1
Abstract The levels of the organic acids in a fruit juice are important to the assessment of authencity. Two separate approaches are rounely used; HPLC linked with either UV or conducvity detecon, or enzymac procedures specific for each acid. HPLC-UV-Vis is used widely in the US for this analysis. There are AOAC (986.13) and Internaonal Fruit Juice (IFU 65) methods available. As the acids absorb in the far UV, it is hard to get a definive UV spectrum that makes posive idenficaon difficult when dealing with low levels, which is a limitaon. The enzymac methods show very good specificity and sensivity, with detecon limits typically around 10 mg/L or higher depending on the diluon used. However, only one acid can be measured at a me, which makes the procedure more me consuming to use. The kits also have limited shelf life, a8er opening, which increases analysis costs with limited numbers of samples. This poster presents a comparison of the AOAC method with a new UPLC-MS/MS procedure that offers much be;er specificity than the UV approach and similar or be;er detecon limits to the enzymac methods. Using Waters UPLC technology it also significantly reduces analysis mes, therefore improving sample throughput. Comparable results were obtained for a number of juices. The repeatability was found to be similar for both methods. The MS/MS method was found to have a smaller linear range than the UV method but the improved selecvity easily offsets this limitaon. HPLC Setup HPLC: Waters 2695 alliance Separaon Module Detector: Waters 2996, Photodiode Array Detector Column: Restek Allure Organic Acids 5µm, 300 x 4.6mm Total Runme: 50 min Injecon Volume: 20µL Column Temperature: 30.0°C Sample Temperature: 10.0°C UPLC Setup UPLC: Waters ACQUITY H-Class Detector: Waters Xevo TQD Column: Waters BEH C18 1.7µm, 2.1 x 100mm Total Runme: 12 min Injecon Volume: 1µL Column Temperature: 30.0°C Sample Temperature: 10.0°C Materials and Methods HPLC with UV-Vis detecon method Mobile Phase: A = 50mM potassium phosphate, pH 2.5 B = 2% methanol in HPLC water C = Acetonitrile Gradient: Time (min) Flow Rate (mL/min) % Mobile Phase A % Mobile Phase C Inial 0.800 100 0 35.00 0.800 100 0 35.10 0.800 0 75 40.00 0.800 0 75 40.10 0.800 100 0 50.00 0.800 100 0 % Mobile Phase B 0 0 25 25 0 0 Detecon: Waters 2996 PDA Wavelength: 226nm Sampling Rate: 0.50 points/s UPLC with MS/MS detecon method Mobile Phase: A = Deionized Water B = 1% Formic Acid in Water C = Acetonitrile Gradient: Time (min) Flow Rate (mL/min) % Mobile Phase A % Mobile Phase C Inial 0.350 50 0 1.15 0.350 50 0 2.30 0.350 30 40 2.90 0.350 30 40 2.91 0.350 0 100 3.45 0.350 0 100 % Mobile Phase B 50 50 30 30 0 0 3.46 0.350 50 50 0 Detecon: Water Xevo TQD Source, Mode: Electrospray, Negave Ion Capillary Voltage: 1kV Dwell Time: 0.020s Channel Cone Coll. Energy Compound 1: 133 > 71 24.0 14.0 Malic Acid (Qualitave) 2: 133 > 115 24.0 10.0 Malic Acid (Quantave) 3: 149 > 73 28.0 16.0 Tartaric Acid (Qualitave) 4: 149 > 87 28.0 12.0 Tartaric Acid (Quantave) 5: 191 > 85 40.0 22.0 Quinic Acid (Quantave) 6: 191 > 87 26.0 18.0 Citric Acid (Qualitave) 7: 191 > 93 40.0 22.0 Quinic Acid (Qualitave) 8: 191 > 111 26.0 10.0 Citric Acid (Quantave) Sample Preparaon All samples used in this study were prepared in duplicate at ENAC. The duplicate set was shipped overnight to Milford, MA and the original kept at ENAC. Fruit Name Sample Name Brix ° Source Apple 1 11.5 Apple concentrate #1 2 11.5 Apple concentrate #2 3 11.5 Commercial apple juice 4 11.5 0.10230g malic + 0.05361g tartaric + 0.05351g quinic + apple concentrate #1 Lemon 1 7.0 Organic Lemon concentrate 2 7.0 Commercial lemon juice 3 7.0 Lemon juice from concentrate 4 7.0 1.00578g citric + 0.10179g malic + organic lemon Orange 1 11.8 Orange concentrate 2 11.8 Organic orange concentrate 3 11.8 Commercial orange juice, pulp-free 4 11.8 0.10058g citric + 0.10803g malic + orange Cranberry 1 7.5 Organic cranberry concentrate 2 7.5 Commercial cranberry juice, 100% pure 3 7.5 Commercial cranberry cocktail 4 7.5 0.10762g quinic + 0.50368g