European Standards Laboratory CALIBRATION-CERTIFICATE Yokogawa Europe Amersfoort, The Netherlands Certificate number Page 1 Applicant Applicant Instrument Instrument Manufacturer Yokogawa Type 760304-40-MV-Q/G6/EC/C7/C5/FL Serial number Serial number Inventory number Inventory number ID numbers used standards CSE820 CSE976 Procedure CALWT3000_2A_2A_2A_2A Version: 1 of 09-05-2012 Calibration method The Powermeter was compared to the Yokogawa Power Calibration System via a phantom technique. During the calibration the distortion of the applied voltage and current were below 0.1 %. Before calibration the device was powered on for at least 24 hours. Environmental conditions Temperature 22.4ºC ±1ºC Relative Humidity 52% ±4% Date of Calibration 10-May-12 Result The results of the calibration are shown on the next pages Traceability The results of the calibration services of Yokogawa European Standards Laboratory are traceable to primary and/or (inter)nationally accepted measurement standards. Date 10-May-12 Name E.J. Kroon Function Supervisor Standards Laboratory Yokogawa Europe BV This certificate is issued provided that Yokogawa European Standards Laboratory does not assume any liability. P.O. Box 163, 3800 AD Amersfoort (NL) Phone +31 (0)88 46410000 Reproduction of the complete certificate is permitted. website http/tmi.yokogawa.com Part of the certificate may only reproduced after written permission of Yokogawa.

European Standards Laboratory CALIBRATION-CERTIFICATE ...cdn6.us.yokogawa.com/content/Example_WT3000_2A_1.01.pdfEuropean Standards Laboratory CALIBRATION-CERTIFICATE Yokogawa Europe

  • Upload
    vocong

  • View
    234

  • Download
    0

Embed Size (px)

Citation preview

European Standards Laboratory CALIBRATION-CERTIFICATEYokogawa EuropeAmersfoort, The Netherlands

Certificate numberPage 1

Applicant Applicant

Instrument InstrumentManufacturer YokogawaType 760304-40-MV-Q/G6/EC/C7/C5/FLSerial number Serial numberInventory number Inventory numberID numbers used standards CSE820 CSE976Procedure CALWT3000_2A_2A_2A_2A Version: 1 of 09-05-2012

Calibration method The Powermeter was compared to the Yokogawa Power Calibration Systemvia a phantom technique. During the calibration the distortion of the applied voltage and current were below 0.1 %. Before calibration the device was powered on for at least 24 hours.

Environmental conditions Temperature 22.4ºC ±1ºCRelative Humidity 52% ±4%

Date of Calibration 10-May-12

Result The results of the calibration are shown on the next pages

Traceability The results of the calibration services of Yokogawa European Standards Laboratoryare traceable to primary and/or (inter)nationally accepted measurement standards.

Date 10-May-12

Name E.J. KroonFunction Supervisor Standards Laboratory

Yokogawa Europe BV This certificate is issued provided that YokogawaEuropean Standards Laboratory does not assume any liability.P.O. Box 163, 3800 AD Amersfoort (NL)Phone +31 (0)88 46410000 Reproduction of the complete certificate is permitted.website http/tmi.yokogawa.com Part of the certificate may only reproduced after written

permission of Yokogawa.

Certificate number Page: 2

E L E M E N T 1

Voltage Calibration DCRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 15.0000 14.9850 15.0003 15.0150 V DC 0.002 Pass

15V -15.0000 -15.0150 -14.9990 -14.9850 V DC -0.006 Pass

100V 100.000 99.900 100.007 100.100 V DC 0.007 Pass

100V -100.000 -100.100 -99.996 -99.900 V DC -0.004 Pass

150V 150.000 149.850 149.995 150.150 V DC -0.003 Pass

150V -150.000 -150.150 -150.007 -149.850 V DC 0.005 Pass

1000V 1000.000 999.000 1000.119 1001.000 V DC 0.012 Pass

1000V -1000.000 -1001.000 -1000.002 -999.000 V DC 0.000 Pass

Voltage Calibration AC at 10 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 100.0000 99.7000 100.0020 100.3000 V AC 0.002 Pass

150V 150.000 149.550 150.005 150.450 V AC 0.003 Pass

Voltage Calibration RMS at 60 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 15.0000 14.9940 15.0000 15.0060 V AC 0.000 Pass

30V 30.0000 29.9880 30.0025 30.0120 V AC 0.008 Pass

60V 60.0000 59.9760 60.0009 60.0240 V AC 0.001 Pass

100V 100.000 99.960 100.003 100.040 V AC 0.003 Pass

150V 150.000 149.940 150.002 150.060 V AC 0.001 Pass

300V 300.000 299.880 300.037 300.120 V AC 0.012 Pass

600V 600.000 599.760 600.021 600.240 V AC 0.004 Pass

1000V 1000.00 999.60 1000.09 1000.40 V AC 0.009 Pass

See Next Page: 3

Certificate number Page: 3

Voltage Calibration RMS at 1 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 15.0000 14.9880 15.0006 15.0120 V AC 0.004 Pass

30V 30.0000 29.9760 30.0046 30.0240 V AC 0.015 Pass

60V 60.0000 59.9520 60.0039 60.0480 V AC 0.007 Pass

100V 100.000 99.920 100.009 100.080 V AC 0.009 Pass

150V 150.000 149.880 150.012 150.120 V AC 0.008 Pass

300V 300.000 299.760 300.056 300.240 V AC 0.019 Pass

600V 600.000 599.520 600.045 600.480 V AC 0.007 Pass

1000V 1000.00 999.20 1000.14 1000.80 V AC 0.014 Pass

Voltage Calibration RMS at 10 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 100.000 99.850 100.053 100.150 V AC 0.053 Pass

150V 150.000 149.775 150.093 150.225 V AC 0.062 Pass

Voltage Calibration RMS at 50 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 100.000 99.600 100.139 100.400 V AC 0.139 Pass

150V 100.000 99.600 100.223 100.400 V AC 0.223 Pass

Voltage Calibration RMS at 100 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 15.000 14.790 15.027 15.210 V AC 0.179 Pass

100V 100.000 98.600 99.523 101.400 V AC -0.477 Pass

150V 100.000 98.600 99.485 101.400 V AC -0.515 Pass

Voltage Calibration Linearity at 60 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 1.000 0.969 0.998 1.031 V AC -0.200 Pass

100V 10.000 9.960 9.999 10.040 V AC -0.011 Pass

100V 50.000 49.920 50.001 50.080 V AC 0.002 Pass

100V 110.000 109.860 110.003 110.140 V AC 0.003 Pass

Current Calibration DCRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100mA 0.10000 0.09990 0.10000 0.10010 A DC -0.003 Pass

100mA -0.10000 -0.10010 -0.10001 -0.09990 A DC 0.005 Pass

1A 1.0000 0.9990 1.0004 1.0010 A DC 0.037 Pass

1A -1.0000 -1.0010 -0.9998 -0.9990 A DC -0.018 Pass

See Next Page: 4

Certificate number Page: 4

Current Calibration AC at 10 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100mA 100.00 99.70 100.00 100.30 mA AC 0.004 Pass

1A 1.0000 0.9970 1.0001 1.0030 A AC 0.008 Pass

Current Calibration AC at 60 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

5mA 5.000 4.996 4.999 5.004 mA AC -0.029 Pass

10mA 10.00 9.99 10.00 10.01 mA AC 0.015 Pass

20mA 20.00 19.98 20.00 20.02 mA AC 0.010 Pass

50mA 50.00 49.96 50.00 50.04 mA AC -0.005 Pass

100mA 100.0 99.9 100.0 100.1 mA AC -0.010 Pass

200mA 200.0 199.8 200.0 200.2 mA AC -0.001 Pass

500mA 500.0 499.8 500.0 500.2 mA AC 0.004 Pass

1A 1.0000 0.9960 1.0000 1.0040 A AC 0.004 Pass

2A 2.0000 1.9992 2.0001 2.0008 A AC 0.005 Pass

Current Calibration AC at 1 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

10mA 10.000 9.992 10.004 10.008 mA AC 0.044 Pass

100mA 100.00 99.92 100.00 100.08 mA AC 0.004 Pass

200mA 200.00 199.84 200.02 200.16 mA AC 0.012 Pass

500mA 500.0 499.6 500.1 500.4 mA AC 0.018 Pass

1A 1.0000 0.9992 1.0001 1.0008 A AC 0.015 Pass

2A 2.0000 1.9984 2.0004 2.0016 A AC 0.018 Pass

Current Calibration AC at 10 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

200mA 200.00 199.70 199.99 200.30 mA AC -0.003 Pass

1A 1.00000 0.99850 1.00029 1.00150 A AC 0.029 Pass

Current Calibration AC at 50 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

