Effect of hemocompatibility on the surface properties of Si incorporated diamond like
carbon films.
R. K. Roy*, S. J. Park*, K.-R. Lee*, D. K. Han**, J.-H. Shin+,
*Future Technology Division, Korea Institute of Science and Technology **Biomaterials Research Center, Korea Institute of Science and
Technology, +Dept. Radiology, Asan Medical Center
Contents:
1. Introduction
2. Experimental
3. Results and Discussion: Surface energetics relation Wettability Blood biomaterial interfacial tension Activated partial thromboplastin time (aPTT) Protein adsorption tests Platelet adhesion and activation XPS analysis
4. Conclusion
1. Introduction
Heart valve
A medical device in cardiovascular application faces the following problems:
• Thrombogenicity : formation of blood clots• Cytotoxity : release of metal ions • Stability in long term use: delamination and spallation during compressive and tensile forces of medical implants
In this respect, diamond like carbon (DLC) has emerged as a promising coating material for cardiovascular application due to its superior tribological, inert and hemocompatible properties.
Improved hemocompatibility is denoted by • higher albumin/ fibrinogen ratio • higher clotting times • low platelet adhesion and activation
To get rid of delamination and improve hemocompatibility, it is necessary to• dope the surfaces of biomaterials with certain elements • use an interlayer which improves its adhesive properties
2. Experimental
Schematic diagram of RF PACVD system.
Coatings : Si incorporated DLC films.Synthesis method: Capacitively coupled rf plasma assisted chemical vapor deposition. Precursor gases : Benzene, silane.Substrates : Si, nitinol. Bias voltage : - 400 VSystem pressure : 1.33 PascalDeposition time : 12 mins Ar cleaning : - 400 V, 0. 49 Pa, 15 mins.
Surface modification:treating the Si incorporated DLC films with plasma of various gases like O, N, H and carbon tetrafluoride at - 400 V, 1.33 Pa, 10 mins.
Interlayer: Si of thickness 5 nm to ensure better adhesion.
Young’s equation:
eslsvlv )(cos
where, θ is the contact angle between liquid and solid, are the free energies of the liquid and solid against their saturated vapor, is the free energy of the interface between liquid and solid, is the equilibrium pressure of adsorbed vapor of the liquid on the solid.
svlvsle
22ll
pl
dllv
22ss
ps
dssv
where the superscript d and p refer to the dispersion and polar force components.
)(2
cos1
)(2)(2cos1
lslslv
lv
pl
ps
lv
dl
ds
(1)
(2)
(3)
(4)
(5)
Surface Energetics Relations:
dss d
ll pss p
ll where ,, , ,
3. Results and Discussion
Wettebility:
Water contact angles of various surfaces:
Liquidαl βl γlv (ergs/cm2)
Water 4.67 7.14 72.8
Formamide 6.28 4.32 58.2
20
40
60
80
Wat
er c
on
tact
an
gle
(in
deg
ree)
SiDLC( O2
treated)
SiDLC(N2
treated)
SiDLC(H2
treated)
SiDLC(CF4
treated)
SiDLCSi Nitinol
Coatings/substrates
Water contact angle
(degree)
Formamidecontact angle
(degree)
O2 treated SiDLC 13.4 ± 1.3 5.0 ± 1.2
Si 35.3 ± 2.7 12.0 ± 1.5
N2 treated SiDLC 42.7 ± 3.7 10.2 ± 1.3
H treated SiDLC 67.2 ± 1.8 45.7 ± 1.4
SiDLC 70.1 ± 3.0 49.5 ± 1.4
Nitinol 76.6 ± 1.1 51.2 ± 0.7
CF4 treated SiDLC 88.4 ± 2.9 68.0 ± 0.5
Coatings α β
SiDLC 5.4 ± 0.5 3.3 ± 0.6
SiDLC (CF4 treated) 5.0 ± 0.4 2.0 ± 0.5
SiDLC (N2 treated) 5.6 ± 0.2 5.2 ± 0.3
SiDLC (O2 treated) 4.2 ± 0.1 7.3 ± 0.1
SiDLC (H2 treated) 5.5 ± 0.3 3.5 ± 0.4
Surface energy of plasma treated Si incorporated DLC films:
Polar component
0
10
20
30
40
50
60
SiDLC SiDLC (O
2
treated)
SiDLC (H
2
treated)
SiDLC (N
2
treated)
SiDLC (CF4
treated)
Su
rfac
e en
erg
y (n
J/cm
2 )
Dispersive component
α (dyne/cm)1/2 β (dyne/cm)1/2
Human blood 3.3 6.0
Human Fibrinogen 4.97 6.35
Human serum albumin 5.6 5.8
0
5
10
15
20
25
SiDLC(O2
treated)
SiDLC(N2
treated)
SiDLC(H2
treated)
Blo
od B
iom
ater
ial i
nter
faci
al t
ensi
on
(dyn
e/cm
)
SiDLC SiDLC(CF4
treated)
Coatings Total SurfaceEnergy
(nJ/cm2)
SiDLC 39.7 ± 8.9
SiDLC (CF4 treated) 29.1 ± 5.8
SiDLC (N2 treated) 58.8 ± 6.1
SiDLC (H2 treated) 42.1 ± 6.0
SiDLC(O2 treated) 71.0 ± 1.1
Interfacial tensions between the implant (phase 1) and blood plasma (phase2) are given by:
221
22112 )()( (6)
Activated Partial Thromboplastin Time (aPTT)
25
26
27
28
29
30
SiDLC(O2
treated)
SiDLC(N2
treated)
SiDLC(H2
treated)
SiDLC(CF4
treated)
aP
TT
(se
c)
1 hour
36
37
38
39
40
41
Act
ivat
ed p
artia
l thr
ombo
plas
tin t
ime
(aP
TT
) in
sec
.