citric + 0.10528g White Grape 1 16.0 White grape concentrate 2 16.0 Organic white grape concentrate 3 16.0 Fresh squeezed white grape juice 4 16.0 0.10568g tartaric + 0.05262g malic + 0.05375g Pomegranate 1 16.0 Organic pomegranate concentrate 2 16.0 Pomegranate concentrate 3 16.0 Commercial pomegranate juice, 100% pure 4 16.0 0.10166g citric + 0.10169g malic + organic Sample Preparaon Each of the 24 samples were diluted and filtered at the two respecve laboratories in Des Moines and Milford. HPLC-UV-Vis UPLC-MS/MS Solvent HPLC grade water Solvent Deionized water Diluons 1:1, 1:10 & 1:50 Diluons 1:100, 1:1000 & 1:10000 Filtraon A8er diluon by 0.45µm syringeless filter device Filtraon Prior to diluon by 0.45µm membrane Standard Preparaon Organic Acid Standards for HPLC-UV-Vis Method Name Stock (ppm) Diluon Series (ppm) L-Tartaric 2048.5 1000, 500, 250, 100, 50, 25 D-Quinic 2033.4 1000, 500, 250, 100, 50, 25 L-Malic 4061.8 3000, 2000, 1000, 500, 250, 100 Citric 2034.2 1500, 1000, 500, 250, 100, 50, 25 Fumaric 533.2* 15, 10, 5, 2, 1 *not injected due to high UV response; only serial diluons used All standards were purchased from ACROS, except L-Tartaric is from TCI. Organic Acid Standards for UPLC-MS/MS Method Individual Stocks 10000 ppm Individual Stocks 1000 ppm Citric Isocitric Malonic Succinic Glutaric Lacc Quinic Tartaric Isoascorbic Malic Shikimic Benzoic Fumaric A 1000 ppm standard mixture was prepared from stocks above and diluon series was made from 10ppm down to 0.1ppm Discussion Overall, a good correlaon between the two methods was seen. As expected, data provided by LC-MS/MS demonstrated slightly lower calculated concentraons owing to the higher specificity of the technique. One notable excepon was the calculated concentraon of quinic acid in citrus juices (orange and lemon) where an obvious interference results in a posive bias to reported values. LC-MS/MS is more accurately able to detect lower concentraons of some organic acids such as citric acid in apple juice and tartaric acid, detected in one specimen of pomegranate juice. Also, notably lower values for malic acid in pomegranate juice were reported by LC-MS/MS likely indicang an interference in the LC-UV-Vis method. Recoveries for the organic acids ranged from 71 –160%, with the majority of calculated recoveries lying in the range 90-105%. Analycal precision based upon replicate injecons (n=6) of the same apple juice demonstrated nominally higher values for LC-MS/MS owing to the lack of an internal standard, which may be easily corrected. Conclusions 1) The LC-UV-Vis method has higher linearity than UPLC-MS/MS procedure (requiring fewer sample diluon steps for an “on scale” result). 2) The LC-UV-Vis method is a well established procedure and a recognized method, AOAC 986.13. 3) UPLC offers be;er resoluon than convenonal HPLC by reducing the chance of matrix-borne interferences. 4) UPLC-MS/MS is faster; > sample throughput = > producvity. 5) UPLC-MS/MS has be;er sensivity than the LC-UV-Vis method allowing the detecon of trace levels of organic acids in juices. 6) UPLC-MS/MS is more specific than the LC-UV-Vis method and mulple MRM channels offer greater analycal confidence in quantave analyses. 7) LC-MS/MS represents a promising alternave to LC-UV-Vis for the quanficaon of organic acids in fruit juices. Contacts: Cassandra Taylor—CassandraTaylor@eurofinsus.com, Kendon Graham—[email protected], Ramin Jahromi—RaminJahromi@eurofinsus.com, David Hammond—DavidHammond@eurofins.com PROS & CONS HPLC-UV Method PROS CONS Low diluons (i.e. 1:1, 1:10, 1:50) therefore smaller diluon errors Low sensivity Large linear range for UV detector Long run mes (50 min) Established method, AOAC 986.13 Large injecon volume (20 µL) HPLC column can be re-condioned to improve separaon and peak shapes LOD significantly degrades below 25ppm (except fumaric acid because of strong UV absorbance) Peak shape is variable from sample to sample Limited number of analytes (≤ 8) can be analyzed in one 50 minute run λ max values for organic acids are too low (200- 