Upload
muhammadumair
View
215
Download
0
Embed Size (px)
Citation preview
8/10/2019 Condumax II Transportable 97149 US Datasheet-V1
1/4
www.michell.com
Process Analyzers
Condumax II TransportableHydrocarbon & Water Dew-Point Analyzer
Transportable natural gas dew-point analyzer system for operation in the field. Suitable for IECZone 1 or 2 hazardous area, NEC Class I, Division 1 Groups B, C, D.
Applications
Spot-check measurements of HC and water dewpoint
Field verification of installed on-line dew-pointanalyzers
Flexibility for periodic measurements at any gas
source
Producer process stage profiling to optimizeefficiency
A complete transportable natural gas dew-point analysis system for field spot-check measurements. Periodic online operationwhen located in a temperature maintained analyzer house. Measurements and functionality are instantly accessible throughintegrated alphanumeric display and touch-glass HMI interface including log of dew-point readings for operator review.
Highlights
Field transportable self-contained dew-pointanalyzer system
Simultaneous HC and water dew-pointmeasurements in a single analyzer
Automatic objective measurements by industry
proven principles
Parameter specific sensors ensure continuity ofHC and water dew-point readings
Self-contained sample handling, conditioning andanalysis package
0.5C (0.9F) HC dew-point accuracy
Fundamental cooled mirror principle
8/10/2019 Condumax II Transportable 97149 US Datasheet-V1
2/4
www.michell.com
Process Analyzers
The Industry Standard HCDP/WDPAnalyzer
The Condumax II fulfills the demands of natural gasproducers and pipeline transmission companies for reliableon-line monitoring of natural gas hydrocarbon and water dewpoint.
Condumax II is the industrys analyzer of choice to
ensure conformance of these key gas quality parameters,from production processing, through trans-continentalpipeline networks into final distribution and utilization bymajor consumers such as power generators. Throughoutthe midstream industry Michell hydrocarbon dew-pointanalyzers are installed as permanent analyzers, providingmeasurements six times every hour 24/7/365.
Transportable Condumax II for FieldValidation
The Condumax II Transportable addresses the furtherneeds of natural gas users as the industry evolves. Withan increased proportion of alternative fuel sources suchas biogas, and greater reliance on LNG imports into gasnetworks which traditionally received gas solely from fixedupstream producers, the attention to ensuring accurate
measurement of dew point has never been more critical.This increased complexity of supply chains has expandedthe number of potential measurement points each requiringconformity to specification. The Condumax II Transportableoffers the flexibility to make investigative spot-checkmeasurements at any location, to even the smallest and mostremote gas source. Furthermore, for the gas producer, theCondumax II Transportable can be used to profile individualprocess stages to ensure optimum efficiency.
Quality Assurance with a New TransferStandard
The Condumax II Transportable also enables establishedusers of permanent on-line dew-point analyzers to carry outperiodic comparative measurements in the field to verifymeasurement performance against the certified factorylaboratory calibration of the Condumax II to all users.
The Condumax II Transportable offers automaticmeasurements of high sensitivity and objectivity whereverthey are needed in a simple to use, self-contained, samplehandling, conditioning and analysis package.
Sample Conditioning System
Complete integration of Condumax II Main Unit
System configuration for connection direct to transmissionpipeline natural gas at typically 70 barg (1015 psig)pressure
Simultaneous water dew-point measurement at processpressure. Integration of water dew-point sample flow in
series for analysis at full line pressure
Process pressure gauge, 100 barg (1450 psig) scale
Coalescing filter with by-pass needle valve and systemdrain
Sample gas pressure reduction via electrically-heatedregulator for analysis of HC dew point at intermediatepressure (cricondentherm condition), adjustment range upto 35 barg (507 psig) fitted (analysis pressure indicatedby Condumax II Main Unit)
Sample flow control with armored rotameter
*To assure correct measurement performance ofthis system in outdoor field use, the ambient air
temperature must be at least 5C (9F) above boththe hydrocarbon dew-point temperature of the
process gas sample at the cricondentherm condition,typically 27 barg (391 psig), and water dew-point at
full process line pressure.