200mA 200.00 199.20 200.05 200.80 mA AC 0.025 Pass

1A 1.00000 0.99600 1.00023 1.00400 A AC 0.023 Pass

Current Calibration AC at 100 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

50mA 50.000 49.300 49.981 50.700 mA AC -0.038 Pass

100mA 100.00 98.60 99.96 101.40 mA AC -0.035 Pass

200mA 200.00 197.20 199.95 202.80 mA AC -0.026 Pass

1A 1.00000 0.98600 0.99979 1.01400 A AC -0.021 Pass

See Next Page: 5

Certificate number Page: 5

Current Calibration Linearity AC at 60 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

1A 0.01000 0.00940 0.00999 0.01060 A AC -0.090 Pass

1A 0.10000 0.09940 0.09998 0.10060 A AC -0.015 Pass

1A 0.50000 0.49940 0.50001 0.50060 A AC 0.003 Pass

1A 1.10000 1.09940 1.09996 1.10060 A AC -0.004 Pass

External Current Calibration DC. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.9500 49.9591 50.0500 mA DC -0.082 Pass

200mV 200.000 199.800 199.989 200.200 mA DC -0.006 Pass

1V 1.00000 0.99900 1.00003 1.00100 A DC 0.003 Pass

2V 2.00000 1.99800 2.00011 2.00200 A DC 0.006 Pass

5V 5.00000 4.99500 5.00030 5.00500 A DC 0.006 Pass

External Current Calibration 10 Hz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.8500 49.9976 50.1500 mA AC -0.005 Pass

200mV 200.000 199.400 200.002 200.600 mA AC 0.001 Pass

1V 1.00000 0.99700 0.99999 1.00300 A AC -0.001 Pass

2V 2.00000 1.99400 1.99998 2.00600 A AC -0.001 Pass

5V 5.00000 4.98500 4.99998 5.01500 A AC 0.000 Pass

External Current Calibration 60 Hz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.8500 49.9945 50.1500 mA AC -0.011 Pass

100mV 100.000 99.960 99.999 100.040 mA AC -0.001 Pass

200mV 200.000 199.920 200.003 200.080 mA AC 0.002 Pass

1V 1.00000 0.99960 1.00000 1.00040 A AC 0.000 Pass

10V 10.00000 9.99600 10.00005 10.00400 A AC 0.001 Pass

External Current Calibration 1 kHz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.9700 50.0110 50.0300 mA AC 0.022 Pass

100mV 100.000 99.940 100.024 100.060 mA AC 0.024 Pass

200mV 200.000 19.988 200.048 200.120 mA AC 0.024 Pass

1V 1.00000 0.99940 1.00020 1.00060 A AC 0.020 Pass

10V 10.00000 9.99400 10.00214 10.00600 A AC 0.021 Pass

See Next Page: 6

Certificate number Page: 6

External Current Calibration 10 kHz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.9250 49.9937 50.0750 mA AC -0.013 Pass

200mV 200.000 199.700 200.100 200.300 mA AC 0.050 Pass

1V 1.00000 0.99850 1.00057 1.00150 A AC 0.057 Pass

2V 2.00000 1.99700 2.00119 2.00300 A AC 0.059 Pass

5V 5.00000 4.99250 5.00310 5.00750 A AC 0.062 Pass

External Current Calibration 50 kHz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.8000 50.0697 50.2000 mA AC 0.139 Pass

200mV 200.000 199.200 200.294 200.800 mA AC 0.147 Pass

1V 1.00000 0.99600 1.00136 1.00400 A AC 0.136 Pass

2V 2.00000 1.99200 2.00270 2.00800 A AC 0.135 Pass

5V 5.00000 4.98000 5.00748 5.02000 A AC 0.150 Pass

External Current Calibration 100 kHz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.3000 50.0981 50.7000 mA AC 0.196 Pass

200mV 200.000 197.200 200.354 202.800 mA AC 0.177 Pass

1V 1.00000 0.98600 1.00155 1.01400 A AC 0.155 Pass

2V 2.00000 1.97200 2.00302 2.02800 A AC 0.151 Pass

5V 5.00000 4.93000 5.00995 5.07000 A AC 0.199 Pass

Power Calibration DCRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 1A 15.0000 14.9770 15.0049 15.0230 W 0.033 Pass

15V 1A -15.0000 -15.0230 -15.0033 -14.9770 W 0.022 Pass

100V 1A 100.0000 99.8500 100.0499 100.1500 W 0.050 Pass

100V 1A -100.0000 -100.1500 -100.0375 -99.8500 W 0.038 Pass

150V 1A 150.000 149.770 150.043 150.230 W 0.029 Pass

150V 1A -150.000 -150.230 -150.075 -149.770 W 0.050 Pass

1000V 1A 1000.000 998.500 1000.557 1001.500 W 0.056 Pass

1000V 1A -1000.000 -1001.500 -1000.373 -998.500 W 0.037 Pass

See Next Page: 7

Certificate number Page: 7

Power Calibration 60Hz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 10mA 1.00000 0.99940 1.00013 1.00060 W 0.013 Pass

100V 20mA 2.00000 1.99880 2.00000 2.00120 W 0.000 Pass

100V 50mA 5.00000 4.99700 4.99958 5.00300 W -0.008 Pass

100V 100mA 10.0000 9.9940 9.9996 10.0060 W -0.004 Pass

100V 200mA 20.0000 19.9880 20.0007 20.0120 W 0.004 Pass

100V 500mA 50.0000 49.9700 50.0027 50.0300 W 0.005 Pass

100V 2A 200.000 199.880 200.014 200.120 W 0.007 Pass

15V 1A 15.0000 14.9910 15.0001 15.0090 W 0.001 Pass

30V 1A 30.0000 29.9820 30.0037 30.0180 W 0.012 Pass

60V 1A 60.0000 59.9640 60.0029 60.0360 W 0.005 Pass

100V 1A 100.000 99.940 100.005 100.060 W 0.005 Pass

150V 1A 150.000 149.910 150.009 150.090 W 0.006 Pass

300V 1A 300.000 299.820 300.047 300.180 W 0.016 Pass

600V 1A 600.000 599.640 599.999 600.360 W 0.000 Pass

1000V 1A 1000.00 999.40 1000.12 1000.60 W 0.012 Pass

Power Calibration 1 kHz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 10mA 1.00000 0.99900 1.00052 1.00100 W 0.052 Pass

100V 100mA 10.0000 9.9900 10.0023 10.0100 W 0.023 Pass

100V 200mA 20.0000 19.9800 20.0048 20.0200 W 0.024 Pass

100V 500mA 50.0000 49.9500 50.0112 50.0500 W 0.022 Pass

100V 2A 200.000 199.800 200.053 200.200 W 0.026 Pass

15V 1A 15.0000 14.9850 15.0029 15.0150 W 0.019 Pass

30V 1A 30.0000 29.9700 30.0078 30.0300 W 0.026 Pass

60V 1A 60.0000 59.9400 60.0115 60.0600 W 0.019 Pass

100V 1A 100.000 99.900 100.020 100.100 W 0.020 Pass

150V 1A 150.000 149.850 150.033 150.150 W 0.022 Pass

300V 1A 300.000 299.700 300.090 300.300 W 0.030 Pass

600V 1A 600.000 599.400 600.132 600.600 W 0.022 Pass

1000V 1A 1000.00 999.00 1000.25 1001.00 W 0.025 Pass

Power Calibration 10 kHz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 200mA 20.0000 17.9700 20.0136 22.0300 W 0.068 Pass

100V 1A 100.000 99.750 100.078 100.250 W 0.078 Pass

150V 1A 100.000 99.625 100.086 100.375 W 0.086 Pass

See Next Page: 8

Certificate number Page: 8

Power Calibration 50 kHz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 200mA 20.0000 15.9400 19.9706 24.0600 W -0.147 Pass

100V 1A 100.000 99.500 100.167 100.500 W 0.167 Pass

150V 1A 100.000 99.625 100.181 100.375 W 0.181 Pass

Power Calibration 100 kHz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 50mA 5.00000 3.43000 4.95791 6.57000 W -0.842 Pass

100V 100mA 10.0000 6.8600 9.9443 13.1400 W -0.557 Pass

100V 200mA 20.0000 13.7200 19.9123 26.2800 W -0.439 Pass

Power Calibration Linearity 60 Hz PF=1 at constant 1 ARange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 1A 1.0000 0.9600 0.9994 1.0400 W -0.059 Pass

100V 1A 10.0000 9.9580 9.9997 10.0420 W -0.003 Pass

100V 1A 50.0000 49.9500 50.0031 50.0500 W 0.006 Pass

100V 1A 110.0000 109.9380 110.0068 110.0620 W 0.006 Pass

Power Calibration Linearity 60 Hz PF=1 at constant 100 VRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 1A 1.0000 0.9600 0.9992 1.0400 W -0.082 Pass