220nm) to use PDA for confirmaon of analytes Poor chromatographic separaon and non-selecve detecon raises the chances of interferences e.g. malic acid in pomegranate juice Large amount of mobile phase is required (i.e. 500mL per 10 samples), which creates more hazardous waste Phosphate buffer degrades instrument parts causing constant maintenance UPLC-MS Method PROS CONS High sensivity High diluons (i.e. 1:100, 1:1000, 1:10000) create greater possibility for diluon errors Short run mes (12 min) Smaller linear range than seen with UV detecon Small injecon volume (1 µL) Further method development required to resolve interference associated with quinic acid Screen ≥ 10 analytes in one 12 min run Mulple MRM’s can be used to confirm analyte’s identy Greater separaon efficiency effected by UPLC LOD <<10ppm Cassandra Taylor 1 , Kendon Graham 2 , Ramin Jahromi 1 and Dr. David Hammond 3 1 Eurofins Nutrion Analysis Center (ENAC), Des Moines, IA 50321, 2 Waters Corporaon, Milford, MA 01757, 3 Eurofins Scienfic France, Nantes Cedex 3, F-44323 Data Precision, Apple (n=6) Quinic Malic Citric LC-UV LC-MS/MS LC-UV LC-MS/MS LC-UV LC-MS/MS Mean Concentraon (ppm) 259 266 2738 2826 106.5 44.9 Standard Deviaon 9.0 11.4 76 115 4.7 1.0 %RSD 3.5 4.3 2.8 4.1 4.4 2.2 Precision, Apple (n=3) Quinic Malic Citric LC-UV LC-MS/MS LC-UV LC-MS/MS LC-UV LC-MS/MS Mean Concentraon (ppm) 259 260 2720 2763 105.7 44 Standard Deviaon 7.3 12 80 91 5.7 1.1 %RSD 2.8 4.8 2.9 3.3 5.4 2.4 Detecon of Citric Acid in Apple Juice by LC-MS/MS (min) Detecon of Tartaric Acid in Pomegranate Juice by LC-MS/MS (min) Background The adulteraon of fruit juices by producers has been a problem for years and remains so to this day. The first successful legal case taken to court in the US was in the early 80's against Beechnut. They were selling a product labeled as “apple juice” but probably contained very li;le juice at all. Although things have significantly improved since then, there are sll cases of adulterated products appearing on the market, e.g. just recently with lemon juice. Many different approaches have been used to try and eradicate these problems, from simple HPLC procedures to look at the individual concentraons of the sugars and organic acids, to complex isotopic methods to assess the internal raos within the juice. The level of tesng sophiscaon required depends on what approach the producer is using and can follow a cyclic pa;ern where people seem to “forget” what can be detected! While the price of fruit juice solids are significantly higher than that of sugar and acids, users of juice must remain vigilant to ensure that they are not using adulterated product. This vigilance should include audits of suppliers and tesng of their products on a regular basis. If there is no experse within the company to carry out these audits then third pares can be contracted (such as www.Eurofinsus.com) or materials can be purchased from an IRMA (Internaonal Raw Material Assurance) approved supplier. IRMA approved companies undergo an independent check every two years to ensure they are producing acceptable products. Further informaon about this scheme can be found here: h;p://www.sgf.org/en/home/_s/halbwarenkontrolle/ As highlighted above, one of the important sets of parameters to look at are the organic acids. These can be measured by HPLC using UV-Vis detecon, ion chromatography or enzymacally, but all these procedures have some limitaons. This poster describes another very specific procedure by which these compounds can be quanfied. This uses chromatography linked with mass spectrometry, which provides a very sensive and specific detecon system. The poster compares the results from this approach using UPLC and mass spectrometric detecon with those for the convenonal AOAC method 986.13. Detecon of Citric Acid in Apple Juice by LC-UV-Vis Detecon of Tartaric Acid in Pomegranate Juice by LC-UV-Vis This work is licensed under the Creave Commons A;ribuon 3.0 Unported License. To view a copy of this license, visit h;p://creavecommons.org/licenses/by/3.0/

