension, acute coronary syndromes, peripheral artery disease, and uremia [4,6-8]. When uremic patients undergo HD, shear stress and the contact between blood and non-human materials could be the mechanisms of PDMPs generation [4]. Because PDMPs contribute to the development of thrombotic complications and atherosclerosis, lower levels of PDMPs are desirable

Biocompatibility characteristics and solute clearance of HD

Abstract

Most polysulfone hemodialysis membranes are hydrophilized to improve biocompatibility; however, platelets are still activated by adhering to the membrane surface. We evaluated a newly developed polysulfone membrane in terms of platelet activation and endothelial function. Twenty-four patients underwent dialysis with a traditional polysulfone membrane (n = 12) or a new polysulfone membrane coated with hydrophilic polymer (n = 12) for 3 months. We analyzed the level of platelet-derived microparticles and flow-mediated dilatation of the brachial artery before and after 3 months. Our results showed that treatment with the traditional membranes did not significantly affect the level of platelet-derived microparticles (11.8 + 2.2 to 12.1 + 3.2 U/mL, p = 0.76) or flow-mediated dilatation (2.3 + 1.5% to 2.7 + 0.9%, p = 0.41). However, treatment with the new polysulfone membrane significantly decreased the level of platelet-derived microparticles (16.7 + 8.1 to 15.0 + 6.1 U/mL, p = 0.03) and increased flow-mediated dilatation (3.0 + 1.8% to 4.1 + 1.9%, p = 0.04).A newly developed polysulfone membrane coated with hydrophilic polymer improves the level of platelet-derived microparticles and endothelial function.

Keywords•Hemodialysis•Polysulfone•Platelet-derived microparticles•Flow-mediated dilatation•Endothelial function

Page 2: Hydrophilic Polymer-Coated Polysulfone … this article: Hidaka S, Kobayashi S,Maesato K, Mochida Y, Ishioka K, et al. (2015) Hydrophilic Polymer-Coated Polysulfone Membrane Improves

Central

Hidaka et al. (2015)Email:

J Clin Nephrol Res 2(2): 1020 (2015) 2/5

membranes are the most important criteria for successful long-term HD [9,10]. The polysulfone (PS) membrane is the mainstay of HD treatment because of its high performance. Most PS membranes are hydrophilized by polyvinylpyrrolidone (PVP) to improve biocompatibility; however, platelets still adhere to the surface of the membrane and activate platelets.

In 2011, a new PS HD membrane coated with a new hydrophilic polymer was developed, which focused on the mobility of adsorbed water close to the membrane surface. Recently, Yamaka et al. reported that platelet activation and adhesion to this membrane were lower than with a traditional PS membrane [11]. Using 2 different types of PS membrane, the traditional and a new, we evaluated differences in biocompatibility by analyzing the level of PDMPs and observing endothelial dysfunction via flow-mediated dilatation (FMD) of the brachial artery [12,13].

MATERIALS AND METHODS

Study design

Twenty-four stable chronic-maintenance HD patients who were receiving dialysis for 4 hours 3 times a week were enrolled. Inclusion criteria were age 50–85 years and clinically stable health condition; hence, all patients had a fistula, and none had a catheter. The dialyses flow rate was 500 mL/min. Patients with malignancy, chronic inflammatory diseases, hematological disorders, or severe liver or lung diseases were excluded.

All patients underwent HD with an APS-SA membrane (Asahi Kasei Medical, Tokyo, Japan), which is composed of a PS membrane, for 6 months as a baseline. Then, the patients were randomized to one of 2 groups: HD for an additional 3 months with an APS-SA membrane or an NV-U membrane (Toray Medical, Tokyo, Japan), a new PS membrane coated with a new hydrophilic polymer. Randomization was conducted using a table of random numbers. Table 1 shows the clinical data of the patients. Age, sex, HD duration, comorbidities, ankle-brachial index, percentage of habitual smoking, and blood pressure were all similar between the 2 groups.

During the study, with the exception of the dialysis membrane, we did not change any other HD condition such as dry weight, blood and dialysate flow rate, dosage or type of anticoagulant used, or the type of dialysate. We also fixed the medications for hypertension, hyperlipidemia, and diabetes mellitus. Blood samples were taken from a peripheral vein at the first HD session of the week at baseline and after 3 months in both groups. The level of PDMPs and FMD were also examined at these same time points.

