47
An Investigation into the World-to-chipInterface World to chip Interface Sammy Datwani, Ph.D June 17, 2009 TFUG Meeting, San Jose, CA Move Liquids With Sound

An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

An Investigation into the “World-to-chip” InterfaceWorld to chip Interface

Sammy Datwani, Ph.DJune 17, 2009

TFUG Meeting, San Jose, CA

Move Liquids With Sound

Page 2: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

HPLC on-a-chip

Overview • Challenges of small scale chromatography• Advantages of microfluidics • Component development for LC on-chipp p p• Separations

Page 3: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Shrinking an HPLC

Goal: Miniaturize HPLC and maintain performance minimize system dispersion

σ2 = σcol2 + σinj

2 + σtube2 + σfittings

2 + σdet2

σext2

Loss of separation efficiency from components

• Large injection volumesd = 1 mm σext2 = 45,300 nL2

• Connection tubing• Fittings and connectors• Detector cell

extd = 0.3 mm σext

2 = 370 nL2

d = 0.1 mm σext2 = 14 nL2

Page 4: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Why Microfabricate?

Benefits of Microfabricated LC• Reduce number of system components• Reduce human intervention • Improve system-to-system reproducibility• Increase performance• Green operation• Scalability of costScalability of cost• Expand capabilities of HPLC

– Sample handling and transferPortability– Portability

• On-line reaction monitoring• Field-portable HPLC systems

Page 5: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Required Technology

• MicrofabricationMicrofabrication

• Precision Fluid Delivery

• Detection

• Integration

• Chip-to-World Interface

Page 6: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Microfabrication

• High pressure compatible structures (>5,000 psi)

• Chemical compatibility

• High temperature operation• High temperature operation

• On-chip functionality:–Valves (pL volumes)–Injectors (precise 5 nL injections)–Columns (variety of phases)–Detectors (UV, Echem, MS, …)–Connectors (high pressure, user-friendly)–Sample Preparation (filters, traps, …)

Page 7: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Chip to World Interface - Connectors

• Double-side etching in silicaC li d i l h l

Injector

– Cylindrical channels

– Separation columns

– Edge connectors

– Injectors

– Detection sensitivity

– Overall system performance

• Better pressure performance – 6000 psi routine Column Connector port

Page 8: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Connectors – Alignment

GasketFixture

Lithography-controlled alignment of multiple simultaneous connections

ChipCapillaryJig

Chip

Jig

Chemically-Etched Shim Stock

A ld k S S i C h “ i fl idi C i ” S A li i # S200 0 6321A1 CD. W. Arnold, K. R. Hencken, P. P. Leung, S. S. Datwani, D. R. Cyr , J. E. Rehm.“Microfluidic Connections” US Application: # US2005056321A1; PCT

Application: # WO2005096751A3; EP Application: # EP1740984A3; CA Application: # 2561508A1.

Page 9: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Connectors – Micro injection molded gasketmolded gasket

Material: Polypropylene for mechanical and chemical properties

Critical Dimensions: 150 um through vias on 1 mm pitchthrough vias on 1 mm pitch

Two per 6-port connection

Brass shim stock for retainer

Excellent performanceFluidic channel

capillaryJig chip

Working fluidic chip

Page 10: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Connectors – Assembly performance

Robust connection for parallel channels to outside world

Excellent performanceExcellent performance• > 7200 psi demonstrated• 50+ make/breaks for 6 connections to

4500 psiMi il i f dd d li• Microcantilever springs for added alignment

precision - excellent• Chip caddy• Micro-injection molded gaskets and j g

lithographically defined retainers

Minor issues with chipping of parts being dd daddressed.

