Data Analysis and Presentation

• Chapter 5- Calibration Methods and Quality Assurance

EXCEL – How To Do

1- least squares and linear calibration

curve/function (chapter 4)

2- standard addition (chapter 5)

3-internal standard (addition) (chapter 5)

Page 2: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

External standard

1- least squares and linear calibration

curve/function (chapter 4)

Page 3: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

CHAPTER 05: Opener B

Page 4: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

CHAPTER 05: Table 5.1

Page 5: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

Calibration function

Q: How the selected instrument responds to the change in quantity (concentration, M ppm ) of the measured analyte?

A: Measure the instrument responses (peak hights, peak areas etc.) from known but different concentrations of the analyte (say 0, 5,10, 15 20 ppm standards). Then construct the response diagram figure (response vs. concentration) with the response or calibration function through the points.

Page 6: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

Calibration Data:

Answer 2

GO TO EXCELL or your calculator

Page 7: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

MFYI: Method of Least Squares/Linear regression 

The method of least squares assumes that the errors in the y values are substantially greater than the errors in the x values. A second assumption is that the uncertainties in all of the y values are similar.

 

Derivation of the Least Squares Method (Equation for a straight line) y = mx + b Vvertical deviation = dii = yi–y = yi–(mxi+b) SSome of the deviations are positive and some are negative. To minimize the magnitude of the deviations irrespective of their signs, we square the deviations to create positive numbers.

Page 8: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

If one assumes that the calibration curve is linear – there are physical reasons for that. There is a linear function that can best represent the response of the instrument.

If points are out of line: error

There are methods based on the propagation of error that can help us calculate the best fit: linear regression of the straight-line calibration curves

OLD SLIDE: Caffeine again: first step calibration

Page 9: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

CALIBRATION: Instrument used to measure signals must be calibrated, must have the calibration relation. Example (linear):

y = m [x ]+ b (general algebraic form for linear equation)

Often given in other similar forms:

Smeasured(total signal) =k nA (concentration) + Sreag ( signal from reagents, from blank)

Calibration is the determination of that relation: it is used for the determination of analyte by using standards and blanks to determine a relationship (function) between concentrations and assay responses.  Validation implies that other labs have approved the analytical method of analysis and produced similar results.  Standards are materials containing accurately known concentrations of the desired analyte. Standardization is the process of determining relationship between the measured signal and the amount of analyte (fittting for k in linear relations).

Blanks are solutions containing all added reagents except the sample analyte.(method, reagent and field blanks) Controls can be an alternate sample, in which the contents are well known.

Primary and Secondary reagents

Page 10: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

Calibration Data:

Answer 2

GO TO EXCELL or your calculator

Page 11: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

CHAPTER 04: Unnumbered Table 4.3

Page 12: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

CHAPTER 04: Unnumbered Table 4.4

Page 13: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

CHAPTER 04: Figure 4.12

Page 14: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

CHAPTER 04: Figure 4.14

Page 15: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

IMPORTANT : BLANK !!

Before you start measuring standards, you need to see the signal from a sample without analyze.

BLANK: everything but analyte

Page 16: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

CHAPTER 04: Table 4.7

Page 17: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

EXAMPLE: CORRECTION FOR nonzero BLANK

Your blank is 0.001 (for zero concentration of analyte)

Your signals are as follows

Signal Corrected signal for blank (!) concentration

0.001 0.001-0.001 0

0.003 0.003-0.001 0.0287

0.006 0.006-0.001 0.0601

0.009 0.009—0.001 0.0889

Page 18: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

2- standard addition (chapter 5)

Page 19: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

Is such external standardization the only way to calibrate and acquire the

accurate and precise values?

• No, there are also other methods that include addition of analyte or some other material into the sample (aliquot)

Page 20: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

Possible problems with external standardization

Page 21: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

Spike

• Sometimes (often) the response to analyte is affected by something else in the sample, which we call MATRIX.

• SPIKE is a known quantity of analyte added directly to the sample to verify if the response to analyte is the same as that expected from pure sample observed in the calibration curve.

