  • Installation Guide

    Single-Channel Analog Input Board on 6890 GC

    Accessory G1556A

  • Agilent Technologies, Inc.2850 Centerville RoadWilmington, DE 19808-1610USA

    © Agilent Technologies 2001

    All Rights Reserved. Reproduction, adaptation, or translation without permission is prohibited, except as allowed under the copyright laws.

    Part number G1556-90307

    First Edition, APR 2001

    Replaces Part No. G1556-90300, Installation Guide, Single-Channel Analog Input Board.

    Printed in USA

    Safety Information

    The Agilent Technologies 6890 Gas Chromatograph meets the following IEC (International Electrotechnical Commission) classifications: Safety Class 1, Transient Overvoltage Category II, and Pollution Degree 2.

    This unit has been designed and tested in accordance with recognized safety standards and designed for use indoors. If the instrument is used in a manner not specified by the manufacturer, the protection provided by the instrument may be impaired. Whenever the safety protection of the Agilent 6890 has been compromised, disconnect the unit from all power sources and secure the unit against unintended operation.

    Refer servicing to qualified service personnel. Substituting parts or performing any unauthorized modification to the instrument may result in a safety hazard. Disconnect the AC power cord before removing covers. The customer should not attempt to replace the battery or fuses in this instrument. The battery contained in this instrument is recyclable.

    Safety Symbols

    Warnings in the manual or on the instrument must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions violates safety standards of design and the intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.

    WARNINGA warning calls attention to a condition or possible situation that could cause injury to the user.

    CAUTIONA caution calls attention to a condition or possible situation that could damage or destroy the product or the user’s work.

    Indicates a hot surface

    ✧ Indicates earth (ground) terminalSound Emission Certification for Federal Republic of Germany

    Sound pressure Lp < 68 dB(A)

    During normal operationAt the operator positionAccording to ISO 7779 (Type Test)

    Schallemission

    Schalldruckpegel LP < 68 dB(A)Am ArbeitsplatzNormaler BetriebNach DIN 45635 T. 19 (Typprüfung)

  • Overview

    OverviewThis section reviews the procedure for installing a single-channel analog input board on an Agilent 6890 Gas Chromatograph (hereafter called the GC.) With the analog board installed, you can connect non-Agilent detectors with 1 V of analog output to the GC. The voltage is then digitized and can be sent to a workstation or INET integrator.

    Before following this procedure, refer to the safety information on the inside front cover.

    Specifications

    Parts List• 1 single-channel analog input board

    • 1 general-purpose analog cable

    Required Tools• Electrostatic protection such as grounded wrist strap (part no.

    9300-0969 for large wrists, part no. 9300-0970 for small wrists)

    • T-20 Torx screwdriver

    Item Description

    Input voltage -10 mV to + 1.05V (between + and - inputs)

    Input noise 5 mV peak to peak, 0.01 to 2.5 Hz frequency

    Input common mode range ±5 V DC (measure + or - to ground)

    Common mode rejection ratio 80 db

    Scaling 1 display count = 15 µV at the input

    1

  • Overview

    Steps1. Preparing the GC

    2. Positioning and securing the analog input board

    3. Connecting the cable

    4. Restoring the GC to operating condition

    Analog input board

    Cable

    2

  • Preparing the GC

    Preparing the GC

    WARNING Hazardous voltages are present in the mainframe when the GC power cord is plugged in. Avoid a potentially dangerous shock hazard by unplugging the power cord before removing the side panels.

    1. Turn off the GC and unplug the power cord.

    2. Remove the electronics side cover. Loosen the two screws with a T-20 Torx screwdriver, slide the cover to the right, and lift it off.

    3. Remove the electronics top cover by disengaging the clips underneath the cover and lifting it up.

    Press clips under cover

    Electronics top cover

    Electronics side cover

    Power switch

    Screws

    3

  • Positioning and securing the AIB

    Positioning and securing the AIB

    Caution Board components can be damaged by static electricity; use a properly grounded static control wrist strap when handling the board.

    1. Remove the board from its static control bag and slide into the front or back detector slot on the main board until it is plugged in. Tighten the screw on the AI board bracket with a T-20 Torx screwdriver.

    2. Make certain switches 1 and 2 are in the On position:

    Back detector slot

    Front detector slot

    Screw

    Analog input board

    Switches 0 and 3are in Off position;switches 1 and 2are in On position

    On Off

    4

  • Positioning and securing the AIB

    3. Attach the signal cable to the board.

    4. Position the cable so that it runs down and out through the first slot to the left of the cryogenic tube outlet opening on the rear panel of the GC. Tie a knot on the cable where it exits for strain relief.

    Cryogenic outlet

    5

  • Connecting the cable

    Connecting the cableConnect the cable to your non-Agilent detector. The following are the cable pinouts. The red, brown, and blue leads are not used and can be cut off, if desired.

    Connector 1 Signal Name Connector 2Quick Disconnects

    1 No connection Brown

    2 1 V (-) White

    3 No connection Red

    4 1 V (+) Black

    6 No connection Blue

    Shield Ground Orange

    G1530-60560

    1

    2

    1 2

    3 4

    5 6

    6

  • Restoring the GC to operating condition

    Restoring the GC to operating condition1. Replace the electronics side panel

    2. Replace the electronics top cover.

    3. Plug in the GC and turn it on.

    4. Press [Front Det] or [Back Det]. Observe the the Output line of the display. If the lines are connected or shorted, and a 0 Volt input is being

    supplied, the Output should be between -25 and + 25.

    BACK DET (AIB)Temp not installed <Output 15

    7

  • Restoring the GC to operating condition

    8

  • Printed on recycled paper.

    This product is recyclable.

    Agilent Technologies, Inc.

    Printed in USA APR 2001

    G1556-90307

    OverviewSpecificationsParts ListRequired ToolsSteps