4
People - Ideas - Solutions Insatech A/S • Algade 133 • DK-4760 Vordingborg • Denmark • Tel. +45 5537 2095 • www.insatech.com • [email protected] CALIBRATION SERVICES Accredited – in laboratory calibrations and on-site calibrations approved by major pharmaceutical industries around the world

Calibration Laboratory Services - · PDF filemajor pharmaceutical industries around the world. The Insatech Calibration Laboratory is accredited by DANAK (Danish Accreditation and

  • Upload
    donhan

  • View
    215

  • Download
    3

Embed Size (px)

Citation preview

Page 1: Calibration Laboratory Services - · PDF filemajor pharmaceutical industries around the world. The Insatech Calibration Laboratory is accredited by DANAK (Danish Accreditation and

People - Ideas - Solutions

Insatech A/S • Algade 133 • DK-4760 Vordingborg • Denmark • Tel. +45 5537 2095 • www.insatech.com • [email protected]

CALIBRATIONSERVICES

Accredited – in laboratory calibrations and on-site calibrations approved by

major pharmaceutical industries around the world

Page 2: Calibration Laboratory Services - · PDF filemajor pharmaceutical industries around the world. The Insatech Calibration Laboratory is accredited by DANAK (Danish Accreditation and

The Insatech Calibration Laboratory is accredited by DANAK (Danish Accreditation and Metrology Fund) and ILAC (International Laboratory Accreditation Cooperation) under accreditation CAL Reg. No. 484.

To achieve this accreditation our laboratory Quality Assurance system, calibration methods and technical competences have been inspected and audited by the Danish authority DANAK, and meets all requirements of ISO17025:2005.

We know that our customers appreciate working with professional and competent partners. Having worked in the fi eld of calibration for a long time, we have built up a large knowledge base and a lot of experience. To document the quality of the work being done, we decided to seek accreditation in the fi eld of conductivity, to enable primarily our customers in the pharmaceutical industry to be able to get a well documented and traceable means of calibrating their conductivity equipment to the standards required by USP (United States Pharmacopeia) or EP (European Pharmacopoeia).

We are since may 2007, the fi rst laboratory in the world to be able to offer accredited conductivity calibration of master meters that will enable users to calibrate their meters for use in the pure water areas, and actually conform to e.g. USP requirements.

The Insatech Calibration Laboratory will ensure you quick and competent handling of your conductivity calibrations with full documentation.

What criteria’s should you consider selecting a supplier for your calibrations?• Professional and appropriate testing laboratory with the right

equipment – calibrated and maintained?• Adequate quality assurance and quality control procedures e.g.

according to ISO 9001?• Appropriate sampling practices/sound testing and inspection

procedures?• Accurate and safe recording and reporting of data?

These are the minimum demands that you should ensure are fulfi lled, but even so you are not confi dent that you have selected a technically competent partner?

What is the difference between traceable calibration and accredited calibration?ISO9001 does not evaluate the technical competence of a laboratory, and does not assure you that the test, inspection and calibration data are accurate and reliable.Your decision is in this case based on the confi dence that you have in the actual supplier and your experience with the other services/products you get from the supplier.

Why use an accredited calibration laboratory?The ISO17025 standard is used for evaluating laboratories, and this standard has the criteria’s relevant to a laboratory to perform precise and accurate test and calibrations.• Technical competency of the personnel• Validity and appropriateness of the procedures and methods

• Traceability of calibrations to national standards• Appropriate application of measurement uncertainty• Suitability, calibration and maintenance of reference

equipment and the laboratory• QA of inspection and calibration data• Continued compliance are ensured by audit from

the national authorities• Suitability, calibration and maintenance of reference equipment,

laboratory environment, Quality assurance of calibration data.

All this gives you the following benefi ts of using an accredited calibration laboratory:

Reduce costs and avoid expensive retesting: Documented knowledge – you save costs and time, minimizing the cost of errors due to inaccurate or invalid measurements, misleading measurement results etc.

Enhance Your Customers and authorities conf idence: Confi dence in your product is enhanced if your customers and authorities know that critical measuring instruments has been evaluated by an independent, competent calibration facility, that has been evaluated by a third party (DANAK) We offer 2 types of conductivity calibration, and electrical calibration of transmitter.Our Conductivity calibration (cell constant) is done using traceable solutions from Danish Fundamental Metrology, and our Comparative conductivity calibrations are carried out according to accredited methods. Why calibrate conductivity?To be able to get a well documented and traceable means of calibrating your conductivity equipment to the standards required by USP (United States Pharmacopeia) or EP (European Pharmacopoeia). With an INSACAL™ master meter you get a very fl exible calibration tool. You can calibrate by closed loop procedures or in an open beaker.This allow you to choose whether your cell is calibrated on-site in-situ or not.If customers use a ”Master Meter” these need to be calibrated.

Issues that need to be def ined are:Accuracy and defi nitionsWhen to do electrical calibrationWhen to do temperature calibrationWhen to use comparative calibrationWhen to use a standard solution to determine the cell constantMeasuring ability.

Fast Service Minimizes Your Down Time for Gasflow Measurements. We Carry Out 95% of All Service and Calibration in Our Own WorkshopIn our gasflow calibration- and service workshop we have the possitility to calibrate all makes of flowmeters.(ISO 17025 accreditation pending).

Standard Calibration: 100%,75%,50%,25%,0% pointsCalibrations are based on air10mln/min < Qmax > 1500 ln/min 2,5mln/min < Qmax >_ 10 mln/minAccuracy up to ±0.5% RD + ±0.1% FS

Page 3: Calibration Laboratory Services - · PDF filemajor pharmaceutical industries around the world. The Insatech Calibration Laboratory is accredited by DANAK (Danish Accreditation and

We can calibrate flowmeters for all gasses. Calibration are done with a conversion factor.Reference equipment is traceable to the Dutch accredited calibration laboratory VSL (Dutch Metrology Institute).On-site calibrations are also performed, please ask us for a quote.

