Upload
others
View
2
Download
1
Embed Size (px)
Citation preview
ServiceManual
Yokogawa Electric Corporation
Model 701820DL708E Digital Scope
SM 701820-01E1st Edition
SM 701820-01E
SM 701820-01E 1
IMPORTANT NOTICE TO THE USERThis manual contains information for servicing the YOKOGAWA Digital Scope DL708E.Check the serial number to confirm that this Service Manual covers the right instrument.Make sure to use the wrong manual.
Before any maintenance and servicing, read all safety precautions carefully.
Only properly trained personnel may carry out maintenance and servicing inaccordance with and to the extent permitted by this Service Manual.
Do not disassemble the instrument or its parts, unless otherwise clearly permitted bythis Service Manual.
Do not replace any part or assembly, unless otherwise clearly permitted by thisService Manual.
YOKOGAWA ELECTRIC CORPORATION (YOKOGAWA) does not in principle supplyparts other than those listed in the Customer Maintenance Parts List in this ServiceManual (mainly modules and assemblies). Therefore if an assembly fails, the user
should replace the whole assembly and NOT components within the assembly (seeNOTE). If the user attempts to repair the instrument by replacing individual componentswithin the assembly, YOKOGAWA assumes no responsibility for any consequences,
such as defects in instrument accuracy, functionality, or reliability, or user safety hazards.
YOKOGAWA does not offer more detailed maintenance and service information than thatcontained in this Service Manual.
All reasonable efforts have been made to assure the accuracy of the content of thisService Manual. However, there may still be errors such as clerical errors or omissions.
YOKOGAWA assumes no responsibility of any kind concerning the accuracy or contentsof this Service Manual, nor for the consequences of any errors.
All rights reserved. No part of this Service Manual may be reproduced in any form or byany means without the express written prior permission of YOKOGAWA. The contents ofthis manual are subject to change without notice.
NoteYOKOGAWA instruments have been designed in a way that the replacement of electronic partscan be done on an assembly (module) basis by the user. YOKOGAWA instruments have alsobeen designed in a way that troubleshooting and replacement of any faulty assembly can bedone easily and quickly. Therefore, YOKOGAWA strongly recommends replacing the entireassembly over replacing parts or components within the assembly. The reasons are as follows:• The instruments use high-performance microprocessors, large scale CMOS gate arrays and
surface-mount components to provide state-of-art performance and functions.• Repair of components can only be performed by specially trained and qualified maintenance
personnel with special tools. In addition, repair of components requires various special partsand components, including costly ones. It also requires facilities where highly-accurate andexpensive maintenance equipment and special tools are provided.
• When taking the service life and cost of the instruments into consideration, the replacement ofassemblies offers the user the possibility to use YOKOGAWA instruments more effectivelyand economically with a minimum in downtime.
Disk No. SM11
1st Edition : October 1998 (YK)
All Rights Reserved, Copyright © 1998, Yokogawa Electric Corporation
2 SM 701820-01E
INTRODUCTIONThis manual contains information for servicing the YOKOGAWA Digital Scope DL708E.
NoteThis manual is the first edition, September 1998.
WARNINGThis Service Manual is to be used by properly trained personnel only. To avoidpersonal injury, do not perform any servicing unless you are qualified to do so.
Refer to the Safety Precautions prior to performing any service.Even in case of servicing according to this Service Manual and carried out byqualified personnel, YOKOGAWA assumes no responsibility for any result
occurring from this servicing.
SAFETY PRECAUTIONSThe following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument. Failure to comply with theseprecautions or with specific WARNINGS given elsewhere in this manual violates safetystandards of design, manufacture, and intended use of the instrument.
YOKOGAWA ELECTRIC CORPORATION assumes no liability for the customer’s failureto comply with these requirements.
General definitions of safety symbols used on equipment and in manuals.
Affixed to the instrument. Indicates danger to personnel orinstrument and the operator must refer to the User's Manual. Thesymbol is used in the User's Manual to indicate the reference.
Functional earth terminal (This terminal should not be used as a“Protective earth terminal”.)
WARNING Describes precautions that should be observed to prevent seriousinjury or death to the user.
CAUTION Describes precautions that should be observed to prevent minor or
moderate injury, or damage to the instrument.
SM 701820-01E 3
WARNINGPower SupplyEnsure the source voltage matches the voltage of the power supply beforeturning ON the power.
Power Cord and PlugTo prevent an electric shock or fire, be sure to use the power cord supplied by
YOKOGAWA. The main power plug must be plugged in an outlet with protectiveearth terminal. Do not invalidate protection by using an extension cord withoutprotective grounding.
Protective GroundingThe protective earth terminal must be connected to ground to prevent an electric
shock before turning ON the power.
Necessity of Protective GroundingNever cut off the internal or external protective earth wire or disconnect thewiring of protective earth terminal. Doing so poses a potential shock hazard.
Defect of Protective Earth and FuseDo not operate the instrument when protective earth or fuse might be defective.
Do not Operate in an Explosive AtmosphereDo not operate the instrument in the presence of flammable liquids or vapors.Operation of any electrical instrument in such an environment constitutes a
safety hazard.
Do not Remove any CoversThere are some areas with high voltages. Do not remove any cover if the powersupply is connected. The cover should be removed by qualified personnel only.
External ConnectionTo ground securely, connect the protective grounding before connecting tomeasurement or control unit. Also, when touching the circuit, turn off the power
to the circuit and check that there is no voltage being generated.
4 SM 701820-01E
HOW TO USE THIS MANUALThis manual is meant to be used by qualified personel only. Make sure to have read thesafety precautions at the beginning of this manual and the warnings/cautions containedin the referred chapter prior to carrying out any servicing.
This manual contains the following chapters:
1 GENERAL INFORMATIONDescribes the introduction and safety considerations.
2 PERFORMANCE TESTDescribes the performance tests for checking performance of the instrument.
3 ADJUSTMENTDescribes the adjustments which can be adjusted by users.
4 PRINCIPLE OF OPERATIONDescribes the function block diagrams and principle of operation.
5 TROUBLESHOOTINGDescribes procedures for troubleshooting and how to handle in case parts need tobe replaced.
6 SCHEMATIC DIAGRAMDescribes the system configuration diagram.
7 CUSTOMER MAINTENANCE PARTS LISTContains exploded views and a list of replaceable parts.
Specifications are not included in this manual; for specifications, refer to IM 701820-01E.
SM 701820-01E 5
1
2
3
4
5
6
7
CONTENTSIMPORTANT NOTICE TO THE USER .............................................................................................................. 1INTRODUCTION ................................................................................................................................................... 2SAFETY PRECAUTIONS .................................................................................................................................... 2
HOW TO USE THIS MANUAL ........................................................................................................................... 4
Chapter 1 GENERAL INFORMATION ....................................................................................................... 1-1
1.1 Introduction ............................................................................................................................ 1-11.2 Safety Consideration ............................................................................................................. 1-1
Chapter 2 PERFORMANCE TEST ............................................................................................................. 2-12.1 Introduction ............................................................................................................................ 2-12.2 Equipment Required .............................................................................................................. 2-1
2.3 Checking of Packaged Contents ........................................................................................... 2-32.4 Initialization ............................................................................................................................ 2-32.5 Checking of Overview ............................................................................................................ 2-3
2.6 Waveform Display Test (Operation Check by Using AUTO SETUP) ..................................... 2-42.7 Built-in Printer Self-diagnosis ................................................................................................ 2-52.8 Internal Memory Self-diagnosis ............................................................................................. 2-5
2.9 Keyboard, Jog Shuttle, and Rotary Knob Self-diagnosis ....................................................... 2-62.10 Built-in FDD Test .................................................................................................................... 2-72.11 Built-in HDD Test (for option /C8) .......................................................................................... 2-7
2.12 Time Base Accuracy Test ...................................................................................................... 2-82.13 High-Speed Isolation Module (701850) Test .......................................................................... 2-92.14 High-Speed Module (701851) Test ...................................................................................... 2-142.15 High-Resolution, High-Voltage, Isolation Module (701852) Test ......................................... 2-18
2.16 High-resolution Isolation Module (701853) Test .................................................................. 2-232.17 Temperature Module (701860) Test ..................................................................................... 2-282.18 Logic Input Module (701870) Test ....................................................................................... 2-31
2.18.1 Using High-Speed Logic Probe (700986) ................................................................. 2-312.18.2 Using Isolation Logic Probe (700987) ...................................................................... 2-33
2.19 Strain Module (701880) Test ............................................................................................... 2-37
Chapter 3 ADJUSTMENT ............................................................................................................................. 3-13.1 Introduction ............................................................................................................................ 3-1
3.2 Test Environment ................................................................................................................... 3-13.3 Equipment Required .............................................................................................................. 3-13.4 High-Speed Isolation Module (701850) Adjustment .............................................................. 3-2
3.5 High-speed Module (701851) Adjustment ............................................................................. 3-93.6 Temperature Module (701860) Adjustment ......................................................................... 3-143.7 Logic Input Module (701870) Adjustment ............................................................................ 3-15
Chapter 4 PRINCIPLE OF OPERATION ................................................................................................... 4-14.1 Introduction ............................................................................................................................ 4-1
4.2 Main Unit (701820) ................................................................................................................ 4-14.3 Modules ................................................................................................................................. 4-34.4 Functions of Each Gate Array ................................................................................................ 4-4
6 SM 701820-01E
Chapter 5 TROUBLESHOOTING ............................................................................................................... 5-15.1 Introduction ............................................................................................................................ 5-15.2 Flow Chart ............................................................................................................................. 5-2
5.3 Self Test ................................................................................................................................. 5-5
Chapter 6 SCHEMATIC DIAGRAM ............................................................................................................ 6-1
Chapter 7 CUSTOMER MAINTENANCE PARTS LIST .......................................................................... 7-17.1 Customer Maintenance Parts List ......................................................................................... 7-1
7.2 Standard Accessories ............................................................................................................ 7-5
CONTENTS
SM 701820-01E 1-1
1
GE
NE
RA
L IN
FO
RM
AT
ION
Chapter 1 GENERAL INFORMATION1.1 Introduction
This manual describes servicing information on YOKOGAWA Digital Scope DL708E.This chapter contains information required for using this manual and information that
must be known before starting servicing of the DL708E.
1.2 Safety ConsiderationThoroughly read the safety precautions at the beginning of this manual without fail. Alsofully read warnings and cautions contained in each chapter.
1.1 Introduction / 1.2 Safety Consideration
SM 701820-01E 2-1
2
PE
RF
OR
MA
NC
E T
ES
T
2.1 Introduction / 2.2 Equipment Required
Chapter 2 PERFORMANCE TEST2.1 Introduction2.2 Equipment Required2.3 Checking of Packaged Contents
2.4 Initialization2.5 Checking of Overview2.6 Waveform Display Test (Operation Check by Using AUTO SETUP)
2.7 Built-in Printer Self-diagnosis2.8 Internal Memory Self-diagnosis2.9 Keyboard, Jog Shuttle, and Rotary Knob Self-diagnosis
2.10 Built-in FDD Test2.11 Built-in HDD Test (for option /C8)2.12 Time Base Accuracy Test
2.13 High-Speed Isolation Module (701850) Test2.14 High-Speed Module (701851) Test2.15 High-Resolution, High-Voltage, Isolation Module (701852) Test
2.16 High-Resolution, Isolation Module (701853) Test2.17 Temperature Module (701860) Test2.18 Logic Input Module (701870) Test
2.18.1 Using High-Speed Logic Probe (700986)2.18.2 Using Isolation Logic Probe (700987)
2.19 Strain Module (701880) Test
2.1 IntroductionThis chapter describes the acceptance inspection of DL708E.
2.2 Equipment RequiredReference operating conditions (Test environment: performance test items)Warm up time More than 30 minutes
Ambient temperature 23 ± 5°CHumidity 55 ± 10%RHVoltage of power supply Specified voltage ± 1% or less
General purpose equipment (Cables)Equipment Qty Mandatory specifications Recommended Remarks
10:1 probe 1 10 MHz or more Yokogawa 700929 —Yokogawa 700998 —
Synthesized signal 1 Output frequency range HP8657A —generator 100 kHz to 150 MHz
Accuracy : ±5ppm
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
Measurement lead 1 For 42 V or more Yokogawa 758917 —
Measurement lead 1 For 42 V or less Yokogawa 366961 Banana to alligator clip
Conversion adapter 1 — HP1250-0780 N plug to BNC jack
Conversion adapter 1 — Yokogawa 366921 BNC to banana(female)
Conversion adapter 2 — Yokogawa 366922 Banana(male) to BNC
50 Ω terminator 1 — HP1251-2277 —
T-shaped adapter 1 — Yokogawa 366923 BNC parallel connection
Alligator clip adapter 1 — Yokogawa 758922 Banana to alligator clip
Formatted floppy disk 1 1.44 / 1.2 MB — —MS-DOS formatted
SM 701820-01E2-2
Test equipment for 701850 / 701851 (High-Speed Isolation module / High-Speedmodule)Equipment Qty Mandatory specifications Recommended Remarks
DC voltage generator 1 DC accuracy : ±0.03% Yokogawa wavetek 9100 Yokogawa wavetek 9100Output : ±32 V or Yokogawa 7651 offers complete compatibility.
Standard signal 1 ±0.5 dB : 10 MHz or less Yokogawa FG200 —generator ±1 dB : more than 10 MHz
100 kHz to 10 MHzSine wave
Power meter 1 ±0.05 dB : 10 MHz or less HP437B —
Power senser 1 100 kHz to 10 MHz HP8482A —
Withstand voltage 1 3700 VAC 10 mA limit Kikusui TOS-8750 —tester
Test equipment for 701852 / 701853 (High-Resolution, High-Voltage, Isolationmodule / High-Resolution, Isolation module)Equipment Qty Mandatory specifications Recommended Remarks
DC voltage generator 1 DC accuracy : ±0.03% Yokogawa wavetek 9100 Yokogawa wavetek 9100Output : ±32 V or Yokogawa 7651 offers complete compatibility.
