Hemocompatibility of Surface Modified Si
Incorporated Diamond-like Carbon Films
R. K. Roy, S.-J. Park, H.-W. Choi, K.-R. Lee
Future Technology Research Laboratories, KIST, Seoul, Korea
T. HasebeTachikawa Hospital, Keio University, Tokyo, Japan
ICMCTF-2007, Apr. 23-27, San Diego, USA
Vascular Stents
• Formation of blood clots Restenosis• Release of metal ions
Hemocompatible and Hermetic CoatingHemocompatible and Hermetic Coating
Clotted Artery
Surface Modification
• Biocompatible Coating : Heparin, PEG, DLC (Hepacoat, Phytis)• Drug Release Coating : Antistenosis, Anticancer, Antibiotic (Cordis)• Isotope Radiation Coating : Radiation therapy
Si-DLC Film
10-14 10-13 10-12 10-11 10-10 10-9 10-8 10-7 10-6 10-5 10-4 10-3 10-2 10-1 100 101
-600
-400
-200
0
200
400
600
800
1000
1200
1400
1600
Po
ten
tial (
mV
vs
SC
E)
Current Density (A/cm2)
Substrate Si-C:H, Bias voltage = -400V a-C:H, Bias voltage = -800V a-C:H, Bias voltage = -400V
Potentiodynamic Polarization
Thin Solid Films, 475, 291-397 (2005)
Sl.No.
ReferencesHemocompatibility
Improves by
1Baier,
Academic Press,New York, 1970.
Critical surface tension of materials ~ 20-30dyne/cm
2Akers,
J.Colloid Interface Sci. 59 (1977) 461.
Zone of biocompatibility
3Ruckensten & Gourisanker,J. Colloid Interface Sci. 101
(1984) 436.
Blood biomaterial interfacial tension of
the order of 1-3 dyne/cm
4
Callow,International
Biodeterioration & degradation, 34 (1994) 333.
Surfaces having initial surface tension
20-30 dyne/cm
5Yu,
Surf. Coat. Technol.128-129 (2000) 484.
Low blood biomaterial interfacial tension
(8.5 dyne/cm)
6Kwok,
Diam. Rel. Mater. 14 (2005) 78.
interfacial tension of about the same magnitude as cell-medium interfacial
tension (1-3 dyne/cm)
Hemocompatibility and Surface Tension
Surface modification of Si-DLC
20
40
60
80
100W
ater
co
nta
ct a
ng
le (
in d
egre
e)
SiDLC( O2treated)
SiDLC(N2treated)
SiDLC(H2treated)
SiDLC(CF4treated)
SiDLCSi Nitinol
Schematic diagram of RF PACVD system.
Film Preparation
• Film Deposition– C6H6 + SiH4
– Pressure : 1.33 Pa– Bias voltage : -400V– Film thickness : ~500nm – Si Concentration in the
film : 2 at.%
• Surface Treatment– O2, N2, H2, CF4
– Pressure : 1.33 Pa– Bias voltage : -400V– 10min
Surface modification of Si-DLC
20
40
60
80
100W
ater
co
nta
ct a
ng
le (
in d
egre
e)
SiDLC( O2treated)
SiDLC(N2treated)
SiDLC(H2treated)
SiDLC(CF4treated)
SiDLCSi Nitinol
Energetics of Surface
(cos ) lv sv sl
22ll
pl
dllv
22ss
ps
dssv
)(2
cos1
)(2)(2cos1
lslslv
lv
pl
ps
lv
dl
ds
Liquid αl βl
γlv
(ergs/cm2)
Water 4.67 7.14 72.8
Formamide 6.28 4.32 58.2
2 : Fowkes' Eq.d dsl sv lv s l
Surface Energy
Polar component
0
10
20
30
40
50
60
SiDLC SiDLC (O
2
treated)
SiDLC (N
2
treated)
SiDLC (H
2
treated)
SiDLC (CF4
treated)
Sur
face
ene
rgy
(dyn
e/cm
)
Dispersive component
Polar Component and Wetting
0 10 20 30 40 50 600
20
40
60
80
100W
ater
co
nta
ct a
ng
le (
deg
ree)
Polar component of surface energy (dyne/cm)