Cassandra Taylor1, Kendon Graham2, Ramin Jahromi1and · PDF file · 2016-06-17Established method, AOAC 986.13 Large injec on volume (20 µL) HPLC column can be re-condi oned to improve

  • Upload
    dolien

  • View
    214

  • Download
    2

Embed Size (px)

Citation preview

Page 1: Cassandra Taylor1, Kendon Graham2, Ramin Jahromi1and · PDF file · 2016-06-17Established method, AOAC 986.13 Large injec on volume (20 µL) HPLC column can be re-condi oned to improve

Abstract

The levels of the organic acids in a fruit juice are important to the

assessment of authen�city. Two separate approaches are rou�nely

used; HPLC linked with either UV or conduc�vity detec�on, or

enzyma�c procedures specific for each acid.

HPLC-UV-Vis is used widely in the US for this analysis. There are

AOAC (986.13) and Interna�onal Fruit Juice (IFU 65) methods available.

As the acids absorb in the far UV, it is hard to get a defini�ve UV

spectrum that makes posi�ve iden�fica�on difficult when dealing with

low levels, which is a limita�on.

The enzyma�c methods show very good specificity and sensi�vity,

with detec�on limits typically around 10 mg/L or higher depending on

the dilu�on used. However, only one acid can be measured at a �me,

which makes the procedure more �me consuming to use. The kits also

have limited shelf life, a8er opening, which increases analysis costs

with limited numbers of samples.

This poster presents a comparison of the AOAC method with a new

UPLC-MS/MS procedure that offers much be;er specificity than the UV

approach and similar or be;er detec�on limits to the enzyma�c

methods. Using Waters UPLC technology it also significantly reduces

analysis �mes, therefore improving sample throughput.

Comparable results were obtained for a number of juices. The

repeatability was found to be similar for both methods. The MS/MS

method was found to have a smaller linear range than the UV method

but the improved selec�vity easily offsets this limita�on.

HPLC Setup

HPLC: Waters 2695 alliance

Separa�on Module

Detector: Waters 2996,

Photodiode Array Detector

Column: Restek Allure Organic

Acids 5µm, 300 x 4.6mm

Total Run�me: 50 min

Injec�on Volume: 20µL

Column Temperature: 30.0°C

Sample Temperature: 10.0°C

UPLC Setup

UPLC: Waters ACQUITY H-Class

Detector: Waters Xevo TQD

Column: Waters BEH C18

1.7µm, 2.1 x 100mm

Total Run�me: 12 min

Injec�on Volume: 1µL

Column Temperature: 30.0°C

Sample Temperature: 10.0°C

Materials and Methods

HPLC with UV-Vis detec�on method

Mobile

Phase:

A = 50mM potassium phosphate, pH 2.5

B = 2% methanol in HPLC water

C = Acetonitrile

Gradient: Time (min)