Standard signal 1 ±0.1 dB : 100 kHz or less Yokogawa FG200 —generator
Withstand voltage 1 3700 VAC 10 mA limit Kikusui TOS-8750 —tester
Test equipment for 701860 (Temperature module)Equipment Qty Mandatory specifications Recommended Remarks
DC voltage generator 1 DC accuracy : ±0.03% Yokogawa wavetek 9100 Yokogawa wavetek 9100Output : ±10 V or Yokogawa 7651 offers complete compatibility.
Standard signal 1 ±0.2 dB : 100 kHz or less Yokogawa FG200 —generator
Thermocouple T 1 Thermocouple T — —
0 °C reference 1 Accuracy ±0.1 °C ZERO-CON ZC-114 —temperature unit Range 0 °C
Test equipment for 701870 (Logic Input module)Equipment Qty Mandatory specifications Recommended Remarks
Standard signal 1 ±0.2 dB : 100 kHz or less Yokogawa FG200 —generator Available of 0 to 5 V,(using 700986) square wave
AC voltage generator 1 30 VAC ± 0.03%, Yokogawa wavetek 9100 —(using 700986) square wave
AC/DC voltage 1 250 VAC ± 0.03%, Yokogawa wavetek 9100 —generator 10 VDC ± 0.03%(using 700987)
Special purpose 1 High-speed logic probe Yokogawa 700986 Test includes logic probe.logic probe
Special purpose 1 Isolation logic probe Yokogawa 700987 Test includes logic probe.logic probe
Connecting lead 1 — Yokogawa B9879PX Converting to alligator clip
Test equipment for 701880 (Strain module)Equipment Qty Mandatory specifications Recommended Remarks
DC voltage/current 1 DC accuracy : ±0.03% with Yokogawa 7651 —generator limiter for voltage and current
Standard signal 1 ±0.2 dB : 100 kHz or less Yokogawa FG200 —generator
Digital multimeter 1 5 1/2-digits equivalent Yokogawa 7552 —
Shielded cable 1 Two cores shielded cable — For maximum input voltageand NDIS connector test
Shielded cable 1 Two cores shielded cable — For auto balance & DCand NDIS connector accuracy test
Measurement lead 1 Two wire lead — For bridge voltage testand NDIS connector
Shielded cable 1 Single core shielded cable — For frequency characteristicswith BNC and NDIS connector test
2.2 Equipment Required
SM 701820-01E 2-3
2
PE
RF
OR
MA
NC
E T
ES
T
2.3 Checking of Packaged ContentsCheck on the following points if the packaged items are complete and do not contradictto the description in the following pages.
(1) Model and suffix code(2) Standard accessories(3) Input module type
(4) Optional accessories
2.4 InitializationReset the settings to the shipped conditions.
[Procedure]Turn on the power source while pressing the RESET key.
2.5 Checking of OverviewTurn on the power source of the DL708E and confirm that the main unit recognizes themodules and optional accessories correctly.
[Procedure](1) Turn on the power source.
(2) Press the MISC key and press the “OVERVIEW” soft key.(3) Confirm that the modules and options are recognized correctly.* The model of the main unit can be confirmed on the nameplate. (cf. 701820-1/M/M1/
C8)
[Display]Model : Model name of the main unit (“DL708E” is always displayed)Version : Version No. of the firmware of the mail unitAcquisition MEM : Installed memory length and memory option
100 kW/CH MAX. 400 kW : Standard memory1 MW/CH MAX. 4 MW /M1 : at option /M12 MW/CH MAX. 8 MW /M2 : at option /M2
4 MW/CH MAX. 16 MW /M3 : at option /M3Math length : Maximum length of computation waveformMath User Def. : Existence of user define math function (“Yes” for option /G2)
Printer : Existence of built-in printer (Always “Yes” for DL708E)External Clock : Existence of External Clock input function (“Yes” for option /F2)SCSI : Existence of SCSI (“Yes” for option /C7 or /C8)
HD : Existence of hard disk (“Yes” for option /C8)Link Date : Date when the firmware is createdModule Configuration : Existence of installed modules
HS_IOSL 701850HS 701851HR_HV 701852
HR 701853TEMP 701860LOGIC 701870
STRAIN 701880
2.3 Checking of Packaged Contents / 2.4 Initialization / 2.5 Checking of Overview
SM 701820-01E2-4
2.6 Waveform Display Test (Operation Check by Using AUTOSETUP)
Check waveform display for each channel.
Run the test on the voltage modules (701850 / 51 / 52 / 53).Use COMP signal as a signal source.
[Equipment required]Test for 701850 / 51Equipment Qty Mandatory specifications Recommended Remarks
10:1 probe 1 10 MHz or more Yokogawa 700929 Isolated probe for 701850
10:1 probe 1 10 MHz or more Yokogawa 700998 Non-isolated probe for 701851
Test for 701852 / 53Equipment Qty Mandatory specifications Recommended Remarks
Measurement lead 1 For 42 V or less Yokogawa 366961 Banana to alligator clip
[Connection]
Probe or measurement lead
[Procedure]Test each channel in the same manner.(1) Send the COMP signal to the channel to be tested.(2) Select AUTO SETUP Channel. Press the SETUP key, then press the “Auto Setup
CH” soft key to select a channel to be tested.(3) Execute AUTO SETUP. Press the SETUP key, then press the “Auto Setup” soft
key.
[Confirmation]Confirm that the COMP signal makes following waveform.
Frequency : 1 kHz ±10%Amplitude : 1 V ±10%* Perform the phase compensation when 10:1 probe is connected.
2.6 Waveform Display Test (Operation Check by Using AUTO SETUP)
SM 701820-01E 2-5
2
PE
RF
OR
MA
NC
E T
ES
T
2.7 Built-in Printer Self-diagnosis[Equipment required]Printer roll paper (B9850NX)
[Procedure](1) Press the MISC key.
(2) Press the “Next 1/2” soft key to display the self test soft key menu.(3) Press the “Self test” soft key to display the self test menu.(4) Press the “Test Item” soft key to select “Printer.”(5) Press the “Exec” soft key to start the printer test.
[Confirmation]Confirm that the following is printed out.
2.8 Internal Memory Self-diagnosis(1) System memory test (Main memory, back-up memory, and ROM sum check)(2) Acquisition memory test (Standard and extended memory (/M1, /M2, or /M3))
[Equipment required]None
2.7 Built-in Printer Self-diagnosis / 2.8 Internal Memory Self-diagnosis
SM 701820-01E2-6
[Procedure]TEST1: System memory test(1) Press the MISC key.
(2) Press the “Next 1/2” soft key to display the self test soft key menu.(3) Press the “Self test” soft key to display the self test menu.(4) Press the “Test Item” soft key to display the test item menu.
(5) Keep pressing the “Test Item” soft key until “Memory” is selected.(6) Press the “Memory Test” soft key to display the memory item menu.(7) Keep pressing the “Memory Test” soft key to select “System.”
(8) Press the “Exec” soft key to start the memory test.
TEST2: Acquisition memory test
(1) Press the MISC key.(2) Press the “Next 1/2” soft key to display the self test soft key menu.(3) Press the “Self test” soft key to display the self test menu.
(4) Press the “Test Item” soft key to display the test item menu.(5) Keep pressing the “Test Item” soft key until “Memory” is selected.(6) Press the “Memory Test” soft key to display the memory item menu.
(7) Keep pressing the “Memory Test” soft key to select “ACQ RAM.”(8) Press the “Exec” soft key to start the memory test.
[Confirmation]Confirm that all test items display “PASS.”* Test2 takes several minutes.
2.9 Keyboard, Jog Shuttle, and Rotary Knob Self-diagnosis[Equipment required]None
[Procedure](1) Press the MISC key.(2) Press the “Next 1/2” soft key to display the self test soft key menu.(3) Press the “Self test” soft key to display the self test menu.
(4) Press the “Test Item” soft key to select “Key Board.”(5) Press the “Exec” soft key to start the key test.(6) To stop the key test, press the SHIFT+ESC key.
[Confirmation]Perform the key operations prompted by a blinking menu item and confirm that the
blinking item shifts one by one.* Operate the shuttle ring very slowly when confirming the Shuttle Left / Right function.
2.8 Internal Memory Self-diagnosis / 2.9 Keyboard, Jog Shuttle, and Rotary Knob Self-diagnosis
SM 701820-01E 2-7
2
PE
RF
OR
MA
NC
E T
ES
T
2.10 Built-in FDD TestExecute save / load of the setup information and confirm that there is no error.
[Equipment required]A formatted floppy disk.
[Procedure](1) Press the MISC key.(2) Press the “Next 1/2” soft key to display the self test soft key menu.
(3) Press the “Self test” soft key to display the self test menu.(4) Press the “Test Item” soft key to select “FDD.”(5) Pressing the “Exec” soft key will start the test.
[Confirmation]Confirm that the read/write test are passed and the floppy disk type is identified correctly.
* The test fails if the floppy disk is not inserted.
2.11 Built-in HDD Test (for option /C8)[Equipment required]None
[Procedure](1) Press the MISC key.
(2) Press the “Next 1/2” soft key to display the self test soft key menu.(3) Press the “Self test” soft key to display the self test menu.(4) Press the “Test Item” soft key and select “HDD.”(5) Pressing the “Exec” soft key will start the test.
[Confirmation]Confirm that the test is passed. (“Write pass” is displayed.)
2.10 Built-in FDD Test / 2.11 Built-in HDD Test (for option /C8)
SM 701820-01E2-8
2.12 Time Base Accuracy Test[Specifications]±0.005%
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
Synthesized signal 1 Output frequency range HP8657A —generator 100 kHz to 150 MHz
Accuracy : ±5ppm
High-speed module 1 — Yokogawa 701850 —or 701851
50 Ω terminator 1 — HP1251-2277 —
Conversion adapter 1 — HP1250-0780 N plug to BNC jack
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
[Connection]
BNC-BNC cable 366924
Synthesizedsignal generator
HP8657A
Conversion adapter HP1250-0780
50 Ω terminator HP1251-2277
To Hogh-Speed module
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each channel in the same manner.
(1) Connect the output lead of the synthesized signal generator to the channel to betested as shown above.
(2) Set the DL708E as shown below.
Record Length 10 kWT/DIV 1 ms/divV/DIV 100 mV Channel to be tested
Position 0 V ↓Probe 1 : 1 ↓MEASURE Freq ↓
(3) Set the synthesized signal generator as shown below.Sine waveAmplitude 400 mVpp
Frequency 1.0005 MHz
[Confirmation]The measured frequency of the beat waveform on screen by using MEASURE must be500 Hz ± 45 Hz.
2.12 Time Base Accuracy Test
SM 701820-01E 2-9
2
PE
RF
OR
MA
NC
E T
ES
T
2.13 High-Speed Isolation Module (701850) TestOffset Voltage Accuracy Test (701850)
[Specifications] ±0.04% of offset voltage range + 1% of set value
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
DC voltage generator 1 DC accuracy ±0.03% Yokogawa wavetek 9100 —
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
Conversion adapter 1 — Yokogawa 366922 Banana(male) to BNC
[Connection]
BNC-BNC cable 366924
DC voltage generator9100
Conversion adapter 366922
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the
“Initialize” soft key.)
[Procedure]Test each channel in the same manner.(1) Connect the output cable of the DC voltage generator to the channel to be tested as
shown above.
(2) Set the DL708E as shown below.T/DIV 1 ms/divACQ Average
Position 0 div Channel to be testedCoupling DC ↓Bandwidth 500 kHz ↓MEASURE Avg(Average) ↓
(3) Set the DC voltage generator as follows.
[Confirmation]Check that the measured value is within the range as specified in the table below.DL708E settings DC voltage generator Permissible rangeV/DIV Offset voltage Output voltage (Test spec.)
100 mV/div 5 V 5 V 5 V ± 100 mV↓ –5 V –5 V –5 V ± 100 mV
2.13 High-Speed Isolation Module (701850) Test
SM 701820-01E2-10
Residual Noise Test (701850)
[Specifications]Larger of ±400 µV or ±0.06 div, typical value
[Connection]
Not connected.
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the
“Initialize” soft key.)
[Procedure]Test each channel in the same manner.Set the DL708E as shown below.
T/DIV 10 µs/div
ACQ NormalV/DIV 5 mV/div Channel to be testedPositon 0 div ↓Coupling GND ↓Bandwidth Full ↓Offset 0 mV ↓MEASURE PP(Peak to Peak) ↓
[Confirmation]Check that the measured value is within 800 µVp-p.
2.13 High-Speed Isolation Module (701850) Test
SM 701820-01E 2-11
2
PE
RF
OR
MA
NC
E T
ES
T
DC Accuracy Test (701850)
[Specifications]±(1.5% of 8 div + offset voltage accuracy)
(offset voltage accuracy = ±0.04% offset voltage range + 1% of set value)
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
DC voltage generator 1 DC accuracy ±0.03% Yokogawa wavetek 9100 —
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
Conversion adapter 1 — Yokogawa 366922 Banana(male) to BNC
[Connection]
BNC-BNC cable 366924
DC voltage generator9100
Conversion adapter 366922
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the
“Initialize” soft key.)
[Procedure]Test each channel in the same manner.(1) Connect the output cable of the DC voltage generator to the channel to be tested as
shown above.(2) Set the DL708E as shown below.
T/DIV 1 ms/divACQ AverageProbe 1 : 1 Channel to be tested
Bandwidth 500 kHz ↓MEASURE Avg(Average) ↓
(3) Set the DC voltage generator as follows.
[Confirmation]Check that the measured value is within the range as specified in the table below.DL708E settings DC voltage genetaror Permissible rangeV/DIV Output voltage (Test spec.)