100V 1A 10.0000 9.9580 9.9994 10.0420 W -0.006 Pass

100V 1A 50.0000 49.9500 50.0021 50.0500 W 0.004 Pass

100V 1A 110.0000 109.9380 109.9973 110.0620 W -0.002 Pass

External Current Power Calibration DC. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 50mV 5.00000 4.99250 4.99824 5.00750 W -0.035 Pass

100V 200mV 20.0000 19.9700 20.0028 20.0300 W 0.014 Pass

100V 1V 100.000 99.850 100.020 100.150 W 0.020 Pass

100V 2V 200.000 199.700 200.029 200.300 W 0.014 Pass

100V 5V 500.000 499.250 500.088 500.750 W 0.018 Pass

External Current Power Calibration 60 Hz. PF=1. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 50mV 5.00000 4.99700 5.00024 5.00300 W 0.005 Pass

100V 200mV 20.0000 19.9880 20.0014 20.0120 W 0.007 Pass

100V 1V 100.000 99.940 100.002 100.060 W 0.002 Pass

100V 2V 200.000 199.880 200.007 200.120 W 0.004 Pass

100V 5V 500.000 499.700 500.018 500.300 W 0.004 Pass

External Current Power Calibration 1 kHz. PF=1. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 50mV 5.00000 4.99500 5.00047 5.00500 W 0.009 Pass

100V 200mV 20.0000 19.9800 20.0052 20.0200 W 0.026 Pass

100V 1V 100.000 99.900 100.018 100.100 W 0.018 Pass

100V 2V 200.000 199.800 200.040 200.200 W 0.020 Pass

100V 5V 500.000 499.500 500.102 500.500 W 0.020 Pass

See Next Page: 9

Certificate number Page: 9

Power Calibration 60Hz PF=0 laggingRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 1A 0.0000 -0.0230 0.0001 0.0230 W - Pass

100V 1A 0.000 -0.015 0.002 0.150 W - Pass

150V 1A 0.000 -0.225 0.001 0.225 W - Pass

Power Calibration 60Hz PF=0 LeadingRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 1A 0.0000 -0.0230 -0.0003 0.0230 W - Pass

100V 1A 0.000 -0.015 -0.005 0.150 W - Pass

150V 1A 0.000 -0.225 -0.002 0.225 W - Pass

Frequency CalibrationRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 60.000 59.970 60.000 60.030 Hz 0.000 Pass

15V 1.00000 0.99950 1.00000 1.00050 kHz 0.000 Pass

15V 10.0000 9.9950 10.0000 10.0050 kHz 0.000 Pass

15V 50.000 49.975 50.000 50.025 kHz 0.000 Pass

See Next Page: 10

Certificate number Page: 10

E L E M E N T 2

Voltage Calibration DCRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 15.0000 14.9850 15.0016 15.0150 V DC 0.011 Pass

15V -15.0000 -15.0150 -14.9988 -14.9850 V DC -0.008 Pass

100V 100.000 99.900 100.010 100.100 V DC 0.010 Pass

100V -100.000 -100.100 -99.993 -99.900 V DC -0.007 Pass

150V 150.000 149.850 150.001 150.150 V DC 0.001 Pass

150V -150.000 -150.150 -150.000 -149.850 V DC 0.000 Pass

1000V 1000.000 999.000 1000.141 1001.000 V DC 0.014 Pass

1000V -1000.000 -1001.000 -999.983 -999.000 V DC -0.002 Pass

Voltage Calibration AC at 10 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 100.0000 99.7000 100.0020 100.3000 V AC 0.002 Pass

150V 150.000 149.550 150.003 150.450 V AC 0.002 Pass

Voltage Calibration RMS at 60 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 15.0000 14.9940 15.0001 15.0060 V AC 0.001 Pass

30V 30.0000 29.9880 30.0024 30.0120 V AC 0.008 Pass

60V 60.0000 59.9760 60.0007 60.0240 V AC 0.001 Pass

100V 100.000 99.960 100.001 100.040 V AC 0.001 Pass

150V 150.000 149.940 150.001 150.060 V AC 0.001 Pass

300V 300.000 299.880 300.035 300.120 V AC 0.012 Pass

600V 600.000 599.760 600.018 600.240 V AC 0.003 Pass

1000V 1000.00 999.60 1000.08 1000.40 V AC 0.008 Pass

See Next Page: 11

Certificate number Page: 11

Voltage Calibration RMS at 1 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 15.0000 14.9880 15.0004 15.0120 V AC 0.002 Pass

30V 30.0000 29.9760 30.0044 30.0240 V AC 0.015 Pass

60V 60.0000 59.9520 60.0037 60.0480 V AC 0.006 Pass

100V 100.000 99.920 100.007 100.080 V AC 0.007 Pass

150V 150.000 149.880 150.010 150.120 V AC 0.007 Pass

300V 300.000 299.760 300.050 300.240 V AC 0.017 Pass

600V 600.000 599.520 600.042 600.480 V AC 0.007 Pass

1000V 1000.00 999.20 1000.13 1000.80 V AC 0.013 Pass

Voltage Calibration RMS at 10 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 100.000 99.850 100.044 100.150 V AC 0.044 Pass

150V 150.000 149.775 150.088 150.225 V AC 0.059 Pass

Voltage Calibration RMS at 50 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 100.000 99.600 100.128 100.400 V AC 0.128 Pass

150V 100.000 99.600 100.230 100.400 V AC 0.230 Pass

Voltage Calibration RMS at 100 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 15.000 14.790 15.026 15.210 V AC 0.173 Pass

100V 100.000 98.600 99.554 101.400 V AC -0.446 Pass

150V 100.000 98.600 99.508 101.400 V AC -0.492 Pass

Voltage Calibration Linearity at 60 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 1.000 0.969 0.996 1.031 V AC -0.400 Pass

100V 10.000 9.960 9.997 10.040 V AC -0.032 Pass

100V 50.000 49.920 49.999 50.080 V AC -0.003 Pass

100V 110.000 109.860 110.002 110.140 V AC 0.002 Pass

Current Calibration DCRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100mA 0.10000 0.09990 0.09999 0.10010 A DC -0.011 Pass

100mA -0.10000 -0.10010 -0.10002 -0.09990 A DC 0.019 Pass

1A 1.0000 0.9990 1.0004 1.0010 A DC 0.042 Pass

1A -1.0000 -1.0010 -0.9998 -0.9990 A DC -0.016 Pass

See Next Page: 12

Certificate number Page: 12

Current Calibration AC at 10 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100mA 100.00 99.70 100.01 100.30 mA AC 0.006 Pass

1A 1.0000 0.9970 1.0001 1.0030 A AC 0.010 Pass

Current Calibration AC at 60 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

5mA 5.000 4.996 5.001 5.004 mA AC 0.022 Pass

10mA 10.00 9.99 10.00 10.01 mA AC 0.016 Pass

20mA 20.00 19.98 20.00 20.02 mA AC 0.010 Pass

50mA 50.00 49.96 50.00 50.04 mA AC -0.003 Pass

100mA 100.0 99.9 100.0 100.1 mA AC -0.007 Pass

200mA 200.0 199.8 200.0 200.2 mA AC 0.001 Pass

500mA 500.0 499.8 500.0 500.2 mA AC 0.006 Pass

1A 1.0000 0.9960 1.0001 1.0040 A AC 0.006 Pass

2A 2.0000 1.9992 2.0001 2.0008 A AC 0.007 Pass

Current Calibration AC at 1 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

10mA 10.000 9.992 10.004 10.008 mA AC 0.043 Pass

100mA 100.00 99.92 100.01 100.08 mA AC 0.007 Pass

200mA 200.00 199.84 200.03 200.16 mA AC 0.015 Pass

500mA 500.0 499.6 500.1 500.4 mA AC 0.020 Pass

1A 1.0000 0.9992 1.0002 1.0008 A AC 0.018 Pass

2A 2.0000 1.9984 2.0004 2.0016 A AC 0.021 Pass

Current Calibration AC at 10 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

200mA 200.00 199.70 200.00 200.30 mA AC -0.001 Pass

1A 1.00000 0.99850 1.00030 1.00150 A AC 0.030 Pass

Current Calibration AC at 50 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

200mA 200.00 199.20 200.05 200.80 mA AC 0.025 Pass

1A 1.00000 0.99600 1.00024 1.00400 A AC 0.024 Pass

Current Calibration AC at 100 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

50mA 50.000 49.300 49.982 50.700 mA AC -0.036 Pass

100mA 100.00 98.60 99.96 101.40 mA AC -0.037 Pass

200mA 200.00 197.20 199.95 202.80 mA AC -0.026 Pass

1A 1.00000 0.98600 0.99978 1.01400 A AC -0.022 Pass

See Next Page: 13

Certificate number Page: 13

Current Calibration Linearity AC at 60 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

