Ninewells Hospital & Medical School Ambulatory & Central BP measurement: What is their Value in the Management of Essential Hypertension Tom MacDonald

Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

Ninewells Hospital & Medical School

Ambulatory & Central BP measurement:

What is their Value in the Management of Essential Hypertension

Tom MacDonald

Page 2: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

Mercury Banned

Page 3: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

Modern Blood Pressure Devices

Page 4: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

More Tests = Improved Precision

Test

Fre

quenc

y

Value Obtained

Page 5: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

Reproducibility of BP measurementABPM systolic

-60

-50

-40

-30

-20

-10

0

10

20

30

40

100 120 140 160 180 200

Day SBP average

Day S

BP

dif

f

CV of repeated measures = 3.3% 5

Page 6: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

SDD values of 12.0 for Clinic SBP and 7.8 for ABPM

Am J Hypertens 2002;15:101-104

For a parallel trial to detect ∆ of 10 mmHg SBP,

51 subjects using Clinic SBP v

28 using ABPM

Page 7: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

‘White Coat’ Effect

135

85

ASLEEPClinic Visit

Page 8: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

140/85

130/80

10-20%

Hypertens Res 2007;30:479-88.

Page 9: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

Cameo:White Coat Hyp

Older, female, smoker with high clinic SBP and variable

daytime readings

Am J Hypertens 2008;21:153-8

Page 10: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

Journal of Hypertension 2007;25:1554–1564

HT = Sustained Hypertension

MH = Masked Hypertension

WHT = White Coat Hypertension

NT = Normotension

Page 11: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

New NICE Guidelines

Hypertension must be diagnosed

using ABPM

Page 12: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

Redefines hypertension

Assumes ABPM every 5 years

Not cost-effective if annual

Lancet 2011;378:1219–30

Page 13: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

No outcome studies of ABPM

Page 14: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

ABPM Diagnostic Thresholds

• Optimal ABPM 115/75mmHg 24h

120/80mmHg daytime

100/65mmHg night time

• Normal ABPM 125/75mmHg 24h

130/80mmHg daytime

110/70mmHg night time

Circulation 2007;115:2145-2152

Page 15: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

NICE Rx Thresholds

• Day-time SABPM >135mmHg and CV risk > 20%

•ORDay-time SABPM >150mmHg

Page 16: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

ABPM Costs• Expensive to buy• Expensive to calibrate/service/repair• Needs Computer Software• Needs batteries• Needs staff to: program/fit/support/remove

• Needs downloaded/stored/printed• Needs interpreted & reported• 1 Machine: 3 ABPM’s / week maximum

Page 17: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

Self Monitoring at Home

Page 18: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

Patient Preference

81% preferred Home BPM to ABPM

Page 19: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

Which BP Monitor?

http://www.bhsoc.org/

Page 20: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

•Disposable patch

•Miniaturized electronics for data collection, processing, and storage

•Eliminates inflating cuff

•Continuous, non-interfering BP measurements

Cuffless Blood Pressure

Bioimpedance: See e.g. S. Grimnes and Ø. G. Martinsen, Bioimpedance and Bioelectricity Basics, AP Oxford 2008

Page 21: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

Blood Pressure-Sensing Underpants

http://www.newscientist.com/article/dn13929-invention-bloodpressuresensing-underpants.html

Page 22: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

Nichols WW, O’Rourke MF. McDonald’s Blood Flow in Arteries (5th Ed.); 2005:88.

Measuredbrachial SBP

and PP are

higher than aortic

root pressure

The difference

diminishes with age

Pressure Pulse Amplification

Page 23: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

Atcor -SphygmoCor

GTF

Static measurements

Omron HEM 9000AI

SBP2

Static measurements

HealthSTATS – BPro, CASPro

NpMA

Static and

Ambulatory measurements

Evolution of Methods to Measure Central Aortic Pressure

Oscillometric Brachial Pulse contour methods

GTF

Static/Ambulatory measurements

Page 24: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

140

70Peripheral Artery

pressure wave

Central Aortic pressure wave

MathematicalConversion

140

70

Bone

Artery

Tonometer

Brachial Blood Pressure

Non-Invasive Derivation of Central Aortic Pressure

Oscillometric BP Device

Brachial Artery wave form detected using oscillometrics

Page 25: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

American Heart Association “Clinical Science Paper of the Year”

Hypertension2013;61:1168-76.

Page 26: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

Brachial & Central Aortic SBP

115

120

125

130

135

140

0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6

Time (Years)

mm

Hg

AtenololAmlodipine

Brachial SBP

Diff Mean (AUC) = 0.7 (-0.4,1.7) mm Hg

Central SBP

Diff Mean (AUC) = 4.3 (3.3, 5.4) mm Hg

AUC

133.9133.2

121.2

125.5

P=.07

P<.0001

Williams B, et al. Circulation. 2006;113:1213-1225

Page 27: Ambulatory & Central BP measurement: What is their Value in the Management … · 2016-06-16 · Evolution of Methods to Measure Central Aortic Pressure Oscillometric Brachial Pulse

Conclusions• ABPM is here to stay

Home BP widely used

• Central aortic BP might valuable for defining hypertension in the young (especially males)

• Both ABPM and Central BP need further research