10 mV/div 38 mV 36.4 to 39.6 mV↓ 0 mV –1.6 to 1.6 mV↓ –38 mV –39.6 to –36.4 mV
200 mV/div 760 mV 732 to 788 mV↓ 0 mV –28 to 28 mV↓ –760 mV –788 to –732 mV
5 V/div 19 V 18.8 to 19.2 V↓ 0 V –0.2 to 0.2 V↓ –19 V –19.2 to 18.8 V
10 V/div 32 V 30.6 to 33.4 V↓ 0 V –1.4 to 1.4 V↓ –32 V –33.4 to –30.6 V
2.13 High-Speed Isolation Module (701850) Test
SM 701820-01E2-12
Frequency Characteristics Test (701850)
[Specifications]DC to 2 MHz (–3 dB)
[Equipment required]Equipment Qty Mandatory specifications Recommended
Standard signal 1 100 kHz to 10 MHz Yokogawa FG200generator Sine wave
Power meter 1 ±0.05 dB : 10 MHz or less HP437B
Power senser 1 100 kHz to 10 MHz HP8482A
T-shaped adapter 1 BNC parallel connection Yokogawa 366923
BNC-BNC cable 1 1 m or less Yokogawa 366924
[Connection]
BNC-BNC cable 366924
Standard signal generatorFG200
T-shaped adapter 366923
Power senserHP8482A
Power meterHP437B
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each channel in the same manner.(1) Connect the output cable of the standard signal generator to the channel to be
tested as shown above.(2) Set the DL708E as shown below.
Record length 10 kW
V/DIV 100 mV/div Channel to be testedProbe 1 : 1 ↓Bandwidth Full ↓MEASURE PP(Peak to Peak) ↓
(3) Set the standard signal generator as shown below.Sine wave
Amplitude 1.2 Vp-p
[Confirmation]Check that the measured value is within the range as specified in the table below.Calculate the difference between the each value at 100 kHz (reference value) and theone at 2 MHz.DL708E settings Standard signal generator DL708E Power meterT/DIV Frequency Measured value PP Measured value (W)
10 µs/div(10 MS/s) 100 kHz <1> <3>↓ 2.002 MHz <2> <4>
Calculation Judgment
20log(<2>/<1>) – 10log(<4>/<3>) –3.0 dB or lower
2.13 High-Speed Isolation Module (701850) Test
SM 701820-01E 2-13
2
PE
RF
OR
MA
NC
E T
ES
T
Withstand Voltage Test (701850)
[Specifications]3700 VAC for 1 minute (between input and chassis GND)
[Equipment required]Equipment Qty Mandatory specifications Recommended
Withstand voltage 1 3700 VAC 10 mA limit Kikusui TOS-8750tester
Banana (male) 1 Short-circuit H and L —-BNC conversion of BNC cable
[Connection]Jig with short-circuited H and L of BNC cable.
Not connected to L/N
Withstandvoltage tester
H
H
L
L
Attached AC cable
(connected tomain unit GND)
[Connection method](1) Prepare the jig with short-circuited H and L of BNC cable.
(2) Connect H of the withstand voltage tester to H/L as shown above.(3) Connect G (GND) of the attached AC cable to L of the withstand voltage tester.(4) The power SW of DL708E should be turned off.
[Withstand voltage tester setting]3700 VAC (60 Hz) 10 mA limit
[Confirmation]Check that the limiter does not operate when voltage is applied for one minute.
WARNINGPerforming this test is very dangerous due to handling high voltage.The withstand voltage test is to be performed only by the experienced personnel
who are thoroughly understand the test method.Be sure not to touch the live-wire.The above connection is provided for explanatory purpose.
To avoid a risk of high voltage, please prepare a jig which has no exposed live-wire.
* The high-speed module (701851) cannot be used for the withstand voltagetest.
2.13 High-Speed Isolation Module (701850) Test
SM 701820-01E2-14
2.14 High-Speed Module (701851) TestOffset Voltage Accuracy Test (701851)
[Specifications] ±0.04% of offset voltage range + 1% of set value
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
DC voltage generator 1 DC accuracy ±0.03% Yokogawa wavetek 9100 —
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
Conversion adapter 1 — Yokogawa 366922 Banana(male) to BNC
[Connection]
BNC-BNC cable 366924
DC voltage generator9100
Conversion adapter 366922
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the
“Initialize” soft key.)
[Procedure]Test each channel in the same manner.(1) Connect the output cable of the DC voltage generator to the channel to be tested as
shown above.
(2) Set the DL708E as shown below.T/DIV 1 ms/divACQ Average
Position 0 div Channel to be testedCoupling DC ↓Bandwidth 500 kHz ↓MEASURE Avg(Average) ↓
(3) Set the DC voltage generator as follows.
[Confirmation]Check that the measured value is within the range as specified in the table below.DL708E settings DC voltage generator Permissible rangeV/DIV Offset voltage Output voltage (Test spec.)
100 mV/div 5 V 5 V 5 V ± 100 mV↓ –5 V –5 V –5 V ± 100 mV
2.14 High-Speed Module (701851) Test
SM 701820-01E 2-15
2
PE
RF
OR
MA
NC
E T
ES
T
Residual Noise Test (701851)
[Specifications]Larger of ±300 µV or ±0.04 div, typical value
[Connection]
Not connected.
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the
“Initialize” soft key.)
[Procedure]Test each channel in the same manner.Set the DL708E as shown below.
T/DIV 10 µs/div
ACQ NormalV/DIV 5 mV/div Channel to be testedPositon 0 div ↓Coupling GND ↓Bandwidth Full ↓Offset 0 mV ↓MEASURE PP(Peak to Peak) ↓
[Confirmation]Check that the measured value is within 600 µVp-p.
2.14 High-Speed Module (701851) Test
SM 701820-01E2-16
DC Accuracy Test (701851)
[Specifications]±(1.5% of 8 div + offset voltage accuracy)
(offset voltage accuracy = ±0.04% of offset voltage range + 1% of set value)
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
DC voltage generator 1 DC accuracy ±0.03% Yokogawa wavetek 9100 —
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
Conversion adapter 1 — Yokogawa 366922 Banana(male) to BNC
[Connection]
BNC-BNC cable 366924
DC voltage generator9100
Conversion adapter 366922
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the
“Initialize” option.)
[Procedure]Test each channel in the same manner.(1) Connect the output cable of the DC voltage generator to the channel to be tested as
shown above.(2) Set the DL708E as shown below.
T/DIV 1 ms/divACQ AverageProbe 1 : 1 Channel to be tested
Bandwidth 500 kHz ↓MEASURE Avg(Average) ↓
(3) Set the DC voltage generator as follows.
[Confirmation]Check that the measured value is within the range as specified in the table below.DL708E settings DC voltage generator Permissible rangeV/DIV Output voltage (Test spec.)
10 mV/div 38 mV 36.4 to 39.6 mV↓ 0 mV –1.6 to 1.6 mV↓ –38 mV –39.6 to –36.4 mV
200 mV/div 760 mV 732 to 788 mV↓ 0 mV –28 to 28 mV↓ –760 mV –788 to –732 mV
5 V/div 19 V 18.8 to 18.2 V↓ 0 V –0.2 to 0.2 V↓ –19 V –19.2 to 18.8 V
10 V/div 32 V 30.76 to 33.24 V↓ 0 V –1.24 to 1.24 V↓ –32 V –33.24 to –30.76 V
2.14 High-Speed Module (701851) Test
SM 701820-01E 2-17
2
PE
RF
OR
MA
NC
E T
ES
T
Frequency Characteristics Test (701851)
[Specifications]DC to 4 MHz (–1 dB)
[Equipment required]Equipment Qty Mandatory specifications Recommended
Standard signal 1 100 kHz to 10 MHz Yokogawa FG200generator Sine wave
Power meter 1 ±0.05 dB : 10 MHz or less HP437B
Power senser 1 100 kHz to 10 MHz HP8482A
T-shaped adapter 1 BNC parallel connection Yokogawa 366923
BNC - BNC cable 1 1 m or less Yokogawa 366924
[Connection]
BNC-BNC cable 366924
Standard signal generatorFG200
T-shaped adapter 366923
Power senserHP8482A
Power meterHP437B
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each channel in the same manner.(1) Connect the output cable of the standard signal generator to the channel to be
tested as shown above.(2) Set the DL708E as shown below.
Record length 10 kW
V/DIV 100 mV/div Channel to be testedProbe 1 : 1 ↓Bandwidth Full ↓MEASURE PP(Peak to Peak) ↓
(3) Set the standard signal generator as shown below.Sine wave
Amplitude 1.2 Vp-p
[Confirmation]Check that the measured value is within the range as specified in the table below.Calculate the difference between the eath value at 100 kHz (reference value) and theone at 4 MHz.DL708E settings Standard signal generator DL708E Power meterT/DIV Frequency Measured value PP Measured value (W)
10 µs/div(10 MS/s) 100 kHz <1> <3>↓ 4.004 MHz <2> <4>
Calculation Judgment
20log(<2>/<1>) – 10log(<4>/<3>) –1.0 dB or lower
2.14 High-Speed Module (701851) Test
SM 701820-01E2-18
2.15 High-Resolution, High-Voltage, Isolation Module (701852) TestOffset Voltage Accuracy Test (701852)
[Specifications]±(0.5% of 8 div) at 200 V/div to 100 mV/div±(1% of 8 div) at 50 mV/div
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
DC voltage generator 1 DC accuracy ±0.03% Yokogawa wavetek 9100 —
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
Conversion adapter 2 — Yokogawa 366922 Banana(male) to BNC
[Connection]
BNC-BNC cable 366924
DC voltage generator9100
Conversion adapter 366922
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the“Initialize” soft key.)
[Procedure]Test each channel in the same manner.
(1) Connect the output cable of the DC voltage generator to the channel to be tested asshown above.
(2) Set the DL708E as shown below.
T/DIV 1 ms/divACQ AveragePosition 0 div Channel to be tested
Coupling DC ↓Bandwidth 400 Hz ↓MEASURE Avg(Average) ↓
(3) Set the DC voltage generator as follows.
[Confirmation]Check that the measured value is within the range as specified in the table below.DL708E settings DC voltage generator Permissible rangeV/DIV Offset voltage Output voltage (Test spec.)
50 mV/div 300 mV 300 mV 300 mV ± 4 mV10 V/div 30 V 30 V 30 V ± 400 mV
2.15 High-Resolution, High-Voltage, Isolation Module (701852) Test
SM 701820-01E 2-19
2
PE
RF
OR
MA
NC
E T
ES
T
Residual Noise Test (701852)
[Specifications]Larger of ±2.5 mV or ±0.01 div, typical value
[Connection]
Not connected.
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the
“Initialize” soft key.)
[Procedure]Test each channel in the same manner.Set the DL708E as shown below.
T/DIV 1 ms/div
ACQ NormalV/DIV 200 mV/div Channel to be testedPositon 0 div ↓Coupling GND ↓Bandwidth Full ↓Offset 0 mV ↓MEASURE PP(Peak to Peak) ↓
[Confirmation]Check that the measured value is within 5 mVp-p.
2.15 High-Resolution, High-Voltage, Isolation Module (701852) Test
SM 701820-01E2-20
DC Accuracy Test (701852)[Specifications]±(0.5% of 8 div) at 200 V/div to 100 mV/div±(1% of 8 div) at 50 mV/div
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
DC voltage generator 1 DC accuracy ±0.03% Yokogawa wavetek 9100 —
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
Conversion adapter 2 — Yokogawa 366922 Banana(male) to BNC
[Connection]
BNC-BNC cable 366924
DC voltage generator9100
Conversion adapter 366922
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each channel in the same manner.(1) Connect the output cable of the DC voltage generator to the channel to be tested as
shown above.(2) Set the DL708E as shown below.
T/DIV 1 ms/divACQ AverageProbe 1 : 1 Channel to be tested
Bandwidth 400 Hz ↓MEASURE Avg(Average) ↓
(3) Set the DC voltage generator as follows.
[Confirmation]Check that the measured value is within the range as specified in the table below.DL708E settings DC voltage generator Permissible rangeV/DIV Output voltage (Test spec.)
50 mV/div 190 mV 186 to 194 mV↓ 0 mV –4 to 4 mV↓ –190 mV –194 to –186 mV
100 mV/div 380 mV 376 to 384 mV↓ 0 mV –4 to 4 mV↓ –380mV –384 to –376 mV
200 mV/div 760mV 752 to 768 mV↓ 0 mV –80 to 80 mV↓ –760 mV –768 to –752 mV
500 mV/div 1.9 V 1.88 to 1.92 V↓ 0 V –0.02 to 0.02 V↓ –1.9 V –1.92 to –1.88 V
1 V/div 3.8 V 3.76 to 3.84 V↓ 0 V –0.04 to 0.04 V↓ –3.8 V –3.84 to –3.76 V
2 V/div 7.6 V 7.52 to 7.68 V↓ 0 V –0.08 to 0.08 V↓ –7.6 V –7.68 to –7.52 V
5 V/div 19 V 18.8 to 19.2 V↓ 0 V –0.2 to 0.2 V↓ –19 V –19.2 to –18.8 V
10 V/div 32 V 31.6 to 32.4 V↓ 0 V –0.4 to 0.4 V↓ –32 V –32.4 to 31.6 V
2.15 High-Resolution, High-Voltage, Isolation Module (701852) Test
SM 701820-01E 2-21
2
PE
RF
OR
MA
NC
E T
ES
T
Frequency Characteristics Test (701852)
[Specifications]DC to 40 kHz (–3 dB) at 200 V/div to 100 mV/div
DC to 30 kHz (–3 dB) at 50 mV/div
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
Standard signal 1 ±0.1 dB : 100 kHz or less Yokogawa FG200 —generator
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
Conversion adapter 1 — Yokogawa 366922 Banana(male) to BNC
[Connection]
BNC-BNC cable 366924
Standard signal generatorFG200
Conversion adapter 366922
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the
“Initialize” option.)
[Procedure]Test each channel in the same manner.
(1) Connect the output cable of the standard signal generator to the channel to betested as shown above.
(2) Set the DL708E as shown below.