1A 0.01000 0.00940 0.00999 0.01060 A AC -0.088 Pass

1A 0.10000 0.09940 0.09998 0.10060 A AC -0.015 Pass

1A 0.50000 0.49940 0.50002 0.50060 A AC 0.005 Pass

1A 1.10000 1.09940 1.09998 1.10060 A AC -0.002 Pass

External Current Calibration DC. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.9500 50.0233 50.0500 mA DC 0.047 Pass

200mV 200.000 199.800 199.993 200.200 mA DC -0.004 Pass

1V 1.00000 0.99900 1.00009 1.00100 A DC 0.009 Pass

2V 2.00000 1.99800 2.00017 2.00200 A DC 0.008 Pass

5V 5.00000 4.99500 5.00038 5.00500 A DC 0.008 Pass

External Current Calibration 10 Hz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.8500 49.9979 50.1500 mA AC -0.004 Pass

200mV 200.000 199.400 200.001 200.600 mA AC 0.001 Pass

1V 1.00000 0.99700 0.99999 1.00300 A AC -0.001 Pass

2V 2.00000 1.99400 1.99999 2.00600 A AC 0.000 Pass

5V 5.00000 4.98500 5.00003 5.01500 A AC 0.001 Pass

External Current Calibration 60 Hz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.8500 49.9999 50.1500 mA AC 0.000 Pass

100mV 100.000 99.960 99.998 100.040 mA AC -0.002 Pass

200mV 200.000 199.920 200.004 200.080 mA AC 0.002 Pass

1V 1.00000 0.99960 1.00001 1.00040 A AC 0.001 Pass

10V 10.00000 9.99600 10.00011 10.00400 A AC 0.001 Pass

External Current Calibration 1 kHz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.9700 50.0095 50.0300 mA AC 0.019 Pass

100mV 100.000 99.940 100.018 100.060 mA AC 0.018 Pass

200mV 200.000 19.988 200.048 200.120 mA AC 0.024 Pass

1V 1.00000 0.99940 1.00021 1.00060 A AC 0.021 Pass

10V 10.00000 9.99400 10.00231 10.00600 A AC 0.023 Pass

See Next Page: 14

Certificate number Page: 14

External Current Calibration 10 kHz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.9250 49.9933 50.0750 mA AC -0.013 Pass

200mV 200.000 199.700 200.094 200.300 mA AC 0.047 Pass

1V 1.00000 0.99850 1.00058 1.00150 A AC 0.058 Pass

2V 2.00000 1.99700 2.00119 2.00300 A AC 0.060 Pass

5V 5.00000 4.99250 5.00313 5.00750 A AC 0.063 Pass

External Current Calibration 50 kHz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.8000 50.0653 50.2000 mA AC 0.131 Pass

200mV 200.000 199.200 200.281 200.800 mA AC 0.141 Pass

1V 1.00000 0.99600 1.00135 1.00400 A AC 0.135 Pass

2V 2.00000 1.99200 2.00267 2.00800 A AC 0.133 Pass

5V 5.00000 4.98000 5.00734 5.02000 A AC 0.147 Pass

External Current Calibration 100 kHz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.3000 50.0927 50.7000 mA AC 0.185 Pass

200mV 200.000 197.200 200.336 202.800 mA AC 0.168 Pass

1V 1.00000 0.98600 1.00152 1.01400 A AC 0.152 Pass

2V 2.00000 1.97200 2.00298 2.02800 A AC 0.149 Pass

5V 5.00000 4.93000 5.00968 5.07000 A AC 0.194 Pass

Power Calibration DCRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 1A 15.0000 14.9770 15.0071 15.0230 W 0.047 Pass

15V 1A -15.0000 -15.0230 -15.0033 -14.9770 W 0.022 Pass

100V 1A 100.0000 99.8500 100.0534 100.1500 W 0.053 Pass

100V 1A -100.0000 -100.1500 -100.0406 -99.8500 W 0.041 Pass

150V 1A 150.000 149.770 150.050 150.230 W 0.033 Pass

150V 1A -150.000 -150.230 -150.064 -149.770 W 0.043 Pass

1000V 1A 1000.000 998.500 1000.595 1001.500 W 0.059 Pass

1000V 1A -1000.000 -1001.500 -1000.466 -998.500 W 0.047 Pass

See Next Page: 15

Certificate number Page: 15

Power Calibration 60Hz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 10mA 1.00000 0.99940 1.00014 1.00060 W 0.014 Pass

100V 20mA 2.00000 1.99880 2.00003 2.00120 W 0.001 Pass

100V 50mA 5.00000 4.99700 4.99970 5.00300 W -0.006 Pass

100V 100mA 10.0000 9.9940 9.9999 10.0060 W -0.001 Pass

100V 200mA 20.0000 19.9880 20.0011 20.0120 W 0.006 Pass

100V 500mA 50.0000 49.9700 50.0035 50.0300 W 0.007 Pass

100V 2A 200.000 199.880 200.018 200.120 W 0.009 Pass

15V 1A 15.0000 14.9910 15.0004 15.0090 W 0.003 Pass

30V 1A 30.0000 29.9820 30.0041 30.0180 W 0.014 Pass

60V 1A 60.0000 59.9640 60.0042 60.0360 W 0.007 Pass

100V 1A 100.000 99.940 100.007 100.060 W 0.007 Pass

150V 1A 150.000 149.910 150.012 150.090 W 0.008 Pass

300V 1A 300.000 299.820 300.049 300.180 W 0.016 Pass

600V 1A 600.000 599.640 600.008 600.360 W 0.001 Pass

1000V 1A 1000.00 999.40 1000.13 1000.60 W 0.013 Pass

Power Calibration 1 kHz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 10mA 1.00000 0.99900 1.00050 1.00100 W 0.050 Pass

100V 100mA 10.0000 9.9900 10.0025 10.0100 W 0.025 Pass

100V 200mA 20.0000 19.9800 20.0050 20.0200 W 0.025 Pass

100V 500mA 50.0000 49.9500 50.0117 50.0500 W 0.023 Pass

100V 2A 200.000 199.800 200.056 200.200 W 0.028 Pass

15V 1A 15.0000 14.9850 15.0027 15.0150 W 0.018 Pass

30V 1A 30.0000 29.9700 30.0086 30.0300 W 0.029 Pass

60V 1A 60.0000 59.9400 60.0126 60.0600 W 0.021 Pass

100V 1A 100.000 99.900 100.022 100.100 W 0.022 Pass

150V 1A 150.000 149.850 150.037 150.150 W 0.025 Pass

300V 1A 300.000 299.700 300.099 300.300 W 0.033 Pass

600V 1A 600.000 599.400 600.147 600.600 W 0.024 Pass

1000V 1A 1000.00 999.00 1000.25 1001.00 W 0.025 Pass

Power Calibration 10 kHz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 200mA 20.0000 17.9700 20.0120 22.0300 W 0.060 Pass

100V 1A 100.000 99.750 100.071 100.250 W 0.071 Pass

150V 1A 100.000 99.625 100.084 100.375 W 0.084 Pass

See Next Page: 16

Certificate number Page: 16

Power Calibration 50 kHz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 200mA 20.0000 15.9400 19.9680 24.0600 W -0.160 Pass

100V 1A 100.000 99.500 100.151 100.500 W 0.151 Pass

150V 1A 100.000 99.625 100.186 100.375 W 0.186 Pass

Power Calibration 100 kHz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 50mA 5.00000 3.43000 4.95865 6.57000 W -0.827 Pass

100V 100mA 10.0000 6.8600 9.9374 13.1400 W -0.626 Pass

100V 200mA 20.0000 13.7200 19.9038 26.2800 W -0.481 Pass

Power Calibration Linearity 60 Hz PF=1 at constant 1 ARange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 1A 1.0000 0.9600 0.9967 1.0400 W -0.334 Pass

100V 1A 10.0000 9.9580 9.9971 10.0420 W -0.029 Pass

100V 1A 50.0000 49.9500 50.0016 50.0500 W 0.003 Pass

100V 1A 110.0000 109.9380 110.0096 110.0620 W 0.009 Pass

Power Calibration Linearity 60 Hz PF=1 at constant 100 VRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 1A 1.0000 0.9600 0.9992 1.0400 W -0.082 Pass

100V 1A 10.0000 9.9580 9.9994 10.0420 W -0.006 Pass

100V 1A 50.0000 49.9500 50.0037 50.0500 W 0.007 Pass

100V 1A 110.0000 109.9380 109.9980 110.0620 W -0.002 Pass

External Current Power Calibration DC. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 50mV 5.00000 4.99250 5.00457 5.00750 W 0.091 Pass

100V 200mV 20.0000 19.9700 20.0039 20.0300 W 0.020 Pass

100V 1V 100.000 99.850 100.019 100.150 W 0.019 Pass

100V 2V 200.000 199.700 200.039 200.300 W 0.019 Pass

100V 5V 500.000 499.250 500.116 500.750 W 0.023 Pass

External Current Power Calibration 60 Hz. PF=1. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 50mV 5.00000 4.99700 5.00015 5.00300 W 0.003 Pass