All subjects gave informed consent, and this study was performed in accordance with the Declaration of Helsinki.

Measurement of PDMP levels

An enzyme-linked immunosorbent assay (ELISA) kit (JIMRO, Tokyo, Japan) was used for the detection of PDMPsas described previously [14]. Briefly, the blood was drawn directly from the vascular access and collected in vacuous tubes containing citrate/ethylenediaminetetraacetic acid (Nipro, Osaka, Japan). The samples were gently mixed by inverting the tubes upside down once or twice and then kept at room temperature for

2–3 hours, followed by centrifugation at 8000 g for 5 minutes at room temperature. Thereafter, 200 µL of the upper layer of supernatant was collected from a 2-mL sample to avoid contamination by platelets. The collected samples were stored at -40°C until analysis. The kit employs 2 monoclonal antibodies directed against platelet GPs, CD42b, and CD42a (GP Ib and IX). The levels of PDMPs were measured twice and mean values were recorded. One U/mL of PDMPs was defined as the amount of PDMPs obtained from 24,000 solubilized platelets /mL in this ELISA system. The performance of this kit has obtained suitable reproducibility such as simultaneous CV (1.1 – 4.0 %) and daily CV (5.2 – 8.8%).

Endothelium-dependent vasodilatation (via FMD)

FMD of the brachial artery was assessed noninvasively using high-resolution ultrasound as described previously in detail [15]. Using a 10-MHz linear-array transducer probe (UNEX, Nagoya, Japan), longitudinal images of the brachial artery on the arm opposite the vascular access were recorded at baseline, and artery diameter was measured after rest in the supine position for >5 minutes. Then, suprasystolic compression (50 mmHg higher than systolic blood pressure) was performed at the same side forearm for 5 minutes, and measurements of artery diameter were performed continuously from 30 seconds until >2 minutes after cuff release. Maximum vasodilatation was evaluated from the change in artery diameter after release of occlusion (%FMD).

Statistical analyses

Normally distributed data are expressed as the mean and standard deviation (SD). Non-normal data are expressed as the median and inter-quartile range (IQR). Categorical variables were compared using the chi-square test or Fisher exact test. The Wilcoxon test was used to compare the 2 groups. Comparisons between data at baseline and after 3 months were performed with the paired Student t test or Wilcoxon signed-rank test. All statistical analyses were performed with JMP program version 10 (SAS Institute Inc., Cary, NC). A p value < 0.05 was considered statistically significant.

RESULTSA total of 24 patients were treated after the observation

period, and all of them completed the study for 3 months without any side effects.

Baseline laboratory data and dialysis efficiency

Table 2 shows the laboratory data, PDMPs levels, and %FMD at baseline. The hemoglobin level was higher in the NV-U group, and the platelet count was higher in the APS-SA group. The other data showed no significant differences between the 2 groups. Laboratory data were not significantly different after 3 months (data not shown).

Change in PDMPs level at baseline and after 3 months

Figure 1a shows the change in PDMPs level in the NV-U and APS-SA groups. The PDMPs level was significantly reduced by dialysis with the NV-U membrane, from 16.7+8.1 to 15.0+6.1 U/mL after 3 months of treatment (p = 0.03), whereas in the APS-SA group, no significant change in PDMPs level was observed

Page 3: Hydrophilic Polymer-Coated Polysulfone … this article: Hidaka S, Kobayashi S,Maesato K, Mochida Y, Ishioka K, et al. (2015) Hydrophilic Polymer-Coated Polysulfone Membrane Improves

Central

Hidaka et al. (2015)Email:

J Clin Nephrol Res 2(2): 1020 (2015) 3/5

(11.8+2.2 to 12.1+3.2 U/mL, p = 0.76). The difference in PDMPs level was defined as the PDMPs level after 3 months minus the baseline level. It was -1.80+2.59 U/mL in the NV-U group, whereas in the APS-SA group, 0.30+3.30 U/mL. This value was lower in the NV-U group (p = 0.049) (Figure 1b).