Probe 1 : 1 Channel to be testedBandwidth Full ↓MEASURE PP(Peak to Peak) ↓
(3) Set the standard signal generator as shown below.Sine waveAmplitude 300 mVp-p (when voltage-axis sensitivity is 50 mV/div.),
20 Vp-p (when voltage-axis sensitivity is 10 V/div.)
[Confirmation]Check that the measured value is within the range as specified in the table below.Calculate the difference between the PP value at 100 Hz (reference value) and the oneat each frequency [<2>, <4>].DL708E settings Standard signal generator DL708ET/DIV V/DIV Frequency Measured value PP
1 ms/div(100 kS/s) 50 mV/div 100 Hz <1>↓ ↓ 30.1 kHz <2>↓ 10 V/div 100 Hz <3>↓ ↓ 40.1 kHz <4>
Calculation Judgment
20 log(<2>/<1>) –3.0 dB or lower
20 log(<4>/<3>) –3.0 dB or lower
2.15 High-Resolution, High-Voltage, Isolation Module (701852) Test
SM 701820-01E2-22
Withstand Voltage Test (701852)
[Specifications]3700 VAC for 1 minute (between input and chassis GND)
[Equipment required]Equipment Qty Mandatory specifications Recommended
Withstand voltage 1 3700 VAC 10 mA limit Kikusui TOS-8750tester
Banana(male) 1 Short-circuit H and L —-BNC conversion of BNC cable
[Connection]Jig with short-circuited H and L of BNC cable.
Not connected to L/N
Withstandvoltage tester
H
H
L
L
Attached AC cable
(connected tomain unit GND)
[Connection method](1) Prepare the jig with short-circuited H and L of BNC cable.
(2) Connect H of the withstand voltage tester to H/L as shown above.(3) Connect G (GND) of the attached AC cable to L of the withstand voltage tester.(4) The power SW of DL708E should be turned off.
[Withstand voltage tester setting]3700 VAC (60 Hz) 10 mA limit
[Confirmation]Check that the limiter does not operate when voltage is applied for one minute.
WARNINGPerforming this test is very dangerous due to handling high voltage.The withstand voltage test is to be performed only by the experienced personnel
who are thoroughly understand the test method.Be sure not to touch the live-wire.The above connection is provided for explanatory purpose.
To avoid a risk of high voltage, please prepare a jig which has no exposed live-wire.
2.15 High-Resolution, High-Voltage, Isolation Module (701852) Test
SM 701820-01E 2-23
2
PE
RF
OR
MA
NC
E T
ES
T
2.16 High-resolution Isolation Module (701853) TestOffset Voltage Accuracy Test (701853)
[Specifications]±(0.3% of 8 div) at 20 V/div to 20 mV/div±(0.5% of 8 div) at 10 mV/div±(1% of 8 div) at 5 mV/div
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
DC voltage generator 1 DC accuracy ±0.03% Yokogawa wavetek 9100 —
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
Conversion adapter 2 — Yokogawa 366922 Banana(male) to BNC
[Connection]
BNC-BNC cable 366924
DC voltage generator9100
Conversion adapter 366922
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the
“Initialize” soft key.)
[Procedure]Test each channel in the same manner.(1) Connect the output cable of the DC voltage generator to the channel to be tested as
shown above.
(2) Set the DL708E as shown below.T/DIV 1 ms/divACQ Average
Position 0 div Channel to be testedCoupling DC ↓Bandwidth 400 Hz ↓MEASURE Avg(Average) ↓
(3) Set the DC voltage generator as follows.
[Confirmation]Check that the measured value is within the range as specified in the table below.DL708E settings DC voltage genetaror Permissible rangeV/DIV Offset voltage Output voltage (Test spec.)
5 mV/div 30 mV 30 mV 30 mV ± 0.4 mV1 V/div 3 V 3 V 3 V ± 20 mV
2.16 High-Resolution Isolation Module (701853) Test
SM 701820-01E2-24
Residual Noise Test (701853)
[Specifications]Larger of ±50 µV or ±0.01 div, typical value
[Connection]
Not connected.
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the
“Initialize” soft key.)
[Procedure]Test each channel in the same manner.Set the DL708E as shown below.
T/DIV 1 ms/div
ACQ NormalV/DIV 5 mV/div Channel to be testedPositon 0 div ↓Coupling GND ↓Bandwidth Full ↓Offset 0 mV ↓MEASURE PP(Peak to Peak) ↓
[Confirmation]Check that the measured value is within 100 µVp-p.
2.16 High-Resolution Isolation Module (701853) Test
SM 701820-01E 2-25
2
PE
RF
OR
MA
NC
E T
ES
T
DC Accuracy Test (701853) (bandwidth limit Auto excluded)[Specifications]±(0.3% of 8 div) at 20 V/div to 20 mV/div±(0.5% of 8 div) at 10 mV/div±(1% of 8 div) at 5 mV/div
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
DC voltage generator 1 DC accuracy ±0.03% Yokogawa wavetek 9100 —
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
Conversion adapter 2 — Yokogawa 366922 Banana(male) to BNC
[Connection]
BNC-BNC cable 366924
DC voltage generator9100
Conversion adapter 366922
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the
“Initialize” option.)
[Procedure]Test each channel in the same manner.(1) Connect the output cable of the DC voltage generator to the channel to be tested as
shown above.(2) Set the DL708E as shown below.
T/DIV 1 ms/divACQ Average
Probe 1 : 1 Channel to be testedBandwidth 400 Hz ↓MEASURE Avg(Average) ↓
(3) Set the DC voltage generator as follows.
[Confirmation]Check that the measured value is within the range as specified in the table below.DL708E settings DC voltage generator Permissible rangeV/DIV Output voltage (Test spec.)
5 mV/div 19 mV 18.6 to 19.4 mV↓ 0 mV –0.4 to 0.4 mV↓ –19 mV –19.4 to –18.6 mV
10 mV/div 38 mV 37.6 to 38.4 mV↓ 0 mV –0.4 to 0.4 mV↓ –38 mV –38.4 to –37.6 mV
20 mV/div 76 mV 75.52 to 76.48 mV↓ 0 mV –0.48 to 0.48 mV↓ –76 mV –76.48 to –75.52 mV
50 mV/div 190 mV 188.8 to 191.2 mV↓ 0 mV –1.2 to 1.2 mV↓ –190 mV –191.2 to –188.8 mV
100 mV/div 380 mV 377.6 to 382.4 mV↓ 0 V –2.4 to 2.4 V↓ –380 mV –382.4 to –377.6 mV
200 mV/div 760 mV 755.2 to 764.8 mV↓ 0 V –4.8 to 4.8 V↓ –760 mV –764.8 to –755.2 mV
500 mV/div 1.9 V 1.888 to 1.912 V↓ 0 V –0.012 to 0.012 V↓ –1.9 V –1.912 to –1.888 V
1 V/div 3.2 V 3.176 to 3.224 V↓ 0 V –0.024 to 0.024 V↓ –3.2 V –3.224 to 3.176 V
2.16 High-Resolution Isolation Module (701853) Test
SM 701820-01E2-26
Frequency Characteristics Test (701853)
[Specifications]DC to 40 kHz (–3 dB) at 20 V/div to 10 mV/div
DC to 30 kHz (–3 dB) at 5 mV/div
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
Standard signal 1 ±0.1 dB : 100 kHz or less Yokogawa FG200 —generator
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
Conversion adapter 1 — Yokogawa 366922 Banana(male) to BNC
[Connection]
BNC-BNC cable 366924
Standard signal generatorFG200
Conversion adapter 366922
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the
“Initialize” option.)
[Procedure]Test each channel in the same manner.
(1) Connect the output cable of the standard signal generator to the channel to betested as shown above.
(2) Set the DL708E as shown below.
Probe 1 : 1 Channel to be testedBandwidth Full ↓MEASURE PP(Peak to Peak) ↓
(3) Set the standard signal generator as shown below.Sine waveAmplitude 30 mVp-p (when voltage-axis sensitivity is 5 mV/div.),
6 Vp-p (when voltage-axis sensitivity is 1 V/div.)
[Confirmation]Check that the measured value is within the range as specified in the table below.Calculate the difference between the PP value at 100 Hz (reference value) and the oneat each frequency [<2>, <4>].DL708E settings Standard signal generator DL708ET/DIV V/DIV Frequency Measured value PP
1 ms/div(100 kS/s) 5 mV/div 100 Hz <1>↓ ↓ 30.1 kHz <2>↓ 1 V/div 100 Hz <3>↓ ↓ 40.1 kHz <4>
Calculation Judgment
20log(<2>/<1>) –3.0 dB or lower
20log(<4>/<3>) –3.0 dB or lower
2.16 High-Resolution Isolation Module (701853) Test
SM 701820-01E 2-27
2
PE
RF
OR
MA
NC
E T
ES
T
Withstand Voltage Test (701853)
[Specifications]3700 VAC for 1 minute (between input and chassis GND)
[Equipment required]Equipment Qty Mandatory specifications Recommended
Withstand voltage 1 3700 VAC 10 mA limit Kikusui TOS-8750tester
Banana(male) 1 Short-circuit H and L —-BNC conversion of BNC cable
[Connection]Jig with short-circuited H and L of BNC cable.
Not connected to L/N
Withstandvoltage tester
H
H
L
L
Attached AC cable
(connected tomain unit GND)
[Connection method](1) Prepare the jig with short-circuited H and L of BNC cable.
(2) Connect H of the withstand voltage tester to H/L as shown above.(3) Connect G (GND) of the attached AC cable to L of the withstand voltage tester.(4) The power SW of DL708E should be turned off.
[Withstand voltage tester setting]3700 VAC (60 Hz) 10 mA limit
[Confirmation]Check that the limiter does not operate when voltage is applied for one minute.
WARNINGPerforming this test is very dangerous due to handling high voltage.The withstand voltage test is to be performed only by the experienced personnel
who thoroughly understand the test method.Be sure not to touch the live-wire.The above connection is provided for explanatory purpose.
To avoid a risk of high voltage, please prepare a jig which has no exposed live-wire.
2.16 High-Resolution Isolation Module (701853) Test
SM 701820-01E2-28
2.17 Temperature Module (701860) Test
2.17 Temperature Module (701860) TestAccuracy Test 1 (701860)
[Specifications]±(0.2% of reading + 1.5 °C), except ±(0.5% of reading + 1.5 °C) at –200 °C to 0 °C, forthermocouple type J, E, and U using on this test.
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
DC voltage generator 1 DC accuracy ±0.03% Yokogawa wavetek 9100 —
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
Conversion adapter 2 — Yokogawa 366922 Banana(male) to BNC
[Connection]
BNC-BNC cable 366924
DC voltage generator9100
Conversion adapter 366922
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each channel in the same manner.
(1) Connect the output cable of the DC voltage generator to the channel to be tested asshown above.
(2) Set the DL708E as shown below.
Record Length 1 kWT/DIV 2 s/divType According to settings below Channel to be tested
Unit °C ↓Bandwidth Full ↓RJC OFF ↓MEASURE Avg (Average) ↓
(3) Set the DC voltage generator as follows.
[Confirmation]Check that the measured value is within the permissible range for each thermocoupletype and applied voltage.DL708E settings DC voltage generator Permissible rangeThermocouple type Output voltage (Test spec.)
J 63792 µV 1096.3 to 1103.7 °C↓ 0 µV –1.5 to 1.5 °C
E –8825 µV –202.5 to 197.5 °C
U 2100 µV 397.7 to 402.3 °C↓ 0 µV –1.5 to 1.5 °C↓ –5603 µV –202.5 to –197.5 °C
SM 701820-01E 2-29
2
PE
RF
OR
MA
NC
E T
ES
T
Accuracy Test 2 (701860)
[Specifications]± Reference junction compensation accuracy (when input terminal temperature is
balanced) ± Measurement accuracyFor thermocouple T,• Reference junction compensation accuracy (when input terminal temperature is
balanced) : ±1.0 °C• Measurement accuracy : ±(0.2% of reading + 1.5 °C)
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
Thermocouple T 1 Thermocouple T — —
0 °C reference 1 Accuracy ±0.1 °C ZERO-CON ZC-114 —temperature unit Range 0 °C
[Connection]
Thermocouple T
0 °C referencetemperature unit
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each channel in the same manner.(1) Connect the thermocouple to the channel to be tested as shown above.
(2) Set the DL708E as shown below.Record Length 100 kWT/DIV 100 mV/div
Type T Channel to be testedUnit °C ↓Bandwidth 2 Hz ↓RJC ON ↓MEASURE Avg (Average) ↓
[Confirmation]Check that the measured value is within ±2.5 °C.
2.17 Temperature Module (701860) Test
SM 701820-01E2-30
Frequency Characteristics Test (701860)
[Specifications]Attenuation ratio for 2 Hz to FULL of the bandwidth : –50 dB or lower at 50/60 Hz
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
Standard signal 1 ±0.2 dB : 100 kHz or less Yokogawa FG200 —generator
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
Conversion adapter 1 — Yokogawa 366922 Banana(male) to BNC
[Connection]
BNC-BNC cable 366924
Standard signal generatorFG200
Conversion adapter 366922
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each channel in the same manner.(1) Connect the output cable of the standard signal generator to the channel to be
tested as shown above.(2) Set the DL708E as shown below.
Record Length 1 kW
T/DIV 50 ms/divType K Channel to be testedUnit °C ↓Bandwidth According to settings below ↓RJC OFF ↓MEASURE PP(Peak to Peak) ↓
(3) Set the standard signal generator as shown below.Sine waveFrequency According to settings below.
Amplitude 40 mVp-p (For FG200, amplitude ; set to 4 Vp-p, and outputattenuator ; set to 1/100.)
Offset 20 mV
[Confirmation]Check that the measured value is within the range as specified in the table below.Standard signal generator DL708E settingsFrequency Bandwidth Measured value PP
50 Hz Full 809 to 1003 °C↓ 2 Hz 1.6 °C or less
2.17 Temperature Module (701860) Test
SM 701820-01E 2-31
2
PE
RF
OR
MA
NC
E T
ES
T
2.18 Logic Input Module (701870) Test2.18.1 Using High-Speed Logic Probe (700986)
Maximum Input Voltage Test (701870 & 700986)
[Specifications]30 Vrms
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
AC voltage generator 1 30 VAC ± 0.03%, square wave Yokogawa wavetek9100—
High-speed logic 1 — Yokogawa 700986 —probe
Connecting lead 1 — Yokogawa B9879PX Converting to alligator clip
[Connection]
AC voltage generator9100
High-speed logic probe700986
Connecting leadB9879PX
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each bit in the same manner.(1) Connect the output lead of the AC voltage generator to the bit to be tested as shown
above.(2) Set the DL708E as shown below.
Record Length 1 kW
T/DIV 10 ms/div(3) Set the AC voltage generator as shown below.
