B ACCREMTED SOUTHWEST RESEARCH INSTITUTE@ 6220 Culebra Road, P 0 Drawer 28510 Institule Quality Systems Institute C allbration Laboratory Phone 210-522-5215 Fax 210-522-4834 Calibration Laboratory Certificate of Calibration Certificate #0972-01 - Cost Center / Customer: DIV20 / DO-N BANNON M:anufacturer/Model: DURO-SENSE / TYPE I( Work Order: 303 103265 Mail Stop: B57 Date Issued: 22-Aug-2011 Date Calibrated: 22-Aug-2011 * Date Due : 22-Aug-2012 Description: 'THERMOCOUPLE Serial Number: 012154 Asset Number: 012154 ** Results: FOUND-LEFT Humidity: 58 %RH Temperature: 73.0 OF Barometer: NIA Procedure: 'TEMPERATURE PROBES - 5 JUN 06 This certificate documents traceability lo the National Institute of Standimis and Technology (NIST) and the International Systim of Units (SI). The Laboratory quali1.y system conforms to ISO/IEC 17025. 2005, ANSI/NCSL Z540-1.1994 and relevant requirements of the IS0 9000-2000 standard. This certificate shall not be reproduced, except in full, without the written approval ofthe Southwest Research Institute Calibration Laboratory. This certificate shall not be used to claim product endorsement by Southwest Research lnst itutr, American Association for Laboratory Accreditation (A2LA) or any agency of the U. S. Government. Results ofthis calibration relate only to the instrument described above at the time of calibration and does not imply any long term stability ofthe instrument. *Determined by the customer, does not imply the: instrument will remain within tolerance as any number offactors may cause :an out-of-tolerance condition before this date. **Data type found in this certificate 01 attached measurement report must be interpreted as: Found-left - adjustment and/or repair was not performed, As-fc'und - data is before unit is adjusted andor repaired, As-lefi - data IS afier adjusted and/or repaired was performed. 'The ciustomer has sole responsibility for determination of in-/out-of-tolerance or conipliarice/nonccmpliance. Measurement uncertainty calculated in accordance with the method described in the IS0 "Guide to the Expression of Uncertainty in Measurement" (GUM). for a confidence level of approximately 95 pcrceiit using a coverage factor of k==2. Remarks: CAL @ 0°C and 150°C Standards Used -_ Asset # Manufacturer 0091 37 HART SCIENTIFIC 013908 HART SCIENTIFIC 015240 HART SCIENTIFIC Model IS75 5028 2566 Descriotion Cal Date Due Date SUPER THERMOMETER 11-Jul-2011 I I-Jan-2012 SPRT 3-Feb-2011 3-Feb-2012 TC SCANNER, 12-CHANNEL MODULE 29-Nov-2010 29-No\-201 I Walt Hill Laboratory Manager m \KIA OCT-08 rpt Page 1 of 1 Metrology Technician

Certificate of Calibration Calibration Laboratory #0972-01 · 2012. 12. 5. · Southwest Research Institute Calibration Laboratory Measurement Report 3031 03265 Asset No. 012154 Serial

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Certificate of Calibration Calibration Laboratory #0972-01 · 2012. 12. 5. · Southwest Research Institute Calibration Laboratory Measurement Report 3031 03265 Asset No. 012154 Serial

B ACCREMTED

S O U T H W E S T R E S E A R C H I N S T I T U T E @ 6 2 2 0 C u l e b r a R o a d , P 0 D r a w e r 2 8 5 1 0

Inst i tule Q u a l i t y S y s t e m s Inst i tute C a l lbra t ion Labora to ry

P h o n e 2 1 0 - 5 2 2 - 5 2 1 5 Fax 2 1 0 - 5 2 2 - 4 8 3 4 Calibration Laboratory Certificate of Calibration Certificate #0972-01 -

Cost Center / Customer: DIV20 / DO-N BANNON

M:anufacturer/Model: DURO-SENSE / TYPE I(

Work Order: 303 103265 Mail Stop: B57 Date Issued: 22-Aug-2011

Date Calibrated: 22-Aug-2011 * Date Due : 22-Aug-2012 Description: 'THERMOCOUPLE

Serial Number: 012154 Asset Number: 012154

** Results: FOUND-LEFT

Humidity: 58 %RH Temperature: 73.0 OF

Barometer: NIA Procedure: 'TEMPERATURE PROBES - 5 JUN 06

This certificate documents traceability lo the National Institute of Standimis and Technology (NIST) and the International Systim of Units (SI). The Laboratory quali1.y system conforms to ISO/IEC 17025. 2005, ANSI/NCSL Z540-1.1994 and relevant requirements of the IS0 9000-2000 standard. This certificate shall not be reproduced, except in full, without the written approval of the Southwest Research Institute Calibration Laboratory. This certificate shall not be used to claim product endorsement by Southwest Research lnst itutr, American Association for Laboratory Accreditation (A2LA) or any agency of the U. S. Government. Results ofthis calibration relate only to the instrument described above at the time of calibration and does not imply any long term stability of the instrument.

*Determined by the customer, does not imply the: instrument will remain within tolerance as any number offactors may cause :an out-of-tolerance condition before this date. **Data type found in this certificate 01 attached measurement report must be interpreted as: Found-left - adjustment and/or repair was not performed, As-fc'und - data is before unit is adjusted andor repaired, As-lefi - data IS afier adjusted and/or repaired was performed. 'The ciustomer has sole responsibility for determination of in-/out-of-tolerance or conipliarice/nonccmpliance.

Measurement uncertainty calculated in accordance with the method described in the IS0 "Guide to the Expression of Uncertainty in Measurement" (GUM). for a confidence level of approximately 95 pcrceiit using a coverage factor of k==2.

Remarks: CAL @ 0°C and 150°C

Standards Used -_ Asset # Manufacturer

0091 37 HART SCIENTIFIC

013908 HART SCIENTIFIC

015240 HART SCIENTIFIC

Model IS75

5028

2566

Descriotion Cal Date Due Date

SUPER THERMOMETER 11-Jul-2011 I I-Jan-2012

SPRT 3-Feb-2011 3-Feb-2012

TC SCANNER, 12-CHANNEL MODULE 29-Nov-2010 29-No\-201 I

Walt Hill Laboratory Manager

m \ K I A OCT-08 rpt Page 1 of 1 Metrology Technician

Page 2: Certificate of Calibration Calibration Laboratory #0972-01 · 2012. 12. 5. · Southwest Research Institute Calibration Laboratory Measurement Report 3031 03265 Asset No. 012154 Serial

Southwest Research Institute Calibration Laboratory Measurement Report

3031 03265 Asset No. 012154 Serial No. 012154

Type Data: Found-left Mfr Model Type.

FunctiodRange Test Point TI Reading Difference +/- Limit +/- IJncertainty Result % Limit Tem pera tu re OC OC OC O C "C

0.1 0.6 0.5 2.2 0.5 Pass 22% 150.0 151.8 1.8 2.2 0.5 Pass 83%

ElVD OF REPORT

IJ:\CC30\ICL\Data Sheets\Temperature\Thermocouple -12 Aug 11 .xIsx Page 1 of 1