8/10/2019 Condumax II Transportable 97149 US Datasheet-V1
3/4
www.michell.com
Process Analyzers
System Connections
Minimessstainless steel braided, micro-bore hoses
Sampling system connection: Minimessquick connect
Set of 3 hoses: Sample inlet, coalescing filter drain /by-pass and sample gas exhaust
Length: 3m (9.8ft) (each hose)
Process connection: 316 stainless steel, 1/4 NPT male
Gas wetted bore: PTFE, 2mm internal diameter
System Package
Custom designed and built carry case of heavy dutyconstruction in 316 stainless steel with six vibrationabsorbing feet on base (for transportation) and on oneedge (for upright analyzer system operation)
Removable case lid secured by four toggle latches. Sixcarry handles for ease of mobility by two persons andtransportation between sites by automobile
Power consumption:
8/10/2019 Condumax II Transportable 97149 US Datasheet-V1
4/4
Process Analyzers
Michell Instruments adopts a continuous development programme which sometimes necessitates specification changes without notice.Issue no: Condumax II Transportable_97149_V1_US_0913
2013 Michell Instruments
Michell Instruments, Inc319 Newburyport Turnpike, Suite 207, Rowley, MA 01969Tel: 978 484 0005, Fax: 978 843 7669, Email: [email protected], Web:www.michell.com/us
Technical Specifications
Condumax II Main UnitAnalyzer
Hydrocarbon Dew-Point Measurement
Measuring technique Dark Spot fixed sample analysis; Directphoto-detection of hydrocarbon condensate athydrocarbon dew-point temperature
Sensor cooling Automatic via 3-stage Peltier effect electroniccooler under adaptive control
Maximum range Up to 55C (131F) measurement depressionfrommain unit operating temperature
Resolution 0.1C, 0.1F
Accuracy 0.5C (0.9F) hydrocarbon dew point(single and multiple condensible componentanalysis)
Sample flow 0.03 to 0.06 m3/hr(0.5 to 1.0 Nl/min / 1.05 to 2.11 scfh)- low flow alarm standard
Measurement frequency 6 cycles/hour (recommended)12 cycles/hour (maximum)
Water Dew-Point Measurement
Measuring technique Michell Ceramic Moisture Sensor
Unit selectionDew point
Moisture contentC and F water dew pointlbs/MMscf; mg/m3, ppmV(conversion method IGT#8 or ISO18453)
Resolution 0.1C, 0.1F dew point0.1 lbs/MMscf, 1 mg/m3 ppmV
Range Calibrated from -100 to +20C (-148 to +68F)dew point
Accuracy 1C (1.8F) from -59 to +20C(-74 to +68F) dew point2C (3.6F) from -100 to -60C(-148 to -76F)dew point
Pressure Measurement(s)Hydrocarbon dew-point analysis pressure
Units MPa, barg, psig
Resolution 0.1 MPa and barg, 1 psig
AccuracyHydrocarbon dew point
Water dew point0.25 barg0.5 barg
Hydrocarbon and Water Dew-Point Analyzer
Sample gas supply Natural gas up to 100 barg (1450 psig),pressure regulated in sampling system
Enclosure Travel case is stainless steel 316 materialconstruction with removable lid for access andrubber feet for transportation / operation;Instrument is EExd cast enclosure with
removable viewing port: internally heated forcondensation protection
Gas connections Sample inlet, filter by-pass outlet, systemdrain.Sample system for: 1/4 NPT female ports,Quick connect Minimesshoses,Process connections: 1/4 NTP male
Operating environment Indoors/outdoors 0 to +40C (+32 to +104F);Max 95% RH*See cautionary note on page 2
Power supply 110 or 230 V AC, 50/60 Hz, 300 W
Weight 60kg (132lbs) (approx)
Logged data
Each log
Visual review
150 data logs, sufficient for minimum 24 hourperiod (stored in volatile memory);HC dew point and analysis pressure, water dewpoint and process line pressure;HMI display with automatic statistical analysis;maximum, minimum, average values forduration of log period
Integrated display/keyboard
Touch screen with vacuum fluorescent display
Outputs Modbus RTU, RS485 @ 9600 baud rate;Two 4-20 mA linear (non-isolated) outputs,user configurable for any combination of dewpoint or pressure parameters;Additional connection junction box requiredexternally
Alarms Process and analyzer status via softwareregister and display annotation. Integratedlow flow alarms for each sample flow Analyzerstatus fault flag 23 mA on mA output 1;Comprehensive analyzer status messagesappear on HMI display;Additional connection junction box requiredexternally
Certification
Hazardous areacertification
ATEX II 2 G EExd IIB + H2 T4 Tamb =-40 to +60C
CCSAUScertified to Class I, Div 1, Groups B, C& D, T4
Transportable CD2 with Water DP (ISO18453) andmaterial certificates ATEX (230 V AC)
CD2PSS-AT-HW-MC
Transportable CD2 with Water DP (IGT#8) andmaterial certificates CSA (110 V AC)
CD2PSS-CS-HW-MC
Accessories
Item Product / Description
SSF-CF-LON-10PK Coalescing Filter Cartridge (pack of 10)
Condumax II Transportable
Order Codes