Flow Rate

(mL/min)

% Mobile

Phase A

% Mobile

Phase C

Ini�al 0.800 100 0

35.00 0.800 100 0

35.10 0.800 0 75

40.00 0.800 0 75

40.10 0.800 100 0

50.00 0.800 100 0

% Mobile

Phase B

0

0

25

25

0

0

Detec�on: Waters 2996 PDA

Wavelength: 226nm

Sampling Rate: 0.50 points/s

UPLC with MS/MS detec�on method

Mobile

Phase:

A = Deionized Water

B = 1% Formic Acid in Water

C = Acetonitrile

Gradient: Time (min)

Flow Rate

(mL/min)

% Mobile

Phase A

% Mobile

Phase C

Ini�al 0.350 50 0

1.15 0.350 50 0

2.30 0.350 30 40

2.90 0.350 30 40

2.91 0.350 0 100

3.45 0.350 0 100

% Mobile

Phase B

50

50

30

30

0

0

3.46 0.350 50 50 0

Detec�on: Water Xevo TQD

Source, Mode: Electrospray, Nega�ve Ion

Capillary Voltage: 1kV

Dwell Time: 0.020s

Channel Cone Coll. Energy Compound

1: 133 > 71 24.0 14.0 Malic Acid (Qualita�ve)

2: 133 > 115 24.0 10.0 Malic Acid (Quan�ta�ve)

3: 149 > 73 28.0 16.0 Tartaric Acid (Qualita�ve)

4: 149 > 87 28.0 12.0 Tartaric Acid (Quan�ta�ve)

5: 191 > 85 40.0 22.0 Quinic Acid (Quan�ta�ve)

6: 191 > 87 26.0 18.0 Citric Acid (Qualita�ve)

7: 191 > 93 40.0 22.0 Quinic Acid (Qualita�ve)

8: 191 > 111 26.0 10.0 Citric Acid (Quan�ta�ve)

Sample Prepara�on

All samples used in this study were prepared in duplicate at ENAC. The duplicate set

was shipped overnight to Milford, MA and the original kept at ENAC.

Fruit Name Sample Name Brix ° Source

Apple

1 11.5 Apple concentrate #1

2 11.5 Apple concentrate #2

3 11.5 Commercial apple juice

4 11.5 0.10230g malic + 0.05361g tartaric + 0.05351g

quinic + apple concentrate #1

Lemon

1 7.0 Organic Lemon concentrate

2 7.0 Commercial lemon juice

3 7.0 Lemon juice from concentrate

4 7.0 1.00578g citric + 0.10179g malic + organic lemon

Orange

1 11.8 Orange concentrate

2 11.8 Organic orange concentrate

3 11.8 Commercial orange juice, pulp-free

4 11.8 0.10058g citric + 0.10803g malic + orange

Cranberry

1 7.5 Organic cranberry concentrate

2 7.5 Commercial cranberry juice, 100% pure

3 7.5 Commercial cranberry cocktail

4 7.5 0.10762g quinic + 0.50368g citric + 0.10528g

White Grape

1 16.0 White grape concentrate

2 16.0 Organic white grape concentrate

3 16.0 Fresh squeezed white grape juice

4 16.0 0.10568g tartaric + 0.05262g malic + 0.05375g

Pomegranate

1 16.0 Organic pomegranate concentrate

2 16.0 Pomegranate concentrate

3 16.0 Commercial pomegranate juice, 100% pure

4 16.0 0.10166g citric + 0.10169g malic + organic

Sample Prepara�on

Each of the 24 samples were diluted and filtered at the two respec�ve laboratories in Des

Moines and Milford.