100V 200mV 20.0000 19.9880 20.0014 20.0120 W 0.007 Pass

100V 1V 100.000 99.940 100.003 100.060 W 0.003 Pass

100V 2V 200.000 199.880 200.009 200.120 W 0.004 Pass

100V 5V 500.000 499.700 500.023 500.300 W 0.005 Pass

External Current Power Calibration 1 kHz. PF=1. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 50mV 5.00000 4.99500 5.00071 5.00500 W 0.014 Pass

100V 200mV 20.0000 19.9800 20.0050 20.0200 W 0.025 Pass

100V 1V 100.000 99.900 100.017 100.100 W 0.017 Pass

100V 2V 200.000 199.800 200.041 200.200 W 0.021 Pass

100V 5V 500.000 499.500 500.104 500.500 W 0.021 Pass

See Next Page: 17

Certificate number Page: 17

Power Calibration 60Hz PF=0 laggingRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 1A 0.0000 -0.0230 0.0002 0.0230 W - Pass

100V 1A 0.000 -0.015 0.002 0.150 W - Pass

150V 1A 0.000 -0.225 0.001 0.225 W - Pass

Power Calibration 60Hz PF=0 LeadingRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 1A 0.0000 -0.0230 -0.0002 0.0230 W - Pass

100V 1A 0.000 -0.015 -0.004 0.150 W - Pass

150V 1A 0.000 -0.225 -0.002 0.225 W - Pass

See Next Page: 18

Certificate number Page: 18

E L E M E N T 3

Voltage Calibration DCRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 15.0000 14.9850 15.0017 15.0150 V DC 0.012 Pass

15V -15.0000 -15.0150 -14.9987 -14.9850 V DC -0.009 Pass

100V 100.000 99.900 100.014 100.100 V DC 0.014 Pass

100V -100.000 -100.100 -99.992 -99.900 V DC -0.008 Pass

150V 150.000 149.850 150.005 150.150 V DC 0.003 Pass

150V -150.000 -150.150 -150.000 -149.850 V DC 0.000 Pass

1000V 1000.000 999.000 1000.154 1001.000 V DC 0.015 Pass

1000V -1000.000 -1001.000 -1000.019 -999.000 V DC 0.002 Pass

Voltage Calibration AC at 10 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 100.0000 99.7000 100.0028 100.3000 V AC 0.003 Pass

150V 150.000 149.550 150.001 150.450 V AC 0.000 Pass

Voltage Calibration RMS at 60 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 15.0000 14.9940 15.0002 15.0060 V AC 0.001 Pass

30V 30.0000 29.9880 30.0028 30.0120 V AC 0.009 Pass

60V 60.0000 59.9760 60.0013 60.0240 V AC 0.002 Pass

100V 100.000 99.960 100.003 100.040 V AC 0.003 Pass

150V 150.000 149.940 150.003 150.060 V AC 0.002 Pass

300V 300.000 299.880 300.038 300.120 V AC 0.013 Pass

600V 600.000 599.760 600.025 600.240 V AC 0.004 Pass

1000V 1000.00 999.60 1000.10 1000.40 V AC 0.010 Pass

See Next Page: 19

Certificate number Page: 19

Voltage Calibration RMS at 1 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 15.0000 14.9880 15.0008 15.0120 V AC 0.005 Pass

30V 30.0000 29.9760 30.0041 30.0240 V AC 0.014 Pass

60V 60.0000 59.9520 60.0036 60.0480 V AC 0.006 Pass

100V 100.000 99.920 100.007 100.080 V AC 0.007 Pass

150V 150.000 149.880 150.011 150.120 V AC 0.007 Pass

300V 300.000 299.760 300.054 300.240 V AC 0.018 Pass

600V 600.000 599.520 600.055 600.480 V AC 0.009 Pass

1000V 1000.00 999.20 1000.16 1000.80 V AC 0.016 Pass

Voltage Calibration RMS at 10 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 100.000 99.850 100.045 100.150 V AC 0.045 Pass

150V 150.000 149.775 150.091 150.225 V AC 0.061 Pass

Voltage Calibration RMS at 50 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 100.000 99.600 100.143 100.400 V AC 0.143 Pass

150V 100.000 99.600 100.233 100.400 V AC 0.233 Pass

Voltage Calibration RMS at 100 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 15.000 14.790 15.035 15.210 V AC 0.231 Pass

100V 100.000 98.600 99.898 101.400 V AC -0.102 Pass

150V 100.000 98.600 99.568 101.400 V AC -0.432 Pass

Voltage Calibration Linearity at 60 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 1.000 0.969 0.994 1.031 V AC -0.600 Pass

100V 10.000 9.960 9.996 10.040 V AC -0.042 Pass

100V 50.000 49.920 49.996 50.080 V AC -0.007 Pass

100V 110.000 109.860 110.004 110.140 V AC 0.004 Pass

Current Calibration DCRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100mA 0.10000 0.09990 0.09998 0.10010 A DC -0.016 Pass

100mA -0.10000 -0.10010 -0.10002 -0.09990 A DC 0.020 Pass

1A 1.0000 0.9990 1.0004 1.0010 A DC 0.035 Pass

1A -1.0000 -1.0010 -0.9998 -0.9990 A DC -0.016 Pass

See Next Page: 20

Certificate number Page: 20

Current Calibration AC at 10 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100mA 100.00 99.70 100.00 100.30 mA AC 0.004 Pass

1A 1.0000 0.9970 1.0001 1.0030 A AC 0.008 Pass

Current Calibration AC at 60 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

5mA 5.000 4.996 4.999 5.004 mA AC -0.028 Pass

10mA 10.00 9.99 10.00 10.01 mA AC 0.012 Pass

20mA 20.00 19.98 20.00 20.02 mA AC 0.007 Pass

50mA 50.00 49.96 50.00 50.04 mA AC -0.005 Pass

100mA 100.0 99.9 100.0 100.1 mA AC -0.009 Pass

200mA 200.0 199.8 200.0 200.2 mA AC 0.000 Pass

500mA 500.0 499.8 500.0 500.2 mA AC 0.004 Pass

1A 1.0000 0.9960 1.0000 1.0040 A AC 0.004 Pass

2A 2.0000 1.9992 2.0001 2.0008 A AC 0.005 Pass

Current Calibration AC at 1 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

10mA 10.000 9.992 10.004 10.008 mA AC 0.040 Pass

100mA 100.00 99.92 100.01 100.08 mA AC 0.006 Pass

200mA 200.00 199.84 200.02 200.16 mA AC 0.009 Pass

500mA 500.0 499.6 500.1 500.4 mA AC 0.018 Pass

1A 1.0000 0.9992 1.0002 1.0008 A AC 0.016 Pass

2A 2.0000 1.9984 2.0004 2.0016 A AC 0.018 Pass

Current Calibration AC at 10 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

200mA 200.00 199.70 199.99 200.30 mA AC -0.003 Pass

1A 1.00000 0.99850 1.00028 1.00150 A AC 0.028 Pass

Current Calibration AC at 50 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

200mA 200.00 199.20 200.05 200.80 mA AC 0.023 Pass

1A 1.00000 0.99600 1.00021 1.00400 A AC 0.021 Pass

Current Calibration AC at 100 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

50mA 50.000 49.300 49.976 50.700 mA AC -0.047 Pass

100mA 100.00 98.60 99.96 101.40 mA AC -0.043 Pass

200mA 200.00 197.20 199.94 202.80 mA AC -0.032 Pass

1A 1.00000 0.98600 0.99975 1.01400 A AC -0.025 Pass

See Next Page: 21

Certificate number Page: 21

Current Calibration Linearity AC at 60 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

1A 0.01000 0.00940 0.00997 0.01060 A AC -0.290 Pass

1A 0.10000 0.09940 0.09998 0.10060 A AC -0.015 Pass

1A 0.50000 0.49940 0.50000 0.50060 A AC 0.000 Pass

1A 1.10000 1.09940 1.09996 1.10060 A AC -0.004 Pass

External Current Calibration DC. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.9500 49.9594 50.0500 mA DC -0.081 Pass

200mV 200.000 199.800 199.955 200.200 mA DC -0.022 Pass

1V 1.00000 0.99900 1.00006 1.00100 A DC 0.006 Pass

2V 2.00000 1.99800 2.00015 2.00200 A DC 0.007 Pass

5V 5.00000 4.99500 5.00053 5.00500 A DC 0.011 Pass

External Current Calibration 10 Hz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.8500 49.9981 50.1500 mA AC -0.004 Pass