Page 11: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Precision Fluid Delivery

Microfluidic Flow Controller (MFC)

• Active Flow Control - feedback control on

Microfluidic Flow Controller (MFC)

rapid, electronically adjusted variable pressure source • Flow meter measures the flow rate (nL/min to 10s of µLs/min)• Flow rate is maintained regardless of system back pressure or viscosity

Hi h f li bl i l fl• High performance, reliable microscale flow source• Fast, reproducible gradients

Page 12: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Microfluidic Flow Controller

402

n) 1 nL/min

398

400

Rat

e (n

L/m

i 1 nL/minstep

2 nL/minstep

0 100 200 300

396

Flow

9000 Qa(nL/min) Qa_profile(nl/min) Qb(nL/min) Qb_profile(nl/min)

/min

)

0 100 200 300

Time (sec)

3000

6000

w R

ate

(nL/

0 10 20 30 400

3000Fl

ow

0 10 20 30 40

Time (min)

Page 13: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Microfabricated Sample Injector

Concept Example - Check Valve High Pressure injector –Two check valves and a pipettorp p

Ch k l W t

Open

Check valve

Actuation

Waste

bypass

Closed

Open

Sample Inlet1mm

Mobile Phasepiston

bypassOpen

40 nL pipettor To column1mmClosed

J. E. Rehm, T. J. Shepodd and E. F. Hasselbrink, MicroTAS 2001 Proceedings, pp. 227-229

Page 14: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Injector performance

High-pressure operation (6000 psi)E ll h i l ibiliExcellent chemical compatibilityExcellent reproducibility (manufacture and operation)

0.8 % RSD peak height1 3 % RSD peak area1.3 % RSD peak area

40 nL injections; 10 mM Thiourea (in MeOH)200,000 injections; on shelf for 6 months; 200,000 injections200,000 injections; on shelf for 6 months; 200,000 injectionsDetection: 239 nm

Page 15: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Detection – UV Absorbance

Microfabricated UV Flow Cell (silica)• Fiber optic illumination/collectionFiber optic illumination/collection• 45 nL cell volume• 4 mm path length• Standard cell for ExpressLC® systems• Standard cell for ExpressLC systems• Improved versions under development

• Higher sensitivity• Lower volume 14000 Butyrophenone

8000

10000

12000

k A

rea

0

2000

4000

6000P

eak

0 50 100 150 200 250 3000

Concentration (μg/mL)

Page 16: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Columns

• Columns microfabricated with 125, 150, 200 and 300 µm diameters• Particles retained by microfabricated and/or synthesized frits• Conventional separation particles

• Various particle manufacturers, phases, sizes p p• Method flexibility• Facilitate method transfer to microfabricated system

• Easy cHiPLC™-to-conventional LC comparison y p• Temperature control facilitated by low thermal mass and

device geometry• Isocratic and gradient methods

Page 17: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Column-to-column performancece 3 columns

Abs

orba

nc 3 columns 4-5 replicate runs each25 nL injections (cHiPLC™ injector)

-2 0 2 4 6 8 10 12 14

Time (min)140000

Phenone mix

Separation conditions:200 µm x 6 2 cm column

80000

100000

120000

es/m

200 µm x 6.2 cm column 3 µm Luna C18 1 µL/min Isocratic 40 % H2O : 60 % ACN

0

20000

40000

60000

Plat

e ACNDetection: 240 nm

0 1 2 3 40

k'

Page 18: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Integrated Inj-Col-DetColumn Diameter

Flow Rate (μL/min)

Injection Volume

(nL)

Detection Volume

(nL)•150 mm – Detection volume too small•300 mm – Serpentined columns have150 μm 1.0+ 10 10

200 μm 1.8+ 20 20

300 μm 3.5+ 40 40

•300 mm Serpentined columns have too much dispersion

•200 mm – Preferred balance of integrated systemintegrated system characteristics

Pipettor

Actuation

Sample Waste

Actuation

Sample• No connections in critical analysis region• Low volume / small dimensions around turn• Gentle expansion into larger diameter region of

Check valves

Mobile Phase

ColumnColumn Fill

column• Balance dispersion and pressure drop

Critical integration region

Page 19: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Integrated Microfabricated System

160

g y

100

120

140 Compound Theoretical Plates/m

Symmetry

Factor

Uracil 75030 0.84

40

60

80

mA

U

Acetophenone 99660 0.94

Propiophenone 103580 0.94

-20

0

20 Butyrophenone 101050 0.97

Valerophenone 98260 0.99

0 2 4 6 8

Time (min)

200 µm x 10 cm column 3 µm Luna C18 1 2 µL/min

Hexanophenone 97030 1.00

Heptanophenone 96420 1.00

1.2 µL/min Isocratic - 30 % H2O : 70 % ACN20 nL injection Detection: 245 nm

Page 20: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Reproducibility of Gradient Separations

