Analysis of Aseptic Loosening of the Acetabular Cup in Cemented Total Hip Replacement Tadhg Adderley¹, Dr. F. McEvoy¹, Dr. Y. Blake¹, Mr. R. Flavin². 1. Mechanical Engineering Department, Institute of Technology Tallaght, Dublin. 2. Orthopaedic Surgeon, Galway University Hospital.

Analysis of Aseptic Loosening of the Acetabular Cup in Cemented Total Hip

Embed Size (px)

DESCRIPTION

biomechanics

Citation preview

  • AnalysisofAsepticLooseningoftheAcetabularCupinCementedTotalHip

    Replacement

    TadhgAdderley,Dr.F.McEvoy,Dr.Y.Blake,Mr.R.Flavin.

    1. MechanicalEngineeringDepartment,InstituteofTechnologyTallaght,Dublin.

    2. OrthopaedicSurgeon,GalwayUniversityHospital.

  • 2

    Analysis of Aseptic Loosening of the Acetabular Cup

    Overview

    1.

    Introduction

    2.

    Project Objectives

    3.

    Pelvis Model Design & Validation

    4.

    Foam Compression Tests & Results

    5.

    Hip Simulator Design & Calibration

    6.

    Future Work

  • 3

    Analysis of Aseptic Loosening of the Acetabular Cup

    1. Introduction

    Over1millionTotalHipReplacements(THR)performedworldwideeachyear. 300,000inAmerica 50,000inEngland&Wales 3,389inIreland(2004)

    Vhipaidout18milliononTHRin04/05. 11,500peroperation

    Between710%ofTHRwillrequirearevision.

    75%ofrevisionscausedbyasepticloosening.TA4

  • Slide3

    TA4 AsepticlooseningofTHRisthemajorreasonforrevisionsurgery.

    Thispaperaimstovalidateapelvismodel,socementedcupscanbetestedandexaminedforcausesofasepticloosening.

    Thepelvismodelwillbeusedonapurposebuilthipsimulatortotestcementedacetabularcups.TadhgAdderley,29/01/2009

  • 4

    Analysis of Aseptic Loosening of the Acetabular Cup

    2. Project Objectives

    Design&build: HipSimulatortoBS&ISOstandards. PelvisModel.

    Findsuitablereplacementmaterialforbone.

    Determine: Theinitiatingfactorsthatcauseasepticloosening. Inwhichzoneorareafailureoccurs.

    Investigatethemigrationofthecupoveritslifespan.

    Identifytheoptimumdesigncriteriafortheacetabularcup.

  • 5

    Analysis of Aseptic Loosening of the Acetabular Cup

    Total Hip Replacement

    3.Pelvis Model Design

  • 6

    Analysis of Aseptic Loosening of the Acetabular Cup

    3. PelvisModel

    0.3 1.7

    15 25

    Majumderetal.(2004)

    2.21

    17.68

    Cilingiretal.(2007)

    Lamvoheeetal.(2006)

    ObtainedResults

    1.2 3.53 Trabecular Bone

    15 16.48 Cortical Bone

    7.27 8.84 7.06 Cement

    {9 12}

    15.3

    {10.34}

    Cup

    VonMisesStress(MPa)

    Cu p

    Cortical Bone

    Cemen t

    Loading Conditions on Pelvis Model

    Von Mises Stress Results

  • 7

    Analysis of Aseptic Loosening of the Acetabular Cup

    0.12g/cc Cellular Rigid Polyurethane Foam

    0.32g/cc Cellular Rigid Polyurethane Foam

    Crack

    4. CompressionTests

  • 8

    Analysis of Aseptic Loosening of the Acetabular Cup

    4. Compression Test Results

    0

    1

    2

    3

    4

    5

    6

    7

    8

    9

    0% 1% 2% 3% 4% 5% 6%

    TrueStrain

    Tru

    eStres

    s(M

    Pa)

    0.2%ProofStress=6.25MPa

    Young'sModulus=298.76MPa

    UltimateStress=7.63MPa

    0.3g/cc Trabecular Bone Youngs Modulus 200 MPa

    Yield Point 8 MPa

    Ultimate Stress 8-10 MPa

    [Bartel et al. (2006) & Nordin et al. (2001)]

    0.32g/cc Cellular Rigid Polyurethane Foam

    Strain Rate = 0.0208s

    -1

  • 9

    Analysis of Aseptic Loosening of the Acetabular Cup

    4. Compression Test Results

    0

    1

    2

    3

    4

    5

    6

    7

    8

    9

    0% 1% 2% 3% 4% 5% 6%

    TrueStrain

    Tru

    eStres

    s(M

    Pa)

    0.2%ProofStress=7.7MPa

    Young'sModulus=345.76MPa

    UltimateStress=8.30MPa

    0.32g/cc Cellular Rigid Polyurethane Foam & Epoxy Sheets

    Strain Rate = 0.0208s

    -1

  • 10

    Analysis of Aseptic Loosening of the Acetabular Cup

    4. Compression Test Results

    Strain Rate Sensitivity

  • 11

    Analysis of Aseptic Loosening of the Acetabular Cup

    CompressionTests

    0

    2

    4

    6

    8

    10

    12

    14

    0 0.5 1 1.5 2 2.5

    Displacement(mm)

    Load

    (kN

    )

    0.01/sCellularFoam 0.001/sCellularFoam 0.01/sSolidFoam 0.001/sSolidFoam

    4. Instron Testing

  • 12

    Analysis of Aseptic Loosening of the Acetabular Cup

    5. Hip Simulator

    Stem Path

    Pneumatic Cylinder

    Gimbal Section

  • 13

    Analysis of Aseptic Loosening of the Acetabular Cup

    5. Calibration of Hip SimulatorRegulatorPressureversusInputVoltage

    [Alldatapoints]

    RegulatorPressure=1.499321178280810*InputVoltage+1.508361740837780

    R2 =0.999992111474258

    0

    1

    2

    3

    4

    5

    6

    7

    8

    9

    1 0 1 2 3 4 5

    InputVoltage(V)

    Reg

    ulatotPress

    ure(bar)

    LoadCellCalibrationInputVoltage10V

    Load=484.136144693950000*Voltage+1.182978741572300

    R2 =0.999998465370372

    0

    500

    1000

    1500

    2000

    2500

    3000

    3500

    4000

    4500

    5000

    0 1 2 3 4 5 6 7 8 9 10

    OutputVoltage(mV)

    Load

    (N)

  • 14

    Analysis of Aseptic Loosening of the Acetabular Cup

    FatiguetestsonInstron. PEcupsamplescementedintofoamblocks.

    ExperimentsoncementedacetabularcupsusingPelvisModelinHipSimulator.

    Investigatemigrationofcups. UsingCMM,measurepositionofcupsbefore&aftertesting.

    UsingCTscans,examinespecimensforfailure.

    6. Future Work

  • 15

    Analysis of Aseptic Loosening of the Acetabular Cup

    ThankyouforListeningAnyquestions?