Topics 15 O-rapid PET for stroke research SPECT for multicenter clinical studies Future directions Quantitative assessment of biological functional parametric images using PET and SPECT Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular Center - Research Institute Osaka, Japan JAPANESE-FINNISH JOINT SYMPOSIUM ICT and Technology in Medical and Health Research 13–15 December 2011, Helsinki, Finland

Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

  • Upload
    others

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future directions

Quantitative assessment of biological functional parametric images using PET and SPECT

Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular Center - Research Institute

Osaka, Japan

JAPANESE-FINNISH JOINT SYMPOSIUM ICT and Technology in Medical and Health Research

13–15 December 2011, Helsinki, Finland

Page 2: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

National Cerebral & Cardiovascular Research Center Ministry of Health, Labor and Welfare (MHLW), Japan

Pre-clinical PET

Clinical PET

NCVC

Hospital Research Institute

Development and Research Initiative Center

Page 3: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

Imaging devices for preclinical research at NCVC GE 3T MRI

Signa3T

Siemens micro PET Focus120

SPECT (inhouse)

Fuji Fuluoro image analyzer

Siemens PET ECAT Accel

Toshiba SPECT

GCA7200A

Angiography

Cyclotron

Hot Labo BioScan

NanoSPECT

Our activity includes application of imaging techniques to preclinical research, and also development of equipment and novel software.

Phantom

Blood radioactivity monitor

Page 4: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

Uniqueness of PET/SPECT

Functional (biological) imaging Physiological functions

▫ perfusion, metabolism, ▫ Receptor BP & occupancy

Tracing therapeutic compound (DDS) Quantitative assessment Kinetic modeling (math, physics)

High sensitivity

Page 5: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

15O-PET for stroke diagnosis 15O-PET provides useful information in relation to ischemic status and neuron survival, but has been limited in clinical environment.

Attributed to: • The present technique is based on the 3-step

approach, requiring 15O2, C15O, and C15O2 (H215O)

administration with TCT, thus long (>1 hr) examination.

• Short half life of 2min, requiring on-site cyclotron, on-site radio-tracer synthesis, and QC for each administration

• Arterial blood sampling needed for determination of arterial input function, which is labor intensive

Page 6: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

15O-PET for clinical diagnosis of stroke

•Cerebral blood flow (CBF) for ischemic status

•Cerebral metabolic rate of Oxygen (CMRO2) for neuron survival

•Study consists of CO2 (H215O), 15O2 and C15O inhalation PET scannings

•Short half life of 15O (approx. 2 min) requires highly advanced logistics

Before and after MCA bypass operation

Page 7: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

15O-PET on cynomolgus monkeys Assessment of irreversible area in acute embolic stroke

Angiography CBF CMRO2 OEF CBV MRI

Pre

M1 throm

M3 throm

JCBFM; 25: 1209–24, 2005

Page 8: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

Goal of NCVC 15

O-PET project

• Single, short PET scan <10 min, for simultaneous quantitation of CMRO2, CBF, CBV, OEF

• Dedicated cyclotron

• Fully automated, united system for rapid production, radio-synthesis, and QC for 15O2, C15O, C15O2 (H2

15O)

• Improved quantitative accuracy in high-sensitive 3D PET

• Quantitation without the arterial blood sampling

“On-demand” PET

Page 9: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

15O-CO Scan 15O-H2O Scan 15O-O2 Scan TCT

PET Protocols for 15O2-based CMRO2/CBF

CBV CBF OEF

CMRO2

Three-step approach (Mintun et al 1984)

Ao(t)

Aw(t) Input Function

Brain TAC

Scan

CMRO2

CBF OEF CBV

15O-O2 15O-H2O (C15O2)

X-CT

Kudomi et al. JCBFM 2005

Rapid 15O-PET approach

Page 10: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

Automated radiosynthesis/QC

PET/CT

Data analysis •Well system •Continuous input func. •Gas anal. etc

15O-dedicated cyclotron

Fully automated united system

•Reconstruction scatter corr. atten corr. random comp. •Eqipments

Automated rapid system 15O2, H2

15O, C15O2, C15O

(2 min half life)

Ultra-Rapid 15O-Gas PET System for acute stroke

Towards “on-demand” PET

Page 11: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

14N(d,n)15O E>3.0 MeV

16O2 in N2 d

15O-16O (15O-oxygen)

Target

15O2 5%H2 in N2

Electric Furnace

H215O vapour

H215O saline

Saline

Production, synthesis, & QC for 15O-PET

H215O infuser

15O2/ C15O2/C15O

QC & inhalation

mCT

15O-dedicated cyclotron

Deuteron at 3.5 MeV

Automated synthesis

Page 12: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

Lung activity Only in blood for C15O2

5 times higher in gas for 15O2

Ventilation(F) 4-20L/min

Clung = F/(F+λ)

Supply 1 L/min 800 MBq/min Air 26L/min

Challenges in 15O2 inhalation PET

Strong radioactivity in the gas supply and in the lung causes

• High single rate, resulting in a large dead-time loss

• High random coincidence rate, resulting in poor image quality

• High scatter events, resulting in inaccurate quantitation

Essential problem in 3D-PET!!