Change in percentage FMD before and after 3 months

Figure 2 shows the change in %FMD in the NV-U and APS-SA groups. Percentage FMD significantly increased by dialysis with the NV-U membrane, from 3.0+1.8% to 4.1+1.9% after 3 months of treatment (p = 0.04). When the APS-SA dialyzer was used, no significant change in %FMD was observed (2.3+1.5% to 2.7+0.9%, p = 0.62). There were no significant differences in %FMD at baseline and after 3 months between the 2 groups.

DISCUSSIONIn this study, we demonstrated that use of an NV-U membrane

for 3 months was associated with a significant decrease in PDMPs and a significant increase in FMD; in contrast, HD with the traditional PS membrane did not show these favorable changes. Differences in biocompatibility might account for the reduction in PDMPs associated with the NV-U membrane.

Oxidative stress is generated when blood components adhere to the surface of HD membranes [3,16]. The NV-U membrane was developed with the goal of reducing the adherence of blood components to the PS membrane [11]. As a result of the new hydrophilic polymer coating, platelets are much less able to adhere to the surface. Therefore, the NV-U membrane is a biocompatible HD membrane with a high membrane performance that considerably prevents fibrinogen and platelets from adsorbing to the surface.

PDMPs promote the expression of adhesion molecules via monocytes and endothelial cells and contribute to the development and progression of atherosclerosis [17]. High shear stress can initiate both platelet aggregation and shedding of procoagulant-containing PDMPs, suggesting that the generation of PDMPs occurs in small diseased arteries and arterioles. The PDMPs levels of HD patients in this study were higher than those of patients with angina (10.8 ± 8.0 U/mL), according to the study by Namba[18]., indicating that patients undergoing HD have a higher risk of thromboembolic and atherosclerotic events than patients with angina. Namba et al. revealed that a high level of PDMPs was an independent predictor for secondary thrombotic events. Because the NV-U membrane activates platelets to a lesser extent than the APS-SA membrane, the level of PDMPs is reduced; therefore, using this HD membrane might be one salient method of preventing thromboembolic conditions.

Endothelial dysfunction, assessed by %FMD of the brachial artery, is thought to be a marker of vascular damage and/or a predictor of further cardiovascular events [12,13]. Kosch et al. reported a decrease in %FMD after dialysis with a cellulosic cuprophane membrane, but not with a synthetic PS membrane [19]. They mentioned that a reduction in serum vitamin E level affected the result. In this study, we did not measure oxidative stress, but the improvement in %FMD associated with use of the NV-U membrane might be associated with a reduction in PDMPs level.

There were several important limitations to our study. First, this was a single center, open-label, and small-scale study. We performed this study as a pilot study. Therefore, a degree of

0

4

8

12

16

20

24

APS-SA NV-U

before

after(U/mL)

p=0.76

p=0.03

Figure 1a Change in plasma PDMP level at baseline and after 3 months. Plasma PDMPs level was significantly reduced by dialysis with the NV-U membrane (p = 0.03). When the APS-SA dialyzer was used, no significant change in plasma PDMPs level was observed (p = 0.76).

Figure 1b The difference in PDMPs levels after using each dialyzer for 3 monthsThe difference in PDMP level (defined as the PDMP level after 3 months minus the baseline level) was significantly lower in the NV-U group (p = 0.49).

0

1

2

3

4

5

APS-SA NV-U

before

afterp=0.41 p=0.04(%FMD)

Figure 2 Change in %FMD at baseline and after 3 months%FMD was significantly increased by dialysis with the NV-U membrane (p = 0.04). When the APS-SA dialyzer was used, no significant change in %FMD was observed (p = 0.62).

Page 4: Hydrophilic Polymer-Coated Polysulfone … this article: Hidaka S, Kobayashi S,Maesato K, Mochida Y, Ishioka K, et al. (2015) Hydrophilic Polymer-Coated Polysulfone Membrane Improves

Central

Hidaka et al. (2015)Email:

J Clin Nephrol Res 2(2): 1020 (2015) 4/5

patient selection bias might have occurred. Second, exercise and diet therapy as well as several drugs often affect FMD. We did not directly assess exercise habits and lifestyle in this study, but we met the patients thrice weekly, and their laboratory data, such as serum HDL-cholesterol levels, did not change significantly during the study period (data not shown). Therefore, the possibility of these factors influencing FMD could be ruled out.