6000 Peak 30.7 % RSD peak height

4000

0.8 % RSD peak area< 0.1 % RSD retention time

Peak 7

Abso

rban

ce 0.7 % RSD peak height0.9 % RSD peak area< 0.1 % RSD retention time

2000

A A: H2O ; B: ACN%B – 90%-100%; 6 sec 200 µm x 10 cm column 3 µm Luna C18

0 2 4 6 8

0

Time (min)

1.2 µL/min 20 nL injection Detection: 245 nm

Time (min)

Page 21: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Test Sample Mix

cHiPLC™ A: 0.1%TFA/H2OB: 0.085% TFA/ACN

ce

Gradient: 0-30 min %B – 10%-90%

200 µm x 6.2 cm cHiPLC™ column3 μm Luna C1825 nL injection

Abs

orba

nc

0 2 4 6 8 10

ExpressLC®

25 nL injection2 μL/minDetection: 214 nm

A: 0.1%TFA/H2OB: 0.08% TFA/ACN

Gradient: 0-30 min %B – 10%-90%

300 µm x 15 cm Eksigent® column3 μm Luna C18

0 2 4 6 8 10 12 14 16 18 20 22 24

Time (min)

μ40 nL injection4 μL/minDetection: 210 nm

( )Theophylline; Sulfamerazine; Amphetamine; Hydrocortisone; Diazepam; Amitriptyline; Bumetanide; Thioridazine; Fenoprofen

Page 22: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Acoustic Droplet Ejection

Page 23: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

History of Acoustic Dispensing

• 1979: Recognition Equipment Inc.D t i ti li ti

• 1920’s: No focus at surface– Document printing applications

• 1985: Xerox Corp.– Document printing applications

– Uncontrolled drop size and trajectory

• 1973: Focus at surface – IBMC t ll d d i d

• 1993: NASA– Part manufacturing

• 2000: Picoliter Inc

– Controlled drop size and trajectory

• 2000: Picoliter Inc.– Life science applications

• Acoustic dispensing

• 2005: Labcyte LLC/Picoliter Inc. merger to form Labcyte Inc.– Instruments offered as well as

acoustically qualified consumablesIBM TECHNICAL DISCLOSURE BULLETIN, Sept. 1973, p. 1168

Page 24: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Non-invasive Acoustic EjectionEjection

Lens

Quartz

TransducerRF ->

AcousticEpoxy

Piezo

Coupling fluid

MHz RF Power

fluid

System

Immersion Non-invasive

(via “coupling Immersion (via coupling fluid”)

Page 25: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Drops

• Drop volume is tunable

– inverse of RF frequency

1 MHz 200 nL

100 MHz 10 pL

• Multiple drops for larger volumes

• Fluid properties have secondary influence oversecondary influence over drop volume

Page 26: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Single Ejection Event, Multi-Strobe CaptureStrobe Capture

• Water Ejection– 5 MHz

Low HighPower

gPower

Page 27: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Changing Rep. Rate

750 Hz 1180 Hz750 Hz 1180 Hz

Page 28: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Well Plate to Well Plate Transfer

Receiving well plate or substrateReceiving well plate or substrate

Fluid transferred

Source well plate

Piezoelectric transducer for the production of focused acoustic waves

Acoustic coupling fluid

•Well plates used: Greiner, Nunc, Whatman, and Corning Costar

Page 29: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Sample Transfer Results

• DMSO 100%, 95%, 90%, 85%, 80%, 75% and 70%• Various fill heights• 5 nL and 50 nL transferred with fluorescein5 nL and 50 nL transferred with fluorescein• PE EnVision fluoresence reader (Excellent Linearity)

Page 30: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Drop Ejection from a 1536 Well PlatePlate

Page 31: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Microfluidic Device Fluid TransferTransfer

Unloading processed materials:Loading diversity:

Well plate to microfluidic device

Microfluidic device to well plate

Microfluidic device to analytical instrument

Microfluidicdevice

Microfluidicdevice

• What do microfluidic device “inlets” and “outlets” look like?