HPLC-UV-Vis UPLC-MS/MS

Solvent HPLC grade water Solvent Deionized water

Dilu�ons 1:1, 1:10 & 1:50 Dilu�ons 1:100, 1:1000 & 1:10000

Filtra�on A8er dilu�on by 0.45µm

syringeless filter device

Filtra�on Prior to dilu�on by 0.45µm

membrane

Standard Prepara�on

Organic Acid Standards for HPLC-UV-Vis Method

Name Stock (ppm) Dilu�on Series (ppm)

L-Tartaric 2048.5 1000, 500, 250, 100, 50, 25

D-Quinic 2033.4 1000, 500, 250, 100, 50, 25

L-Malic 4061.8 3000, 2000, 1000, 500, 250, 100

Citric 2034.2 1500, 1000, 500, 250, 100, 50, 25

Fumaric 533.2* 15, 10, 5, 2, 1

*not injected due to high UV response; only serial dilu�ons used

All standards were purchased from ACROS, except L-Tartaric is from TCI.

Organic Acid Standards for UPLC-MS/MS Method

Individual Stocks 10000 ppm Individual Stocks 1000 ppm

Citric Isocitric Malonic Succinic

Glutaric Lac�c Quinic Tartaric

Isoascorbic Malic Shikimic

Benzoic

Fumaric

A 1000 ppm standard mixture was prepared from stocks above and dilu�on series

was made from 10ppm down to 0.1ppm

Discussion

Overall, a good correla�on between the two methods was seen. As expected,

data provided by LC-MS/MS demonstrated slightly lower calculated concentra�ons

owing to the higher specificity of the technique. One notable excep�on was the

calculated concentra�on of quinic acid in citrus juices (orange and lemon) where an

obvious interference results in a posi�ve bias to reported values. LC-MS/MS is more

accurately able to detect lower concentra�ons of some organic acids such as citric

acid in apple juice and tartaric acid, detected in one specimen of pomegranate juice.

Also, notably lower values for malic acid in pomegranate juice were reported by

LC-MS/MS likely indica�ng an interference in the LC-UV-Vis method. Recoveries for

the organic acids ranged from 71 –160%, with the majority of calculated recoveries

lying in the range 90-105%. Analy�cal precision based upon replicate injec�ons

(n=6) of the same apple juice demonstrated nominally higher values for LC-MS/MS

owing to the lack of an internal standard, which may be easily corrected.

Conclusions

1) The LC-UV-Vis method has higher linearity than UPLC-MS/MS procedure

(requiring fewer sample dilu�on steps for an “on scale” result).

2) The LC-UV-Vis method is a well established procedure and a recognized method,

AOAC 986.13.

3) UPLC offers be;er resolu�on than conven�onal HPLC by reducing the chance of

matrix-borne interferences.

4) UPLC-MS/MS is faster; > sample throughput = > produc�vity.

5) UPLC-MS/MS has be;er sensi�vity than the LC-UV-Vis method allowing the

detec�on of trace levels of organic acids in juices.

6) UPLC-MS/MS is more specific than the LC-UV-Vis method and mul�ple MRM

channels offer greater analy�cal confidence in quan�ta�ve analyses.

7) LC-MS/MS represents a promising alterna�ve to LC-UV-Vis for the

quan�fica�on of organic acids in fruit juices.

Contacts: Cassandra Taylor—[email protected], Kendon Graham—[email protected], Ramin Jahromi—[email protected], David Hammond—[email protected]

PROS & CONS

HPLC-UV Method

PROS CONS

Low dilu�ons (i.e. 1:1, 1:10, 1:50)

therefore smaller dilu�on errors

Low sensi�vity

Large linear range for UV detector Long run �mes (50 min)

Established method, AOAC 986.13 Large injec�on volume (20 µL)

HPLC column can be re-condi�oned to

improve separa�on and peak shapes

LOD significantly degrades below 25ppm (except

fumaric acid because of strong UV absorbance)

Peak shape is variable from sample to sample

Limited number of analytes (≤ 8) can be analyzed in

one 50 minute run

λmax values for organic acids are too low (200-

220nm) to use PDA for confirma�on of analytes

Poor chromatographic separa�on and non-selec�ve

detec�on raises the chances of interferences e.g.