CONCLUSIONA newly developed PS membrane coated with hydrophilic

polymer improves the level of PDMPs and endothelial function. These improvements might have desirable effects on preventing the development and progression of atherosclerosis in HD patients.

REFERENCES1. Kahn MR, Robbins MJ, Kim MC, Fuster V. Management of cardiovascular

disease in patients with kidney disease. Nat Rev Cardiol. 2013; 10: 261-273.

2. Levey AS, Eknoyan G. Cardiovascular disease in chronic renal disease. Nephrol Dial Transplant. 1999; 14: 828-833.

3. Nusair MB, Rajpurohit N, Alpert MA. Chronic Inflammation and Coronary Atherosclerosis in Patients with End-Stage Renal Disease. Cardiorenal Med. 2012; 2: 117-124.

4. Daniel L, Fakhouri F, Joly D, Mouthon L, Nusbaum P, Grunfeld JP, et al. Increase of circulating neutrophil and platelet microparticles during acute vasculitis and hemodialysis. Kidney Int 2006; 69: 1416-1423.

5. Nomura S, Ozaki Y, Ikeda Y. Function and role of microparticles in various clinical settings. Thromb Res. 2008; 123: 8-23.

Table 1: Basic clinical characteristics of hemodialysis patients using the APS-SA and NV-U dialyzer membranes.

Parameter APS-SA(n = 12)

NV-U(n = 12) p-value

Age (years) 70.6+8.5 72.0+9.5 0.58

Male/Female 8/4 9/3 > 0.99

HD duration (months) 90.9+95.1 138.0+28.3 0.84

Hypertension (%) 83.3 91.7 > 0.99

Diabetes mellitus (%) 41.7 33.3 > 0.99

Angina pectoris/OMI (%) 16.7 33.3 0.64

CVA (%) 16.7 0 0.48

ABI 1.08+0.15 1.16+0.10 0.31

Ex and current smoker (%) 66.7 66.7 >0.99

Dialyzer surface area (m2) 1.88+0.29 1.89+0.26 0.63

Blood flow rate (mL/min) 215 + 30 221+35 0.84

Dry weight (Kg) 55.0+7.8 55.3+8.5 0.91

Systolic BP (mmHg) 136.2+25.9 136.2+17.4 0.9

Diastolic BP (mmHg) 70.3+12.1 69.0+9.9 0.77Data are expressed as mean+SD. HD, hemodialysis; OMI, old myocardial infarction; CVA, cerebral vascular accident; ABI, ankle-brachial index; BP, blood pressure. p- values determined by Wilcoxon and chi-square test.

Table 2: Laboratory data of hemodialysis patients using the APS-SA and NV-U dialyzer membranes at baseline.

Parameter APS-SA (n = 12)

NV-U(n = 12) p-value

Scr (mg/dL) 10.98+1.46 10.25+2.61 0.54

UN (mg/dL) 65.6+10.8 61.9+13.1 0.4

β2-mg (mg/L) 25.7+5.4 27.7+5.8 0.45

kT/v 1.49+0.20 1.55+0.15 0.54

Hb (g/dL) 10.7+1.4 11.7+1.3 0.03*

WBC (/µL) 5858+1692 4575+1642 0.07

Plt (x104/µL ) 16.8+3.1 12.6+4.0 0.01*

CRP (mg/dL) 0.18 (0.01-0.76) 0.09 (0.03-0.26) 0.75

PDMPs (U/mL) 11.4 (10.4-13.3) 13.9 (9.5-23.4) 0.16

FMD (%) 2.6 (0.7-3.7) 3.2 (1.2-4.8) 0.3Data except CRP, PDMPs level, and %FMD are expressed as mean+SD. Serum CRP level, PDMP level, and %FMD are expressed as median (IQR). Scr, serum creatinine; UN, urea nitrogen; β2-mg, β2-microglobulin; Hb, hemoglobin; WBC, white blood cell; Plt, Platelet count; CRP, C-reactive protein; PDMPs, platelet-derived microparticles; FMD, flow-mediated dilatation. Serum CRP levels were analyzed statistically after converting logarithmic value. p- values determined by Wilcoxon test. *p < 0.05 between the APS-SA group and the NV-U group.