Interfacial Tension with Human Blood
α (dyne/cm)1/2
β (dyne/cm)1/2
Human Blood
3.3 6.0
221
22112 )()(
0
5
10
15
20
25
SiDLC(O2
treated)
SiDLC(N2
treated)
SiDLC(H2
treated)
Blo
od B
iom
ater
ial i
nter
faci
al t
ensi
on
(dyn
e/cm
)
SiDLC SiDLC(CF4
treated)
α β
Si-DLC 5.4 ± 0.5 3.3 ± 0.6
Si-DLC (CF4 treated) 5.0 ± 0.4 2.0 ± 0.5
Si-DLC (N2 treated) 5.1 ± 0.2 5.5 ± 0.3
Si-DLC (O2 treated) 4.2 ± 0.1 7.3 ± 0.1
Si-DLC (H2 treated) 5.5 ± 0.3 3.5 ± 0.4
100 102 104 106
4
8
12
16
Inte
nsi
ty (
x1
0 )
Binding energy (eV)
Si-C
Si 2pSiDLC
Si2O3
SiO2
(b)
280 284 288 2920
10
20
30
40
50
Inte
nsi
ty (
x1
0 2
)
Binding energy (eV)
C=C
C-C
Si-DLC
C 1s(a)
280 284 288 2920
10
20
30
40
Inte
nsity
( x
10
2 )
Binding energy (eV)
C=C
C-CC-OH
H2-Si-DLC
C 1s(c)
100 102 104 106
4
8
12
16In
tens
ity (
x10
)
Binding energy (eV)
Si-C
Si-Si
Si2O3
Si 2pH2-Si-DLC(d)
XPS Anaysis
280 284 288 2920
5
10
15
20
25
Inte
nsity
( x
10 2 )
Binding energy (eV)
C-CFn
CF-CFn
CF2 CF3
1
2
C 1sCF4-Si-DLC(e)
100 102 104 106
4
6
8
10
Inte
nsi
ty (
x 1
0 )
Binding energy (eV)
Si-C
Si2O3
SiO2
Si-Si
Si 2pCF4-Si-DLC(f)
684 688 692
10
20
30
40
Inte
nsity
( x
10 2
)
Binding energy (eV)
F 1s
CF4-Si-DLCCFn(g)
_
_
100 102 104 106
5
10
15
20
25
Inte
nsity
(x1
0)
Binding energy (eV)
Si-N
Si-CSiO2
Si 2pN2-Si-DLC(i)
280 284 288 2920
5
10
15
20
25In
tens
ity (
x 1
0 2
)
Binding energy (eV)
C 1sN2-Si-DLCC=C
C=N
C=N
C-C
C-N
(h)
394 396 398 400 402 4044
8
12
16
Inte
nsity
( x
10
2 )
Binding energy (eV)
N 1s
N2-Si-DLC
C-N
C=N
N1
N2
(j)
N1 : Si-NN2 : C=N
280 284 288 2920
4
8
12
16
Inte
nsity
( x
10 2
)
Binding energy (eV)
C=C
C-C C-O C=O
C 1sO2-Si-DLC(k)
100 102 104
20
40
60
80
Inte
nsi
ty (
x10
)
Binding energy (eV)
Si 2pO2-Si-DLC
Si2O3
SiO2
Si-C
(l)
528 530 532 534 536
10
20
30
40
50
Inte
nsi
ty (
x1
0 2
)
Binding energy (eV)
O 1sO2-Si-DLC
Si=O
C=O
(m)
_
_
100 102 104 106
4
8
12
16
Inte
nsi
ty (
x1
0 )
Binding energy (eV)
Si-C
Si 2pSiDLC
Si2O3
SiO2
(b)
280 284 288 2920
10
20
30
40
50
Inte
nsi
ty (
x1
0 2
)
Binding energy (eV)
C=C
C-C
Si-DLC
C 1s(a)
XPS Anaysis
280 284 288 2920
4
8
12
16
Inte
nsity
( x
10 2
)
Binding energy (eV)
C=C
C-C C-O C=O
C 1sO2-Si-DLC(k)
100 102 104
20
40
60
80
Inte
nsity
(x1
0)
Binding energy (eV)
Si 2pO2-Si-DLC
Si2O3
SiO2
Si-C
(l)
XPS Analysis
FilmsChemical bonds present on surface
(XPS analysis)
Si-DLC C=C, C-C, Si-C, Si-O
Si-DLC(CF4 plasma treated)
C=C, C-C, C-CFn, Si-C, Si-O
Si-DLC(N plasma treated)
C=C, C-C, C-N, Si-N, Si-O
Si-DLC(H plasma treated)
C=C, C-C, Si-C, Si-O
Si-DLC(O plasma treated)
C=C, C-C, C-O, Si-O
aPTT Measurement
• Soaking for 60min in platelet poor plasma (PPP: 7x103/l) using human whole blood from healthy volunteer
• aPTT measurement system by Sysmex Instrument
25
26
27
28
29
30
SiDLC(O2
treated)
SiDLC(N2
treated)
SiDLC(H2
treated)
SiDLC(CF4
treated)
aPT
T (
sec)
1 hour
Plasma Protein Adsorption
• Done by treating the samples with albumin (3mg/ml) and fibrinogen (0.2mg/ml) solution.