SiDLC(30 mins)
O2 treated SiDLC
(30 mins)
SiDLC (1 hr)
O2 treated SiDLC
(1 hr)
The aPTT determines the ability of blood to coagulate through the intrinsic coagulation mechanism. It measures the clotting time from the activation of factor XII through the formation of fibrin clot.
Incubation time: 30 mins and 1 hour in platelet poor plasma (PPP).
A higher aPTT was noted in case of O plasma treated SiDLC films.
0
20
40
60
5mins
Am
ou
nt o
f ad
sorb
ed
alb
um
in (g
/cm
2 )
SiDLC (O
2
treated)
60 mins
SiDLC (CF4
treated)
SiDLC (N
2
treated)
SiDLC (H
2
treated)
SiDLC 0
20
40
60
5 mins
Am
ou
nt o
f ad
sorb
ed
fib
rino
ge
n (g
/cm
2)
60 mins
SiDLC SiDLC (CF4
treated)
SiDLC(N2
treated)
SiDLC(O2
treated)
SiDLC (H2
treated)
Plasma Protein Adsorption:
Incubation time: 5 mins, 60 mins.
The plasma protein adsorption tests were done by treating the samples with albumin and fibrinogen solution and measuring the absorbances through ELISA (Enzyme linked immunosorbent assay) analysis.
The O and carbon tetraflouride plasma treated SiDLC films showed increased rate of albumin adsorption with time, while H plasma treated SiDLC films minimized rate of fibrinogen adsorption.
Platelet Adhesion and Activation:
Si
SiDLC
(N treated)
Incubation time in fresh human platelet rich plasma (PRP): 2 hours
6
7
8
9
10
No
. of a
dh
ere
d p
late
lets
/ cm
2 ( x
10
3 )
Si SiDLC(N2 treated)
SiDLC(O2 treated)
SiDLC
(H treated)
SiDLC
(O treated)
These results show that the adherence and activation of platelets is less in case of O and H plasma treated SIDLC films, compared to N plasma treated SiDLC films and Si substrate.
X-Ray Photoelectron Spectroscopy (XPS) studies:
280 282 284 286 288 290
XP
S in
tens
ity
Binding Energy (eV)
C1s C-C
98 99 100 101 102 103 104
XP
S In
tens
ity
Binding Energy (eV)
Si 2pSi-C
680 685 690 695
XP
S in
ten
sity
Binding energy (eV)
F 1sF-F
394 396 398 400 402 404
XP
S in
ten
sity
Binding energy (eV)
N 1sC-N
528 530 532 534 536 538
XP
S In
tens
ity
Binding Energy (eV)
O 1sSi-O Bonding structure of carbon and fluorine show low polarizability, which gives rise to the low surface energy and increased hydrophobicity of carbon tetrafluoride treated SiDLC films.
Si-O and C-N bonds are polar in nature, which results in higher polar component in the surface energy of O and N plasma treated SiDLC films.
C-C bonds are present in all the films.
F-F, C-N and Si-O bonds are seen on the surface of carbon tetraflouride, N and O plasma treated SiDLC films.
4. Conclusions:
The plasma treatment of the Si incorporated DLC films with various gases like N,O,H and carbon tetrafluoride has a considerable effect on its hemocompatibility and surface property.
The O plasma treated SiDLC films have shown higher aPTT, higher albumin adsorption and minimized platelet adhesion and activation. The H plasma treated SiDLC films have reduced fibrinogen adsorption.
The XPS studies revealed the presence of C-N, Si-O and F-F bonds in case of N,O and carbon tetrafluoride treated SiDLC films.
The oxygen and H plasma treated SiDLC films can serve as effective coating on endovascular SMART nitinol stents.