Square wave
Amplitude 30 VACFrequency 50 Hz
[Confirmation]Confirm that 50-Hz square waves appear for all bits.
2.18 Logic Input Module (701870) Test
SM 701820-01E2-32
Response Time Test (701870 & 700986)
[Specifications]1 µs or less
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
Standard signal 1 ±0.2 dB : 100 kHz or less Yokogawa FG200 —generator
High-speed logic 1 — Yokogawa 700986 —probe
High-speed module 1 — Yokogawa 701851 —
Connecting lead 1 — Yokogawa B9879PX Converting to alligator clip
Conversion adapter 1 — Yokogawa 366921 BNC to banana(female)
T-shaped adapter 1 — Yokogawa 366923 BNC parallel connection
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
[Connection]
Standard signal generatorFG200
High-speed logic probe700986
Connecting leadB9879PX
Conversion adapter366921
T-Shaped adapter366923
BNC-BNC cable 366924
To 701851
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each bit in the same manner.
(1) Install the high-speed module 701851 into the CH5 slot.(2) Connect the output lead and cable of the standard signal generator to the bit to be
tested as shown above.
(3) Set the DL708E as shown below.Record Length 1 kWT/DIV 200 ns/div
Trigger Source : CH5, Level : 1.4 V, Slope : Falling(4) Set the standard signal generator as shown below.
Square wave
Frequency 100 kHzAmplitude 5 Vp-pOffset 2.5 V
(5) Set the high-speed module 701851 as shown below.Probe 1 : 1V/DIV 2 V/div
[Confirmation]Check that the falling edge of the logic waveform appears within ±1µs of the trigger point.
2.18 Logic Input Module (701870) Test
SM 701820-01E 2-33
2
PE
RF
OR
MA
NC
E T
ES
T
2.18.2 Using Isolation Logic Probe (700987)
Response Time Test for DC Inputting (701870 & 700987)
[Specifications]1 ms or less
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
AC voltage generator 1 250 VAC ± 0.03% Yokogawa wavetek 9100 —
Isolation logic probe 1 — Yokogawa 700987 —
High-speed module 1 — Yokogawa 701851 —
Measurement lead 1 — Yokogawa 758917 —
Alligator clip adapter 1 — Yokogawa 758922 Banana to alligator clip
Probe 1 1 m or less Yokogawa 700929 —
[Connection]
AC voltage generator9100
Isolation logic probe700987
Measurement lead758917
Alligator clipadapter758922Probe 700929
To 701851
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each bit in the same manner.(1) Install the high-speed module 701851 into the CH5 slot.
(2) Connect the output lead and cable of the AC voltage generator to the bit to be testedas shown above.
(3) Set the DL708E as shown below.
Record Length 1 kWT/DIV 500 µs/divTrigger Source : CH5, Level : 100 V, Slope : Falling
(4) Set the Isolation logic probe 700987 as shown below.Input switching DC
(5) Set the AC voltage generator as shown below.
Square waveFrequency 50 HzAmplitude 250 V
Offset 125 V(6) Set the high-speed module 701851 as shown below.
Probe 100 : 1
V/DIV 100 V/div
[Confirmation]Check that the falling edge of the logic waveform appears within ±1 ms of the fallingedge of the 701851’s waveform.
2.18 Logic Input Module (701870) Test
SM 701820-01E2-34
Response Time Test for AC Inputting (701870 & 700987)
[Specifications]20 ms or less
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
AC voltage generator 1 250 VAC ± 0.03% Yokogawa wavetek 9100 —
Isolation logic probe 1 — Yokogawa 700987 —
High-speed module 1 — Yokogawa 701851 —
Measurement lead 1 — Yokogawa 758917 —
Alligator clip adapter 1 — Yokogawa 758922 Banana to alligator clip
Probe 1 1 m or less Yokogawa 700929 —
[Connection]
AC voltage generator9100
Isolation logic probe700987
Measurement lead758917
Alligator clipadapter758922Probe 700929
To 701851
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the
“Initialize” option.)
[Procedure]Test each bit in the same manner.(1) Install the high-speed module 701851 into the CH5 slot.(2) Connect the output lead and cable of the AC voltage generator to the bit to be tested
as shown above.(3) Set the DL708E as shown below.
Record Length 1 kW
T/DIV 10 ms/divTrigger Source : CH5, Level : 100 V, Slope : Falling
(4) Set the Isolation logic probe 700987 as shown below.
Input switching AC(5) Set the AC voltage generator as shown below.
Square wave
Frequency 10 HzAmplitude 250 VOffset 125 V
(6) Set the high-speed module 701851 as shown below.Probe 100 : 1V/DIV 100 V/div
[Confirmation]Check that the falling edge of the logic waveform appears within ±20 ms of the falling
edge of the 701851’s waveform.
2.18 Logic Input Module (701870) Test
SM 701820-01E 2-35
2
PE
RF
OR
MA
NC
E T
ES
T
Threshold Level Test for DC Inputting (701870 & 700987)
[Specifications]6V ± 50% (when H level is 9 V or more and L level is 3 V or less)
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
DC voltage generator 1 9 VDC ± 0.03% Yokogawa wavetek 9100 —
Isolation logic probe 1 — Yokogawa 700987 —
Measurement lead 1 — Yokogawa 758917 —
Alligator clip adapter 1 — Yokogawa 758922 Banana to alligator clip
[Connection]
DC voltage generator9100
Isolation logic probe700987
Measurement lead758917
Alligator clipadapter758922
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each bit in the same manner.(1) Connect the output lead and cable of the DC voltage generator to the bit to be tested
as shown above.(2) Set the DL708E as shown below.
Record Length 1 kW
T/DIV 500 µs/divTrigger AUTO
(3) Set the Isolation logic probe 700987 as shown below.
Input switching DC(4) Set the DC voltage generator as follows.
[Confirmation]Confirm that the Isolation logic probe 700987’s LED operates properly depending on theoutput voltage of the DC voltage generator as shown in the table below.DC voltage generator Corresponding bit’s Corresponding bit'sOutput voltage Signal level LED condition
9 V H level ON3 V L level OFF
2.18 Logic Input Module (701870) Test
SM 701820-01E2-36
Threshold Level Test for AC Inputting (701870 & 700987)
[Specifications]50VAC ± 50% (when H level is 75 VAC or more and L level is 25 VAC or less)
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
AC voltage generator 1 75 VAC ± 0.03% Yokogawa wavetek 9100 —
Isolation logic probe 1 — Yokogawa 700987 —
Measurement lead 1 — Yokogawa 758917 —
Alligator clip adapter 1 — Yokogawa 758922 Banana to alligator clip
[Connection]
DC voltage generator9100
Isolation logic probe700987
Measurement lead758917
Alligator clipadapter758922
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each bit in the same manner.(1) Connect the output lead and cable of the AC voltage generator to the bit to be tested
as shown above.(2) Set the DL708E as shown below.
Record Length 1 kW
T/DIV 500 µs/divTrigger AUTO
(3) Set the Isolation logic probe 700987 as shown below.
Input switching AC(4) Set the AC voltage generator as follows.
[Confirmation]Confirm that the Isolation logic probe 700987’s LED operates properly depending on theoutput voltage of the DC voltage generator as shown in the table below.AC voltage generator Corresponding bit’s Corresponding bit’sOutput voltage Signal level LED condition
75 VAC H level ON25 VAC L level OFF
2.18 Logic Input Module (701870) Test
SM 701820-01E 2-37
2
PE
RF
OR
MA
NC
E T
ES
T
2.19 Strain Module (701880) Test
* uSTR stands for 10–6 strain.
Maximum Input Voltage Test (701880)
[Specifications]5 V (DC+ACpeak)
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
DC voltage / current 1 DC accuracy : ±0.03% with Yokogawa 7651 —generator limiter for voltage and current
Shielded cable 1 Two cores shielded cable — For maximum input voltageand NDIS connector test
[Connection]
Shielded cable
DC voltage generator7651
NDIS connector
Strain module
Shielded cable
Connected toshield wire
1) Connection diagram for testing the Input+ (pin D)
H output terminal
L output terminal
G terminal
Bridge+
Sense+
Sense–
Bridge–
Input+
Input–
Floating common
Case (connector shell)
< <A
< <F
< <G
< <C
< <D
< <B
< <E
NDISconnector
DC voltage generator
Strain module
Shielded cable
2) Connection diagram for testing the Input– (pin B)
H output terminal
L output terminal
G terminal
Bridge+
Sense+
Sense–
Bridge–
Input+
Input–
Floating common
Case (connector shell)
< <A
< <F
< <G
< <C
< <D
< <B
< <E
NDISconnector
DC voltage generator
Connected toshield wire
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the
“Initialize” soft key.)
[Procedure]Test each channel in the same manner.(1) Prepare the shielded cable and NDIS connector.(2) Connect the output cable of the DC voltage generator to the channel to be tested as
shown above.
2.19 Strain Module (701880) Test
SM 701820-01E2-38
(3) Set the DL708E as shown below.Range 1000 uSTR Channel to be testedExcitation 2 V ↓
(4) Set the DC voltage generator as follows.
[Confirmation]Confirm each terminal’s allowable voltage range and protective circuit operation.
• Confirming the allowable voltage range of Input+ (pin D)
(1) Short Input– (pin B) and Floating common (pin E).(2) Set the DC source to voltage mode and set the current limit to 5 mA.(3) Apply +5 V between Input+ (pin D) and Floating common (pin E), and check that
the current limit of the DC source does not operate.(4) Apply –5 V between Input+ (pin D) and Floating common (pin E), and check that
the current limit of the DC source does not operate.
• Confirming the protective circuit of Input+ (pin D)(1) Short Input– (pin B) and Floating common (pin E).
(2) Set the DC source to current mode and set the voltage limit to 5 V.(3) Allow +1 mA to flow between Input+ (pin D) and Floating common (pin E), and
check that the voltage limit of the DC source operates.
(4) Allow –1 mA to flow between Input+ (pin D) and Floating common (pin E), andcheck that the voltage limit of the DC source operates.
• Confirming the allowable voltage range of Input– (pin B)(1) Short Input+ (pin D) and Floating common (pin E).(2) Set the DC source to voltage mode and set the current limit to 5 mA.
(3) Apply +5 V between Input– (pin B) and Floating common (pin E), and check thatthe current limit of the DC source does not operate.
(4) Apply –5 V between Input– (pin B) and Floating common (pin E), and check that
the current limit of the DC source does not operate.
• Confirming the protective circuit of Input– (pin B)
(1) Short Input+ (pin D) and Floating common (pin E).(2) Set the DC source to current mode and set the voltage limit to 5 V.(3) Allow +1 mA to flow between Input– (pin B) and Floating common (pin E), and
check that the voltage limit of the DC source operates.(4) Allow –1 mA to flow between Input– (pin B) and Floating common (pin E), and
check that the voltage limit of the DC source operates.
2.19 Strain Module (701880) Test
SM 701820-01E 2-39
2
PE
RF
OR
MA
NC
E T
ES
T
Auto Balance Test (701880)
[Specifications]±10000 uSTR
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
DC voltage generator 1 DC accuracy : ±0.03% Yokogawa 7651 —
Shielded cable 1 Two cores shielded cable — For auto balance testand NDIS connector
[Connection]
Shielded cable
DC voltage generator7651
NDIS connector
Resistance
Strain module
Shielded cable
Connection diagram
H output terminal
L output terminal
G terminal
Bridge+
Sense+
Sense–
Bridge–
Input+
Input–
Floating common
Case (connector shell)
< <A
< <F
< <G
< <C
< <D
< <B
< <E
NDISconnector
DC voltage generatorResistance : 350 Ω ± 5%
Connected toshield wire
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each channel in the same manner.
(1) Connect the output cable of the DC voltage generator to the channel to be tested asshown above.
(2) Set the DL708E as shown below.
T/DIV 10 ms/divACQ NormalRange 5000 uSTR Channel to be tested
Excitation 5 V ↓Bandwidth Full ↓
(3) Set the DC voltage generator as follows.
[Confirmation]Confirm the operation in order according to the table below.No. DC voltage generator
Output voltage Check item Description
1 0 mV Perform calibration*1 Check that no errors occur.*2
2 ↓ Perform balancing*1 Check that balancing error does not occur.*3
3 29 mV Perform balancing Check that balancing error does not occur.
4 –29 mV Perform balancing Check that balancing error does not occur.
*1 See User's Manual IM701820-01E for the operating procedures on calibration and balancing.*2 Confirm that no error messages appear after the calibration.*3 Confirm that “Automatic balancing is complete” appears on the screen when the balancing completes.
2.19 Strain Module (701880) Test
SM 701820-01E2-40
DC Accuracy Test (701880)
[Specifications]±1.0% of FS ± 40 uSTR
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
Digital multimeter 1 5 1/2-digits equivalent Yokogawa 7552 —
Measurement lead 1 Two wire lead — For bridge voltage testand NDIS connector
DC voltage generator 1 DC accuracy : ±0.03% Yokogawa 7651 —
Shielded cable 1 Two cores shielded cable — For DC accuracy testand NDIS connector
[Connection]
Measurement lead
Digital multimeter7552
NDIS connector
Resistance
Strain module
Connection diagram 1
H output terminal
L output terminal
Bridge+
Sense+
Sense–
Bridge–
Input+
Input–
Floating common
Case (connector shell)
< <A
< <F
< <G
< <C
< <D
< <B
< <E
NDISconnector Digital multimeter
Resistance*
* Resistance• 120 Ω ± 5% for bridge voltage 2 V• 350 Ω ± 5% for bridge voltage 5 V
Shielded cable
DC voltage generator7651
NDIS donnector
Resistance
Connection diagram 2
Strain module
Shielded cable H output terminal
L output terminal
G terminal
Bridge+
Sense+
Sense–
Bridge–
Input+
Input–
Floating common
Case (connector shell)
< <A
< <F
< <G
< <C
< <D
< <B
< <E
NDISconnector
DC voltage generatorResistance : 350 Ω ± 5%
Connected toshield wire
2.19 Strain Module (701880) Test
SM 701820-01E 2-41
2
PE
RF
OR
MA
NC
E T
ES
T
[[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure](1) Connect the measurement leads according to connection diagram 1, and set the
Resistance to 120 Ω.(2) Set the excitation voltage to Excitation 2 V, and measure the bridge voltage
(Excitation) with a digital multimeter.(3) Calculate the Scaling factor (SF2) when the excitation voltage is set to Excitation 2 V.