Page 5: Hydrophilic Polymer-Coated Polysulfone … this article: Hidaka S, Kobayashi S,Maesato K, Mochida Y, Ishioka K, et al. (2015) Hydrophilic Polymer-Coated Polysulfone Membrane Improves

Central

Hidaka et al. (2015)Email:

J Clin Nephrol Res 2(2): 1020 (2015) 5/5

Hidaka S, Kobayashi S,Maesato K, Mochida Y, Ishioka K, et al. (2015) Hydrophilic Polymer-Coated Polysulfone Membrane Improves Endothelial Function of He-modialysis Patients: A Pilot Study. J Clin Nephrol Res 2(2): 1020.

Cite this article

6. Tan KT, Tayebjee MH, Lynd C, Blann AD, Lip GY. Platelet microparticles and soluble P selectin in peripheral artery disease: relationship to extent of disease and platelet activation markers. Ann Med. 2005; 37: 61-66.

7. Ridker PM, Buring JE, Rifai N. Soluble P-selectin and the risk of future cardiovascular events. Circulation. 2001; 103: 491-495.

8. Inami N, Nomura S, Takahashi N, Isami Y, Nakamura E, Tsuda N, et al. Correlation between platelet-derived microparticles and soluble L-selectin in patients undergoing hemodialysis. Thromb Haemost 2004; 92: 1452-1454.

9. Hakim RM, Held PJ, Stannard DC, Wolfe RA, Port FK, Daugridas JT, et al. Effect of the dialysis membrane on mortality of chronic hemodialysis patients. Kidney Int 1996; 50: 566-570.

10. Leypoldt JK, Cheung AK, Carroll CE, Stannard DC, Pereira BJ, Agodoa LY, et al. Effect of dialysis membranes and middle molecule removal on chronic hemodialysis patient survival. Am J Kidney Dis. 1999; 33: 349-355.

11. Yamaka T, Ichikawa K, Saito M, Watanabe K, Nakai A, Higuchi N, et al. Biocompatibility of the new anticoagulant dialyzer TORAYLIGHT?R NV. Science Postprint 2014; 1.

12. Widlansky ME, Gokce N, Keaney JF Jr, Vita JA. The clinical implications of endothelial dysfunction. J Am Coll Cardiol. 2003; 42: 1149-1160.

13. Moens AL, Goovaerts I, Claeys MJ, Vrints CJ. Flow-mediated vasodilation: a diagnostic instrument, or an experimental tool? Chest.

2005; 127: 2254-2263.

14. Nomura S, Shouzu A, Taomoto K, Togane Y, Goto S, Ozaki Y, et al. Assessment of an ELISA kit for platelet-derived microparticles by joint research at many institutes in Japan. J Atheroscler Thromb. 2009; 16: 878-887.

15. Tomiyama H, Matsumoto C, Yamada J, Teramoto T, Abe K, Ohta H, et al. The relationships of cardiovascular disease risk factors to flow-mediated dilatation in Japanese subjects free of cardiovascular disease. Hypertens Res. 2008; 31: 2019-2025.

16. Miyazaki H, Matsuoka H, Itabe H, Usui M, Ueda S, Okuda S, et al. Hemodialysis impairs endothelial function via oxidative stress: effects of vitamin E-coated dialyzer. Circulation. 2000; 101: 1002-1006.

17. Nomura S, Tandon NN, Nakamura T, Cone J, Fukuhara S, Kambayashi J. High-shear-stress- induced activation of platelets and microparticles enhances expression of cell adhesion molecules in THP-1 and endothelial cells. Atherosclerosis 2001; 158: 277–287.

18. Namba M, Tanaka A, Shimada K, Ozeki Y, Uehata S, Sakamoto T, et al. Circulating platelet-derived microparticles are associated with atherothrombotic events: a marker for vulnerable blood. Arterioscler Thromb Vasc Biol. 2007; 27: 255-256.

19. Kosch M, Levers A, Fobker M, Barenbrock M, Schaefer RM, Rahn KH, et al. Dialysis filter type determines the acute effect of haemodialysis on endothelial function and oxidative stress. Nephrol Dial Transplant. 2003; 18: 1370-1375.