200mV 200.000 199.400 200.000 200.600 mA AC 0.000 Pass

1V 1.00000 0.99700 0.99999 1.00300 A AC -0.001 Pass

2V 2.00000 1.99400 1.99998 2.00600 A AC -0.001 Pass

5V 5.00000 4.98500 4.99999 5.01500 A AC 0.000 Pass

External Current Calibration 60 Hz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.8500 50.0015 50.1500 mA AC 0.003 Pass

100mV 100.000 99.960 99.999 100.040 mA AC -0.001 Pass

200mV 200.000 199.920 199.998 200.080 mA AC -0.001 Pass

1V 1.00000 0.99960 1.00000 1.00040 A AC 0.000 Pass

10V 10.00000 9.99600 10.00005 10.00400 A AC 0.000 Pass

External Current Calibration 1 kHz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.9700 50.0073 50.0300 mA AC 0.015 Pass

100mV 100.000 99.940 100.018 100.060 mA AC 0.018 Pass

200mV 200.000 19.988 200.037 200.120 mA AC 0.018 Pass

1V 1.00000 0.99940 1.00018 1.00060 A AC 0.018 Pass

10V 10.00000 9.99400 10.00200 10.00600 A AC 0.020 Pass

See Next Page: 22

Certificate number Page: 22

External Current Calibration 10 kHz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.9250 49.9828 50.0750 mA AC -0.034 Pass

200mV 200.000 199.700 200.073 200.300 mA AC 0.036 Pass

1V 1.00000 0.99850 1.00052 1.00150 A AC 0.052 Pass

2V 2.00000 1.99700 2.00110 2.00300 A AC 0.055 Pass

5V 5.00000 4.99250 5.00289 5.00750 A AC 0.058 Pass

External Current Calibration 50 kHz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.8000 50.0611 50.2000 mA AC 0.122 Pass

200mV 200.000 199.200 200.257 200.800 mA AC 0.129 Pass

1V 1.00000 0.99600 1.00127 1.00400 A AC 0.127 Pass

2V 2.00000 1.99200 2.00252 2.00800 A AC 0.126 Pass

5V 5.00000 4.98000 5.00695 5.02000 A AC 0.139 Pass

External Current Calibration 100 kHz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.3000 50.0851 50.7000 mA AC 0.170 Pass

200mV 200.000 197.200 200.314 202.800 mA AC 0.157 Pass

1V 1.00000 0.98600 1.00143 1.01400 A AC 0.143 Pass

2V 2.00000 1.97200 2.00279 2.02800 A AC 0.140 Pass

5V 5.00000 4.93000 5.00909 5.07000 A AC 0.182 Pass

Power Calibration DCRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 1A 15.0000 14.9770 15.0070 15.0230 W 0.047 Pass

15V 1A -15.0000 -15.0230 -15.0038 -14.9770 W 0.025 Pass

100V 1A 100.0000 99.8500 100.0470 100.1500 W 0.047 Pass

100V 1A -100.0000 -100.1500 -100.0294 -99.8500 W 0.029 Pass

150V 1A 150.000 149.770 150.060 150.230 W 0.040 Pass

150V 1A -150.000 -150.230 -150.071 -149.770 W 0.047 Pass

1000V 1A 1000.000 998.500 1000.613 1001.500 W 0.061 Pass

1000V 1A -1000.000 -1001.500 -1000.387 -998.500 W 0.039 Pass

See Next Page: 23

Certificate number Page: 23

Power Calibration 60Hz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 10mA 1.00000 0.99940 1.00017 1.00060 W 0.017 Pass

100V 20mA 2.00000 1.99880 1.99997 2.00120 W -0.001 Pass

100V 50mA 5.00000 4.99700 4.99968 5.00300 W -0.006 Pass

100V 100mA 10.0000 9.9940 9.9997 10.0060 W -0.003 Pass

100V 200mA 20.0000 19.9880 20.0015 20.0120 W 0.008 Pass

100V 500mA 50.0000 49.9700 50.0034 50.0300 W 0.007 Pass

100V 2A 200.000 199.880 200.017 200.120 W 0.009 Pass

15V 1A 15.0000 14.9910 15.0004 15.0090 W 0.003 Pass

30V 1A 30.0000 29.9820 30.0040 30.0180 W 0.013 Pass

60V 1A 60.0000 59.9640 60.0038 60.0360 W 0.006 Pass

100V 1A 100.000 99.940 100.007 100.060 W 0.007 Pass

150V 1A 150.000 149.910 150.009 150.090 W 0.006 Pass

300V 1A 300.000 299.820 300.048 300.180 W 0.016 Pass

600V 1A 600.000 599.640 600.006 600.360 W 0.001 Pass

1000V 1A 1000.00 999.40 1000.13 1000.60 W 0.013 Pass

Power Calibration 1 kHz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 10mA 1.00000 0.99900 1.00048 1.00100 W 0.048 Pass

100V 100mA 10.0000 9.9900 10.0021 10.0100 W 0.021 Pass

100V 200mA 20.0000 19.9800 20.0048 20.0200 W 0.024 Pass

100V 500mA 50.0000 49.9500 50.0114 50.0500 W 0.023 Pass

100V 2A 200.000 199.800 200.054 200.200 W 0.027 Pass

15V 1A 15.0000 14.9850 15.0026 15.0150 W 0.018 Pass

30V 1A 30.0000 29.9700 30.0083 30.0300 W 0.028 Pass

60V 1A 60.0000 59.9400 60.0116 60.0600 W 0.019 Pass

100V 1A 100.000 99.900 100.020 100.100 W 0.020 Pass

150V 1A 150.000 149.850 150.035 150.150 W 0.023 Pass

300V 1A 300.000 299.700 300.094 300.300 W 0.031 Pass

600V 1A 600.000 599.400 600.137 600.600 W 0.023 Pass

1000V 1A 1000.00 999.00 1000.25 1001.00 W 0.025 Pass

Power Calibration 10 kHz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 200mA 20.0000 17.9700 20.0123 22.0300 W 0.062 Pass

100V 1A 100.000 99.750 100.071 100.250 W 0.071 Pass

150V 1A 100.000 99.625 100.080 100.375 W 0.080 Pass

See Next Page: 24

Certificate number Page: 24

Power Calibration 50 kHz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 200mA 20.0000 15.9400 19.9709 24.0600 W -0.146 Pass

100V 1A 100.000 99.500 100.166 100.500 W 0.166 Pass

150V 1A 100.000 99.625 100.188 100.375 W 0.188 Pass

Power Calibration 100 kHz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 50mA 5.00000 3.43000 4.95991 6.57000 W -0.802 Pass

100V 100mA 10.0000 6.8600 9.9463 13.1400 W -0.537 Pass

100V 200mA 20.0000 13.7200 19.9151 26.2800 W -0.424 Pass

Power Calibration Linearity 60 Hz PF=1 at constant 1 ARange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 1A 1.0000 0.9600 0.9947 1.0400 W -0.534 Pass

100V 1A 10.0000 9.9580 9.9959 10.0420 W -0.041 Pass

100V 1A 50.0000 49.9500 49.9989 50.0500 W -0.002 Pass

100V 1A 110.0000 109.9380 110.0082 110.0620 W 0.007 Pass

Power Calibration Linearity 60 Hz PF=1 at constant 100 VRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 1A 1.0000 0.9600 0.9972 1.0400 W -0.282 Pass

100V 1A 10.0000 9.9580 9.9984 10.0420 W -0.016 Pass

100V 1A 50.0000 49.9500 50.0017 50.0500 W 0.003 Pass

100V 1A 110.0000 109.9380 109.9991 110.0620 W -0.001 Pass

External Current Power Calibration DC. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 50mV 5.00000 4.99250 5.00088 5.00750 W 0.018 Pass

100V 200mV 20.0000 19.9700 20.0017 20.0300 W 0.009 Pass

100V 1V 100.000 99.850 100.023 100.150 W 0.023 Pass

100V 2V 200.000 199.700 200.045 200.300 W 0.023 Pass

100V 5V 500.000 499.250 500.132 500.750 W 0.026 Pass

External Current Power Calibration 60 Hz. PF=1. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 50mV 5.00000 4.99700 5.00040 5.00300 W 0.008 Pass

100V 200mV 20.0000 19.9880 20.0008 20.0120 W 0.004 Pass

100V 1V 100.000 99.940 100.003 100.060 W 0.003 Pass

100V 2V 200.000 199.880 200.009 200.120 W 0.005 Pass

100V 5V 500.000 499.700 500.024 500.300 W 0.005 Pass

External Current Power Calibration 1 kHz. PF=1. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 50mV 5.00000 4.99500 5.00042 5.00500 W 0.008 Pass

100V 200mV 20.0000 19.9800 20.0041 20.0200 W 0.021 Pass

100V 1V 100.000 99.900 100.018 100.100 W 0.018 Pass

100V 2V 200.000 199.800 200.040 200.200 W 0.020 Pass

100V 5V 500.000 499.500 500.103 500.500 W 0.021 Pass

See Next Page: 25

Certificate number Page: 25

Power Calibration 60Hz PF=0 laggingRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 1A 0.0000 -0.0230 0.0002 0.0230 W - Pass