(SF2 = Digital multimeter’s measurement value ÷ 2.0000)
(4) Connect the measurement leads according to connection diagram 1, and set theResistance to 350 Ω.
(5) Set the excitation voltage to Excitation 5 V, and measure the bridge voltage
(Excitation) with a digital multimeter.(6) Calculate the Scaling factor (SF5) when the excitation voltage is set to Excitation 5 V.
(SF5 = Digital multimeter’s measurement value ÷ 5.0000)
(7) Connect the output cable of the digital voltage generator to the channel to be testedas shown connection diagram 2.
(8) Set the DL708E as shown below.
T/DIV 1 ms/divACQ Average (Count=8)Record length 1 kW
Gauge Factor 2.00 Channel to be testedMEASURE Avg(Average) ↓Set the other setting items as follows. ↓
(9) Set the DC voltage generator as follows.Set the output voltage of the DC voltage generator to the value adjusted by theScaling factor.
[Confirmation]Perform the following tests in order and check that the results are within the permissiblerange. Perform balancing each time you change the Strain module’s settings.No. DC voltage generator Strain module Permissible range
Output voltage Range Excitation Bandwidth Scale (Test spec.)Upper / Lower
[mV] [uSTR] [V] [Hz] [uSTR] [uSTR]
1 0.000 1000 2 10 1350 / –1350 (Only perform balancing)2 1.000 × SF2 ↓ ↓ ↓ ↓ 940 to 10603 0.000 ↓ ↓ ↓ ↓ –60 to 604 –1.000 × SF2 ↓ ↓ ↓ ↓ –1060 to –940
5 0.000 10000 2 Full 13500 / –13500 (Only perform balancing)6 10.000 × SF2 ↓ ↓ ↓ ↓ 9760 to 102407 0.000 ↓ ↓ ↓ ↓ –240 to 2408 –10.000 × SF2 ↓ ↓ ↓ ↓ –10240 to –9760
9 0.000 1000 5 100 1350 / –1350 (Only perform balancing)10 2.500 × SF5 ↓ ↓ ↓ ↓ 940 to 106011 0.000 ↓ ↓ ↓ ↓ –60 to 6012 –2.500 × SF5 ↓ ↓ ↓ ↓ –1060 to –940
13 0.000 2000 5 100 2700 / –2700 (Only perform balancing)14 5.000 × SF5 ↓ ↓ ↓ ↓ 1920 to 208015 0.000 ↓ ↓ ↓ ↓ –80 to 8016 –5.000 × SF5 ↓ ↓ ↓ ↓ –2080 to –1920
17 0.000 5000 5 100 6750 / –6750 (Only perform balancing)18 12.50 × SF5 ↓ ↓ ↓ ↓ 4860 to 514019 0.000 ↓ ↓ ↓ ↓ –140 to 14020 –12.50 × SF5 ↓ ↓ ↓ ↓ –5140 to –4860
21 0.000 10000 5 100 13500 / –13500 (Only perform balancing)22 25.00 × SF5 ↓ ↓ ↓ ↓ 9760 to 1024023 0.000 ↓ ↓ ↓ ↓ –240 to 24024 –25.00 × SF5 ↓ ↓ ↓ ↓ –10240 to –9760
25 0.000 20000 5 1k 27000 / –27000 (Only perform balancing)26 50.00 × SF5 ↓ ↓ ↓ ↓ 19560 to 2044027 0.000 ↓ ↓ ↓ ↓ –440 to 44028 –50.00 × SF5 ↓ ↓ ↓ ↓ –20440 to –19560
2.19 Strain Module (701880) Test
SM 701820-01E2-42
Frequency Characteristics Test (701880)
[Specifications]DC to 20 kHz (–3 dB)
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
Standard signal 1 ±0.2 dB : 100 kHz or less Yokogawa FG200 —generator
Shielded cable 1 Single core shielded cable — For frequency characteristicswith BNC and NDIS connector test
[Connection]
Shielded cable
Standard signal generatorFG200
NDIS connector
Strain module
Shielded cable
Connection diagram
Bridge+
Sense+
Sense–
Bridge–
Input+
Input–
Floating common
< <A
< <F
< <G
< <C
< <D
< <B
< <E
NDISconnector
Standard signal generatorBNC connector
Connected toshield wire
[Initialization]Before performing this test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each channel in the same manner.
(1) Connect the output cable of the standard signal generator to the channel to betested as shown above.
(2) Set the standard signal generator’s output voltage to 0 V and start the generator.
(3) Set the DL708E as shown below.T/DIV 1 ms/divACQ Normal
Range 20000 uSTR Channel to be testedExcitation 5 V ↓Gauge Factor 2.00 ↓Bandwidth Full ↓MEASURE PP(Peak to Peak) ↓
(4) Perform balancing.
(5) Set the standard signal generator as shown below.Sine waveFrequency According to settings below
Amplitude 100 mVp-pOffset 0 V
2.19 Strain Module (701880) Test
SM 701820-01E 2-43
2
PE
RF
OR
MA
NC
E T
ES
T
[Confirmation]Check that the measured value is within the range as specified in the table below.Calculate the difference between the PP value at 100 Hz (reference value) and the one
at each frequency [<2>, <3>, <4>].Standard signal generator DL708EFrequency Measured value PP Calculation Judgement
100 Hz <1> — —1 kHz <2> <2>/<1> 1 ± 0.0220 kHz <3> 20log(<3>/<1>) –1.5 to –3 dB200 kHz <4> 20log(<4>/<1>) –30 dB or lower
2.19 Strain Module (701880) Test
SM 701820-01E 3-1
3
AD
JUS
TM
EN
T
3.1 Introduction / 3.2 Test Environment / 3.3 Equipment Required
Chapter 3 ADJUSTMENT3.1 Introduction3.2 Test Environment3.3 Equipment Required
3.4 High-Speed Isolation Module (701850) Adjustment3.5 High-Speed Module (701851) Adjustment3.6 Temperature Module (701860) Adjustment
3.7 Logic Input Module (701870) Adjustment
3.1 IntroductionThis chapter describes the adjustment of DL708E.Adjustment procedures for the following units are omitted because they need noadjustment.
701820 DL708E main unit701752 High-Resolution, High-Voltage, Isolation module
701853 High-Resolution, Isolation module701880 Strain module
3.2 Test EnvironmentReference operating conditions (Test environment: performance test items)Warm up time More than 30 minutes
Ambient temperature 23 ± 5°CHumidity 55 ± 10%RHVoltage of power supply Specified voltage ± 1% or less
3.3 Equipment RequiredGeneral purpose equipment (Cables)Equipment Qty Mandatory specifications Recommended Remarks
Measurement lead 1 For 42 V or more Yokogawa 758917 Banana to alligator clip and 758922
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
Conversion adapter 2 — Yokogawa 366922 Banana(male) to BNC
Equipment for 701850 / 701851 (High-Speed Isolation module / High-Speedmodule)Equipment Qty Mandatory specifications Recommended Remarks
DC voltage generator 1 DC accuracy : ±0.03% Yokogawa wavetek 9100 Yokogawa wavetek 9100or Yokogawa 7651 offers complete compatibility.
Standard signal 1 ±0.5 dB : 10 MHz or less Yokogawa FG200 —generator ±1 dB : more than 10 MHz
Equipment for 701860 (Temperature module)Equipment Qty Mandatory specifications Recommended Remarks
DC voltage generator 1 DC accuracy : ±0.03% Yokogawa wavetek 9100 Yokogawa wavetek 9100or Yokogawa 7651 offers complete compatibility.
SM 701820-01E3-2
3.3 Equipment Required / 3.4 High-Speed Isolation Module (701850) Adjustment
Equipment for 701870 (Logic Input Module)Equipment Qty Mandatory specifications Recommended Remarks
DC voltage generator 1 DC accuracy : ±0.03% Yokogawa wavetek 9100 —
AC voltage generator 1 50 VAC ± 0.03% Yokogawa wavetek 9100 —
Special purpose 1 Isolation logic probe Yokogawa 700987 Test includes logic probe.logic probe
3.4 High-Speed Isolation Module (701850) AdjustmentDAC (D/A converter) Span Adjustment (701850)
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
DC voltage generator 1 DC accuracy : ±0.03% Yokogawa wavetek 9100 —
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
Conversion adapter 1 — Yokogawa 366922 Banana(male) to BNC
[Connection]
BNC-BNC cable 366924
DC voltage generator9100
Conversion adapter 366922
[Initialization]Before performing the test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Adjust each channel in the same manner.(1) Connect the output cable of the DC voltage generator to the channel to be tested as
shown above.(2) Set the DL708E as shown below.
T/DIV 200 µs/div
ACQ AverageV/DIV 200 mV/div Channel to be testedProbe 1 : 1 ↓Bandwidth 500 kHz ↓Offset +5 V ↓MEASURE Avg (Average) ↓
(3) Set the output of the DC voltage generator to +5 V.
SM 701820-01E 3-3
3
AD
JUS
TM
EN
T
3.4 High-Speed Isolation Module (701850) Adjustment
(4) Adjust the RV30 channel to be tested so that the measured value becomes 5 V ± 5mV.
RV30 CV1 to 4 CV5 CV101
High-speed isolation moduleAdjustment point location diagram
CV100RV31
SM 701820-01E3-4
3.4 High-Speed Isolation Module (701850) Adjustment
Frequency Characteristics 1/1 Range Adjustment (701850)
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
Standard signal 1 ±0.5 dB : 10 MHz or less Yokogawa FG200 —generator ±1 dB : more than 10 MHz
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
[Connection]
BNC-BNC cable 366924
Standard signal generatorFG200
[Initialization]Before performing the test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each channel in the same manner.
(1) Connect the output cable of the standard signal generator to the channel to betested as shown above.
(2) Set the DL708E as shown below.
T/DIV 200 µs/divV/DIV 10 mV/div Channel to be testedPosition 0 div ↓Coupling DC ↓Bandwidth Full ↓Offset 0 mV ↓
(3) FG settingsOutput Waveform Square waveFrequency 1 kHz
Amplitude 60 mVp-p (FG settings: 1/100, 6 Vp-p)(4) Adjust the CV5 channel to be tested so that the rising point of the waveform
becomes flat as shown in the figure below.
(Flatness: amplitude ±1% or less)
Waveform rising point
Adjust CV5 so that this rising point reaches an amplitude of ±1% or less.
SM 701820-01E 3-5
3
AD
JUS
TM
EN
T
3.4 High-Speed Isolation Module (701850) Adjustment
Frequency Characteristics 1/10 Range Adjustment (701850)
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
Standard signal 1 ±0.5 dB : 10 MHz or less Yokogawa FG200 —generator ±1 dB : more than 10 MHz
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
[Connection]
BNC-BNC cable 366924
Standard signal generatorFG200
[Initialization]Before performing the test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each channel in the same manner.
(1) Connect the output cable of the standard signal generator to the channel to betested as shown above.
(2) Set the DL708E as shown below.
T/DIV 200 µs/divV/DIV 100 mV/div Channel to be testedPosition 0 div ↓Coupling DC ↓Bandwidth Full ↓Offset 0 mV ↓
(3) FG settingsOutput Waveform Square waveFrequency 1 kHz
Amplitude 600 mVp-p (FG settings: 1/10, 6 Vp-p)(4) Adjust the CV1 and CV2 channels to be tested so that the rising point of the
waveform becomes flat, refer to the figure in the preceding page.
(Flatness: amplitude ±1% or less)
SM 701820-01E3-6
3.4 High-Speed Isolation Module (701850) Adjustment
Frequency Characteristics 1/100 Range Adjustment (701850)
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
Standard signal 1 ±0.5 dB : 10 MHz or less Yokogawa FG200 —generator ±1 dB : more than 10 MHz
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
[Connection]
BNC-BNC cable 366924
Standard signal generatorFG200
[Initialization]Before performing the test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each channel in the same manner.
(1) Connect the output cable of the standard signal generator to the channel to betested as shown above.
(2) Set the DL708E as shown below.
T/DIV 200 µs/divV/DIV 1 V/div Channel to be testedPosition 0 div ↓Coupling DC ↓Bandwidth Full ↓Offset 0 mV ↓
(3) FG settingsOutput Waveform Square waveFrequency 1 kHz
Amplitude 6 Vp-p(4) Adjust the CV3 and CV4 channels to be tested so that the rising point of the
waveform becomes flat, refer to the figure in the preceding page.
(Flatness: amplitude ±1% or less)
SM 701820-01E 3-7
3
AD
JUS
TM
EN
T
3.4 High-Speed Isolation Module (701850) Adjustment
Square Wave Adjustment (701850)
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
Standard signal 1 ±0.5 dB : 10 MHz or less Yokogawa FG200 —generator ±1 dB : more than 10 MHz
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
[Connection]
BNC-BNC cable 366924
Standard signal generatorFG200
[Initialization]Before performing the test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each channel in the same manner.
(1) Connect the output cable of the standard signal generator to the channel to betested as shown above.
(2) Set the DL708E as shown below.
T/DIV 10 µs/divInterpolate Straight lineV/DIV 50 mV/div Channel to be tested
Position 0 div ↓Coupling AC ↓Bandwidth Full ↓Offset 0 mV ↓
(3) FG settingsOutput Waveform Square wave
Frequency 100 kHzAmplitude 300 mVp-p (FG settings: 1/10, 3 Vp-p)(The above settings is for without 50 Ω termination. If the cable is short and is not
influenced, no termination is required.)(4) Adjust CV100 channel to be tested so that the waveform overshoot reaches a
maximum.
(5) Adjust RV31 channel to be tested so that the waveform attenuation amplitudereaches a minimum.
(6) Adjust CV100 channel to be tested so that the waveform overshoot reaches a
minimum.(7) Repeat the procedures of 5 and 6 so that the waveform overshoot is 3% or less.
(Exceeding 3% instantaneously is allowable because of few sampling points, but do
not exceed 5%.)(8) Change the input signal as follows.