Page 32: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Drop Ejection into a Pipette

Water ejection into 1550 μm hole of Pasteur pipetteWater ejection into 1550 μm hole of Pasteur pipette- Outer diameter 1840 μm

Images captured with multiple strobe flashes of a single droplet event

9.2 nL7 MHz

19.2 nL5 MHz

Images captured with multiple strobe flashes of a single droplet event.

Page 33: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Drop Ejection into a Caliper

Page 34: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Drop Ejection into a Capillary

Page 35: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Materials we can transfer

• What can we transfer?– DMSO (100% - 0%)– Aqueous solutions

• Water• Water• Salt solutions, buffers• Surfactant solutions (Triton X-100, SDS)• DNA Primers ProbesDNA, Primers, Probes• Proteins (BSA, Lysozyme), Fetal Calf Serum• Glycerol• Alcoholco o

– Polymeric solutions• Acrylamide• OthersOthers

– Ionic Liquids

Page 36: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Low Surface Tension Fluids

Ethanol 0.005% Triton X-100 Water

γ = 25 mN/m γ = 45 mN/m γ = 73 mN/m

Page 37: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Acrylamide/ Mixed Fluid Systems

Less Dense More Dense

Page 38: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Interesting Phenomena

Page 39: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Fluid Detection

• Acoustics enable “echo• Acoustics enable echo detection” of fluid

• Drop formation• Drop formation– Delivery of energy to the focal

spotspot

• Echo detectionE fl t d b k t th– Energy reflected back to the transducer

Page 40: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Acoustic Path: Attenuation

Aco stic energ is lost as the• Acoustic energy is lost as the waves travel from the source to the focusto the focus

a

Af Af Attenuation through fluid

A = Af + Ap + Ac

af

aw

apAp

A af

ApAc

Attenuation through plate Attenuation through coupler

Fluid attenuation (dB/mm for frequency F)Ac af

ac

ap

Fluid attenuation (dB/mm for frequency F)Plate attenuation (dB/mm for frequency F)Coupler attenuation (dB/mm for frequency F)

Page 41: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Acoustic Path: Reflections

• Acoustic energy is reflected gyfrom interfaces where acoustic impedance of materials do not matchmaterials do not match

Z

Ra/f R1/2 =( Z1 - Z2 )2

(Z1 + Z2 )2Zf

Zc

Zp

Rf/p

R /

Ra/fRf/pR

Reflection from air/fluid interfaceReflection from fluid/plate interface

(Z1 Z2 )

Rp/c

ZfZp

Rp/c Reflection from plate/coupler interface

Fluid acoustic impedancePlate acoustic impedance

Zc

pCoupler acoustic impedance

Page 42: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Acoustic Path: Transmission

• Acoustic energy is gytransmitted through interfaces based on degree of acoustic impedance match

ZT4 Z1 Z2

2T1/2 =Zf

Zc

Zp

Tp/c

Tf/p

Tf/pTp/c

Reflection from fluid/plate interfaceReflection from plate/coupler interface

(Z1 + Z2 )21/2

Zf

Zc

Zp

p pFluid acoustic impedancePlate acoustic impedanceCoupler acoustic impedancec Coupler acoustic impedance

Page 43: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Acoustic Path

• Drop formationp

• Detection

Page 44: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Fluid Depth Detection

• Acoustic Echo Signal Singlepoint

dit– 384 well in polypropylene plate

– 10 μL DMSO fill

audit

0.1

0.15

e [V

] Well Bottom DMSO

0 05

0

0.05

mpl

itude

-0.1

-0.05

25.00 25.50 26.00 26.50 27.00 27.50

Am

0.751 us = 0.595 mm

Time [us]

Page 45: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Meniscus Detection

• Acoustic Echo Signal Multiplepoint

dit– 384 well in polypropylene plate

– 10 μL DMSO fill

audit

Acoustic Echo Image of Meniscus Surface [Vertical Axis = Echo Delay time]

1 2 3 4 5 6 7

S1 S

4 S726.3

Page 46: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Well Plate Sonogramg

1536 Well Plate

h

o t

ime

E

ch

M43 M44 M45 M46 M47 M48M42

Page 47: An Investigation into the “World-to-chip” Interface...Drop Ejection into a Pipette Water ejection into 1550Water ejection into 1550 μm hole of Pasteur pipettem hole of Pasteur

Acknowledgements

Rick Stearns, Ph.DRich Ellson, CTO