100V 1A 0.000 -0.015 0.002 0.150 W - Pass

150V 1A 0.000 -0.225 0.001 0.225 W - Pass

Power Calibration 60Hz PF=0 LeadingRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 1A 0.0000 -0.0230 -0.0002 0.0230 W - Pass

100V 1A 0.000 -0.015 -0.003 0.150 W - Pass

150V 1A 0.000 -0.225 -0.005 0.225 W - Pass

See Next Page: 26

Certificate number Page: 26

E L E M E N T 4

Voltage Calibration DCRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 15.0000 14.9850 15.0015 15.0150 V DC 0.010 Pass

15V -15.0000 -15.0150 -15.0000 -14.9850 V DC 0.000 Pass

100V 100.000 99.900 100.013 100.100 V DC 0.013 Pass

100V -100.000 -100.100 -100.002 -99.900 V DC 0.002 Pass

150V 150.000 149.850 150.006 150.150 V DC 0.004 Pass

150V -150.000 -150.150 -150.015 -149.850 V DC 0.010 Pass

1000V 1000.000 999.000 1000.186 1001.000 V DC 0.019 Pass

1000V -1000.000 -1001.000 -1000.062 -999.000 V DC 0.006 Pass

Voltage Calibration AC at 10 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 100.0000 99.7000 100.0045 100.3000 V AC 0.005 Pass

150V 150.000 149.550 150.004 150.450 V AC 0.003 Pass

Voltage Calibration RMS at 60 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 15.0000 14.9940 15.0003 15.0060 V AC 0.002 Pass

30V 30.0000 29.9880 30.0030 30.0120 V AC 0.010 Pass

60V 60.0000 59.9760 60.0021 60.0240 V AC 0.003 Pass

100V 100.000 99.960 100.004 100.040 V AC 0.004 Pass

150V 150.000 149.940 150.005 150.060 V AC 0.004 Pass

300V 300.000 299.880 300.041 300.120 V AC 0.014 Pass

600V 600.000 599.760 600.031 600.240 V AC 0.005 Pass

1000V 1000.00 999.60 1000.10 1000.40 V AC 0.010 Pass

See Next Page: 27

Certificate number Page: 27

Voltage Calibration RMS at 1 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 15.0000 14.9880 15.0009 15.0120 V AC 0.006 Pass

30V 30.0000 29.9760 30.0044 30.0240 V AC 0.015 Pass

60V 60.0000 59.9520 60.0046 60.0480 V AC 0.008 Pass

100V 100.000 99.920 100.009 100.080 V AC 0.009 Pass

150V 150.000 149.880 150.011 150.120 V AC 0.007 Pass

300V 300.000 299.760 300.059 300.240 V AC 0.020 Pass

600V 600.000 599.520 600.064 600.480 V AC 0.011 Pass

1000V 1000.00 999.20 1000.17 1000.80 V AC 0.017 Pass

Voltage Calibration RMS at 10 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 100.000 99.850 100.039 100.150 V AC 0.039 Pass

150V 150.000 149.775 150.082 150.225 V AC 0.055 Pass

Voltage Calibration RMS at 50 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 100.000 99.600 100.101 100.400 V AC 0.101 Pass

150V 100.000 99.600 100.184 100.400 V AC 0.184 Pass

Voltage Calibration RMS at 100 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 15.000 14.790 15.019 15.210 V AC 0.127 Pass

100V 100.000 98.600 99.859 101.400 V AC -0.141 Pass

150V 100.000 98.600 99.463 101.400 V AC -0.537 Pass

Voltage Calibration Linearity at 60 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 1.000 0.969 0.997 1.031 V AC -0.300 Pass

100V 10.000 9.960 9.999 10.040 V AC -0.014 Pass

100V 50.000 49.920 49.999 50.080 V AC -0.003 Pass

100V 110.000 109.860 110.004 110.140 V AC 0.004 Pass

Current Calibration DCRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100mA 0.10000 0.09990 0.09998 0.10010 A DC -0.020 Pass

100mA -0.10000 -0.10010 -0.10002 -0.09990 A DC 0.021 Pass

1A 1.0000 0.9990 1.0004 1.0010 A DC 0.040 Pass

1A -1.0000 -1.0010 -0.9998 -0.9990 A DC -0.019 Pass

See Next Page: 28

Certificate number Page: 28

Current Calibration AC at 10 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100mA 100.00 99.70 100.00 100.30 mA AC 0.004 Pass

1A 1.0000 0.9970 1.0001 1.0030 A AC 0.008 Pass

Current Calibration AC at 60 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

5mA 5.000 4.996 5.001 5.004 mA AC 0.025 Pass

10mA 10.00 9.99 10.00 10.01 mA AC 0.015 Pass

20mA 20.00 19.98 20.00 20.02 mA AC 0.009 Pass

50mA 50.00 49.96 50.00 50.04 mA AC -0.006 Pass

100mA 100.0 99.9 100.0 100.1 mA AC -0.008 Pass

200mA 200.0 199.8 200.0 200.2 mA AC -0.005 Pass

500mA 500.0 499.8 500.0 500.2 mA AC 0.004 Pass

1A 1.0000 0.9960 1.0001 1.0040 A AC 0.005 Pass

2A 2.0000 1.9992 2.0001 2.0008 A AC 0.006 Pass

Current Calibration AC at 1 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

10mA 10.000 9.992 10.004 10.008 mA AC 0.042 Pass

100mA 100.00 99.92 100.01 100.08 mA AC 0.005 Pass

200mA 200.00 199.84 200.02 200.16 mA AC 0.011 Pass

500mA 500.0 499.6 500.1 500.4 mA AC 0.019 Pass

1A 1.0000 0.9992 1.0002 1.0008 A AC 0.015 Pass

2A 2.0000 1.9984 2.0004 2.0016 A AC 0.020 Pass

Current Calibration AC at 10 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

200mA 200.00 199.70 200.00 200.30 mA AC -0.002 Pass

1A 1.00000 0.99850 1.00028 1.00150 A AC 0.028 Pass

Current Calibration AC at 50 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

200mA 200.00 199.20 200.05 200.80 mA AC 0.024 Pass

1A 1.00000 0.99600 1.00023 1.00400 A AC 0.023 Pass

Current Calibration AC at 100 kHzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

50mA 50.000 49.300 49.981 50.700 mA AC -0.037 Pass

100mA 100.00 98.60 99.96 101.40 mA AC -0.039 Pass

200mA 200.00 197.20 199.95 202.80 mA AC -0.027 Pass

1A 1.00000 0.98600 0.99980 1.01400 A AC -0.020 Pass

See Next Page: 29

Certificate number Page: 29

Current Calibration Linearity AC at 60 HzRange Target Lowlimit Measured Highlimit Unit Deviation % Result

1A 0.01000 0.00940 0.00995 0.01060 A AC -0.490 Pass

1A 0.10000 0.09940 0.09995 0.10060 A AC -0.045 Pass

1A 0.50000 0.49940 0.50000 0.50060 A AC 0.000 Pass

1A 1.10000 1.09940 1.09996 1.10060 A AC -0.003 Pass

External Current Calibration DC. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.9500 49.9887 50.0500 mA DC -0.023 Pass

200mV 200.000 199.800 200.015 200.200 mA DC 0.008 Pass

1V 1.00000 0.99900 1.00010 1.00100 A DC 0.010 Pass

2V 2.00000 1.99800 2.00011 2.00200 A DC 0.006 Pass

5V 5.00000 4.99500 5.00053 5.00500 A DC 0.011 Pass

External Current Calibration 10 Hz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.8500 49.9980 50.1500 mA AC -0.004 Pass

200mV 200.000 199.400 199.997 200.600 mA AC -0.001 Pass

1V 1.00000 0.99700 0.99999 1.00300 A AC -0.001 Pass

2V 2.00000 1.99400 1.99997 2.00600 A AC -0.001 Pass

5V 5.00000 4.98500 5.00002 5.01500 A AC 0.000 Pass

External Current Calibration 60 Hz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.8500 49.9970 50.1500 mA AC -0.006 Pass

100mV 100.000 99.960 99.999 100.040 mA AC -0.001 Pass

200mV 200.000 199.920 200.000 200.080 mA AC 0.000 Pass

1V 1.00000 0.99960 0.99999 1.00040 A AC -0.001 Pass

10V 10.00000 9.99600 10.00013 10.00400 A AC 0.001 Pass

External Current Calibration 1 kHz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.9700 50.0098 50.0300 mA AC 0.020 Pass