Output Waveform Sine wave
Amplitude 300 mVp-p (FG settings: 1/10, 3 Vp-p)Frequency 100 kHz
(9) Determine the value at 100 kHz and 2.003 MHz.
(10)Check that the value made up of 20Log (DATA (at 100 kHz)/DATA (at 2 MHz)) iswithin the range from +0.5 dB to –2.5 dB.
SM 701820-01E3-8
3.4 High-Speed Isolation Module (701850) Adjustment
CMRR Adjustment (701850)
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
Standard signal 1 ±0.5 dB : 10 MHz or less Yokogawa FG200 —generator ±1 dB : more than 10 MHz
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
[Connection]
BNC-BNC cable 366924
Standard signal generatorFG200
[Initialization]Before performing the test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each channel in the same manner.
(1) Connect the output cable of the standard signal generator to the channel to betested as shown above.
(2) Set the DL708E as shown below.
T/DIV 10 µs/divInterpolate Straight lineV/DIV 5 V/div Channel to be tested
Position 0 div ↓Coupling DC ↓Bandwidth Full ↓Offset 0 mV ↓
(3) FG settingsOutput Waveform Sine wave
Frequency 100 kHzAmplitude 100 Vrms
(4) Adjust CV101 so that the waveform of the main unit jig tube surface reaches a
minimum.
SM 701820-01E 3-9
3
AD
JUS
TM
EN
T
3.5 High-Speed Module (701851) Adjustment
3.5 High-speed Module (701851) AdjustmentDAC (D/A converter) Span (701851) Adjustment
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
DC voltage generator 1 DC accuracy : ±0.03% Yokogawa wavetek 9100 —
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
Conversion adapter 1 — Yokogawa 366922 Banana(male) to BNC
[Connection]
BNC-BNC cable 366924
DC voltage generator9100
Conversion adapter 366922
[Initialization]Before performing the test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Adjust each channel in the same manner.
(1) Connect the output cable of the DC voltage generator to the channel to be tested asshown above.
(2) Set the DL708E as shown below.
T/DIV 200 µs/divACQ AverageProbe 1 : 1 Channel to be tested
Bandwidth 500 kHz ↓MEASURE Avg (Average) ↓
(3) Set the output of the DC voltage generator to +5 V.
SM 701820-01E3-10
3.5 High-Speed Module (701851) Adjustment
(4) Adjust the RV30 channel to be tested so that the measured value becomes 5 V ± 5mV.
High-speed moduleAdjustment point location diagram
RV30 CV1 to 2 CV5CV4
SM 701820-01E 3-11
3
AD
JUS
TM
EN
T
3.5 High-Speed Module (701851) Adjustment
Frequency Characteristics 1/1 Range Adjustment (701851)
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
Standard signal 1 ±0.5 dB : 10 MHz or less Yokogawa FG200 —generator ±1 dB : more than 10 MHz
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
[Connection]
BNC-BNC cable 366924
Standard signal generatorFG200
[Initialization]Before performing the test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each channel in the same manner.
(1) Connect the output cable of the standard signal generator to the channel to betested as shown above.
(2) Set the DL708E as shown below.
T/DIV 200 µs/divV/DIV 10 mV/div Channel to be testedPosition 0 div ↓Coupling DC ↓Bandwidth Full ↓Offset 0 mV ↓
(3) FG settingsOutput Waveform Square waveFrequency 1 kHz
Amplitude 60 mVp-p (FG settings: 1/100, 6 Vp-p)(4) Adjust the CV5 channel to be tested so that the rising point of the waveform
becomes flat as shown in the figure below.
(Flatness: amplitude ±1% or less)
Waveform rising point
Adjust CV5 so that this rising point reaches an amplitude of ±1% or less.
SM 701820-01E3-12
Frequency Characteristics 1/10 Range Adjustment (701851)
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
Standard signal 1 ±0.5 dB : 10 MHz or less Yokogawa FG200 —generator ±1 dB : more than 10 MHz
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
[Connection]
BNC-BNC cable 366924
Standard signal generatorFG200
[Initialization]Before performing the test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each channel in the same manner.
(1) Connect the output cable of the standard signal generator to the channel to betested as shown above.
(2) Set the DL708E as shown below.
T/DIV 200 µs/divV/DIV 100 mV/div Channel to be testedPosition 0 div ↓Coupling DC ↓Bandwidth Full ↓Offset 0 mV ↓
(3) FG settingsOutput Waveform Square waveFrequency 1 kHz
Amplitude 600 mVp-p (FG settings: 1/10, 6 Vp-p)(4) Adjust the CV1 and CV2 channels to be tested so that the rising point of the
waveform becomes flat, refer to the figure in the preceding page.
(Flatness: amplitude ±1% or less)
3.5 High-Speed Module (701851) Adjustment
SM 701820-01E 3-13
3
AD
JUS
TM
EN
T
Frequency Characteristics 1/100 Range Adjustment (701851)
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
Standard signal 1 ±0.5 dB : 10 MHz or less Yokogawa FG200 —generator ±1 dB : more than 10 MHz
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
[Connection]
BNC-BNC cable 366924
Standard signal generatorFG200
[Initialization]Before performing the test, initialize the settings. (Press the SETUP key and select the“Initialize” option.)
[Procedure]Test each channel in the same manner.
(1) Connect the output cable of the standard signal generator to the channel to betested as shown above.
(2) Set the DL708E as shown below.
T/DIV 200 µs/divV/DIV 1 V/div Channel to be testedPosition 0 div ↓Coupling DC ↓Bandwidth Full ↓Offset 0 mV ↓
(3) FG settingsOutput Waveform Square waveFrequency 1 kHz
Amplitude 6 Vp-p(4) Adjust the CV4 channel to be tested so that the rising point of the waveform
becomes flat, refer to the figure in the preceding page.
(Flatness: amplitude ±1% or less)
3.5 High-Speed Module (701851) Adjustment
SM 701820-01E3-14
3.6 Temperature Module (701860) Adjustment
3.6 Temperature Module (701860) AdjustmentCalibration Value Write (701860)
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
DC voltage generator 1 DC accuracy : ±0.02% Yokogawa wavetek 9100 —
BNC-BNC cable 1 1 m or less Yokogawa 366924 —
Conversion adapter 1 — Yokogawa 366922 Banana(male) to BNC
[Connection]
BNC-BNC cable 366924
DC voltage generator9100
Conversion adapter 366922
[Initialization]Before performing the test, initialize the settings. (Press the SETUP key and select the
“Initialize” option.)
[Procedure]Adjust each channel in the same manner.(1) Connect the output cable of the DC voltage generator to the channel to be tested as
shown above.
(2) Set the DL708E to service mode.Press the MISC key.Press the “NEXT 1/2” soft key to display the self test soft key menu.
Press the “Self test” soft key to display the self test menu.Press the “Test Item” soft key to select “Key Board”.Press the “Soft Key” soft key to display the keyboard screen.
Input “********” using the jog shuttle and the SELECT key.Enter “ENT” to enter the string.Press the “Test Item” soft key to select “Module.”
(3) Press the “Select CH” soft key to select the channel to be tested. Press the “ModuleID” soft key to select Temp. Press the “Exec” soft key once.
(4) Input the full range voltage corresponding to each reference voltage values as the
following table. Press the “Pre-cal” soft key to acquire the calibration value at eachvoltage.
Reference voltage DC voltage generatorOutput voltage
20 mV 20 mV60 mV 60 mV2 V 2 V
(5) Press the “Pre-cal” soft key at all reference voltage before pressing the “Exec” softkey.
SM 701820-01E 3-15
3
AD
JUS
TM
EN
T
3.7 Logic Input Module (701870) Adjustment
3.7 Logic Input Module (701870) AdjustmentIsolation Logic Probe DC Adjustment (700987)
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
DC voltage generator 1 DC accuracy : ±0.03% Yokogawa wavetek 9100 —
Isolation logic probe 1 — Yokogawa 700987 —
Measurement lead 1 — Yokogawa 758917 Banana to alligator clipand 758922
[Connection]
DC voltage generator9100
Isolation logic probe700987
Measurement lead758917 and 758922
[Procedure]Adjust each bit in the same manner.(1) Connect the output cable of the DC voltage generator to respective signal input
terminal as shown above.(2) Switch the probe switch to DC.(3) Set the output of the DC voltage generator to 6 VDC.
(4) Adjust trimmer resistor R (C) so that the frequency measured at test point (B)corresponding to each bit (A) as the table below fits within 2.5 V ± 0.1 V.
Correspondence of bits and test points to the trimmer resistance for adjustment
Bit (A) Test point (B) R (C)
1 CON303-2pin R0192 CON303-3pin R0483 CON303-4pin R0784 CON303-5pin R1085 CON303-6pin R1696 CON303-7pin R1987 CON303-8pin R2288 CON303-9pin R258
Pin numbers 1 and 10 for CON303 are for GND use.
R019 R048 R078 R258R108 R169 R198 R228
Isolation logic probe DC adjustmentAdjustment point location diagram
Test pointCON303
SM 701820-01E3-16
3.7 Logic Input Module (701870) Adjustment
Isolation Logic Probe AC Adjustment (700987)
[Equipment required]Equipment Qty Mandatory specifications Recommended Remarks
AC voltage generator 1 50 VAC ± 0.03% Yokogawa wavetek 9100 —
Isolation logic probe 1 — Yokogawa 700987 —
Measurement lead 1 — Yokogawa 758917 Banana to alligator clipand 758922
[Connection]
AC voltage generator9100
Isolation logic probe700987
Measurement lead758917 and 758922
[Procedure]Adjust each bit in the same manner.(1) Connect the output cable of the AC voltage generator to respective signal input
terminal as shown above.(2) Switch the probe switch to AC.(3) Set the output of the AC voltage generator to 50 VAC, 40 Hz.
(4) Display the waveform by the DL708E.Adjust the trimmer resistor for adjustment (B) corresponding each bit (A) as the tablebelow so that the AC waveform fits as shown in upper figure on the next page.When applying 50 VAC and 35 Hz at this point, the waveform becomes as middle
figure on the next page.
Correspondence of bits to the trimmer resistance for adjustment
Bit (A) R (B)
1 R0232 R0523 R0824 R1125 R1736 R2027 R2328 R262
Locations of the trimmer resistance for adjustment are shown on the next page.
SM 701820-01E 3-17
3
AD
JUS
TM
EN
T
3.7 Logic Input Module (701870) Adjustment
The waveform should always be in H level.
Displaywaveform
L level ofdisplay form
DL708E screen (display only for the 1 bit)
The rectangular wave should be displayed.
Displaywaveform
L level ofdisplay form
DL708E screen (display only for the 1 bit)
R023 R052 R082 R262R112 R173 R202 R232
Isolation logic probe AC adjustmentAdjustment point location diagram
SM 701820-01E 4-1
4
PR
INC
IPL
E O
F O
PE
RA
TIO
N
4.1 Introduction / 4.2 Main Unit (701820)
Chapter 4 PRINCIPLE OF OPERATION4.1 Introduction4.2 Main Unit (701820)4.3 Modules
High-Speed Isolation Module (701850)High-Speed Module (701851)High-Resolution, High-Voltage, Isolation Module (701852)
High-Resolution, Isolation Module (701853)Temperature Module (701860)Logic Input Module (701870)
Strain Module (701880)4.4 Functions of Each Gate Array
Acquisition Data Processor Gate Array (ADP)
Acquisition Control Logic Gate Array (ACL)Graphic Control Gate Array (GCG)AD (701850)
Temperature Module Interface Controller (701860)
4.1 IntroductionThis chapter describes functions of DL708E. Descriptions for the main unit of DL708E(701820) and modules (701850, 701851, 701852, 701853, 701860, 701870, and701880) are separately described. Their functions and specifications are described
briefly. Functions of the various gate arrays used for DL708E are also briefly described.
4.2 Main Unit (701820)This section describes functions of the main unit of DL708E (701820). The blockdiagram of the 701820 is shown on the next page.
(1) CPU Board AssemblyThis is equipped with a CPU capability, data acquisition capability and external input andoutput capability. Most of the digital signals are processed on this board.
(2) SUB-CPU Board AssemblyThis connects modules to the CPU.
(3) INLET AssemblyPower is supplied to the main unit via this assembly. A filter and switch are also
installed.
(4) Power Board AssemblyPower is supplied to each board via this assembly.
(5) LCD CONNECT Board AssemblyThe LCD Board Assembly relays the connectors of the LCD signals.
(6) Key/Sub-key AssemblyThe key switches, LEDs, the rotary encoders and the jog shuttle and COMP terminal areinstalled on the Key/Sub-key assembly.
(7) Waveform Memory Module (optional)This is an extension memory for data acquisition. The capacity may vary depending onthe option.
SM 701820-01E4-2
4.2 Main Unit (701820)
(8) SCSI CONNECT Assembly (optional)External connector and terminator for SCSI are installed.
(9) SCSI DCDC Assembly (optional)Power supply board of HDD to change from 24 VDC to 12 VDC.
(10)LCD Assembly10.4 inch TFT LC display. Full display resolution: 640 × 480. Waveform displayresolution: 501 × 432.
(11) Printer AssemblyThe printer for 112 mm width (8 dots per mm) is equipped with a line dot thermal
sensitivity printer head.
(12)FDD AssemblyThe FDD Assembly supports the formats of 640 KB, 720 KB, 1.2 MB and 1.44 MB on3.5-inch disk.
(13)HDD Assembly (optional)Built in 3.5-inch HDD.
Inlet board Assy
Power board Assy
AC
DC
ADD
ADD
Module701850701851701852701853701860701870701880
AcquisionData
ProcessingASCI
CH1 to 4
AcquisionData
ProcessingASCI
CH5 to 8
FDD
SH2CPU
Main Memory4MB
DisplayASIC
SCSI InternalHDD
GP-IB
SerialI/F
Serial Port (RS-232)
ExternalPrinter I/F
DisplayRAM
10.4 ColorTFT Display
Built inprinter
Key/Sub keyBoard Assy
StandardACQ-Memory
400kW/CH
Acquision ControlLogic ASIC
Optional Memory1MW/CH to 4MW/CH
StandardACQ-Memory
400kW/CH
Optional Memory1MW/CH to 4MW/CH
SCSIPort
GP-IB Port
External Printer Port(Centro)
Ext trig in
Ext clk in
Trig out
SM 701820-01E 4-3
4
PR
INC
IPL
E O
F O
PE
RA
TIO
N
4.3 Modules
4.3 ModulesSpecifications of modules (701850, 701851, 701852, 701853, 701860, 701870, and701880) for DL708E are described in this section. These modules are shown on next
page.