Page 13: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

C15O DARG

(15O2 + C15O2)

Physiological model for metabolite Reconstruction from images

Vasculature radioactivity information for non-invasive arterial input function

External detector

15O2

Quantitative assessment in Rapid 15O-PET 3D-PET (Biograph-mCT, Siemens)

Page 14: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

Medical Physicist’s Brain - CMRO2

study date : 20110208

Page 15: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

Clinical trials for patients with acute stroke using rt-PA with additional devices

• 0.6 mg/kg of Alteplase extending to 4.5 hrs

• Desmoplase (<9 hrs) DIAS III, DIAS-Japan

• Stroke MRI (PDM, MRA)… J-ACT II

• Neuroprotectives…NXY-059, Edaravone

• US-Thrombolysis…CLOTBUST, tPA/US/MB,

TCT-LoFUT

• Devices(Fructure, removal)

• Local rt-PA(alone, combined)… MELT

• Others

Page 16: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

Role of PET in stroke ・Pre-operative evaluation of ischemic status &neuron survival

at >3hrs of onset (Rapid 15O-PET required)

・Identification of effective care

Neuron-, glial cell-, microglia-, ion channel-specific pathophysiology

・ Pre-operative prediction of hemorhage

Ultra-rapid,

automated,

on-demand PET

for acute stroke

United system

(cyclotron, radiosynthesis,

acquisition, data processing,

data analysis)

New indecies

CBF

CBV

OEF

Molecular imaging

Hayashi T, et al. JCBFM 2002

Effective oxygen diffusivity

0

50

100

150

200

250

300

350

400

450

500

0 10 20 30 40 50 60

On-going research

Understanding pathophysiol

and treatment outcome

Page 17: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

Use of SPECT in clinical research

• Large scale availability in clinical institutions • Variety of radio-tracers available • Inter-institutional & inter-equipment reproducibility

well verified • But quantitative accuracy not well supported

A novel software (QSPECT) developed at NCVC enables quantitative SPECT reconstruction for most of existing SPECT cameras. This software is considered to make SPECT suitable for multicenter clinical studies.

Page 18: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

Quantitative reconstruction

Functional parametric calculation

Quantitative SPECT (QSPECT) Reconstruction

Functional parametric images

>12k clinical scans/year at 170 institutions

Page 19: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

100

%

0

Rel

ativ

e C

BF

Institute A Rest

Institute C Rest

3 institutes Rest

Institute B Rest

Inter-institutional consistency of SPECT-CBF Averaged at each institution

Each institute with different SPECT systems

Page 20: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

Application of QSPECT On going multicenter clinical research

• Outcome evaluation of EC-IC bypass surgery in MCA occlusion (Japan)

• Drug evaluation (cerebral flow reactivity) (Japan)

• Traumatic brain injury (Japan)

• Pre-synaptic dopamine reuptake – normal data base (EU)

• Evaluation of carotid artery stent graft

• others

Page 21: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

• Multiple functional and anatomical images

• Shortening the positron range for improved spatial resolution

Hybrid PET/MRI System

E=mc2 (511 keV)

E=mc2 (511 keV)

Positron Range (sub millimeter)

Positron Emitter (eg 11C, 13N, 15O, 18F, etc)

Page 22: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

Reduction of positron range in high-field MRI

-10 10 mm 0 -10

10

0

-10 10 mm 0 -10

10

0

-10 10 mm 0 -10

10

0

-10 10 mm 0 -10

10

0

Iida et al., IEEE TNS, 1986

15O

Page 23: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

0 T

3 T

7 T

9.4 T

3D-Brain Phantom in Magnetic Field 120I (Maximum energy : ~ 4 MeV)

Jeulich Institute, Germany

Page 24: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

Future directions in PET/SPECT

Multi-modality imaging Software fusion CT/PET, CT/SPECT MRI/PET, MRI/SPECT

Shortening scan duration & time resolution Multiple functional imaging Transient change detection

Standardization among institutions/equipment setup

Page 25: Hidehiro Iida, PhD, DSc National Cerebral and Cardiovascular … · 2012. 2. 16. · Topics • 15O-rapid PET for stroke research • SPECT for multicenter clinical studies • Future

Summary

• Rapid 15O-PET is being setup for stroke

research, but still requires further works.

• SPECT is of use in multicenter clinical study

• Multi-modality, multi-functional imaging is

future direction.