100mV 100.000 99.940 100.020 100.060 mA AC 0.020 Pass

200mV 200.000 19.988 200.036 200.120 mA AC 0.018 Pass

1V 1.00000 0.99940 1.00019 1.00060 A AC 0.019 Pass

10V 10.00000 9.99400 10.00215 10.00600 A AC 0.021 Pass

See Next Page: 30

Certificate number Page: 30

External Current Calibration 10 kHz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.9250 49.9887 50.0750 mA AC -0.023 Pass

200mV 200.000 199.700 200.071 200.300 mA AC 0.035 Pass

1V 1.00000 0.99850 1.00049 1.00150 A AC 0.049 Pass

2V 2.00000 1.99700 2.00100 2.00300 A AC 0.050 Pass

5V 5.00000 4.99250 5.00269 5.00750 A AC 0.054 Pass

External Current Calibration 50 kHz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.8000 50.0558 50.2000 mA AC 0.112 Pass

200mV 200.000 199.200 200.237 200.800 mA AC 0.119 Pass

1V 1.00000 0.99600 1.00114 1.00400 A AC 0.114 Pass

2V 2.00000 1.99200 2.00225 2.00800 A AC 0.113 Pass

5V 5.00000 4.98000 5.00632 5.02000 A AC 0.126 Pass

External Current Calibration 100 kHz. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

50mV 50.0000 49.3000 50.0773 50.7000 mA AC 0.155 Pass

200mV 200.000 197.200 200.298 202.800 mA AC 0.149 Pass

1V 1.00000 0.98600 1.00132 1.01400 A AC 0.132 Pass

2V 2.00000 1.97200 2.00255 2.02800 A AC 0.128 Pass

5V 5.00000 4.93000 5.00859 5.07000 A AC 0.172 Pass

Power Calibration DCRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 1A 15.0000 14.9770 15.0065 15.0230 W 0.043 Pass

15V 1A -15.0000 -15.0230 -15.0042 -14.9770 W 0.028 Pass

100V 1A 100.0000 99.8500 100.0494 100.1500 W 0.049 Pass

100V 1A -100.0000 -100.1500 -100.0411 -99.8500 W 0.041 Pass

150V 1A 150.000 149.770 150.046 150.230 W 0.030 Pass

150V 1A -150.000 -150.230 -150.071 -149.770 W 0.047 Pass

1000V 1A 1000.000 998.500 1000.638 1001.500 W 0.064 Pass

1000V 1A -1000.000 -1001.500 -1000.509 -998.500 W 0.051 Pass

See Next Page: 31

Certificate number Page: 31

Power Calibration 60Hz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 10mA 1.00000 0.99940 1.00016 1.00060 W 0.016 Pass

100V 20mA 2.00000 1.99880 2.00002 2.00120 W 0.001 Pass

100V 50mA 5.00000 4.99700 4.99973 5.00300 W -0.005 Pass

100V 100mA 10.0000 9.9940 9.9999 10.0060 W -0.001 Pass

100V 200mA 20.0000 19.9880 20.0012 20.0120 W 0.006 Pass

100V 500mA 50.0000 49.9700 50.0040 50.0300 W 0.008 Pass

100V 2A 200.000 199.880 200.019 200.120 W 0.010 Pass

15V 1A 15.0000 14.9910 15.0005 15.0090 W 0.004 Pass

30V 1A 30.0000 29.9820 30.0044 30.0180 W 0.015 Pass

60V 1A 60.0000 59.9640 60.0050 60.0360 W 0.008 Pass

100V 1A 100.000 99.940 100.008 100.060 W 0.008 Pass

150V 1A 150.000 149.910 150.013 150.090 W 0.008 Pass

300V 1A 300.000 299.820 300.053 300.180 W 0.018 Pass

600V 1A 600.000 599.640 600.015 600.360 W 0.002 Pass

1000V 1A 1000.00 999.40 1000.14 1000.60 W 0.014 Pass

Power Calibration 1 kHz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 10mA 1.00000 0.99900 1.00044 1.00100 W 0.044 Pass

100V 100mA 10.0000 9.9900 10.0020 10.0100 W 0.020 Pass

100V 200mA 20.0000 19.9800 20.0030 20.0200 W 0.015 Pass

100V 500mA 50.0000 49.9500 50.0114 50.0500 W 0.023 Pass

100V 2A 200.000 199.800 200.056 200.200 W 0.028 Pass

15V 1A 15.0000 14.9850 15.0030 15.0150 W 0.020 Pass

30V 1A 30.0000 29.9700 30.0082 30.0300 W 0.027 Pass

60V 1A 60.0000 59.9400 60.0121 60.0600 W 0.020 Pass

100V 1A 100.000 99.900 100.022 100.100 W 0.022 Pass

150V 1A 150.000 149.850 150.037 150.150 W 0.025 Pass

300V 1A 300.000 299.700 300.095 300.300 W 0.032 Pass

600V 1A 600.000 599.400 600.152 600.600 W 0.025 Pass

1000V 1A 1000.00 999.00 1000.26 1001.00 W 0.026 Pass

Power Calibration 10 kHz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 200mA 20.0000 17.9700 20.0109 22.0300 W 0.054 Pass

100V 1A 100.000 99.750 100.065 100.250 W 0.065 Pass

150V 1A 100.000 99.625 100.076 100.375 W 0.076 Pass

See Next Page: 32

Certificate number Page: 32

Power Calibration 50 kHz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 200mA 20.0000 15.9400 19.9638 24.0600 W -0.181 Pass

100V 1A 100.000 99.500 100.127 100.500 W 0.127 Pass

150V 1A 100.000 99.625 100.143 100.375 W 0.143 Pass

Power Calibration 100 kHz PF=1Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 50mA 5.00000 3.43000 4.95465 6.57000 W -0.907 Pass

100V 100mA 10.0000 6.8600 9.9376 13.1400 W -0.624 Pass

100V 200mA 20.0000 13.7200 19.9049 26.2800 W -0.476 Pass

Power Calibration Linearity 60 Hz PF=1 at constant 1 ARange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 1A 1.0000 0.9600 0.9977 1.0400 W -0.234 Pass

100V 1A 10.0000 9.9580 9.9989 10.0420 W -0.011 Pass

100V 1A 50.0000 49.9500 50.0009 50.0500 W 0.002 Pass

100V 1A 110.0000 109.9380 110.0084 110.0620 W 0.008 Pass

Power Calibration Linearity 60 Hz PF=1 at constant 100 VRange Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 1A 1.0000 0.9600 0.9952 1.0400 W -0.482 Pass

100V 1A 10.0000 9.9580 9.9964 10.0420 W -0.036 Pass

100V 1A 50.0000 49.9500 50.0029 50.0500 W 0.006 Pass

100V 1A 110.0000 109.9380 110.0000 110.0620 W 0.000 Pass

External Current Power Calibration DC. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 50mV 5.00000 4.99250 5.00378 5.00750 W 0.076 Pass

100V 200mV 20.0000 19.9700 20.0073 20.0300 W 0.036 Pass

100V 1V 100.000 99.850 100.028 100.150 W 0.028 Pass

100V 2V 200.000 199.700 200.044 200.300 W 0.022 Pass

100V 5V 500.000 499.250 500.141 500.750 W 0.028 Pass

External Current Power Calibration 60 Hz. PF=1. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 50mV 5.00000 4.99700 4.99996 5.00300 W -0.001 Pass

100V 200mV 20.0000 19.9880 20.0010 20.0120 W 0.005 Pass

100V 1V 100.000 99.940 100.003 100.060 W 0.003 Pass

100V 2V 200.000 199.880 200.010 200.120 W 0.005 Pass

100V 5V 500.000 499.700 500.030 500.300 W 0.006 Pass

External Current Power Calibration 1 kHz. PF=1. For current 1V/A is used.Range Target Lowlimit Measured Highlimit Unit Deviation % Result

100V 50mV 5.00000 4.99500 5.00069 5.00500 W 0.014 Pass

100V 200mV 20.0000 19.9800 20.0047 20.0200 W 0.024 Pass

100V 1V 100.000 99.900 100.018 100.100 W 0.018 Pass

100V 2V 200.000 199.800 200.042 200.200 W 0.021 Pass

100V 5V 500.000 499.500 500.111 500.500 W 0.022 Pass

See Next Page: 33

Certificate number Page: 33

Power Calibration 60Hz PF=0 laggingRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 1A 0.0000 -0.0230 0.0001 0.0230 W - Pass

100V 1A 0.000 -0.015 0.003 0.150 W - Pass

150V 1A 0.000 -0.225 0.002 0.225 W - Pass

Power Calibration 60Hz PF=0 LeadingRange Target Lowlimit Measured Highlimit Unit Deviation % Result

15V 1A 0.0000 -0.0230 -0.0003 0.0230 W - Pass

100V 1A 0.000 -0.015 -0.004 0.150 W - Pass

150V 1A 0.000 -0.225 -0.002 0.225 W - Pass

The above measurements are within the Manufacturer Specifications.

Comments :