• ELISA analysis method to characterize the proten adsorption
0.0
0.5
1.0
1.5
2.0
SiDLC (H2
treated)
SiDLC(O2
treated)
SiDLC(N2
treated)
SiDLC (CF4
treated)
SiDLC
Alb
umin
/ f
ibri
noge
n ra
tio
5 mins
60 mins
Platelet Adhesion Measurement
• PRP (1.5x1015/ml) from human whole blood from healthy volunteer• Soaked in PRP for 60min • Adherent platelet are fixed and dehydrated for observation under OM
and SEM
Si-DLC Si-DLC(CF4) Si-DLC(N2) Si-DLC(O2)0
10
20
30
40
50
60
70
80
90
100
Pla
tele
t Ad
he
sio
n A
rea
Ra
tio (
%)
Specimen
Platelets on Surface
Goodman and Allen et al.
On a-C:H surface
Platelets on Si-DLC
Platelets on Si-DLC (CF4)
Platelets on Si-DLC (N2)
Platelet on Si-DLC (O2)
Nitrogen or Oxygen Plasma Treatment
25
26
27
28
29
30
SiDLC(O2
treated)
SiDLC(N2
treated)
SiDLC(H2
treated)
SiDLC(CF4
treated)
aPT
T (
sec)
1 hour
0.0
0.5
1.0
1.5
2.0
SiDLC (H2
treated)
SiDLC(O2
treated)
SiDLC(N2
treated)
SiDLC (CF4
treated)
SiDLC
Alb
umin
/ fib
rinog
en r
atio
5 mins
60 mins
Si-DLC Si-DLC(CF4) Si-DLC(N2) Si-DLC(O2)0
10
20
30
40
50
60
70
80
90
100
Pla
tele
t A
dh
esi
on
Are
a R
atio
(%
)
Specimen
Sl.No.
ReferencesHemocompatibility
Improves by
1Baier,
Academic Press,New York, 1970.
Critical surface tension of materials ~ 20-30dyne/cm
2Akers,
J.Colloid Interface Sci. 59 (1977) 461.
Zone of biocompatibility
3Ruckensten & Gourisanker,J. Colloid Interface Sci. 101
(1984) 436.
Blood biomaterial interfacial tension of the order of 1-3 dyne/cm
4Callow,
International Biodeterioration & degradation, 34 (1994) 333.
Surfaces having initial surface tension20-30 dyne/cm
5Yu,
Surf. Coat. Technol.128-129 (2000) 484.
Low blood biomaterial interfacial tension (8.5 dyne/cm)
6Kwok,
Diam. Rel. Mater. 14 (2005) 78.
interfacial tension of about the same magnitude as cell-medium interfacial
tension (1-3 dyne/cm)
Interfacial Tension?
0
5
10
15
20
25
SiDLC(O2 treated)
SiDLC(N2 treated)
SiDLC(H2 treated)
Blo
od B
iom
ater
ial i
nter
faci
al te
nsio
n (d
yne/
cm)
SiDLC SiDLC(CF4 treated)
Conclusions
• Hemocompatibility of Si-DLC film would be improved by surface treatment using nitrogen and oxygen plasma.– Large surface energy (large polar component)– Low interfacial energy with blood
Acknowledgement