malic acid in pomegranate juice

Large amount of mobile phase is required

(i.e. 500mL per 10 samples), which creates more

hazardous waste

Phosphate buffer degrades instrument parts causing

constant maintenance

UPLC-MS Method

PROS CONS

High sensi�vity High dilu�ons (i.e. 1:100, 1:1000, 1:10000) create

greater possibility for dilu�on errors

Short run �mes (12 min) Smaller linear range than seen with UV detec�on

Small injec�on volume (1 µL) Further method development required to resolve

interference associated with quinic acid

Screen ≥ 10 analytes in one 12 min

run

Mul�ple MRM’s can be used to

confirm analyte’s iden�ty

Greater separa�on efficiency effected

by UPLC

LOD <<10ppm

Cassandra Taylor1, Kendon Graham

2, Ramin Jahromi

1and Dr. David Hammond

3

1Eurofins Nutri�on Analysis Center (ENAC), Des Moines, IA 50321,

2 Waters Corpora�on, Milford, MA 01757,

3Eurofins Scien�fic France, Nantes Cedex 3, F-44323

Data

Precision, Apple (n=6) Quinic Malic Citric

LC-UV LC-MS/MS LC-UV LC-MS/MS LC-UV LC-MS/MS

Mean Concentra�on (ppm) 259 266 2738 2826 106.5 44.9

Standard Devia�on 9.0 11.4 76 115 4.7 1.0

%RSD 3.5 4.3 2.8 4.1 4.4 2.2

Precision, Apple (n=3) Quinic Malic Citric

LC-UV LC-MS/MS LC-UV LC-MS/MS LC-UV LC-MS/MS

Mean Concentra�on (ppm) 259 260 2720 2763 105.7 44

Standard Devia�on 7.3 12 80 91 5.7 1.1

%RSD 2.8 4.8 2.9 3.3 5.4 2.4

Detec�on of Citric Acid in Apple Juice by LC-MS/MS

(min)

Detec�on of Tartaric Acid in Pomegranate Juice by LC-MS/MS

(min) Background

The adultera�on of fruit juices by producers has been a problem

for years and remains so to this day. The first successful legal case

taken to court in the US was in the early 80's against Beechnut. They

were selling a product labeled as “apple juice” but probably contained

very li;le juice at all. Although things have significantly improved since

then, there are s�ll cases of adulterated products appearing on the

market, e.g. just recently with lemon juice.

Many different approaches have been used to try and eradicate

these problems, from simple HPLC procedures to look at the individual

concentra�ons of the sugars and organic acids, to complex isotopic

methods to assess the internal ra�os within the juice. The level of

tes�ng sophis�ca�on required depends on what approach the

producer is using and can follow a cyclic pa;ern where people seem to

“forget” what can be detected!

While the price of fruit juice solids are significantly higher than that

of sugar and acids, users of juice must remain vigilant to ensure that

they are not using adulterated product. This vigilance should include

audits of suppliers and tes�ng of their products on a regular basis. If

there is no exper�se within the company to carry out these audits then

third par�es can be contracted (such as www.Eurofinsus.com) or

materials can be purchased from an IRMA (Interna�onal Raw Material

Assurance) approved supplier. IRMA approved companies undergo an

independent check every two years to ensure they are producing

acceptable products. Further informa�on about this scheme can be

found here: h;p://www.sgf.org/en/home/_s/halbwarenkontrolle/

As highlighted above, one of the important sets of parameters to

look at are the organic acids. These can be measured by HPLC using

UV-Vis detec�on, ion chromatography or enzyma�cally, but all these

procedures have some limita�ons. This poster describes another very

specific procedure by which these compounds can be quan�fied. This

uses chromatography linked with mass spectrometry, which provides a

very sensi�ve and specific detec�on system. The poster compares the

results from this approach using UPLC and mass spectrometric

detec�on with those for the conven�onal AOAC method 986.13.

Detec�on of Citric Acid in Apple Juice by LC-UV-Vis

Detec�on of Tartaric Acid in Pomegranate Juice by LC-UV-Vis

This work is licensed under the Crea�ve Commons A;ribu�on 3.0 Unported License. To view a copy of this license, visit h;p://crea�vecommons.org/licenses/by/3.0/