Adjustment – Cleaning – ServiceApart from calibrations, we perform all kinds of repairs*, service andcleaning with ultrasonic bath, etc. re-ranging of instruments is also offered (this is, of course, limited of the instrument capacity).On account to the fact that we perform all repairs and calibrations in our own workshop, we are able to minimize the delivery time.If you wish to sign up for a monthly or yearly review or calibration, we gladly offer you a maintenance contract tailored to your company´s requirements.

Other services we can offer:

Contact us for a specif ic and detailed quotation based on your requirements

Denmark: Insatech Calibration LaboratoryAlgade 133 – DK 4760 Vordingborg.Contact: Morten Kristensen, tel. 0045 5535 [email protected] - www.insatech.com

International gensidig anerkendelse, f.eks. NIST, UKAS

Let us help you manage your calibration intervals – make sure that your calibration equipment always is calibrated and up to date before the period of validity expires • Training courses and education of your staff to ensure optimal

effi ciency• Consultation and advice in setting up and maintaining calibration

facilities• Facilities management

On-site Calibration• On-site calibration of conductivity measuring loops approved by

major pharmaceutial industries.• On-site and laboratory calibrations of fl ow, temperature, pressure, pH

Measure Measuring rangeU(CMC)

LowLimit

U(CMC)HighLimit

WorkingStandard

Method Remark

Temperature 0,01˚C...90˚C ± 0,041˚C ± 0,041˚CPt resistancethermometre

Insatech KF5.5

Conductivity (cellconstant) 100 μS/cm (10 mS/m) 0,34% 0,34% DFM100 Insatech KF5.6 At T.sol. 25˚C ± 1˚C

Conductivity (cellconstant) 1mS/cm (100 mS/m) 0,27% 0,27% DFM1000 Insatech KF5.6 At T.sol. 25˚C ± 1˚C

Conductivity (cellconstant) 10 mS/cm (1 S/m) 0,26% 0,26% DFM10000 Insatech KF5.6 At T.sol. 25˚C ± 1˚C

Conductivity (cellconstant) 100 mS/cm (10 S/m) 0,24% 0,24% DFM100000 Insatech KF5.6 At T.sol. 25˚C ± 1˚C

Conductivity (comparisation) 1,300 μS/cm ≤ Act. ≤ 99,9 μS/cm) 0,96% af Act. 0,83% af Akt. Insatech KF5.7

Conductivity (comparisation) 100 μS/cm ≤ Act. ≤ 239 mS/cm) 0,53% af Act. 0,23% af Akt. Insatech KF5.7

Electrical (DC/AC) 0,28% 0,022% Multimeter Insatech KF5.9 Decadebox - measuring with multimeter

Electrical (DC/AC) 0,022% 0,081% Multimeter Insatech KF5.9 Decadebox - measuring with multimeter

Electrical (DC) 4 mA - 20 mA 3,4 μA 5 μA mA - loopcalibrator Insatech KF5.9 Transmitter Output

FlowMassflowmeters

Volume flow rate ln/sLow Limit 0,011 mln/sHigh Limit 1,67 mln/s

1% 0,38% Pistonprover 1 ln/h Insatech KF5.36

Calibration media: AirMeasurement is mass flowgiven as volume flow mln/s

FlowMassflowmeters

Volume flow rate ln/sLow Limit 0,00167 ln/sHigh Limit 28,1 ln/s

0,83% 0,33% Pistonprover 10 ln/h Insatech KF5.36

Calibration media: AirMeasurement is mass flowgiven as volume flow ln/s

FlowMassflowmeters

Flow rate g/sLow Limit 0,0144 mg/sHigh Limit 2,17 mg/s

1% 0,38% Rotary meter25m3/h Insatech KF5.36 Calibration media: Air

FlowMassflowmeters

Flow rate g/sLow Limit 0,00217 g/sHigh Limit 36,3 g/s

0,83% 0.33% Rotary meter100 m3/h Insatech KF5.36 Calibration media: Air

Learn more: www.insacal.com or contact Morten Kristensen, [email protected]

Page 4: Calibration Laboratory Services - · PDF filemajor pharmaceutical industries around the world. The Insatech Calibration Laboratory is accredited by DANAK (Danish Accreditation and

Mennesker - Ideer - Løsninger

Insatech A/S • Algade 133 • 4760 Vordingborg • Tlf. +45 5537 2095 • www.insatech.com • [email protected]

The Insatech Calibration Laboratory is accredited by DANAK (Danish Accreditation and Metrology Fund ) under accreditation CAL Reg.No. 484.

To achieve this accreditation our laboratory Quality Assurance system, calibration methods and technical competences have been inspected and audited by the Danish authority DANAK, and meets all requirements of ISO17025:2005.

DANAK is a Service Company handling the administration of accreditation and metrology in Denmark based on a contract with The Danish Safety Technology Authority which is part of The Danish Ministry of Economics and Business Affairs and a member of ILAC.

ILAC – the international Laboratory Accreditation Cooperation – is an international cooperation of laboratory and inspection accreditation bodies formed more than 30 years ago to help remove technical barriers to trade.

Furthermore all our activities in Insatech A/S are certifi ed according to ISO/IEC 9001:2008, and has been, since 1994.

You can fi nd more informations on the web sites:

www.insatech.com www.danak.dk

www.insacal.com www.ilac.org20