High-Speed Isolation Module (701850)Maximum input voltage : 850 V(DC+ACpeak)Maximum allowable common mode voltage : 400 VrmsInput frequency band : 2 MHz
DC accuracy : ±1.5% (except for the case of 5 mV/div)
High-Speed Module (701851)Maximum input voltage : 250 V(DC+ACpeak)Input frequency band : 4 MHz (at attenuation of –1 dB)DC accuracy : ±1.5% (except for the case of 5 mV/div)
High-Resolution, High-Voltage, Isolation Module (701852)100 kS/s, 16 bits.
Maximum input voltage : 850 V(DC+ACpeak)Maximum allowable common mode voltage : 400 VrmsDC accuracy : ±0.5% (except for the case of 50 mV/div)
High-Resolution, Isolation Module (701853)100 kS/s, 16 bits.
Maximum input voltage : 100 V(DC+ACpeak)Maximum allowable in-phase voltage : 400 VrmsDC accuracy : ±0.3% (except for the case of 5 mV/div and
10 mV/div)
Temperature Module (701860)Isolation input. May correspond to 12 thermocouples.
Logic Input Module (701870)16-bit input by two 8-bit probes. A selection between the non-isolated type (700986) andthe isolated type (700987) is available for an input probe.
Strain Module (701880)Isolation inputBridge voltage : Selectable 2 V or 5 V
Auto balance range : ±10000 × 10–6 strainBridge voltage sensing : ApplicableGauge factor input : Applicable
Recommended bridge head : 700932, 700933
ATT ADFLT 16
Isolation Block
H
L
+OFFSET
AMP
+OFFSET
AMPATT AD
Isolator
FLT10
Isolation Block
ATT ADFLT
10+
OFFSET
AMP
ADFLT
Isolation Block
H
L
AMP
RJC
1chip
CPU
168
8
POD-A
POD-B
Bridge Excitation
ADFLT 16
Isolation Block
Floating Common
+–
AMP
DIFA+–
Strain Module 701880
High-Speed Isolation Module 701850
High-Speed Module 701851
High-Resolution, High-voltage, Isolation Module 701852 /High-Resolution, Isolation Module 701853
Anti-aliasing filter
Temperature Module 701860
Logic Input Module 701870High-Speed Logic Probe 700986
Isolation Logic Probe 700987
SM 701820-01E4-4
4.4 Functions of Each Gate Array
4.4 Functions of Each Gate ArrayAcquisition Data Processor Gate Array (ADP)
This is a CMOS gate array which aims to process the data from the A/D converters sothat the data can be written to the acquisition memories. One ADP gate array supports 4
channels. It also performs triggering, average, box-average and envelope functionprocessing.
Acquisition Control Logic Gate Array (ACL)This is a CMOS gate array which aims to generate and control memory addresses to beused for Acquisition, P-P compression, Histogram, Average processing and the accessby CPU. It also controls CPU, peripheral devices and the keyboard.
Graphic Control Gate Array (GCG)This is a CMOS gate array which performs the general graphic control as follows:• This enables the CPU to perform read/write control of VRAM for graphics and SRAM
for text.• To supplement rasters• To interpret commands for the waveform display area
• The pallet control (40 colors out of 512 are available.)• The control of VRAM refresh• The control of text and printers
AD (701850)This is a small sized CMOS IC. A variable gain amplifier, filter and AD converter ,etc. areincorporated in one chip.
Temperature Module Interface Controller (701860)This is a small sized CMOS gate array. This is used for parallel conversion of serial datafrom the analog section and interface with the CPU.
SM 701820-01E 5-1
5
TR
OU
BL
ES
HO
OT
ING
5.1 Introduction
Chapter 5 TROUBLESHOOTING5.1 Introduction5.2 Flow Chart5.3 Self Test
Self-diagnosis of Built-in PrinterSelf-diagnosis of Internal MemorySelf-diagnosis of Keyboard, Jog Shuttle, and Rotary Knob
5.1 IntroductionThis chapter describes troubleshooting items to solve presumable errors. In such cases,
assembly removal may be required. Please heed the following warning.
WARNINGReplacing assembly must be performed only by qualified service personnel who
are aware of the hazard involved (such as fire and electrical shock).
NoteIf an error message is displayed, the error may have been caused by incorrect operation of theunit. Check the User’s Manual, and perform the correct operation.
SM 701820-01E5-2
5.2 Flow Chart
5.2 Flow Chart
YES
YES
NO
NO
(2)
(1)Display?
Power ON
Message displayed?
START
YES
NO(4)
Execute PerformanceTest
Performance test
END
YES
NO(3)
Excute Self-test
Self-test normal
Refer to section 5.3.
SM 701820-01E 5-3
5
TR
OU
BL
ES
HO
OT
ING
5.2 Flow Chart
(1)
NO
YES
*1
*5
*1
NO
YES
Check fuse
Replace fuse
Check secondary sidevoltage
Check power supplyvoltage
FUSE OK?
Voltage OK?
END
START
NO
YES
Replace powersupply unit
Voltage OK?
(1)’
NO
YES
Execute INITIALIZE *2
LCD displaynormal?
NO
YES
Check video signal
*4
*3
Replace LCD Assembly
Video signal OK?
NO
YES
LCD DisplayOK?
Replace CPUAssembly
Maintenance Service is required.
Replace LCD Unit
*1 Refer to Check Procedure of Power Supply Voltage and System Clock.
*2 Turn power on and press the [SHIFT] key and then [INIT] key.*3 Connect the RGB VIDEO OUT signal to a VGA supported monitor.*4 Check the display of the monitor.*5 A fuse is built in, therefore, cannot be accessed.
SM 701820-01E5-4
5.2 Flow Chart
(1)’A short-circuit may occur in an assembly other than the power supply unit. To check inwhich voltage line a short circuit has occurred, investigate each assembly to whichvoltage is supplied, using a circuit test. The table below shows the relationship between
assemblies and voltage supplied to them.
Voltage Assembly Part No.
+5V CPU Board Assembly B9947LA/LBSUB CPU Board Assembly B9946LB/TTKey Board Assembly B9946LC/TYSUB Key Board Assembly B9946LJ/TZSCSI-DCDC Board Assembly (Option) B9946LFSCSI Connect Board Assembly (Option) B9947LDLCD Board Assembly A1054VAHDD Assembly (Option) A1110UNFDD Assembly A1092UNThermal Printer Assembly A1207UDSIMM Module (Option) A1418LQ, A1419LQ, A1420LQ701850 —701851 —701852 —701853 —701860 —701870 —701880 —
–5V CPU Board Assembly B9947LA/LBSUB CPU Board Assembly B9946LB/TT701850 —701851 —
+12V CPU Board Assembly B9947LA/LBSUB CPU Board Assembly B9946LB/TTFAN Assembly B9946WHSUB FAN Assembly B9946WJInverter Assembly A1459UP701850 —701851 —701852 —701853 —701860 —701870 —701880 —
–12V CPU Board Assembly B9947LA/LBSUB CPU Board Assembly B9946LB/TT701870 —
+24V CPU Board Assembly B9947LA/LBSUB CPU Board Assembly B9946LB/TTSCSI-DCDC Board Assembly (Option) B9946LF
(2)When a problem has occurred, it is necessary to refer to the Instruction Manual todetermine whether the problem is caused by incorrect operation or by a defect in the
hardware. The table below shows where a problem occurred and which part ofhardware had a defect.
Message Assembly
604 and 607 Media failure.606 Media is protected.611 and 612 Media is full.
FDD Assembly
621 File is damaged.Floppy Disk
646-654 Media failure.
901 Data backup failure. Initialization is performed. BatteryCheck the state of the battery.
906 Fan stopped. Turn off immediately. Fan
SM 701820-01E 5-5
5
TR
OU
BL
ES
HO
OT
ING
(3)When a problem has occurred, check the test item displaying FAIL and select therelevant defective item from the table below. If necessary, replace the relevantassembly.
Test item Assembly
Memory (System) CPU Board
Memory (ACQ RAM) CPU BoardSIMM
Key Board CPU BoardKeyboardSub-keyboard
FDD FDD AssemblyCPU Board
Printer Thermal PrinterCPU Board
(4)When a problem has occurred, check the non-confirming test item and select therelevant defective assembly from table below. If necessary, replace the relevant
assembly.
Test Item Assembly
2.6 Waveform Display Test CPU BoardSub-CPU Board
2.10 Built-in FDD Test CPU Board
2.11 Built-in HDD Test CPU BoardSCSI DCDC BoardHDD
5.3 Self TestSelf-diagnosis of Built-in Printer
Refer to Section 2.7.
Self-diagnosis of Internal MemoryRefer to Section 2.8.
Self-diagnosis of Keyboard, Jog Shuttle, and Rotary KnobRefer to Section 2.9.
5.2 Flow Chart / 5.3 Self Test
SM
701820-01E6-1
6SCHEMATIC DIAGRAM
Ch
apter 6
SC
HE
MA
TIC
DIA
GR
AM
system
system
option: included B9947BD
option
option
to
SM 701820-01E 7-1
7
CU
ST
OM
ER
MA
INT
EN
AN
CE
PA
RT
S L
IST
Chapter 7 CUSTOMER MAINTENANCEPARTS LIST
7.1 Customer Maintenance Parts List7.2 Standard Accessories
7.1 Customer Maintenance Parts List
7.1 Customer Maintenance Parts List
SM 701820-01E7-2
1715
Complete Set
14
1213
11
9
87
654
3
21
18
10
16
Cap (no SCSI)1A9065ZB18
1Description
21
Item
1SheetB9947CA
Part No. Qty
345
678910
A9079ZGA9076ZGB9946BRB9946BX
B9946BKY9408LBY9306LBB9946GZB9946GX
121
142
168
KnobKnobKnobSheet
Front BezelB.H.Screw,M4x8B.H.Screw,M3x6ScrewBlind (Accessory)
1112131415
16
Y9408LBB9946BHB9946BTB9946BPB9946CC
41211
B.H.Screw,M4x8Top CoverFootPRN CoverHandle
17B9946BMY9306LB
12
Side CoverB.H.Screw,M3x6
7.1 Customer Maintenance Parts List
SM 701820-01E 7-3
7
CU
ST
OM
ER
MA
INT
EN
AN
CE
PA
RT
S L
IST
7.1 Customer Maintenance Parts List
1
2
3
6
7
8
9 1011 12
20
13
1514
16
2221
17
19
4
5
18
*2 701820 -/C7
Note:
*1 701820 -/C8
22 A1212WF 1 Multiconductor*1,*2SCSI Connect Assembly*1,*21B9947LC21
(select)
(select)
18
17
1615
14
13
B9946TT
A1418LQB9947LAB9947LBY9318LB
B9946WHB9946VBB9946VAB9946CZ
1
1114
1113
Sub CPU Assembly
HSC3208E-6(/M1)CPU Assembly*1,*2CPU Assembly(no SCSI)B.H.Screw,M3x18
Fan AssemblyPower Unit(200V)Power Unit(100V)Screw
1
Description
Printer1211
Item
A1207UD
Part No. Qty
109876
543
B9946DTA1120UNB9946LFB9946WEY9314LB
B9946WJB9946CE -Y9306LB
41114
1118
Screw*1Memory System*1SCSI-DCDC Assembly*1Cable Assembly*1B.H.Screw,M3x14
Sub Fan AssemblyCover AssemblySub CPU AssemblyB.H.Screw,M3x6
1DescriptionFront Assembly (see page 4)
21
ItemB9946DAPart No. Qty
HSC3216E-6(/M2)1A1419LQA1420LQ 1 HSC3232E-6(/M3)
19 B9946WF 1 Cable Assembly*1
20 B9946WG 1 Cable Assembly*1
(select)
SM 701820-01E7-4
7.1 Customer Maintenance Parts List
Front Assembly
15
20
21
16
14
13
12
11
8
9
45
6
7
173
2
118
1019
22
23
MulticonductorB.H.Screw,M2x4Memory SystemMulticonductorSpacer KGLS-10RT
Cable AssemblyCXA-L0612-VJLSpacerClampB.H.Screw,M3x6
BracketKeysub AssemblyB.H.Screw,M3x6
13111
11221
113
A1214WFY9204LBA1092UNA1212WFB9292CE
B9917TBA1459UPB9946DGA9094KYY9306LB
B9946DEB9946TZY9306LB
1514131211
109876
543
Key Assembly1B9946TY1
DescriptionLCD
21
ItemA1054VAPart No. Qty
1617
B9946DBB9930DQ
31
BumperFFC
B9946WA 1 Cable AssemblyB9946ZY 1 Ferrite CoreB9914ZF 1 Ferrite Core
181920
LCD Connect Board Assembly1B9946LS21Stud2B9947CB22B.H.Screw,M3x64Y9306LB23
SM 701820-01E 7-5
7
CU
ST
OM
ER
MA
INT
EN
AN
CE
PA
RT
S L
IST
7.2 Standard Accessories
1
2
3
4
5
6
7
8
9
11
10
Stopper2 A9088ZM10
701820 - -R*5
(select)
98
B9946EA -
11
Front CoverInstruction ManualSoft CaseRoll Chart
11
11
B9946EBB9850NX
A1024WDA1054WD
76
543 A1009WD 1
3P-2P Adapter *11A1253JZ1
Description
21
ItemA1006WDPart No. Qty
Power Supply Code (UL.CSA standard) *1.*2
Power Supply Code (VDE standard) *3Power Supply Code (BS standard) *4Power Supply Code (SAA standard) *5
701820 - -Q701820 - -F701820 - -D
*4*3*2*1
Note:701820 - -M
11 B9946GX 8 Blind (see page 7-2)
7.2 Standard Accessories