Page 22: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

2. Standard additionSStandard addition: Known quantities of analyte are added to the unknown, and the

increased signal lets us deduce how much analyte was in the original unknown. Typically we use the method of standard addition when unknown sample matrix is sufficiently complex. The blank and standards are not representative of the unknown sample and lead to analysis error. This method requires a linear response to analyte.

 For a single spike, one trial 

[X]i / [X]f + [S]f = Ix / I s+x  [X]i = unknown initial concentration of analyte

Ix = signal from first solution[S]f = concentration of standard in second solution[X]f = diluted concentration of analyteI s+x = signal from second solution

 

Page 23: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

CHAPTER 05: Equation 5.7

Page 24: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

EExample 2 A blood serum sample containing sodium ion gives a signal of 4.27 mV on a light intensity meter in an atomic emission experiment. The Na+ concentration in the serum is then increased by 0.104 M by a standard addition, without significantly diluting the sample. This "spiked" serum sample gives a signal of 7.98 mV in atomic emission. Find the original concentration of Na+ in the serum. [X]i = unknown initial concentration of Na+

Ix = 4.27 mV[S]f = 0.104 M[X]f = diluted concentration of analyte = [X]i I s+x = 7.98 mV [X]i = 0.120 M

Page 25: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

FYI: (2) Sampledirect addition

a b

S sample = S spike

a b

Page 26: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

(1)Sampleseparate aliquots (portions)

a b

S sample = S spike

a b

FYI: Accurate description

Ssample = signal sampleCS= concentration spikedCA= concentration analyte

Page 27: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

WWe use standard addition methods when our sample has matrix effects which are hard to duplicate in a blank. Let's again find the amount of Pb+2 in a blood sample. A 5.00 mL blood sample containing lead yielded a signal of 0.712 units. Afterwards, the sample was spiked with 5.00 L of 1560 ppb Pb+2 standard. (note: a small addition) This spiked solution gave a reading of 1.546 units. Find the concentration of Pb+2.

 [X]i / [X]f + [S]f = Ix / I s+x or [C]i / [C]f + [Cs]f = Sx / S s+x

Be careful with volume!

Example 3.

CA=1.33ppb

Page 28: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

3. Internal Standard

• In addition to the analyte you measure add another similar compound (not analyte ) that has similar response ( sensitivity ) as the analyte.

Page 29: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

Internal Standard is sometimes added to an unknown sample. The reason may be to verify the signal response in situations where instrument response varies slightly from run to run. For example, an analysis is preformed on different days or different instruments or under different operating conditions. Typically, the internal standard resembles the analyte. Let's say we are separating isomers of octane and determining their concentration on a gas chromatograph (GC). To verify how much is present we might add a known amount of cyclohexane as an internal standard

Page 30: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

Internal Standard

• An internal standard is a known amount of a compound, different from the analyte, that is added to the unknown sample. A Standard mixture of analyte and standard is prepared before hand.

• Internal standards are desirable whenever losses of sample are likely to occur during handling or analysis. They are also used to calibrate the instrument when same analysis is done on different days. Use the following relation:

Page 31: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

Internal Standard

• SA(signal due to analyte) =k A CASIS(signal due to int. stand) =kIS CIS

Page 32: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

CHAPTER 05: Equation 5.11

Page 33: Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function

Example Example on internal standard: Note we are not diluting the sample significantly! A lab uses Cu+2 as a internal standard for a Pb+2 analysis in blood. Known solution standards for Pb+2 (1.75 ppm) and Cu+2 (2.25 ppm) give readings such that the signal ratio of Pb to Cu is 2.37. A blood sample was spiked with the Cu standard (2.25 ppm). We assume that the blood sample volume does not change significantly. The ratio of the readings for Pb to Cu in the unknown spiked solution is 1.80. Based on the response for the known amount of Cu+2 added to a blood sample determine, the amount of Pb+2 present.

SA(signal due to analyte) = k A CA

SIS(signal due to int. stand) = kIS CIS

CA

CIS= K

K=3.05

CA=1.33ppb Pb 2+

CA=CIS SA

SISK

First measurement

Second measurement